U.S. EPA Watershed Management Research: Experimental Stream Facility and Associated Field Studies - Standard Operating Procedure TITLE: Total Phosphorus and Total Dissolved Phosphorus Wet Chemistry SOP No. ESF-SOP 030 Last Revision Date: 2/18/2011 Last Review Date/Initials: ________/_________ PI Approval for Use: ___________________________________________ Date: _____________ QA Concurrence: _______________________________ __________ Date: ____________

Upload: others

Post on 13-Jan-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

U.S. EPA Watershed Management Research: Experimental Stream Facility and Associated Field Studies - Standard Operating Procedure

TITLE: Total Phosphorus and Total Dissolved Phosphorus Wet Chemistry

SOP No. ESF-SOP 030

Last Revision Date: 2/18/2011

Last Review Date/Initials: ________/_________

PI Approval for Use: ___________________________________________ Date: _____________

QA Concurrence: _______________________________ __________ Date: ____________

Page 2: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

Watershed Management Research SOP 030TP-TDP - Wet Chemistry

Version Date 02/18/11Page 2 of 14

Revision Summary

Revision Name Date Description of Change

0 Christopher Nietch, Ph.D. 3/15/2005 Developed SOP

1 Christopher Nietch, Ph.D. 02/21/2006 Reformat SOP

2 Don Brown 2/18/2011 Editorial clarifications

Page 3: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

Watershed Management Research SOP 030TP-TDP - Wet Chemistry

Version Date 02/18/11Page 3 of 14

TABLE OF CONTENTS

SECTION NUMBER SECTION TITLE PAGE

NUMBER

1.0 Scope and Applicability 4

2.0 Summary of Method 4

3.0 Definitions 4

4.0 Health and Safety 6

5.0 Cautions 6

6.0 Interferences 7

7.0 Personnel Qualifications 7

8.0 Equipment and Supplies 7

9.0 Procedures 8

10.0 Calculation 11

11.0 Data and Records Management 11

12.0 Quality Assurance/Quality Control 11

13.0 References 12

Table 1. Reactive Phosphate-TDP/TP Wet Chemistry Reagent , Standard, and Spike Recipes, Holding Times, Container Specs, Disposal Method, Notes, and Tracking

Appendix A: Analytical Schedule Worksheets for TDP and TP Analysis

12-13

Page 4: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

Watershed Management Research SOP 030TP-TDP - Wet Chemistry

Version Date 02/18/11Page 4 of 14

1.0 Scope and Applicability

1.1 Total Phosphorus (TP) and/or Total Dissolved Phosphorus (TDP) in surface water and intercobble (pore) water is analyzed as orthophosphate after oxidation of all phosphorus species in the sample under an acid persulfate digest procedure. Once samples have been digested this standard operating procedure is equivalent to ESF-SOP-029 where the orthophosphate ion (PO43-) reacts with ammonium molybdate and antimony potassium tartrate under acidic conditions to form a complex. This complex is reduced with ascorbic acid to form a blue complex which absorbs light at 880 nm.

1.2 The applicable concentration range for this SOP is 0 to 5 mg P/L as PO43-. The method detection limit was determined at 0.008 mg/L for a 0.5 mg/l phosphate standard. The SOP is also applicable over an equivalent range for wastewater and groundwater samples. The SOP throughput is approximately 75 injections per hour.

2.0 Summary of Method

2.1 This SOP covers procedures that take place in the wet chemistry laboratory after samples have been processed according to ESF-SOP-007. The TP and TDP determinations are made on raw or filtered filtered water samples, respectively, after sample digestion using the Latchat Quickchem 8000, Flow Injection Analysis, Autoanalyzer (Hach Co.).

2.2 The wet oxidation digestion procedure is designed to convert all organic phosphorus species quantitatively to orthophosphate with and acid digestion reagent. The orthophosphate ion (PO43-) reacts with ammonium molybdate and antimony potassium tartrate under acidic conditions to form a complex. This complex is reduced with ascorbic acid to form a blue complex which absorbs light at 880 nm.

2.3 The project principle investigator has optimized the Latchat instrument for all nutrient analyses required. Four independent channels are set with mixing manifolds in line with a colorimeter (spectrophotometer). For each sampling event analytical schedules are run that cover all nutrient species of interest. For the TP-TDP determination the third schedule is of interest and is set to analyze these constituents on a 5 ml sample split (5 ml each). A mixed standard series is prepared and used to calibrate the instrument for both TP/TN and TDP/TDN analyses. (see Table 1). The digestion reagent may be adjusted chemically to analyze these nutrients simulatenously following the method of Patton and Kyrskalla, 2003 with volumetric adjustments made based on sample volumes.

2.4 Sampling preparation, processing, and analytical blanks are included in each schedule, as are Matrix spikes, calibration check standards, recovery standards, and analytical duplicates as indicated in the example sample schedule provided in Appendix A.

2.5 This nutrient analysis protocol represents a unique application of the Latchat Flow Injection Analysis System and was developed specifically for the ESF-East Fork Watershed Study to meet the sample handling and loading requirements.

