some, novel analytical atom cells and detectors for …

254
SOME, NOVEL ANALYTICAL ATOM CELLS AND DETECTORS FOR ATOMIC SP7CTROSCOPI AT LOW 1:A7E=GTHS by Michael James Adams, Grad. R.I.C. A Thesis submitted for the Degree of DOCTOR OF PHILOSOPHY of the University of London September 1975 Chemistry Department, Imperial College of Science and Technology, London, S.4T.7.

Upload: others

Post on 30-Jan-2022

1 views

Category:

Documents


0 download

TRANSCRIPT

SOME, NOVEL ANALYTICAL ATOM CELLS AND DETECTORS FOR

ATOMIC SP7CTROSCOPI AT LOW 1:A7E=GTHS

by

Michael James Adams, Grad. R.I.C.

A Thesis submitted for the Degree of

DOCTOR OF PHILOSOPHY

of the University of London

September 1975

Chemistry Department,

Imperial College of Science and Technology,

London, S.4T.7.

ABSTRACT

The construction and operation of small, graphite tube and

rod atomisers for AAS at wavelengths below 200nm is described.

The use of these electrothermal furnaces for the direct

determination of iodine and sulphur using their vacuum ultra-

violet resonance lines is described also.

The molecular absorption spectra of

some commonly occurring inorganic salts vaporised in non-flame

cells are presented and shown to be similar to the spectra

observed with cool-flame cells.

The temperature-time profiles of a

commercial graphite furnace and the atomic vapour produced in

this cell are discussed.

Finally, a preliminary evaluation is

undertaken of the use of photoionisation chambers as detectors

for non-dispersive atomic spectrometry in the vacuum ultraviolet.

AC K NOWLED EdENTS

The work presented in this thesis was undertaken in

the Chemistry Department at Imperial College of Science and

Technology between October 1972 and September 1975. Except

where due reference is made it is entirely original, and no

part has been submitted for any other degree.

I would like to thank my supervisors,

Dr. G.F.Kirkbright and Prof.T.S.West, for their advice and

encouragement throughout the course of this work. I would

also like to thank my fellow researchers in the Analytical

Department for many useful discussions and suggestions.

I am indebted to the Science Research

Council and the British Steel Corporation for their support

of this work under the C.A.P.S. scheme, and the Paul Instrument

Fund Committee for much of the apparatus employed in the research.

Finally, I wish to thank my wife Janet

for her invaluable assistance in the preparation of this

thesis — completed despite the attempted 'aid' of 'Merlin'.

CONTENTS

Pa e

Abstract 1

Acknowledgements 2

Contents

3

CHAPTER ONE, INTRODUCTION

6

1.1 The Discovery Of The Vacuum Ultraviolet

7

Region Of The Spectrum.

1.2 Emission Sources For The Far Ultraviolet. 11

1.3 Window Materials And Spectral Filters. 23

1.4 Dispersion Elements. 27

1.5 Radiation Detectors. 28

1.6 Low Wavelength Analytical Spectrometry. 35

CHAPTER TWO,NON-FLAI'E ELECTROTHERMAL ATOMI=S 42

FOR AAS IN TiE FAR U.V.

2.1 Introduction. 43

2.2 Furnace Atomisers. 47

2.3 Filament Atomisers. 53

2.4 Miscellaneous Atomisers. 54

CHAPTER THREE, THE DIRECT Dr:TERMINATION OF

58 IODINE.

3.1 Introduction 59

3.2 Apparatus-D.C. Measurement Techniques. 6o

3.3 Iodine AAS Results. 70

3.4 Apparatus-Mechanical Modulation And A.C. 75 Measurement Techniques.

3.5 Iodine AAS Results. 80

3.6 The Effect Of Foreign Ions. 84

3

CHAPTER FOUR, THE DIRECT DETERMINATION OF

88 SULPHUR.

4.1 Introduction. 89

4.2 Apparatus-Electronic Modulation And A.C. 91 Measurement Techniques.

4.3 Sulphur AAS Results. 99

4.4 Discussion-Faults With The Simple Graphite 103 Tube Atomiser.

CHAPTER FIVE, THE DIRECT DETERMINATION OF IODINE 104 AND SULPHUR USING A NEW GRAPHITE

TUBE FURNACE AD A 1RAPHITE ROD.

5.1 Introduction.

5.2 The Design And Construction Of A New Graphite

Tube Atomiser.

5.3 A Mini-I"Iassmann Rod Atomiser.

5.4 A Demountable Hollow Cathode Lamp.

5.5 Iodine AAS.

5.6 Sulphur AAS.

5.7 Conclusion.

CHAPTER SIX, IODINE AAS AT 206.1nm AND NON-SPECIFIC

MOLECULAR ABSORPTION IN NON-FLAME:

ATOMISERS BY SIMPLE INORGANIC SALTS.

6.1 Introduction

6.2 Apparatus.

6.3 The Nature Of The Iodine Absorption Signals

Observed In The HGA 2000 At The 206.1nm Iodine Line.

105

105

114

116

118

126

148

151

152

152

154

6.4 Absorption By Other Common Salts In The HGA 2000. 158

6.5 Absorption Spectra Of Common Inorganic Salts In 162 The Small Graphite Furnace Atomiser.

6.6 Conclusion 166

4

CHAPIZR SEVEN, TEMPERATUR71-TI1E PROF= IN TEE; 168

HGA 2000 g:LAPHIT: =ACE AT O!

AND THE ATOMIC VAPOUR PRODUCED IN

THIS CELL.

7.1 Introduction. 169

7.2 Terminal Temperatures Attained By The 169

HGA 2000 Graphite Furnace.

7.3 Temperature-Time Profiles Of The HGA 2000. 171

7.4 Measurement Of Electronic Excitation Temperatures 174

Of Atomic Species By A Two-Line AAS Method.

7.5 The Electronic Excitation Temperature Of The 180

Atomic Vapour Of Ga And In In The HGA 2000.

7.6 Temperature-Time Profiles Of An Atomic Vapour 186

In The HGA 2000.

7.7 Discussion And Conclusion. 190

CHAPTER EIGHT, PHOTOIONISATION DETECTORS. 193

8.1 Introduction 194

8.2 Theory Of Photoionisation Detectors. 196

8.3 Construction Of A Photoionisation Detector. 206

8.4 Vacuum Ultraviolet Line Emission Sources. 208

8.5 Selecting The Filling Gas For A PID. 219

8.6 Filling The Photoionisation Detector. 223

8.7 The Spectral Response Characteristics Of 223

The Ethylamine PID.

8:8 The Effect Of Ethylamine Vapour Pressure And 228

Applied Voltage On The Response Of The PID.

8.9 Quantitative Analysis With The Ethylamine PID. 230

8.10 The Xylene PID 233

8.11 The N1:1-di-ethyl-p-toluidine PID. 233

8.12 Conclusions. 235

CHAPTER :TI1T:, CO:LLUSIOIIS. 237

9.1 Conclusion 238

9.2 Suggestions For Further ':ork. 243

5

ONE

1.1 The Discovery Of The Vacuum Ultraviolet Region Of The Spectrum.

The origin of modern spectroscopy may be attributed to

the work of Fraunhofer who, in the early part of the nineteenth

century, was the first to observe and appreciate the importance

of the dark lines, now bearing his name, contained in the sun's

spectrum (1). In 1860 Bunsen and Kirchoff (2) discovered that

many elements, when introduced into a flame, emitted radiation

characteristic of the element and were able to relate this emission

to the absorption lines reported by Fraunhofer.

Before Bunsen et al., however, the electromagnetic

spectrum was known to extend beyond the visible regions from

the research by Herschel (3) and Ritter (4) who had detected

infra—red and ultraviolet radiation respectively. Ritter, in

1801, had studied with the aid of a spectroscope, radiation of

wavelengths shorter than 400nm and was able to demonstrate a

chemical action caused by some energy form in this region; the

ultraviolet region. He discovered that silver chloride blackened

and was decomposed more rapidly by radiation in this region than

that in the visible part of the spectrum.

The study of spectroscopy progressed rapidly during

the nineteenth century and it was realised that the atoms of

many more elements could be made to emit by excitation in an

electrical discharge than in a flame. Using a spark source,

quartz optics and the recently discovered technique of photo-

graphic recording of spectra, Stokes (in 1862) was able to extend

the limit of the known ultraviolet region to 183.0nm, (5). The

next major advance into the far u.v. was by Schumann (6) who,

between 1885 and 1903, made three major discoveries. He was

the first to realise that the opacity of air to radiation of

wavelengths shorter than 190nm was due largely to the molecular

oxygen in the atmosphere, and constructed the first vacuum

spectrograph to overcome this. To detect the short wavelength

radiation he invented the Schumann photographic plate; containing

little gelatin binder and still in use today for recording far

u.v. spectra. With most of the u.v.-absorbing gelatin removed

from the plate it is sensitive throughout the extreme ultra-

violet. Finally,to extend the ultraviolet limit beyond that of

Stokes, Schumann used fluorite in place of quartz for the optical

components. These discoveries allowed Schumann to extend

the region of study to below 130nm and in his honour the region

125nm to 185nm is frequently referred to as the Schumann U.V.

The early vacuum spectrographs were fluorite

prism instruments and as the constants determining the dispersion

of radiation for fluorite were unknown Schumann was unable to

assign a wavelength scale to the spectra he obtained.

Lyman (7) constructed the first vacuum spectro-

graph using a grating as the dispersion element and was able to

fix a wavelength scale to the Schumann spectra. In 1914 he

discovered the hydrogen series now bearing his name. The major

proportion of Lyman's work was confined to helium, discovering

the He I series limit at 50.Lnm and its continuum beyond this

to ca. 23nm. Even shorter wavelengths were explored by Millikan(8).

Using a vacuum hot-spark source he detected atomic line emission

down to 14mm and discovered that the spectra of many highly ionised

atoms could be arranged in isoelectronic sequences; these findings

greatly advanced the acceptability of the Bohr-Sommerfield theory

of atomic line spectra.

8

The remaining gap between X-rays and extreme u.v. radiation

was finally closed in 1927 when Osgood (9) photographed atomic

emission lines down to 4.4nm and Dauvillier extended the X-ray

limit to 12.1nm. From 1925 to 1941 the study of atomic and mol-

ecular spectroscopy underwent rapid expansion; the optical

properties of the materials used in the vacuum u.v. were well

known and the importance of the photoelectric effect in monitoring

radiation was realised. This period of fundamental discoveries

in spectroscopy has been reviewed in detail by Boyce(10). Interest

in the extreme u.v. advanced further after 1945 with the development

of the new technologies of space astronomy, rocketry, high

temperature plasmas and solid-state phenomena.

The fundamental fact that the infra-red, visible, ultra-

violet and X-rays were all of the same basic nature had been

established. Because of the nature of the physical and chemical

effects they each produce, and because of the special techniques

required for their production, detection and measurement, it is

convenient to separate these forms of radiation into distinct

groups. Fig. 1.1 presents the more common classification in use

today for the ultraviolet. The division between middle and far

ultraviolet arises from the absorption of radiation of shorter

wavelengths than 200nm by the molecular oxygen and water vapour

in the atmosphere. Since the work of Schumann, lithium fluoride

(not a naturally occurring material) has been found to be trans-

parent down to 105nm. Below this limit no solid material, with

the exception of very thin films, has been discovered allowing

the transmission of radiation and the region is usually termed

the extreme ultraviolet. Studies in this region employ window-

less systems and special experimental techniques are required.

Near

U.V.

Middle

U.V.

Far U.V.

Vacuum U.V.

100 200

Wavelength, ( nm.)

1 400 300

ULTRA - VIOLET

Extreme U.V.

Soft X

rays

Lyman

region

Schumann

region

1.2 Emission Sources For The Far Ultraviolet.

At the beginning of this century the most popular

source for vacuum ultraviolet radiation emission was the spark

source. Indeed, Lyman, in the second edition of his now classic

monograph on vacuum spectroscopy (11), provides the atomic emission

wavelengths obtained with this source for over thirty elements.

Because of the variety of the sources available today it is

necessary to classify them, by considering the spectral region

to which the source is best adapted, the method used for exciting

the radiation or the type of spectrum emitted. In examilj.ing

some of the emission sources used in vacuum u.v. studies three

kinds of emission spectra may be differentiated; continuum, band

and line spectra.

a) Continuum and Band Emission.

Continuum emission spectra are produced by incan-

descent solids or liquids and, under special conditions, from

individual atoms or molecules. Band emission spectra are produced,

in general, by an electrical discharge through a polyatomic gas

or vapour. Examination under high resolution of band spectra

shows that they consist of a series of lines, so closely spaced

as to give the appearance of bands.

A classical example of continuum emission is the

radiation produced on heating a blackbody; a substance which

absorbs all incident radiant energy and whose emission spectrum

is due to its temperature. The spectral intensity curve from a

blackbody emitter from the low energy side gradually rises to a

maximum and falls sharply toward shorter wavelength. The position

11

of this maximum emission intensity is given by Wien's law,

2.898 x 106 Xmax.(nm) T

where T is the temperature in degrees Kelvin.

The total radiant energy may be calculated by means

of the Stefan-Boltzmann law,

W (watts.cm-2) = d.T4

... 1.2

d (Stefan-Boltzmann Constant) = 5.672 x 10 12watts.cm-2.deg 4

and the wavelength distribution of this emitted radiation is

provided by Planck's distribution law:

EA.dA = c2 AT - $ e

... 1.3

= spectral radiant intensity over the spectral bandpass dA(cm).

A = area of emitting surface

c1,c2 = first and second radiation constants.

Tables 1 and 2 show the total radiant power emitted

by a blackbody as a function of temperature and the position, at

various temperatures, of the maximum emission intensity. Fig.1.2

presents a series of energy distribution curves at different

temperatures. Although the shift in wavelength of maximum emission

intensity with increasing temperature is toward shorter wavelengths,

it is clear that at temperatures readily attainable in the lab-

oratory the ultraviolet portion of the radiation is only a small

part of the whole. As a source of far u.v. radiation, therefore,

blackbody emission is an inefficient energy conversion and

12

Fig.1.2

4000K

3000X

2000K

1000K

-o

.Lt E

E 108—

10 >N 7

L C 6 W 10 - a) 0 105 :-- — 0

Energy Distribution For A Black—Body At

Various Temperatures.

I I I 0 1 2 3 4 5

Wavelengt h (p)

Table 1.1 Total Energy Radiated per Unit Area

Of Radiating Surface At Various Temperatures

Temperature (K)

Watts.cm-2

1000

5.74

1500

29.

2000

92

2500

224

3000

464

Table 1.2 Wavelength Of Maximum Radiant Emission.

Temperature (K) 1000 1500 2000 2500 3000

Wavelength (a) 2.898 1.94 1.45 1.16 0.967

13

rarely used for studies at wavelengths shorter than 300nm.

One of the few sources (if not the only source)

capable of providing continuum emission from the visible to the

extreme u.v. is the Lyman discharge source (12). The emission

is obtained by discharging a capacitor through a gas, at low

pressure, contained within a glass capillary (typically 1mm internal

diameter). An external spark-gap is used to trigger the discharge.

This source has been extensively modified by Garton(13),

overcoming the problem of capillary erosion by increasing the

tube diameter to about 10mm, Fig.1.3. The necessary high current

densities (ca. 30,000A.cm2) are obtained by keeping the inductance

to a minimum and operating the discharge at upto 15kV, through

a gas pressure of less than one torr. The pulse of radiation

obtained is of the order of a few milliseconds duration and the

emission spectral distribution approximates that from a blackbody

of 30,000K. It is a particularly useful source for transient

species absorption measurements in the vacuum u.v., especially

in kinetic studies of gaseous reactions.

A more popular ultraviolet continuum source is

that produced by hydrogen (or deuterium) in a low pressure discharge

lamp or low pressure arc. The hydrogen lamp comprises a fused

silica envelope (or a glass envelope with a quartz window)

containing the gas at a low pressure (typically 10torr). The

discharge is operated at about 100 watts. The continuum emission

arises from an electronic transition from a stable excited state

to a lower state in which the molecule dissociates. The emission

extends from 160 nm to about 400 nm. The deuterium continuum is

produced in the same manner and its use provides for an increase

14

Trigger Connector

Ceramic Capillary

P

Garton-Type

Flash Tube

Pumping

Tube

Fig.1.3

Cathode Gas Exhaust

Gas In

Fused Silica Envelope

Gas In

Anode

Arc Region'

Vortex Stabilised Radiation Source

15

in intensity of two to three times that obtained in hydrogen.

Other u.v. arc-sources include the short-arc lamps and those of

carbon, argon and mercury (14,15). The argon arcs, and those

in other rare gases, are in wide use and have been extensively

described by Golman and others, (16,17). The arc is struck between

a tungsten rod cathode and copper plate anode; operating currents

of 400amps are common. Very high temperatures are obtained and

Dickerman et al. were able to obtain emission from a variety of

materials by introducing suitable powders into the gas feeding

the arc, (18). A commercially available system based on this is

the vortex-stabilised radiation source (V.S.R.S.), shown in Fig.1.4.

In general, as the gas pressure in a lamp is increased the emission

tends to a continuum. This is clearly demonstrated by mercury

arc lamps. In addition to mercury the lamp usually contains a

rare gas to assist in starting the lamp. Initiation of the discharge

causes mercury, contained in a reservoir, to evaporate and the arc

to strike. The mercury vapour pressure rises during the lamp

warm-up period and its continuum emission efficiency correspondingly

increases.

The rare gas continua are often excited by repetitive

capacitor discharges (typically 10kV pulses at a frequency ranging

from 100 to 1600 pulses.sec 1). Wilkinson et al. (19) used

microwave excitation, 125watts at 2450MHz, to obtain these continua.

The gas,at approximately 200torr, is contained within a quartz

tube fitted with a LiF window and held in a i-wave resonant cavity.

The addition of a gettering side-arm to the tube served to remove

most of the impurities from the discharge.

The useful spectral ranges of emission continua produced

16

by the rare gases and hydrogen are shown in Fig. 1.5. The continuum

emission produced by the noble gases is considered to be produced

by electronic transitions from the repulsive ground states of

diatomic rare gas molecules.

Continuum radiation also may be produced by the acceleration

(synchrotron radiation) or deceleration (bremsstrahlung radiation)

of free electrons. A review of such sources is included in the

vacuum u.v. spectroscopy text by Samson, (17).

b) Line Emission.

Line radiation is produced by transitions between

the electronic energy levels of neutral atoms, ions and molecules.

Although called line emission the lines do have a finite width

depending on the lifetime of the excited state involved in the

transition (natural broadening) and external conditions such as

temperature and pressure and the presence of electric and magnetic

fields. Typical line widths range between 10-2 and 10-4hm. The

problems of producing radiation of a given wavelength are concerned

with creating the appropriate conditions for exciting atoms or

ions. In general, the shorter wavelength radiation is produced

by the more highly ionised species. Milliken (8) was the first

to study the spectra of metals in the extreme ultraviolet using

a vacuum spark source. When high potentials were applied between

two electrodes less than a millimetre apart in a high vacuum

(10-5 to 106torr), a brilliant spark occurred. Using this spark

as a line source Milliken was able to extend the extreme ultraviolet

down to 14nm.

The most popular line emission sources, especially

in atomic absorption studies, are those produced by a glow discharge.

The cold cathode discharge tubes are operated by increasing the

17

Xe

" lD Kr

Ul

Ar

Ne

SOnm 100nm 150nm

VacuumU.v. Rare Gas Continua

I Aston dark space space

cathode glow

Fig.1.6

~ve

glow

space +ve

column

I anode dark space

anode glow

Emission Characteristics of a Glow Discharge

18

200n

voltage across the tube, containing gas at low pressure (ca. ltorr),

until the discharge strikes. This striking voltage is greatly

in excess of the potential required to maintain the discharge,

usually about 600V at between 100 and 500mA. Such a discharge

consists of a series of alternate light and dark regions as shown

in Fig.1.6. The presence of all these features depends on the

nature of the gas, its pressure and the operating conditions.

The cathode features, however, tend to be more permanent and

reproducible as they are involved in the maintenance of the discharge.

If the discharge is confined to a capillary (typically 100-200mm

long, 3-5mm i.d.) the positive column fills the entire capillary

and may be used as a line source. Fig. 1.7 shows such a lamp,

consisting of a water-cooled glass or quartz capillary sealed into

a hollow cathode and anode by 0-rings, (17). This type of source

was used by Hunter to obtain an intense hydrogen line spectrum

in the range 80nm to 170nm. (20).

A more common line emission source is the hollow

cathode lamp (sometimes referred to as a Schiller lamp) which emits

intense, narrow lines of selected elements. The emission is from

the negative glow of the discharge which, at certain gas pressures

and operating conditions, fills the entire cavity of the hollow

cathode. During the discharge the cathode becomes hot and some

vaporisation of the cathode material is to be expected; however,

ion-bombardment by the filler rare gas ions is the primary mechanism

for the cathode material entering the discharge. By constructing

the cathodes of different materials it is possible to excite lines

of many species and commercial hollow cathode lamps are available

for more than fifty elements. Paschen and Scffiler (21) were the

19

first to use this source which was later redesigned by Sullivan

and Walsh (22) who obtained an increase in intensity of two orders

of magnitude over the conventional source, without increasing the

line-width. A typical hollow cathode lamp source is shown

schematically in Fig. 1.8. The use of this type of source for

far ultraviolet studies is determined largely by the window

transmitting characteristics. However, because the elements

whose principal emission lines lie in the vacuum ultraviolet are

mainly the non-metals of high volatility, no commercial hollow

cathode lamps exist for this region. To overcome this problem

a demountable hollow cathode lamp may be constructed, allowing

a variety of cathode materials and windows to be used. A variety

of these systems have been proposed and used for emission studies

in the vacuum ultraviolet region. Milazzo (23) and Newburgh et al.

(24,25) have used a graphite cathode lamp to excite atomic emission

lines below 150nm. Milazzo has also studied graphite and aluminium

cathode lamps for the emission by aluminium, carbon and iodine (26,

27). A demountable, flow-through, water-cooled hollow cathode

lamp has recently been described by Kirkbright et al. as a line

emission source for atomic sulphur, phosphorus and iodine, (28).

Electrodeless discharge lamp (EDL) sources are also

a popular line emission source in the ultraviolet. The high

frequency DL source was discovered by Hittorf in 1884 (29) but

gained little attention in analytical spectrometry until the 1960's

when work by Winefordner (30) and West (31) led to their general

acceptance as suitable line emission sources for atomic absorption

and atomic fluorescence spectrometry. The excitation of this

type of source is achieved using radiowave or microwave frequency

20

:r: 0 --0 ~

. (")

0 /

~ ...... , ::r ,

I 0 I a. J co I

0 Ul n ::r 0 .,

Ul

water ou.t

water in

Positive Column Source

Fig.1.7

Fig. 1.8

r"---"'. r-------, I I I I I I I I

I I I I

0 I I 0 I

I ...... I ::r I 0 1 a. I co I

21

~~gas

In

electromagnetic fields in conjunction with wave-guide cavities,

coils or antennae. Although the excitation frequency may range

from 30-3000MHz the most common is 2450MHz because of the avail-

ability of comparatively inexpensive diathermy units which operate

at this frequency. Many reports have been published on the optimum

design, construction and operating conditions of EDL sources, (30,

32). In general, the lamp consists of a cylindrical bulb, ca.

30mm long, formed at the end of a 8mm i.d., 1mm wall thickness,

silica tube. This bulb contains the element, or its halide, of

interest along with about 5torr of inert gas, usually argon.

Such an arrangement only allows for the study of emission of

greater wavelengths than the cut-off limit of the quartz envelope

(ca. 170nm). For the transmission of shorter wavelength radiation

a more elaborate construction is required. Braun and Davies (33)

obtained intense atomic emission lines from O,N,S,C, and the

halogens using a flow-through EDL; helium, with a trace of the

species to be examined, was passed continuously through a quartz

tube fitted with a LiF window and supported in a 71,-wave Eveson type

cavity. The gas pressure was about ltorr and applied microwave

power typically 50watts. Various other vacuum ultraviolet EDL

sources are referred to in the literature (34,35) and the emission

is characteristically narrow spectral line emission of high .

intensity. They are used especially in photochemical studies,

laboratory aeronomy, Raman spectroscopy and, increasingly, anal-

ytical atomic spectrometry.

Other, less common, line emission sources such as the

hot filament arc discharge and duaplasmatron are discussed in the

reviews on vacuum ultraviolet sources by Samson, (17,35).

22

1.3 Window Materials and Spectral Filters.

Compared with the other regions of the electromagnetic

spectrum the choice of window materials and filters for use in

the far ultraviolet is limited. The more important crystalline

window materials are listed in Table 1.3 with their short wavelength

cut-off limits. This limit is taken here as corresponding to that

wavelength at which the transmission of the material reaches a

value of 1C% for a thickness of 1mm, at room temperature. The

short wavelength limit is not the only criterion which determines

the choice of a window material for a particular application; the

slope of the transmission curve with decreasing wavelength is

equally important. The transmittance of all these materials is

temperature dependant and, in general, as the temperature increases

over the range 77K to 650K the transmission limit shifts to longer

wavelengths at the rate of 0.025 to 0.03nm.deg-1. At room temperature

the hydrogen Lyman-Kline at 121.6nm is totally absorbed by a

1mm calcium fluoride window,whereas at 77K the cut-off is shifted

to 117nm and the radiation is transmitted, (37). Purity of the

crystal and differing methods of manufacture may alter the trans-

mission characteristics of a window considerably. Smaller thicknesses

obviously extend the cut-off limit to shorter wavelengths. Thin

films of silica (ca. 10mg.cm 2), for example, will transmit 80%

of the radiation of wavelengths longer than 100nm (38) but these

are very fragile and have little practical value in normal use.

The means by which the window may be sealed to

a lamp or optical system also needs to be considered. Quartz

and sapphire have the advantage of being capable of providing

graded seals to glass which are resistant to thermal and mechanical

shock. The other materials such as CaF2, SrF2, BaF2, - MoF2' etc.

23

cannot be sealed in this manner and less permanent; and often

less satisfactory, means of making a seal are necessary. Lithium

fluoride is an exception and a graded seal to glass may be made

by means of silver chloride-silver-silver chloride discs between

the window and glass (38). The assembly, if constructed carefully,

is stable to thermal shock and will withstand temperatures as high

as 700K. Many of the materials listed in Table 1.3 are rarely used

because of their solubility in water. Attack by water vapour in

the atmosphere may cause a loss in transmission, especially at

the lower wavelengths approaching the transmission limit.

Below the lithium fluoride transmission cut-off

(105nm) windowless, differential pumping systems are necessary

as no crystalline window material is available capable of transmitting

this short wavelength radiation. Very thin films of aluminium

and plastics are transparent below 100nm but their fragile nature

prevents their common use.

Many gases and gas mixtures, because of their

discreet absorption band systems in the vacuum ultraviolet, may

be used as spectral filters in this region and Table 1.4 (39,40)

lists several examples. All the constituent gases of the atmosphere,

with the exception of helium and neon, exhibit wide spectral

regions of strong band or continuum absorption between 50nm and

200nm. The principal absorber, as previously stated, is molecular

oxygen. The Schumann-Runge oxygen absorption bands, commencing

at ca.195nm, show intense absorption by 185nm and converge with

a strongly absorbing dissociation continuum at 175nm. Maximum

absorption is observed at 142.5nm. A localised "window" exists

between 120nm and 135nm and Garton has utilised this region of

24

Table 1.3 Crystalline Window Materials.

Material

Limit (nm) Remarks

Window glass 300 Impurities effect transmission

Pyrex 270 PY

Vitreosil 180

Synthetic fused quartz 160

Cultured crystal quartz 145

Sapphire 142

Barium fluoride 135 Sol. 0.12g.100m1-1

Strontium fluoride 128 ,, 0.011, ,,

Calcium fluoride 123 „ 0.002

Magnesium fluoride 112 „ 0.008

Lithium fluoride 105 yp 0.270

Table 1.4 Gaseous Spectral Filters.

Gas Filter Thickness

At NTP (cm)

Regions Of Relative Transmission

(um)

02 6.0 110 —111, 121.6,>180

CH3Cl 0.1 142— 146, >185

CH3Br 0.05 152 —157, >180

H2S 0.1 160 —170, , 230

CH4

0.025 7140

CC14

0.0025 116 —120, 7155

C2H4

0.025 7185

(CH3)3CH 0.1 >106

25

6.1:6!.d

light

Photovoltaic Cell

Fig.1.10

Absorption Coef. (cm-1

o = Ni Ni

0

..-.1. --a. 0 1C:) 1

IN

1 .-a

transparent film

selenium co

__.a. (A) — CD

C:)

( WL.

1 ) q

;Bue

lano

m

(J1 — CD

0

0

.__ CO — 0

1..) 0 - 0

Photocell

Fig.1.11

relatively high transparency in monitoring the emission from

hydrogen at 121.6nm, (41). At wavelengths shorter than 120nm the

absorption continuum and intense absorption bands recommence. The.

absorption of vacuum ultraviolet radiation by water vapour is

frequently as serious in analytical spectrometry as that due to

oxygen. The water vapour absorption continuum occurs between

183nm and 150nm, where a relatively transparent region exists

down to 138nm. Below this wavelength strong, diffuse absorption

bands occur, (42) Fig.1.9.

Most other common gases are relatively transparent

in the Schumann ultraviolet region of the spectrum. The band

absorption spectrum of nitrogen commences at 145nm but is weak

until 110nm and carbon dioxide is a stronger absorber in the 115nm

to 140nm region. Hydrogen band absorption does not commence until

110nm and becomes a continuum only at wavelengths shorter than

85nm. The absorption characteristics of many polyatomic gases

and vapours are given in the review by Sponer and Teller, (43)

1.4 Dispersion Elements.

The dispersion device is a major element in any

spectrochemical technique as it provides the means by which the

radiation emitted by a source may be separated into its differing

wavelength, or energy, components. Two types of dispersion element

are in common use; prisms and gratings. Their efficiency and use

in the ultraviolet has been the subject of several reviews,(17,36).

The prism serves as a dispersion device because

of the phenomenon of refraction. The linear dispersion of a prism,

D, depends on the prism angle, 20C, the focal length of the

spectrographlf, the wavelength of the incident radiation and

27

constants of the prism material.

2 1 D = ()-K

1 )2 (1-sin 002 ...1.4

2K2

f sin oC

i.e. D depends upon the square of the wavelength of the incident

radiation; making for difficulties in assigning precise wavelengths

to an emission or absorption spectrum.

The concave diffraction grating was first described

by Rowland (51) in 1882. He showed that the linear dispersion

depended only on the radius of curvature of the grating, the angle

of emergent radiation p and the groove separation on the grating

surface,d,

D = n Rc x 109 ...1.5

d cos

i.e. the linear dispersion is uniform over the whole wavelength

range.

Because of the necessity of constructing prisms

from LiF or CaF2 for use in the Schumann ultraviolet, and the loss

in transmission of incident radiation of shorter wavelengths by

absorption with these materials, the diffraction grating is the

most commonly used dispersion element in use today for studies in

this region.

1.5 Radiation Detectors.

The results of the interaction of ultraviolet

radiation with matter provides for a variety of detector types;

classified according to whether the reaction is basically a chemical

or physical process.

28

a) Chemical Detectors.

The oldest and most common of the chemical detectors

is the photographic plate. This serves as a detector according to

the effect of the radiation on silver salts held on a glass plate

with a gelatin binder. The degree of blackening of the emulsion

is a measure of the incident radiation intensity. Unfortunately,

the gelatin base of this emulsion absorbs radiation of wavelengths

shorter than 280nm and is opaque below 200nm. This may be overcome

by removing most of the gelatin (the Schumann plate), but this

renders the plate very fragile and sensitive to abrasion. A more

satisfactory process is to coat the normal plate with a substance

capable of absorbing the incident far ultraviolet radiation and

re-emitting radiation of longer wavelengths than 280nm; fluorescence

sensitisation. This technique was first examined and reported by

Ducleaux and Jealet (45) who coated a photographic plate with

light machine oil which fluoresced under exposure to vacuum ultra-

violet radiation. Since their work a variety of sensitisers have

been examined including the dihydrocollidine ethyl carboxylic

ester, manganese activated willemite phosphor (sensitive to radiation

above 145nm), sodium chloride (sensitive below 30nm) and sodium

salicylate (17,46). Sodium salicylate is by far the most common

in use today. It is a finely crystalline powder readily soluble

in methanol. A plate is sensitised by spraying it with a saturated solution

in methanol. A hot-air blower continually playing over the surface

assists in the evaporation of the solvent. In this manner a thin

layer of sodium salicylate is produced which may be built on until

the required thickness is obtained (1 to 2 mg.cm-2). Its fluorescence

spectrum has a maximum intensity at around 420nm and its absolute

29

quantum efficiency (ca. 65%) is constant between 50nm and 220nm,

although in some circumstances ageing may decrease its response

below 160nm. Diphenylstilbene (DPS) and tetraphenylbutadiene (TPB)

both have greater quantum efficiencies than sodium salicylate

but have found little practical application as sensitisers as both

readily sublime under vacuum at room temperature.

