I
I <k
NBSIR 75-759 (R)
Standard Reference Materials and
Data For Raman Spectroscopy,
Electron Paramagnetic Resonance,
and Magnetic Moment Measurements
H. S. Bennett, G. A. Candela, T. Chang, R. E, Mundy, and G, J. Rosasco
National Bureau of Standards
institute for Materials Research
Solid State Materials Section
Washington, D.C. 20234
July 1975
u s. DEPARTMENT OF COMMERCE
NATIONAL BUREAU OF STANDARDS
1
NBSIR 75-759
STANDARD REFERENCE MATERIALS AND
DATA FOR RAMAN SPECTROSCOPY,
ELECTRON PARAMAGNETIC RESONANCE,
AND MAGNETIC MOMENT MEASUREMENTS
H. S. Bennett, G. A. Candela, T. Chang, R. E. Mundy, and G. J. Rosasco
National Bureau of Standards
Institute for Materials Research
Solid State Materials Section
Washington, D.C. 20234
July 1975
U.S. DEPARTMENT OF COMMERCE, Rogers C.B. Morton, Secretary
James A. Baker, III, Under Secretary
Dr. Betsy Ancker-Johnson. Assistant Secretary for Science and Technoiogy
NATIONAL BUREAU OF STANDARDS, Ernest Ambler, Acting Director
STANDARD REFERENCE MATERIALS AND DATA FOR RAMAN SPECTROSCOPY,
ELECTRON PARAMAGNETIC RESONANCE, AND MAGNETIC MOMENT MEASUREMENTS
ABSTRACT
This publication contains the analyses of three survey-questionnaires to assess the needs and markets for standards in
Raman spectroscopy, electron paramagnetic resonance, and magneticmoment measurements. These analyses give the requirements andsuggest the strategies to be followed for making each of the abovethree measurement methods more quantitative than they are atpresent. Some of the requirements are data cataloging, standardsfor data presentation, standard reference materials, and standardreference data.
Keywords: Electron paramagnetic resonance; magnetic moment; Ramanspectroscopy; standard reference data; standard reference materials.
Table of Contents
1.0 Introduction 1
2.0 Summary of Findings and Highlights 1
2.1 Raman Spectroscopy 1
2.2 Electron Paramagnetic Resonance 2
2.3 Magnetic "oment Measurements 2
3.0 Survey Report on Standards for Raman Spectroscopy . 3
3 . 1 Summary 3
3.2 Domestic and Foreign Responders 6
3.3 Fields and Areas of Application 7
3.4 Data Base Problems 11
3.5 Standard Reference Materials 11
3.6 Reference Data 12
3.7 Formats for Raman Data Presentations 13
3.8 Problem Definition and Research Strategy 13
4.0 Survey Report on Standards for Electron Paramagnetic Resonance .... 16
4.1 Introduction 16
4.2 Findings and Discussion 18
4.3 Additional Comments 25
5.0 Survey Report on Standards for Magnetic Moment Measurements 25
5.1 Introduction 25
5.2 Data Base and Findings 28
5.3 Possible Users 33
5.4 Summary 33
5.5 Technical References 34
6.0 Acknowledgments. . 34
1.0 INTRODUCTION
The Solid State Materials Section at the National Bureau of Standards conducts researchon solid state materials with emphasis on their technologically important behavior whichresults from their optical, electrical, and magnetic properties. It interprets the mechan-isms for such properties in terms of physical, structural, and chemical factors anddevelops measurement methods, data, and standards required to support selected technologies.
To continue our efforts on the provision of standard reference materials and data,we have used three survey-questionnaires to assess the needs and markets for standardsin Raman spectroscopy, electron paramagnetic resonance (EPR) , and magnetic moment measure-ments. In addition, we have used these surveys to determine the requirements for makingsuch measurements more quantitative than they are at present. The detailed analysesof these three surveys are given, respectively, in parts 3.0, 4.0, and 5.0 of thispublication
.
Our general approach for these three surveys was to obtain mailing lists of workerswho had a reasonable probability of using one or more of the above three measurementtechniques. After the mailings were completed, we analyzed those questionnaires whichwere voluntarily returned to us. Because of limitations on the resources availableto us, we did not seek responses from those who had not returned the questionnaires.Hence, our analyses were based upon volunteer responders and as such represented a
lower bound on the size and needs of the potential beneficiaries of our standardsservices
.
2.0 SUMMARY OF FINDINGS AND HIGHLIGHTS
We mention briefly in the next several paragraphs some highlights from these threesurveys
.
2.1 RAMAN SPECTROSCOPY
The magnitude of the response was 111 out of 285, or 39%. Some of our findings
are listed below:
(1) Many respondents encounter problems because the Raman frequencyshifts, relative band intensities, and depolarization ratios are not reportedaccurately.
(2) A definite interest in data compiling efforts exists among the
respondents
.
(3) Raman standard reference materials elicit strong, positive comments. Our
analysis reveals two problems: namely, how do we measure the standardspectrum of a standard reference material? and, how do we transfer the neces-
sary measurement technology to the user of the SRM? Standard referencematerials will make Raman spectroscopy a more quantitative measurement sciencethan it is at present.
(4) The setting of standards for data presentation is essential for the success
of any SRM effort.
- 1 -
This survey is a first step for the National Bureau of Standards in its efforts tohelp promote standardization in the important measurement science of Raman spectroscopy.
Please direct inquiries on standards in Raman spectroscopy to Gregory J. Rosasco(telephone: 301-921-2780).
2.2 ELECTRON PARAMAGNETIC RESONANCE
Our EPR survey assesses the need for a standard reference material (SRM) by whichto improve intensity measurements in electron paramagnetic resonance (EPR) . As in Ramanspectroscopy, such an EPR SRM will make EPR a more quantitative technique. The number of
answered questionnaires is 175 out of 503 or 35%.
From our analysis, about one quarter of the EPR users is in physics, one quarteris in biomedicine, and the remaining half is in chemistry. By references to journals in
which researchers in physics and in biomedicine publish we discern a trend for the numberof EPR users in physics to decrease and the number of EPR users in biomedicine to increase.We cite here highlights of the survey:
(1) The EPR respondents need and want an intensity EPR SRM (93% say yes)
.
Everyone who has attempted quantitative EPR measurements stated that the
lack of standards presents great handicaps and difficulties.
(2) A standard reference material which results in EPR line intensity measurementsaccurate and precise to within 10% will satisfy the present needs of over halfof the respondents.
(3) The purchasers of EPR SRM's will require technical help and advice, andparticularly, detailed and extensive instructions on how to use the EPR SRM's;i.e., a challenge exists to transfer our technology in EPR to the EPR usercommunity.
Please direct inquiries on standards for electron paramagnetic resonance toTe-Tse Chang (telephone: 301-921-2820).
2.3 MAGNETIC MOMENT MEASUREMENTS
The number of answered questionnaires is 297 out of 781 or 38%. The magneticmoment survey reveals that 87% of the 124 workers who returned questionnaires, and whostated that they need a magnetic moment SRM, already use some type of reference materials.Nickel is the most frequently used calibrant by this subset of workers. None of theworkers who used a reference material state the accuracy of their own calibrant.
The value of the saturation magnetization for nickel given in the literatvire
varies by 8% among reliable sources. Workers do not know which value is the correct oneto use.
And finally, this survey shows that if NBS were to issue a magnetic moment SRM,then it should have the following properties:
(a) an accioracy within ±1%,(b) more than one size (perhaps a 1-mm and a 3-mm sphere)
,
(c) nickel will most likely be one of the calibrants,(d) the price should be less than $150.00.
Please direct inquiries on standards for magnetic moment measurements to George A.Candela (telephone: 301-921-2780).
- 2 -
1
3.0 SURVEY REPORT ON STANDARDS FOR RAMAN SPECTROSCOPY
3.1 SUMMARY
The survey "Standards for Raman Spectroscopy" was distributed in January 1973,
The numerical results, the analysis of these results, and conclusions derived fromthe survey are presented here.
