Simulation and Feedback control in Atomic Force Microscope. Michal Hrouzek, Alina Voda, Martin Stark, Joël Chevrier Laboratoire d’Automatique de Grenoble, INP/UJF Grenoble The European Synchrotron Radiation. Outlines of the presentation. AFM description Feedback control in AFM - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Simulation and Feedback control in  Atomic Force Microscope

1

Simulation and Feedback control in

Atomic Force Microscope

Michal Hrouzek, Alina Voda, Martin Stark, Joël Chevrier

Laboratoire d’Automatique de Grenoble, INP/UJF Grenoble

The European Synchrotron Radiation

Page 2: Simulation and Feedback control in  Atomic Force Microscope

2

Outlines of the presentation

AFM description Feedback control in AFM Sources of noise in AFM Cantilever model Driving loop with controller Interaction forces Thermal noise in AFM Conclusion

Page 3: Simulation and Feedback control in  Atomic Force Microscope

3

Schema of Dynamic Force Microscopy

• Head of the AFM– Driving loop– Head positioning loop

• Stage with a sample– x axis positioning loop– y axis positioning loop

(excitation

(set-value of

Page 4: Simulation and Feedback control in  Atomic Force Microscope

4

Detection techniques in dynamic AFM

• Amplitude Modulation (AM)– Original operation technique, Developed by Y. Matin – J. Appl. Phys. 61 (10)– Driver is exciting the cantilever with constant driving signal.– Interaction forces affect the cantilever and lower the vibration amplitude.– Change in amplitude depends directly on interaction force.

• Frequency Modulation (FM)– Newer technique, Developed by T.R. Albrecht – J. Appl. Phys. 69 (2)– Driver with controller is exciting the cantilever to constant vibration amplitude.– Interaction forces affect the cantilever and change the resonant frequency.– Frequency shift depends directly on interaction force. f– More sensitive compare to AM-technique

– Further would be treated only FM technique.

Page 5: Simulation and Feedback control in  Atomic Force Microscope

5

Outlines of the presentation

AFM description Feedback control in AFM Sources of noise in AFM Cantilever model Driving loop with controller Interaction forces Thermal noise in AFM Conclusion

Page 6: Simulation and Feedback control in  Atomic Force Microscope

6

Feedback control in AFM

• Loop controlling position of AFM head in z axis– Maintaining constant set-value of frequency shift

– Low frequency response (1 – 30 kHz)

• Cantilever driving loop (cantilever excitation)– Exciting the cantilever and ensuring that contact with surface is not lost

– High frequency response (1kHz – 1 MHz)

– Nonlinear behavior of the driver (due coupling with surface)

– Directly maintaining cantilever vibration amplitude

– Could be used for possible attenuation of thermal noise perturbation and influence of another noises.

Page 7: Simulation and Feedback control in  Atomic Force Microscope

7

Position of the head and lever excitation

rzd(t), rzp(t) – desired excitation and set-value of frequency shift

mz(t) – measured deflection of the cantilever

ezd(t), ezp(t) – regulation errors of the bimorph and head

zdri(t), zpos(t) – driving signal and head positioning signal

z(t) – real deflection of the cantilever

Page 8: Simulation and Feedback control in  Atomic Force Microscope

8

Feedback control in AFM

• Loops controlling position of AFM stage in x-y axes

– Scanning• Positioning of the stage with sample• Simple movement in straight lines under the cantilever with tip

– Manipulation• Particles manipulation at the surface• Complex movement with many possible shapes• Control of the applied force onto the particle is crucial

Scanning Manipulation

Page 9: Simulation and Feedback control in  Atomic Force Microscope

9

Loops controlling stage position

rx,y(t) – desired position piezo-electric stack

mx,y(t) – measured (estimated) position of the stage

ex,y(t) – regulation error

ux,y(t) – driving signal

yx,y(t) – real position of the stage

Page 10: Simulation and Feedback control in  Atomic Force Microscope

10

Loops controlling stage position

• Accuracy problems with piezo-electric actuators– Positioning nonlinearities (getting bigger with increasing speed)

– Usually are used piezo stacks with hysteresis 10% - 15% of max. displacement

(Harder stacks have smaller hysteresis but smaller displacement range)

– Drift due to creep

(Could be reduced to curtain level by careful design of the stage)

• Measurement problems of stage position– High level of noise of LVDT detectors get relevant at nano-meter resolution

– Observer based regulator can achieve better positioning resolution

Page 11: Simulation and Feedback control in  Atomic Force Microscope

11

Outlines of the presentation

AFM description Feedback control in AFM Sources of noise in AFM Cantilever model Driving loop with controller Interaction forces Thermal noise in AFM Conclusion