Ultraviolet radiation may also be detected by the

acetone-methylene blue reaction. Aqueous solutions of the reagent

are bleached by radiation below 320nm and the degree of bleaching

is proportional to the incident radiation flux. The reaction,

however, is very slow and rapidly reverses in the dark. A better

reaction is the decornpostion of oxalic acid, by ultraviolet

radiation, in the presence of uranyl sulphate. The amount of

decomposition, proportional to the incident radiation level, may

be determined by titrating the remaining acid with potassium

permanganate, (46).

Chemical means of radiation detection, although

useful in providing an integrated response with respect to time

of the incident radiation flux, are too slow for most applications

and too sensitive to temperature and the presence of small amounts

of impurities.

b) Physical Detectors.

The majority of these detectors, whose response

to incident radiation is a change in some physical property, belong

to one of two types; the radiometers and the photoelectric devices.

Radiometers detect and monitor the heating effect produced by

absorbed radiation. Included in this class are the thermopile and

balometer. The former comprises a junction of two dissimilar

30

metals which, when heated by absorbing radiation, generates an

emf. This emf is proportional to the energy of the incident

radiation rather than the radiation flux. The balometer is a

delicate Wheatstone-bridge arrangement, one arm being heated by

the radiation of interest; the change in resistance upon irradiation

being the measured quantity. Semiconducting balometers (thermistors)

are commercially available and are more sensitive than the older,

metal-arm systems. The Golay cell (47) is also a heat sensitive

detector and is basically a differential gas thermometer. The

expansion of gas causes movement in a deflecting membrane; the

degree of deflection is measured by its defocussing effect on a

light beam falling onto a photocell. Radiometers are probably the

only detectors having constant response throughout the infra-red,

visible and ultraviolet regions and they are frequently used as

laboratory standards. Practical difficulties, however, prevent

their widespread use for routine determinations.

The second type of physical detectors (photoelectric

devices) are the most commonly used detectors today. Photovoltaic

cells, resistance cells, photodiodes and phototubes all have found

application in detecting and monitoring ultraviolet radiation.

The construction of a typical photovoltaic cell is shown schematically

in Fig. 1.10. Radiation incident on the surface of the cell

causes a flow of electrons from the semiconductor to the metal

and a current to flow in the external circuit from the iron to

the selenium. The magnitude of this current is proportional to

the intensity of the incident radiation; although below 27Cnm

the spectral response of this type of detector is very poor.

The most important detector for ultraviolet radiation

31

is the vacuum phototube. The absorption of a quantum of radiation

at the surface of a suitably prepared metal or alloy may cause

the ejection of an electron from the surface if the energy of the

radiation is greater than some threshold value. The quantum

efficiency for this photoelectric process. (the ratio of photoelectrons

to incident photons) at photon energies above the threshold value

varies in a complex, but in most cases well known, manner. In

general, it rises to a maximum with increasing incident energy

and falls off in a manner determined by the optical absorption

properties of the photosurface and any window material present.

A typical photocell is shown in Fig.1.11. A separate anode is

mounted inside the evacuated envelope to collect the photoelectrons

and currents generated by such devices lie in the range 10-6 to

- 10 16 amps, (48). The small magnitude of these signals usually

necessitates one of two basic amplification procedures. The

cell may contain argon as a filler gas and in such a gas free

electrons may exist without negative ion formation. At high

anode voltages the electrons, in crossing from the photocathode

to the anode, may attain sufficient energy to cause ionisation

of the filling gas by collisions and secondary electrons, also

collectable by the anode, are formed. Increasing the voltage

even further may cause more than one ionising collision per

photoelectron and further amplification is achieved. For 'nt

such ionising collisions the amplification (gas-multiplication)

factor, /I, will be given by 2n. The more common method of signal

amplification, by secondary emission, is used in photomultiplier

tubes. The photoelectron ejected by the photocathode has kinetic

energy the magnitude of which is related to the excess energy of

the incident photon, over and above that necessary to achieve

32

the threshold limit of the cathode, and the potential gradient

between the cathode and anode. The photoelectron incident on

the collecting anode surface passes the excess energy to electrons'

bound to the solid-vacuum interface and a few of these may approach

the interface with sufficient energy to overcome the potential

barrier and emerge as secondary electrons. These secondary electrons

may pass to a new anode for collection or for them to emit, in a

similar manner, further electrons. By increasing the number of

these secondary emitting surfaces (dynodes), each at a potential

higher than the previous one, amplification is achieved. If the

number of secondary electrons produced per incident electron is Ix'

and the number of dynodes or stages is 'n', then the total gain

in the photomultiplier may be expressed by xn. This is frequently

of the order of 106 to 108. Fig.1.12 illustrates a typical photo-

multiplier arrangement.

The short wavelength spectral response of such

devices is determined largely by the transmission characteristics

of the envelope or window. Using LiF, MgF2 or CaF2 windows the

spectral response of the photomultiplier tube is extended well

into the vacuum ultraviolet, although these tubes are rarely used

for analytical spectrometry owing to their high cost. For work

below 100nm no window material is available and windowless tubes,

operated in high vacuum surroundings, are necessary. A more

common technique to extend the wavelength response of photomultiplier

tubes is to coat the window of the tube with a sensitiser as

described for photographic plates. Again, the most common is

sodium salicylate. The development of photomultiplier tubes with

cathodes insensitive to radiation of wavelengths longer than

33

E.HT.

anode

cathode light 1 2 3 4 5 6 7 8 9

shield

readout 0

-1-

Photomultiplier System

ca. 300nm (solar-blind tubes) has been the subject of much research

and these devices are discussed in the review by Dunkelman (49).

With decreasing wavelength the energy associated

with radiation increases, and in the vacuum ultraviolet this

energy is sufficient to ionise many molecular species. This

ionising ability forms the basis of the photoionisation techniques

of detecting ultraviolet radiation. The photoionisation detector

comprises a gas or vapour contained within a chamber between two

electrodes, across which a potential difference is applied. The

spectral response of such a detector is determined on the low

energy (long wavelength) side by the ionisation characteristics

of the filling gas and on the high energy (short wavelength) side

by the transmission characteristics of the window material present

on the chamber. The use of these detectors in monitoring far

ultraviolet radiation is discussed in more detail in Chapter Eight.

1.6 Low Wavelength Analytical Spectrometry.

Spectrochemical analysis may be both qualitative

and quantitative. If the concentration of the analyte remains

constant and the wavelengths of emission or absorption are scanned

the spectrum obtained is characteristic of the species being

examined. Selecting a suitable wavelength and varying the analyte

concentration provides quantitative information regarding the

sample.

Both qualitative and quantitative, molecular and

atomic, spectrochemical analysis is possible in the vacuum u.v.

a) i,',olecular Spectra.

Despite the high absorptivities and characteristic

35

spectra of many gases in the far ultraviolet; relatively little

analytical molecular absorption analysis has been reported.

The main reasons for this are the absorption by the atmosphere

of radiation below 200nm and the limited transmission of most

spectrophotometers in this region. Kaye (50) modified Beckman

DK-1 and DK-2 spectrophotometers for use at wavelengths down to

170nm. This was accomplished using high quality silica for the

prisms and optical components and a hydrogen continuum lamp, with

a thin quartz window as the source. The attenuation of the

continuum radiation by the atmosphere was eliminated by purging

the optical path with nitrogen. Using a 1mm cell the absorption

spectrum of CS2 (0.2torr) between 170nm and 220nm was recorded.

Absorption of vacuum u.v. radiation by molecules occurs through

electronic transitions and any fine structure observed is due to

vibrational transitions. The overlapping of vibrational and

electronic states is evident from the far ultraviolet spectruM

of CS2. In a similar manner the spectra of methyl iodide vapour,

iodine vapour and ammonia were obtained (51). The possibility

of quantitatively analysing for ammonia in nitrogen and air was

examined by Gunther (52) and using the absorption bandhead at

204.3nm he was able to detect as little as 7ppm NH3. Ammonia

also has a bandhead at 193.6nm and this would provide for even

lower detection limits but the presence of oxygen could not be

tolerated.

The absorption spectra of liquids and solids in the

far ultraviolet are not so discreet as electronically excited

states are readily perturbed by neighbouring molecules. Vibrational

fine structure, therefore, is almost always absent on going from

vapour to solution spectrometry. Two solvents of widely varying

36

polarity, water and n-heptane, are among the most transparent

liquids at wavelengths above 172nm. A library of reference absorption

spectra in the far ultraviolet of organic and inorganic compounds

is presented in the reviews by Kaye (51) and Sponer (43).

Molecular emission is rarely of analytical use in

the far ultraviolet as the conditions necessary to attain a

significant excited population of a species frequently cause

decomposition of the molecule. Most of the molecules whose emission

spectra have been recorded are strongly bonded diatomic systems

eg. H2,N2,02,S2,CO3etc. (10).

b) Atomic Spectra.

The greatest amount of observed and recorded data

obtained in the far ultraviolet is concerned with atomic spectra.

Although all the elements show some characteristics in this region,

analytically the far ultraviolet is mainly concerned with the non-

metals. Table 1.5 lists those elements whose resonance transitions

occur in the Schumann ultraviolet.

Analytical atomic emission in the vacuum u.v.

is normally achieved using a vacuum spark source. The anode of

copper or carbon is set approximately 2mm above the sample, the

cathode, and the spectrum obtained using an overdamped discharge.

Spark sources have been used for the determination of carbon,

phosphorus, sulphur, selenium and arsenic in iron, steel and a

variety of alloys, (53,54).

Several workers have described the use of the

hollow cathode lamp as a source for quantitative atomic emission

analysis. Konovalov and Frisch (55) were able to detect nitrogen

and argon to a few thousand ppm and McNally et al. (56) developed

37

Table 1.5 Vacuum Ultraviolet Resonance Transitions

Atomic Species Wavelength (nm) E. - F (cm 1)

Chlorine 118.88 0 - 84116

Bromine 129.30 0 - 77330

Bromine 129.40 '0 -.77260

Sulphur 129.59 0 - 77166

Sulphur 131.66 0 - 75955

Chlorine 133.57 0 - 74861

Chlorine 134.72 0 - 74221

Sulphur 140.15 0 - 71353

Sulphur 142.52 0 - 70168

Bromine 144.99 0 - 68970

Sulphur 147.42 0 - 67829

Bromine 148.86 0 - 67190

Bromine 149.50 0 - 66877

Iodine 150.70 0 - 66355

Iodine 151.47 0 - 66021

Bromine 154.08 0 - 64901

Bromine 157.65 0 - 63430

Iodine 158.26 0 - 63187

Iodine 161.76 0 - 61819

Iodine 164.21 0 - 60895

Phosphorus 167.17 0 - 59820

Phosphorus 167.46 0 - 59716

Phosphorus 167.97 0 - 59535

Phosphorus 177.50 0 - 56340

Iodine 178.28 0 - 56093

Phosphorus 178.29 0 - 56090

Phosphorus 178.77 0 - 55939

Sulphur 180.73 0 - 55331

Iodine 183.04 0 - 55171

(From Ref. 97 and 98)

38

a spectroscopic method of analysis for fluorine, chlorine and

• sulphur at the ppm level using such a source. More recently

Milazzo (23,27) has described a hollow cathode source capable

of determining iodine at the nanogram level in a matrix of KC1.

Using a cathode sputtering cell, Kirkbright and Wilson have also

determined iodine, down to 0.02pg by atomic emission at 183.0nm(57).

The radiofrequencypinduction-coupled plasma has

been demonstrated to be an excellent excitation source for non-

metal emission analysis. Using such a plasma Ward was able to

detect iodine, sulphur, phosphorus and carbon at the ppm level,

(58 59). Dreher and Frank have described an induction-coupled

plasma powered from a 270W R.F. generator at 27.5MHz used for

the emission spectrometry of small samples of arsenic (at 189.0nm)

and iodine (at 183.0nm). The detection limit was 1pg for both

elements.

Atomic absorption spectrometry (AAS), as a sensitive

and quantitative analytical technique, developed from the work

by Walsh (60). If the shape of the atomic absorption profile

is due entirely to Doppler broadening, the absorption coefficient

(ky) at the line centre measured with a line emission source having

a negligable half-width is given by,

ky 2 (11.1 1r 2_. ire

---G) C 454T • " 1 ' 6

i.e. the absorption coefficient is dependant on the number of

atoms (N) capable of absorbing the incident radiation. Atomic

absorption follows an exponential law for the intensity of

transmitted radiation (I) against absorbing volume length and

39

absorption coefficient,

I = Io e-k

... 1.7

Where Io is the intensity of the incident radiation. In practice

the absorbance (A) is measured, defined as,

A = log (I0A) ... 1.8

and, from Eq. 1.8,

A = kyl/2.303 ... 1.9

i.e the absorbance is directly proportional to the absorption

coefficient.

The most common atom cells in analytical AAS are flames.

At wavelengths shorter than 200nm, however, few flames have

sufficient transparency to allow atomic absorption measurements

to be undertaken. Kirkbright et al. (61,62,63) have demonstrated

that the nitrogen-separated nitrous oxide-acetylene flame permits

radiation above 175nm to be transmitted and have used this flame

as the atomiser for the direct determination of mercury at 184.9nm,

iodine at 183.0nm, sulphur at 180.7nm and phosphorus at 177.5nm,

178.2nm and 178.8nm. A typical instrumental arrangement employed

is shown in Fig.1.13.

The use of non-flame atomisers for AAS in the vacuum

ultraviolet is described in more detail in Chapter 2.

40

N20 -C2H2 burner

vacuum monochrom.

.... ..."

(0 EDL ._. .__ 0.) p.m.t

/ modulator

amp.+ p.s. E.H.T.

readout

VacuumU.V. Flame AAS System

TWO

2.1 Introduction

Atomic absorption spectrometry (AAS) has developed

rapidly into a powerful analytical technique, complementing the

older practice of atomic emission spectrometric analysis (AgS).

Because of its development from AFS it is not surprising that

flames, the original excitation cells for emission analysis,

became the most common and widely accepted atom cells for AAS.

As the atom cell is responsible for the production, from the

analyte, of the atoms on which qualitative and quantitative meas-

urements are made the cell must be capable of desolvating the

sample, dissociating the molecular species present and, for AAS,

producing ground-state atoms. These conditions usually necessitate

high temperatures and a chemically reducing atmosphere. The atomisation

process should be independAnt of the nature of the sample, i.e.

no matrix effects, and be adaptable to a variety of techniques

(AAS,AES and AFS). There should be no spectral interference from the cell

itself, and, finally, the atoms once formed should have sufficiently

long enough residence time in the optical path for analytical

measurements to be made. Although no atom cell fulfills all of

these criteria, flames provide most of the requirements necessary

for practical AAS as well as simplicity of operation, ease of

sample introduction and a variety of available types. Because

of this progression from emission to absorption and the use of

the readily available flame systems little research was undertaken

on non-flame cells until the late 1960's. Today many such cells

are available commercially and their application is becoming

rapidly routine in many analytical laboratories. This expanding

field of non-flame AAS has prompted several reviews of these

systems, (64,65). It is necessary here only to mention the

relative merits of flame and non-flame cells and discuss the

more common non-flame atomisers; especially those having been

employed for the detection and determination of the non-metals

and those capable of possible far u.v. AAS.

The more important advantages of using electro-

thermal atom cells over flames for AAS may be summarised as follows.

1) Non-flame cells provide for greater sensitivity than flames.

Winefordner (66) has shown that the atomic concentration in

electrothermal devices is greater than in flames; no gas expansion

as occurs with flames takes place with non-flame cells.

2) Much smaller samples (typically 0.5 to 10911) are capable of

being analysed using electrothermal cells. The sensitivity

of the technique may be increased by applying larger volumes

to the atom cell-each sample being dried usually before a

second is applied.

3) The reactions of analyte atoms with hot gas molecules in flames

often leads to complex and thermally stable species being formed

eg. monoxides, monohydroxides,etc. As non-flame cells use

inert-gas purging,oxide formation is a much less serious problem;

although. reactions with the cell material eg. graphite may

cause a loss in sensitivity due to stable compound formation

eg. carbides.

4) Compared with many flame systems the background emission from an

electrically heated cell is negligible in the wavelength region

most commonly employed for AAS; below 400nm. Any interfering

emission that is present may be overcome with little difficulty

by limited-field viewing or source-emission modulation.

5) The use of potentially explosive gas mixtures as employed

with flames makes their use unsuitable for many applications.

6) Solid samples may be analysed directly in non-flame cells in

a number of applications. Solid sample nebulisation into

flames is a notoriously unreliable technique.

7) For the AAS study of non-metals, whose resonance lines occur

at wavelengths shorter than 185nm, all flames suffer the

disadvantage of absorbing much of the incident radiation. The

nitrogen-separated nitrous oxide-acetylene flame has proved

to be the most suitable for the AAS determination of the non-

metals at wavelengths above 175nm, (470,G2,63). The inert

gas separation of this flame increases its reducing nature

and decreases the partial pressure of oxygen, the principal

absorbing species of radiation below 200nm. Wilson investigated

the transparency of the nitrogen-separated nitrous oxide-acetylene

flame and a variety of other flames at the iodine 183.0nm

resonance line and the results are summarised in Table 2.1.

Non-flame cells do not suffer from this loss in

transmission below 200nm as the inert gas purging employed

with such devices serves to remove the atmospheric oxygen,

water vapour, etc. from the cell. However, at these shorter

wavelengths background absorption and scatter of incident

radiation by vaporised and particulate matter from the furnace

may present similar problems as those observed with flames.

The major disadvantages of non-flame cells are due

to the transient nature of the signals produced by these devices.

Such signals are typically 0.1 to 10sec in duration depending on

the cell geometry, and sophisticated electronic processing systems

45

Table 2.1 Transmission Characteristics of Various Flames

at 183.0nm

Flame % Transmission

Nitrous oxide — Acetylene 25%

Nitrogen Separated 75%

Nitrous oxide — Acetylene

Air — Acetylene 10%

Nitrogen Separated

Air — Acetylene 28%

Hydrogen — Nitrogen 3%

Table 2.2 Non—'lame AAS Results Reported by Lvov (Ref. 71)

Element Wavelength (nm) Sensitivity (pg)

Iodine 183.0 36.7

Sulphur 180.7 81.4

Sulphur 182.0 105.0 -

Sulphur 182.6 315.0

Phosphorus 177.5 3.4

Phosphorus 178.3 4.4

Phosphorus 178.8 7.4

for 1% absorption

46

are often required especially with 0.1 to 0.5sec signals. Because

of the shorter residence time of the analyte sample in the heating

zone of non-flame atomisers interference effects caused by non-

specific absorption from thermally stable inorganic salts are

encountered frequently. This molecular absorption interference

tends to be far more severe with electrothermal cells than with

flames and background correction techniques are widely employed

to counter these effects. As mentioned previously, some elements

may react with the material from which the cell is made, forming

stable species and lowering the sensitivity of the technique.

2.2 Furnace Atomisers

The King Furnace. The majority of non-flame atomisers in use

today for AAS are constructed of graphite and resistively heated

by low voltage, high current power supplies. The original cell

on which most are based was due to King in 1908, (67).

The King furnace comprised a graphite

tube, 150 to 200mm long, 16mm o.d., of wall thickness 0.5mm except

at the ends which were thicker to take the metal supports serving

as electrodes. It was heated from a 25V, 200A a.c. supply and was

surrounded by a water jacket. The hydrogen purging employed

prevented the oxidation of the graphite. Using this apparatus

King observed the emission and absorption spectra produced by a

variety of metals (Na, Ca, Cu, Cs, etc.) under carefully controlled

conditions. The sample, usually as the metal or its halide, to

be investigated was placed inside the furnace in a small porcelain

boat. Working temperatures of 2900K were obtained and the spectra,

after dispersion with a spectrograph, recorded on photographic plates.

47

A similar but much larger furnace has been constructed

by Collins et al. (GS). The graphite tube was resistively heated

by a 42.5V, 5000A power supply and maintained for over 100 hours

at 2200K without signs of decomposition.

The L'vov Furnace. The first furnace reported to be employed

successfully for analytical purposes was developed in Russia by

L'vov. A topical L'vov furnace is shown in Fig.2.1. He was

one of the first to examine the effect of the size of the furnace

on the analytical signals obtained and tubes 20 to 50mm long and

1 to 6mm i.d. have been reported, (6q,7o). The tube is resistively

heated by an a.c. current at 10V from a 4kW transformer. The

temperature attained by the furnace is regulated by the voltage

supply to the primary circuit of the transformer. The sample

is placed on the tip of an auxilliary 6mm diameter graphite rod

electrode and dried; the tip of this rod is shaped to fit the

orifice in the tube wall. Five or six electrodes with samples

are positioned below the tube in an evacuated chamber. Argon is

introduced into the chamber until the desired pressure, upto 8

or 9 atm., is achieved. The tube is heated for 20 to 30sec until

the required temperature is attained and the auxilliary electrode

moved into the orifice in the tube wall. This electrode is then

heated by a separate a.c. current from the secondary windings

of a 1kW step-down transformer (220/15V). The sample on the rod

is vaporised and the atomic absorption signal recorded. With

an argon pressure of 1.2atm, the tube at 1900K and using LiF

windows and lenses, a vacuum monochromator and electrodeless

discharge lamp line emission sources L'vov determined iodine,

48

4cm.

crucible

graphite electrode with sam•le contacts

L'vov Furnace. Fig. 2.1

300mm

[7- Furnace

Sample Electrode

Woodriff Furnace Fig.2 2

49

sulphur and phosphorus by AAS at their vacuum ultraviolet resonance

lines, (71). The sensitivities achieved are shown in Table 2.2.

The Woodriff Furnace. Independently of L'vov's work a similar

furnace atomiser arrangement was developed by Woodriff, (65,/2„).

The Woodriff furnace (Fig.2.2) consists of an insulated graphite

tube containing an inert gas, usually argon, at atmospheric pressure.

The sample port has a constriction near the atomisation tube

making a seal with the rim of the sample containing crucible.

Unlike L'vov's system the temperattmof this crucible is raised

by heat transfer from the furnace. The rate of heating, therefore,

depends on the heat capacity of the crucible, the rate of insertion

of the crucible into the furnace and the rate of heat transfer

from the tube to the crucible. No vacuum ultraviolet AAS studies

have been reported using this system; because of its complex

design and large physical size relatively few applications for

metal element AAS by workers other than Woodriff have been reported.

However, the use of a large furnace does assist in eliminating

sample matrix effects and this system may be of value in future

non-metal AAS studies. As with the L'vov furnace the fact that

the sample in the Woodriff system is completely enclosed inside

the tube, except for the tube ends, increases the atomic residence

time in the optical path providing for increased sensitivity.

The Massmann Furnace. A more popular graphite furnace has been

described by Massmann (73) and is available commercially (74);

the HGA 2000 (Perkin-Elmer Corptn., U.S.A.). This is shown in

Fig.2.3. The graphite tube is 53mm long, 6.5mm i.d., has a wail

thickness of approximately 1.5mm and two purge gas holes and a

central sample introduction orifice in the wall. The furnace is

50

UU

DW

SS

DIA

J -

NIN

Electrical

Removable Window . Gas- In

Water In

■••■••••

Graphite Tube

Optical Axis

—Insulator

Water Out

Massmann Furnace Fig.2.3

sp 3 3

rn

3 3 II

(r)

3 -o fp

Fig.2.5

Fig.2.4

51.

held between conical graphite end-pieces in water-cooled metal

electrodes inside an inert-gas purged chamber, open to the atmosphere

at the tube ends. The sample solution (typically 5 to 20q41) is

placed in the cold tube via the central hole using a micropipette.

The tube is heated gently to dry the sample and remove the solvent

from the furnace and then heated more strongly, with a current of

upto 400A, to atomise the sample. If an organic matrix is to

be analysed an ashing stage may be introduced between the drying

and atomising steps to decompose the matrix and prevent smoke

formation during atomisation. At full power (12V at 400A) the

tube may attain a temperature of upto 3000K within a few seconds.

The water-cooled supporting electrodes allow for rapid cooling

of the furnace and reduce the dead-time between samples. This

arrangement also makes for a thermal gradient along the length of

the tube when it is heated and because of the cooler tube ends

this system occasionally suffers from severe background absorption

and molecular absorption from the sample matrix. Wilson (is)

modified a Perkin-Elmer 305B double beam atomic absorption spectro-

photometer with HGA2000 furnace and background correction

facilities to enable the determination of iodine at 183.Onm using

an iodine EDL as the line emission source.

Morrow et al. (76) have recently described the

modification of carbon rod atomisers to accept Massmann type graphite

tubes. Their tube design is shown in Fig.2.4.

The Mini-Ilassrnann Furnace. A smaller carbon furnace atomiser

was reported by Amos and Moutasek (/7). It consists of a short

length of carbon tube (ca. 8mm) held between the ends of two carbon

rods, Fig.2.5. With a small hole in the centre of the tube it

52

may be used in a similar manner to that described for the Massmann

furnace. This carbon rod atomiser, often referred to as a mini-

Massmarn, is also available commercially (Varian Inst. Pty. Aust.).

In general this small tube has lower background absorption and

greater sensitivity than the larger furnaces but shows the most

severe matrix effects.

Becker-Ross and Falk (-78) have recently

described the direct determination of bromine by AAS using a bromine

resonance line in the vacuum u. v.; 148.86nm. Their atom cell

was a carbon tube oven fitted at the ends with LiF lenses. A

hollow cathode lamp was employed as the source and non-specific

absorption corrected for by utilising the nearby nitrogen 149.26nm

line.

2.3 Filament Atomisers.

In addition to the graphite tube atom cells,

graphite filament atomisers have been employed extensively for

the determination of trace quantities of metals by AAS. These

filaments are based generally on the system designed by Alder and

West (19) and comprise a short carbon rod (ca. 25mm long and 2mm

diameter) clamped between two water-cooled supporting pillars,Fig.2.6.

Power is transferred to the rod via the usual Variac and trans-

former arrangement. Because of the open nature of this type of

cell volatile elements and those species requiring a long residence

time at elevated temperatures are unsuitable for AAS analysis

with this system. Marshall (80) examined such a cell for the

direct determination of sulphur but was unable to detect any

atomic sulphur signal from elemental sulphur or its compounds.

Graphite is the most common material from which

53

filaments and furnaces have been made. Its electrical conductivity,

low thermal expansion, excellent thermal stability and chemical

reducing nature all combine to enhance its value as the primary

furnace construction material. It is not,however, the only material

from which non-flame atomisers have been produced. Cantle (81)

has recently described a tantalum filament, used in a similar

manner to the graphite filament of Alder and West, and many workers

have employed tantalum liners inside graphite furnaces to eliminate

carbide formation, (82). A tantalum furnace which has received

little study and may be of value in vacuum ultraviolet AAS is the

boat arrangement described by Donega and Burgess (83). The

tantalum boat, 50mm long and 6mm wide, containing the dried sample

is held inside a closed chamber, Fig.2.7. The filament is heated

electrically with a current of 30 to 50A at 12V and is capable

of achieving about 2500K in less than 0.1sec. The chamber, of

quartz tubing, may be purged with inert gas at pressures between

1 and 760torr.

2.4 Miscellaneous Atomisers.

Several workers have investigated the potential

of cathode sputtering cells as atom generating systems for AAS.

Walsh (810 used such a cell to provide phosphorus atoms from

metal samples for the direct determination of this element at

its vacuum ultraviolet resonance lines. More recently Wilson (57)

used a Skogerboe type sputtering cell, Fig.2.2, for the direct

determination of iodine at 183.0nm. A detection limit 0.05pg

was achieved but the magnitude of the absorption signal was

dependant on the nature of the iodine containing species and was

suppressed by the presence of other halides.

54

JI0A

-leSei

WO

4V

41J

aWD

]ld

UOCI

JIDO

m fl

CD

Fig. 2.6

boat

quartz

light path

vacuum/gas

Absorption Chamber by Donega & Burgess Fig.2. 7

55

Fig.2.9

Trigger —01 Unit —o

Pyrolysis Flash Tube

L1

-100yF

Cap.

Garton Flash Cell

Trigger

Unit

Background Flash Tube

10pF Cap.

10 kV supply

56

slit

1 00 mm O

Gas In. To Vacuum

raphite

Teflon

Rubber

O

Skogerboe Sputtering Cell Fig. 2.8

Photolytic dissociation and atomisation has been

used successfully for atomic absorption studies in the far ultra-

violet. Garton (35) investigated the absorption spectra of flash-

vaporised solids by placing the finely divided solid sample inside

a special flash tube fired by ca. 4000J from a capacitor bank,Fig.2.9.

The radiant fluxes obtained were as high as 60kW.cm 2 causing

flash-evaporation of the sample. A second flash, delayed by 0.3m.sec,

provides the emission source against which the absorption spectrum

is measured. Using this apparatus Garton was able to monitor the

absorption spectrum of tungsten in a wavelength range extending

well into the Schumann region. Nelson and Keubler (EIG) used a

similar technique to measure the absorption spectra from many

flash-vaporised metals. Although no direct quantitative analyses

were made sensitivities in the low ppm range were anticipated.

The radio frequency induction-coupled argon plasma

and the microwave argon plasma have both been reported as being

used as atom cells for absorption studies (BI;BS) and could be

useful for vacuum ultraviolet studies. However, the high temp-

eratures and high excitation conditions present in such systems

makes them better as atomic emission devices.

The vacuum ultraviolet AAS studies reported in

this work were conducted using two types of graphite electrothermal

atomiser. The first was a small tube furnace similar to the

Massmann furnace and the second a mini-Massmann rod system.

57

THREE

58

3.1 Introduction.

Iodine is a biologically important element,

occurring in trace quantities in all body tissues as inorganic

iodide and protein-bound iodine. Typical protein-bound levels

in serum range from 0.005 to 0.04tg.m1 iodine, depending on

the functioning of the thyroid glands, (8q).

The AAS determination of iodine is accomplished

normally via an indirect determination. Thus, the ion-association

complex formed between tris (1,10-phenanthroline) Cd II and iodide

may be extracted into nitrobenzene and the cadmium determined

by AAS at its 228.8nm resonance line using an air-acetylene flame,

(90,91). In a second procedure iodide reduces Cr VI to Cr III

in acid medium, the excess Cr VI may be extracted into MIBK and

the absorbance by chromium in the aqueous layer then increases

in a linear manner with increasing iodide concentration, (n).

Alternatively, Se IV may be reduced by iodide to elemental selenium,

the solution may be filtered and the decrease in absorbance of

the selenium in the filtrate is proportional to the iodide ion

concentration in the sample solution, (92.). In acid solutions

iodate will oxidise Fell to Fe III which will extract into diethyl

ether from the acid medium and the absorbance by iron in either

solution may be determined as a measure of the iodate concentration.

Indirect determinations are often unsatisfactory

as the chemical pretreatment before AAS is affected often by the

sample matrix, eg. the solution pH, interfering ions, etc. These

methods require a detailed knowledge of the sample before analysis,

and are frequently laborious techniques.

Below 200nm iodine is the first non-metal element

59

to have a resonance line, at 183.0nm, in the Schumann ultraviolet.

An extensive survey of the spectrum of iodine has led Kiess and

Corliss (15) to list over nine-hundred lines emitted by neutral

iodine atoms in the wavelength region 119.5nm to 230.7nm. The

Grotrian diagram for iodine is illustrated in Fig.3.1 and the

transition probabilities of the more important transitions in

Table 3.1, (94.).

This chapter describes a method for the direct determination

of iodine by non-flame AAS using its resonance lines in the far

ultraviolet; 183.0nm and 173.2nm.

3.2 Apparatus-D.C. Measurement Techniques.

The instrumental arrangement employed is shown

in Fig.3.2. An iodine electrodeless discharge lamp (ELL) was

used as the line emission source. This was made from silica

tubing (i.d. 8mm, 1mm wall thickness) to form a bulb 200mm in

length containing a few milligrams of iodine and approximately

5torr of argon. The EDI, was supported in a 4-wave resonant cavity

(Model 210L, Electromedical Supplies Ltd., Wantage, U.K.) by

means of a metal holder which allowed reproducible positioning

of the lamp. The base of the cavity was sealed and a 25mm diameter

Pyrex tube fitted over the viewing port to allow efficient purging

of the cavity and optical path. Most EDL sources used for AAS

studies have bulbs typically 30 to 50mm long, enabling the whole

of the bulb to be contained within the cavity and thus achieving

thermal equilibrium over the length of the lamp. The highly

volatile nature of iodine, however, prevents such small bulbs

being used with this element. The heat generated by the discharge

causes a high vapour pressure of iodine to rapidly form inside

60

Fig. 3.1 Partial Grotrian Dia,7ram For Iodine.