2.6 This TP-TDP SOP was adopted and slightly modified from the EPA approved Latchat QuikChem Method 10-115-01-4-E (Boren, 2001), which was developed from AWWA standard method 4500-P B.5. Persulfate Digestion Method, 1995).

3.0. Definitions

Dissolved (nutrient form) Dissolved analytes are qualified based on a glass fiber filter size of .47 um (Whatman GF/F).

Page 5: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

Watershed Management Research SOP 030TP-TDP - Wet Chemistry

Version Date 02/18/11Page 5 of 14

Total (nutrient form) Total analytes are qualified based on an unfiltered aliquot of the

original environmental sample. The sample collection bottle is shaken vigorously for homogenization before the aliquot is taken for

DI- water Water taken from a Reverse Osmosis, Ion Exchange Resin, or Distillation system and then passed through a second ion exchange system. Water of this quality should be no more than 18 ohm resistivity. The water meets ASTM II criteria and is considered reagent grade.

Analytical Batch The set of samples extracted/distilled/or digested at the same time.

Calibration Standard A solution prepared from the primary dilution standard solution or stock standard solutions. The solutions are used to calibrate the instrument response with respect to analyte concentration. Calibration standards are run through the digestion procedure thereby implicitly accounting for dilution of the original sample volume with the addition of the digestion reagent.

Field Blank An aliquot of reagent water or equivalent neutral reference material treated as a sample in all aspects, including exposure to a sample bottle holding time, preservatives, and all preanalysis treatments. The purpose is to determine if the field or sample transporting procedures and environments have contaminated the sample.

Field Duplicate Two samples taken at the same time and place under identical circumstances which are treated identically throughout field and laboratory procedures. Analysis of field duplicates indicates the precision associated with sample collection, preservation, and storage, as well as with laboratory procedures.

Lab Blank An aliquot of reagent water or equivalent neutral reference material treated as a sample in all aspects, except that it is not taken to the sampling site. The purpose is to determine if the analytes or interferences are present in the laboratory environment, the reagents, or the apparatus.

Laboratory Duplicate Two aliquots of the same environmental sample treated identically throughout a laboratory analytical procedure. Analysis of laboratory duplicates indicates precision associated with laboratory procedures but not with sample collection, preservation, or storage procedures.

Method Detection Limit The lowest level at which an analyte can be detected with 99 percent confidence that the analyte concentration is greater than zero.

Recovery Standard A solution prepared containing a known quantity of an organic nitrogen species. In this case Adenosine Triphosphate and/or Glutamic Acid are used. The recovery standard tests the efficiency of the digest procedure to convert the organic nitrogen species to nitrate.

4.0 Health and Safety Warnings

4.1 Standard laboratory personal protective equipment (i.e., lab coat, gloves, safety glasses, and safety shoes) are required in wet chemistry lab where analysis is performed.

Page 6: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

Watershed Management Research SOP 030TP-TDP - Wet Chemistry

Version Date 02/18/11Page 6 of 14

4.2 The toxicity or carcinogenicity of each reagent used in this method has not been fully established. Each chemical should be regarded as a potential health hazard and exposure should be as low as reasonably achievable. Cautions are included for known extremely hazardous materials.

4.3 Each laboratory is responsible for maintaining a current awareness file of the Occupational Health and Safety Act (OSHA) regulations regarding the safe handling of the chemicals specified in this method. A reference file of Material Safety Data sheets (MSDS) should be made available to all personnel involved in the chemical analysis. The preparation of a formal safety plan is also advisable.

4.4 The following chemicals have the potential to be highly toxic or hazardous, for detailed explanations consult the MSDS.

4.4.1. sulfuric acid4.4.2. antimony4.3.3. potassium persulfate

4.5 Safety precautions must be taken into consideration for operating an autoclave. The culture tubes used in the digest are capped tightly. Therefore, an appropriate time for cooling of the samples after autoclaving must be given to avoid an explosion hazard. Typical precautions for handling hot liquids must also be taken into consideration.

5.0 Cautions

5.2. Wear gloves.

5.3. Samples run in mixed batches with Nitrate and Ammonium wet chemistries can not be preserved and must be run within 48hrs. For phosphate only, the maximum holding time is then 48 hrs.

5.4. Samples should be collected in plastic or glass bottles. All bottles must be thoroughly cleaned and rinsed with reagent water. Volume collected should be sufficient to insure a representative sample, allow for replicate analysis (if required), and minimize waste disposal.

5.5. Add reagents in the order that they appear on the manifold to reduce staining.

5.6. See Table 1 for appropriate disposal methods for unused reagent solutions. The effluent line from the phosphate manifold shall be directed into its own waste container.

6.0 Interferences

6.1. Silica forms a pale blue complex which also absorbs at 880nm. This interference is generally insignificant as a silica concentration of approximately 30 mg/l would be required to produce a 0.007 mg p/l positive error in orthophosphate.

6.2. Glassware contamination can be a problem in low level phosphorus determinations. Glassware shall be washed with 10% HCL solution and rinsed with deionized water. Commercial detergents should never be used.

