shimadzu icpms-2030 trace elements in clinical samples...clinical samples using quadrupole icp-ms ....

1
But still as Argon is one of the major element present within the mass spectrometer, the Argon-Argon polyatomic interference needs to be considered, especially when the major components of the sample is changing. In this case for urine, serum and whole blood, the changing amount of Carbon gives some further difficulties. The reason is, that carbon forms polyatomic ions with Argon as well. So as more carbon is present, as more Carbon-Argon-Ions are formed, thus it influences the equilibrium of the formation of Argon-Argon-Ions and affects the determination of 78 Se. At the same time with the increasing amount of Carbon, other elements like 52 Cr are interfered by 40 Ar 12 C, which gives the idea that easy matrix matching by adding a carbon source to the internal standard could solve these troubles without the need for applying reaction gases which would save additional measurement time. General Method Parameters The ICPMS-2030 has been used in basic Minitorch setup, enabling drastically lower flow rates of Argon. Even for high matrix samples the total consumption is below 10 l/min. At the same time the Argon purity requirements for the ICPMS-2030 are on a very low level (min. Argon 3.5 = 99.95 %). The basic method parameters are summarized in table 1. Tab. 1: ICPMS-2030 Configuration and Method Parameters Sample Measurement - Urine For Urine, the preparation of samples is very easy. All samples have been diluted by a factor of 10 using acidified purified water (1 Vol-% HNO 3 ). As the internal standard solution is added automatically, the sampler just have to be placed to the sampler. Urine typical consist of 95% water. After applying factor 10 dilution the amount of TDS cannot exceed 0,5%. Thus, the determination of trace elements using ICPMS technique is not most challenging. For evaluation of the method and instrument capabilities, certified reference material has been analysed (Table 2). The result mentioned is an average of 2 single measurements, whereat both single results meet the certified range. Sample Measurement – Serum and Whole blood For the measurement of these two types of sample it now starts that the influence of Carbon is increasing. The samples are more and more different compared to the reference calibration solutions, which do not containing any carbon. For example Serum contains appr. 91% water and about 7 % proteins (albumin and globulins). Easy and Fast Determination of Trace Elements in Clinical Samples using Quadrupole ICP-MS Jan Knoop 1 , Nils Garnebode 2 , Konstantin Kartaschew 2 , Uwe Oppermann 1 , Ludivine Fromentoux 3 1 Shimadzu Europa GmbH, Germany; 2 Shimadzu Deutschland GmbH, Germany; 3 Shimadzu France SAS, France Overview and Objective The analysis of trace elements in clinical samples is a frequent required objective nowadays and oftentimes connected to the use of ICPMS techniques. The first question then is, which technique exactly can meet the user requirements as different measurement principle are available within commercial quadrupole ICPMS devices to overcome interferences and achieving correct results. In addition to standard measurement mode these commonly are kinetic energy discrimination (KED) using Helium as collision gas or special reaction modes with gases like Hydrogen or Ammonia. Whereat the kinetic energy discrimination mode is very easy to handle, as it affects the resulting mass spectra in basically the same way and no account has to be taken to specific behaviours to specific isotopes, the reaction modes with each different gas always needs to be improved for different isotopes individually in order to reduce interferences. For each mode mentioned above, the ICPMS collision/reaction cell needs to be stabilized. Thus, as more and more different modes are applied, the measurement times are drastically extended. Within this poster the ease of uses of matrix matching is shown with the goal to use as less measurement modes as possible in order to save analysis time. Figure 1: Inductively Coupled Plasma Mass Spectrometer ICPMS-2030 (Shimadzu) Clinical Samples and Typical Interferences For the determination of trace elements in urine, serum and whole blood, different