1 0 -

8

6-

0)

-

179.909

184.445

187.641 178.276 206.163

183.038

2

Table 3.1 Iodine Resonance and Non-Resonance Lines.

Wavelength (nm) Transition Prob. (seC1 ) xl09

E. 1

Eic (cm 1 )

178.276 2.71 0 56093

179.909 2.11 7600 63187

183.038 0.16 0 55171

184.445 0.07 • 7600 61819

187.641 0.03 • 7600 60895

206.163 0.03 7600 56093

61

pump

IN) Variac trans- former

N21 2

source cell

m onochromator •■■•■•• .11■■•

e.h.t. readout

rn C)

FA)

O

1.3

CD

C') I)

O

Cf3

CD

7D

:71 •

the bulb causing an unstable condensed discharge. The source

was powered by a 2450MHz Microtron Mk III microwave generator

(EMS Ltd., U.K.), using a reflected-power meter (EDT Research

Ltd., London W.4.) connected in series between the generator and

cavity to allow the lamp and cavity assembly to be tuned to minimum

reflected power. The discharge was initiated with a Testa coil

and the optimum applied power found to be 15watts; with a reflected

power reading of near zero. At higher applied powers the cavity

and lamp became appreciably warmer and the iodine inside the lamp

was seen to condense onto the wall of the silica tubing not contained

in the cavity. At powers of 60 to 70watts the discharge became

visibly unstable and the iodine line emission intensity much

reduced.

Radiation from the source was focussed into the graphite

tube atomiser by a biconvex silica lens (25mm diameter, 80mm focal

length). The radiation emerging from the atomiser was brought

then to a focus on the entrance slit of the monochromator by a

similar silica lens. The optical path was formed from glass

tubing (25mm i.d.) placed between the source and graphite atomiser

and between atomiser and monochromator entrance slit; this path

was maintained under a slightly positive pressure of oxygen-free

nitrogen to remove atmospheric oxygen, water vapour, etc. Each

lens was positioned inside the glass tubing using polythene holders

and the complete optical path, including the source and atomiser,

was mounted onto a one-metre optical bar to provide normal incidence

at the monochromator entrance slit.

A one-metre, normal incidence, concave grating, vacuum

monochromator (type E760, Rank Precision Instruments Ltd., U.K.)

63

was used for these studies, Fig.3.3. The groove separation of

the concave grating used was 1.67)1m and provided a working range

of 30 to 500nm in the first order, a dispersion rate (plate factor)

of 1.6nm.mm-1 and theoretical resolving power of approximately

30,000. Its effective numerical aperture ratio was f/12. The

monochromator was operated in conjunction with a rotary roughing

and backing pump and an oil diffusion pump, Fig.3.4. Pressures

of less than 10 5torr were possible using this arrangement, although

for those studies at wavelengths greater than 175nm an operating

pressure of ca. 0.ltorr was sufficient to provide good transmission

of radiation. The positions of both exit and entrance slits were

fixed, and wavelength scanning was controlled either manually or

by a motor driving through a gear box. A nine-speed gearbox was

fitted providing nominal forward and reverse scanning speeds of

0.05, 0.1, 0.2, 0.5, 1.0, 2.0, 5.0, 10.0 and 50.0nm.sec-1. A

25mm diameter, 3mm thick, calcium fluoride window was mounted

across the entrance slit of the monochromator to retain the low

pressure. A vacuum tight flange was constructed from aluminium

alloy and fitted to the monochromator exit slit. The photomultiplier

tube housing was bolted to this flange, holding the pmt such that

its window provided a vacuum-tight seal against a rubber 0-ring

recessed into the flange. The photomultiplier tube employed was

a Model 6256B, 50mm diameter silica end-window tube, with an

effective cathode diameter of 10mm, (EMI Electronics Ltd., U.K.).

This was a 13-stage venetian blind system of very high gain, ca.

10g. The photomultiplier tube was operated with a Brandenburg,

Model 475R, EHT unit at the voltage which gave the best signal to

noise ratio with. the radiation levels available; about 1600V.

64

P gauge

••■••■■••

monochromator valve

P gauge

valve

backing pump

diff. pump

air valve

Pumping System

Fig. 3. 4

66

The graphite tube atom cell employed is shown in

Fig.3.5 and Fig3.6 and is similar to that described by Dagnall,

Johnson and West, (15). A high purity graphite tube of 6mm o.d.,

4mm i.d. and 20mm in length was machined from graphite rod (Grade

RW003, Ringsdorf Company Ltd.). A sample introduction port was

provided by a 2mm diameter hole drilled through the wall of the

tube. The graphite tube was supported by a graphite split-ring

at each end and the assembly was then retained by two stainless-

steel L-shaped holders. One holder was secured to a stainless-

steel support column and the second rested on a second column

and was earthed via a length of copper braiding attached to the

base of the unit. One holder was thus allowed to 'float' to

simplify the alignment procedure and prevent fractures in the

graphite tube owing to its expansion on heating. Both support

columns were water cooled.

The graphite tube was heated electrically; power

was supplied via a Variac transformer (20A Claude Lyons Ltd, U.K.)

and a large power transformer (rating 10kW, Foster Transformers Ltd.,

U.K.) to provide 1CV and over 500A across the furnace. The entire

cell was maintained under a constant flow of oxygen-free nitrogen

or argon by means of a glass housing fitted with a removable cover

to allow the sample to be transferred to the tube, and silica

windows (1mm thickness) to permit transmission of the source

radiation. The atom cell was inert -gas purged by a supply separate

from that used for the optical system, so that the flow-rate

could be adjusted without affecting the transmission of the optical

path.

The output signal from the photomultiplier tube

67

steel electrod

support

graphite tube

split

ring .--2 0 mm---■

graphite tube

g electrode

I I 1: base

water 20mm

water

0

0

I ; 44. : ;

4... 1 j I I

.9mm- --6m m•— • shield gas

15mm

Fig. 3.5 Fig. 3.6

was led to either a potentiometric chart recorder (RE511, Servoscribe

Ltd., U.K.) if spectra were being recorded or to a storage type

oscilloscope (DM64, Telequipment Ltd., U.K.) if transient absorption

signals were being monitored.

All real signals carry an unwanted component

termed noise. This serves to produce variations in the measured

analytical signal and it is necessary to extract the signal from

this noise as efficiently as possible. For a d.c. signal, such

as the EDL emission used in these studies, the major noise component

is produced by a time variation in the emission intensity and the

noise inherent in the photomultiplier tube radiation measuring

system, (96,97,98). The d.c. signal may be smoothed by using a

long readout time-constant or by shunting a capacitor across the

voltage output terminals of the detector. This capacitance smoothing

technique serves to filter from the required signal all frequencies

higher than that of the zero frequency of the d.c. signal. This

procedure necessitates a compromise between noise reduction and

the effective time-constant of the measuring system. Thus, an

increase in capacitance increases the efficiency of filtering

and hence the signal-to-noise ratio but increases the instrumental

time-constant, i.e. the time required for a signal to be measured.

The use of non-flame atomisers for AAS implies relatively short

atomic residence times (typically 0.1 to 5.Osec) in the optical

path and this limits the capacitance which may be applied to smooth

the signal. The iodine AAS studies described here were conducted

with a 0.41F capacitor across the signal input terminals of the

oscilloscope having an impedance of 114 ; providing for a time-

constant of 200m.sec. This capacitance value was determined by

69

trial and error as providing maximum signal-to-noise ratio without

overdamping the transient absorption signals.

3.3 Iodine AAS Results.

The apparatus was assembled as described

- . above. Aqueous stock solutions of 10,000pg.m1 1 Iodine as potassium

iodide, iodate and periodate and ammonium iodide were prepared

using analytical grade reagents and distilled water. These solutions

were diluted for use as necessary with distilled water.

The line emission spectrum from the iodine

EDL operated at 15watts applied microwave power in the wavelength

range 175nm to 210nm with monochromator slit widths of 1091m is

shown in Fig.3.7. The signal-to-noise ratio and relative intensity

of each line is tabulated. It was observed that the relative

intensities of the resonance lines and non-resonance lines altered

by upto approximately 25% from day to day although the stability

of each line remained within 5 to 10% of its mean value over an

operating-time period of eight hours. The change in relative

intensities appeared to be related to the repositioning of the

EDL in the resonant cavity. The cavity and optical path were

purged of atmospheric oxygen using a positive flow of oxygen-

free nitrogen at 11.min 1. The atomiser chamber was purged independently

again using 11.min-1 of nitrogen.

The potassium iodide stock solution was

diluted using distilled water to provide a working solution containing

50.11g.m1-1I. A icAl aliquot was applied to the graphite atomiser

using an Eppendorf micropipette and dried using a Variac setting

of 1.5V for about one minute; the removal of water vapour from

70

I2 -Ar EDL

Ig.3.7

183-Onm Fig.3.8

cn

Abso

rban

ce

I I I 20 40 60

ng I

71,

0 80 100

1.0—

0.8—

0.6—

0.4—

0.2—

N H4 I

KI

-n

the cell could be monitored as an increase in the transmission

of the source radiation as observed on the storage oscilloscope

screen. The power to the furnace was switched off and the Variac

set at 8V, the sample was atomised at 8V for 3 to 4sec. This

procedure was conducted at each line between 175nm and 200nm and

absorption was only observed at the two iodine resonance lines;

50% and 75% absorption at the 183.Onm and 178.2nm lines respectively.

This served to indicate that atomic absorption by iodine atoms

produced in the furnace was occurring. Using 10sl aliquots of

— 1 to 19.1g.m1

1 iodine as aqueous potassium iodide the system was

optimised for peak atomic absorption measurements at the iodine

resonance line at 183.Onm. An absorbance vs iodine concentration

calibration graph plot was constructed, Fig.3.8. The absorbance

varied with iodine concentration but deviated from linearity at

approximately 0.2 absorbance unit. The linear working range was

limited by this curving—off towards the concentration axis and

a means of raising this point of deviation to higher absorbance

values was sought.

The factors responsible for bending of analytical

curves may be classified into two groups. The first concerns

faults in the apparatus employed. The second group includes

factors related to the spectral line itself, eg. hyperfine structure,

the ratio of the absorption and emission line profiles and resonance

broadening and line shift in the absorbing medium. The instrumental

arrangement employed was examined and possible optical or apparatus

faults investigated.

It was observed that the voltages applied

across the furnace to achieve atomisation of the iodide sample

72

caused emission from the hot tube to register at the photomultiplier

tube. The variation of this furnace emission intensity at 183.0nm

with applied voltage is shown in Fig.3.9, and its wavelength

dependence in Fig.3.10. As discussed in Chapter One the proportion

of radiation emitted in the far ultraviolet by a blackbody is very

small for temperatures below 5000K. That the radiation detected

by the PT was not of wavelengths shorter than 200nm was demonstrated

by the emission intensity being independent of whether the optical

path between the furnace and the monochromator was nitrogen

purged or not. The radiation incident at the FMT was attributed

to scattered light inside the monochromator. The emission was

not detectable on the 183.Onm line over the EDL emission, its

presence was shown by stopping the radiation from the DL or

changing wavelength. The emission signal was not observed until

approximately 1.5sec after the atomisation power (7 to 9V) was

applied. The bending of the iodine analytical curve was imputed

to the high temperatures necessary to volatilise and decompose

KI (B.Pt. 1493K) and possible interference from the hot tube emission

causing a depression of the atomic iodine absorption peak heights.

Ammonium iodide has a sublimation temperature of 824K and was not

expected to suffer from this high temperature interference.

Working solutions of 1 to 10pg.m1 1 I as ammonium iodide were

prepared and the absorbance vs iodine concentration calibration

curve plotted using the same instrumental parameters as for KI.

The analytical curve obtained using ammonium iodide solutions is

compared with that from KI solutions in Fig.3.8. The two calibration

curves were identical at low concentrations but the ammonium

iodide plot had a much longer linear range; linearity was observed

73

Optimisation For Iodine AAJ D.C. System.

10—

4- o tn (r) 2- .E

0 I I I I 7.0 7.5 8.0 8.5 9.0. 9.5

Voltage across tube

Fig. 3.9--

10—

8 _

6 — c 0 4— En

2—

0 200 300

Wavelength, nm

400

Fig. 3.10

74

to approximately 0.7 absorbance unit, covering a working range

of about orders of magnitude (0.3 to 6.04g.m1-1). The

sensitivity for 1% absorption was 0.04g.m1-1I (4 1x10 °gI).

Two methods of eliminating the effects of the

emission from the hot furnace were considered. Johnson et al.

used the technique of limited-field viewingin which a small aperture

was placed across the exit of the tube. This prevented the radiation

from the atomiser reaching the monochromator but limited the

aperture of the optical path and caused a slight reduction in

source emission intensity, i.e. worsening the signal-to-noise

ratio and, hence, the detection limit of the technique. The

second method examined was modulation of the ;DL emission and the

apparatus was modified to achieve this.

3.4 Apparatus - Mechanical Modulation and A.C. Measurement Techniques.

The smoothing technique mentioned previously

for increasing the signal-to-noise ratio is a simple and useful

signal treatment procedure. However, most noise has a strong

1/f component (flicker noise), in which the power of the noise

is proportional to the reciprocal of the frequency, and is common

in almost all instrumental systems. Therefore, it is usually

advantageous to locate the desired signal away from zero frequency,

i.e. d.c., where 1/f noise is greatest. This is achieved by

imposing the signal informatioh on an a.c. wave at a desired

frequency, (110). This process is called modulation.

The 3L emission intensity may be modulated

by a radiation chopper at a frequency fo and the signal be detected

at this frequency. This not only minimises the 1/f noise but

increases the filtering efficiency as less noise is included in

75

the filter bandwidth. To serve as the filter and eliminate frequency

drift a lock-in or phase-sensitive amplifier is employed. Phase-

sensitive detection, occasionally referred to as synchronous

rectification, is a method of demodulating an a.c. signal such

that only signals in phase with a reference signal are extracted

from the modulated signal. The majority of p.s.d. systems operate

on the principle of the synchronous switch. With reference to

Fig.3.11(a), the signal is periodically switched into the load

resistor at a frequency and phase determined by the reference

voltage. If the signal and reference voltages have the same frequency

and phase the voltage across the load resistor will have the form

shown in Fig. 3.12(a). The switch acts as a half-wave rectifier

for signals of the same frequency and phase as the reference.

In practice a double-switching arrangement is employed, Fig.3.11(b);

this balanced two-way switch provides full-wave rectification,

Fig.3.13, improving the signal to noise ratio.

If the signal is out-of-phase with the reference

the waveform will be of the form shown in Fig.3.12(b), and

rectification will not occur. The response of the system to

noise on the signal depends on the time constant (T = RC) of the

d.c. measuring circuits, and the effective bandpass (ilf) of the

p.s.d. is given by,

ilf = _1- = 1

... 3.1 4RC 41-

To enhance the signal-to-noise ratio and

eliminate the d.c. emission signal from the heated atomiser the

emission from the EDL was modulated. Because of the transient

nature of the atomic absorption signal, as high a chopping frequency

76

V2

V

ref.voltage

signal voltage

V1

Vi -V2 b) balanced switch

Fig. 3.11

Fig. 3.13

Fig. 3.12

ref. voltage

signal voltage

load to

resistor d.c. recorder

a) single switch

time

a) in phase b) out of phase

as possible was used. The best available was a 285Hz Techtron

chopper (Techtron Prod. Pty., Aust.), modified to match the phase-

sensitive detector used. The modified apparatus is shown in

Fig.3.14. The mechanical chopper was placed between the resonant

cavity and the first lens. To achieve optimum signal-to-noise

ratio of the EDL emission the radiation should have been focussed

at the chopper blades to obtain as sharp a modulation cut as

possible; the radiation then focussed again into the graphite

atomiser. This arrangement would have necessitated a lens between

the EDL and the chopper, increasing the optical path length. The

purging efficiency was found to decrease with increasing path

length due to entrained oxygen along the optical path, and the

signal-to-noise ratio made worse. The chopped radiation was

unfocussed therefore. The reference signal was provided by a

12V lamp irradiating a phototransistor, through the rotating

chopper blades. The 285Hz a.c. reference signal from the photo-

transistor was led to the reference channel input of the Model

411 phase-sensitive detector (Brookdeal Electronics Ltd.,U.K.).

The EDL emission was focussed, after the rotating sector, into

the graphite atomiser and then again onto the entrance slit of

the monochromator using 3mm thick calcium fluoride lenses, of

focal length 50mm, shortening the optical path still further.

From the PMT the signal was led to a Model 450 low-noise amplifier

(Brookdeal Electronics) where it was amplified to provide an

output of 1V r.m.s.„ in the frequency range 100 to 1000Hz. This

signal was passed to the p.s.d. which selected the component in-

phase and of the same frequency as the reference signal, and

amplified the rectified signal obtained to produce an output of

78

Variac transformer

amp. e.h.t

mon ochromator source

choppe r

p.s.di

(readout pump

Cn

cf-

tn

b"

Cl :3" 0

CD i-s

18.

approximately 10V d.c. The d.c. signal was led to the storage

oscilloscope as before.

3.5 Iodine AAS Results.

The optimum operating conditions for the

determination of iodine with the graphite furnace and instrumental

assembly described were established via atomisation and measurement

of the peak absorption obtained at 183.Onm with 1011p1 aliquots

of aqueous potassium iodide solution. Fig.3.15(a) shows the

variation in absorbance at 183.Onm for 2pg.m1-1 of iodine with

variation in voltage applied across the graphite furnace. All

measurements were made with an applied voltage of 8.5V. Fig.3.15(b)

shows the variation in absorbance at 183.Onm with flow-rate of

nitrogen in the furnace chamber. A nitrogen flow-rate as high

as 1.51.min 1 could be used before any decrease in absorbance

was detected in the atomisation of 1qpi1 of 2yg.m1 1 iodine solution.

Above this flow-rate, cooling and dilution effects were observed

and the peak absorbance decreased and became less reproducible.

-1 When a nitrogen flow-rate of less than 0.51.min was used the

purging time required between samples and between the drying and

atomisation steps was lengthened considerably. A flow-rate of

11.min 1 was maintained in all further work.

The exit and entrance slits of the mono-

chromator were set at 30pm (spectral bandpass 0.05nm) and a PMT

voltage of 1620V was employed. These settings were found to give

optimum signal-to-noise and signal-to-background intensity ratios.

The photomultiplier output was led to the amplifier; adjusted to

produce a signal for the phase-sensitive detector such that its

d.c. output voltage was 10V at 100% transmission. The variable

80

time-constant of the phase-sensitive detector was optimised to

produce the best signal to noise ratio without distortion of the

analytical signal by overdamping. The value of the time-constant

used was 300msec.

With the optimum operating conditions established,

calibration graphs for iodine atomic absorption at 123.0nm were

constructed, Fig.3.16; the samples being introduced as aqueous

solutions of potassium iodide, ammonium iodide, potassium iodate

and potassium periodate. Calibration, graphs of identical slope

and linear range were obtained in each case; 10111 samples produced

calibration graphs linear up to 4,tg.m1-1 and the sensitivity

(for 1% absorption) was 0.041g.m1-1 iodine, i.e. 4 x 10 1°g I.

For the 178.2nm iodine resonance line the sensitivity (for 1%

absorption) was found to be better by a factor of two,i.e. 0.02yz.m1

in a 1Cyl sample; an absolub detection limit of 2 x 10 1°g I. -

Although higher sensitivity was obtained at 178.2nm the recorded

emission intensity from the source was lower at this line than

at 183.0nm; the precision of the measurement was thus lower and

a similar practical detection limit of 2 x 10 9g I was observed.

The absorption signals recorded were wholly atomic in nature; no

non-specific absorption from aqueous sample solutions was observed

at the 179.9nm, 184.4nm and 187.6nm iodine non-resonance lines

-1 from the source when 10p1 aliquots of 241g.m1 iodide solutions

were atomised.

A typical drying, purging and atomisation cycle

at 183.Onm for an iodine containing sample is illustrated in

Fig.3.17. This trace was recorded using the potentiometric chart

recorder. The operation may be summarised as follows.

81

0.3—

0

a)

Optimisation For Iodine AAS At 183.0nm.

7.0 7.5 8.0 8.5 Voltage across tube

9.0

b) a) U 0C 0.2— -o C5 (n _o

0.1

0 0.5 1.0 1.5 2.0 2.5

N2 f low through chamber ( 1.min1)

Fig. 3.15 82

83

0 1 2 3 time (min)

60 80 100

sample introduced 100%

absorption

atomisation

4 5

Fig. 3.17 Analysis Cycle For Iodine At 183.0nm.

A bs

orba

nce

Fig.3.16 Iodine A.S.

178.2nm 183.0nm

m CD (A)

1) t = Osec, the lid of the atomisation.chamijer was romovd

and a lop' sample introduced into the furnace. 2) t = 10sec, with a voltage of 1 to2V across the tube, the •

sample was dried.

3) t = 9Osec, the power to the furnace was switched off and

the chamber lid replaced. Purging of the

chamber continued until the known 100% trans-

mission level was achieved.

4) t = ca. 180sec, the voltage across the tube was set to 8.5V

and the sample atomised at this voltage for

2 to 3 sec.

5) t = 183sec, the furnace was allowed to cool to ambient

temperature.

6) t = ca. 250 a second sample was examined.

to 300sec,

3.6 The Effect of Foreign Ions.

In many sample types in which the determination

of traces of iodide is of interest the predominating inorganic

species present from the matrix are the salts of the alkali metal

and alkaline earth ions. Thus, for example, in biological and

botanical samples, foodstuffs and water, large excesses of sodium

and potassium (and frequently calcium and magnesium) are present.

After sample pretreatment to destroy organic matter by wet oxidation,

these ions may be present in conjunction with one or more common

anions such as chloride, sulphate, nitrate or phosphate. The effect

on the iodine determination of different weights of several common

inorganic salts likely to be present after sample pretreatment was

investigated.

84

Table 3.2 TJ:ffect of Common Inorganic Salts On Iodine Absorption.

Salt Concentration 183.0nm

Studied ratio, Change in Increase in Absorbance

MX MX:iodide absorbance,% absorbance at 184.4nm

NaC1

KC1

Na2SO4

1:1 0 0 0

10:1 +25 0.05 0.05

100:1 +100 0.22 0.22

1000:1 100% absorption of radiation

1:1 0 0 0

10:1 0 0 0

100:1 +40 0.10 0.10

1000:1 1005 absorption of radiation

1:1 0 0

10:1 0 0

0

100:1 +40 0.10 0.10

1000:1 1005 absorption of radiation

NaNO3

1:1 0 0 0

10:1 0 0 0

100:1 +15 0.03 0.03

1000:1 +200 0.40 0.40

Na2HPO4 1:1 0 0 0

10:1 +25 0.05 0.05

100:1 +75 0.15 0.15

1000:1 1005 absorption of radiation

85

The following compounds were examined at weight ratios

of 1:1, 10:1, 100:1 and 1000:1 with 2pg.m1 1 iodine as aqueous

potassium iodide; potassium chloride, sodium chloride, sodium

nitrate, sodium mono-hydrogen phosphate and sodium sulphate.

The effects on the recorded iodine absorbance values at 183.0nm

are shown in Table 3.2 and may be summarised as follows.

a) At 1:1 no species was observed to interfere.

b) At 10:1, potassium chloride, sodium nitrate and sodium sulphate

had no effect. Sodium chloride and sodium phosphate gave the

same enhancement; +25% of the absorbance signal.

c) At 100:1 all salts showed an enhancement of the absorbance

signal and the relative effect was:

NaNO3, Na2SO4,KCl, Na2HP04111aC1

Increasing enhancement

d) At 1000:1, all salts absorbed 100% of the incident radiation,

with the exception of sodium nitrate which increased the absorbance

value by 200%.

The positive interference effects caused by these

foreign ions were attributed wholly to non-specific absorption

at the 183.0nm line by molecular species. This was demonstrated

by measurement of the absorption at the nearby 184.4nm iodine

non-resonance line for similar sample solutions. In each case

the same absorbance was obtained at this line as that at the

183.0nm line for blank solutions containing only the salt investigated,

and this was the same as the increase in absorbance recorded at

this wavelength for iodide solutions when the foreign salt was

added. It was possible, therefore, to correct for the non-specific

86

absorption interference observed in the presence of moderate

concentrations of the salts investigated, by subtraction of the

absorbance obtained at 184.4nm from that at 183.0nm. In the

presence of 1C00-fold weight excesses of most of the salts studied,

however, virtually complete absorption of incident radiation at

both lines was obtained and such a correction could not be made.

The exception to this was sodium nitrate. Conversion of alkali

metal salts into the nitrates before atomisation would thus be

preferable in methods developed for examination of biological

and botanical samples.

A more detailed discussion as to the nature

of the observed non-specific background absorption observed with

simple inorganic salts in non-flame AAS is presented in Chapter

Six.

87

FOUR

88

4.1 Introduction.

Sulphur , like iodine, is determined frequently

by indirect AAS procedures. Following oxidation of the sample,

converting all sulphur present to sulphate, barium chloride is

added. The precipitate of barium sulphate may be dissolved in

EDTA solution and the concentration of barium determined by AAS

at 553.5nm in air-acetylene flame. Alternatively, a known excess

of Ba2+ may be added to the oxidised sample solution and the remaining

2+ . Ba in solution determined by AAS. These methods have been used

to determine sulphur in urine and other biological tissues, after

oxidation, (101,102). An inhibition titration method for sulphate

determination has been developed by Looyenga et al. (103). The

sulphate solution is titrated, at a constant delivery rate, with

magnesium chloride solution. The titrated solution is simultaneously

nebulised into an air-hydrogen flame. The atomic absorption of

Mg is inhibited by its precipitation as the sulphate until the

end-point of the titration is reached, when the Mg absorbance

rises rapidly. Concentrations as low as 14g.ml-1 sulphate may

be determined by this technique. Alkali metals and commonly

occuring anions do not interfere at the 0.1M level although many

other metals, phosphate and silicate must be removed. Sulphur

dioxide may be determined in a similar manner; it is oxidised by

hydrogen peroxide to the sulphate which may be determined by the

addition of a known excess of barium or magnesium ions as discussed

above, or lead nitrate and the unprecipitated lead determined

by AAS at 283.3nm, (10*). Sulphite may be determined by adding

the sulphite sample to a suspension of mercury (II) oxide leading

to the formation of the soluble Hg(S03)22 complex, (10S). The

89

amount of oxide dissolved may be monitored at the 253.7nm Hg

resonance line and is proportional to the sulphite concentration.

Salet (tob), in 1869, observed that an aerosol

of a sulphur compound when injected into a low temperature, fuel-

rich hydrogen flame produced an intense blue emission if a cold

object was placed near the cone of the flame. This blue emission

is due to molecular S2, excited by chemiluminescent reactions in

cool flames. Veillon et al. (107) have reviewed the history of

this phenomenon as an analytical technique for the determination

of sulphur. Using a cold glass sheath around an air-hydrogen

flame Veillon has monitored the S2 emission spectrum in the wavelength

range 320 to I1Cnm, and used the emission maximum at 384nm to

determine sulphur, at the ppm level, in oils. Dagnall et al. (10B)

have used a similar technique for determining sulphur in organic

and aqueous matrices by employing a simple filter photometer.

Everett (OM) has described the use of a carbon filament reservoir

resistively heated in a hydrogen-nitrogen diffusion flame for

the determination of sulphur via this emission. ly1 of the sulphur

containing solution was applied to the filament and dried by

heating the filament at a low voltage. The flame was lit and

the sample evaporated into the cool flame; the S2 emission was

recorded at 394nm. Townshend et al. (!10) have examined S2 emission

by the technique of molecular emission cavity analysis (MECA).

The sample is contained, within a small cavity positioned in an

air-hydrogen flame and the emission from excited S2 molecules

again monitored in the wavelength. range 310nm to 430nm.

This chapter is concerned with the direct determination

of sulphur by non-flame AAS using its resonance and near-resonance

90

transitions in the Schumann ultraviolet. The graphite tube atom

cell described in Chapter Three was used for these studies. The

Grotrian diagram for sulphur and data relating to the analytically.

useful lines is presented in Fig.4.1.

4.2 Apparatus - Electronic Modulation and A.C. Measurement Techniques.

The apparatus employed for the direct determination

of sulphur by AAS is shown schematically in Fig.4.2.

The source used was a sulphur containing al.

Dagnall et al. (111 ) have described the preparation and use of EDL

tubes containing organo-sulphur and phosphorus compounds for the

qualitative identification of these elements in organic matrices.

Marshall (1M), for the determination of sulphur by AAS in a nitrogen-

separated nitrous oxide-acetylene flame at 180.7nm, used an EDL

source containing ca. 1mg of sublimed sulphur and 3torr of argon

in a 60mm fused silica bulb. For the studies described here a

variety of sulphur-containing compounds were examined in EDL

tubes as possible line emission sources for sulphur AAS. Sulphur

dioxide and hydrogen sulphide were investigated at ratios of

1:10 and 1:1 with 3torr of argon in EDL tubes of 25mm to 30mm

length. A lamp containing a few milligrams of sublimed elemental

sulphur with about 5torr argon, similar to that used by Marshall,

was also examined. All the tubes were prepared in the normal

manner. They were operated in the 1-wave resonant cavity and all

emitted the analytically useful lines between 180nm and 183nm.

To evaluate the performance of each lamp and determine the lamp

having the best signal-to-noise and signal-to-background ratios these

sources were electronically modulated.

91

Fig. 4.1 Partial Grotrian Dia:ram For Sulphur.

10-

6- 178.225

4- 180.731

182.037

182.625

2

0

Wavelength (run)

178.225

180.731

• 182.037

182.625

190.03

Transition Prob. (sec 1 ) s log

S. Ek

1 .50 22181 78290

4.10 0 55331

2.20 397 55331

0.73 574 55331

0.00 o 52624

92

signal generator

p.s.d amp e.h.t.

Variac transformer microwave generator

r--

monochromator

pump readout'

rf:J

Fl

CF RI C

c-+

CD 0 c+ 0

ci

0

C)

r o.)

(_)

to 0

C)

t-1

In Chapter Three a mechanically modulated-emission

source AAS instrumental arrangement was described. Whilst mechanical

modulation provides the required a.c. source emission intensity,

several disadvantages exist with chopping radiation. The presence

of a rotating sector in the optical path between source and atomiser

cell may limit the aperture of the instrument, decreasing the

signal-to-noise ratio. Most mechanical choppers are difficult

to purge of atmospheric oxygen and for non-metal AAS, at wavelengths

below 200nm, this increases the noise of the system. Mechanical

modulation is confined to a single frequency usually, characteristic

of the motor employed to drive the rotating sector; variable

speed choppers are usually bulky and very expensive. Thus,

optimisation of modulation frequency, to operate away from mains

frequency or other interfering signals, is not possible.

Electronic modulation, with a tuned a.c. detecting

system, does not suffer these disadvantages and greatly simplifies

the instrumental arrangement. Phillips (it3) has described the

electronic modulation of flow-through microwave plasmas for photo-

chemical kinetic studies. He obtained 10C% modulation of the

Lyman-ochydrogen line at frequencies upto 150kHz and modulated

iodine in argon plasmas and mercury in helium plasmas at upto

110kHz. Source emission intensity modulation for AAS or AFS is

at much lower frequencies. Browner and Dagnall 0110 examined

the effects of electronic modulation on EDL sources for the AAS

of a variety of metals. The power output to the lamps from the

microwave generator was controlled by varying the anode voltage

of the magnetron valve. When the generator used by Dagnall et al.

(Microtron 200, E.M.S. Ltd.) was operated in the modulated mode,

94

a 50Hz modulation signal was introduced into the anode circuit of

the magnetron by means of a modulation transformer. This super-

imposed a 50Hz component on the d.c. potential of the anode. It

was demonstrated that electronic modulation had little effect

on the behaviour of the EDL sources except to improve the stability

and signal-to-noise ratio of the emission. Silvester (0%5) used

electronically modulated EDL tubes for AFS studies and investigated

the effects of modulation frequency and the shape of the imposed

waveform. Maximum fluorescence signal was observed with a square-

wave modulation frequency of approximately 20kHz.

