Shielding for Diagnostic X-rays:UK Guidance
Jerry WilliamsRoyal Infirmary of EdinburghRoyal Infirmary of Edinburgh
![Page 2: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/2.jpg)
BIR/ IPEM Working Party ReportBIR/ IPEM Working Party Report
British Institute of Radiology
Institute of Physics and Engineering in Medicine
Working Party:
David Sutton Jerry Williams Colin Martin Don McIntosh Tony
Institute of Physics and Engineering in Medicine
David Sutton, Jerry Williams, Colin Martin, Don McIntosh, Tony Cotterill, Graham Hart, David Gallacher
Publication date: 2000
![Page 3: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/3.jpg)
BIR/ IPEM Working Party Report
Content:
BIR/ IPEM Working Party Report
• Design criteria and dose constraints
• Primary & Secondary radiations• Primary & Secondary radiations
• Building materials
• X-ray transmission factorsay t a s ss o acto s
• Assessment of shielding
• Worked examples
Working Party:
David Sutton Jerry Williams Colin Martin Don McIntosh Tony David Sutton, Jerry Williams, Colin Martin, Don McIntosh, Tony Cotterill, Graham Hart, David Gallacher
Publication date: 2000
![Page 4: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/4.jpg)
dRadiation Sources
• Radiography (film/ screen)• Radiography + Fluoroscopy• Radiography + Fluoroscopy• Angiography
CT ( i l li )• CT (single slice)• Mammography• Dental• DEXA• DEXA
![Page 5: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/5.jpg)
h lkThis talk
• Design criteria• Transmission factors• Transmission factors• Shielding materials
P i d S d R di ti• Primary and Secondary Radiations
![Page 6: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/6.jpg)
Design Criteria: Design Criteria: Dosimetric Quantities
• Dose limits–Effective dose, E mSvEffective dose, E mSv
• Dose monitoringOperational quantity–Operational quantity
–Personal dose equivalent, Hp(d) mSv
Shi ldi l l ti• Shielding calculations–Air kerma, K mGy
![Page 7: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/7.jpg)
Design Criteria
• Dose constraint (people)– Based on public dose limit (1 mSv/ year)
A li d ( ff bli )– Applied to everyone (staff + public)– 0.3 mGy/ year (6 µGy/ week)
• Occupancy factor • Occupancy factor – ≥5%– Dose to be < 6 mGy/ year– (Controlled area)
• Dose constraint (film or CR)/ d ( / k)– 0.4 µGy/ day (2 µGy/ week)
![Page 8: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/8.jpg)
fOccupancy factors
• Based on individual occupancy• Examples:p
Office 100%Reception area 100%Reception area 100%X-ray control room 100%Patient examination room 50%Corridor 20%Corridor 20%Toilets/ bathrooms 10%Outdoor area with seating 10%Store rooms 5%Store rooms 5%Unattended waiting rooms 5%
![Page 9: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/9.jpg)
fX-ray Transmission factors
• Adopted Archer’s empirical model1−
⎤⎡ ⎞⎛ γγα
αβ
αβ1 ⋅⋅
⎥⎦⎤
⎢⎣⎡ −⎟
⎠⎞⎜
⎝⎛ += teB
• Used Simpkin’s α, β, γ values• Added data for brick and barium Added data for brick and barium
plaster
Archer et al (1983) Health Physics, 44, 507.