7.0 Personnel Qualifications

Page 7: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

Watershed Management Research SOP 030TP-TDP - Wet Chemistry

Version Date 02/18/11Page 7 of 14

7.1 Each scientist/technician is responsible for following this SOP as written and notifying the author, QA officer, and or the Project Leader, as appropriate, of equipment or procedural problems found during the use of this SOP.

7.2 The techniques of a first-time scientist/technician shall be reviewed by a senior scientist/technician prior to initiating this SOP alone.

7.3 In house training on the Latchat Quikchem 8000 is required for sample analysis.

8.0 Equipment and Supplies

8.1. Balance -- analytical, capable of accurately weighing to the nearest 0.0001 g.

8.2. Glassware -- Class A volumetric flasks, pipettes or plastic containers as required. Samples may be stored in plastic or glass.

8.3. Flow injection analysis equipment designed to deliver and react sample and reagents in the required order and ratios.

8.3.1. Sampler8.3.2. Multichannel proportioning pump8.3.3. Reaction unit or manifold8.3.4. Colorimetric detector8.3.5. Data system

8.4. Special Apparatus8.4.1. Heating unit Latchat Part No. A85X00 (X=1 for 110V, X=2 for 220V)

8.5. Vortex Mixer

8.6. Digital Pippettor for dilution of stock standard solutions and delivery of spike solutions

8.7. Helium tank, regulator, and attached degassing tube.

8.8. Disposable glass culture tubes; serve as Latchat system sampling ‘cups’. See ESF-SOP-007 for specifications.

8.9 Chemicals used to make reagents and standards in Table 1.

8.10 Autoclave

9.0 Procedure

9.1. Prepare Reagents

9.1.1. Use ASTM Type II water for all reagents and standard solutions.

9.1.2. Degas reagents with helium, except for the digestion solution solution:To prevent bubble formation, degas all solutions except the standards with helium. Use He at 140kPa (20 lb/in2) through a helium degassing tube (Lachat Part No. 50100.). Bubble He through the solution for one minute.

9.1.3. Reagent 1. Stock Ammonium Molybdate Solution (See Table 1) Mix with a magnetic stirrer for at least four hours.

Page 8: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

Watershed Management Research SOP 030TP-TDP - Wet Chemistry

Version Date 02/18/11Page 8 of 14

9.1.4. Reagent 2. Stock Antimony Potassium Tartrate Solution (See Table 1)Mix until dissolved store in a dark bottle.

9.1.5. Reagent 3. Molybdate Color Reagent (See Table 1)Degas with helium

9.1.6. Reagent 4. Ascorbic Acid Reducing Solution (See Table 1)Degas with helium then add 1 g dodecyl sulfate. Prepare fresh weekly

9.1.7. Reagent 5. Carrier, .13M Sulfuric Acid (See Table 1).

9.1.8. Reagent 6. Sodium Hydroxide – EDTA Rinse (See Table 1)

9.1.9. Reagent 7. Digest Solution-Potassium Persulfate (See Table 1).

9.7.10 Reagent 8. Digest Solution 2 – Sulfuric Acid (See Table 1).

9.2. Prepare Calibration Standards, Recovery Standards and Spike Solution as described in Table 1 of ESF SOP 028. 5 ml of calibration and recovery standards are pipetted into digestion tubes.

9.3. Prepare samples for analysis following sample processing procedures outlined in ESF-SOP-007. At the end of this step there should be a ‘digestion’ tray of culture tubes ready for addition of digestion reagent (See example sample schedule in Appendix A).

9.4. Prepare Matrix Spikes (See Table 1). After the spike has been added to the sample, vortex the sample cup for 5 seconds and return to the digestion tray.

9.5. Add specified amount of digestion solution to each 5 ml sample aliquot and QC samples. Cap the each sample tightly with a Teflon faced phenolic screw cap, and invert to mix.

9.6. Autoclave the unknown samples and QC samples at 110oC for at least 45 minutes. Allow time for the samples to cool. If using the Community Autoclaves in the AWBERC facility they are programmed to allow for appropriate cool down.

9.7. Transfer all digested samples into autosampler sample cups following the sequence specified in the analytical schedule (See Appendix A for an example).

9.8. Set up Phosphate manifold as specified in Prokopy 2000. At the start of a new ESF experiment,

9.9. Turn on the Latchat Quickchem Colorimeter Unit.

9.10. Input analytical system parameters in the latchat software program as specified in Diamond, 2001. This occurs upon first application of the method. If the manifold specifications do not change or the manifold is not switched out between runs then this information remains saved in the software system.

9.11. Attach reagent lines and condition pump tubes by rolling on a flat surface for several seconds. Replace pump tubes every 6 months or earlier as required.

9.12. Fill a container with fresh DI-water or AWBERC house RO water. The reagent lines are placed in this container to serve as the pre-analysis rinse of the manifold. Fill a separate container with DI-water and use this for the probe rinse.

Page 9: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

Watershed Management Research SOP 030TP-TDP - Wet Chemistry

Version Date 02/18/11Page 9 of 14

9.13. Turn pump on and select the “manual” button. If needed spray rollers with silicone and wipe

with kim-wipe or equivalent. The pump rollers should be greased at least quarterly.