To summarize the results, the five major subject areas are presented in a
graphical format below. Each subject area is briefly described in a paragraph followingthe graph. In the graphs, the n\mber of respondents or the number of responses of a
given type are plotted by the horizontal bars. Percentages (indicated by %) of relevantresponse totals are often provided; however, only numbers of respondents or responsesare plotted. Categories in which a single respondent could give more than one replyare indicated by an asterisk (*)
.
1. Kcspoasa
O.SJl.
M
OriJir.al .".ail jr.;
100. I
120 140I
160I
180I
200
Rgspcns 40%
CisAda. Europe
,
Seipoases 35 i 33 %
Iknu of Scien-
tific tsdeavor
al«C3l«r Physics
tad CTieaistry of
all phases of
utter
AsalyticalChratstry
riiTSical
Oacsistxy
aoUd fiMttfbfsics
OrfaaleOaaistzy
locheoistry/Blo^ysies
Zaer^anicOcmittry
Iislyeer Physics
•ad Oieaistry
jBatnsent Kanu-
tactuze and 9ata
eatalo<3inq
The magnitude of the response is 39%. The breakdovm of the areas of scientific endeavour
is based on the descriptive titles most used by the respondents.
- 3 -
3. to you tncountorprobltms In con-pArlion of spec-tra froB other•ourees?
*tvp«» of Problemi
Kelatlve/AbsoluteIntensity
frequency Shift/Linewldth
rolarlzatlnn Char-acter of a Band
Inoonplate Specif,
of Relevant Exp.VaiaDatari
J! I10 40 SO •0 70 •0 10 100
Beplleii 90»r«plled qoo/lll)
Of the s e 100 who repliedTea 84 9,
There are significant problems encountered in the comparison of Raman data from "othersources." Interestingly, a large number of respondents have encountered problems becausethe Raman frequency shifts are not accurately reported. Since accurate frequency calibra-tion is generally a routine matter, this observation suggests that improvements in the
data base can be achieved by spectroscopists being a little more conscientious.
1. Oata Catalog*
Xitportant toyour work
TaaoBewhato
Z«portant to h^tmSpaetroteopistsin Cemral
y«alooawhatMo
Kaaa catalogs youknow
RoneOnaTwoThree or
Bora
Should •efforts" in
this area b*•spanded
y«t
10
J_30 40 50
B3* replied (9:/lll)
Of these 92
J21_"I
76 gjreplied (87/111)
Of these 87
76%
0%
83 ^replied (92/111)
7S% replied (83/111)
Of these 83
95^"
5%
BO
_I_100
There is definite interest in data cataloging efforts. In replying to the question ofincreased "efforts" in this area, many respondents cited "efforts" such as "systematization"and "standardization." There were warnings to avoid a repetition of the history of IRdata cataloging ventures. The nature and importance of this problem are generally under-stood. These concerns have been communicated to a variety of groups whose activitiesinvolve "reference data."
.- 4 -
4. Keeoisner.datlonifor Presentation of«*nan Spectra forCataloging a.-.d Doc-cscntatlon inFarranent DataCollections
Additionaliu73esti':ns 7
ao>e«
Il<jcb«r of typesof suggestions
Specific eca-B«nt of coa-patUble Ranan-tH foraat
YasBo
10 20 30 40 SO 60 70 60 90 100
5B% replied [65/111)
Of these 65
5:%
This subject area was included in the questionnaire in order to assess the reaction tothe "Recommendations" and, further, to assess the need for simultaneous presentationof Raman and infrared data in data collections. The number of positive replies (34 of111) is not large enough to allow a resolution of these questions.
S. standard Reference~ Materials
Important/Useful toyour work?
teaSecevhatHo
Important/Useful to
Jtasan SpectroscopistsIn General
TesSooe-what
Ho
toggast Materials forOse as SKM^General Categories:"Coraon" SolventsAqueous Solutions•Polyoers"Other Orgvdc ''.iquids
and SolidsInorganic Solids a.-id
tlquldsGasas
*Sp<elfic Substances:
CCiuCtHgC9B7
"2•AIR"
Single crystal a-5i02rW|
,
02
CjoHa
Single crystal XF2
X Ca,Ba,MnCOj
Internal Standard:Fluorescent or ResonantBaaan Scattering
10 50 £0 90 100
83% replied (92/111)
Of these 92
74%
9^
8:% replied (91/111)
Of these 91
8-! %
2 %
77 Responses
94 rifferer.t Materials
Igualified 'Yes" .iS %
*Rainan Newsletter, No. 29, 1971. - 5 -
The area of standard reference materials elicited strong, positive comment over-all. A very important consideration which is not easily quantified "emerged" from theanalysis of the survey. This consideration involves two problems! (1) How do we measurethe standard spectrum of a standard reference material, and (2) how do we transfer the
necessary measurement technology to the user of the SRM? The definition and/or developmentof this technology involves (1) accurate laser power measurement (at the sample) , (2)
selection of reproducible scattering geometry and laser focusing parameters, (3) accuratedetermination of collection and transfer efficiency of the fore optical system, (4) measure-ment of the spectrometer transfer efficiency, (5) measurement of the absolute detectorresponse function, (6) measurement of the polarization character of the incident andscattered light, and (7) measurement and deconvolution of the- spectral effects of theinstrument function.
The use of standard reference materials will, hopefully, allow laboratories to measurea spectrum truly representative of the material without the need of performing all the
calibrations listed above. The development and implementation of simple, accurate tech-niques for calibrations such as the above items are first priorities for establishinga useful standard reference material.
The information gained in this survey has been communicated to a number of organiza-tions involved in standards for Raman spectroscopy. These activities are detailed inSection 3.8. Positive steps pursuant to the promulgation of useful standards fordata presentation are being taken in cooperation with the Joint Committee on Atomicand Molecular Physical Data. Developments relevant to the measurement of the absolutedetector response function (item 5 above) have been communicated to Raman spectroscopists.Progress in this area will follow the level of the developments described above. Around-robin test designed to establish the magnitude of the errors encountered in datacomparisons is also recommended.
This survey is a first step for the National Bureau of Standards in its effortsto help promote standardization in Raman spectroscopy* It is importantto continue interacting with the scientific community in the area of standardizationfor Raman spectroscopy.
3.2 DOMESTIC AND FOREIGN RESPONDERS
The questionnaire was sent to 285 people, 192 of these in the U.S.A., and 93
in Canada, Europe, and Japan. The list for mailing was assembled from:
1. Mailing list of the RAMAN NEWSLETTER obtained from the Optical Society ofAmerica,
2. "Users" lists from Raman instrviment manufacturers:
Spex Industries, Inc.
Jarrel-Ash Division, Fisher ScientificGary Instruments Division, Varian AssociatesBeckman Instruments
3. Lists of participants: The International Conference on Light Scatteringin Solids, 1968 (New York) ; 1971 (Paris).
The U.S. mailing can be broken down into two categories:
A. national laboratoriescommercial/industrial research laboratoriesinstrument manufacturersgovernment agencies and laboratories
B. universities.
- 6 -
57% (110/192) of the surveys were sent to those in category A. We note also that there
are 69 organizations represented in category A.
The total number of responders was 111 or 39%; 4C% (76/192) of the U.S, mailingand 38% (35/93) of the rem.ainder. Various estim^ates by the Raman Technical Group of
the Optical Society of America concluded that there are from 50C to 800 Ranan facilitiesin the world and from 250 to 400 Raman facilities in the U.S.A. Her.ce we have sampledroughly one-c'uarter to one-half of the domestic Raman facilities.
Under category A (governr.ent, industry, etc.) of the U.S. mailing, 37% (41/110)
responded, whereas 43% (35/82) of those in the -universities responded. The responseby number of organization in category A was 48% (33/69) . There was a significant responsefrom those involved in the comir.erciai design and manufacture of R.aran instrum.entation
.
