small-field dosimetryseaapm.org/meetings/2017/fri_0805_das-small-field... · 2017-03-09 · summary...
TRANSCRIPT
IJDas (1)
Indra J. Das, PhD, FAAPM, FACR, FASTRODepartment of Radiation OncologyNew York University Langone Medicine CenterNew York, NY
Small-Field Dosimetry
IJDas (2)
Treatment Fields
Advance Therapy FieldsSRS/SRTGamma KnifeCyber-KnifeTomotherapyIMRT
4x4 cm2 0.3x0.3 cm2
Magna-FieldsTraditional Fields
40x40 cm2 4x4 cm2200x200 cm2
Small Field
IJDas (3)
403530252015100.80
0.82
0.84
0.86
0.88
0.90
0.92
0.94
0.96
0.98
1.00
1.02
WEHU-ArizonaTemple UU PennErlanger Serago et al.Fan et al.Zhu et al.
Total scatter factor from different institutions
Cone Diameter (mm)
Cone
Fac
tor (
St)
Das et al, J Radiosurgery, 3, 177-186, 2000
Small Field Dosimetry Problem
Institutional variability in 6 MV Radionics SRS dosimetry
12% diff
IJDas (4)
403530252015100.80
0.82
0.84
0.86
0.88
0.90
0.92
0.94
0.96
0.98
1.00
Pinpoint (par)Pinpoint (per)0.125ion(par)0.125ion(per)
Diamond
TLD
MC(1mm)MC(5mm)
CEA (Film)Kodak(Film)
Total scatter factor with various detectors
Cone Diameter (mm)
Cone
Fac
tor (
St)
Das et al, J Radiosurgery, 3, 177-186, 2000
Dosimetric Variation with Detectors
14%
IJDas (5)
Wrong detector used for BrainLab cone calibration
Misadministration Media Coverage
IJDas (6)
Misadministration Media Coverage
IJDas (7)
What is a Small Field? Lack of charged particle Dependent on the range of secondary
electrons Photon energy
Collimator setting that obstructs the source size
Detector is comparable to the field size
IJDas (8)
CPE & Electron Range
Electron range= dmax in forward direction Electron range in lateral direction
Nearly energy independent Nearly equal to penumbra (8-10 mm)
Field size needed for CPE Lateral range 16-20 mm
CPE, Charged Particle Equilibrium
dmax
IJDas (9) Das et al, Med. Phys. 35, 206-215, 2008
Definition of Small Fields
IJDas (10)
�1.5�1.0�0.5 0.0 0.5 1.0 1.50.0
0.2
0.4
0.6
0.8
1.0
x �cm�2.0x2.0 cm2
Source sizescm
---- 0.01---- 0.35---- 0.50
�6.5�6.0�5.5�5.0�4.5�4.0�3.50.0
0.2
0.4
0.6
0.8
1.0
x �cm�
Ener
gyFl
uenc
eSc�
norm
alized
Energy Fluence Penumbra/Output
�1.5�1.0�0.5 0.0 0.5 1.0 1.50.0
0.2
0.4
0.6
0.8
1.0
x �cm�1.0x1.0 cm2
�1.5�1.0�0.5 0.0 0.5 1.0 1.50.0
0.2
0.4
0.6
0.8
1.0
x �cm�0.5x0.5 cm2
Large field
Nyholm et al Radiother Oncol 78, 347, 2006 and Phys Med Biol 51,6245, 2006
IJDas (11)
Cavity
Bragg-Gray Cavity: Range and Cavity Size
Dose = .S/ (number/cm2)(MeVcm2/g) = MeV/g= J/kg (Gy)
Range
Dose is calculated only by charged particle crossers in the cavity &Cavity should be non-perturbing
IJDas (12)
12
34 1
32
Lateral Electronic Equilibrium, LEE
Large Field Small Field
IJDas (13)
Density Effect
Water Vapor
Liquid Water
Dense Water
Bouchard et al, Med Phys, 42, 6063-6047, 2015
Bea
m o
n ca
vity
Bea
m o
ff ca
vity
IJDas (14)
Source Size
Jaffray et al, Med Phys 20, 1417-1427 (1993)90%, 70%, 50%, 30%, 10% iso-intensity line
IJDas (15)
Varian TrueBeam
Sawkey et al, Med Phys, 40, 332, 2013
Dimensionsx, y (mm)0.9, 0.7
0.8, 0.7
0.9, 0.9
0.9, 0.8
T2 T3T1
6 MV
6 MVFFF
10 MV
10 MVFFF
Source Size, Modern Machines
IJDas (16) Alfonso, et al. Med Phys 35, 5179-5186 (2008)
IAEA/AAPM Proposed Pathway
IJDas (17)
Why So Much of Fuss? Reference (ref) conditions cannot be achieved for most SRS