9.14. Attach pump compression brackets to each reagent line, probe rinse line, and sample line, and then to the pump.

9.15. Pump DI water through all reagent lines and check for leaks and smooth flow. Switch lines to the reagent bottles, making sure effluent line from the colorimeter is directed into the appropriate waste container. Allow reagent to pass through the manifold for at least 3 minutes, and then switch the valve on the manifold to put the cadmium column in-line. Allow the system to equilibrate until a stable baseline is achieved. Baseline stability is viewed by selecting the “preview” button in the software system.

9.16. Turn on the autosampler.

9.17. Place standards and samples in the autosampler. Input the schedule information required by the software system (See Hach 2003). Follow the sample ID scheme outlined in ESF-SOPs-007 and 010. See Appendix A for an example. Sample Schedules can be imported from an excel file saved in csv format. The format for the csv file is: sample number, sample id, cup number.

9.17.1. Digested Calibration standards are scheduled to begin each run. The software system will then associate the concentrations with the peak area for each standard to determine the calibration curve. Calibration curves with r-squared values less then 0.95 are un-acceptable. If this occurs standards need to be prepared again and re-run.

9.17.2. Digested Quality control Blanks are scheduled to run immediately after the calibration curve is verified. Blanks returning concentrations greater than 2.5 times the method MDL (0.008 mg/l) are considered contaminated and a correction of the final concentration will have to be applied to the sample concentrations. The equation for the blank correction factor is:

Blank Corrected Concentration = D-Bave

Where D = Determined concentration of analyte and Bave = is the average concentration of blank replicates. Blanks are run at 2% of an analytical batch.

9.17.3. Quality control Analytical Duplicates are scheduled to run on every 10th sample. After a run compute the relative percent difference (RPD) between two sample results using the following equation:

Where, D1 = Concentration of analyte in the sample, D2 = Concentration of analyte in the second (duplicate) sample. RPDs for nitrate-nitrite wet chemistry falling outside of ±7% fail QC and the data must be flagged accordingly before submission to the ESF database administrator.

9.17.4. Analyses of matrix spike samples are required to demonstrate the method accuracy and to monitor matrix interferences (interferences caused by the sample matrix). The laboratory must spike a minimum of 10% of all samples (one sample in each batch of ten samples) in an analytical batch. The concentration of the spike in the sample shall be determined as 1 to 5 times higher than the background concentration (see Table 1).

Page 10: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

Watershed Management Research SOP 030TP-TDP - Wet Chemistry

Version Date 02/18/11Page 10 of 14

Calculate the percent recovery (P) of the analyte in each aliquot using the following equation.

Where, A = Measured concentration of analyte after spiking, B = measured background concentration of analyte, T = True concentration of the spike. Spikes for nitrate wet chemistry falling outside of ±15% fail QC and the data must be flagged accordingly before submission to the ESF database administrator.

9.17.5 Verify calibration using a midrange digested calibration standard every ten samples or once every analytical batch if fewer than 10 samples exist in a batch. Compute the percent recovery at the end of the run using the following equation:

%Recovery = D/K * 100

Where, D = Determined concentration of analyte in the calibration standard, K = Actual concentration of the analyte in the calibration standard. If % recovery exceeds +/-10%, the analytical system is judged to be out of control, and the problem must be immediately identified and corrected and the analytical batch reanalyzed.

9.17.6 Verify efficiency of the digestion procedure by analyzing recovery standards at least 10% of an analytical batch Compute the percent recovery at the end of the run using the following equation:

%Recovery = D/K * 100

Where, D = Determined concentration of analyte in the recovery standard, K = Actual concentration of nitrogen in the organic nitrogen recovery standard. If % recovery exceeds +/-20%, the analytical system is judged to be out of control, and the problem must be immediately identified and corrected and the analytical batch reanalyzed.

9.18. Configure the Autosampler and Autodiluter according to Latchat Instruments User Manual (Hach Co. 2003).

9.19. Before starting the run make sure that the pump selection button “normal operation” has been selected. This causes the pump to turn to a low speed setting when the run schedule has finished and minimizes wasting reagents. Prompt data system to start analysis.

9.20. During analysis check progress periodically to insure steady baseline and to look for out of range values. If returned values fall outside the range of calibrants (>5 mg/l). Re-run the sample by copying and appending this row to the end of the schedule and triggering an autodilution of the sample at 1:10 in the software. At the end of the run, review the peak profiles for each sample noting any irregularities (drifting base line or air spikes). Re-run identified irregular samples if required.

9.21. Return the reagent lines to the DI-Water rinse and pump for several minutes.

9.22. Remove the reagent lines from the DI-Water rinse and pump air through the manifold until dry (several minutes).

P=( A−B)100T

Page 11: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

Watershed Management Research SOP 030TP-TDP - Wet Chemistry

Version Date 02/18/11Page 11 of 14

9.23. Export run schedule output to a csv file (See Hach 2003) and import into an Excel Worksheet.

Run QC calculations/corrections described above in step 9.14 and flag data as appropriate.