The magnitude of the response (39%) is quite good, but even more important was the detailand quality of the completed questionnaires. The survey covered a broad range of topics.Being somewhat "free-form," it also attempted to elicit a great deal of individual response.The efforts of the responders to relate valuable information in many areas of theirexpertise and their eagerness in many cases to assist in standards efforts are veryencouraging indications to the significance of a "standards" effort.
V7e enclose a copy of the questionnaire as figure 1 for the convenience of the
reader
.
3.3 FIELDS AND AREAS OF APPLICATION
Questions II. 1.A and II.l.C assess the areas of scientific interestresponders. These areas are best described by the set of descriptive titlesnamed by the responders. (In many cases, one person mentioned more than one
The breakdown is as follows:
Molecular physics and chemistry of gases.liquids and solids 31
Analytical chemistry 29
Physical chemistry 29
Solid state physics 25
Organic chemistry 15
Biochemistry/biophysics 14
Inorganic chemistry 12
Polymer physics and chemistry 10
InstruTient manufacture/data cataloging 8
The scientists involved in these areas of research utilize data from, a greatvariety of experimental disciplines (Question II.l.B). Infrared spectra and structuraldata (x-ray, neutron, and electron diffraction) naturally predominate among those men-tioned. NI-'J^ spectra are also widely er.ployed.
The wide variety of physical, chem.ical and biological disciplines enccunteredin the responses to this question suggests:
1. The survey results should reflect the needs and interests of a great manyscientists
.
2 . The inter-disciplinary nature of light scattering research and its wideapplication to problems in diverse areas.
3. The importance of a useful and m.eaningful data base for Raman spectroscopy.
of themost regularlyarea .
)
- 7 -
0MB No. 41-S72080Figure 1. approval expires JUNE 30, 1973
NBS-732 U. 8. DEPARTMENT OF COMMERCE(1.73) NATIONAL BUREAU OF STANOAnOS
STANDARDS FOR RAMAN SPECTROSCOPY QUESTIONNAIRE
PART I Please complete identification material below
Organization/Affi liationName (Last, first. Middle!
Address tStroel. Cilv. Slate, ZIPI
PART II Please answer the following questions
1 A. To what areas of science and technology do you apply the techniques of Raman spectroscopy?
B. What other types of data do you utilize in conjunction with Raman spectroscopy?
C. What type of problem do you attempt to solve by means of Ramrn spectroscopy?
2. Do you encounter problems in comparison of spectra from other sources that you would trace to lack of proper standardization
procedures? (e.g., inaccuracies in frequency calibration, depolarization ratios and relative intensities.)
3. Standard Reference Materials. These materials could be used in Raman instruments for frequency and intensity calibration,and possibly will allow determination of instrument functions.
A. How would you assess the importance of standard reference materials?
(1 ) to your specific efforts?
(2) for the users of Raman spectroscopy?
B. What type of standard materials would be most useful? Please suggest specific candidates (useful to you or useful in leral)
in each category.
(1 ) Gas
(2) Liquid
(3) Solid
- 8 -CONTINMD
Figure 1 (continued)C. Woj'd you prefef an internal (trace additivel standard (e.g., a species which shows sharp ime fluorescence or resonar.;
Raman scanenngi'
Arty Suggestions''
D. What types of sampling geometries and cells would you recommend' (Please be as specific as possible.)
E. In what form would the companion Standard spectrum be most useful to you: graphical/digital/tabular?
4. Catalog of reference Raman spectra.
A. How would you assess the impo<rtance of a catalog of Randan spectra?
(1 1 to your specific efforts'
(2) to the efforts of scientists as a whole?
(3) what Raman spectral catalogs are you pfesently aware of?
B. Should efforts in this area be expanded?
C. What types of materials would be important' Please specify some important rraterials in each category.
(1 ) Solvents (laboratory and/or commercial)?
(2) Substrate, window and cell materials?
(3) Conmon gases (pollutants)?
(4) Important solids'
5. lUPAC has approved a set of "Recom.T>endat,ons for Presentation of Rarnan Spectra for Cataloging and Documentation ,n
Permanent Data Collections" (Raman Newslener, No. 29, .Way 1971),
A. Do you have additional suggestions for the format in which the data is presented, such as presentation of Raman and infra-
red data simultaneously?
B. Do you have a preference for graphical, tabular, or digital presentation of standard spectra?
- 9 -CONTINUED
Fiyure 1 (cont:iniiPri^
Du you have siiggostioiis coiiccniiny tho need to ycnerate Ramon standard referon(;e materials which can be used as
standards in other fields of spectroscopy, chemical analysis, and physical measurements' IPIoJSC be siiccilic)
Does this need for compatibility affect all types of Raman Standards?
7A. What specific problems must be overcome in order to
|1 ) define a standard spectrum?
(2) make it practically useful?
B. What suggestions can you offer to meet these difficulties?
8. In what capacity can the National Bureau of Standards be most useful to spectroscopists in an effort to establish standards
for Raman spectra?
PART III We would appreciate any comments you may wish to add
- 10 - USCOHM-NBS-OC
3.4 DATA BASE PROBLEMS
Question II. 2 examines the magnitude of the problems encountered in comparisonof Raman data from various sources. One hundred (90%) of the responders replied tothis question and of these only 16% experienced "no" problems. In some of these cases,there were no problems because very little literature data was available with whichto compare their data. The 84% who experienced problems cited a variety of types ofinaccuracies. The following lists the problems and the number of times cited (more
than one area was typically mentioned in a response)
:
Accuracy (relative and absolute) ofmeasurement of intensity of scattered 64
light
Accuracy of frequency (shift) measurements(and in some cases also linewidth 58measurements)
Accuracy of determination of polarizationproperties 51
Incomplete specification of relevantexperimental parameters and sample 18
characteristics
The responses to this question indicate that serious problems are encounteredin the measurement and reporting of all the vital parameters of Raman spectrum. Someof these inaccuracies are primarily rooted in the difficult nature of certain measurementsand calibration, e.g., intensity measurements and linewidth deconvolutions . It is somewhatsurprising that frequency inaccuracies are encountered so often. This may stem fromthe fact that in many experiments accurate frequency values are not particularly important.An important inference is that the problem areas cited by the responders may in manycases be remedied by selecting standard techniques and reference materials.
3.5 STANDARD REFERENCE MATERIALS
Question II. 3 addresses the use of Standard Reference Materials, described in
the questionnaire by: "These materials could be used in Raman instruments for frequencyand intensity calibration, and possibly will allow determination of instrument functions."The question of the importance of SRM's to "your specific efforts" (Question II.3.A.1)
and "for the users of Raman specrroscopy" (Question II. 3. A. 2) was answered asfollows
:
To Question II.3.A.1, 83% (92/111) replied, and of these 74% (68/92) thoughtthat an SRM would be important or very important to their work, 17% (16/92) saw someimportance, and 9% (8/92) said not important.
To Question II. 3.A. 2, 82% (91/111) replied, and of these 84% (76/91) thoughtthat an SRM would be important or very important to users in general, 14% (23/91) sawsome importance, and 2% (2/91) said not important.
Many responders used the word "useful" and the word "important" should be under-stood in this context.
Question II. 3. B asks for specific candidates for reference materials. A numberof responders emphasized that frequency calibration is "best" accomplished by meansof "known" atomic standards, i.e., Ne, Ar, Kr lamps, or possibly the non-lasing emissionlines from the laser plasma. The important characteristics (spectral properties and
- 11 -
material properties) of a standard reference material were outlined by others (20-30responders) . Such characteristics were also mentioned in response to Questions II.7.A&B.For example, the Raman spectrum of the material should exhibit strong, sharp lines,cover a wide frequency range, and be insensitive to temperature and excitation wavelength.The material itself should be non-toxic, non-hygroscopic, easy to purify and easy to
measure
.
