spectroscopic measurement (intro, lit. review, conc, ref)
TRANSCRIPT
1
ABSTRACT
The total transition rate summed over all decay channels can be measured
either through frequency-resolved studies of the level width, or through
time-resolved studies of the level lifetime. In order to determine the natural
line width in a field-free spectroscopic measurement, either the lifetime must
be very short, or the Doppler, pressure, and instrumental broadenings must
be made very small. Line widths have been determined using Fabry-Perot
spectrometry at very low temperatures and pressures, and in beam-foil
studies of radioactive transitions in which the lower level decays very
rapidly via auto-ionisation. The line-width can also be determined through
the use of the phase shift method. Here modulated excitation is applied to
the source, thus producing similarly modulated emitted radiation, and the
width can be specified from the phase shift between the two signals. Other
methods involve resonance fluorescence techniques, where sub-Doppler
widths are obtained because the width of the exciting radiation selects a
subset of particle motions within the sample. Resonance fluorescence
techniques that can be used to determine level widths include zero-field level
crossing (the Hanle effect), high-field level crossing, and double optical
resonance methods, but the Hanle effect is the most common.
2
INTRODUCTION
TIME-RESOLVED DECAY MEASUREMENTS
The most direct method for the experimental determination of level lifetimes
is through the time-resolved measurement of free decay of the fluorescence
radiation following a cut-off of the source of excitation. An important factor
limiting the accuracy is the repopulation of the level of interest by cascade
transitions from higher-lying levels. For this reason, decay curve
measurements fall into two classes: those that involve selective excitation of
the level of interest, thus eliminating cascading altogether; and those that use
correlations between cascades connected decays to account for the effects of
cascades.
SELECTIVE EXCITATION
Lifetime measurements accurate to within a few parts in 103 have been
obtained through selective excitation produced when appropriately tuned
laser light is incident on a gas cell or a thermal beam, or on a fast ion beam.
With a gas cell or thermal beam, the timing is obtained by a pulsed laser and
delayed coincidence detection. With a fast ion beam, time-of-flight methods
are used. In either case, after removal of the background, the decay curve of
intensity vs. time is a single exponential, and the lifetime is obtained from its
semi logarithmic slope. Laser-excited time-of-flight studies were first
3
carried out by observing the optical decay of the ion in light following
excitation using a laser beam, which crossed the ion beam. In these studies,
the laser light was tuned to the frequency of the desired absorption transition
either through the use of a tunable dye laser, or by varying the angle of
intersection to exploit the Doppler effect. Recent measurements have
utilised diode laser excitation in this geometry.
A number of adaptations of this technique have been developed in which the
laser and ion beams are made to be collinear, and are switched into and out
of resonance within a segment of the beam by use of the Doppler effect.
The collinear geometry can provide a longer excitation region and less
scattering of laser light into the detector than occurs in the crossed beam
geometry. In one adaptation, excitation occurs within an electrostatic
velocity switch, and the time resolution is obtained by physically moving the
velocity switch. In another adaptation the ion beam is accelerated with a
spatially varying voltage ramp. The resonance region is moved relative to a
fixed detector by time-sweeping either the laser tuning or the ramp voltage.
While these selective excitation methods totally eliminate the effects of
cascade repopulation, they are generally limited to levels in neutral and
singly ionised atoms that can be accessed from the ground state by strongly
absorptive E1 transitions, and the selectivity itself is a limitation. Many very
4
precise measurements have been made by these techniques, but they have
primarily involved ¨Q = 0 resonance transitions in neutral alkali atoms and
singly ionised alkali-like ions.
NONSELECTIVE EXCITATION
Much more general access can be obtained by non-selective excitation
methods, such as pulsed electron beam bombardment of a gas cell or gas jet,
or in-fight excitation of a fast ion beam by a thin foil. Pulsed electron beam
excitation can be achieved either through use of a suppressor grid, or by
repetitive high frequency deflection of the beam across a slit so as to chop
the beam. Particularly in the case of weak lines, the high frequency
deflection technique offers the advantages of high current and sharp cut-off
times. The high currents yield high light levels, so that high-resolution
spectroscopic methods can be used to eliminate the effects of line blending.