The electronic modulation apparatus used in

the studies on sulphur is shown in Fig.4.2. The Microtron

microwave generator used in this work contained an error-sensing

circuit around the magnetron valve; this served to stabilise the

power output from any fluctuations in the mains supply. Imposing

an a.c. signal into this error-sensing circuit caused the output

from the magnetron valve to follow this modulation signal. With

the Mark III generator it was only necessary to supply the a.c.

signal to the input socket provided for this purpose. The generator

thus supplied output power containing an a.c. component whose

frequency, size and waveform of modulation depended upon the

a.c. voltage applied. The modulation control in the generator

allowed the degree of modulation to be varied between 0 and 80.

In all studies maximum (80) modulation was used.

MD provide the a.c. voltage to the microwave

generator an AIN supply (Aim Electronics, U.K.) was employed.

This consisted of a clock generator unit (CGU 102),pulse-width

delay unit (PU]) 103), rise and fall unit (VHF 107) and power

05

supply unit (PSU 101). The use of this system enabled various

signal parameters to be examined and the best signal in terms

of lamp emission signal-to-noise ratio was found to be a 1000Hz,

square-wavesignal of mark-to-space ratio of 1:1 and 20V peak-to-peak.

This signal was led to the microwave generator and to the reference

input of the phase-sensitive detector.

For electronic modulation of EDL sources

to be effective the lamp must be capable of responding to the

applied a.c. power, i.e. the intensity of the source emission

should follow the modulation signal. The ability of the lamp

to react to the modulated microwave power is best determined by

trial and error, although a guide to its performance may be obtained

froM the emission intensity vs applied microwave power curve.

Fig.4.3 shows the variation in emission intensity of the 1:1

hydrogen sulphide - argon EDL at the sulphur resonance line at

180.7nm with different applied microwave powers. The effect of

a modulated microwave power signal centred around 45watts was

to produce an a.c. intensity output from the EDL. This was

confirmed in practice. The other sulphur atomic transitions of

interest in the Schumann ultraviolet behaved in a similar manner,

Fig.4.4. Those lamps containing sulphur as sulphur dioxide and

elemental sulphur also responded to the modulated microwave power.

The spectra from each source was recorded using an instrumental

time-constant of 100m.sec and the signal-to-noise and signal-to-

background ratios measured for each of the three analytical lines;

180.7nm, 182.0nm and 182.6nm. The results are shown in Table 4.1

and the 1:1 hydrogen sulphide and argon lamp selected as the best

for sulphur AAS studies.

96

10— H2S/Ar EDL 180.7nm

C a) 4-

6- co Xs

LAO 0 4- Ti)

input 'modulation

I I I I 20 40 60 80

Applied power (watts)

2—

0

0

CI.0 . component of emission

100

Fig. 4.3

Fig. 4.4

H2S/Ar EDL 10—

8- U) c -71 W

UD 6—

4 — 0

cr 2—

182.6nm

182.0nm

I I I 0 2.0 40 60 80 100

Applied power (watts

97

Fig. J.5 Emission Spectrum From H S.Ar 2

182.625

182.037

180.731

190.03

178.225

I I I i 200 190 180 170

Wavelength ( nm)

Wavelength (nm) Transition

180.731 3501 - 3P2

182.037 3501 - 3P1

182.625 350

- 3P0 0

190.03 5502 - 3P2

98

The remainder of the apparatus, calcium fluoride

lenses, atom cell, monochromator and signal readout, were similar

to that described in Chapter Three for the direct determination

of iodine.

Stock solutions (10,000pg.m1-1 S) were prepared

from thiourea, potassium thiocyanate, potassium sulphate and

sodium sulphate. These soluticns were diluted, using distilled

water,as necessary.

4.3 Sulphur AAS Results.

Aqueous solutions of sulphur (25pg.m1-1 S) as

the sulphate and thiocyanate of potassium and as thiourea were

prepared from the stock solutions. Using a Finipipette micropipette

1.11 aliquots of each solution were examined for sulphur atomic

absorption at 180.7nm, 182.0nm and 182.6nm, using the graphite

tube atomiser. Each aliquot was pipetted into the furnace, dried

at 1.5V for 30 to 60sec and then atomised at 8.0V across the

graphite tube. Absorption was recorded at all three sulphur lines;

0.3, 0.2 and 0.1 absorbance unit at 180.7nm, 182.0nm and 182.6nm

respectively. The absorbance was independent of the nature of

the sulphur containing species. To check for non-specific absorption

identical samples were examined at the 179.9nm and 183.0nm iodine

lines. No absorption was observed at these wavelengths.

A monochromator entrance and'exit slit width

of 30ym was found to provide for maximum signal-to-noise ratio

and maximum absorption at 180.7nm from sulphur containing samples

and this slit width was used for all sulphur AAS studies. The

apparatus was then optimised for maximum absorbance using 2pg.m1 1

99

sulphur as K2SO4

at 180.7nm. The effect of nitrogen flow-rate

through the furnace chamber is shown in Fig.4.6 and the effect

of applied voltage across the graphite tube in Fig.4.7. All

further studies were undertaken with a chamber nitrogen flow-

rate rate of 11.min and furnace applied voltage of 9.5V. It was

observed that at this applied voltage a background absorption

signal from the graphite furnace was present occasionally and

was often as great as 20, absorption with some tubes. Because

of distortion of the L-shaped tube holders it was found necessary

to replace these more often for sulphur determinations than with

the iodine studies. This was attributed to the higher voltages

necessary to form atomic sulphur in the graphite furnace. Calibration

curves of absorbance vs sulphur concentration at 180.7nm, 182.Onm

and 182.6nm were determined, however, and are presented in Fig.4.8.

These results are summarised in Table 4.2.

The calibration curves obtained at 180.7nm

and 182.0nm bend towards the concentration axis at absorbance

values of approximately 0.25. This was shown not to be due to

emission interference from the heated tube, and decreasing the

time constant to 30m.sec had no effect on increasing the linear

range of the analytical curves, although the signal-to-noise

ratio was severely decreased. Sulphur AAS at 182.6nm was less

reproducible than at the other two lines examined and at sulphur

concentrations above 15Ong S evidence of non-specific absorption

was observed by checking at the 183.0nm iodine line.

Although only 180.7nm is a resonance line

for sulphur, the observation of sulphur atomic absorption at

182.0nm and 182.6nm may be explained by the low energy of the

100

OSI. OOL S

S 6u

9.Z21,

(net. L.081,

I- 0 1.

Absorbance N

Absorbance EL.)

I O

3 5- Ni- 1

•••■■■■•

GJ-

-n O

r

O

(0-

-n (r)

O

angdIn 8'47 *ZTA

co

Table 4.1 Characteristics of Sulphur 172)L Sources.

S (element)

S:N S:B

Wavelength (nm)

S:N

H2S . . .

S:B : . S:N

SO2

S:B

. . . • : .

180.7 40:1 100:1 : 28:1 100:1 : 30:1 100:1 •

182.0 60:1 100:1 : 45:1. 100:1 : 50:1 100:1

182.6 100:1 100:1 : 60:1 100:1 : 90:1 100:1

S:N Signal-to-noise ratio

S:B Signal-to-background ratio

Table 4.2 Sulphur AAS Results With The Graphite Tube Furnace.

Wavelength (nm) Sensitivity (ng) Det. Limit(ng) Linear Range (ng)

180.7

182.0

182.6

1.2

1.8

4.8

6.0

6.0

5.0

to 75

to 90

to ca. 150

* for 1% absorption

102

of the lower states for these transitions; 398cm-1 and 573cm-1

respectively. At the temperatures attained in the graphite

atomiser these states would be populated to an appreciable extent.

4.4 Discussion - Faults With The Simple Graphite Tube Atomiser.

The problems associated with using this graphite

furnace for sulphur atomic absorption studies may be summarised

as follows:

1) Higher temperatures were necessary to produce atomic sulphur

from such species as sulphates and thiocyanates than were required

for iodine AAS. Fracturing of the graphite tubes was more frequently

observed and the useful lifetime of each tube was reduced consid-

erably.

2) The inefficient nature of the water-cooling arrangements in

the cell provided for a long dead-time between samples and could

not prevent the metal electrodes from overheating. This often

caused considerable damage to the atom cell.

3) The manner in which the graphite tube was held in the electrodes

by the carbon split-rings made it necessary for these end-pieces

to be machined to within fine tolerances for efficient electrical

and mechanical contact. 3rrors in the production of these rings

often produced sparking at the ends of the tube and the electrodes

would become welded to their support pillars.

To investigate further the direct determination

of sulphur by AAS and obtain more reproducible results, a more

efficient and reliable graphite tube atomiser was designed and

constructed. The details of this new furnace and the results

obtained are presented in Chapter Five.

103

FIVE

104

5.1 Introduction.

Chapters Three and Four described the direct

determination of iodine and sulphur at their vacuum u.v. resonance

lines using a small graphite tube atomiser. Despite the many

faults with this furnace the results obtained have shown that

the small electrothermal graphite atomiser was a practical and

useful system for the production of atomic vapour from the non-

metals for AAS in the far ultraviolet. To eliminate or minimise

the faults discussed in Chapter Four a completely new system was

designed and constructed. This chapter describes the development

of this atomiser and its application to the direct determination

of iodine and sulphur by AAS. A mini-Massmann rod atomiser is

described also for the direct determination of iodine; sulphur

AAS was not possible with this furnace. A comparison of the

results obtained with the two non-flame devices is made. Finally,

the use of a demountable hollow cathode lamp source as a line

emission source for iodine AAS is discussed.

5.2 The Design and Construction of a New Graphite Tube Atomiser.

The new assembly, which was to provide the basis

for further modifications, and the cell ultimately used, is shown

in Fig.5.1. In a similar manner to the original system this new

cell was housed inside a chamber on the optical bar, between

source and monochromator, and could be purged with an inert gas

to remove atmospheric oxygen. Contact between the graphite tube

and stainless-steel electrodes was again via graphite end-pieces,

machined from 12mm diameter graphite rod providing a tight fit

in the electrodes, and a 4mm diameter hole was drilled through

the piece to provide an optical path. The water-cooled stainless-

105

(a)

/ steel

electrode -l. \ 0 0"

( b)

Fig.5.2

tube

graphite contact

power -+--~::--T--t-~~ term ina 1 s

baseplate

inert gas

Fig.S.1 .

steel electrodes were mounted on a spring-loaded carriage mechanism,

allowing for any expansion by the furnace during heating and

enabling tubes of various lengths to be examined. The normal

graphite tube length was 20mm and, as before, was machined from

a rod of high purity Ringsdorff graphite to provide a tube of

6mm diameter and 1mm wall thickness. The electrical power transfer

to the tube and water cooling efficiency of the metal electrodes

was far superior to that obtained with the original system. The

power to the furnace was supplied by the Variac and transformer

arrangement previously reported.

Using the H2S-Ar EDL source the performance

of this cell was examined. A background absorption signal of

approximately 0.2 absorbance unit was always present when more

than 75% power was used to heat the furnace, although no sample

(even a blank) was present in the cell. Firing the-tube at peak

voltage more than twenty times had no effect in removing this

non-specific absorption which was present at all wavelengths

below 200nm. Various experiments were undertaken to discover

and eliminate the cause of this interference. Oxygen-free nitrogen

was the usual purging gas employed and the possibility of CN being the

absorbing species was considered. Changing to argon as the purging

gas, however, did not decrease the observed absorption. Furthermore,

no effect was observed on the signal when the purge gas flow-rate

was changed, except at very high flow-rates (in excess of 2 to

3 1.min-1 ) when cooling of the tube by the gas occurred.

L'vov (1W) has described a technique of cutting

slits into tube furnaces to provide an alternative exit from the

tube other than the ends of the tube for species capable of molecular

107

absorption, i.e. condensation or recombination reactions were

encouraged to occur out of line with the optical axis. A similar

arrangement was investigated with this furnace. A slit, 1mm wide,

was cut into the tuba to a depth of 2 to 3mm on each side of the

sample introduction port. This did succeed in reducing the non-

specific absorption to 0.10 to 0.15 absorbance unit. To reduce

this signal further, attempts were made to line the furnace with

tantalum foil, a technique used by other workers when faced with

problems such as carbide formation, (117). Using thin (ca. 0.1mm)

tantalum foil, the inside of the graphite tube was lined and the

furnace and foil positioned between the graphite end-pieces and

metal electrodes. The results obtained with this were not reproducible.

On some occasions no background absorption signal was observed

and at other times the signal was unaffected. Even when no back-

ground signal was detected neither iodine nor sulphur AAS was

possible. The liner served to cool the furnace and decreased

the heating rate of the sample, so allowing for volatilisation

and escape from the tube without decomposition to the atomic state.

The next series of experiments was directed

towards the geometry of the system and size of the graphite tube.

Several of the commercially available EGA 74 Perkin-Elmer tubes

(28mm long, 9mm o.d. and 1.5mm wall thickness) were examined but

a similar absorption signal was observed at wavelengths below

195nm. An identical wavelength dependence was observed with

these tubes as with the lab. made tubes. Varying the length of

the tubes had no effect on this signal either.

The graphite-graphite and graphite-steel

108

contacts at the ends of the tube were examined and.modifications

to the stainless-steel supports undertaken. Two new electrode

designs were examined, Fig.5.2, and the main characteristics may •

be summarised as:

a) The carbon-steel contact surface area was much increased so

reducing heat dissipation at these points.

b) The carbon end-pieces were much smaller in thickness and the

ends of the tube projected beyond the external face of the piece,

i.e. the carbon-carbon interfaces were no longer directly in the

optical path.

Studies with both systems revealed the background

absorption signal to be a function of the carbon-carbon contact

and the assembly shown in Fig.5.2 (b) eliminated this signal.

Sulphur AAS using the redesigned furnace was now

examined. 1941 aliquots of pure (distilled) water were placed

in the furnace using an Eppendorf micropipette, the tube was

heated at 1 to 27 until all signs of water vapour had disappeared

and finally the furnace was fired at 8 to 97 for 2 to 3sec. A

large and irregular absorption peak was observed, Fig.5.3; water

vapour was condensing along the optical path during the drying

stage and vaporising during atomisation. Although sulphur AAS

could be observed over this non-specific molecular absorption,

Fig.5.3 (b), without background-correction facilities no analytically

useful results could be obtained.

The only purge gas entry into the atomisation

chamber was directly below the centre of the tube, nitrogen (or

argon) flow through the furnace was only a slow diffusion process

1C9

C 5 bid

% Absorption

0-1 0

0 0

a

3 CD

(f) CD 0

OH-

• ogn,i, og.Tticr—Tro, ow, ti10.1!' pana.o:;q0 uoT. q.ciaescix pimaizyto-ca-

via the ends of the tube. There was little positive gas flow

to remove the water vapour quickly from the furnace. A method

of passing nitrogen through the tube during the drying stage was

required and the assembly shown in Fig.5.4 was examined. This

apparatus was operated outside of the purged chamber and nitrogen

flowed continuously through the furnace via silica side-arms

clipped onto the metal blocks. To protect the graphite tube from

the atmosphere and prevent its oxidation, the furnace was surrounded

by an asbestos shield. Whilst this arrangement succeeded in

removing all water vapour during the drying stage, when sulphur

or iodine samples were examined these too were swept from the

tube and no atomic absorption signals were observed. Even a gas

1 flow-rate of a few ml.min was sufficient to suppress any atomic

absorption signal. Attempts to stop the gas flow during the atomisation

stage met with no succes as air quicklyre-entered the system

causing irregular, non-specific absorption peaks. An extension

of this technique was to rehouse this flow-through assembly inside

the purged chamber. This was accomplished by fixing 10mm diameter

calcium fluoride windows onto the ends of the stainless-steel

electrodes and providing for a nitrogen flow inside the furnace,

Fig.5.5. The application of a 1911 water sample into the tube

and drying caused a loss in transmission of the emission source

radiation, even after several minutes of drying. Firing the tube

at 8 to 9V to atomise this blank produced a non-specific absorption

signal again. It was observed that water vapour was condensing

on the windows decreasing their transmission characteristics and

causing the base line from the source emission to drift.

The condensation of water vapour, with its subsequent

111

water

power connection

~steel electrode o contact

o

insu lat ion

graphite contac t

I

normal tube

slit tube

Fin9.l Graohi te Tuce-:lectrode !l.ss8:.lbl"r =::-.1ploved ?or

Sulohur and lad inc A4\S.

Fig.5.6

evaporation and absorption of radiation during the.atomisation

stage in heating the furnace, was a serious problem because the

evaporation occurred in the optical path. The relatively large

mass of cooled stainless-steel encouraged the condensation and

it was necessary to reduce this whilst maintaining efficiently

cooled electrodes. This was carried out to provide the final,

successful, electrothermal graphite cell for the atomisation of

non-metallic species, Fig.5.6. The electrodes are similar to

the original system (Fig.3.5) but are larger and individually

water-cooled.

5.3 A Mini-Massmann Rod Atomiser.

As referred to earlier,the 20mm graphite

tube furnace was not the only non-flame absorption cell examined

for vacuum ultraviolet AAS. The second graphite furnace, a mini-

Massmann rod atomiser, is shown in Fig.5.7. As can be seen, it

was much smaller than the tube furnace but fitted into the same

metal electrodes via its carbon end-pieces and was used with the

assembly turned through 900; typical of the graphite-filament

type cells. The rod was made from the same 6mm diameter, high

purity graphite rod as used for the tube furnaces. Because of

its smaller size, the rod atomiser was purged more easily of

oxygen and water vapour, and smaller samples (typically 1 to 5.p1)

could be employed. However, to prevent radiation reaching the

detector without first passing through the atom cell it was necessary

to employ limited-field viewing. This was achieved by small-bore

collimating tubes, made from brass, fixed in the optical path

so stopping-down the incident radiation from the source.

114

coll imator

~> .. ~ (2)

furnace

sample port? . c=flOll

--20mm-

optical ax is

Graphite Rod Atomiser

Fig.5.7

11 5

5.4 A Demountable Hollow Cathode Lamp.

The iodine and sulphur AAS experiments referred

to in Chapters Three and Four were undertaken using 1-11)L sources

as commercial hollow cathode lamps were not available for these

elements. A limiting factor with the use. of EDL sources, however,

is that being prepared from silica tubing, radiation of shorter

wavelengths than ca. 175nm is absorbed to a great extent by the

quartz bulb. A second possible source is the demountable hollow

cathode lamp. The lamp examined here was a commercially available

device (IMiniglowt, Spectral Products Inc, New Haven, Conn., U.S.A.)

and consisted of a replaceable hollow cathode retained within

a water-cooled brass support inside a brass body. Argon, at

pressures between 2 and 20torr, was passed continuously across

the face of the cathode, through the body of the lamp, and a

calcium fluoride window (3mm thickness, 25mm diameter) was fitted

to allow transmission of radiation of wavelengths down to ca. 125 to

130nm. The lamp is shown in Fig.5.8. The continuous flow of

inert gas through the lamp was expected to reduce the self-absorption

of the emitted resonance radiation and the water-cooling of the

cathode lowered the operating temperature of the lamp so providing

sharper line emission.

A variety of salts were examined as cathode

materials and spectra obtained from chlorides, bromides, iodides,

sulphur and phosphorus. The chlorine and bromine emissions are

discussed in more detail in Chapter sight.

The power to the lamp was supplied via

a commercial Techtron hollow cathode lamp source unit and to

compare its performance with ?DL sources for iodine AAB, the

116

water-cooled brass block

window r

gas in brass cathode

holder

- L I

\\\1\IP- water

cathode

anode vac. pump

MINIGLOW Hollow Cathode Lamp

supply to the lamp was electronically modulated. This was achieved

in a similar manner to that described for the electronic modulation

of the sulphur EDL in Chapter Four, i.e. a 1000Hz, 20V peak-

to-peak, square-wave a.c. modulating signal was led to the trigger

unit of the hollow cathode lamp supply unit and the same signal

was passed to the phase-sensitive detector as a reference signal.

The hollow cathode line emission was observed to follow well this

modulating signal.

5.5 Iodine AAS1(a) The Graphite Tube Atomiser.

Using the new electrothermal graphite tube

atomiser, the direct determination of iodine by AAS was again

examined. In Chapter Three a mechanically modulated a.c. system

was described. The iodine EDL employed for these studies did

not respond to electronic modulation. The emission intensity

vs applied microwave power curve for this r_LA, at 183.Onm is shown

in Fig.5.9, and as can be seen a variation in the applied power

about the optimum 15watts produced negligible modulated emission.

Decreasing the applied power to about ca. 10watts and modulating

to produce an on-off signal caused the lamp emission intensity

to be unstable. To achieve electronic modulation of the source

for iodine AAS, and the inherent advantages of this technique,

a new iodine containing EDL was necessary. A variety of metal

iodide lamps were examined and the best in terms of signal-to-

noise ratio and signal-to-background ratio for the iodine 183.Onm

resonance line was an EDL containing a few milligrams of mercuric

iodide and ca. 4torr of argon. The 183.Onm line emission intensity

vs applied microwave power curve for this lamp is shown in Fig.5.10,

and it responded well to the modulation signal used for the sulphur

studies, i.e. 1000Hz, square-wave, mark-to-space ratio of 1:1.

118

Performance Of Iodine LDL Sources.

HgI2-Ar 183.0nm

10—

I I I I I 0 20 40 60 80

100

Applied power (watts)

Fig. 5.10

Fig. 5.9

I2-Ar EDL 183.0nm

10—

O.C. component

— of emission

tr) C C)

a)

cc

input [modulation

0 0 10 20 30 40 50

Applied power (watts)

119

The chamber nitrogen flow-rate was optimised

at 11.min 1, the slits at 301.1m and the atomisation voltage at

8.0V across the graphite tube. With an instrumental time-constant.

Of 300m.sec iodine calibration plots were determined at 178.2nm

and 183.0nm for *1 aliquots of iodine as aqueous potassium iodide,

iodate and periodate. The absorption signals observed at the

iodine resonance lines were independent of the nature of the

iodine bearing species and furnace purging gas, (nitrogen or

argon). The analytical curves of absorbance vs concentration

were both linear upto an absorbance value of approximately 0.6,

as found previously with the simple graphite tube apparatus.

The sensitivities (for 1% absorption) were improved, however, by

ca. 50%, providing a linear working range at 178.2nm of 1.5 to

15.Ong I and at 183.0nm of 2.0 to 5Ong I. This improvement in

sensitivity was attributed to the more efficient heating rates

obtained with the new graphite furnace, and the mercuric iodide

EDL source suffering less pressure broadening than the iodine

EDL source due to the lower volatility of mercuric iodide and,

hence, its lower vapour pressure in the lamp. The improved detection

limits were due to the better signal-to-noitB ratios obtained for

the electronically modulated source as compared with the mechanical,

rotating sector technique.

The relative effects of simple foreign

ions were identical to those described in Chapter Four, with one

exception. Using this new tube atomiser, phosphate, as di-sodium

hydrogen phosphate, did not cause such a large enhancement of the

absorption signal at 183.0nm. No interference was observed from

a 10-fold excess of phosphate on a 5)11 sample aliquot containing

120

1Ong I. A 100-fold excess enhanced the 183.0nm absorption signal

by 30%, and, again, a 1000-fold weight excess caused 100% absorption

of the incident radiation. The effect of the 100-fold weight

excess of phosphate could be monitored at the 124.4nm line, as

before, i.e. the 184.4nm line could be employed to correct for

non-specific background absorption. No explanation could be found

to explain the different effects of the phosphate in the two

furnace atomisers.

The Demountable Hollow Cathode Lamp Source.

The iodine AAS calibration graphs were repeated

using the demountable hollow cathode lamp source. Before any

AAS measurements were conducted a study was made of the emission

characteristics of a variety of cathode materials; carbon, stainless-

steel, copper, brass and aluminium. The spectra, recorded between

170nm and 200nm with an argon flow pressure of 2torr and lamp

power of 15mA, are shown in Fig.5.11. For all further studies,

aluminium blank cathodes were employed because of its spectral

purity in the wavelength region of interest. The cathode for

the atomic emission of iodine was prepared as follows. A few

grams of mercuric iodide were mixed with an equal weight of powdered

tungsten, and the mixture was compressed into a blank aluminium

cathode. The mixture was then drilled out of the cathode, using

a N° 33 drill, leaving a thin layer, ca. 1mm thick, around the

inner wall of the cathode. The cathode was then pushed into its

water-cooled brass holder and the assembly operated as referred

to previously. An emission spectrum from the HgI2/W/A1 hollow

cathode source was recorded and Table 5.1 compares the results

with the emission from the ii=7I2 EDL.

121

Table 5.1 Comparison of ilia, Lamps.

Wavelength (nm) Relative Intensity

2DL DHCL

178.3 0.3 0.8

179.9 0.2 • 0.6

183.0 1.0 1.0

184.4 1.0 0.3

DHCL Demountable Hollow Cathode Lamp

Table 5.2 Ron-Flame Iodine AAS Results.

(a) 20mm Graphite Tube Atomiser

Wavelength (nm) EDL DHCL

S D.L L.R S D.L L.R

178.2 0.1.0 1.5 to 15 0.3 4.0 to 40

183.0 0.3 . 2.0 to 50 0.4 2.5 to 55

(b) Graphite Rod Atomiser.

Wavelength (nm) EDL DHCL

S D.L L.R S D.L L.R

178.2 0.06 4.0 to 20 0.1 4.0 to 15

183.0 0.10 1.0 to 40 0.26 1.5 to 40

DHCL Demountable Hollow Cathode Lamp

S = Sensitivity (ng) for 1% absorption.

D.L. = Detection Limit (ng).

L.R. = Upper Limit (

of Linear Range.

122

0.2—

U C

.n0.1-

stainless-steel

1 70

8.0 9.0

copper Voltage across rod

jk.■■••■•■■■■■

w0.2— C L.)

0" .2 0.1—

brass

aluminium "A-

200 190 180 170 A inm

Fig. 5.11

2.0 1.0

N2 flow through. chamber

Fig. 5.12

Calibration plots of iodine concentration vs

absorbance were determined using the 20mm graphite tube atomiser

and the demountable hollow cathode lamp source at 183.0nm and

178.2nm. The purge flow-rates, monochromator spectral bandpass

and furnace atomising voltage were as for. the EDL source studies.

The sensitivity (for 1% absorption) at 183.0nm was 0.4ng I and

at 178.2nm 0.3ng I; the linear working ranges were 2.5 to 55ng I

and 4 to Ong I at the 183.0nm and 178.2nm resonance lines

respectively.

5.5 Iodine AAS, (b) The Mini-Massmann Rod Atomiser.

The mini-Massmann rod atomiser was arranged

as described in Section 5.3. The source emission, EDL or hollow

cathode lamp, was focussed by means of the calcium fluoride lens

into the furnace through the collimator to reduce stray radiation

reaching the detector. As a further precaution, a second collimator

was employed between the furnace and monochromator. With standard

solutions of iodine as potassium iodide, iodate and periodate

the usual optimisation of purge flow-rate, spectral bandpass and

atomising voltage were undertaken. Fig.5.12 shows the effect

of varying the nitrogen flow-rate and applied voltage on the

absorbance by iodine at 183.0nm using the rod atomiser. For

further studies the nitrogen flow-rate through the chamber was

maintained at 0.51.min 1 and an atomisation voltage of 9.07 employed.

Because of the much smaller volume of the rod furnace compared

with the tube, it was necessary to use an instrumental time-constant

of 100m.sec to prevent damping of the signals.

Fig.5.13 shows the analytical curves obtained

at 183.0nm using the mini-Massmann rod with the mercuric iodide

124

0.8 EDL

(a)

DHCL

(b)

(a)rr •=100m.sec

( b) T =300 m.sec

20 ng I

125

10 30 40 50

Iodine Calibration Curves With The Rod Atomiser.

178.2nm Minimassmann rod

0.8— DHCL

cD

CTI

EDL

0.4- L O tr)

0 0

10 15 20

25

n g I

Fig. 5.14

Fig. 5.13

183.Onm Minimassmann rod

demountable hollow cathode lamp source. Fig.5.13 also shows

the effect of the instrumental time-constant (') on the calibration

curves. The curves were independent of the nature of the iodine

sample and independent of the nature of the purge gas employed;

nitrogen and argon.

Fig.5.14 shows the calibration curves at the

178.2nm iodine resonance line. The sharp curvature of the analytical

curve using the EDL source was due to the spectral bandpass of

the monochromator being at 0.16nm compared with the 0.05 nm band-

pass employed with the hollow cathode lamp at 178.2nm and with

both sources at 183.0nm. The increased slit width for the EDL

source studies at 178.2nm was necessary because of the poor signal-

to-noise ratio at this line from the EDL with this small atomiser.

The results obtained for the iodine AAS studies

at 178.2nm and 183.0nm, using the 20mm graphite tube atomiser

and the mini-Massmann rod furnace are summarised in Table 5.2.

The Effect Of Foreign Ions With The Rod Atomiser.

As in Chapter Three the effects of 10:1, 100:1,

and 1000:1 weight excesses of sodium chloride, potassium chloride,

sodium sulphate, sodium nitrate and di-sodium hydrogen phosphate

on 1Ong iodine (5p1 aliquot of 2pg.m1-1 iodide) at 183.Onm were

investigated. The results are summarised in Table 5.3. As before,

corrections could be made for the non-specific molecular absorption

by employing the 179.9nm and 184.4nm iodine non-resonance lines.

5.6 Sulphur AAS, (a) The New Graphite Tube Atomiser.

Using the new electrothermal graphite tube

atomiser, the direct determination of sulphur by AAS using its

126

Table 5.3 effect Of Common Salts On Iodine Absorption Usine.

The Graphite oc Atomiser.

1S3.0nm Salt Concentration Chan

At ge Increase Absorption

Studied ratio in in at

MX MX; iodide absorbance,% absorbance 184.4nm

NaC1 10:1 0 0 0

100:1 +100% 0.1 0.1

1000:1 100% absorption of radiation

KC1

Na2HP04

NaNO3

Na2SO4

10:1 0 0

100:1 0 0

0

1000:1. +1705 0.17

0.17

10:1 0 0

0

100:1 +601, 0.06

0.06

1000:1 +100% 0.1

0.1

10:1 0 0 0

100:1 +100% 0.1 0.1

1000:1 +120% 0.12 0.12

10:1 0 0 0

100:1 0 0 0

1000:1 +2005 0.2 0.2

127

0

0 U) 0 — Z 0 c,-

0

(r)

o 01

Absorbance

0

0

01 CD

Absorbance

U, 0

CY)

51.'5'2

91-'5 • 2T3

CO -

t.0 —

cl)

3 •

S 6u 09 07

0Z

0 I

U 0 Z 1.

w u L 0 9 I.

• STV anticiln5

far ultraviolet resonance and near-resonance lines was undertaken.

The remainder of the apparatus was as described in Chapter Four,

i.e. an iI2S/Ar EDL, electronically modulated; was used. The

optimum slit width was found to be 30pm (0.05nm bandpass) and

the sulphur atomic absorption results were optimised for nitrogen

flow-rate through the chamber and applied voltage across the

furnace, Fig.5.15 (a) and (b) respectively. Employing a nitrogen

1 flow-rate of 0.751.min and an applied voltage of 97, calibration

curves for sulphur as potassium and sodium sulphate, potassium

thiocyanate and thiourea were constructed at 180.7nm, 182.Onm

and 182.6nm, Fig.5.16. The results are summarised in Table 5.4.

The sensitivities obtained with the new graphite

tube atomiser were a factor of 2 to 3 times greater than with

the older system.

As can be seen in Fig.5.16„ the calibration

plots of absorbance vs concentration of sulphur all deviate from

linearity towards the concentration axis at low absobance values

(ca. 0.3). Identical curvature of calibration curves was obtained

whether the sulphur EDL was made from H2S/Ar, S02/Ar or elemental

sulphur/Ar. This may have resulted from a variety of causes,

as mentioned in Chapter Three; these will be discussed in more

detail here.

5.6 (b) The Bending Of Analytical Curves In AAS.

(i) Apparatus Faults.

In their discussion concerning the bending

of analytical curves in AAS, Rubeska et al. (11g) classify the

causes of bending into two groups; apparatus faults and properties

of the spectral line itself. The first includes such factors

129

as a fraction of the source radiation falling on the detector

without passing through the absorbing medium. Some lamps emit

a continuum with the line spectra or, perhaps, a non-absorbed

line that is not resolved from the resonance line. In quantitative

AAS analysis, a direct proportionality is assumed to exist between

concentration and absorbance, the absorbance (A) is given by,

A = log Io/I ... 5.1

as discussed in Chapter One.