measures must be considered to achieve correct results. While it is quite easy to determine trace elements in urine after dilution using acid media (x10), the same is getting more challenging for serum or whole blood matrix, due to the higher amount of carbon and subsequent appearing interferences for example for the determination of selenium. In general for quantification of trace elements the most abundant isotope is selected. Taking Selenium into consideration, this would be 80 Se (49.61%). But at the same time, one of the most frequent polyatomic interferences affect m/z 80, in this case it is the polyatomic ion of 40 Ar 40 Ar from the Argon used as plasma gas. That is why the 2 nd most abundant isotope, 78 Se is most commonly selected (23.78 %). Here polyatomic ions of Argon interfere as well ( 38 Ar 40 Ar). But as the 38 Ar is much less abundant than the 40 Ar, the level of interference to 78 m/z is appr. factor 1500 less than to 80 m/z. As a result, the signal to noise ratio of 78 Se is several times better leading to higher sensitivities compared to 80 Se, even 78 Se is less sensitive. ____________________ [1] Nelms, Simon M., Inductively Coupled Plasma Mass Spectrometry Handbook, CRC Press LLC, 2005 Shimadzu Europa GmbH Albert-Hahn-Str. 6-10 D-47269 Duisburg +49(0)203/7687-0 www.shimadzu.eu [email protected] Tab. 2: Urine quantification results (all in µg/l) Tab. 3: Serum quantification results (all in µg/l) Tab. 4: Whole Blood quantific. results ( A µg/l B mg/l) A similar concentration of water (90%) is the main component of whole blood. Anyway – in accordance to diagnostic sensitivity requirements, different dilution factors have been applied. The serum was diluted similar to urine by a factor of 10 and the whole blood by a factor of 20. For both dilutions again the applied media was 1 Vol-% nitric acid. To cover additional interferences caused by carbon, the internal standard solution contains a certain level of Ethanol and Triton X-100 in order to have a final concentration of 3 vol.-% Ethanol and 0,3 weight-% Triton X-100 present after mixing by T-piece. Summary As indicated by table 2-4 all results meet the certified range of the reference material by just using standard and/or the very common KED mode. Any additional reaction gases have not been necessary. So the ICPMS-2030 is easy to use for the measurement of clinical samples by e.g. adding a carbon source to the internal standard solution. Parameter Setting RF generator power 1.2 kW Plasma gas 8 l/min Auxilliary gas 1,1 l/min Carrier gas 0.7 l/min Nebulizer coaxial Sampling depth 6 mm Spray Chamber temperature 5 °C Collision Cell Gas flow (Helium) 6 ml/min Cone material Copper Internal Standard Automatic Addition (T-piece) Sample tube Black-black, PVC Internal Standard tube Orange-blue, PVC Optional Fast Rinse Unit ASXPress Plus (Catac Teledyne) Isotope Mode Internal Std. Element Recipe Level I Recipe Level II Certified Result Certified Result Min Max Average Min Max Average 52 Cr KED Ge 3.26 4.89 4.17 15.9 23.8 19.9 65 Cu KED Ge 29.4 44.1 38.7 73.5 110 91.3 55 Mn KED Ge 3.13 4.69 3.93 15.5 23.2 19.5 60 Ni standard Ge 4.73 7.10 5.91 34.4 51.7 41.6 75 As standard Ge 34.4 51.6 42.7 66.6 99.9 79.8 111 Cd standard Te 1.97 2.95 2.46 11.5 17.2 14.5 206 Pb KED Ho 19.2 28.8 20.2 52.0 78.0 54.9 203 Tl KED Tb 5.79 8.69 6.40 15.2 22.8 17.3 78 Se KED Rh 24.0 35.9 28.7 66.6 99.9 87.4 27 Al standard Be 26.4 39.6 31.6 68.7 103 84.9 59 Co standard Ge 1.63 2.44 1.94 27.8 41.7 32.2 68 Zn KED Y 163 245 192 428 642 515 127 I KED Y 89.9 150 136 373 622 528 57 Fe KED Ge 30.8 46.3 42.5 179 268 232 Isotope Mode Internal Std. Element Recipe Level I Recipe Level II Certified Result Certified Result Min Max Average Min Max Average 63 Cu KED Ge 681 921 821 1140 1540 1410 195 Pt KED Lu 7.11 10.7 10.1 68.8 103 917 78 Se KED Rh 52.9 79.3 67.9 84 126 109 64 Zn KED Y 1120 1520 1400 1730 2350 2280 Isotope Mode Internal Std. Element Recipe Level I Recipe Level III Certified Result Certified Result Min Max Average Min Max Average 52 Cr A KED Ge 0.94 1.57 1.55 8.74 13.1 10.9 63 Cu B KED Ge 0.54 0.82 0.61 1.34 2.00 1.39 56 Fe B KED Sc 303 454 322 302 452 317 55 Mn A KED Ge 6.41 9.61 6.72 17.1 25.7 17.4 78 Se A KED Rh 61.4 92.1 81.5 135 203 170 66 Zn B KED Y 3.75 5.62 5.41 6.57 9.85 8.95