devices (cyberknife, gammaknife, tomotherapy etc)
Machine Specific reference (msr) needs to be linked to ref
Ratio of reading (PDD, TMR, Output etc) is not the same as ratio of dose
2
1
2
1
MM
DD
msrclin
msrclin
ffQQk
MM
DD ,
,2
1
2
1 1
2
2
1,, M
MDDk msrclin
msrclin
ffQQ
Alfonso et al. Med Phys 35, 5179-5186 (2008)
IJDas (18)
Relative Dosimetry
rel
relf
Qf
Q
fQ
fQff
QQ adingOutput
MDMD
k msrclin
msrclin )(Re)(
//
clin
msr
clin
msr
clin
clin
msr
clin
,w,w
,w,w,,
refmsr
msr
msr
msr
msr
msr
,,,,,
ffQQQoQQoDW
fQ
fQw kkNMD
msrclin
msrclinmsr
msr
clin
clin
msr
msr
msr
msr
clin
clin
clin
clin
msr
msr
clin
clinmsrclin
msrclin
,,
,w
,w
// ff
QQfQ
fQ
fQ
fQ
fQ
fQ
fQ
fQff
QQ kMM
MDMD
MM
msrfmsrairw
fclinfclinairwffQQ PS
PSk msrclin
msrclin
)()(
,
,,,
IJDas (19)
Small Field Dosimetry Publications
IAEA-TRS 483, 2017
+ TG-155 (AAPM)Relative dosimetry
IJDas (20)
Meaning of k in Micro-Chambers
Kumar et al, PMB, 60, 8187, 2015
Bouchard et al, Med Phys, 36, 4654, 2009
IJDas (21)
Stopping Power Ratio
IJDas (22) Sanchez-Deblado et al, Phy. Med. Biol., 48, 2081, 2003
Radiological Parameters
IJDas (23) F. Araki, Med. Phys. 33, 2955-2963 (2006).
Cyber Knife Dosimetry
0.5%
IJDas (24)
Sensitivity vs Volume of Detectors
0
10
20
30
40
50
60
70
80
90
100
110
120
130
140
1.0E-04 1.0E-03 1.0E-02 1.0E-01 1.0E+00
Volume (cm3)
Rel
ativ
e se
nsiti
vity
PFD
SFD
A-16 PinPoint Markus IC40.125cc
0.3cc
0.6 cc
Detector Response
IJDas (25)
Correction Factors
Francescon, et al Med Phys 35, 504, 2008
Correction Factor depends on:Field sizeSource size (FWHM; machine type)Detector type
IJDas (26)
0.90
0.92
0.94
0.96
0.98
1.00
1.02
1.04
1.06
1.08
1.10
1.12
1.14
1.16
0 5 10 15 20 25 30 35
Corre
ction
facto
r
Field size (mm)
Siemens; PTW diode 60012Elekta; PTW diode 60012Siemens; Exradin A16Elekta; Exradin A16Siemens; Sun Nuclear Dedge"Elekta; Sun Nuclear Dedge"Siemens; PTW Pinpoint 31014Elekta; PTW Pinpoint 31014Siemens; PTW microLionElekta; PTW microLion
Francescon et al Med Phys,38, 6513, 2011
Published data on
IJDas (27)
Monte Carlo Derived
Kumar et al, PMB, 60, 8187, 2015
IJDas (28) Med. Phys. 38 (12), 6992, 2011
of Linear Accelerators (Varian)
IJDas (29)
Radiotherapy and Oncology 100 (2011) 429–435
Med. Phys. 38 (12), 6513-6527, 2011
of Linear Accelerators
IJDas (30)
Cyber Knife
Pantelis et al, Med Phy. 37, 2369-2379, 2010
IJDas (31)
Med Phys, 41(4), 041711, 2014
IJDas (32)
IJDas (33)
Correction in Profile (OAR)
Francescon et al, Med Phys, 41(10), 101708, 2014
5 mm cone 25 mm cone
IJDas (34)
Profile & Source Size
Sham et al, Med Phys, 35, 3317-3330, 2008
IJDas (35)
Effect of Focal Spot
0.5
0.6
0.7
0.8
0.9
1.0
1.6 1.8 2.0 2.2 2.4 2.6 2.8
Out
put (
Mon
te C
arlo
)
Full Width at Half Maximum, FWHM (mm)
20 mm Cone10 mm cone7.5 mm cone5 mm cone
Moignier et al, Med Phys 41(7), 071702, 2014
IJDas (36)
Effect of Beam Energy
0.5
0.6
0.7
0.8
0.9
1.0
6.4 6.6 6.8 7.0 7.2 7.4 7.6
Out
put (
Mon
te C
arlo
)
Effective Photon Energy (MeV)
20 mm Cone10 mm cone7.5 mm cone5 mm cone
Moignier et al, Med Phys 41(7), 071702, 2014
IJDas (37) Papaconstadopoulos, et al, Phys. Med. Biol. 59 (2014) 5937–5952
Good News with Modern Microdetectors
IJDas (38)
IJDas (39)
Machine, Cone & Depth?