9.24. Send data report to ESF Database administrator for upload to ESF-Database.

10.0 Calculation

10.1. Calibration is done by injecting digested calibration standards. The data system will then prepare a calibration curve by plotting response versus standard concentration. Sample concentration is calculated from the regression equation. A linear or 2nd order polynomial fit to the calibration curve may be used.

10.2. Refer to the procedures section for calculations associated with quality control samples

11.0 Data and Records Management

11.1. The Analytical Schedule Worksheets define the sample IDs for inclusion in the Software program that compiles the carbon and nutrient analytical data. Any occurrences that may affect data quality during the bottle preparation, sampling, and sampling splitting or filtration procedures shall be noted on these sheets. Hard copies of these worksheets shall be tracked by the manager of the Wet Chemistry Laboratory conducting the analyses.

11.2. Report only those values that fall between the lowest and highest calibration standards. Samples exceeding the highest standard should be diluted and reanalyzed.

11.3. Report sample results for nitrite-nitrate in mg N/L as NO2-3 to three significant figures.

11.4. Flag data failing QC criteria and include a metadata file explaining the reason for the flagged values.

12.0 Quality Control and Quality Assurance

12.1 Analyses of matrix spike samples are required to demonstrate method accuracy and precision and to monitor matrix interferences (interferences caused by the sample matrix). The procedure and QC criteria for spiking are described in section 9.14.

12.2. Analyses of laboratory blanks are required to demonstrate freedom from contamination.

12.3. The laboratory shall, on an ongoing basis, demonstrate through calibration verification that the analysis system is in control. Calibration is verified using a midrange calibration standard every ten samples or every analytical batch.

12.4. The laboratory should maintain records to define the quality of data that is generated.

12.5. Compute the relative percent difference (RPD) between two sample results. The RPD for duplicates shall meet the current laboratory acceptance criteria. If the criteria are not met, the analytical system is judged to be out of control, and the problem must be immediately identified and corrected and the analytical batch reanalyzed.

Page 12: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

Watershed Management Research SOP 030TP-TDP - Wet Chemistry

Version Date 02/18/11Page 12 of 14

12.6. All laboratory equipment and instrumentation used in this SOP that does not have specified

calibration procedures herein shall be calibrated and maintained in the manner specified under the QC procedures outlined in ESF-SOP-039.

13.0. References

AWWA 1995. Standard methods for the examination of water and wastewater. 19th Edition. APHA, AWWA, WEF, Washington, DC 20005.

Prokopy, W.R. 2000. QuikChem Method 10-115-01-1-B. Latchat Instruments, 5600 Lindburgh Drive, Loveland CO 80539.

ESF-SOP-007. Analytical Schedules, Sample Bottle Preparation and Sample Filtration for Carbon and Nutrient Wet Chemistry Analyses. On file in the Experimental Stream Facility System Control Room or available upon request from Christopher T. Nietch, U.S. EPA, National Risk Management Research Laboratory, 26 West Martin Luther King Drive, MS 690, Cincinnati, OH 45268.

ESF-SOP-039. Verification of Experimental Stream Facility Support Equipment and Deionized Water System. On file in the Experimental Stream Facility System Control Room or available upon request from Christopher T. Nietch, U.S. EPA, National Risk Management Research Laboratory, 26 West Martin Luther King Drive, MS 690, Cincinnati, OH 45268.

Hach Company. 2003. QuickChem FIA+ Automated Ion Analyzer: User Manual. Lachat Instruments a Hach Company Brand, USA.

Patton, C.J., and Kryskalla, J.R., 2003, Methods of analysis by the U.S. Geological Survey National Water Quality Laboratory--Evaluation of alkaline persulfate digestion as an alternative to Kjeldahl digestion for determination of total and dissolved nitrogen and phosphorus in water: U.S. Geological Survey Water-Resources Investigations Report 03-4174, 33 p.

Boren, K. 2001. QuikChem® Method 10-115-01-4-E. TOTAL PHOSPHORUS IN MANUAL PERSULFATE DIGESTS. Latchat Instruments, 5600 Lindburgh Drive, Loveland CO 80539.

Page 13: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

Watershed Management Research SOP 030TP-TDP - Wet Chemistry

Version Date 02/18/11Page 13 of 14

Table 1. TP-TDP followed by Nitrite-Nitrate Wet Chemistry Reagent, Standard, and Spike Recipes, Holding Times, Container Specs, Disposal Method, Notes, and Tracking. Posted in wet chemistry laboratory.

Reagent # Name Prep Flask Chemical Ingredients Efficacy Storage

ContainerDisposal Method Notes

983 g DI water40 g Ammonium MolybdateStir 1 hr or until dissolved995 g DI water3 g Antimony Potassium Tartrate HemihydrateStir until dissolved72 ml Stock Antimony213 ml Stock Molybdate21 ml Concentrated Sulfuric AcidDilute to mark with DI water

4 Ascorbic Acid Solution

1 liter Volumetric

60 g Granular Ascorbic Acid, disslove completely on stir plate before adding: 1 g dodecyl sulfate (sodium lauryl sulfate)

1 week square glass sink

This solution needs to be degassed with helium until it clears. Store in

refrigerator. Discard if the solution becomes yellow

500 ml DI water7.2 ml Concentrated Sulfuric Aciddilute to mark and invert to mix65 g NaOH 6 g Tetrasodium EDTA. Dissolve in 1 liter or 1kg of DI water

7Digest

Solution 1 for TP/TDP only

250 ml volumetric

15.25 g Potassium Persulfate dilute to mark and stir to dissolve.