Page 12: Simulation and Feedback control in  Atomic Force Microscope

12

Sources of noise in AFM

• Electronic parts

– Photo detector• Optical path noise• Thermal gradient• Laser intensity noise• Shot noise• Laser mode noise• Laser phase noise

– Electronic circuits• Electrostatic noises• Noise of amplifiers

• Mechanical parts

– Cantilever and tip• Thermally induced cantilever noise• Mechanical vibration• Relaxation• Air turbulences and acoustic waves• Magnetic noises • Electrostatic noises• Chemical noises

– Bimorph• Thermally induced noise• Electrostatic noises

Page 13: Simulation and Feedback control in  Atomic Force Microscope

13

Thermal noise

• Thermal noise is limiting AFM sensitivitySome noises could be lower by appropriate design and construction of AFM.

• The minimum detectable interaction force. (dynamic mode)

k Spring constant, stiffness

T Temperature kB Boltzmann constant

Measurement bandwidth

A0 Vibration amplitude

0 Resonance frequency

Q Quality factor

Page 14: Simulation and Feedback control in  Atomic Force Microscope

14

Outlines of the presentation

AFM description Feedback control in AFM Sources of noise in AFM Cantilever model Driving loop with controller Interaction forces Thermal noise in AFM Conclusion

Page 15: Simulation and Feedback control in  Atomic Force Microscope

15

Cantilever model

• Second order differential equation model

• Q 0.1-1 liquid environment

1-1000 air environment

1000-100000 vacuum environment

• L 10-500m length

• w 10-50m width

• t 0.1-5m thick

• E 130-180GPa for silicon cantilevers

• k 0.01-100N/m stiffness

Page 16: Simulation and Feedback control in  Atomic Force Microscope

16

Cantilever model

• Multimode model of the cantilever

• E Modulus of elasticity

• I Area moment of inertia

• m Mass per unit length

• L Cantilever length

Page 17: Simulation and Feedback control in  Atomic Force Microscope

17

Cantilever model

• Multimode model of the cantilever

• E Modulus of elasticity

• I Area moment of inertia

• m Mass per unit length

• L Cantilever length

Page 18: Simulation and Feedback control in  Atomic Force Microscope

18

Computer Simulation

• The cantilever properties used for multimode model

• Computed properties of separated harmonic modes

Page 19: Simulation and Feedback control in  Atomic Force Microscope

19

Computer Simulation

• Spectral analysis schema of the multi mode cantilever model

Page 20: Simulation and Feedback control in  Atomic Force Microscope

20

Computer Simulation - results

• Thermal noise was the only excitation of the Cantilever.(displayed spectra is an average over 100 FFT)

Page 21: Simulation and Feedback control in  Atomic Force Microscope

21

Measured spectra

• non-contact silicon cantilever NSC12/50 (cantilever F)(displayed spectra is an average over 1000 FFT)

Page 22: Simulation and Feedback control in  Atomic Force Microscope

22

Computer Simulation - results

• Time response of thermally driven cantilever-Time sequence 0 to 2ms -Time sequence 0 to 0.2ms (Zoom)

Modeled thermal excitation (blue)

(normal distribution)

Position of the cantilever (red)

Zoom

Modelinitialization

Page 23: Simulation and Feedback control in  Atomic Force Microscope

23

Computer Simulation - results

• Time response of artificially driven cantilever at resonanceblue curve – exciting displacement zdriv(t)=Z0sin(0t)red curve – displacement of the cantilever at its free end

- Time sequence 0 to 10ms - Time sequence 0 to 0.4ms (zoom)

Zoom

Page 24: Simulation and Feedback control in  Atomic Force Microscope

24

Outlines of the presentation

AFM description Feedback control in AFM Sources of noise in AFM Cantilever model Driving loop with controller Interaction forces Thermal noise in AFM Conclusion

Page 25: Simulation and Feedback control in  Atomic Force Microscope

25

Computer Simulation with controller

– Driving with the PID regulator

Page 26: Simulation and Feedback control in  Atomic Force Microscope

26

Computer Simulation with controller

• Driver with phase shift

• Band pass filter – selects the frequencies of our interest (10kHz-100kHz)

• Amplitude detector – gives numerical value of vibration amplitude A(t)

• Controller – select the gain(t) that is multiplied by filtered cantilever position

• Low pass filter – eliminates high frequency noise

• Gain – amplification of the signal

• Phase shift – optimal value is /2

Page 27: Simulation and Feedback control in  Atomic Force Microscope

27

Computer Simulation with controller - results

– Gain of the PID controller gain(t)

Page 28: Simulation and Feedback control in  Atomic Force Microscope

28

Computer Simulation with controller - results

– Displacement of the driving bimorph zbimorph(t) = kbimorph zdrive(t)

Page 29: Simulation and Feedback control in  Atomic Force Microscope

29

Computer Simulation with controller - results

– Displacement of the cantilever (red curve)– Displacement of the driving bimorph (blue curve)

Page 30: Simulation and Feedback control in  Atomic Force Microscope

30

Outlines of the presentation

AFM description Feedback control in AFM Sources of noise in AFM Cantilever model Driving loop with controller Interaction forces Thermal noise in AFM Conclusion

Page 31: Simulation and Feedback control in  Atomic Force Microscope

31

Interaction forces

• Properties– They are non-linear

– Can be long-range or short-range, and attractive or repulsive.