Pulsed electron excitation methods are well suited to measurements in
neutral and near neutral ions (although for very long lifetimes in ionised
species, the decay curves can be distorted if particles escape from the
viewing volume through the Coulomb explosion effect).
However, for highly ionised atoms, the only generally applicable method is
thin foil excitation of a fast ion beam. Nonselective excitation techniques
can also be applied to measurements such as the phase shift method and the
5
Hanle effect, in which case cascade repopulation also can become a serious
problem. However, most of the attempts to eliminate or account for cascade
effects have occurred in decay curve studies.
ANDC METHOD
Precision lifetime values have been extracted from cascade-affected decay
curves by a technique known as the arbitrarily normalized decay curve
(ANDC) method, which exploits dynamical correlations among the cascade-
related decay curves. These correlations arise from the rate equation that
connects the population of a given level to those of the levels that cascade
directly into it. The instantaneous population of each level n is, to within
constant factors �n involving the transition probabilities and detection
efficiencies, proportional to the intensity of radiation In (t) emitted in any
convenient decay branch. In terms of these intensities, the population
equation for the level n can be written in terms of its direct cascades from
levels i as
Thus, if all decay curves are measured at the same discrete intervals of time
of time tp, the population equation provides a separate independent linear
relationship among these decay curves foe each value of tp, with common
6
constant coefficients given by the lifetime 2n and the normalization
parameters �i. Although the sum over cascades is formally unbounded, the
dominant effects of cascading from highly excited states are often accounted
for by indirect cascading through the lower states, in which case the sum can
be truncated after only a few terms. ANDC analysis consists of using this
equation to relate the measured Ik(tp) (using numerical differentiation or
integration) to determine 2k and the �i through a linear regression. If all
significant direct cascades have been included, the goodness-of-fit will be
uniform for all time sub-regions, indicating reliability. If important cascades
have been omitted or blends are present, the fit will vary over time sub
regions, indicating a failure of the analysis. Very rugged algorithms have
been developed that permit accurate lifetimes to be extracted even in cases
where statistical fluctuations are substantial, and studies of the propagation
and correlation of errors have been made. Clearly the ANDC method is most
easily applied to systems for which repopulation effects are dominated by a
small number of cascade channels.
There are three rather different ways of measuring lifetimes of excited states:
delay time, beam foil spectroscopy, and Hanle effect. None of these
methods is universally applicable. Some work only for states that can be
7
excited directly from the ground state, others are no good if the lifetime is
too short (strong line) or the intensity too low (weak line) for accurate
measurement. Amongst them, however, they cover a very wide range of
transitions, both atomic and molecular, and they have in common the one
immensely important advantage that there is no need to know the population
of the initial level. It must be remembered that a lifetime does not yield
directly an oscillator strength, unless there is only one allowed transition
from the excited state. However, if relative values of A2j can be obtained
from emission, they can be put on an absolute scale by a measurement of 2.
(where A2j is the Einstein coefficient for spontaneous emission and 2 is the
time constant).
In the description of the delay methods that follows (and also in that of the
Hanle effect) reference is made to excitation by resonance radiation. It is
now becoming possible to use tuneable dye laser as an exciting source. Its
high intensity makes it ideal for selective population of excited levels, and
its short pulse duration allows good time resolution. New ways of
exploiting these advantages are constantly being devised at the present time.
8
1. DELAY METHODS
The delayed coincidence method is shown schematically below (Figure 1).
The gas is excited by a pulsed electron beam (or occasionally by a pulse of
resonance radiation), which also triggers off the time-to-pulse-height
converter. The stop signal for this is provided by the emitted radiation, via a
monochromator and photo-multiplier, so that the pulse height is proportional
to the time between the excitation and decay of the relevant excited level.
Building up statistics for dN/dt as a function of t should result in an
exponential of time constant 2.