Seventy-seven (77) of the responses included some specific recommendationsfor a reference material. In the remaining portion of this paragraph the numbers in
parentheses represent the number of times a material was cited. Eighty-four differentsubstances were named in the categories of common solvents (12) , aqueous solutions (10)
,
polymers (5), "other" organic liquids and solids (12), inorganic liquids and solids(30), and gases (15). Some frequently mentioned compounds were: CCli^ (28), C5H5 (20),
C9H7 (Indene) (16), N2 (15), "Air" (8), single-crystal a-Si02 (8), CHi+ (7), O2 (7),
CeHi2 (7), CioHe (Napthalene) (7), single-crystal XF2 (X = Ca, Mn, Da) (6), and CO2 (6).
In many cases the "popular" materials do not display some of the important characteristicsof a "reference" material. A set of "best" candidates has not emerged from the evaluationof the survey.
Question II. 6 addresses the need for compatibility of reference materials amongthe various physical and chemical disciplines. Only 20% (22/111) of the responderscommented on the need for compatibility. From these answers there was no strong indicationof the necessity to design strictly compatible standards although such efforts "mightbe helpful."
Question II.S.CfiD explored two further areas with regard to reference materials.The possibility of using a "trace" additive internal standard was the subject of QuestionII. 3. C. The trace additive standard, which was suggested in the question, might displayeither sharp-line fluorescence or resonant-Raman scattering. There were several objectionsto the use of fluorescent or resonant standards; namely, the strong wavelength dependence,unknown solvent effects, and some basic uncertainties with regard to the nature of the
effect in the case of resonant scattering. Objections were also raised because of possiblechemical interferences. Over-all, then, 52% of the 83 responses were against the ideaof an "internal" standard. The remaining 48% gave a qualified yes.
Part D of Question II. 3 addressed the problems of sample cells and scatteringgeometry. There were 69 responses to this question. 90° geometry was suggested 50
times vs. 13 suggestions of 180" scattering. Capillary tubes (melting point type) weresuggested 23 times as sample cells for liquids. Cylindrical cells illuminated alongtheir axes (8 responses) and rectangular cells (7 responses) were mentioned among the
other "popular" cells. Single pass was explicitly mentioned ten times and multi-passin five responses.
3.6 REFERENCE DATA
In Question II. 4 the survey attempted to assess the present level of interestin data cataloging efforts in the field of Raman spectroscopy. The results of theseinquiries are as follows:
Question II.4.A.1 Are data catalogs important to your specific efforts?
83% (92/111) indicated that the catalogs were important or very important;17% (16/92) said "some importance"; and 23% (21/92) indicated "not
important .
"
Question II. 4, A. 2 Are the data catalogs important to users of Raman spectroscopyin general?
- 12 -
78% (67/111) conzer.red; 78% (cS/37) of rhcse who zo—.er.zed said ir.pcrtar.t
or very inportant; ar.d 22% (19/87; said "sorse ir^crtar.ce.
"
This question — I". 4. A. 3 — also asked -r.e respor.der zo r.ar.e z'r.e caralcgs cf
Ranar. spectra of which he was aware. 83% 02/111; cc—.er.ted. The resul-s v/ere as follows:
25%, none; 3"%, one; 21%, frfo; 17%, three or -ere. Zr. z'r.Ls quesricn zr.e respondersmentioned the veri-- lir^ced editions of certain instr-— er.- rar.uf ac^ jrers
.
The svibject of Question II. 4. 3 was need for further efforts in this ar&a.73% (83/111) contoented; and of these 95% (79/S3) felt that these "efforts" should beexpanded. However, the nature of these "efforts" was, in nany cases, suggestions to"organize," "syster^tize ,
" and "evaluate" the speczral catalogs. A significar.i r.u.-her cfcoments concerned z'r.e need to avoid proliferaiion , duplication, and incc-pa -:Jcili-y of
the cataloging efforts in^Ranan spectroscopy.
Question II. 4. C. asked for specific candidate materials for spectral cataloging.The response to trJ.s question was not as large as for the siriilar q'jes-ion wLzs. regard to
SSM's (II. 3. 3). There was a large '/ariety in the responses -.".zs. no trends.
3.7 FCR:-L-.rs TZ?. ?j-2c-.:: ^ata ppzsenta.ioks
Question II. 5 ccncernsd the recently p'jhlished 'IVPAC approved) "RecoEssenda-ions
for Presentation of ?^-an Spectra for Cataloging and Ioc'^.entaticn in r£rr.anent -ataCollections." We asksd for additional suggestions for the fcrrat of the data and, in
particular, q'aestioned if the ?.ar=.n and infrared fort::at3 should he identical and if the
"Permanent Data Collections" should present both -he Ranan and infrared spectra of a
coi^pc^jnd. 53% ;£5/lll) restcnded to the question and of these 48% (31/65) had noadditional suggestions. Fifteen Z'/zes of suggestions were -ade; the most significant cf
these was: 17 people wanted co-patihle I?.-?.ar.an forra.ts whereas 5 were against thisposition (or at least wanted sor.e :h=r.r5E ir. the cm"- region z people wanted the
spectra presented after correction ftr ir.str _-er.t5.1 effects. Cur irtression is that r.any
of those who ref.i2med the ~^estionnaire vrere not aware either of the "P-etc—.endations"
or of their intent.
Z'r.e 00— ents associated vrith Question II. £ are included in the disc-ussicn ofQuestion II . 3
.
3.8 PSDBLE2'i czji^.:iric:; a:- pzsza-P.c:-: siraizsy
Question 11." is extreoely vital to the evaluation cf any efforts to providea standard reference spectr-JE of a standard reference material . Z'r.e conr-jents on the natureof the material chosen as a standard have been included -jinder the discussion of Question11.3. The important conclusion to be dravm from the ans-rfers to Question II. 7 and th^cughourelevant sections of the q-jesticrj-.aire is that there are f-undar-ental probler.s involved in
the measurement of a "standard" spe:tr-^-. , i.e., a spectr-jr, tr'ily representative only cf the
material, not of the instr-~.ent used tc r.ai^e the -eas-urer.ents . Tnis latter req-uirer.ent is
essential to the definition of a standard reference material. rh_i= problem area v,-ill bedealt with in the s-m—ary of options discussed below.
Question II. S asked "In what capacity can the National Bureau of Standards bemost useful to spectroscopists in an effort to establish standards for Raman spectra?"The responses to this question and many coments from other parts of the que3tior.naire
are relevant tc the choice cf directions to be p-ursued by the ::atitn=l Bureau of Standardsand others concerned with standards efforts. These responses oar. 'ze = z be discussed -under
two headings: I. Assuring efficient production cf (oc—.eroial) spectral catalogs, and
II. Proceeding toward the generation of standard reference spectra.
Under heading I, there were a wide variety of recommendations. Some ofthese have been discussed in question II. 4. B. It is suggested that a classification ofRaman spectra along the same lines as the Coblentz designations IR - I, II, III wouldbe appropriate for spectral catalogs and standard reference spectra. The production anddistribution of commercial spectral catalogs is seen as important by the responders.However, these catalogs must be extensive to be useful and at the same time inexpensiveto produce if they are to be practical for both the producer and purchaser. Thus, theywould generally fall into a category similar to Coblentz IR-III. The responders weregenerally in agreement that, within the limits of this class, efforts should be directedtoward assuring the maximum, practical quality of the data (sample quality and spectralquality) . Relevant spectral parameters must be specified (see discussion below) , andthe format of presentation should be uniform both within a data compilation and for
different compilations. The specification of relevant spectral parameters and questionsof format for data presentation have been addressed in the "Recommendations" of the Ramansubcommittee of the Joint Commission on Atomic and Molecular Physical Data (NSRDS) andapproved by lUPAC. At least a minimal subset of these recommendations should be adoptedas a uniform requirement for Class III collections. As mentioned above, there is a needto "organize," "systematize," and "standardize" data catalogs. The Bureau is seen bythe responders as having varying degrees of responsibility towards realization of thesegoals. The concerns of the responders have been communicated to the Raman Technical Groupof the Optical Society of America (Spring Meeting, Denver, Colorado, March 1973) and havebeen discussed with the chairman of the Raman Technical Group; Dr. Ted Becker,National Institutes of Health, Chairman of the Raman Committee of the CoblentzSociety; and Dr. Stephen Rossmassler, National Bureau of Standards, Office of StandardReference Data.