2 months

same as prep flask

dilute sink

Store at room temperature. If precipitate forms stir under low heat until redissolved. To 5 ml

sample add 1 ml.

8

Digest Solution 2 for

TP/TDP-TAHP/DAHP

100 ml volumetric

50 ml Di Water, add 31 ml H2SO4, dilute to mark. To 5 ml sample add 100 ul for Both AHP and TP samples

2 months

same as prep flask

dilute sink

To 5 ml sample add 100 ul for Both AHP and TP

samples

Spike Sol 100 mg P l-1 250 ml

volumetric

0.1099 g primary standard grade anhydrous potassium phosphate monobasic; dried for 1 hr at 105 oC dissolve and dilute to mark

1 month square nalgene sink

10 ul of this solution to 6 ml sample is added for matrix spike; results in

166.6 ppb (ug l-1) spike.0.1099 g primary standard grade anhydrous potassium phosphate monobasic; dried for 1 hr at 105 oCdissolve and dilute to mark10 ml of Stock 1 into 1 literdissolve and dilute to mark80 ml of Stock 2 into 100 mldilute to mark and invert to mix40 ml of Stock 2 into 100 mldilute to mark and invert to mix20 ml of Stock 2 into 100 mldilute to mark and invert to mix10 ml of Stock 2 into 100 mldilute to mark and invert to mix4 ml of Stock 2 into 100 mldilute to mark and invert to mixDI water onlydilute to mark and invert to mix

sink

sink

sink

sink

Method 10-115-01-1-B/4-E: Orthophosphate/Total Phosphate by Ascorbic Acid (Reagents used for DRP, TP, and/or TDP)

5Carrier:

Sulfuric Acid 0.13 M

1 liter volumetric 2 week square

nalgenedilute sink

store in refrigerator. Make 2 liters.

Stock 2 250 ug P l-1 1 liter Volumetric 1 month square

nalgene

25 mg P l-1 1 liter Volumetric

Store in refrigerator.

Store in refrigerator.

contract

Store in refrigerator.

STDF PO4-StdE-0.0 ug P l-1

100 ml volumetric 2 week

square nalgene

Store in refrigerator.

STDE PO4-StdE-10 ug P l-1

100 ml volumetric

Store in refrigerator.STDD PO4-StdD-25 ug P l-1

100 ml volumetric 2 week

STDB square nalgene

STDC PO4-StdC-50 ug P l-1

100 ml volumetric 2 week

100 ml volumetric 2 week

square nalgene

Store in refrigerator.

square nalgene Store in refrigerator

2

STDA PO4-StdA-200 ug P l-1

100 ml volumetric 2 week square

nalgene

Store in refrigerator.

3 Molybdate Color Reagent

1 Liter Volumetric 1 week

1Stock

Ammonium Molybdate

1 Liter Nalgene

2 months

same as prep

Store in plastic and refrigerate

Stock Antimony Potassium

Tartrate

1 Liter Boston Amber

2 months

same as prep

Store in a dark bottle and refrigerate

sink

contract

2 month

sink

sink

sink

2 week square nalgene

square nalgene

square nalgene

1 month sink

square nalgene

dilute sink

Used as rinse to manifold tubing post analysis

Digest Solutions for TP, TDP, TAHP, and/or DAHP when analyzed separately; i.e., when not using the Dual Method.

6 NaOH-EDTA Rinse

1 liter Nalgene

Store in refrigerator.

Last Made

Working Standards and Spike Solution for low sensitivity DRP only (use combined standard (see Nitrate-Nitrite chart) when also analyzing for NO2-3 and NH4).

Stock 1

PO4-StdB-100 ug P l-1

Page 14: Standard Operation Procedure (SOP) for …gce-lter.marsci.uga.edu/public/uploads/ESF-SOP-030-2011... · Web view9.17. Place standards and samples in the autosampler. Input the schedule

Watershed Management Research SOP 030TP-TDP - Wet Chemistry

Version Date 02/18/11Page 14 of 14

Appendix A. Analytical Schedule Worksheets for 1)TP-TDP analyses.