Figure 1. Schematic arrangement for measurement of lifetimes by
delayed coincidence
9
A variant of this is to use a coincidence counter in place of the time-to-
pulse-height converter and to record coincidences between the exciting
electron pulse and the emitted radiation as a function of a variable delay time
t introduced to the signal from the electron pulse. In yet another variant,
shown schematically below (Figure 2), the light pulses from transitions into
and out of a given level are fed into a coincidence counter and the number of
coincidences recorded as a function of the delay time introduced into the
signal from the first transition. In practice it is usually difficult to find a
suitable pair of transitions to which to apply this variation.
The phase shift method uses modulated instead of pulsed excitation, usually
in the form of resonance radiation but sometimes as an electron beam, to
excite the relevant level.
Figure 2. Delayed coincidence method using cascading transitions. The
delay between the emission of photons hv1 and hv2 measures the lifetime
of E2.
10
The emitted radiation is then also modulated, but with a phase shift θ
relative to the exciting radiation which depends on the lifetime of the excited
state according to tanθ = 2πvm2, where vm is the modulating frequency. The
figure below (Figure 3) shows the method schematically. The modulation
may be done with a Kerr cell or with a standing ultrasonic wave. The
accuracy can be improved by modulating the voltage on the photo-multiplier
cathode so that θ is measured at the beat frequency of a few kilocycles rather
than at the modulation frequency, which has to be in the megacycle range.
The delay methods can be used for lifetimes from a few microseconds to a
few nanoseconds. Apart from difficulties associated with scattered light and
insufficient resolution (they are usually used with broad band-pass
monochromators or filters to obtain enough intensity in the fluorescent
light), they have two principal potential sources of error. The first is
cascading: unless the relevant level is excited by resonance radiation or by a
tuned dye laser, a number of higher levels may be excited at the same time
and subsequently decay to the level under investigation. The latter is
therefore re-populated from above, so that its apparent lifetime is prolonged.
In principle, even with electron excitation it should be possible to excite only
the required level, but in practice the cross-section near the threshold energy
is so small that considerably higher electron energies have to be used. The
11
second principal difficulty affects all levels emitting to the ground or a
metastable state and is usually known as imprisonment of radiation.
Figure 3. Schematic arrangement for measurement of lifetime by phase
shift.
If the gas is not optically thin, a number of photons will be absorbed and re-
emitted one or more times before they eventually get out, and the effect is,
again, to prolong the apparent lifetime of the excited state. Measuring
apparent lifetime as a function of pressure can check for the effect. A
further source of error that may affect the longer-lived states is collisional
depopulation; this of course shortens the apparent lifetime. Both these last
12
troubles may be avoided by going to sufficiently low pressures (atomic
beams have been used for this reason), but one may then run into difficulties
with the low light intensities. The accuracy claimed for lifetime
measurements is usually around 10%, is often better, but may in fact be
considerably worse if systematic errors have not been eliminated.
2. BEAM FOIL METHOD
A beam of ions of various degrees of ionisation emerges from the thin foil in
different states of excitation. The excited states decay as the beam travels
downstream from the foil, and the rate of decrease of intensity of any
particular line as a function of distance from the foil gives directly the
lifetime of the relevant excited state.
Historically, this method is the successor to the experiments of Wien in
1920s with canal rays, in which the lifetime was measured from the decay of
emitted radiation as the excited ions in a discharge tube travelled beyond the
cathode. In practice, in the beam foil method, the spectrometer is kept fixed,
and the intensity of the required lines is measured while the foil is moved
upstream. It is necessary to monitor the constancy of the beam while this is
going on, either by measuring the total charge collected at its end or by using
a second photo-multiplier.
13
Figure 4. Reason for Doppler broadening in beam foil spectroscopy.
The excited ions have a velocity component in the R-direction of νν sin θθ
away from the spectrometer, giving a red-shift. For the B-direction
there is an equal blue shift.
The principal difficulties of the method are cascading from higher excited
states, low light intensity, and large Doppler broadening. The reason for this
last is illustrated above (Figure 4). The velocity of the beam is of order
5x108 cm sec
-1, and because of the finite acceptance angle of the
spectrometer the direction of observation is not strictly perpendicular to the
beam direction. For light travelling along ray B the ions have a component
of velocity along B of v sin θ, resulting in a blue shift, whereas for ray R the
velocity is v sinθ away from the spectrometer, giving a red shift. Since all
rays in the cone RB contribute to the image, the result is a broadened line.