Recently, at the September 12, 1974, meeting of the Executive Committee of
the Joint Committee on Atomic and Molecular Physical Data, the topic of standards for
Raman data presentation was addressed. The results of this survey were presented as
part of the meeting agenda. As a result of this interaction, G. Rosasco, with the consulta-tion of the JCAMPD and a panel of Raman experts, will be preparing a manuscript whichwill serve to establish a preferred set of specifications for Raman data presentation.
Concerning heading II, the results of the survey generally are positive onthe usefulness of determining standard reference spectra (of standard reference materials)
,
and indicate a definite role for NBS in such an effort. The areas of effort are:
(1) Specification of standards and techniques for calibration of frequency,intensity and instrument response functions.
(2) Development, evaluation, and recommendation of standard methods for obtainingreference quality (Class II and I) spectra.
(3) Selection, measurement, and distribution of reference spectra of referencematerials.
The order of importance (and frequency) of suggestion is as numbered above.This order is based on a systematic approach to the standardization problem and thusconstitutes a reasonable framework for structuring a measurement strategy. Within thisstrategy provision should be made for multi-lab measurements of "standard" samples;this point was raised by a number of responders, and, in fact, some volunteered to cooperatein such an effort.
In the context of a research strategy based essentially on items (1) through(3), some initial efforts are under way by G. Rosasco in search of spectroscopic standardsand techniques for frequency and intensity calibration. As an example of this effort,we are inquiring into Bureau calibration of radioactively pumped phosphors as intensity
- 14 -
standards. The present intensity standard is a photometric standard whose specificproperties and certain concomitant difficulties in technical procedures (and high cost)cause it to be less than ideal for routine standardization of Raman spectrometers.
The experimental procedures to be addressed in item (2) involve questions asto laser power measurement, determination of polarization properties of incident andscattered light, and techniques for measurement and deconvolution of the effects ofinstrument functions of Raman spectrometers. G. Rosasco visited the MBS Boulder Labora-tories in March 1973 specifically to learn more about accurate laser power measurement.We have informed Raman spectroscopists via the RAMAN NEWSLETTER (August 1973) of theBureau's efforts (in conjunction with the ASTM Committee F-1.02) in a Measurement AssuranceProgram for laser power measurements. G. Rosasco is attempting to keep abreast of develop-ments at NBS (Boulder) and^ from ASTM F-1.02 in order to inform Raman spectroscopists.
Within the context of item (2) , it appears useful for NBS to communicate signifi-cant information from the "experts" in an associated problem area to Raman spectroscopists.The "experts" in many cases will come from the field of Raman spectroscopy, and in anyselection of a "standard" technique the particular needs of Raman spectroscopists mustbe considered. These efforts seem appropriate for NBS and follow from a significantnumber of suggestions from the responders.
A natural evolution from significant achievements under items (1) and (2) wouldbe specific efforts on the part of NBS and other interested parties toward fulfillmentof item (3) . There are a variety of strategies one can apply to the selection of referencematerials. Because of the many interests of Raman spectroscopists, it may he necessaryto select a number of materials. In some cases a "material" may have to be defined,e.g., a mixture of liquids or gases. A list of minimum requirements for a referencematerial should be selected and possible candidates chosen. In this process the expertsin each area (gas phase, aqueous solutions, etc.) should be consulted. Careful evaluationof the candidates should be made by a group of Raman spectroscopists; for example, theAdvisory Board of the RAMAN NEWSLETTER.
In addition to a role in the communication of important developments towardsthe realization of "standards" for Raman spectroscopy as outlined above, another importantcontribution could be made at this time. This effort would be a first round of multi-lab testing of a selected sample for the purposes of evaluation of routine Raman measure-ments. This idea evolved from a discussion at the Spring 1973 meeting of the RamanTechnical Group. A number of variations to the nature of the test were suggested.However, a simple, straightforward test specification seems most useful at this point.The procedure for the multi-lab test seems relatively straightforward and G. Rosascois willing to implement the testing, i.e., sample preparation, distribution and compilationof results. Further details will be specified after the basic concept is approved.
This development sequence is based on the knowledge of problem areas relatedto Raman spectroscopy. Solution of each level of problem and an adequate evaluationof routine test spectra will be invaluable to the design, measurement and successfulimplementation of a standard reference material.
- 15 -
4.0 SURVEY REPORT ON STANDARDS FOR
ELECTRON PARAMAGNETIC RESONANCE
4.1 INTRODUCTION
We report here the findings of our survey-questionnaire to assess the needfor a standard reference material (SRM) by which to improve intensity measurements inelectron paramagnetic resonance (EPR) . This report summarizes the requirements of thepotential users of EPR SRM's and reveals possible areas of difficulties where we mayrender technical assistance.
Table 1 gives an analysis of the survey sampling. The list for mailing wasassembled from the mailing list of an EPR spectrometer manufacturer and from our owncontacts. We enclose a copy of the questionnaire as figure 2, We sent out 518 question-naires. Of this number 15 either were returned by the Post Office as not deliverableor were sent to persons who are no longer performing EPR measurements. We thereforetake 503 as the base number. The number of answered questionnaires is 175. This is35% (175/503) of the base number. We received answers from 24 different companies, 26different government laboratories, and 90 different universities.
Table 1. Analysis of Sampling
NumberTotal number of questionnaires sent out 518
Returned by Post Office as not deliverable 13
No longer in EPR 2
Base nxjmber of questionnaires sent out 503
Answered questionnaires 175 (35%)
We estimate that there are about 1000 EPR groups in the world, and that thenumber of domestic EPR groups is between 400 and 500. Most of the foreign EPR groupslisted in Table 1 are from Europe, Canada, and several from Japan, Australia, and SouthAmerica. We did not reach east European countries, Russia, China, India, or Arabia.We know of several groups in these countries.
The background of the responders is given in Table 2. We prepared Table 2 fromour knowledge of the organizations of the responders. We divide the EPR users into threemajor disciplines; namely, physics, chemistry, and biomedicine. For this analysis weplace biologists, medical researchers, and pharmacists in the biomedical category. Werecognize that some physicists and chemists also work on biological problems. Thus,one quarter of the EPR users is in physics, one quarter is in biomedicine, and the remaininghalf is in chemistry. We find by reference to the journals in which researchers in physicsand biomedicine publish that there is a trend for the number of EPR users in biomedicine
- 16 -
FicMre ^ 0MB NC. 41-S72080
IS-721
U. S. DEPARTVEM OF CCV.MERCEMAT:CNAL BUREAU OF StAVUAKU>
QUESTIONNAIRE FCR ELECTRON PARAMAGNETIC RESONANCE (EPR)
STANDARD RErERL\CE VATEF.IALS
TO:
r
L J
PLEASE RETURN! THIS
QUESTIONNAIRE TO:
Dr. Te—Tse ChangNational Bureau of Staniiards
Bldg. 223, Roon A- 253
Washington, . C. 20234
la Do yOu think National Bureau of Starviards certified EPR stanoard referervce r%ater:ais are needed? YES
b. If rvo, d's^egati t"e 'z ow"-: c.'sst o-s S"' re:j-~. t-e c_5s: c--a -e
NO
2a. Have you ever ar.e-iptec to -r.easare tne njroer cf sdi-s r, YO_r experirrents'
b. If yes, what refererice sarr.pie did you use?
YES NC
c. How accurate f-»d precise v.as tne reference samole you used? Accuracy =reciSior-
% .
d. If the experiments are describee - s-r\' p_s cat.o-s, piease ser« a rep'int. pr cite re-ere-ce
3. if yoo v,e'e p-rc-3s -5 a set pf Erfi stancard reference materials, list your preference for the foliowir.g;
a. shape b. size C. concentration range of
paramagnetic ions
<!•. Within wtvat Krr.perature range wouid you use the EPR standard reference materiaiS'
e. List frequency DanMs) you:
(1 ) now operate in (2) expect to operate .n, in the future
f. How accurate would you want the EPS stanoard reference materiais to oe'
4, Any additional comrr>ents would be weico<T*ed
- 17 -
USCOMM-NSS -oc
to Increase and for the number of EPR users In physics to decrease.