Sample# Collection Bottle Analytical Base Sample-ID Cup # Auto-Tray-ID Digest # Di-Tray-ID

standard 07/18/2006-QAC-STD-A-1-DI-CAL 76 TDN-TN 1 TDN-TN-STDS/QCstandard 07/18/2006-QAC-STD-B-1-DI-CAL 77 TDN-TN 2 TDN-TN-STDS/QCstandard 07/18/2006-QAC-STD-C-1-DI-CAL 78 TDN-TN 3 TDN-TN-STDS/QCstandard 07/18/2006-QAC-STD-D-1-DI-CAL 79 TDN-TN 4 TDN-TN-STDS/QCstandard 07/18/2006-QAC-STD-E-1-DI-CAL 80 TDN-TN 5 TDN-TN-STDS/QCstandard 07/18/2006-QAC-STD-F-1-DI-CAL 81 TDN-TN 6 TDN-TN-STDS/QCstandard 07/18/2006-QAC-STD-G-1-DI-CAL 82 TDN-TN 7 TDN-TN-STDS/QCstandard 07/18/2006-QAC-STD-H-1-DI-CAL 83 TDN-TN 8 TDN-TN-STDS/QC

QC 07/18/2006-QAC-BLK-A-1-DI-LAB 61 TDN-TN 9 TDN-TN-STDS/QCQC 07/18/2006-QAC-BLK-B-1-DI-LAB 62 TDN-TN 10 TDN-TN-STDS/QC

Plastic Tube QC 07/18/2006-QAC-BLK-A-1-DI-ESF 63 TDN-TN 11 TDN-TN-STDS/QCPlastic Tube QC 07/18/2006-QAC-BLK-B-1-DI-ESF 64 TDN-TN 12 TDN-TN-STDS/QC

1 Plastic Tube water column 07/18/2006-ERS-0-C-1-SW-UKN 1 TDN 1 TDN-TN-12 Plastic Tube water column 07/18/2006-ERS-0-C-2-SW-UKN 2 TDN 2 TDN-TN-13 Plastic Tube water column 07/18/2006-ERS-0-C-3-SW-UKN 3 TDN 3 TDN-TN-14 Plastic Tube water column 07/18/2006-E04-9.6-C-1-SW-UKN 4 TDN 4 TDN-TN-15 Plastic Tube water column 07/18/2006-E04-9.6-C-2-SW-UKN 5 TDN 5 TDN-TN-16 Plastic Tube water column 07/18/2006-E04-9.6-C-3-SW-UKN 6 TDN 6 TDN-TN-17 Plastic Tube water column 07/18/2006-E02-9.6-C-1-SW-UKN 7 TDN 7 TDN-TN-18 Plastic Tube water column 07/18/2006-E02-9.6-C-2-SW-UKN 8 TDN 8 TDN-TN-19 Plastic Tube water column 07/18/2006-E02-9.6-C-3-SW-UKN 9 TDN 9 TDN-TN-1

10 IG-Sampler porosity 07/18/2006-E01-C13-R-1-IC-UKN 10 TDN 10 TDN-TN-1QC 07/18/2006-E01-C13-R-1-IC-DUP 10 TDNQC 07/18/2006-QAC-STD-D-1-DI-CHK 85 TDN 19 TDN-TN-STDS/QC

11 sed/biota 07/18/2006-E01-C02-L-1-IC-UKN 11 TDN 11 TDN-TN-1QC 07/18/2006-E01-C02-L-1-IC-SPK 38 TDN 26 TDN-TN-STDS/QC

12 IG-Sampler sed/biota 07/18/2006-E01-C09-C-1-IC-UKN 12 TDN 12 TDN-TN-113 IG-Sampler porosity 07/18/2006-E02-C13-R-1-IC-UKN 13 TDN 13 TDN-TN-114 IG-Sampler sed/biota 07/18/2006-E02-C02-L-1-IC-UKN 14 TDN 14 TDN-TN-115 IG-Sampler sed/biota 07/18/2006-E02-C09-C-1-IC-UKN 15 TDN 15 TDN-TN-116 IG-Sampler porosity 07/18/2006-E03-C13-R-1-IC-UKN 16 TDN 16 TDN-TN-117 IG-Sampler sed/biota 07/18/2006-E03-C02-L-1-IC-UKN 17 TDN 17 TDN-TN-118 IG-Sampler sed/biota 07/18/2006-E03-C09-C-1-IC-UKN 18 TDN 18 TDN-TN-119 IG-Sampler porosity 07/18/2006-E04-C13-R-1-IC-UKN 19 TDN 19 TDN-TN-120 IG-Sampler sed/biota 07/18/2006-E04-C02-L-1-IC-UKN 20 TDN 20 TDN-TN-1

QC 07/18/2006-E04-C02-L-1-IC-DUP 20 TDNQC 07/18/2006-QAC-STD-D-1-DI-CHK 85 TDN SAME

21 IG-Sampler sed/biota 07/18/2006-E04-C09-C-1-IC-UKN 21 TDN 21 TDN-TN-1QC 07/18/2006-E04-C09-C-1-IC-SPK 39 TDN 27 TDN-TN-STDS/QC