Upload: others

Post on 04-Apr-2020

20 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Shimadzu ICPMS-2030 Trace Elements in Clinical Samples...Clinical Samples using Quadrupole ICP-MS . Jan Knoop1, Nils Garnebode2, Konstantin Kartaschew2, Uwe Oppermann1, Ludivine Fromentoux3

But still as Argon is one of the major element presentwithin the mass spectrometer, the Argon-Argonpolyatomic interference needs to be considered,especially when the major components of the sampleis changing. In this case for urine, serum and wholeblood, the changing amount of Carbon gives somefurther difficulties. The reason is, that carbon formspolyatomic ions with Argon as well. So as morecarbon is present, as more Carbon-Argon-Ions areformed, thus it influences the equilibrium of theformation of Argon-Argon-Ions and affects thedetermination of 78Se.

At the same time with the increasing amount ofCarbon, other elements like 52Cr are interfered by40Ar12C, which gives the idea that easy matrixmatching by adding a carbon source to the internalstandard could solve these troubles without the needfor applying reaction gases which would saveadditional measurement time.

■ General Method Parameters

The ICPMS-2030 has been used in basic Minitorchsetup, enabling drastically lower flow rates of Argon.Even for high matrix samples the total consumptionis below 10 l/min. At the same time the Argon purityrequirements for the ICPMS-2030 are on a very lowlevel (min. Argon 3.5 = 99.95 %). The basic methodparameters are summarized in table 1.

Tab. 1: ICPMS-2030 Configuration and Method Parameters

■ Sample Measurement - Urine

For Urine, the preparation of samples is very easy.All samples have been diluted by a factor of 10 usingacidified purified water (1 Vol-% HNO3). As theinternal standard solution is added automatically, thesampler just have to be placed to the sampler. Urinetypical consist of 95% water. After applying factor 10dilution the amount of TDS cannot exceed 0,5%.Thus, the determination of trace elements usingICPMS technique is not most challenging.For evaluation of the method and instrumentcapabilities, certified reference material has beenanalysed (Table 2). The result mentioned is anaverage of 2 single measurements, whereat bothsingle results meet the certified range.

■ Sample Measurement – Serum and Whole blood

For the measurement of these two types of sample itnow starts that the influence of Carbon is increasing.The samples are more and more different comparedto the reference calibration solutions, which do notcontaining any carbon. For example Serum containsappr. 91% water and about 7 % proteins (albuminand globulins).

Easy and Fast Determination of Trace Elements inClinical Samples using Quadrupole ICP-MS Jan Knoop1, Nils Garnebode2, Konstantin Kartaschew2, Uwe Oppermann1, Ludivine Fromentoux3

1Shimadzu Europa GmbH, Germany; 2Shimadzu Deutschland GmbH, Germany; 3Shimadzu France SAS, France

■ Overview and Objective

The analysis of trace elements in clinical samples isa frequent required objective nowadays andoftentimes connected to the use of ICPMStechniques.

The first question then is, which technique exactlycan meet the user requirements as differentmeasurement principle are available withincommercial quadrupole ICPMS devices to overcomeinterferences and achieving correct results. Inaddition to standard measurement mode thesecommonly are kinetic energy discrimination (KED)using Helium as collision gas or special reactionmodes with gases like Hydrogen or Ammonia.

Whereat the kinetic energy discrimination mode isvery easy to handle, as it affects the resulting massspectra in basically the same way and no accounthas to be taken to specific behaviours to specificisotopes, the reaction modes with each different gasalways needs to be improved for different isotopesindividually in order to reduce interferences.

For each mode mentioned above, the ICPMScollision/reaction cell needs to be stabilized. Thus, asmore and more different modes are applied, themeasurement times are drastically extended.

Within this poster the ease of uses of matrixmatching is shown with the goal to use as lessmeasurement modes as possible in order to saveanalysis time.

Figure 1: Inductively Coupled Plasma Mass Spectrometer ICPMS-2030 (Shimadzu)

■ Clinical Samples and Typical Interferences

For the determination of trace elements in urine,serum and whole blood, different measures must beconsidered to achieve correct results.While it is quite easy to determine trace elements inurine after dilution using acid media (x10), the sameis getting more challenging for serum or whole bloodmatrix, due to the higher amount of carbon andsubsequent appearing interferences for example forthe determination of selenium.

In general for quantification of trace elements themost abundant isotope is selected. Taking Seleniuminto consideration, this would be 80Se (49.61%). Butat the same time, one of the most frequentpolyatomic interferences affect m/z 80, in this case itis the polyatomic ion of 40Ar40Ar from the Argon usedas plasma gas. That is why the 2nd most abundantisotope, 78Se is most commonly selected (23.78 %).Here polyatomic ions of Argon interfere as well(38Ar40Ar). But as the 38Ar is much less abundantthan the 40Ar, the level of interference to 78 m/z isappr. factor 1500 less than to 80 m/z. As a result, thesignal to noise ratio of 78Se is several times betterleading to higher sensitivities compared to 80Se,even 78Se is less sensitive.