TaUsle 2. Background of Responders
Physics Chemistry Biomedicine Not determined Total
35 71 29 • 40 175
4.2 FINDINGS AND DISCUSSION
4.2.1 Question 1 — Need
Question 1 assesses the need for EPR standard reference materials.Table 3 shows that 93% (163/175) of the responders answered "yes." Hence the EPR communityneeds and wants a standard reference material.
^
Table 3. Is a National Bureau of Standards certified SRM for EPR needed?
U.S.A. Foreign Total Percent
Yes 121 . 42 163 93%
No 10 2 12 7%
4.2.2 Question 2 — Present Reference Samples
Question 2 concerns the existing samples by which EPR users measure lineintensities. These samples are not calibrated nor certified. They estimate the spinconcentration by knowing the weight of the sample and its chemical composition.
Table 4 shows that 87% (129/157) of the responders to question 2 triedsome quantitative EPR measurements. Everyone who attempted quantitative EPR measurementsmentioned that the lack of standards presents great handicaps and difficulties. Amongthose who had not attempted quantitative EPR line intensity measurements, some statedthat they plan to do so in the near future. Others stated that they cannot make quantita-tive line intensity measurements because there are no standards. The general assessmentis that a good standard reference material is necessary for EPR intensity measurements.
Table 4 . Quantitative Measurement Attempted
U.S.A. Foreign Total
Yes 96 33 129
No 20 8 28
Table 5 lists the commonly used reference samples in order of the numberof times each sample was cited. In order to shorten this table we have combined someof the entries. For example, the item "copper salts in various forms" represents
- 18 -
CuSOii, CuCl2» and other Cu bearing salts in single crystal, powder, or solution forms.The copper content of such samples may also be diluted. This table shows that convenienceand availability are important factors in selecting reference samples.
Tcible 5. Existing Reference Samples
Reference Sample Number of times cited
DPPH 53
Pitch in KCl and Varian samples 37
Cu salt in various forms 37
Mn salt in various forms 16
Various free radials (except DPPH) 16
Spin label, in various forms 14
P-doped Si (Bell Laboratories sample) 8
Charcoal (including radiated carbon, burned^
sugar, etc.)
vo'''"'" in various forms 4
Ruby or Cr ions in various forms 4
02 2
F-center in KCl 1
Fe'*"'"'^:Al203 1
Table 6 presents the assessments by the responders themselves of theaccuracy and precision of their present reference samples. This table shows that a standardreference sample which results in EPR line intensity measurements accurate and preciseto within 10% will satisfy the present needs for over half of the responders. In addition,comments on the answered questionnaires, follow-up conversations with some of the respond-ers, and this Table 6 all indicate that we must educate some of the potential users ofEPR SRM's to understand the difference between accuracy and precision.
Table 6. Quality of Existing Reference Samples
Percent accuracy or precision Number of times cited for
stated by responders . Accuracy Precision
0.1% to 4.9% 14 15
5.0% 17 17
6.0% to 14.9% 25 19
15% to 50% 8 24
greater than 50% 2 11
Responders stated that accuracy andprecision for the number of spins is:
Not known 14 16
Poor 4 4
Known to an order of magnitude 6 6
- 19 -
If the EPR spectrometer has reliable and stable units, and if the experi-mental conditions are kept the same or are monitored for changes and the data are correctedaccordingly, the reproducibility of the EPR measurements may be precise to within 1%.
This is good precision but the correlation of the experimental results with the truenumber of spins is not straightforward, particularly when the correlation is not welldefined. Also, very accurate reference samples may be prepared, but transferring thisaccuracy to the EPR measurements may not be achieved readily. For example, the CuSOi|
solution may be prepared with a concentration known to a ±0.01% accuracy; however, thetransfer of this accuracy becomes questionable when the solution is placed in a quartztube and inserted into the microwave cavity without additional precautions. The mannersin which the reference sample and the unknown sample perturb the electric and magneticfields of the cavity must be determined first. That is, the user must know what quantityis being measured.
4.2.3 Question 3 — Sample Characteristics
Question 3 considers the shape, size, and concentration range of paramag-netic ions, temperature range, and frequency range. Table 7 summarizes the shapes and
sizes which the responders preferred. The item "cylinder" in this table includes alsorods and tubing. The item "cube" includes cxibes, squares, rectangles, and needles.Because most users place samples in quartz tiJoes, they would prefer a cylindrical standardsample. Most responders want sample standards 4 mm in size or smaller, so that placingthe standards near the unknown sample in the cavity does not disturb the measurement.The second largest number requested cube-shaped samples. We conclude that EPR SRM'swhich are small cylinders or cubes will satisfy the needs of most people.
Table 7. Shape or Form
Number of Times CitedCylinder 77
Cube 21
Single crystal 9
Powder 17
In solution 7
Size — 4 mm or smaller
Figures 3 to 6 are bar graphs and illustrate the distributions of concen-tration ranges for the paramagnetic ions (spins) expressed by the responders. Each figurecorresponds to one of the several ways in which responders expressed the concentration.Each vertical bar in these figures represents the range of concentrations requested byan individual responder. The average value for the concentration in the mole/liter(figure 3) is about 5 x 10"^ moles/liter. If we use a quartz tube with a 2-mm insidediameter in a 25-mm length, the volume is about 0.8 x 10"** liters and the number ofparamagnetic ions is in the 1 x 10^^ range. The average number for the paramagneticions in each of the remaining three figures is also about 10^^. Hence, from these datawe conclude that the first EPR SRM to be issued should contain about 10^ ^ spins and thatsucceeding EPR SRM's should then possess concentrations both greater and lower than thisvalue.
- 20 -
Figure 3. Molar concentration of paramagnetic ions
Molar (moles/liter)
10
10-1
10-2
10-3
10"
10-5
10"
10"
10-&
II
10-
Each vertical line represents the range of concentrationrequested by an individual responder.
Figure 4. Number of spina/gm
10^
lOll
10l3
10l9
1020
10^2
1023
Each vertical line represents the range ofnumber of spins per gram requested by anindividual responder.
- 22 -
Figure 5 . Number of spins/cm
10'
10-
10 10
10
10
11
12
10
10
13
lU
10
10
15
16
10
10
17
10
18
19
]020
1021
1022
II
1023
Each vertical line represents the range ofthe number of spins per cubic centimeterrequested by an individual responder.
- 23 -
Figure 6. Total number of spins in sample
108
lolo
10
10
10
11
12
13
lOl**
10l5
10l6
10
10
10
17
18
19
1020
1021
1022
10 23
Each vertical line represents the range inthe total number of spins requested by anindividual responder.
- 24 -
Most responders stated that we should provide the most accurate samplepossible. Our present knowledge and experience suggest that our first goal be an SRMwhich is 20% to 50% accurate. In the future we should work towards the 5% to 15% rangeindicated by the survey.
4.3 ADDITIONAL COMMENTS
Most of the answered questionnaires contained enthusiastic comments. Some respondersalso offered to assist us, and others were pleased that someone finally is engaged to
develop an EPR SRM.
The analysis of the answers to this questionnaire establishes that the EPR commu-nity needs standards. The first standard could be a single crystal in the shape of a
cylinder or cube which is small enough to fit into 4-mm inside-diameter tubing and whichhas about 10^^ spins. We should consider later issuing other concentrations and othersizes and exploring alternative materials for standards. We should develop continuingresearch and calibration efforts. And, finally, the purchasers of EPR SRM's should beprovided with technical help and advice, and, particularly as indicated by Tables 2 and
6, with detailed and extensive instructions on how to use EPR SRM's.
5.0 SURVEY REPORT ON STANDARDS FOR
MAGNETIC MOMENT MEASUREMENTS5.1 INTRODUCTION
At present, there are no reliable and accurate standard calibrants for magneticmoment measurements. For this reason, we at NBS are considering the preparation andcertification of a Standard Reference Material (SRM) for magnetic moment. In order toassess the need and to determine the characteristics for such a standard reference material,we have undertaken a survey-questionnaire. Figure 7 is a copy of the questionnaire.