22 IG-Sampler porosity 07/18/2006-E05-C13-R-1-IC-UKN 22 TDN 22 TDN-TN-123 IG-Sampler sed/biota 07/18/2006-E05-C02-L-1-IC-UKN 23 TDN 23 TDN-TN-124 IG-Sampler sed/biota 07/18/2006-E05-C09-C-1-IC-UKN 24 TDN 24 TDN-TN-125 IG-Sampler porosity 07/18/2006-E06-C13-R-1-IC-UKN 25 TDN 25 TDN-TN-126 IG-Sampler sed/biota 07/18/2006-E06-C13-R-1-IC-UKN 26 TDN 26 TDN-TN-127 IG-Sampler sed/biota 07/18/2006-E06-C09-C-1-IC-UKN 27 TDN 27 TDN-TN-128 IG-Sampler porosity 07/18/2006-E07-C13-R-1-IC-UKN 28 TDN 28 TDN-TN-129 IG-Sampler sed/biota 07/18/2006-E07-C02-L-1-IC-UKN 29 TDN 29 TDN-TN-130 IG-Sampler sed/biota 07/18/2006-E07-C09-C-1-IC-UKN 30 TDN 30 TDN-TN-1

QC 07/18/2006-E07-C09-C-1-IC-DUP 30 TDN31 IG-Sampler porosity 07/18/2006-E08-C13-R-1-IC-UKN 31 TDN 31 TDN-TN-1

QC 07/18/2006-E08-C13-R-1-IC-SPK 40 TDN 28 TDN-TN-STDS/QCQC 07/18/2006-QAC-STD-D-1-DI-CHK 84 TDN SAME

32 IG-Sampler sed/biota 07/18/2006-E08-C02-L-1-IC-UKN 32 TDN 32 TDN-TN-133 IG-Sampler sed/biota 07/18/2006-E08-C09-C-1-IC-UKN 33 TDN 33 TDN-TN-1

QC 07/18/2006-QAC-REC-A-1-DI-ATP 73 TDN 20 TDN-TN-STDS/QCQC 07/18/2006-QAC-REC-B-1-DI-ATP 74 TDN 21 TDN-TN-STDS/QCQC 07/18/2006-QAC-REC-C-1-DI-ATP 75 TDN 22 TDN-TN-STDS/QCQC 07/18/2006-QAC-REC-A-1-DI-GUP 87 TDN 23 TDN-TN-STDS/QCQC 07/18/2006-QAC-REC-B-1-DI-GUP 88 TDN 24 TDN-TN-STDS/QCQC 07/18/2006-QAC-REC-C-1-DI-GUP 89 TDN 25 TDN-TN-STDS/QC

Plastic Tube QC 07/18/2006-QAC-BLK-A-1-DI-FLD 65 TDN 13 TDN-TN-STDS/QCPlastic Tube QC 07/18/2006-QAC-BLK-B-1-DI-FLD 66 TDN 14 TDN-TN-STDS/QC

QC 07/18/2006-QAC-BLK-A-1-DI-LAB 67 TN SAMEQC 07/18/2006-QAC-BLK-B-1-DI-LAB 68 TN SAME

Plastic Tube QC 07/18/2006-QAC-BLK-A-1-DI-ESF 69 TN 15 TDN-TN-STDS/QCPlastic Tube QC 07/18/2006-QAC-BLK-B-1-DI-ESF 70 TN 16 TDN-TN-STDS/QC

1 Plastic Tube water column 07/18/2006-ERS-0-C-1-SW-UKN 28 TN 28 TDN-TN-12 Plastic Tube water column 07/18/2006-ERS-0-C-2-SW-UKN 29 TN 29 TDN-TN-13 Plastic Tube water column 07/18/2006-ERS-0-C-3-SW-UKN 30 TN 30 TDN-TN-1

QC 07/18/2006-ERS-0-C-3-SW-DUP 30 TN4 Plastic Tube water column 07/18/2006-E02-9.6-C-1-SW-UKN 31 TN 31 TDN-TN-1

QC 07/18/2006-E02-9.6-C-1-SW-SPK 41 TN 29 TDN-TN-STDS/QC5 Plastic Tube water column 07/18/2006-E02-9.6-C-2-SW-UKN 32 TN 32 TDN-TN-16 Plastic Tube water column 07/18/2006-E02-9.6-C-3-SW-UKN 33 TN 33 TDN-TN-17 Plastic Tube water column 07/18/2006-E04-9.6-C-1-SW-UKN 34 TN 34 TDN-TN-18 Plastic Tube water column 07/18/2006-E04-9.6-C-2-SW-UKN 35 TN 35 TDN-TN-19 Plastic Tube water column 07/18/2006-E04-9.6-C-3-SW-UKN 36 TN 36 TDN-TN-1

QC 07/18/2006-QAC-REC-A-1-DI-GUP 87 TN SAMEQC 07/18/2006-QAC-REC-B-1-DI-GUP 88 TN SAMEQC 07/18/2006-QAC-REC-C-1-DI-GUP 89 TN SAME

Plastic Tube QC 07/18/2006-QAC-BLK-A-1-DI-FLD 71 TN 17 TDN-TN-STDS/QCPlastic Tube QC 07/18/2006-QAC-BLK-B-1-DI-FLD 72 TN 18 TDN-TN-STDS/QC

red=blanks, blue=duplicates, purple=calibration standards, green=check standards, orange=recovery standards, pink= spikes.

Nutrient Wet Chemistry Analytical Schedule for Sampling Event-07/18/2006. QC at 51% for Digested Analytes.