Radiotherapy Oncology, 112, 442-446, 2014
IJDas (40)
PTW-Pinpoint-
per
PTW-Pinpoint-
par
IBACC01-per
IBACC01-par PTW-LIC
PTW-Micro
DiamondSI PSD IBA-EFD IBA-SFD
PTW-Unshielded Diode
IBA-PFDPTW-
ShieldedDiode
Series1 14.7 10.2 8.1 5.0 1.1 0.2 0.0 -0.9 -2.0 -5.0 -5.3 -9.0
-14-12-10-8-6-4-202468
10121416
% D
iffer
ence
from
1.0
in k
Fac
tor
k Factor for Varian IX Machine, 5x5 mm2 Field
Benmakhlouf et al, Med Phys 41(4), 041711, 2014
IJDas (41)
k Factor of CyberKnife 5 mm Cone
Moignier et al, Med Phys 41(7), 071702, 2014
PTWPinPoint
Diamond(old) SI-PSD EBT2 MicroTLD IBA-SFD
PTWunshielded
Diode
PTWshieldedDiode
SN-Edge
Series1 12.0 7.5 0.1 -1.4 -1.4 -3.0 -5.0 -6.0 -6.0
-14-12-10-8-6-4-202468
10121416
% D
iffer
ence
from
Mon
te C
arlo
in
k F
acto
r
IJDas (42)
0.90
0.92
0.94
0.96
0.98
1.00
1.02
1.04
1.06
0.0 0.5 1.0 1.5 2.0 2.5 3.0
Cor
rect
ion
Fact
or
Field Size (cm x cm)
- MC-Francescon et al (2011)
- SFD-Azangwe et al (2014)
- MLIC-Azangwe et al (2014)
10.3%
1.5%0.2%
for Detectors msrclin
msrclin
ffQQk ,
,
IJDas (43)
for select detectors
0.90
0.92
0.94
0.96
0.98
1.00
1.02
1.04
1.06
0.0 0.5 1.0 1.5 2.0 2.5 3.0
Cor
rect
ion
Fact
or
Field Size (cm x cm)
- microLion - Diode D1V - Scintillator, W1 (parallel) - Scintillator, W1 (perpendicular) -microDiamond (T-60019)
6.9%
0.8%1.8%
msrclin
msrclin
ffQQk ,
,
IJDas (44)
for Detectors and Devicesmsrclin
msrclin
ffQQk ,
,
Francescon, Beddar & Das, ESTRO, 2014
IJDas (45)
Tomotherapy
MLC, Linac
Cyber Knife Field size (cm)
TRS-483, 2016
of Different Devices and Selected Detectorsmsrclin
msrclin
ffQQk ,
,
IJDas (46)
MicroDiamond-Multi-institution
Masi et al, Physica Medica,32, 1637-1643, 2016
CyberKnife
IJDas (47)
Plastic Scintillator-Multi-institution
Masi et al, Physica Medica,32, 1637-1643, 2016
CyberKnife
IJDas (48)
0.00.10.20.30.40.50.60.70.80.91.01.1
0 1 2 3 4 5 60.00.10.20.30.40.50.60.70.80.91.01.1
0 1 2 3 4 5 60.00.10.20.30.40.50.60.70.80.91.01.1
0 1 2 3 4 5 60.00.10.20.30.40.50.60.70.80.91.01.1
0 1 2 3 4 5 6
Validity of Proposed Method
Field Size (cm)
Rel
ativ
e do
se a
t dm
ax
IJDas (49)
Summary Understand the limit of small field, in terms of focal
spot & spectral issues Detectors that are water equivalent like, MicroLion,
MicroDiamond, Plastic Scintillators and EBT are best suited
Use proper correction factors to correct detector response to correct for the dose
Use data from for IAEA TRS-483 and TG-155 guidelines when published
IJDas (50)
Thanks