During October 1974, 781 questionnaires were mailed to addresses in the UnitedStates. These addresses were obtained from the mailing list maintained by the AmericanInstitute of Physics for the annual Conference on Magnetism and Magnetic Materials.In this report the following classification is used: industry, university, government,private individual, and unknown. The classification "unknown" is for those questionnaireswhich were returned without a name or address on them.
Table 8 shows the mailing distribution of the questionnaire.
Table 8. Mailing Distribution
Classification Number Percentage
Industry 337 43
University 276 35
Government 116 15Private 52 7
Total 781 100
- 25 -
Figure 7.0MB i4l-S7'4077
APPROVAL EXPIRES; June. 1975
NBS-79'+
(9-7^4)
U. S. DEPARTMENT OF COMMERCE
National Bureau of Standards
QUESTIONNAIRE FOR MAGNETIC MOMENTSTANDARD REFERENCE MATERIALS
TO: PLEASE RETURN THIS
QUESTIONNAIRE TO:
r ~1 Office of Standard ReferenceHater lALS
Chemistry Bloc, Room 8311
National Bureau of StandardsWashington, 0. C. 2023^^
_l (a pre-aooressed, postage-paidenvelope is enclosed for your
convenience.)
The National Bureau of Standards is proposing to prepare ano certify a Standard Ref-
erence Material (SRM) for magnetic moment. The proposed SRM is for a 3mm-diameter
nickel sphere certified for saturation ma gn e t i za t i on a t room temperature to within
_+ 5 percent accuracy.
The purpose of this questionnaire is to determine the extent of the need FOp SUCH AN
SRM. Because questionnaires may not reach all interested parties, you are encouraged
TO submit the names of colleagues whom you feel would wish to receive the questionnaire
1. Do you need a standard reference material for magnetic moment to calibrate
magnetometers? (a) Yes (b) No (c) Not in my field of interest
IF (b) Or"(c )', D I SREGAR D the FOLLOwTng" QUEST I o'^^Ti^TeTuYn" THE QUEST I ONNA I RE
2. Do you now use a calibrant with your apparatus? (a) Yes (s) No
If "Yes^' please answer the following questions:
(A) What calibrant do you use?
(B) Is your apparatus adaptable to various sizes of samples? (a) Yes (b) No
(C) What is the maximum magnetic field obtained with your magnet?
(D) Please cite the references that describe your experiments, or send reprints.
- 26 -
, FigurR 7 .(cogtinuad] .
3i Please list your preferences for an SRM for magnetic hohent.
(a) Shape (if other than spherical) (e) Magnitude of magnetic moment
(b) Size (f) Temperature (other than room)
(c) Material (g) Other Specifications (list)
(d) Accuracy (if better than + 5%)
If the proposed SRM (3mm-diameter nickel sphere) were available for approximately$150, would you purchase it? (a) Yes (b) No
5. Please list the names and addresses of any persons you feel would be interested in
an SRM FOR magnetic moment.
additional comments
- 27 -USCOMM-NrtS-D,C
Table 9 lists the returns as of December 2, 1974.
Teible 9. Returns
Normalized Returns
Percent ofNumber of Number of Number Sent Out Relative
Classification Returns Returns (Table 8) Percent
Industry 69 120 36 40
University 71 124 45 42
Government 22 38 33 13
Private 9 15 33 5
Unknown 126
Total 297 297 100
We have normalized the returns by assuming the 126 unknown returns have the
same distribution among classifications as those returns with a name or address on them.
Both in number and percentage, a slightly larger response was obtained from universities.
5,2 DATA BASE AND FINDINGS
In this report, each question will be stated, followed by a summary of tlie
response to it, and then analyzed.
Question 1. Do you need a standard reference material for magnetic moment to
calibrate magnetometers?
(A) Yes (B) No (C) Not in my field of interest
Table 10. Response to Question 1
Classification Number of Returns
Yes No Not in my field of interest
Industry 35 25 9
Universities 38 is 15
Government 9 5 8
Private 2 .2 5
Unknown 40 43 43
Total 124 93 80
Percent of response 42 (124/297) 31 (93/297) 27 (80/297)
Percent of thosein field 57 (124/217) 43 (93/217)
A majority (57%) of the responses from workers in the field claims that a need for anSRil exists. The remainder of this report will analyze the 124 questionnaires on which the
response to question 1 was yes.
- 28 -
Question 2. Do you now use a calibrant with your apparatus?
Yes No
Yes - 103 No - 16 Percent Yes - 87
Of the 124 who claimed they need an SRM/ 103 or 83% already are using a calibrant. Thisresponse and that to Question 1 indicate that these workers are not completely satisfiedwith their own reference material and would like an SRM certified by a reputable standardslaboratory.
Question 2, continued
If "Yes," please answer the following questions:
2A. What calibrant do you use?
Table 11. Response to Question 2A
Number Percent
Ni 78 60Area-turns 15a 12Paramagnetic salts lib 9
Fe 8 6
Ferrites 5 4
Iron oxides 4 3
Superconductors 4 3
Others 4c 3
Total 129
of Total
^Includes reference magnets (5) , reference voltage (1)
^Persons informed of existing SRM' s for paramagnetism^h, Ge, Ta, Mo
The majority of those who use a calibrant use nickel.
2B. Is your apparatus adaptable to various sizes of samples?
Yes No
Yes - 98 No - 12
Although 89 percent (98/110) answered yes to this question, a size preference is shown
later in question 3B.
- 29 -
2C. What is the maximum field obtained with your magnet?
Table 12. Response to Question 2C
Field (tesla) Number Percent of Total
0-0.1 7 6
0.1 - 0.5 9 7
0.5 - 1.0 25 21
1.0-2.0 34 28
2.0 - 3.0 17 14
..,^3.0 - 5.0 7 . • 6
....^.t** 5.0 - up 22 18
i
Total 121
Most of the fields above 3 teslas are produced by superconducting magnets. Approxi-mately 13% of the responses need a moment characterized in a field of less than 0.5 tesla.
2D. Please cite references that describe your experiments, or send reprints.
References cited 51
Reprints sent 22
The responses from questions 2A and 2D show that the majority of the workers (124)
are active in the field of magnetism.
Question 3. Please list your preferences for an SRM for magnetic moment.
There are seven parts to this question (A-G) . These are discussed separately.
3A. Shape (if other than spherical)
Table 13. Response to Question 3A
Shape Number Percent
sphere* 96 • 72cylinder 20 15
film or disc 15 11
ellipsoid 3 2
Total 134
Includes those who wrote in sphere (22) and those who left3A blank. See questions 2A and 4. The vast majority choseel sphere which is easy to fabricate and whose demagnetizationfactor is well established.
- 30 -I
3B. Size
Table 14 . Response to Question 3B
Size (diameter mm)* Number Percent
1 9 162 15 27
3 20 36
4 4 7
5 4 7
10 4 7
Total 56
*Only the size for a sphere has been tabulated.
More than one size sphere may be needed to satisfy the requests.
3C. Material ^
Table 15. Response to Question 3C
Material Numbsr Percent
Ni 38 75
Fe 6 12
Other* 7 13
Total 51
*Fe203 (2) , nonmetallic (2) , ferrite (1) , Permalloy (1)
,
superconductor (1)
.
There was a low response to this question; however, question 2A showed that mostworkers used nickel.
3D. Accuracy (if better than ±5 percent)
Table 16. Response to Question 3D
Accuracy (%) Number Percent
.1 - .5 12)
.6 - 1.0 30> 49
1.1 - 4.0 16)4.1 - 5.0 _61 51
Total 119
This is a very important question and clearly shows that about one-half of the workerswould like an SRM certified to better than ±5 percent accuracy. One notes that our high
- 31 -
response for a ±5 percent accuracy may have been prejudiced by our phrasing of the question.However, the responses to question 4 and the additional comments given below indicate thata ±1 percent SRM is needed.