With the above apparatus and optical

faults the effect is to add an intensity term to both the numerator

and denominator, so that the measured absorbance may be represented

by,

' " A = log I + I ... 5.2

Where Io and I" are the signals due to the resonance line and

the non-resonance line radiation, respectively, when blank solutions

are atomised, and It and I" are the readings due to the resonance

and non-absorbed radiation, respectively, when a sample is atomised.

The effect is to introduce a non-linearity into the absorbance vs

concentration curve.

The optical arrangement employed for the

sulphur AAS studies was identical to that used for iodine AAS.

Deviations from linearity at low absorbance values were not observed

with iodine, therefore, the possibility of radiation reaching

the detector in sulphur AAS without passing through the furnace

was small. As a check, however, the sulphur calibrations were

130

re-examined using collimators, of the type shown in Fig.5.7, to

limit the field of observation. Apart from worsening the signal-

to-noise ratio of the source emission, identical results were

observed with or without the collimators present.

Decreasing the width of the entrance and exit

slits to less than 3C-nn did not increase the linear range of the

analytical curve. The presence of any continuum emission passing

through the atom cell could be disregarded therefore as the sensitivity

of AAS using a continuum source is dependent on the slit width

of the monochromator. The possibility of an impurity having an

emission line so close to the sulphur lines and being present

in the EDL could be ignored also.

Emission from the heated atomiser served to

dampen absorption signals, as discussed in Chapter Three. Employing

modulation techniques, however, eliminated the effect of this

interference and no emission signal was observed using the a.c.

detection apparatus.

(ii) Bending Due To The Nature Of The Spectral Line.

The second group of causes of bending of analytical

curves, those due to the spectral line, are more difficult to

deal with. Included are hyperfine structure of the line, the

ratio of the absorption and emission line widths and the resonance

line broadening and line shift in the atom cell.

Hyperfine Splitting.

The hyperfine line structure is determined

by the interaction between the spin of the nucleus and the resultant

spin of the electron shell. According to these spin moments a

magnetic interaction is set up between the nucleus and electron

131

splitting the electron energy levels and consequently the spectral

line. A nucleus with an even number of protons and an even number

of neutrons has a moment of zero and there is no hyperfine splitting

of the line. Elemental sulphur is 95.1% 32S, comprising 16 neutrons

and 16 protons, i.e. there is no resultant spin moment and negligible

hyperfine splitting from the other isotopes; 33S (0.700 and

"S (4.2%).

Ratio Of Emission and Absorption Line Widths.

In many cases, the width of the lamp emission

line cannot be neglected in comparison with the width of the

absorption line as this often causes bending of analytical curves,

i.e. the profile of the emission line as well as that of the

absorption line must be considered. These line profiles may be

a Gaussian function if the Doppler effect determines the line

shape or a Lorentzian function if collisional broadening is the

major component. (A combination of the two effects results in

a Voigt function).

The extreme case of Doppler broadened lines

has. been dealt with by Mitchell and Zemansky (IA). They have

tabulated values of the total absorption, defined as,

1 - T = 1 - I/I0 2

for different optical depths for line profiles with different

line-width ratios.

Defining,

Emission line width , Doppler broadened ... 5.3 Absorption line width

and converting absorption to absorbance, curves of absorbance

132

therefore,

and as

A))14 = 0.653 x 108 sec-1

>■ A X =

against optical depth for various values of XI; are'shown in Fig.5.17(a).

Rubeska et al. have treated the case of collisional

broadened profiles. Their results, expressed as absorbance vs

optical depth for various values of XL, are presented in Fig.5.17(b).

Where,

Emission line-width XL = , Lorentz broadening ... 5.4

Absorption line-width

To achieve the bending of the analytical

curve observed with the sulphur calibration, it may be seen that

the emission line-width to absorption line-width ratio needs to

be greater than unity. The relative effects of the Doppler and

Lorentz broadening processes in the absorption cell and lamp

source may be estimated for the 180.7nm sulphur resonance line.

Natural Broadening.

Natural broadening of a spectral line is

related to the lifetime (t) of the excited electronic state and

the natural width of a resonance line may be defined as,

Al)" - zrrt A

2.1T

-1 sec ... 5.5

Where A2,1

is the transition proability for the line, and for

the 180.7nm S line,

A2,1 = 4.1 x 108 sec

AXINI = 7.1 x 10-6 nm

133

Absorbance

O

I I I Absor bance

0■1% WW1.

ti

X X r

x 11 r r I CA)II O

O (A)

O

01—

1

O

........

'6!.d

0

ooTweli L19,1P-M—JTEll ouTrl 014I DUF saAJno ruoTI-:[ruK JO zuTPuaa

T

In comparison with the Doppler and Lorentz widths

the natural width may be ignored; it is identical for the absorption

and emission lines.

Doppler Broadening.

The Doppler half-width .of the spectral line

(Ali) is given by,

Avo= zyo 2 In 2 RT C A ... 5.7

Where A.is the atomic weight, R the universal gas constant, T

the temperature and C the velocity of light.

Substituting the values for R and C, andVo = 1.67 x 1015 sec-11

= i-2. 10

and, 60 D 31 • IC; *FA? "" ... 5.8

Table 5.5 presents values of LX at various

temperatures, for the 180.7nm sulphur resonance line.

Lorentz Broadening.

Lorentz broadening of spectral lines is due

to collisions by the analyte atoms with foreign atoms or molecules.

It is dependant on the experimental conditions employed and increases

with increasing pressure in the atom cell. The Lorentz half-

width (L))L) of an atomic transition is given by,

13 2 • (:).02.• 10 . P "2 -1 3.zec

1r miti

62 is the effective collisional cross-section, and M1 and M

2 the

atomic and molecular weights of the analyte atom and foreign

sec

135

gas species respectively. If the pressure (P) is measured in

torr, then substituting for the constants,

AA- = 1022. P. 61 N1. 4_

c-1 M% / se

Winefordner (12 62 1) has discussed the choice of value for for

flames and has suggested that it lies within the limits of 20

and 100 x 1516cm2. The maximum value of 10714cm2 is assumed here.

Thus,

= 4- sec-1

The pressure in the graphite furnace was atmospheric (ca. 760torr)

and in the EDL source 3torr. Assuming a furnace temperature of

2000K and EDL operating temperature of 500K the relative Lorentz

half-widths may be determined.

Graphite Tube Furnace.

For sulphur AAS, using nitrogen as the cell

purge gas, M1 is 32 and 142 is 28. Therefore,

1.9. log (2 sec-1

and,

Z.13 x 10 4 nri

EDL Source

Again, M1 is 32 and M2, using argon as the lamp filling

gas, is 40.

Therefore, &VL =14-•2106 2( sec-1

and converting to wavelength units,

6XL. = 1.55 x 1076 nm

136

Collisions between the gaseous atoms and molecules

not only cause broadening of the spectral line but also displaces

the line from its initial position. This effect may be ignored

here as the shift is usually much smaller than the broadening

effect. Both nitrogen and argon, being heavy gases, cause a shift

towards longer wavelengths. The displacement effect, therefore,

in the furnace and lamp source is similar and has little effect

on the ratio of the emission and absorption line widths.

Resonance Broadening.

This is often referred to as Holtzmark broadening

and is a particular case of collisional broadening. It is caused

by collisions between atoms of the same kind. The half-width

of the spectral line, therefore, increases with increasing concentration

of the analyte atoms; causing bending of the analytical curve.

Rubeska et al. (Ile) used the following

relationship to express the resonance broadened half-widthl Wik.

= X . sec-1

2m ).7c,

e and m are the charge and mass of the electron, respectively,

and N is the number of analyte atoms per unit volume. f is the

oscillator strength of the transition and for the 180.7nm S line

is 0.12. Several values have been proposed for the constant X 1

although they differ only insignificantly. A value of 0.32 is

assumed here.

Substituting for the constants and expressing the

half-width in terms of wavelength units,

.6.Xs = 4.5 x 10-14. X3. f ri cm

137

and for the 120.7nm S line,

ISXR = 3.2 x 10 22 x N nm

Graphite Tube Furnace

At 2000K the effective volume of the graphite

tube is 1.68cm3. For a 50ng sample of S and assuming 10C% atomisation

of the sample, the concentration of sulphur atoms in the furnace

is 5.63 x 1014 atoms S.cm-3.

Therefore, = 1.8 x 1 0 7nra

EDL Source.

Assuming 100% dissociation of 3torr H2S contained within

a 2cm3 EDL bulb, the number of sulphur atoms in the lamp is

1.1 x 1017 atoms S.cm-3.

Therefore) /SXR = 3.52 x 105nm

For an assumed EDL gas temperature of 500X

and a graphite furnace atom temperature of 2000X, the various

half-widths are summarised in Table 5.6.

As the pressure broadening and the resonance

broadening both produce Lorentzian profiles for the spectral line

the values may be summed and XD and XL determined.

= 0.51

XL = 0.17

Three observations may be made concerning

these results:

1) Resonance broadening in the atom cell (flame or furnace) is

138

Table 5.4 Non-Flame Sulphur AAS Results.

Wavelength (nm) Sensitivity (ng) Det. Limit (ng) Lin. 1:ange (ng)

180.7 0.42 2.0 to 33

182.0 0.68 2.0 to 45

182.6 1.50 2.0 to ca. 100

* for 15 absorption.

Table 5.5 Doppler Half-Widths For 180.7nm S Line.

Temperature (K)

500 1000 2000 3000

Doppler half-width (104nm) 5.18 7.34 10.2 12.7

Table 5.6 Half-Widths For 180.7nm S Line.

Natural Doppler Lorentz Resonance

Broadening Broadening Broadening Broadening (nm) (nm) (nm) (nm)

EDL 7.1 x10 ( 5.18 x 10 4 1.55 x 10 ° 3.5 x 10 5

Furnace 7.1 x106 1.02 x 10 3 2.13 x 10 4 1.8 x 10 7

EDL temperature of 500K, argon pressure of 3torr.

Furnace temperature of 2000K, 5Ong S sample.

139

referred to often as a cause of bending of calibration curves

in AAS, as the broadening process is proportional to the analyte

concentration. However, the estimation of the resonance broadening

here indicates that this process is the least serious of the

broadening processes. Where actual experimental measurements

have been possible on resonance broadening the results obtained

show the theoretical values to be an order of magnitude too low (118),

but this still has little effect here. Furthermore, when AAS

studies are undertaken using a source whose line emission is not

truly monochromatic,as here, resonance broadening in the atom

cell would assist in straightening the analytical curves as XL

would tend to decrease with increasing analyte concentration.

Livov (illo) has demonstrated the effect on

the AXIS calibration curve of broadening the absorption line profile

with respect to the emission line profile. By increasing the

pressure in the graphite furnace chamber to 9atm. the linearity

of a cadmium calibration plot was increased significantly, although,

of course, the sensitivity at the higher pressure was worse.

Unfortunately, high pressure studies were not possible with our

graphite tube furnace arrangement.

2) It appears from Table 5.6 that the major cause of broadening

of spectral lines in both the graphite furnace and the DL is

the Doppler effect. Although little information is available

concerning atom temperatures in non—flame cell and plasma gas

temperatures (as opposed to electronic excitation temperatures

in plasmas) itisda±tfel that the relative temperatures are such

to cause the Doppler broadening ratio (X,) to be greater than

unity. In Chapter Seven a two—line AAS technique is employed

140

to determine the temperatures attained by an atomic vapour in

a commercial graphite atomiser; the results obtained indicate

the atomic vapour temperature follows the graphite wall temperature

quite closely. However, it is douletfla whether these results

may be applied here as non-metal dissociation and atomisation

for AAS cannot follow the same mechanisms and kinetics as are

assumed to exist for normal metal analyses with graphite furnaces.

3) The preceeding estimates of the relative broadening effects

occurring in the furnace and EDL source ignored the effects of

self-absorption in the source. As the mechanism of the plasma

discharge is not fully understood and no measurements of self-

absorption broadening in EDL sources have been undertaken, it

is not possible to assess the magnitude of this effect with the

sulphur EDL sources. If self-absorption in the sulphur containing

lamps is the limiting case here, i.e. self-absorption broadening

is greater than the effect of pressure broadening, then this must

be a special case as such bending of analyical curves was not

observed for iodine calibrations. Two other points also may be

made. Marshall (122) employed a similar sulphur EDL source for

the direct determination of this element at 180.7nm using a nitrogen-

separated nitrous oxide-acetylene flame and reported calibration

curves linear to absorbance values of 0.7 to 0.8. L'vov (V LB)

investigated the non-flame determination of iodine, sulphur and

phosphorus using a graphite cuvette assembly and reported linear

calibration graphs upto absorbance values of 0.3 to 0.5. Unfortunately,

he provides no details as to which element provided the least

linear analytical curve. However, it would appear that the non-

metals, and sulphur in particular, do not provide for long linear

141

working ranges when determined by graphite tube AAS. Before

turning from this problem of non-linear calibration curves, the

results obtained from a study of the molecular species formed

on the volatilisation of sulphur samples will be discussed.

In Chapter Six results are presented from

an examination of the non-specific molecular absorption spectra

of some simple inorganic salts volatilised in the small, 20mm

graphite tube atomiser. The apparatus described for these experiments

was employed here to study the decomposition products of the

sulphur containing species, i.e. a deuterium arc continuum lamp

emission source, monochromator bandpass of 0.25nm and instrumental

time-constant of 50m.sec. Using 53t1 aliquots of sample solution,

calibration curves were determined for potassium sulphate, potassium

thiocyanate and thiourea, at 180.0nm using the continuum lamp

source. These curves are shown in Fig.5.18 (a), (b) and (c).

As may be seen, these curves deviate from linearity but in a

direction opposite to that observed for the atomic sulphur absorbance

curves.

Spectrometric absorption studies rely for their analytical

usefulness on a direct relationship existing between absorbance

and concentration, i.e.

A o< CX

Where x = 1 for a typical Beer's Law curve, and if a plot of

log A vs log C is computed the slope of this line is unity. These

log-log curves were determined for the atomic sulphur results

and for the molecular absorption curves in Fig.5.18 and are shown

in Fig.5.19 (a) and (b). The gradient (Gx) of each line was

142

a) sulphate b)thiocyanate c) thiourea

1.0—

Abs

orba

nce

0.5—

0 I I I I I

500 1000 0 200 400 0 100 pg.ml-1 S

Fig.5.18

logC 0.5 1.0

< -0.5- 0) 0

0

(i)= thiourea

(ii)=thiocyanate

<-0.5- rn 0

(iii) = sulphate

-1.0—

144

Fig.5.19

determined and the results were,

Gs (atomic sulphur calibration) = 0.81

Gso4 (sulphate calibration)

= 1.25

G (thiocyanate calibration) = 1.22

Gcs (thiourea calibration) = 1.22

It was of interest that the reciprocal of the sulphur AAS line

gradient was similar to the results for the molecular species,

1/Gs = 1.22

This served to indicate that the species responsible for the

molecular absorption at 180.0nm was responsible for the deviation

in linearity of the sulphur AAS calibration curves determined

at 180.7nm.

With the deuterium arc continuum source the non-specific

molecular absorption spectra were determined from 5,1tl aliquots

of aqueous solutions of potassium sulphate, potassium thiocyanate

and thiourea at 5 and 10nm intervals between 180nm and 290nm

using the small graphite tube. These spectra are shown in Fig.5.20

(a), (b) and (c). An atomising voltage of 8.5V was used throughout

and a rnonochromator spectral bandpass of 0.25nm. Fig.5.20 (a)

also shows the spectrum observed by Fuwa (V2*) on spraying sulphur

containing solutions (as sulphuric acid or cysteine) into a long-

path absorption tube containing the exhaust gases from an air-

acetylene flame. Fuwa attributed this spectrum to SO2 absorption

and proposed the use of the absorption maximum at 207nm for the

direct determination of sulphur. Thompson et al. (MS) have

145

0.6—

0.4— U C 0 .0 0 U)

0.2—

o Graphite. tube

o Flame

--- SO2

220 N,nm

180 200 240 260

146

Fig. 5.20 (b) and (c)

• 250pg.mr1 S (KSCN)

o 100pg.ml 1 S (NH2)2S

a)

C 0

0 tn 03—

Molecular Absorption From Sulphur Containinc species.

180 200 220

240 A s nth

Fig. 5.20 (a)

E —50

L46 0

0 eL

—25 0

0 U)

determined the SO2 absorption spectrum between 180nm and 240nm

using 0.3torr of the gas in a 100mm cell, their results are also

shown in Fig.5.20 (a). Sponer and Teller (4.3) have reported the

SO2

band spectrum between 175nm and 244nm. The spectrum is degraded

to the red and the transition is thought to be due to a relatively

non—bonding electron localised on the oxygen atom.

The thermal decomposition of sulphates may

be schematically represented as, (M29),

MSO4

MO + S03

and

SO3

SO2 + 2 - 2

At the high temperatures experienced with the graphite furnace

atomiser, the dissociation of the trioxide to sulphur dioxide

would be complete.

AAS analysis is concerned normally with the

determination of metal species, and the metal oxide would be

reduced by the hot graphite according to the mechanism,

MO + C M + CO

as discussed by Ottoway et al. (127). Once formed the metal

would rapidly volatilise and collisions with the hot graphite

walls and purge gas molecules and the low partial pressure of

oxygen in the furnace would prevent further metal oxide formation

and favour atomisation. The sulphur dioxide, however, being

gaseous, would rely almost exclusively on collisions with the

graphite furnace walls and purge gas molecules for reduction and

decomposition, It may be expected that this gaseous reaction

147

would be concentration dependant and may cause the deviation

from linearity of the sulphur AAS calibration curves and sulphate

etc. molecular absorption curves.

5.6 (c) The Ilini-Massmann Rod Atomiser.

The apparatus was assembled as described

in Section 5.5 (b) and 5111 aliquots of sulphur containing samples

applied to the graphite rod atomiser. No sulphur atomic absorption

signals were observed from 5Ong of sulphur as potassium or sodium

sulphate, potassium thiocyanate or thiourea, at 180.7nm, employing

a H2S/Ar EDL line emission source.

5.7 Conclusion.

The design and construction of a small graphite

tube electrothermal atomiser has been described; it has been

employed for the direct determination of iodine and sulphur by

AAS at their vacuum ultraviolet resonance lines. It has been

shown that the non-specific background absorption encountered

with graphite tube atomisers at wavelengths shorter than ca. 200nm

is a function of the geometry of the graphite-graphite contacts

employed with the furnace. The problem was surmounted here by

designing a new furnace with the tube projecting beyond the graphite

end-pieces.

A mini-Massmann graphite rod atomiser has been examined

also as an atom cell for non-metal AAS in the far u.v. Atomic

absorption studies with this small furnace have enabled the following

comparison between this small furnace and the graphite tube to

be made.

1) The smaller dimensions of the rod atomiser allowed faster

148

heating rates to be achieved; cooling to room temperature was

faster also, so reducing the necessary dead-time between samples.

2) The smaller furnace was more easily purged of atmospheric

oxygen and water vapour during the sample drying stage. As no

part of the electrode assembly lay along the optical path the

apparatus was free from non-specific absorption due to condensed

water vapour,etc.

3) The size of the sample aliquot used with the rod atomiser

was in the range 1 to 5pl compared with 5 to 15)11 used with the

small tube furnace.

4) The sensitivity (expressed as 1% absorption), for iodine,

was superior with the graphite rod atomiser by a factor of about

two.

5) Sulphur AAS was not possible with the rod atomiser; this was

attributed to the much smaller volume of this furnace and the

reduced residence time of the analyte vapour in the furnace.

6) Because of the smaller aperture of the rod furnace compared

with the tube the emission from the source reaching the detector

was reduced so decreasing the signal-to-noise ratio.

7) The faster heating rate and smaller dimensions of the graphite

rod atomiser decreased the atomic vapour residence time in the

optical path necessitating a smaller instrumental time-constant,

(1C0m.sec compared with typically 300m.sec with the small tube)..

A reduction in the signal-to-noise ratio again resulting.

Although non-specific absorption was encountered

from moderate amounts of simple inorganic salts, nearby non-

resonance lines could be employed for its correction.

. Apart from the usual )L sources employed

149

for non-metal AAS a demountable hollow cathode lamp has been

examined as a line emission source for iodine AS. The results

obtained with this source were similar to the results obtained

using the LL emission source. Unfortunately, time did not permit

an investigation of sulphur AAS with the demountable hollow cathode

lamp.

Finally, a discussion has been presented concerning the

nature of the causes of bending of analytical curves and the

deviation from linearity and Beer's Law in non-flame AAS. The

results obtained for sulphur AAS indicate that the production

of sulphur atomic vapour in the graphite tube furnace was a complex

function of SC2 reduction in the tube, when sulphate samples

were examined. Although a similar atomic vapour production

mechanism appears to occur when the sulphur is originally present

as thiourea or thiocyanate, the molecular spectra observed with

these salts could not be identified to indicate the nature of the

absorbing species.

150

SIX

151

6.1 Introduction

In previous chapters the direct determination

of iodine and sulphur has been discussed, employing their low

wavelength resonance lines. Where interferences due to inorganic

salts were studied, the effect of the foreign ions on the absorption

signal was shown to be an enhancement of the absorption signal,

which could be corrected for by repeating the measurement at a

nearby non-resonance line. The validity of this technique, however,

relies on a foreknowledge of the nature of the interference so

that a suitable non-resonance line may be chosen. This chapter

is concerned with the study of the nature of these interference

absorption signals to assist in selecting the non-resonance line,

and possibly gain an insight into the mechanism of sample dissociation

in graphite furnace. atomisers.

The non-specific absorption spectra

produced by some simple inorganic salts using a commercial HGA 2000

graphite furnace have been investigated in the wavelength range

190nm to 360nm. Using the smaller graphite tube furnace described

in Chapter Five, timabsorption spectra have been determined in the

wavelength range 180nm to 250nm.

The absorption signals obtained

at the iodine 206.1nm non-resonance line observed with potassium

iodide, iodate and periodate in the HGA 2000 furnace are shown

to be due to the molecular iodide.

6.2 Apparatus.

To examine the non-specific absorption spectra observed

in graphite furnace atomisers, two non-flame AAS instruments

were employed.

152

(a) In the wavelength range 190nm to 360nm a Perkin-:lmer Model

305B Atomic Absorption Spectrophotometer fitted witha deuterium

background corrector and HG A 2000 graphite furnace atomiser was

used for the absorption measurements. The furnace was purged,

using automatic stop-flow conditions, with argon (meter reading

4) and the absorption signals obtained, using slit setting 3,

recorded on a conventional potentiometric.chart-recorder (type

112 511, Servoscribe Ltd.,). Sample solutions were transferred

to the furnace using a icpi Eppendorf pipette, and dried at 100°C

for 30see prior to atomisation for 15sec at 2200°C; unless

otherwise stated.

Molecular absorption spectra were obtained using

the deuterium background corrector as the continuum source. With

the deuterium lamp on, the spectrometer operates in the single-

beam mode and, if no line emission source is present in the apparatus,

the absorption signal recorded is the molecular absorption produced

by the sample in the graphite furnace. The absorption signals

were measured, against distilled water blanks, at 5nm or 10nm

intervals between 190nm and 360nm. each absorption measurement

was repeated several times. As the absorption zero was set at

each wavelength the spectra obtained were independent of any

variation in photomultiplier response in the wavelength region

examined.

(b) The absorption spectra in the 180nm to 250nm range were

obtained with the small graphite tube system used for the direct

determination of sulphur and iodine as described in Chapter Five.

The continuum source employed was a deuterium

153

arc lamp (type C70-6V-H, Cathodcon Ltd., U.K.) fitted with a

vitreosil window. Power to the lamp was supplied via a Model

504 APT Regulated Power Supply (lectronic industries Ltd., U.K.)

at 250mA and 807. To reduce the length of the optical path the

window of the lamp was fitted up against the chamber of the furnace

housing. The apparatus was used in the d.c. mode and to enhance

the signal-to-noise ratio, a 50,000pF capacitor was connected

across the input terminals of the oscilloscope used to record

the absorption signals; this provided for a time-constant of

50m.sec. Both entrance and exit slit widths were maintained at

150Rm ( spectral bandpass of 0.25nm) during the absorption meas-

urements.

Sample solutions, containing 10,000,11g.m11, of each

anion were prepared from the chloride, bromide, iodide, nitrate,

sulphate and phosphate of potassium and sodium using analytical

reagent grade salts; these were diluted as required using distilled

water. Stock solutions of iodine as potassium iodate and potassium

periodate were also prepared.

6.3 The Nature Of The Iodine Absorption Signals Observed In The

IDA 2000 At The 206.1nm Iodine Line.

The non-resonance 206.1nm iodine line has been

reported by several workers to be suitable for the direct determination

of iodine by AAS, (iX8,17.ci ). This line corresponds to the trans-

ition 5P5.2Pi - 6S

2 P5/2 and has a transition probability of

3 x 106sec

-1. As its lower energy level is only 0.94eV above

the ground state for iodine it may be expected to be populated

to an appreciable extent at the temperatures attained in typical

flames and electrothermal atomisers commonly employed for AAS.

154

At 2700:: the population of this level would be about of the

total atomic iodine population in the atom cell. The use of this

line for the direct determination of iodine is attractive as a

purged optical path and vacuum monochrOmator are not necessary.

Lvov et al. (1n) used their conventional furnace

system to study the iodine absorption at 206.1nm; with an argon

pressure in the chamber of 3atm. A detection limit of 2 x 10-9g I

was claimed. Thompson (I2A) reported the determination of iodine

at this wavelength using an air-acetylene flame, but quoted a

poor sensitivity (1300ug.m171 for 1% absorption). Kirkbright

and Wilson WIC have more recently examined this absorption at

206.1nm using an HGA 2000 graphite furnace and iodine IDL source

and obtained a sensitivity, for 1% absorption, of 35ng iodine.

The analytical curve was linear to 7.5,pg I, and was independant

of whether the iodine was present as the iodide, iodate or per-

iodate of potassium. However, when the deuterium arc background

corrector was employed during the examination of iodine absorption

at 206.1nm no signal was observed. As no absorption signal for

.iodine has been reported at the 184.4nm and 187.6nm non-resonance

lines, whose lower energy levels are similar to that of the 206.1nm

line and which have greater_ oscillator strengths; the absorption

signal observed at 206.1nm would appear to be anomolous.

The nature of this absorption signal was

examined using the Perkin-Elmer AAS spectrometer with the HGA

2000 furnace and the results are presented here.

Results.

The stock solutions of potassium iodide, iodate and periodate

were diluted with distilled water to provide working solutions

155

each containing 200Fg.m1 1 iodine. The molecular absorption spectra

obtained on the vaporisation of 14t1 aliqots of these solutions

are shown in Fig.6.1. It is evident that the same species was

responsible for the molecular absorption in each case and accounts

for the appearance of similar signals for these salts at 206.1nm

using an iodine atomic line emission source. Furthermore as

206.1nm is close to an absorption maximum at 200 in the molecular

absorption spectra the relatively high sensitivity previously

reported for iodine absorption at this wavelength is explained.

In an attempt to assign the species responsible

for this absorption the spectra were compared with those obtained

by nebulising an aqueous solution of potassium iodide into a

cool H2 - N2

diffusion flame burning at a long slot burner, (131).

Comparison was also made with the spectra obtained by Koirtyohann

and Pickett (132.) using an 02 - H2 flame and long path, Fuwa-type

absorption tube. In all cases although differences in relative

peak heights were evident, the wavelengths of the absorption

peak maxima were very similar. The spectra obtained were similar

to those reported by Muller (t33) who used a closed furnace device,

and assigned the absorption to the diatomic molecular alkali metal

iodide. Koirtyohann et al. (134) considered the possibility of

the radiation loss,on aspirating inorganic salt solutions into

the oxy-hydrogen flame, to be due to scattering by particulate

matter in the flame. A theoretical and practical study of the

possible scattering effect, however, indicated a loss of radiation

by scattering of less than 0.25 and little wavelength dependance.

Iodate and periodate appear to undergo thermal

decomposition to iodide in the furnace to produce the molecular

156

KI Absorption Spectra In The EGA 2000

0.2 KI

0.1—

0.1—

200 250 300

Wavelength, nm

Fig.6.1 157

absorption spectrum of the alkali metal iodide. Alkali metal

iodates are known to decompose on heating to give the periodate

and iodide, and as the periodates are powerful oxidising agents,

.like their parent acids, it is probable that in the presence of

the hot graphite at the furnace wall reduction of the periodate

to the iodide takes place.

6.4 Absorption By Other Common Salts In The HGA 2000.

The corresponding non-specific, molecular

absorption effects observed with other alkali metal salts in

the same wavelength range(190nm to 360nm) were investigated also.

To enable comparison with the results obtained for potassium iodide)

stock solutions of sodium and potassium sulphate, nitrate, chloride,

bromide and iodide were diluted to provide solutions containing

200pg.m1-1 of the anion. For the sulphates and nitrates of sodium

and potassium no absorption was detected at this concentration in

the wavelength range 190nm to 360nm using the deuterium arc lamp

source. It is probable that these salts decompose on heating

by a mechanism such as, (Section 5.5):

14504

MO + SO

and

MNO3 YO + NO

and subsequent reduction of the metal oxide by the hot graphite,

MO + C _______4.M + CO

as discussed by Ottoway et al. (1.1).

For the halide salts, however; no such mechanism

is available and they vaporise without decomposition. It is for

158

0.4

Fig. 6.4

Fig. 6.3

0.2.— o Na Br

• K Br

200 2510

1 300 350nm

Inorganic Salts In The EGA 2000

0

1 200

250 300 350

Wavelength, nm.

160

I 1 I 7 8 9 10

Voltage

UOIS

SIW

] O

N

Relative Emission Intensity

0.8— a)

0.6-

0.2—

I 1 1 1 0 100 200 300 400

[NaCl]

0.4— b)

a _o 0"

.1g 0.2—

6

this reason, and their high absorption coefficients, that the

alkali metal halides are a common cause of interference in many

applications of electrothermal atomisation to AAS. The molecular

absorption spectra obtained for the vaporisation of 10p1 aliquots

of the bromides, chlorides and iodides of sodium and potassium

are shown in Figs. 6.2, 6.3 and 6.4, respectively. Again, a

comparison between these spectra obtained here using the HGA 2000

graphite furnace and those of Wood (H2 - N

2 flame) and Koirtyohann

and Pickett (02 - H2 flame) serve to indicate that it is the molecular

diatomic halide which was responsible for the absorption. The

wavelengths of maximum absorption obtained in the molecular spectra

are listed in Table 6.1.

6.5 Absorption Spectra Of Common Inorganic Salts In The Small

Graphite Furnace Atomiser.

The emission spectrum obtained from the

deuterium arc operated under the conditions previously stated

1 and with a nitrogen flow-rate of 11.min through the atomiser

chamber is shown in Fig.6.5. Because of the vitreosil window

fitted to this lamp, no radiation of wavelengths shorter than

180nm was transmitted.

The stock solutions of the inorganic salts

were diluted as required and absorbance vs concentration analytical

curves determined. In all cases (with the exception of the

sulphates, discussed in Section 5.5) no deviation from linearity

was observed for absorbance values below 1.0. A typical calibration

curve is shown in Fig. 6.6 (a) for sodium chloride at 180nm and

the effect of atomising voltage on the absorption from 204tg.m1-1

Cl- as NaC1,at 180nm,in Fig.6.6 (b).

162

The chlorides, bromides and iodides of sodium and

potassium were diluted to provide working solutions of concentrations

such as to provide well defined absorption spectra from the inorganic

salts, (200,pg.m1 1 Cl as NaC1 and KC1, 500,11g.m1 1 Br- as KBr

and 1000,pg.m1-1 Br as NaBr, and 400 and 601491g.m1-1 I as NaI and

KI, respectively). The,absorbance vs wavelength molecular absorption

spectra observed from these salts using the small graphite tube

furnace at 8.5V are shown in Figs.6.7, 6.8 and 6.9. The wavelengths

of maximum absorption are tabulated in Table 6.1.

Lvov, as discussed in Chapter Five, suggested

that much of the background absorption observed with non-flame

electrothermal atomisers could be eliminated by cutting slits

into the graphite tube, enabling condensation processes to occur

out of line with the optical axis. The effect of this'technique

on the magnitude and nature of the molecular absorption signals

observed with the salts and apparatus used here was examined.

Two slits, ca. 1mm wide and 2.to 3mm deep, were cut into the tube

walls, one on either side of the sample introduction port. The

tube was then operated in the furnace in the normal manner. The

spectra of potassium chloride and sodium bromide were typical

of the effects of these slits and are shown in Fig.6.7 and 6.8.