– Forces are very sensitive to environmental conditions such as temperature, humidity, surface chemistry, and mechanical and electrical noises.

– Depend on the material, geometry and size of nano entities.

– Curtain forces are getting dominant in specific environments and some diminish.

• Interaction forces– van der Waals (analogous to the gravity at the nano-scale)

– Casimir

– Thermal motion: Exist for any material and depends only on temperature

– Capillary, Hydrogen and Covalent bonding, Steric, Hydropobolic, Double layer, …

Page 32: Simulation and Feedback control in  Atomic Force Microscope

32

Interaction forces

• Intensity of the van der Waals and repulsive forces.

Page 33: Simulation and Feedback control in  Atomic Force Microscope

33

Interaction forces

• Approach curve of the cantilever.

1) Non-contact

2) “Snap on” point, spring constant is smaller than attractive force.

zero) Equilibrium point, lever isn’t deflected in any direction.

3) Repulsive interaction is dominant.

4) Maximum positive deflection.

5) Capillarity holds the tip onto the surface

6) “snap off” point, spring constant overcomes the capillarity

Page 34: Simulation and Feedback control in  Atomic Force Microscope

34

Interaction forces model

• Intensity of the interaction forces as a function of the separation distance.

• Mathematical equations describing interaction forces between the tip and surface

a0 – intermolecular distance

AH – Hamaker constant

RS – Sphere radius (end of the tip)

E* – Effective stiffness

• Numerical values used from reference: S. I. Lee, Physical Review B, 66(115409), 2002.

Page 35: Simulation and Feedback control in  Atomic Force Microscope

35

Interaction forces model - results

Approach curve with the cantilever, Q=1 Approach curve with the cantilever, Q=100

• Approach curve is simulated without any excitation, chip with the cantilever is slowly approaching the surface

• This behavior has been observed at the experiment. (Martin Stark, Frederico Martins)

Page 36: Simulation and Feedback control in  Atomic Force Microscope

36

Interaction forces model - results

– Cantilever has been excited with constant driving signal zbimorph(t) = Adrivesin(0t)

– Harmonic outputs of separated modes has been recorded.

– Amplitude of first harmonic mode is decreasing with smaller separation distance.

Separation

distance

Vibration

amplitude

Page 37: Simulation and Feedback control in  Atomic Force Microscope

37

Interaction forces model - results

– Amplitude of higher harmonic modes are increasing with smaller separation distance.

– Time sequences of one period are shown

(second mode – red curve, third mode – green curve, fourth mode – blue curve )

Page 38: Simulation and Feedback control in  Atomic Force Microscope

38

Outlines of the presentation

AFM description Feedback control in AFM Sources of noise in AFM Cantilever model Driving loop with controller Interaction forces Thermal noise in AFM Conclusion

Page 39: Simulation and Feedback control in  Atomic Force Microscope

39

Outlines of the presentation

AFM description Feedback control in AFM Sources of noise in AFM Cantilever model Driving loop with controller (model of PLL) Interaction forces Thermal noise in AFM Conclusion

Page 40: Simulation and Feedback control in  Atomic Force Microscope

40

Conclusion

• Multimode cantilever model has been developed. Simulations have shown that the model is correct approximation of the dynamic system.

• Model of the interaction forces have been implemented into Matlab Simulink environment.

• Dynamic interaction between both models has been simulated and compared with measurements.

• Driving controller has been employed to control the excitation of the cantilever interacting with the surface.

Page 41: Simulation and Feedback control in  Atomic Force Microscope

41

Acknowledgements

• Frederico Martins• Mario Rodrigues• Raphaelle Dianoux

Page 42: Simulation and Feedback control in  Atomic Force Microscope

42

Future work

• Development of microscope stage controllers that are responsible for the sample positioning under the head with cantilever. This controller has to fulfill new requirements for speed and accuracy due to application of the AFM as a nano-manipulator.

• Thermal noise is second field of further work. Driving controller has to be redesigned to lower the nose signal ration to achieve better results in weak forces measurements.