 Varian 500-MS LC Ion Trap Advantage Statement: SelecTemp allows the Varian 500-MS to deliver increased sensitivity and selectivity for all analytes, including thermally labile compounds. Other benets include reduced interface contamination, improved throughput, simplied setup, and automatically recorded parameters for GLP purposes. The patent pending, temperature programmable API drying gas control is a unique feature of the Varian 500-MS. Positive and negative temperature programming ensures the optimum drying gas temperature for all mobile phase compositions in gradient separations resulting in greater sensitivity and productivity . The drying gas temperature is specied based on the mobile phase composition during the run, as shown in Figure 1. Generally gradient LC analysis increases the percent of organic phase and decreases the percent of aqueous phase during the run. Therefore the optimum drying gas temperature should also decrease during t he run. Constant temperatures would result in excess heating of the electrospray droplets and may cause destruction of the thermally labile compounds. However, at the end of a gradient run when the column is being regenerated and the aqueous composition of the mobile phase is high, it is advantageous to have a high drying gas temperature to ensure complete desolvation of droplets to reduce contamination of the API interface. Using SelecTemp, start and end temperatures are specied with their hold time, and the rate is calculated based on the full run time. The temperature prole created by these specications can be overwritten by lower or higher values specied for the individual time segments as shown in Figure 2. Set point Specications, apply to the whole ru n; while Segment Specication applies only for the duration of the specied time segment (Figure 3). Figure 1 Schematic of SelecTemp Figure 2 ESI and SelecTemp set points for total analysis run Figure 3 ESI and SelecTemp set points for chromatograph ic  segment time ADVANTAGE NOTE SelecTemp

Upload: aptureinc

Post on 31-May-2018

216 views

Category:

Documents


0 download

TRANSCRIPT

8/14/2019 SI-0243 10 28 05 good jb

http://slidepdf.com/reader/full/si-0243-10-28-05-good-jb 1/2

 Varian 500-MS LC Ion Trap

Advantage Statement:  SelecTemp allows the Varian

500-MS to deliver increased sensitivity and selectivity

for all analytes, including thermally labile compounds.

Other benefits include reduced interface contamination,

improved throughput, simplified setup, and automatically

recorded parameters for GLP purposes.

The patent pending, temperature programmable API drying

gas control is a unique feature of the Varian 500-MS.

Positive and negative temperature programming ensures

the optimum drying gas temperature for all mobile phase

compositions in gradient separations resulting in greater

sensitivity and productivity.

The drying gas temperature is specified based on the

mobile phase composition during the run, as shown in

Figure 1. Generally gradient LC analysis increases the

percent of organic phase and decreases the percent of 

aqueous phase during the run. Therefore the optimum

drying gas temperature should also decrease during the

run. Constant temperatures would result in excess heating

of the electrospray droplets and may cause destruction of 

the thermally labile compounds. However, at the end of a

gradient run when the column is being regenerated and

the aqueous composition of the mobile phase is high, itis advantageous to have a high drying gas temperature

to ensure complete desolvation of droplets to reduce

contamination of the API interface.

Using SelecTemp, start and end temperatures are specified

with their hold time, and the rate is calculated based on

the full run time. The temperature profile created by these

specifications can be overwritten by lower or higher values

specified for the individual time segments as shown in

Figure 2. Set point Specifications, apply to the whole run;

while Segment Specification applies only for the duration

of the specified time segment (Figure 3).

Figure 1 Schematic of SelecTemp

Figure 2 ESI and SelecTemp set points for total analysis run

Figure 3 ESI and SelecTemp set points for chromatographic

 segment time

ADVANTAGE NOT

SelecTemp™

8/14/2019 SI-0243 10 28 05 good jb

http://slidepdf.com/reader/full/si-0243-10-28-05-good-jb 2/2

SelecTempTM

 Varian, Inc.www.varianinc.comNorth America: 800.926.3000, 925.939.2400Europe The Netherlands: 31.118.67.1000Asia Pacific Australia: 613.9560.7133Latin America Brazil: 11.3845.0444 Other sales offices and dealers throughout the world

GC • LC • MS• AA • ICP • ICP-MS • UV-Vis-NIR • FT-IR • Raman • Fluorescence • Dissolution • NMR • MRI • Consumables • Data Systems

SI-0243 10/05 Printed in U.S.A.

The controlled drying gas temperature also increases sensitivity for thermally

labile compounds. To demonstrate this, 1 ng of Naphtyl Glucuronide (Figure 4),

was injected on column at different drying gas temperatures, with otherwise

identical conditions. The mobile phase was 0.1% acetic acid in water (solvent

A) and 0.1% acetic acid in acetonitrile (solvent B) at 0.2 mL/min. The gradient

conditions are given in Table 1.

The temperature program of the drying gas started at 330º C with a 2 minute

hold time, followed by programming the temperature from 330º C to 300º C in 3

minutes.

The reconstructed ion chromatogram for m/z=199 from positive ESI MRM,

Precursor 343‡ product 175-205, is shown in Figure 5. Red trace: Naphtyl

Glucuronide with programmed drying gas, (down ramped) green trace with

isothermal drying gas. Note the signal intensity increase and noise decrease

using the temperature programming of the drying gas, ultimately resulting in

an S/N increase from 146 to 462. Although the temperature of the drying gas

was only lowered by 30º C from the initial value, SelecTemp provided a factor of 

three improvement in signal-to-noise.

Figure 5 Use of SelecTemp directly translates into improved S/N ratiofor Naphtyl-Glucuronide

Table 1 HPLC conditions used for the study 

Figure 4 Chemical structure of 

Naphtyl-Glucuronide

Naphtyl Glucuronide

1 ng on column

Isothermal

Temperature

programmed

1 ng S/N=462

S=192184 N=416

1 ng S/N=146

S=110049 N=752

Timemin:sec %A %B

Flow ratemL/min

0:00 95 5 0.2

1:30 95 5 0.2

2:00 57 43 0.2

3:45 5 95 0.2

4:30 5 95 0.2

4:45 95 5 0.2

5:00 95 5 0.2