As may be seen, the spectra obtained with

the smell graphite tube furnace are similar to those obtained with

the HGA 2000. The effect of the slits in the tube on the molecular

absorption was to decrease the magnitude of the absorption signal.

The presence of the slits in the

tube decreased also the atomic absorption signal from iodine as

KI and no sulphur A_1S signals were observed using tubes with slits.

163

164

2.20 180 190 200 210 Wavelength, nm

1 I 190 200

Wavelength, nm 220 210 180

Fig. 6.8

0.3 ® NaCI o KCI o KCI (slit tube)

200,pg.ml -1 CI

0 cn

0.1

Fig. 6.7

o NaBr o NaBr(slit tube)

1000„pg.m1-1Br

0.3—

KBr 1

500jug.ml Br

260 180 200 220 240

0.8

• K1 , I

o Na I , 400pg.ml I 0.6

0.2

Wavetength,nm

6.6 Conclusion.

Because of the common occurrence of the alali metal

halides, particularly sodium chloride, in sample matrices or in

solutions after dissolution of samples, it is frequently necessary

to separate the analyte element from these salts before analysis

by AAS. It is clear from the results presented here that the absorption

by such species has definite wavelength dependence, so that the

severity of the observed non-specific absorption by alakali metal

halides varies with the wavelength of the resonance line of the

analyte element to be determined. The use of background correction

minimises the errors observed provided that the interferant is not

present at excessively high concentrations. Ediger and Kerber (135)

have recently reported the minimisation of interferences caused

by molecular absorption in the presence of high concentrations of

sodium chloride by the use of ammonium nitrate addition to the sample,

viz.

NaC1 + NH4NO3

NaNO3

+ NH4C1

The ammonium chloride sublimes in the furnace at 335°C and, as

confirmed in this work, the residual sodium nitrate produces no

absorption over the wavelength range 190nm to 360nm.

Following the completion of the work described

here a paper by Surles and Culver (t36) has been published. This

paper discusses the molecular absorption spectra in the range 200nm

to 400nm, produced by the vaporisation of fluorides, chlorides and

iodides of sodium and potassium using a carbon rod atomiser. Where

comparisons are possible the molecular spectra obtained are similar

to those presented here and are included in Table 6.1.

166

Table 6.1 .Molecular Absorption Spectra Of Some Alkali Metal Halides.

Wavelengths Of Maximum Absorption (nm)

Salt Small

Tube

HGA

2000

Graphite

Rod'

:2 - N22

Flame

H2 - 02 Flamec

NaC1 197 N.O. N.R. 195 N.R. N.R. 238 235 237 240

NaBr 197 195 N.R. 200 N.R. N.R. 248 N.R. 247 N.R. N.R. 290 N.R. 285 N.R. N.R. N.O. N.R. 315 N.R.

Nal 195 195 N.R. 198 N.R. 220 222 220 228 N.R. N.R. 260 257 260 N.R. N.R. 325 325 325 N.R.

KCI 195 200 N.R. 200 N.R. N.R. 248 246 250 246

KBr N.O. 215 210 215 N.R. N.R. 280 278 275 275

KI 200 200 N.R. 205 N.R. N.R. 240 N.H. 238 238 N.R. 260 N.R. 260 N.O. N.R. 320 N.R. 325 320

N.R. = Not Reported.

N.O. = Not Observed. •

a = Ref. 13e

b = Ref. 131

c = Ref. 132.

167

SEVEN

168

7.1 Introduction.

The graphite furnace atomiser is an attractive

atom cell to the analyst employing AAS techniques because of its

often negligible background absorption and emission compared with

flame cells and its capability of atomising many refractory, oxide-

forming elements. However, despite the now widespread and routine

'application of such devices few measurements have been made of the

temperature attained by the atomic vapours produced in a non-flame

cell. Where studies relating to the kinetics and atom producing

mechanisms have been undertaken using graphite atomisers (137,v59),

it has been customary to accept the manufacturers calibration of

the temperature of the furnace and no results have been reported

concerning the variation in furnace temperature within its useful

lifetime. As the performance of a graphite furnace for AAS depends

ultimately on its heating rate and terminal temperature attained

during atomisation this lack of data concerning these systems is

surprising.

This chapter describes the determination of the terminal

temperature attained by the HGA 2000 graphite furnace atomiser at

various applied voltages and the heating rates associated with these

temperature programmes. Finally, a method of monitoring the temperature

of the atomic vapour contained within the graphite furnace is examined

and applied to the atomisation of gallium and indium.

7.2 Terminal Temperatures Attained By The HGA 2000 Graohite Furnace.

The HGA 2000 graphite furnace atomiser employed

in these studies was as described previously. The electrical power

to the furnace was supplied via a commercial transformer system

(Perkin-Elmer Ltd.) and furnace temperature selection achieved by

169

monitoring the readout-meter provided with this unit. This set

temperature was the approximate terminal temperature attained by

the furnace during its heating-cycle. Because of the change in

electrical resistance of a graphite tube throughout its useful

lifetime and because the open nature of this furnace gives rise

to a temperature gradient in the length of the tube it was necessary

to determine independently this terminal furnace temperature and

compare the observed values with the temperature indicated at the

meter.

At temperatures in excess of ca. 600°C a hot body

begins to emit radiation in the visible regions of the spectrum,

a dull red colour being observed. As the temperature increases the

apparent colour changes through yellow to the 'white' associated

with incandescent lamps. The optical pyrometer employs this visible

emission from a hot surface to measure the temperature of the surface

by comparison with a standard, calibrated filament; the disappearing-

filament pyrometer. A Northrup-Leeds optical pyrometer was employed

here.

The terminal temperature achieved by the graphite

furnace varies throughout its length (e.g. the centre is hotter

than the ends) and a reproducible viewing point was selected for

study with the pyrometer. As the sample for AAS is placed in the

furnace at its centre, and it is here that the sample remains until

volatilised, this was the point selected for viewing. By removing

the observation tube from the furnace housing the inside surface

of the graphite tube, directly opposite the sample introduction

port, was visible. The temperature of this spot in the furnace

was determined at various set temperatures ranging from 1000°C to

170

2200°C, in 200° steps, by focussing the radiation from the inner

wall of the furnace at the viewing lens cf the optical pyrometer.

These furnace temperature settings and the pyrometer temperature

measurements were repeated many times during the useful lifetime

of several graphite tubes whilst the furnace was employed for normal

analysis.

A comparison of the set and observed furnace temperatures

is presented graphically in Fig.7.1. It was apparent that the

observed temperature of the furnace deviated from the temperature

selected by upto ± 10% depending on the age of the graphite tube

and the set temperature. The mean observed temperature, throughout

the useful lifetime of the furnace, was between 3% and 7% greater

than the set temperature; the deviation increasing with increasing

temperature.

Because of the deviation in observed and set temperatures

and to avoid confusion the term 'furnace temperature' used in the

remainder of this chapter refers to the optical pyrometer temperature

and not the temperature as provided by the meter on the furnace

power supply. Where reference to the power settings is necessary

the voltage applied across the graphite tube is quoted as this

is independent of the condition of the furnace.

7.3 Temperature—Time Profiles Of The IRA 2000 Furnace.

Findlay et al. Om ) have described a method for

monitoring the furnace temperature with respect to time using

thermocouples placed inside the graphite tube. With this apparatus

temperatures upto 1700°C were measured and considerable temperature

gradients were observed to exist, both radially and axially, in the

HGA 2000 furnace. Most atomic absorption studies, however, are

171

/- 0

3000–

/

/A

/ /0

0

.1 (D 2000– / /

, // E / / A = Towards end of tube lifetime

0 / / / B = Mean, throughout lifetime

>, ‘ Ci / . --

/ •

o 1000 1000 2 000 3000

Meter (Perkin-Eimer) Temp. °C Fig. 7.1

N

L_

/ / C = New tube

2–

E

0

0 E w 0 —1–

Blackbody Emission From The HGA 2000

5.0 5.15

1 1 4.0 4.5

1/ T x104

-2–

Fig. 7. 2 172

undertaken at furnace temperatures in excess of 1800°G and a means

of monitoring the temperature-time characteriStics of the graphite

furnace applicable to these higher temperatures was sought. Johnson

014.0) has recently conducted a study of the temperature-time profiles

of a carbon filament atomiser using the Planck's law relationship

between emission intensity of a hot body and its temperature to

determine these profiles. The application of this method to the

HGA 2000 will be described here.

The intensity, I, of the radiation emitted at

wavelength by a blackbody at temperature T is given by Planck's

law,

B -lih c i.e. ... 7.1

For any particular instrumental arrangement

and measuring the emission intensity from the blackbody as a voltage

(VV) produced by a PMT, then,

Vi, = A>, . cy ... 7.2

where A). is a constant for a given instrumental system. For the

wavelength of 600nm and the temperatures considered here (2000K to

3000K), Eq. 7.2 reduces to,

VA = A A. e -ricr ... 7.3

and

t0.9 V 103 A>, - >c,t1, k4.7. ... 7.4

i.e. a plot of log Vx vs 1/t is linear and allows the temperature

173

of the emitter to be determined from the emission intensity measured

as voltage from a suitable radiation detector system.

The Perkin-Elmer Model 305B Spectrometer was

operated in the emission mode with a slit-setting N° 3 and constant

gain applied to the PMT. At a wavelength of 600nm the emission

from the furnace was monitored with respect to elapsed time from

the start of the heating cycle and at various applied voltages across

the graphite tube. The emission intensity vs time curves were recorded

on a potentiometric chart recorder the scale of which was set such

that no damping of the signal was observed by the recorder time-

constant. The terminal temperature attained by the furnace was

determined at each applied voltage using the disappearing-filament

optical pyrometer as previously described.

A plot of log(peak emission intensity, mV) vs

1/T (K) is shown in Fig.7.2 and is a straight line. Employing this

linear relationship it was possible to convert the emission intensity-

time curves to temperature-time curves and the results are presented

in Fig. 7.3. Temperatures belcw 10000C were difficult to measure

with an acceptable accuracy using the optical pyrometer and linear

extrapolation to room temperature was assumed. The validity of this

assumption was verified by a heating rate (°/sec) vs (applied voltage)2

plot for this region, Fig. 7.4, which was linear as expected.

7.4 Measurement Of Electronic Excitation Temoeratures Of Atomic

Species By A Two-Line Atomic Absorption Method.

Browner and Winefordner(144 )have described

a two-line atomic absorption technique for the measurement of flame

temperatures which is applicable to non-flame atomisation devices.

It is a straightforward atomic absorption method, using an uncalibrated

174

20. 10

Fig.7.3 1000—

Furnace Heating Rate vs(Applied Voltage)2

V2 0 I

0 50

Fig.7.4 175

100

3000

ci E 2000

a) a) E 0 '5,1000 C-

0 U

Ct. 0

8.0V 6.75V

5.50V

4.25V 3.0V

HGA 2000 Temperature—Time Curves.

time(sec)

Y2.

La e.2 1. c)a. n a_ X3)]

c c,2

..o 7.7

continuum light source, that is relatively simple, rapid and

reasonably accurate. The technique is applicable to any atomic

absorption spectrometer without any modification. The theoretical

basis of this two-line atomic absorption temperature measurement

technique is briefly discussed.

The absorption factor,M, is defined as

the fraction of incident radiation absorbed by the analyte atoms,

i.e.

Absorbed radiation intensity = I0 -I =AI ... 7.5 Incident radiation intensity

Io I

For atomic absorption using a continuum emission source two limiting

cases may be considered. In the limiting case of low optical density,

oc --= ... 7.6

where e and m are the charge and mass of the electron respectively,

c is the velocity of light, 1 is the absorption path length,ikois

the peak wavelength of absorption, n is the concentration of atoms

at the lower energy level involved in the transition, f is the

absorption oscillator strength and s is the spectral bandwidth of

the monochromator.

At high optical densities Eq.7.6 becomes,

whereapis the Doppler half-width of the line and a. is the damping

constant, defined by,

0_ = #

176

Ct. = 11,2 r AX + NA4

... 7.8

As discussed in Chapter Five the natural half-width of a spectral

transition may be neglected in comparison with the other broadening

factors and substituting for a in 1:q.7.7,

0.< 1 2.-Tr a2t -n. S 2 rn

V2

... 7.9

Equations 7.7 and 7.9 do not hold at absorbance values greater than

0.3 as parts of the absorption line wings may extend beyond the

spectral bandwidth of the monochromator.

When an atomic vapour is in a state of thermal

equilibrium with its surroundings, the relative populations no and

n1 of two electronic levels 0 and 1 are related by,

Tt I no

90 emc (E° E kV ... 7.10

Eo and '

1 are the energy levels of the states 0 and 1 and go

and -

g1 their statistical weights, k is the Boltzmann constant, and T

is the absolute temperature.

For an atomic species having one or more

electronic levels close to the ground state it is possible to relate

the relative absorption at the resonance line and at a non-resonance

line to the relative populations of these levels by combining Eqs. 7.6

and 7.10.

177

i.e. At low absorbance values,

T Et — E.o

k In 1( 91c-1 (L)2(_c_=.21 9000 A X01 04, Ij

... 7.11

and at high absorbance values,

T Et — Eo

IC trIPLi)(112V4212( AXL‘ 9 -fp Xb 04, ... 7.12

where Ocand V. referto the relative absorption signals at the lines

of wavelength Xo and XI whose lower states have energies Eo and E1

and whose Lorentz half-widths are LXL. andOkLi. To achieve optimum

conditions for measuring p( the energy separations of 0 and 1 should

be less than 1eV for flame temperatures below 4000K.

Of those elements with suitable energy levels

and f-values the lines of the first term of the sharp series of

gallium (403.30nm and 417.20nm), indium (410.18nm and 451.13nm), and

thallium (377.57nm and 535.05nm) were the most suitable for practical

measurements. The physical constants relating to these transitions

are shown in Table 7.1. The ratio of 06/0c, for each line pair

is an exponential function of temperature, Fig.7.5. In practice °C/,

may be measured with good precision over the range 1 G (44es 4. 5

and this restricts the working ranges of temperature for the elements

to:

Gallium, 650 to 2200K

Indium, 1220 to 3500K

Thallium, 3650 to 7750K

Without knowledge of the Lorentz half-

widths the low absorbance limiting case must be used for temperature

178

Ga

Variation Ofc< With Absolute Temperature,

8000—

6000—

a)

' 4000— 5 a)

E a)

2000—

T1

0 2 4 6 GC

0/X,

1

Fig.7.5

179

measurements and as this implies absorption measurements of less

than the 1 level great inaccuracies are to be expected with most

non-flame cello due to background absorption, etc.

In order to determine temperatures by the

measurements of ccit at high optical densities it is necessary to

know the ratio 4.1,1 ,/,' 1.0. It has been shown (OW that if the growth

curves (logo( vs log concentration) for the two wavelengths employed

for each element are parallel and if the intersection point of low

(slope = 1) and the high (slope =. 0 absorption asymptotes occurs

at the same concentration then Eq. 7.12 reduces to Eq. 7.11. Under

these conditions more accurate and repi.oducible absorption measurements

may be made.

7.5 'The Electronic Excitation Temperature Of The Atomic Vapour

Of Gallium And Indium Produced In The HGA 2000.

A double-beam atomic absorption spectrometer

(Model 305B, Perkin-Elmer Ltd.) was employed for all absorption •

measurements. The continuum radiation source employed was an

uncalibrated, 150W, quartz-halogen lamp. The instrumental slit -

width was set at N° 2 setting as this allowed adequate absorption

and sensitivity without high PMT gain. The absorption signals from

the spectrometer were recorded on a potentiometric chart-recorder.

The commercial, non-flame HGA 2000 graphite furnace atomiser unit

was employed as the atom cell. The continuum source was positioned

in the lamp housing, in place of the normal hollow cathode lamp,

such that the radiation from the source was focussed in the centre of

the graphite furnace. Unless otherwise stated, argon was used as

the furnace purge gas, under automatic stop-flow conditions at flow-

rate N° 4.

Stock solutions of gallium and indium were

180

prepared from the pure metals by dissolution in ca. 3C% analytical

% grade nitric acid. The stock solutions (10,009mg.m1 1) were diluted

as necessary using distilled water. A 10Al aliquot of the required

sample was transferred to the graphite tube using an Eppendorf

micropipette. The solution was dried at 100°C for ca. 30sec prior

to atomisation with a voltage of 5.5V applied across the furnace.

As a precaution against monochromator drift during

the absorption studies with a continuum source the following procedure

was employed for the indium solution analysis. Using an indium

hollow cathode lampline source the line at 410.18nm was selected

and the monochromator set. The HCL was replaced by the continuum

lamp and the absorption from an indium solution determined from two

samples of equal indium concentration. The 451.13nm indium line

was then selected, again using the HCL, and the absorption by the

same concentration of indium solution determined. The 410.18nm

line was re-selected and a second concentration of indium solution

examined. By this procedure of constant checking of the wavelength

with the aid of an indium HCL it was expected problems with wavelength

drift to be negligible. Alternate measurements ofa at the resonance

line and non-resonance line ensured that any change in the heating

rate of the furnace associated with its use were smoothed-out.

Because of the change in furnace properties with use (as described

in Section 7.2) the absorption measurements were repeated many times

(more than ten) over a period of several weeks and the mean absorption

peak heights recorded for various indium solution concentrations.

An identical procedure was employed for the gallium absorption

studies.

Growth curves were constructed for both

181

metals at each line and these are shown in Figs.7.6 and 7.7 for

gallium and indium respectively. The inflection points of these

curves are shown in Table 7.2.

The inflection points for both metals

occurred at the same concentration for the resonance and non-resonance

lines; therefore the approximation to reduce the high absorption

equation (Eq.7.12) to the form of the low absorbance equation (Eq.7.11)

could be applied. Substituting the constants from Table 7.1 into

Eq. 7.11 for gallium and indium prOvides Eqs. 7.13 and 7.14.

Indium

16.3_ ___ T(K) = 7;;/ ) log(1.08 06

Gallium ... 7.13

T(K) — 1383 ... 7.14 log(2.637%)

The growth curve plots shown in Figs.7.6

and 7.7 include the atomic vapour temperatures determined using the

points of the curves and Eqs. 7.13 and 7.14. The results are

summarised in Table 7.3.

Using both gallium and indium as the sensing

elements the atomic vapour temperatures in each case were similar

and were ca. 15% below the furnace wall temperature. As mentioned

previously the working temperature range for gallium is 65000 to

220000 and hence the poorer precision of the results obtained with

this metal are understandable. The two-line atomic absorption

temperature measurements in Table 7.3 were determined by recording

the absorption peak heights. Thus, no information relating to the

variation in temperature of the atonic vapour produced ins the7rapnits

furnace with respect to time was obtained. From the above results

it cannot be said that the recorded temperatures were the maximum

182

1975 K 162—

lo 10 10

10 —

417.20nm 2300 K

2200 K 2200 K

2050 K

2000 K

1950 K

1950 K

1950 3 10

10-1

Gallium

101 —

103

10 -1 10 10

p 1g.m1 Ga

Curves Of Growth For 14L1 Samples Of Gallium

Fig. 7.6 183

(a) 410.18 nm

(b) 451.13nm

oc

-2 10

1900 K

Indium

-3 10 10 1 1 1 2 -1

10 10 10 10

pg.mt 1 In

Curves Of Growth For 10,1 Samples Of Indium.

84.

Table 7.1 Physical Constants For Ga, In and Tl.

Element Wavelength (nm) gf Transition (cm-1)

Gallium 403.29 0.258 0 - 24,789

417.20 0.405 825 - 24,789

Indium 410.18 0.288 0 - 24,373

451.13 0.628 2,213 - 24,373

Thallium 377.57 0.250 0 - 26,478

535.05 0.540 7,793 - 26,478

Table 7.2 Inflection Points For Growth Curves.

Element

Wavelength (nm) Inflection Point (pg;m1-1)

Gallium 403.29 1.5 417.20 1 . 5

Indium 410.18 1.0 451.13 1.1

185

atomic temperatures but merely the atom temperatures corresponding

to maximum absorption.

An experimental method for the determination

of the temperature-time characteristic profiles of an atomic vapour

produced in a graphite furnace and the results obtained are presented

in the next section.

7.6 Temperature-Time Profiles Of An Atomic Vapour In The HGA 2000.

Because of the useful working range (1220K to

3500K) indium was selected as the study element. A standard solution

of.2yg.m1 -1 indium was employed for all absorption measurements.

A logl aliquot of the 2yg.m1-1 indium solution

was pipetted into the graphite tube and following the drying stage

(30sec at 100°C) the potentiometric chart recorder and atomisation

programme were activated simultaneously. Thus an absorption vs time

profile was recorded. The alternate resonance and non-resonance

line wavelength selection previously discussed in Section 7.5 was

employed to minimise instrumental drift.

The absorption-time profiles were obtained

for the indium resonance line (410.18nm) and non-resonance line (451.13nm)

at various applied voltages across the graphite tube. 04 was

measured at various times from the absorption-time profiles and

hence an atomic vapour temperature-time profile determined at each

of the applied furnace voltages.

The resultant temperature-time profile curves

of the indium atomic vapour in the HGA 2000 at applied voltages

across the furnace of 4.25V, 5.507, 6.757 and 8.00V are shown in

Figs. 7.8 (a), (b), (c) and (d) respectively. Also included with

these figures are the temperature-time curves for the furnace as

determined with the optical pyrometer. Finally, as a further comparison,

186

187

10 12 I i I 1 2 4 6 8

time (sec)

—0.03

—0.02

—0.01

2 000—

1000—

2 4 6 8 time(sec)

10 12

3000—

2000

6-1 000 - E a)

451.13nm

(.)

0 _D

—0.02

—0.01

4.25 V

Fig. 7.8

0 21 4

I 61

time (sec) Fig.7.8

8 10 12

188

6.75 V Furnace Temp.

2000—

Atom Temp.

1000—

410.18nm

451.13rim 1 8 10

Furnace Temp.

ci.. E 0 a) 3 000

2 4 8.0 V

6

2000— Atom Temp.

1000—

410.18nm

451.13nm

—0.03

_0.02

—0.01

12

—0.03

—0.02

—0.01

Abs

orba

nce

Table 7.3 Atomic Vapour and Furnace Temperature Measurements.

element Applied Voltage Set Temp.a Furnace Temp.b

Atom Tempc

R.S.D.

Gallium 5.5 2273 2413 2047 6.06%

Indium 5.5 2273 2413 2050 4.72%

a = Temperature as indicated on the instrument meter (K)

b = Temperature as determined using optical pyrometer (K)

c = Temperature as determined using two-line AAS technique (K)

Table 7.4 Some Temperature Measurements With The HGA 2000

Applied Voltage Set Temp.a Furnace Temp.b Max. Atom Temp.c

4.25 2073 2173 1830

5.50 2273 2413 1950

6.75 . 2473 2633 2070

8.00 2673 2853 2100

a = Terminal meter temperature (K), for the applied voltage.

b = Terminal temperature (K) determined using an optical pyrometer.

c = Maximrm indium atom temperature determined by two-line AAS.(K)

189

the absorbance-time profiles are included for the resonance and

non-resonance absorption processes. Some significant results obtained

from these curves are presented in Table 7.4.

The atomic vapour temperature-time profiles

were of similar distribution to the absorption-time curves,i.e.

rising rapidly to a maximum value, corresponding to maximum absorption

or temperature and falling to lower absorption or temperature values

more gradually with increasing time. This decay, however, was less

pronounced for the temperature-time curves than for the absorption-

time profiles.

7.7 Discussion And Conclusion.

The temperature-time characteristics of

a resistively heated graphite electrode (such as a rod or tube

furnace) depend on the applied voltage, electrode length and cross-

sectional area, and the density, specific heat, thermal conductivity,

resistivity and emissivity of the electrode material. The electrical

power supplied to a graphite furnace, PIN, may be dissipated by

conduction, Pc, and radiation, PR, as well as supplying power to

raise the furnace temperature, PT, i.e.

PIN = PT + PR + PC

Many of the functions controlling the

magnitude of these power terms are themselves dependent and any

theoretical solution of the nature of the temperature-time profiles

of graphite furnaces must take this into account. Cresser and Mullins

Oka) applied a simplified theoretical approach to calculate these

profiles for graphite and tungsten filaments. The temperature-time

characteristic curves shown in Fig.7.3 for the HGA 2000 graphite

190

tube atomiser were similar to those obtained by Cresser, i.e. a

linear relationship existed between furnace temperature and time

until temperatures were reached when power losses due to radiation

and thermal conduction became significant.

The twoline atomic absorption method of

temperature determination has been applied to the study of the atom-

isation of gallium and indium in the HGA 2000 graphite furnace.

Bratzel and ChaRrabarti ( i4-3) applied this method from peak-absorption

measurements to the determination of temperatures of atomic vapours

produced by graphite filament and graphite rod atomisers. Their

studies indicated that gallium gave higher atomic vapour temperatures

than indium; for indium the concentrational location of the inflection

points between the low and the high absorption portions of the growth

curves were not the same for the resonance and non-resonance lines

and low absorption figures only could be used. The higher temperature

achieved for gallium was attributed by these workers to the greater

boiling point (2676K) of this metal than that of indium (2353K),

implying that because gallium is in direct contact with the surface

of the rod for a longer time, the temperature attained was greater.

The temperature studies reported here using

the EGA 2000 graphite tube show similar temperatures for both gallium

and indium (Table 7.3) and both species fulfilled the equilibrium

condition of similar concentrational inflection points between

the resonance and non-resonance transitions in their growth curves.

These facts probably relate to the increased tendancy to thermal

equilibrium using the larger graphite tube atomiser and the increased

atom residence times associated with these systems.

Temperature measurements derived solely

from peak absorption methods provide no information concerning the

191

change in atom temperature with time during the heating cycle of a

furnace. Using indium as the study element temperature-time profiles

of the atomic vapour were determined for various applied voltages

across•the graphite tube, Fig.7.8. In all cases the mean temperature

of the atomic vapour in the furnace increased with the increase

in furnace wall temperature, achieved a maximum temperature at the

same time as the absorption maximum occurred and finally tailed-off

in a manner similar to, but less dramatic than, the absorption curves.

L'vov (116 ) has shown that the atomic absorption signal duration

using a tube furnace is controlled mainly by the rate of atom-loss

by diffusion from the furnace. Because of the open nature of the

HGA 2000 tube furnace a severe temperature gradient is likely to

exist along the length of the hot tube and the decreasing rate of

atomic vaoour temperature following its maximum value may be attributable

to the diffusion process controlling the absorption curve. That

this decreasing rate is less severe than observed for the absorption

process may be due to collisions with the hot furnace walls or

purge-gas molecules or atoms.

A study of the temperature-time curves employing

various purge-gases (e.g. N2, Al.., Hel Ar-H2, etc) would provide

further information on atom formation and residence processes in

non-flame, tube furnaces.

192

EIGHT

193

8.1. Introduction.

Analytical atomic spectrometry depends for its

sensitivity and selectivity on the separation of the required spectral

lines from any background spectrum and the detection and monitoring

of this radiation of interest. For example, the selectivity of

atomic emission analysis is subject to the ability to isolate the

analyte emission from any source background emission or sample

matrix emission. The sensitivity of A7_,S is controlled ultimately

by the capability of the detector to monitor this required emission.

Although in AAS the atomic vapour produced in the flame or electrothermal

cell serves as a very narrow bandpass filter such that absorption

only occurs at the resonance lines emitted by the source, a dispersion

element or filter is required to isolate these resonance transitions.

As discussed in Chapter One the most common dispersion

element and detector assembly for analytical spectrometry is the

monochromator and photomultiplier—tube arrangement. Whilst these

systems are relatively simple when concerned with visible and middle

ultraviolet radiation, in the vacuum ultraviolet far more sophisticated

apparatus is required. Several workers have described the adaptation

of commercial spectrophotometers to the region 180nm to 200nm,

usually by simply removing the atmospheric oxygen from the monochromator

by inert gas purging. However, in general, these techniques for

extending the wavelength range of the apparatus are of little value

at wavelengths shorter than 180nm. At these low wavelengths ent-

rainment of oxygen and water vapour in the monochromator and optical

path causes a severe loss in transmission and vacuum systems are

normally required.

The atoms of many elements, including the halogens,

194

carbon, sulphur, oxygen, nitrogen, phosphorus and mercury emit

intense resonance line radiation in the region 120nm to 200nm.

The possibility of providing a simple, non-dispersive system for

the detection of radiation from the atomic line emission of these

elements would be attractive for their detection and determination

at these low wavelengths by the techniques of atomic emission,

absorption and fluorescence spectrometry. By analogy with the

advantages of the adoption of non-dispersive detection systems

rather than grating or prism monochromator-detectors in analytical

atomic spectroscopy in the near u.v. and visible regions of the

spectrum, the use of a non-dispersive detection system in the vacuum

ultraviolet might be expected to yield advantages of simplicity,

large aperture and suitability for operation in simultaneous multi-

element analysis.

This chapter discusses the theory and describes

the construction, use and preliminary evaluation of photoionisation

detectors for the detection of radiation in the Schumann U.V. from

a microwave excited argon plasma at atmospheric pressure or from

a flow-through demountable hollow cathode lamp source. The photo-

iOnisation detector (PID) appears to offer the possibilities for

non-dispersive spectrometry in the vacuum ultraviolet region.

These detectors can be constructed to detect radiation in the far

ultrviolet regions of the spectrum and their useful properties

include high spectral selectivity, high quantum efficiencies and

very low noise levels. A typical PID consists of a chamber containing

a gas or vapour at low pressure and two electrodes across which

a constant d. c. potential is applied. When the chamber is irradiated

by photons whose energy is sufficient to cause ionisation of the

195

molecules of the gas or vapour an ionisation current is produced

and may be registered in the external circuit. The magnitude of

the current is proportional to the incident photon flux at those

wavelengths capable of causing ionisation. The spectral response

of a PID is controlled on the short wavelength (high energy) side

by the transmission characteristics of the material from which the

window is made and on the long wavelength (low energy) side by the

photoionisation threshold of the filler gas or vapour. By careful

selection of the window material and the filler gas, detectors with

a spectral bandpass of only a few nanometres may be constructed

to detect radiation in the wavelength range 105nm to 200nm. They

may be made to have large optical aperture, are 'solar—blind' and

may be rugged and compact.

To date the use of such detectors has been

limited to rocket use 045.) including observations of the solar

spectrum, and the determination of molecular oxygen densities in

the atmosphere by absorption measurements and the intensity of

solar Lyman—gemission and adjacent u.v. emission lines, (144,14(7).

Although Chubb and Friedman (141) have described the determination

of water vapour by its absorption at 121.6nra in air and Strober et

al. (448) have described commercially available PID chambers to

detect radiation in the region 105nm to 140nm.

8.2 Theory of Photoionisation Detectors.

It has been previously mentioned that the spectral

response of a PID is dependent on the transmission characteristics

of the window material employed and the photoionisation characteristics

of the filling gas or vapour. Fig.8.1 shows schematically a typical

parallel plate photoionisation chamber. A quantum of radiation

196

Fig. 8.1 The Action Of A ?ID.

by

I. C

C 0

PID Action

Fig.8,2 Current—Voltage Characteristics.

gas gain"

plateau'

C

Applied Voltage

197

transmitted by the window, of sufficient energy to promote ionisation

in the gas contained within the chamber, will produce an ion-pair

if the radiation is absorbed. Provided a minimum voltage (V1) is

applied between the plate electrodes these charged species will be

collected by the electrodes and a current will flow in the circuit.

If the applied voltage (V) is less than V1 recombination of the

ion-pairs is probable and the ion-collecting efficiency is impaired.

a) The Current Produced by a PR).

We can define the photoionisation

efficiency, or photoionisation yield, (t), as the number of ion-

pairs produced per photon absorbed and, for incident radiation of

wavelengths shorter than the. ionisation threshold of the gas,

is approximately unity for many gases and vapours. (It is interesting

to note that should the absorbed radiation be sufficiently energetic

to cause doubly ionised species then r may be greater than unity). Thus the current, i (amperes), generated by the BID is proportional

to the number of ion-pairs, Ni, produced by the incident radiation

which, in turn, is proportional to the number of photons per second

absorbed, N , by the gas and the photoionisation yieldlt.

i.e.

8.1

The absorption of the radiation by the gas or vapour in the PID is defined

by the Beer-Lambert Law,

I = Io. exp ( - dt.n.l ) 8.2

. Where I (photons.sec 1 ) is the intensity of the incident radiation

entering the gas, I is the transmitted radiation and n is the number

of atoms or molecules of gas per unit volume in the cell of length 1.