____________________[1] Nelms, Simon M., Inductively Coupled Plasma Mass Spectrometry Handbook, CRC Press LLC, 2005

Shimadzu Europa GmbHAlbert-Hahn-Str. 6-10

D-47269 Duisburg+49(0)203/7687-0www.shimadzu.eu

[email protected]

Tab. 2: Urine quantification results (all in µg/l)

Tab. 3: Serum quantification results (all in µg/l)

Tab. 4: Whole Blood quantific. results (Aµg/l Bmg/l)

A similar concentration of water (90%) is the maincomponent of whole blood. Anyway – in accordanceto diagnostic sensitivity requirements, differentdilution factors have been applied. The serum wasdiluted similar to urine by a factor of 10 and thewhole blood by a factor of 20. For both dilutionsagain the applied media was 1 Vol-% nitric acid.

To cover additional interferences caused by carbon,the internal standard solution contains a certain levelof Ethanol and Triton X-100 in order to have a finalconcentration of 3 vol.-% Ethanol and 0,3 weight-%Triton X-100 present after mixing by T-piece.

■ Summary

As indicated by table 2-4 all results meet the certifiedrange of the reference material by just usingstandard and/or the very common KED mode. Anyadditional reaction gases have not been necessary.So the ICPMS-2030 is easy to use for themeasurement of clinical samples by e.g. adding acarbon source to the internal standard solution.

Parameter SettingRF generator power 1.2 kW

Plasma gas 8 l/minAuxilliary gas 1,1 l/min

Carrier gas 0.7 l/minNebulizer coaxial

Sampling depth 6 mmSpray Chamber temperature 5 °C

Collision Cell Gas flow (Helium) 6 ml/min

Cone material Copper

Internal Standard Automatic Addition (T-piece)

Sample tube Black-black, PVC

Internal Standard tube Orange-blue, PVC

Optional Fast Rinse Unit ASXPress Plus (Catac Teledyne)

Isotope Mode

Internal

Std.

Element

Recipe Level I Recipe Level II

Certified Result Certified Result

Min Max Average Min Max Average

52Cr KED Ge 3.26 4.89 4.17 15.9 23.8 19.965Cu KED Ge 29.4 44.1 38.7 73.5 110 91.3

55Mn KED Ge 3.13 4.69 3.93 15.5 23.2 19.560Ni standard Ge 4.73 7.10 5.91 34.4 51.7 41.675As standard Ge 34.4 51.6 42.7 66.6 99.9 79.8

111Cd standard Te 1.97 2.95 2.46 11.5 17.2 14.5206Pb KED Ho 19.2 28.8 20.2 52.0 78.0 54.9203Tl KED Tb 5.79 8.69 6.40 15.2 22.8 17.378Se KED Rh 24.0 35.9 28.7 66.6 99.9 87.427Al standard Be 26.4 39.6 31.6 68.7 103 84.959Co standard Ge 1.63 2.44 1.94 27.8 41.7 32.268Zn KED Y 163 245 192 428 642 515127I KED Y 89.9 150 136 373 622 528

57Fe KED Ge 30.8 46.3 42.5 179 268 232

Isotope Mode

Internal

Std.

Element

Recipe Level I Recipe Level II

Certified Result Certified Result

Min Max Average Min Max Average

63Cu KED Ge 681 921 821 1140 1540 1410

195Pt KED Lu 7.11 10.7 10.1 68.8 103 917

78Se KED Rh 52.9 79.3 67.9 84 126 109

64Zn KED Y 1120 1520 1400 1730 2350 2280

Isotope Mode

Internal

Std.

Element

Recipe Level I Recipe Level III

Certified Result Certified Result

Min Max Average Min Max Average

52Cr A KED Ge 0.94 1.57 1.55 8.74 13.1 10.963Cu B KED Ge 0.54 0.82 0.61 1.34 2.00 1.3956Fe B KED Sc 303 454 322 302 452 317

55Mn A KED Ge 6.41 9.61 6.72 17.1 25.7 17.478Se A KED Rh 61.4 92.1 81.5 135 203 17066Zn B KED Y 3.75 5.62 5.41 6.57 9.85 8.95