14
The low intensity makes a wide aperture essential, and the Doppler
broadening may be as high as
On the other hand, the density in the beam is so low that there are no
difficulties with imprisonment of radiation or collisional depopulation.
More importantly, this is the only method other than that of emission line
intensity, which can be applied to ionic spectra. It is best suited for lifetimes
of about 10-8
sec, during which time the ions travel a few cm. With shorter
lifetimes the decay is too fast to measure accurately, bearing in mind that it
is usually necessary to accept light from an appreciable length of beam ∆l to
obtain sufficient intensity, so that the space resolution is rather poor. If the
lifetime is very long, so few ions decay in the path ∆l that inadequate
intensity again becomes a problem.
SUMMARY
In beam-foil studies, the excitation is created in the dense solid environment
of the foil, after which the ions emerge into a field free, collision free, high
vacuum region downstream from the foil. A time-resolved decay curve is
obtained by translating the foil upstream or downstream relative to the
detection apparatus. The beam is a very tenuous plasma, which has both
15
advantages and disadvantages. The low density avoids the effects of
collisional de-excitation and radiation trapping, but also produces relatively
low light levels. This requires fast optical systems with a corresponding
reduction in wavelength dispersion, and care must be taken to avoid
blending of these Doppler broadened lines. Methods have been developed
by which grating monochromators can be refocused to a moving light
source, thus utilizing the angular dependence of the Doppler effect to narrow
and enhance the lines.
3. HANLE EFFECT
This technique again is a resurrection from the 1920s, when the effect,
investigated by Hanle, was known as magnetic depolarisation of resonance
radiation. It is now also known as zero-field level crossing, being a special
case of the level-crossing experiments for the measurement of hyperfine
structure. For a very brief qualitative description of the effect it is simplest
to use a semi-classical model, but when there is hyperfine structure or close
fine structure in the excited level a proper quantum-mechanical treatment is
required. A typical experimental layout is shown below (Figure 5). Light
from the source of resonance radiation travelling in the y-direction is plane-
polarized in the x-direction before entering the resonance vessel where it is
absorbed and re-emitted. The emitted radiation has the same polarization as
16
the exciting radiation, and a photo-multiplier on the x-axis therefore records
no signal. If now a magnetic field is applied to the resonance vessel in the z-
direction, the resonance radiation is partly de-polarized, and a signal is
registered, rising at strong field to half the intensity in the absence of the
polariser.
Figure 5. Schematic arrangement for Hanle effect. The photomultiplier
in the x-direction registers a signal where field B is applied in the z-
direction.
This is because the direction of oscillation of the electrons, originally
parallel to the x-axis, now precesses about the z-axis. If the precession is
sufficiently rapid, half the oscillations may be considered as in the y-
direction, thus radiating in the x-direction. The lifetime enters into the story
17
because if the precession is not very fast some of the atoms will have
decayed before the precessional cycle is completed, so that the intensity
component in the y-direction is less than the original x-component.
Figure 6. Classical interpretation of Hanle effect. (a) represents small
damping (ωωB>1/ττ) and (b) represents large damping �&Ba��ττ). The
length of each spoke in these figures is proportional to the intensity of
the radiation at the corresponding direction of polarization.
The signal thus depends on the relative magnitudes of the precessional and
decay times, as shown schematically above (Figure 6): (a) represents the
situation when the precessional frequency ωB is large compared to the
damping constant γ = 1/τ and (b) shows them comparable. Simply by
treating the system as an oscillator precessing with angular frequency ωB and
decaying amplitude exp (- ½ γ / t) it can be shown that the signal is given by
18
Figure 7. Signal as a function of magnetic field in Hanle effect. The
curve is an inverted Lorentzian, and the field at the half-value points,
B1/2, is given by B1/2 = Y /2µµB 1/gJττ.
where the Larmor precessional frequency ωB = gJµBB/Y (µB is the Bohr
magneton and gJ the Lande g-factor). This expression for I has the inverted
Lorentzian form shown above (Figure 7). The signal has half its maximum
value when
19
2 can therefore be determined from the half-width of the signal-versus-field
curve if gJ is known.