198

n = no P.To

8.3

P0 .T

and no = Loschmidt's number (2.69 x 1019 atoms or molecules.cm3 )

P and T are the pressure and temperature of the gas or vapour

and, Po and To are the S.T.P. pressure and temperature.

dt is the total absorption cross-section which is related to the

photoionisation cross-section) d., by

di = . t

assuming that scattering of radiation is negligible.

As

Np = Io - I

then from Eq. 8.2

... 8.4

... 8.5

N = Io (1 - exp( -d .n.1)) ... 8.6

and substituting for dt from Eq. 8.4,

Np = I0 (1 - exp( -diAt.n.1)) ... 8.7

To take into account the transmission characteristics of the window

material Io in Eq. 8.7 may be replaced by,

Io = Io'.T ... 8.8

Where I' is the intensity of the incident radiation (photons.sec 1)

at the window outer face and T is the transmittance of the window

material at the wavelengths of interest.

Substituting in Sq. 8.7,

N=II.T(1-exg-d.'e.1)) ... 8.9

199

From Eq. 8.1 and 8.9,

Ni = I.T. (1 - exp(-6i/t.n.1)) ... 8.10

If the current produced by the PID is measured in amperes and e

is the electronic charge,

N. = i/e ... 8.11

61 . for the gas or vapour contained within the chamber is not always

known. However, if the parameters of the chamber and the gas pressure

are chosen such that the incident radiation is totally absorbed

by the gas then,

exp (-6i/t.n.1) ---r 0 as n.1 ---r 00

and under these conditions,

1 , i/e = T.Io.0 sec-1

or i = Il.T.e.t amperes ... 8.12

Equation 8.12 represents the total current flowing in the ion chamber

when all the ions formed are collected and provided the applied

potential difference between the electrodes is not so large as to

cause secondary, collisional ionisation; ion-multiplication. Davis

and Braun (101) investigated a flow-through, reduced pressure

microwave plasma source in helium as a vacuum ultraviolet atomic

line source and reported absolute line emission intensities of

1013 to 10

15 quanta.sec

-1. Assuming these values to be typical

of microwave plasma emission intensities it may be seen from Llq.8.12

that currents as great as 10 4 amperes could be expected from a PID.

200

b) Ion-Y.u2tielication.

Fig.8.2 shows a typical ion-current against

applied voltage curve for a PID. The geometry of the PID, the

pressure of the gas or vapour in the chamber and the incident radiation

intensity are all constant. Referring to Fig.8.2, in the applied

voltage range between A and B the ions formed by the reaction,

M + hv + e-

are collected with increasing efficiency as the fieldstrength in

the chamber increases with increasing voltage and the ion-current

rapidly saturates at an applied potential of a few volts. Further

increase in applied voltage, in the range B to C, can not result

in further collection efficiency for the given geometry and cell

dimensions and a plateau region of relatively constant output current

is obtained. In practice this plateau region may extend over a

voltage range of 100 to 200 volts. At higher applied voltages, as

in the region of C and above, the electrons produced by the initial

photoionisation of the gas or vapour may achieve sufficient energy

in the stronger electric field to cause secondary ionisation on

collision with a gas molecule, i.e. the energy of a fraction of the

photoelectrons becomes greater than the ionisation potential of the

filler gas. As the applied voltage is increased the increasing

field strength thus leads to electron multiplication by secondary

ionisation and a consequent rapid increase in the current produced

in the external circuit. This region is often referred to as the

'gas-gain' or ion-multiplication region and can be employed to

increase the sensitivity of the detector system.

The most effective design for a PID operating in

201

the gas-gain mode is a cylindrical outer electrode with a thin

central wire. The electric field of such a coaxial arrangement

is given by,

E = V

... 8.13

r In (a/b)

Where E is the electric field at a distance r from the centre of

the PID, V is the applied voltage and a and b are the radii of the

outer cylindrical electrode and thin wire electrode respectively.

It can be seen that the field strength increases rapidly as r approaches

b, i.e. the electron, or positive ion, is accelerated towards the

central wire.

If the PID is operated in the plateau region it is

often immaterial whether the ions are collected or the electrons.

However, when operating in the gas-gain region it is best for the

applied voltage to be arranged such that electrons and not ions

are driven to the central wire as electrons have greater mean-free

paths than ions and are responsible for the initiation of the ion-

multiplication process. If the polarity of the applied field is

reversed higher voltages are necessary to provide ion-multiplication.

c) The Spectrn1 Response of a PID at Wavelengths Near The

Ionisation Threshold.

The theoretical treatment of the variation in

the efficiency of photoionisation in the region where the energy

of the ionising photons exceeds the ionisation potential of the

gas or vapour by only a few eV has been discussed by several workers,

(1st), i51052 ). A qualitative description is sufficient here.

For a covalently bonded molecule R1R2

three

different photoprocesses need to be considered OW,

202

i. Ionisation to the ground state of the ion,

R1R2

R1R2.

+ e

or ionisation to an excited state of the ion,

R1R2

2 + e

_

ii. Fragmentation of the parent ion,

R1R2

R1R2 e

R+1 + R

2

or dissociative ionisation,

R1R2

n + 1 + R2

Preionisation of the excited molecule,

e-

The

12 _ _ R R+

-1R P 1 1 2 +

The spectral thresholds of these processes are

different in general, usually that for the simple photoionisation

being of the lowest energy. We may define two different values of

ionisation potential, IP.

i. Adiabatic IP - This corresponds to a transition from the zero-

vibration level of the ground state molecule to the zero-vibration

level of the ground state of the ion.

ii. Vertical IP- This corresponds to the most probable transition

from the ground state of the molecule to that of the ion.

If the interatomic distance in the molecule

is widely different from in the ion then the vertical transition

between the respective potential energy surfaces will lead to a

203

nitric oxide Fig. 8.3

120 130 X nm

140

benzene

Fig. 8.4 (a)

eV

9

10 liT

aniline

-6%

Fig. 8.4 (b)

eV

9 10 11

204

vibrationally excited ground state of the positive ion; in

accordance with the Franc-Condon principle.

When photoionisation proceeds only from a single

vibrational level of the molecule to a single level of the ion the

threshold law for the ionisation efficiency is expressed by a step

function. The photoionisation efficiency is zero when hv - IP < 0

(where hv is the energy of the photon). A jump occurs when hv = IP

and it remains constant for hv - IP > 0. Fig. 8.3 shows the variation

of the photoionisation cross-section, 61, as a function of incident

radiation wavelength for nitric oxide, (050). Several well marked •

steps corresponding to the positions of the vibrational levels of

the ion are evident and the portion of the curve within the limits

of one vibrational level is well defined by a step function. The

jump in cfi at the longest wavelength corresponds to the 0-0 transition,

i.e. the adiabatic IP of nitric oxide. The other steps on the di

curve correspond to transitions to the higher vibrational levels

of the ion.

In the ionisation of more complex molecules the curves

for the photoionisation efficiency become considerably more complex,

as at temperatures considerably above OK the molecules always have

a supply of vibrational energy. The occupancy of these vibrational

levels will obey approximately the Boltzmann Law. Ionisation from

these vibrational levels gives rise to an ion current at photon

energies lower than the first adiabatic ionisation potential. If

the probabilities of ionisation from all vibrational levels are

equal the increase in the photoionisation efficiency must follow

an exponential law for increasing photon energy. In general, these

levels are not resolved and the ionisation originating from them

205

results in a smoothing-out of the abrupt jumps in the ion current,

both in the region of the transitions to the various excited levels

of the ion and in the region of the adiabatic ionisation potential.

In the case in which the interatomic distances

in the molecule and the ion differ little the probability of tran-

sition to the higher. vibrational levels is smaller than to the

lowest level and an abrupt increase in the ion current at the- ionisation

threshold is observed which tapers off as the photon energy is

increased, eg. benzene, Fig. 8.4a. In the converse case, in which

the probability of transition to the higher levels is greater,

corresponding to the case of differing interatomic distances in the

ion and molecule, the observed rise in ion current with increasing

photon energy is more gradual becoming steeper with increasing

energy, eg. aniline, Fig. 8.4b. In both cases, at higher photon

energies, the region of the ionisation continuum is achieved and

the ionisation yield varies more smoothly with energy and the nature

of the variation is determined by the structure of the energy levels

and the probability of transitions between them.

8.3 Construction of a Photoionisation Detector.

The PID assembly used in this work is shown

in Fig.8.5. A Pyrex cylinder (90mm in length, 29mm diameter, 2mm

wall thickness) fitted with a B14/23 ground glass cone and socket

on one end and a side-arm fitted with a three-way 2mm bore tap was

employed as the detector envelope. The electrode assembly, mounted

onto glass-to-metal sealing rods through the ground glass cone,

consisted of a cylindrical sheet copper cathode (80mm in length,

25mm internal diameter, 0.19mm thickness) along the central axis

of which a 1mm tungsten wire anode was located. The electrodes

206

three way tap

ground glass joints ethylamine envelope

p.t.f.e. supports Ca F2

window

glass to metal seal central tungsten

wire electrode cylindrical copper electrode

O ^-2

0

0 (D

UO

T '4

.r.7. S

I LI O

T 0 q

.07.4

c1 e

tT a,

were insulated from each other by means of the glass and PI T-2 anode

supports shown in Fig.8.5. A polished calcium fluoride window

(3mm thickness, 25mm diameter) was mounted in a cylindrical aluminium

holder sealed with epoxy resin cement onto the end of the PID envelope.

The aluminium window holder was fitted externally with an annular

depression in which an 0-ring was positioned to permit vacuum sealing

of the detector at the exit slit of the vacuum monochromator employed

in some of the work. All the components of the PID were carefully

cleaned and degreased before assembly. The ground glass joints

were sealed with high temperature wax.

A d.c. voltage (5 to 400 volts) was provided

across the detector from a dry battery supply. The photoionisation

current was registered using a picoammeter (type P24, Knick, Berlin)

and the voltage output from this device was displayed on a potentio-

metric chart recorder.

To evaluate the potential of this assembly

as a detector of far u.v. radiation it was necessary first to examine

a variety of ultraviolet sources before selecting suitable filling

gases for the PID.

8.4 Vacuum Ultraviolet Line Emission Sources.

a) Microwave Plasma Source.

Davis and Braun (141) have described an

intense vacuum u.v. atomic line emission source produced by microwave

excitation of various gases in helium under flow conditions at

reduced pressures. The line emission from excited carbon, nitrogen,

oxygen, sulphur, selenium, chlorine, bromine, hydrogen and krypton

were examined and the spectra observed between 120nm and 200nm

were free from spectral impurities. Reduced pressure, flow-through

208

plasmas, however, require more elaborate construction and operation

than .atmospheric pressure systems. The plasma used in the studies

described here was a low-powered, microwave excited argon plasma

supported in a fused silica tube (150mm length, 2mm i.d.) within

a 4-wave resonant cavity. This assembly is shown in Fig.8.6. The

. 1 argon flow-rate was variable between 0.5 and 1.0 litre.= by

means of a Teter-Rate type C flow-meter and the plasma discharge

was viewed end-on along the central axis of the silica tube. Power

(ca. 50 watts) was supplied at 24501,21z from the 200 watt microwave

power supply previously described.

The spectral emission characteristics of the

plasma source were established using the 1-metre normal incidence

vacuum grating monochromator. This was equipped with an end-window

photomultiplier tube (=I 6256 B) whose silica window had been

sensitised using sodium salicylate. Kriastianpoller and Knapp (1600)

and Samson (17 )have described in great detail the use of this material

for sensitising photomultiplier tubes to radiation of low wavelengths.

The flat response to incident radiation between 80nm and 300nm

and relatively high quantum efficiency make sodium salicylate the

most common sensitising agent in use today. The fluorescence emission

peak of sodium salicylate is at 390nm and this corresponded with

the peak response of the 6256B photomultiplier tube, Fig.8.7a.

Using a compressed inert-gas spray-gun a saturated solution of

sodium salicylate was sprayed onto the face of the photomultiplier

window to provide a coating of between 1 and 2mg.cm 2; the optimum

layer density for this material. A hot-air blower was

employed to evaporate the solvent. The weight of the organic salt

deposited on the window was determined by calibrating the spray

gun emission. This was achieved by spraying several pre-weighed

209

detection system

uartz 2mm. o.d.

argon

Variac

coaxial connector

coupling adjustment

Atmospheric Microwave Plasma.

The At mospher

ic Microwave P

lasma.

(WU

) lN5U

819AID

M

Quantum Efficiency % O O

N.) 01 rn Co

O O

0 - 0

— 0

.uoTssTm7 aesmsues pi e GAano .KoueToTKra Lk/. L.B

rn (D-O

I o

uo!ss!w3 lueoseJonH

Fig. 8.8 Spectrum From Argon Microwave Plasma.

G

F

C C

A

I I I 1 1 1 I 130 140 150 160 170 180 190

Wavelength,(nm ). Line. WavelengthInm Species. L.ine Wavelength,nm Species

A 127.75 Carbon 148.18 Carbon

B 130.22 Oxygen F 156.10 Carbon

130.49 G 165.72 Carbon

130.60 H 175.19 Carbon

C 132.93 Carbon I 193.04 Carbon

D 146.33 Carbon

212

thin glass slides with the salicylate-methanol solution and re-

weighing the slides after evaporation of the solvent. The number

of passes of the spray...gun over the face of the slides necessary

to provide the required deposit of sodium salicylate was thus determined.

The signals from the photomultiplier tube

were led directly to a potentiometric chart recorder.

Spectral Emission Characteristics Of The Argon Microwave Plasma.

The atmospheric pressure, microwave excited

argon plasma apparatus was assembled as described above. With an

. - 1 argon flow-rate of 1.0 to 1.51itres.min and an applied microwave

power of ca. 100-watts the plasma was initiated using a Tesla discharge

coil. The flow-rate was lowered to 1.0 litre.min 1 argon and the

microwave power to 50 to 60 watts. With the plasma discharge immed-

iately up against the entrance window of the vacuum monochremator

and employing a spectral bandpass of 0.08nm the spectrum from the

plasma was recorded between 120nm and 200nm. The voltage to the

PMT was maintained at 1600V. The spectrum obtained and the atomic

lines observed, with their assigned transitions, is shown in Fig.8.8.

The emission lines observed were thought to be due to impurities

in the argon and pick-up from the walls of the silica tubing.

Sharp (154) has shown that in a low-powered

microwave argon plasma that although the electronic temperature

is in excess of 5000K a gas temperature of only ca. 1000K is obtained.

This difference may be attributed to the inefficient transfer of

kinetic energy from the electrons to gas atoms or molecules. Many

compounds do not atomise at such a low temperature and it is usually

necessary for any sample to be introdacedinto the plasma as a gas or

vapour, i.e. the plasma is employed solely as a means of excitation.

213

To investigate the vacuum ultraviolet emission characteristics

from the non-metallic elements the vapour introduction techniques

were used.

A one litre flask was positioned in the argon flow line

between the flow-meter and i.,-wave cavity. To investigate the low

wavelength spectrum of iodine a few milligrams of the solid were

placed in the flask and the plasma operated as discussed previously.

Chlorine atomic emission was examined by passing the argon supply

over a few millilitres of CC14 contained within the flask. Atomic

emission from sulphur was observed by bleeding a 1% SO2 in argon

mixture into the plasma-argon supply line. The emission spectra

from these iodine,sulphur and chlorine containing plasmas recorded

in the wavelength range 120nm to 200nm are shown in Figs.8.9, 8.10 and 8.11.

The spectra were not as free from impurities as

those observed by Braun et al. using the reduced pressure plasma.

and similar sample introduction techniques but the spectra obtained

here did serve to indicate the intense emission lines from the

non-metal atoms in the vacuum ultraviolet.

b) A Demountable Hollow Cathode Lamp Source.

The atmospheric microwave plasma was not the

only vacuum u.v. atomic emission source examined, a commercial

demountable hollow cathode lamp source was employed also. The

details of the construction and use of this lamp as a line emission

source have been discussed in Chapter Five. Its operation here

was identical, i.e. a 3min thick calcium fluoride window was used

and aluminium cathodes because of their spectral purity employed.

To examine the spectral atomic emission characteristics

of chlorine and bromine a variety of salts containing these elements

were investigated. The compounds producing the most intense and

214

LO O N

LO

7-4 CO CO Lf) LO Lc)

1-1 C.)

I I 1 1 1 I I 130 140 150 160 170 180 190 200

Wavelength, nm.

Fig. 8.9 Argon licrowave Plasma Containing Iodine Vapour.

If)

U

215

CO

0

(5)

0

N co

N co

O CO •c-

LD

(i)

130 140 150 160 170 180 190

Wavelength, nm.

Fig. 8.10 Argon Microwave Plasma With Sulphur Dioxide.

216

87110 971.10

ZE1.1 0 8Z1. 1 3

C61.0 - 0)

--- CO

591.,3

95113

— N. s--

Wa v

elen

gth,

nm

.

0 - (0

•-•

—In

o —

L'6C1.10'6C1.:0"9EVE7.61 41

} 91.Z'SCVL'ICV9'661. I

Fig. 8.11 Argon Microwave Plasma With CG14 Vapour.

Fig. 8.12 (a) Hollow Cathode Lamp With Mercuric Chloride Cathode.

0 ••••••• Q.

•■••■ft,

Wave

leng

th, n

m.

217

0 I-

0 - CO

0

0 I-

Fig. 8.12 (b) Hollow Cathode Lamp With Lead Bromide Cathode.

Fig. 8.12 (c) Hollow Cathode Lamp With Lead Chloride Cathode.

Cr \

Wave

leng

th, n

u.

218

cleanest soectra in the vacuum ultraviolet were mercuric chloride

and lead bromide, Figs. 8.12 a and b. As before, mixtures of the

sample and tungsten powder (1 to 1 w/w) were compressed into the

blank aluminium cathodes and drilled out to provide a deposit of

the mixture, ca. 1mm thickness, on the inside of the hollow cathode.

As a comparison, Fig. 8.12 c shows the spectrum, in the same wavelength •

range, of a lead chloride/tungsten/aluminium cathode lamp. The

spectral bandpass of the monochromator and voltage to the PNT was

the same in each case and the.same as employed during the atmospheric

microwave plasma studies. The hollow cathode lamp source was mounted

up against the entrance window of the monochromator and the 30mm

path between the source and window purged free of air using argon

passing through a 30mm, 20mm diameter glass tube forming the optical

path.

In general the spectra obtained using the

demountable hollow cathode lamp were of greater spectral purity

than those obtained from the microwave plasma, however, the atomic

line emission intensities were typically an order of magnitude

inferior.

8.5 Selecting The Filling Gas For A PID.

The number of crystalline window materials

suitable for use as cut-off filters in the vacuum ultraviolet is

limited. The PID described in Section 8.3 was fitted with a 3mm

thickness calcium fluoride window providing for a short wavelength

cut-off of ca. 125nm. To limit the spectral bandpass of the detector,

and because of the intensity of the oxygen atomic triplet emission

from the microwave plasma at 130nm and the profusion of atomic

chlorine emission lines in the wavelength range 130nm to 140nm from

219

both sources examined, a filling gas or vapour was sought having

an adiabatic ionisation potential of about 140nm; approximately 8.8eV.

A study of the extensive literature available

relating to the ionisation potentials of organic and inorganic.

vapours showed that a variety of compounds could be employed for

the PID used here. Table 8.1 provides some typical compounds whose

adiabatic ionisation potentials occur in the region 8.50eV to 9.00eV

(ca. 145nm to 138nm). It is obvious from Table 8.1 and the vast

amount of material accumulated in the literature that a direct

relationship exists between the value of the adiabatic ionisation

potential and the structures of molecules; simple quantitative rel-

ationships have been established in some cases.

The vapour of ethylamine, which has an ionisation

potential of 8.8eV was employed as the PID filler material. The

availability of ethylamine and its low boiling point (289.7K) made

this an ideal vapour for filling the chamber and examining the PID

as a narrow bandpass radiation detector. The photoionisation efficiency

curve for ethylamine and the transmission characteristics of the

calcium fluoride window employed are shown in Fig. 8.13 (155115C* ).

From the curves shown in Fig.8.13 the spectral bandpass of the PID

for the use of ethylamine with this window material was predicted

to be between 125nm and 140nm. It is interesting to note that the

photoionisation efficiency curve for ethylamine varies between the

adiabatic ionisation (appearance) potential and the vertical ionization

potential in a manner similar to that described in Section 8.2(c)

for aniline, i.e. typically a transition between a molecule and

ion of greatly differing internuclear spacing. Hurzeler et al.

have suggested that the electron removed by ionisation is not a

a non-bonding one as expected, but possibly a bonding electron from

the C bond.

220

Table 8.1 Lome ::.aterials :aving ionisation Poteintials In

The Re7ion 8.55e7 to 8.8CeV

Vapour Ionisation Pot. (eV) Vapour Ionisation Pot.

o-xylene 8.56 iso-propylamine 8.72

m-xylene 8.59 iso-butylamine 8.70

ruethylarnine 8.97 t-butylamine 8.64

ethylamine 8.80 toluene 8.82

n-propylamine 8.78 ethyl benzene 8.76

n-butylamine 8.71 n-propyl benzene

n-butyl benzene

8.72

8.69

Table 8.2 Some Materials Having Ionisation Potentials In

The Region 6.90eV to 7.0CeV

Vapour

Ionisation Pot. (eV) Vapour Ionisation Pot.

(eV)

N1N-di-n-butylaniline 6.95 N,N,-di-n-propylaniline 6.96

N,N-diethyl-p-toluidine 6.93 N,N,-diethyl-p-bromoaniline 6.96

p-bis(dimethylaminobenzene)6.90 N,N,-diethylaniline 6.99

221

pump

detector

cold trap

0-20 torr pressure guage

argon

Et NH2

ballast Dewar

Fig. 8.14 The PID Filling System.

Fig. 8.13 Predicted Response Of The PID.

>1 100- 0 C a)

L.6 C

U) "E .05

40—

0 _c 0- 20—

1mmCaF2...

2mm raF 2

—100

—80

—60

—40

—20

C2H5NH2

UO

! SSIW

SU

DJ±

%

120 130 140 150

Wavelength,nm.

222

8.6 The Photoionisation Detector.

A simple vacuum and purging system was constructed

to permit the filling of the PID with ethylamine (or other gases

or vapours). This is shown in Fig.8.14. Anhydrous ethylamine

(10 mis.) was placed in the pear-shaped flask and frozen by immersion

of the flask in liquid nitrogen. The complete system was then

evacuated using a two-stage rotary pump and flushed out several

times with argon. On a further evacuation of the complete system

tap A was closed to isolate the flask containing the ethylamine.

The liquid nitrogen was then removed from beneath the flask and

the ethylamine allowed to warm to produce a steady increase in ethylamine

vapour pressure. By first closing tap B and then opening tap A

the system was flushed out several times with ethylamine vapour,

followed each time by re-evacuation to a pressure of ca. 3 torr,

Tap B was then finally closed and a preselected pressure of ethylamine

vapour admitted to the system via tap A. With taps A and B both

closed the detector was sealed by closing tap C and removed from

the system for use. oven with the simple PID system constructed

for this study the PID envelope would maintain a fill pressure as

low as 1 torr for periods in the excess of eight hours and was

simply and rapidly refilled when necessary.

8.7 The Spectral Response Characteristics Of The ahvlamine P=D.

The spectral response characteristics of the

ethylamine PID were studied using both available sources, i.e. a

microwave excited argon plasma and the demountable hollow cathode

lamp system.

The PID was filled with ethylamine vapour at 2 torr

and positioned at the exit slit of the vacuum monochromator; the

223 •

PID was mounted such that the window of the ionisation chamber

formed the vacuum.seal with the monochromator. The argon microwave

plasma, as described previously, was viewed by the monochromator

using a spectral bandpass of 1.6nm. The emission spectrum of the

argon plasma was recorded using the PIL at 85V applied potential;

the signal from the PID was recorded on a potentiometric chart recorder.

The spectrum recorded is shown in Fig. 8.15. As a comparison Fig.8.15

also shows the spectrum of the argon plasma obtained using the vacuum

monochromator at the same spectral bandpass (1.6nm) with the photo-

multiplier detector, at 700V. It is clear from a comparison of these

spectra that the PID responds only to radiation in the wavelength

range predicted. Whereas the carbon and oxygen line emissions in

the region 128nm to 132nm are observed, the more intense longer

wavelength carbon lines between 140nm and 193nm are not detected.

As a further test of the response characteristics

of the ethylamine PID the demountable hollow cathode lamp source

with argon filler gas and an aluminium cathode containing a mixture

of tungsten and mercuric chloride was employed. The spectrum from

this source in the range 130nm to 200nm was obtained using the

monochromator-photomultiplier detection system and is shown in

Fig. 8.12 a. The spectrum consisted of atomic line emission from

chlorine, oxygen, carbon and mercury and was obtained using an argon

purged optical path (30mm) between the source and monochromator

entrance window. The experiment was then repeated with this 30mm

path-length purged with a 1% methane in argon gas mixture. As

methane absorbs strongly at wavelengths less than 140nm (1567) the

emission intensity from the source at low wavelengths is attenuated.

The spectrum obtained under these conditions is shown in Fig. 8.16

224

Fig. 8.15

Emission Spectra From Microwave Plasma

(a) recorded with ethylamine PID

(b) recorded with 6256B PMT.

120 130 140 150 160170 180

Ol o (b) co 0

LI)

225

If) N-

ro

Fig. 8.16

(1) DHCL + HgC12 Spectrum, Argon Purged Path

Inte

nsity

, ar b

. un i

ts

LO DHCL + HgC12 Spectrum, Argon + 1% CH4

Purged Path.

I I II I I I I 130 140 150 160 170 180 190 200

Wavelength, nm.

226

10-

109_

U)

E ci -10

4E: 10 — a) J

C.)

Fig. 8.17 The Effect Of 1;"; Methane In The Optical Path

Between The DHCL + HgC12 and The PID.

10

3cm. path of argon.

3cm. path of argon +1% methane.

10 1

Background, no radiation.

10-13

227

and clearly- demonstrates the absorption of the oxygen and chlorine

atomic line emission between 130nm and 140nm'and the unaffected

intensities at longer wavelengths. These experiments were repeated

using the PID in the non-dispersive mode, i.e. with the radiation

falling directly on to the detector via the 30mm purged path without

using the vacuum monochromator. The response obtained in the presence

and absence of 1%. methane in the argon purge gas is shown in Fig.8.17.

The decrease in PID response observed in the presence of methane

correlates well with the decrease in intensity of the emission recorded

between 130nm and 140nm shown in Fig.8.16 in the presence of methane

using the PIE detector. It was thus apparent that it was the incident

radiation below 140nm to which the PID responded.

8.8 The Effect Of Ethylamine Vapour Pressure And Applied Voltage

On The Response Of The PID.

The radiation from the demountable hollow

cathode source with an aluminium cathode containing a mixture of

tungsten powder and solid mercuric chloride was allowed to fall

directly onto the PID in the non-dispersive mode of operation. The

effect of the fill pressure of ethylamine vapour and the applied

d.c. potential on the response of the detector to incident radiation

in the range 130nm to 140nm was examined under these conditions.

The picoammeter response was determined at applied voltages between

5 volts and 400 volts for a number of ethylamine filler vapour

pressures. The current-voltage curves obtained are shown in Fig.8.18.

The sharp current rise regions at low applied voltages, the plateau

regions and the regions of gas-gain, ion-multiplication, as described

in Section 8.2 (b) are clear. It can be seen that at the higher

voltages, in the gas-gain region, an increase in' he sensitivity

of the detector by two to three orders of magnitude is possible.

228

2 torr Et NH2

5torr EtNH

Cur

ren t

(am

p s)

DHCL 5mA.

100 200 300 Voltage, (volts).

. Fig. 8.19

10torrEtNH2

20torrEtNH

5 torrEtNH2

760 tore He

10 —

Current,(amps)

10-8—

I I 0 100 200 300

Voltage (volts)

Fig. 8.18

400

The effect shown in Fig. 8.18,where the ion current is inversely

proportional to the pressure of the ethalamine present over the

range studied, can be explained on the basis of the mean free-path

of the electrons formed on ionisation. Thus, as the ethylamine

pressure was increased the mean free-path of the photoelectrons

decreased so that there was insufficient time to permit their

acceleration to energies sufficient to cause secondary ionisation

on collision with an ethylamine molecule. This effectively delayed

the onset of the ion-multiplication region to higher applied voltages.

The limiting case shown in Fig. 8.18 corresponds to the addition

of helium as an inert filler gas at atmospheric pressure in the

presence of 5 torr of ethylamine. Under these conditions the plateau

region extended to greater than 300V without observation of secondary

ionisation.

Fig. 8.19 demonstrates the dependance of the PID response

on the incident radiation intensity from the source employed. An

increase in operating current from 5mA to 15mA d.c. resulted in

an increase in current of ca. one order of magnitude due to the

equivalent increase in intensity from the oxygen and chlorine atomic

line emission between 125nm and 140nm. The proportionality between

incident intensity and PID response is observed in both the plateau

and ion-multiplication regions.

8.9 Quantitative Analysis With The ahvlamins, PTD.

It has been shown how a narrow spectral

bandpass detector may be constructed and operated to monitor radiation

in the wavelength region 125nm to 140nm using ethylamine vapour'

as the ionisation medium and a calcium fluoride window as the short

wavelength cut-off filter. Attempts to employ the ethylamine PID

230

as a quantitative ultraviolet detector for atomic spectrometry met

with little success but the experiments will be reviewed here to

indicate the problems accompanying the use of a PID with the atomic

emission systems examined.

The atmospheric pressure argon microwave plasma atomic

emission source was observed to become very unstable when samples

such as CC141 SO

2' 02, N

2' etc. were introduced into the argon supply.

This perturbation of the plasma may be attributed to the low gas

temperature attained within the plasma and the high collisional

cross-sections of these molecules makes them very efficient in quenching

the discharge. To overcome this problem of instability a variety

of techniques have been proposed and employed by many workers to

introduce samples into the plasma without disturbing the discharge.

Johnson 057) used a radiofrequency, inductively heated nickel

boat assembly to accomodate up to 100 microlitres of sample solution

or ca. 50mg of powdered solid sample. The sample was dried in the

boat, contained within the argon stream and then the sample volatilised

into the argon flow and carried into the plasma. However, many

disadvantages and interference effects were observed with this

technique. Dagnall et al. have examined a platinum loop filament

to introduce the sample into the plasma. The sample, typically

1/41, is dried and then volatilised from an argon sheathed, resistively

heated platinum filament and fed into the discharge. A platinum

loop assembly in conjunction with the argon microwave plasma was

examined as an atomic emission source for the determination of

chlorine using the ethylamine PIT). ":7:xperiments conducted with

such an assembly, however, gave no clear indications as to whether

chlorine emission from sodium chloride and ammonium chloride

231

volatilised into the plasma was detected by the PID. Unfortunately,

the intense oxygen emission at 130nm from the plasma served to

provide a large background signal (typically 10 7 amperes) at the

detector and no further signal was observed above this on volatilising

the chlorine containing sample from the platinum filament.

To further limit the spectral bandpass

of the ethylamine PIE, a barium fluoride window (135nm cut-off,

2mm thickness, 25mm diameter) was positioned between the plasma and PID.

The currents obtained from the PID were now limited to ca. 5 x 10-12

amperes and reached only ca. 10 10 amperes when GC14

vapour was

continuously flowing into the argon stream into the plasma. The

low response by the detector to the atomic chlorine emission from

the plasma was attributed to the poor transmission characteristics

of the double (GaF2-BaF2) window arrangement employed. No further

experiments were undertaken with the atmospheric argon plasma and

ethylamine PID system.

The radio-frequency, induction-coupled, atmospheric

argon plasma is a far more efficient atomic emission source than

the microwave plasma. Gas temperatures in excess of 8000K are

attainable and the discharge is not so susceptible to perturbation

by sample introduction. However, air-borne radio-frequency interference

with such devices is a problem. The PID containing 2 torr ethylamine

was assembled alongside a radio-frequency plasma, connected by an

argon purged optical path. 0.1M HG1 solution was nebulised into

the plasma but no atomic emission signals were detectable by the

PID owing to the very high noise level, typically 105 to 10 7 amperes,

generated by the plasma system and received by the PID. Screening

the cables from the detector to the readout device from such large

spurious signals was not accomplished; a complete redesign of the

232

whole apparatus to match the two instruments would have been necessary.

8.10 The Xylene PID.

Xylene (a mixture of ortho, meta and para)

was examined also as a PID filling vapour for chlorine atomic emission

studies. As shown in Table 8.1 the use of this vapour would provide

for a detector responding to radiation in the wavelength range

135nm to 147nm when used with a BaF2 window.