3E. Magnitude (emu/g)* Number
< .1 10> .1 12
Only 22 persons answered this question, but it appears froifl other questions aboutmaterials that the majority of the workers need an SRM with amoment > .1 emu. (See
questions 2A, 3C, and 4. Also note that the saturization magnetization for nickel is
approximately 50 emu/g.)
*Most workers who responded to this questionnaire use(emu/g) for the "cgs" units of magnetic moment. To convertthis td,. SI units, multiply by 47r x l5'* p, where p is the
denisity of the material in g/cm^
.
3F. Temperature (other than room)
Temperature Number
4.2 K 26
77 K 13
4.2 K to 293 K 4
43
Approximately 35% of the workers would like an SRM to check or calibrate theirequipment at low temperatures
.
3G. Other Specifications (list)
Only one person responded, and the response did not pertain to magnetic momentconsiderations.
Question Ai If the proposed SRM (3 mm diameter nickel sphere) were availablefor approximately $150, would you purchase it?
Yes No
Table 17. Response to Question 4
Yes, if accuracy PercentClassification No Yes is better than 5% Total Total Yes
(write in)** Yes (normalized)
Industry 5 23 7 30 46
University 11 21 6 27 44
Government 4 6 6 9
Private 2 2 1
Unknown 9 26 4 30
Total 29 78* 17 95
Percent 23 63 (78/124) 14 77
*This includes 19 maybe 's.
Seventeen people who answered this question wrote in "Yes, if accuracy is
better than 5%."
- 32 -
The responses to this question show that 63% of the 124 responses would buy the 5% SRMfor $150 even though many prefer a more accurate SRM. Response to question 3D showed that50% of the workers claimed they needed an SRM to better than 5%. A total of 77% (95/124)
would purchase the SRM if it were certified to better than 5% accuracy. Approximately 46%
of the potential customers are companies, 44% universities, and 9% government. Only 14%
(17/124) would not buy an SRM with an accuracy of 5%.
Question 5. Please list the names and addresses of any persons you feel would beinterested in an SRM for magnetic moment. This question brought 44 responses.
5.3 POSSIBLE USERS
We mention here who the users of these standard reference materials might be
.
5.3.1 Industry
Users of vibrating sample magnetometers in industry require such standardreference materials for calibration purposes. Vibrating sample magnetometers are commer-cial instruments and are widely used. In fact, two SRM's with different moments have beensuggested. They would be used to determine the linearity of magnetometers.
Producers and customers of rare earth cobalt magnets need such standardreference materials to certify field measurements traceable to NBS.
5.3.2 Universities
Researchers at universities also request SRM's for vibrating samplemagnetometers. Others state a requirement for magnetization standards with momentsas small as 10"^ emu/g. The latter standards interest researchers concerned with the
magnetism of rocks (both terrestrial and lunar) . Some investigators suggest standardsto calibrate mutual inductance systems for measuring differential susceptibility at
liquid helium temperatures. This latter suggestion involves a non-metallic sample.Preferably the sample should be an optically transparent, magnetic insulator with a
known Faraday rotation.
5.3.3 Government and Others
Additional' potential users mention the need to calibrate sensitive Faradaymagnetometers, Faraday microbalances , torque magnetometers, and hysteresigraphs by
insertion of known samples
.
5.4 SUMMARY
The 124 responses were from workers who are active in the field of magnetism.
Approximately 87% of these 124 workers use a calibrant (mainly nickel) . Approximately
one-half of the workers want a certified SRM. However, some workers have confused
precision and accuracy. Also, none of the workers who have used a reference material
volunteered to state the accuracy of their own calibrant. Some workers thought the price
of $150 was too high.
There is an uncertainty in the value of saturation magnetization for nickel in
the literature. Values given by reliable sources vary by about 8%. Workers do not know
what value to use.
- 33 -
If NBS issues an SRM for magnetic moment, it should probably have the followingproperties
:
1. Accuracy within ±1%.
2. More than one size (perhaps a 1-mm and a 3-mm sphere)
.
3. Material most likely of nickel.4. A price less than $150.00.5. The actual magnetic moment determined at known values of magnetic
field (H) greater than 5000 oersteds.
5.5 TECHNICAL REFERENCES
(1) "Calibration of Vibrating-Seunple Magnetometers," J. Res. Natl. Bur. Stand.
(U.S.) C. Engineering and Instrumentation, Vol. 70C, No. 4, October-December1966. W. E. Case and R. D. Harrington.
(2) "The Magnetization of Pure Iron and Nickel," Proc. Roy. Soc. London A321 ,
447-491 (1971). J. Crangle and G. M. Goodman.
(3) "New Determinations of the Saturation Magnetization of Nickel and Iron,"J. Appl. Phys. 39^, 2 (1968). H. Danan, A. Herr, and A. J. P. Meyer.
6.0 ACKNOWLEDGMENTS
We wish to acknowledge and to thank the many people both at the National Bureau ofStandards and outside the Bureau who have assisted us in these three surveys. We thankthe staffs of the Office of Standard Reference Materials and the Office of StandardReference Data. In particular, we acknowledge Mrs. Ruth Meyer for her help with the
over-all design and format of the questionnaires. We gratefully thank Frances Fussingerfor administrative assistance and for typing and preparing this report.
USCOMM-NBS Dt - 34 -
KeS-n4A :f!Ev. 7-73)
U.S. DEP-. CP COMM.BIBLIOGRAPHIC DATA
SHEET
1. PUBLICATION OR REPORT NO.
NBSIR 75-759
2. Gov't Accf.ssionNo.
3. Rec ipicnt '.s Acccs.sion No.
4. TITLE .AND SLBTlTLii
STANDARD REFERENCE MATERIALS AND DATA FOR RAMANSPECTROSCOPY, ELECTRON PARAMAGNETIC RESONANCE, ANDMAGNETIC MOMENT MEASUREMENTS
5. Publication Date-
July 19756. Performing Organization Code
7. .AUTHOR S-
E. S. Bennett, G. A. Candela, T. Chang, R. E. Mundy, and G. J. I
8. Performing Organ. Report No.
^osasco NBSIR 75-7599. PERFOR.MLNG ORGANIZ.ATION NAME AND .ADDRESS
NATIONAL BUREAU OF STANDARDSDEPARTMENT OF COMMERCEWASHINGTON, D.C. 20234
10. Project Task '''ork Unit No.
11. Contract /Grant No.
12. Sponsoring Organization Name and Complete Address (Street, City, State, ZIP)
Same as No. 9
13. Type of Report & PeriodCovered
Final
14. Sponsoring Agency Code
15. SUPPLE.VIENTARY NOTES
16. -ABSTR.ACT fA 200--Aord or less factual summary of most si^ificant information. If document includes a significant
bibliography or literature survey, mention it here.)
This publication contains the analyses of three survey-questionnaires to assessthe needs and markets for standards in Raman spectroscopy, electron paramagneticresonance, and magnetic moment measurements. These analyses give the requirementsand suggest the strategies to be followed for making each of the above threemeasurement methods more quantitative than they are at present. Some of therequirements are data cataloging, standards for data presentation, standardreference materials, and standard reference data.
17. KEY UORDS (six to twelve entries; alphabetical order; capitalize only the first letter of the first key word unless a proper
name; separated by semicolons
)
Electron paramagnetic resonance; magnetic moment; Raman spectroscopy;
standard reference data; standard reference materials.
18. AVAILABILITY [3 Unlimited 19. SECURITY CLASS(THIS REPORT)
21. NO. OF P.AGES
X For Official Distribution. Do Not Release to NTISUNCL ASSIFIED 38
Order From Sup. of Doc, U.S. Government Printing OfficeWashington. D.C. 20402. SD Cat. No. C13
20. SECURITY CLASS(THIS PAGE)
22. Price
Order From National Technical Information Service (NTIS)Springfield, \'irginia 22151 UNCLASSIFIED
USCOMM.DC 29042-P74