Looked at from the quantum-mechanical point of view, all the magnetic sub-
levels of the upper state are populated by the incident radiation in zero field,
since they are then degenerate. The emitted radiation from these sub-levels
is coherent and interferes destructively in the x-direction. To lift the
degeneracy it is necessary to separate the sub-levels by an amount ∆EB
greater than their natural width ∆Enat ~ ��τ leading to gJµB%!���τ; this is
fulfilled beyond the half-width point. However, any hyperfine structure or
close fine structure within this energy range must also be brought into
account. Further degeneracies, with consequent changes of polarization,
occur whenever the field is such that the energies of two hyperfine
components overlap, as already seen in level-crossing spectroscopy. The
Hanle effect is concerned primarily with the zero-field degeneracies, but
because the field must be raised to 10-3
– 10-2
Wb m-2
in order to establish
the width these other degeneracies may have to be taken into account.
20
In contrast to the other lifetime techniques, this one has no cascading
problems. The ‘line width’ is determined entirely by the lifetime, with no
Doppler or power broadening. Apart from limitations on the possible
transitions to which the method can be applied, the main difficulties are
associated with self-absorption in the original resonance source and
coherence narrowing of the ‘line’, resulting from coherent trapping of the
radiation.
SUMMARY
In its most commonly used form, the Hanle effect makes use of polarized
resonance radiation to excite atoms in the presence of a known variable
magnetic field. The magnetic substates of the sample are anisotropically
excited, and the subsequent radiation possesses a preferred angular
distribution. By applying the magnetic field in a direction perpendicular to
the anisotropy, the angular distribution is made to precess, producing
oscillations in the radiation observed at a fixed angle. At infinite
precessional frequency the intensity would be proportional to the
instantaneous average angular intensity, by at finite precessional frequency it
depends upon the decay that has occurred during each quarter rotation.
Measured as a function of magnetic field, the emitted intensity has a
21
Lorentzian shape centered about zero field with a width that depends on the
lifetime and g-factor of the level.
22
REFERENCES
1. Thorne A. P. (1974) Experimental determination of transition
probabilities, Spectrophysics, Chapman and Hall & Science Paperbacks,
London, 332.
2. Curtis L. J. Lifetimes: In Part B, 17, Precision Oscillator Strength and
Lifetime Measurements, 262,
http://www.physics.utoledo.edu/~ljc/drake17.pdf, Department of Physics
and Astronomy, University of Toledo, Toledo, Ohio 43606.
23
BIBLIOGRAPHY
1. (Lifetimes) Corney, A. The Measurement of Lifetimes of Free Atoms,
Molecules and Ions, Adv. In electronics and Electron Phys. 29, 115,1970
2. (Beam foil) Bickel, W. S. Mean Life Measurements using Beam Foil
Light Source Appl. Optics 7,2367,1968.
3. (Hanle) De Zafra, R. L. and Kirk, W. Measurement of Atomic Lifetimes
by the Hanle Effect, Amer. J. Phys. 35, 573, 1967.
4. Curtis L.J. (2003) Atomic Structure and lifetimes: A Conceptual
approach, Cambridge Univ. Press, Cambridge.
5. Thorne A., Litzen U., Johansson S. (1999) Spectrophysics: Principles
and Applications, Springer, Heidelberg.
6. Andra H. J. (1976) Beam-Foil spectroscopy, Vol. 2, ed., By Sellin I. A.,
Pegg D. J., Plenum, New York, p. 835.
7. Curtis L.J. (1976) Beam-Foil Spectroscopy, Heidelberg, ed., By S.
Bashkin, Springer, Berlin, Heidelberg, p. 3
8. Bashkin, S. (1968) ‘Beam Foil Spectroscopy’, ed., Gordon and Breach.
9. Jin J., Church D. A. (1993) Phys. Rev. 47, 132