The PID containing 2 torr of xylene was employed

with the atmospheric microwave plasma and platinum loop assembly.

As before, to reduce the incident oxygen emission intensity at

130nm from the plasma a barium fluoride window was positioned between

the detector and the plasma. The ion-current readings produced

by the plasma itself were again very small (ca. 10 11 amperes),

however) small increases in current were observed when mixtures

of argon plus carbon tetrachloride were passed through the plasma.

Similarly, small increases in ion-current were observed when small

quantities (10-6g) of ammonium chloride were vaporised into the

plasma using the platinum loop filament. Unfortunately, the magnitude

of the signals did not allow for any quantitative assessment of the

chlorine emission intensity.

8.11 The N,N-Diethyl7p-toluidine PID.

The PID chambers described previously were

constructed to detect radiation in the wavelength range 125nm to

145nm. Because of the many problems with such low wavelength studies

a longer wavelength PID was examined. The literature was consulted

for a substance with an ionisation potential in the region of 6.9eV

(180nm), having a reasonable vapour pressure (a few torr) at room

temperature. It was hoped to examine with the new device the emission

233

lines from excited iodine, sulphur and phosphorus atoms in the

wavelength region 175nm to 181nm. Possible compounds having an

ionisation potential in the 6.9e7 region are shown in Table 8.2

(158 ,05q). II,N,-diethyl-p-toluidine was available and was used

as the PID filling gas. The usual chamber, with calcium fluoride

window, was employed and the lower wavelength limit was set by the

transmission characteristics of the quartz-bulb EDL sources examined

with this detector. The PID was filled (to ca. 2 torr) with the

organic vapour and operated in a similar manner as for ethylamine

and xylene.

An iodine EDL source was supported in a iwave

resonant cavity and operated at ca. 50 watts applied microwave

power. A current of 3 x 10-10 amperes was obtained with an argon

purged path between the detector and 0ZOL and a PID applied voltage

of 85V. As a similar current was using a high pressure mercury

vapour discharge lamp as the source it was evident that the 11,N,-

diethyl-p-toluidine was capable of detecting radiation of wavelengths

as high as the 184.9nm mercury resonance line; 6.7eV. With no

source emission falling on the detector or a stop placed between

the source and PID, currents of ca. 1012 amperes were obtained.

Cold-Vapour nercury AAS.

A mercury iJ)L was operated at 12W applied

microwave power in a i-wave resonant cavity. The optical path

between the source and detector was provided by a 40mm long, open-

ended cylindrical glass cell (20mm i.d.). With oxygen-free nitrogen

purging of the EDL cavity and optical path the measured current

- was 1.1 x 10

10 amperes. When nitrogen, at the same flow-rate,

was passed over mercury before being introduced into the glass the

234

- - current dropped to 0.7 x 10 10 amperes, but rose again to 1.1 x 10 10

amperes on switching the pure nitrogen directly into the glass cell.

The decrease in current was shown to be due to mercury atomic

absorption of the 184.9nm mercury resonance line by repeating the

experiment using a deuterium arc continuum source; no decrease in

current was observed when the nitrogen was passed over the mercury

pool prior to the glass cell. The small currents observed with

this PID did not allow its spectral bandpass to be determined using

the vacuum monochromator.

8.12 Conclusions.

The preliminary work described here indicated that

simple, high efficiency, narrow spectral bandpass, solar-blind

photoionisation detectors may offer advantages for detection of

radiation at short wavelengths in analytical atomic emission spect-

rometry. They should prove most valuable when used for non-dispersive,

on-line detection systems. The most useful wavelength range for

such detectors lies between 120nm and 200nm and detectors operating

in this region using different window and filler materials have

been described in this chapter.

The two major advantages of the PID over the

alternative and more common vacuum monochromator - photomultiplier-

tube system are the great savings in both cost and space. Correct

and careful assembly of the PID is essential to achieve optimum

results and construction features should include a guard ring to

prevent leakage currents, a thin window to provide high transmittance

over the required spectral bandpass of the PID, and adequate shielding

to prevent air-borne interference causing spurious noise signals.

Current vs applied voltage curves were obtained

235

for the ethylamine PID at various pressures of ethylamine and it

was observed that gas-gains of the order of 103 could be obtained.

For studies of weak ultraviolet sources it would be desirable to

operate the detectors in the gas-gain mode.

By careful selection of window material and filler

gas a large selection of banciasses in the region 105nm to 200nm

are available. As many of the compounds with ionisation potentials

above 150nm are solids at normal temperature a heated chamber is

necessary to achieve adequate vapour pressure of the material.

These experiments used primarily emission techniques

to evaluate the PID. The use of the argon atmospheric microwave

plasma as a vacuum ultraviolet emission source with the PID had

several disadvantages. The high background emission from the plasma

and the difficulties experienced in introducing samples into the

plasma are practical problems not easily solved and contrast with

the intense u.v. atomic emission obtained with this source. The

demountable hollow cathode lamp appears to have a greater potential

as a suitable line emission source for use in the region 105nm to

200nm. However,a more detailed study is required into the possible

use of this source as a quantitative emission system for elemental

analysis.

236

NINE

237

9.1 Conclusion.

This thesis has described the direct. determination

by AAS of iodine and sulphur in aqueous solutions using non-flame

electrothermal graphite atomisers. The vacuum ultraviolet resonance

transitions of these elements have been employed for these studies

and the sensitivities achieved are comparable to those for trace

metal analysis by non-flame AAS.

The major difficulties encountered with atomic

spectrometry in the vacuum ultraviolet have been reviewed and for

the AlS of iodine and sulphur were observed to be:

a) achieving a sufficiently intense line emission source to permit

an acceptable signal-to-noise ratio to be obtained.

b) achieving a sufficiently transparent optical path to allow

the transmission of the source radiation.

c) the construction of a suitable non-flame cell to provide the

atomic species of interest without the associated problem of

interference from background absorption by the hot cell.

Two types of line emission source have been

examined. For the AAS of iodine and sulphur a variety of EDL

sources were considered and the best in terms of signal-to-noise

ratio are those containing iodine as mercuric iodide and sulphur

as hydrogen sulphide. Both lamps are capable of responding to

electronic modulation of the applied microwave power and a comp-

arison between mechanical (chopper) modulation and electronic

modulation has shown the latter to be superior in terms of signal-

to-noise ratio of the emitted resonance radiation. The major

disadvantage of this technique is the need for a specially prepared

(mercuric iodide) lamp in the case of iodine which would follow

the modulating signal. For the iodine studies a water-cooled

238

demountable, hollow cathode lamp has been employed also as a line

emission source. The sensitivities (for 15 absorption) and detection

limits achieved at the two resonance lines examined (183.0nm and

178.2nm) using this source are comparable to those obtained using

the ML source. Fitted with a calcium fluoride window or lithium

fluoride window this HOL may have great potential as a line emission

source for the AAS of phosphorus, bromine and chlorine whose

resonance lines are difficult to obtain from EDL sources.

To achieve a relatively transparent optical

path for the vacuum ultraviolet two techniques have been employed.

The one-metre, grating monochromator was a vacuum system and,

as previously stated, with the rotary and diffusion pumps is capable

of achieving and maintaining pressures of less than 10-6 tors..

The use of an evacuated optical path between the source and entrance

slit of the monochromator was not undertaken. The problems associated

with the construction and operation of such a system would have

far outweighed the few advantages with the apparatus. Instead,

for the optical path and atom cell an inert-gas purging system

has been employed. By passing nitrogen or argon at a relatively

low flow-rate continuously through the glass and aluminium tubing

forming the optical path the majority of atmospheric oxygen and

water vapour is successfully eliminated from the system. Although

this technique requires a constant supply of inert gas for purging

that no vacuum seal must be broken to introduce samples into the

atom cell provided the technique with a simplicity of operation

not achieved with vacuum techniques.

The major problems encountered with the direct

determination of iodine and sulphur by AAS are related to the

239

construction and operation of the atom cell. It has been shown

that the graphite tube furnace employed by Johnson et al. (95 )

for the direct determination of metals at wavelengths in the

middle and near ultraviolet regions is not suitable for routine

vacuum ultraviolet AAS. The background absorption obtained with

this atomiser was severe at 183.Onm and at the sulphur resonance

line (180.7nm) prevented a serious study of this elements AAS

characteristics. It has been demonstrated that the background

interference from the graphite tube atomiser is a function of the

graphite tube - graphite end-piece contact and for the elimination

of this interference no part of this contact can remain in the

optical path. To reduce the dead-time between samples and reduce

the time necessary for drying and purging of the atom chamber a

relatively open atom cell is necessary. The problems associated

with the construction of a tube furnace combining these points

have been discussed and the construction and operation of a furnace

fulfilling these rquirements is described in Chapter Five.

The graphite rod atomiser has been examined

also as an atom cell for sulphur and iodine AAS. The characteristics

of this furnace are midway between those of the small graphite

tube and the graphite filament atomisers. The major advantage

of this cell for vacuum ultraviolet AAS is the ease of purging

the cell of oxygen and sample solvent; this reduces the time per

analysis. Unfortunately, whilst satisfactory results are obtained

with this device for iodine AAS at 183.Onm no atomic sulphur

could be detected at 180.7nm, 182.0nm and 182.6nm. The failure

to perform sulphur AAS with the graphite rod atomiser may be attributed

to the relatively short residence-time of the analyte species

in the hot furnace. Indeed, the volatile nature of the simple

240

sulphur thermal decomposition products (e.g. SO2) appears to be

responsible for the deviations in linearity of the sulphur analytical

curves obtained with the graphite tube furnace. As discussed in

Chapter Five the successful determination of sulphur by AAS appears

to require a long residence-time in the hot environment of the

atomiser to produce a reasonably linear working range for analytical

purposes.

Whilst the sensitivity attained in AAS using

electrothermal atomisers is frequently far superior to that with

flames the non-flame cells, in general, exhibit greater interference from

sample matrix effects e.g. non-specific absorption by thermally

stable salts. Because of the common occurrence of these salts

in the samples frequently encountered with non-flame AAS the lack

of data concerning the nature of this non-specific absorption

is surprising. Chapter Six of this thesis describes a study of

the non-specific molecular absorption observed with some simple

alkali metal salts vaporised in the small graphite tube atomiser

and the commercial HGA 2000 tube furnace. Although the discussion

relates more to the qualitative spectral characteristics of the

molecular absorption, rather than the magnitude and relative

severity of each potential interferant, it is of interest that

the spectra obtained are very similar to those observed using

cool flame cells. The wavelengthsof maximum absorption of incident

radiation corresponded, within experimental error, to the wavelengths

calculated by Herzburg for the electronic transitions of the

gaseous ionic molecules. Thus, the absorption by the salts examined

exhibit a wavelength dependence. To correct for the molecular

absorption by the common inorganic salts studied as interferants

241

in the ASS studies of iodine and sulphur a nearby non-resonance

line has been employed. The lines employed are within 2nm of

the wavelength of the resonance transitions and no changes in the

molecular absorption spectra were observed over the wavelength

region of interest, i.e. in this case the technique of employing

a non-resonance line for background correction is justified. At

the concentration of the salts examined in the molecular absorption

studies no absorption by oxy-anion salts (e.g. nitrates, sulphates,

phosphates, etc.) is observed in the HGA 2000 graphite furnace.

At low wavelengths (in the vacuum ultraviolet)

the complexity and cost of the instruments necessary for quantitative

atomic spectrometry frequently prohibits their use as common,

routine analytical techniques. Partly to simplify the instrumental

arrangements employed in atomic spectrometry the study of non-

dispersive systems has stimulated much research; especially with

atomic fluorescence spectrometry in the middle ultraviolet. The

heart of any non-dispersive apparatus is the detector, which must

have a limited spectral-response bandpass. In the vacuum ultra-

violet such a detector is the photoionisation detector and Chapter

Eight has described the construction, operation and evaluation

of such a device. With intense emission sources (1013 to 1015

quanta.sec-1) currents as high as 10 4 amperes may be generated.

Although this may be considered an upper limit, using a microwave-

excited argon plasma at atmospheric pressure, the maximum current

observed was ca. 10 6 amperes, and more typically 108 amperes.

The spectral selectivity of PID chambers is shown to be dependent

on the nature of window material forming the entrance to the

chamber and the ionisation characteristics of the filling gas

242

or vapour. The use of these detectors for quantitative analysis

was limited in the studies described in this thesis by the emission

systems employed. No emission system capable of producing an

intense, quantitative emission signal from an analyte was available.

9.2 SuFestions For Further Study-.

Following the studies described in this

thesis relating to the non-flame AAS of iodine and sulphur, the

application of the furnace and the techniques described here to

the analysis of 'real' samples would be of interest. This work

could follow two paths. The graphite tube furnace and vacuum

spectrometer assembly of the type discussed in this thesis could

be employed directly or a commercial system, such as the HGA 2000

graphite atomiser and 305B spectrometer, could be modified to enable

the vacuum ultraviolet studies to be conducted. The basic modif-

ications to the Perkin-lmer 305B apparatus for low wavelength

studies have been described by Wilson (75) and were merely the

sealing of the monochromator housing to facilitate inert-gas

purging of the monochromator and extension arms fitted to the ends

of the furnace to prevent the entry of atmospheric oxygen into

the system. To reduce the background absorption observed with

this furnace at wavelengths less than 200nm the graphite end-cones

employed with the furnace could be modified to remove the contact

area from the optical path. An end-window photomultiplier tube

of high gain (e.g. al 6256B) would provide for increased signal-

to-noise ratio of the source emission. Again only minor modifi-

cations to most commercial instruments would be necessary to achieve

this. The background-correction facility of many commercial

instruments is undertaken using a deuterium-arc continuum source

fitted usually with a Vitreosil window. The low wavelength cut-

243

off limit of this material (180nm) and the solarisation of the

window by the ultraviolet radiation with increasing use provides

for only low intensity radiation of wavelengths shorter than ca.

190nm. A continuum source with a thin quartz window or (better)

a sapphire window would not suffer from these disadvantages. The

use of such a background corrector for eliminating or reducing

non-specific absorption is very important at low wavelengths and

would greatly enhance the practical value of non-metal, non-flame

AAS. The lamp housing of most commercial atomic absorption spect-

rometers will allow the use of EDL sources or a demoumtable, hollow

cathode lamp such as described in Chapter Five with little or no

modification.

Wilson (28) has described the use of the Leipert

amplification procedure for increasing the sensitivity of determining

iodine by /US. Iodide in aqueous solutions is oxidised to iodate

by bromine water, the solution is acidified and excess iodide

added. The iodate reacts with the added iodide to liberate six

equivalents of iodide for each equivalent of iodate (or iodide)

in the original sample. The liberated iodine may be separated

from the excess iodide by extraction into an organic solvent (e.g.

MIBIC) and the organic solution nebulised directly into a nitrogen-

separated nitrous oxide-acetylene flame. Unfortunately, the highly

volatile iodine and organic solvent would both be removed from

the furnace during the drying stage if this procedure were applied

to non-flame 112,S. The use of organic solvents with non-flame

techniques at low wavelengths requires a special study. Many

organic solvents have greater absorption coefficients in the vacuum

ultraviolet than water and trace quantities could cause complete

244

abporption of the incident radiation, presenting many problems

should condensation occur along the optical path, as discussed

in Chapter Five for the case of water vapour. Thus, for example,

the determination of sulphur in oils would be best undertaken

probably by oxidation of the sample prior to non-flame AAS.

The direct determination of phosphorus

by non-flame AAS at its vacuum ultraviolet resonance lines has

been demonstrated by L'vov and should be possible using the graphite

tube furnace described in Chapter Five. Studies of phosphorus

EDIL sources (160 have indicated that elevated temperatures are

necessary for the required intense line emission to be achieved

i.e. the use of a thermostated resonant cavity. A preliminary

examination of the demountable hollow cathode lamp containing an

aluminium cathode with a lining of red phosphorus showed the resonance

emission to be reasonably intense but very unstable during operation

for several hours.

At wavelengths shorter than 200nm a serious

problem with non-flame AAS is the intense absorption produced by

the vaporisation of modest amounts of common inorganic salts.

Following the absorption spectra presented in this work a further,

more detailed study as to the nature and magnitude of more inter-

fering salts would be of interest.

The past decade has seen a dramatic growth in

the technique of atomic absorption spectrometry and the field of

atomic emission spectrometry has remained relatively quiet. However,

the use of the radio-frequency, induction-coupled argon plasma

as a high temperature atom cell for atomic emission appears to

promise a renaissance for the atomic emission techniques. An

245

interesting field of study would be the use of such a plasma with

PID chambers for selective, non-dispersive, multi-element atomic

emission spectrometry. a)rokin et al. (162) have examined the

use of organometallic compounds as filler vapours in photoionisation

chambers. Although elevated temperatures (typically 75° to 150°C)

are necessary to achieve a sufficient vapour pressure of the material

the relatively low ionisation potentials of these compounds (typically

160nm to 200nm) enables the construction of PID chambers for

monitoring phosphorus, iodine, sulphur and mercury radiation.

These detectors would be useful for special applications such as

GLC detectors etc .

0000000000000

246

References.

1. Fraunhofer, J., Gilberts Ann. 56, 264 (1817)

2. Kirchoff, G. and Bunsen, R., Fogg. Ann., 110, 161 (1860)

3. "Practical Spectroscopy", Harrison,G.R., Prentice Hall (1942)

4. "Practical Spectroscopy", Cutting, T.A., W.Heinemann Ltd. (1952)

5. Stokes, G.G., Collected Papers, Univ. Press Camb. /1 267 (1880)

6. Schumann, V., Ann. de Phys. 5, 349 (1901)

7. Lyman, T., Mem. Am. Acado Arts Sci., 13, 3 (1906)

8. Millikan, R.A. and Bowen, I.S., Phys. Rev., 23, 1 (1924)

9. Osgood, T.H., Phys. Rev., 22, 567 (1927)

10. Boyce, J.C., Revs. Mod. Phys., 13, 1 (1941)

11. Lyman, T., "The Spectroscopy Of The Extreme Ultraviolet."

Longmans, Green, New York, 2nd. Ed. (1928)

12. Lyman, T., Astrophys J., 60, 1 (1924)

13. Garton, U.R.S., J. Sci. 36, 11 (1959)

14. Cann, M.W.P., Appl. Opt., 8, 1645 (1969)

15. Samson, J.A.R., Planetry Aeronomy V , N.A.S.A. CR-17 (1.963)

16. Golman, K., Physica, 29, 1024 (1963)

17. Samson, J.A.R., "Techniques Of Vacuum U.V. Spectroscopy"

J.Wiley and Sons Inc., (1967)

18. Dickerman, P.J. and Deul, R.W., J..iant. Rad. Transfer,

Al 807 (1964)

19. Wilkinson, P.G. and Byam, E.T., Appl. Opt., 581 (1965)

20. Hunter, W.R., Proc. 10th Colloq. Spec. Int., 247 (1962)

21. Schuler, H., Physik Zeitschr., 22, 264 (1921)

22. Sullivan, J.V. and Walsh, A., Spec. Acta, 21, 721 (1965)

23. Milazzo, G., Appl. Opt., 21, 135 (1967)

24. Newburgh, R.G., Appl. Opt., 2, 864 (1963)

25. Newburgh, R.G., Heroux, L. and Hinteregger, H.E.1 Appl. Opt., 1,.733 (1962)

247

26. Mazzo, G. and Spranzi, N., Appl. Spec., 21, 172 (1967)

27. Milazzo, G. and Sopranzi, N., Appl. Spec., 21, 256 (1967)

28. Wilson, P.J., Ph.D. Thesis Univ. of London (1974)

29. Hittorf, W., Wiedermann's Ann., XXI, 138 (1884)

30. Mansfield., J.M. and Winefordner, J.D., Anal. Chem., 37, 1049 (1965)

31. Dagnall, R.M., Thompson, K.C. and West, T.S.) Talanta, 11A, 551 (1967)

32. Browner, R.F. and Winefordner, J.D., Spec. Acta, 28B, 263 (1973)

33. Davis, D. and Braun, W., Appl. Opt., 7, 2071 (1968)

34. Warneck, P., Appl. Opt., 1, 721, (1962)

35. Samson, J.A.R., Vacuum U.V. Light Sources, N.A.S.A.-395 (1963)

36. Mann, C.K., Vickers,T.J. and Gulick, w.n., Inst. Analysis,

Harper and Row, (1974)

37. Tousey, R., Appl. Opt., 1, 679 (1962)

38. McNasby, J.R. and Okabe, H., Adv. in Photochem., 3, 157 (1964)

39. Carver, J.H. and Mitchell, P., J.Sci. Inst., LI., 555 (1964)

40. Garton, W.R.S., Adv. in At. and Mol. Phys., 2, 93 (1966)

41. Kaye, W.I., Appl. Spec., 15, 130 (1961)

42. Thompson, B.A., Harteck, P. and Reeves, R.R., J.Geophys. Res.68,6431 (1963)

43. Sponer, H. and Teller, E., Rev. of Mod. Phys., 13, 75 (1941)

44. Rowland, H.A., Phil. Mag., 16, 197 (1883)

45. Ducleaux, J. and Jealet, P., J. Phys. Radium, 2, 156 (1921)

46. Koller, L.R., "Ultraviolet Radiation" Chapman and Hall Ltd.,

London, (1952)

47. Golay„ M.J.E., Rev. Sci. Inst., 18, 347 (1947)

48. Sharpe, J., E.M.I. Doc., R/P006 (1961)

49. Dunkelman, L. Fowler, W.B. and Hennes, J., Appl. Opt., 1, 695 (1961)

50. Kaye, W.I., Appl. Spec., 15, 89 (1961)

51. Kaye, W.I., Appl. Spec., 15, 130 (1961)

52. Gunther,F.A., Barkley, J.H., Kolbezen, R.0 and Stagg, R.A.

Anal. Chem., 28, 1985, (1952)

248

53. Malamand, F., Moth. Phys. Analyse., 6, 349 (1970)

54. Gots,'i., Ikeda, S., Hirokawa, A, Senoh, H. and Tokoku, A.,

Japan Analyst, 13, 661 (1964)

55. Konovalov, V.A. and Frisch, S.B., J. Tech. Phys., h, 523 (1934)

56. McNally, J.R., Harrison, G.K. and Rowe, E., J. Opt. Soc. Am.,

37, 93 (1947)

57. Kirkbright, G.F. and Wilson, P.J., Anal. Chim. Acta, 68, 462 (1974)

58. Kirkbright, G.F., Ward,A.F. and West,T.S., Anal. Chim. Acta, 62 (1972)

59. Kirkbright,G.F., Ward, A.F. and*West, T.S., Anal. Chim.Acta,L (1973)

60. Walsh, A., Spec, Acta, 7, 108 (1955)

61. Kirkbright,G.F., Marshall,M. and West,T.S., Anal. Chem., 12379(1972)

62. Kirkbright,G.F., Marshall,M. and West,T.S., Anal. Chem.,4401288(1972)

63. Kirkbright,G.F., Wilson,P.J. and West,T.S.,At. Abs. News.111,53(1972)

64. Kirkbright, G.F., Analyst, 96, 609 (1971)

65. Woodriff, R., Appl. Spec., 28, 413 (1974)

66. Winefordner, J. D. and Vickers T.J., Anal. Chem., ll, 150R (1972)

67. King, A.S., Astrophys. J., 27, 353 (1908) 68. Collins, A.D. and Blackwell, D.E., Appl. Opt., 2, 1606 (1970) 69. L'vov, B.V., Spec. Acta, 17, 761 (1961)

70. L'vov, B.V., Spec. Acta, .2413., 53 (1969)

71. Lvov, B.V., Pure Appl. Chem., 23, 11 (1970)

72. Woodriff, R. and Ramelow, G., Spec. Acta, 23B, 665 (1968)

73. Massmann, H., Spec. Acta, 23, 215, (1968)

74. Manning, D.C., and Fernandez, F., At. Abs. News., 9, 65 (1970)

75. Kirkbright,G.F. and Wilson,P.J., At. Abs. News., 11, 140 (1974)

76. Morrow, R.W. and McElhaney, R.J., Appl. Spec., 27, 386 (1973)

77. Amos, N.D. and Matousek, J.P., Paper J2, Pitts. Conf., (1972)

78. Becker-Ross, H. and Fp1k, 4th Int. Conf. Appl. Spec.,

Toronto, Can. (1973)

79. Alder, J.F., and West, T.S., Anal. Chim. Acta, 51, 365 (1970)

80. Marshall, M., Ph.D. Thesis, London Univ., (1972)

249

81. Cantle, J.E. and West, T.S., Talanta, 20, 5 (1973)

82. Baird, R.2. and Gabrielian, S.M., Appl. Spec., 28, 273 (1974)

83. Donega, H.M. and Burgess, T.E., Anal. Chem., 1521 (1970)

84. Russell, B.J. and Walsh,A., Spec. Acta, 15, 883 (1959)

85. Garton, Adv. in Atomic and Phys., 2, 93 (1966)

86. Nelson, L.S. and Kuebler, N.A., Spec. Acta, 19, 781 (1963)

87. Wendt, R.R. and Tassel, V.A., Anal. Chem., 337 (1966)

88. Veillon, C. and Margoshes, M., Spec. Acta, 23B, 503 (1968)

89. Biochemists Handbook, Ed. C. Long, Spon (1961)

90. Yamamoto, Y., Humumani, T., Hayashi, Y. and Otani, Y.,

Bull. Chem. Soc. Japan, 38, 2204 (1965)

91. Kumumani,*T., Bull. Chem. Soc. Japan, .142, 956 (1969)

92. Christian, G.D. and Feldman, F.J., Anal. Chem. Acta, 40, 173 (1968)

93. Kiess, C.E. and Corliss, C,H., J. Res. Nat. Bur. Std.,

63A, 1 (1959)

94. Lawrence, G.M., Astrophys. J., 148, 261 (1967)

95. Dagnall, R., Johnson, D. and West, T.S., Anal. Chin. Acta,

67, 79 (1973)

96. Hieftje, G.M., Anal. Chem., 81A (1972)

97. Hieftje, G.M., Anal. Chem., /L, 69A (1972)

98. Sharpe, J., R/P021 Y 70, En' Symposium,(1964)

99. Wiese, W.L., Smith,M.W. and Miles, B.N.,

N.S.R.D.S. — N.B.S. 22, At. Trans. Probs., II, Washington D.C. (1969)

100. Lawrence, G.M., Astrophys J., 148, 261 (1967)

101. Magyar,B., Santos, F., Hely. Chien. Acta, 52, 820 (1969)

102. Dunk, R., Mostyn, R.A. and Hoare, P.C., At. Abs. News.,

8, 79 (1969)

103. Looyenga, R.U. and Huber, C.O., Anal. Chin. Acta, 55 , 179 (1971)

104. Rose, S.A. and Holtz, D.T., Anal. Chin. Acta, Lo 239 (1969)

105. Jugneis, E. and Anavi, Z., Anal. Chin. Acta, 55, 179 (1971)

250

106. Salet, J. Bull. Soc. Chim. Fr., 68, 404 (1869)

107. Veillon, C. and Park, J.Y., Anal. Chim. Acta, 60, 293 (1972)

108. Aldous, K.M., Dagnall, R.M. and West, T.S., Analyst, 15, 417 (1970)

109. Everett, G.L. and West, T.S., Anal. Chia:. Acta, 68, 387 (1974)

110. Belcher, R., Bogdanski, S.L. and Townshend, A.,

Anal. Chim. Acta, 67, 1 (1973)

111. Aldous, K.M., Dagnall, R.M., Pratt, S.J. and West, T.S.,

Anal. Chem., j, 1851 (1969)

112. Kirkbright, G.F. and Marshall, M., Anal. Chem., ZA, 1288 (1972)

113. Phillips, L.P., Rev. Sci. Inst., 2,2, 7 (1971)

114. Browner, R.F., Dagnall, R.M. and West, T.S., Anal. Chim. Acta,

Z11 163, (1969)

115. Dagnall, R.M., Silvester, M.D. and West, T.S., Talanta, 18, (1971).

116. L'vov, B.V., "Atomic Absorption Spectrochemical Analysis"

Adam Hilger Ltd., (1970)

117. Baird, R.B. and Gabrielion, Appl. Spec., 28, 273 (1974)

118. Rubeska, I. and Svoboda, V., Anal. Chim. Acta, 2z, 253 (1965)

119. Mitchell, A.C.G. and Zemansky„ M.N.,

"Resonance Radiation and Excited Atoms", Cambridge Press (1971)

120. Wiese,W.L., SmithIM.W. and Miles,B., IS.R.D.S.-U.B.S.22,II, (1969)

121. Parsons,M.,McCarhy,W. and Winefordner,J.,Appl.Spece,20,2231(1966)

122. Kirkbright, G.F. and Marshall, M., Anal. Chem., LA, 7 (1972)

123. L'vov, B.V., AAS - Sheffield Conf., 11 (1969)

124. Fuwa, K. and Vallee, B.L., Anal. Chem., Z1.1 188 (1969)

125. Thompson, B.A., Harteck, P. and Reeves, R.R.,

J. Geophys. Res., 68, 6431 (1963)

126. High Temperature Properties and Decomposition of Inorganic Salts.

Part 1. Sulrthates, - :1.13.S. 7 (1966)

127. Ottoway, J.M. and Campbell, W.E., Talanta, 21, 837 (1974)

128. L'vov, B.V. and Khartsyzov, A.D., Zh. Prikl. Spectrosk., 11, 9 (1969)

129. Thompson, K.C., Spectroscopy Letts., 3, 59 (1970)

251

130. Kirkbright, G.F. and Wilson, P.J., At. Abs. News., 13, 140 (1974)

131. Wood, S.J., Ph. 1). Thesis, London Univ., (1971)

132. Koirtychann, S.R. and Pickett, E.E., Anal. Chem., 38, 585 (1966)

133. Muller, L.A., Ann. Physik, 82, 39 (1927)

134. Koirtyohann, S. and Pickett, E.E., Anal. Chem., 35 1C37 (1966)

135. Ediger, R.D. Peterson, G.E. and Kerber, J.D.,

At. Abs. News., 13, 61 (1974).

136. Culver, B.R. and Surles, T., Anal. Chem., Z2, 920 (1975)

137. Fuller, C.W., Anal. Chico. Actal 62, 442 (1972)

138. Clark, D., Dagnall, P.M. and West, T.S., Anal. Chim. Acta,

63, 11 (1973)

139. Findlay, W.j., Zdrojewski,A. and ':cuickest, N., Specroscopy Letts.,

7, 63 (1974)

140. Johnson, D.J., Sharp, B.L. and West, T.S., Anal. Chem., L7, 1234 (1975)

141. Browner, R.F. and Winefordner, J.D., Anal. Chem., 44, 247 (1972)

142. Crosser, M.S. and Mullins, C.E., Anal. Chim. Acta, 68, 377 (1974)

143. Bratzel, M.D. and Chakrabarti, C.L., Anal. Chim. Acta, 63, 1 (1973)

144. Carver, J.H. and Mitchell, P., J. Sci. Inst., 555 (1964)

145. Byrann, A., Astrophys. J., 121, 480 (1956)

146. Byrann, A., Astrophus. J., 128, 738 (1958)

147. Chubb, T.A. and Friedman, H., Spec. Acta, 8, 121 (1956)

148. Strober, A., Scolnik, R. and Hennes, J.P., Appl. Opt., 2, 735 (1963)

149. Davis, D. and Braun, W., Appl. Opt., 7, 2071 (1968)

150. Vilesov, F.I., Sov. Phys. Usp., 6, 888 (1964)

151. Wigner, E.P., Phys. Rev., 73, ICO2 (1948)

152. Coltman, S., Phys. Rev., 102, 171 (1956)

153. Terenin, and Vilesov, F.I., Adv. in Photochem., 2, 385 (194)

154. Sharp, B., Ph.D. Thes-is, Univ. of London (1972)

155. Hurzeler, H., Inghram, M.G. and Morrison, J.D., J.Chem. Phys.,

28, 76 (1958)

252

156. Wilkinson, P.G., Johnston, H.L., j.Chem. Phys., 18, 190 (1950)

157. Johnson, D., M.D. Thesis, Univ. of London (1973)

- 158. "Ionisation Potentials, Appearance Potentials and heats Of

Formation Of Gaseous Positive Ions" U.S.N.B.S. (1969)

159. Handbook Of Chemistry and'Physics, E62, Chemical rubber Co.

160. Kristianpoller,N. and KnappIR.A., 53rd. Ed. (1973)

Appl. Opt., 915 (1964)

161. Bartley, D., Private Communication.

162. Sorokin, L.S., Adamchuk, V.K. and Dmitriev, A.B.,

Prib. i Tekhnika 2ks., .1216 (1970)

253