Simpkin (1995) Health Physics, 68, 704
![Page 10: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/10.jpg)
h ld lShielding material
UK Standard lead thicknesses
Code t mm3 1.324 1.805 2.246 2.65
![Page 11: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/11.jpg)
Primary and Secondary Primary and Secondary Radiations
• Leakage radiationg• Primary/ transmitted beams• ScatterScatter
![Page 12: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/12.jpg)
X-ray Tube LeakageX ray Tube Leakage
Assumptions:•Tube leakage factor: 150 kV/ 3.3 mA
•Field size: 1000 cm2
•Distance: 1 m from tube/ patient
WP Recommendation:WP Recommendation:
•Leakage contribution may be ignored
![Page 13: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/13.jpg)
Primary & Transmitted Primary & Transmitted Radiation
• Fluoroscopy– Beam restricted to
image intensifier/ flat X-raybeam
image intensifier/ flat panel detector
– Detector has > 2mm lead equivalent
Patient
lead equivalent
• Mammography
Imagingdevice
![Page 14: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/14.jpg)
Radiography: Film dose method
• Dose to film/ imaging plate– 400 speed system– Dose ≤10 µGy
Dose ≤10 µGy
• Lead equivalence (100 kV)– Cassette
• 0 19 mm• 0.19 mm– Cassette + grid
• 0.26 mm– Cassette/grid/tableCassette – Cassette/grid/table
• 0.8 mm– Cassette/grid/chest Bucky
• 0 7 mm
Bucky system
• 0.7 mm
Lead equivalence data derived from Dixon (1994) Med Phys, 21, 1785
![Page 15: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/15.jpg)
Radiography: Film dose method
• Dose to film/ imaging plate
– 400 speed system– Dose ≤10 µGyDose ≤10 µGy
• Lead equivalence (100 kV)– Cassette
• 0 19 mm• 0.19 mm– Cassette + grid
• 0.26 mm– Cassette/grid/table– Cassette/grid/table
• 0.8 mm– Cassette/grid/chest Bucky
• 0 7 mm• 0.7 mm
![Page 16: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/16.jpg)
Attenuated primary: limiting HVLAttenuated primary: limiting HVL
Limiting HVL: = ln(2) / α1
Limiting HVL: ln(2) / α
0.1 to eDD ⋅α−=
0 2
0.3
VL m
m
0.01
Tran
smis
sion
0.1
0.2
Lim
iting
H
0.001
T
040 60 80 100 120
kV0.0001
0 0.5 1 1.5 2 2.5 3
Lead thickness mm
![Page 17: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/17.jpg)
Radiography: ESD methodRadiography: ESD method
• Unattenuated primary beam
p y– Beam outside patient
• Entrance Surface Dose (ESD)• Situations include
– Beam not collimated to cassette– Beam not directed to Bucky
• Examples– Chest radiography– Cross table radiography– Out of Bucky radiography (e.g.
extremities)
extremities)• Area of wall irradiated
– Not easy to predict– Variable– Variable
![Page 18: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/18.jpg)
lExamples
![Page 19: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/19.jpg)
Chest Radiography (film dose)Chest Radiography (film dose)
• ParametersParameters– 100 films/ week– 90 kV– ESD = 0.15 mGy– Film dose = 10 µGy– FFD = 1.8 m– FSD = 1.4 m
Fil ll 0 7 – Film-wall = 0.7 m
Best Radiographic Practice
![Page 20: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/20.jpg)
Calculation (Film-dose method)
• Attenuated kerma at wall per week:Film dose x Workload x ISL correction
mGy52.07.08.1
8.110010102
3 =⎥⎦⎤
⎢⎣⎡
+××× -
⎦⎣
• No of HVLs required (6 µGy/ week) = 6.5 • Limiting HVL of lead @ 90 kV = 0.23 mmLimiting HVL of lead @ 90 kV 0.23 mm• Total equivalent lead thickness = 1.5 mm• Additional lead shielding =
0 8 0.8 mm
![Page 21: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/21.jpg)
Chest Radiography (ESD)Chest Radiography (ESD)
• ParametersParameters– 100 films/ week– 90 kV– ESD = 0.15 mGy– Film dose = 10 µGy– FFD = 1.8 m– FSD = 1.4 m
Fil ll 0 7 – Film-wall = 0.7 m
![Page 22: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/22.jpg)
Calculation (ESD method)( )• Worst case assumption:
• Beam size greater than patient and Bucky system
• Unattenuated kerma at wall per week:ESD x Workload x ISL correction
mGy7.47.08.1
4.110015.02
=⎥⎦⎤
⎢⎣⎡
+××
• Maximum transmission permitted= 6x10-3/ 4 7= 6x10 3/ 4.7= 0.0013
• Lead shielding @ 90 kV1.4 mm
![Page 23: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/23.jpg)
Summary
• Best practice –Beam collimated to patient/ image Beam collimated to patient/ image
plate/ Bucky–0.8 mm lead
• Poor practice–Beam outside patient and BuckyBeam outside patient and Bucky–1.4 mm lead
![Page 24: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/24.jpg)
Scatter
![Page 25: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/25.jpg)
Scatter ModelNCRP 49 method
400FKaK us ⋅⋅=
Ks – scatter kerma; Ku – primary kermaF– Field size (cm2); a – scatter factorTrout and Kelly (1972), Radiology, 104, 161
Ku . F ⇒ Area-air kerma product (AKP) orArea-air kerma product (AKP) orDose area product (DAP)
DAPSK DAPSKs ⋅=
![Page 26: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/26.jpg)
hWhy DAP?
• No assumptions on field size• Availability of DAP data• Availability of DAP data
–Requirement in UK legislation to record patient doserecord patient dose
–DAP preferred patient dose metric (fluoroscopy and radiography)(fluoroscopy and radiography)
–National surveys of patient dose
![Page 27: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/27.jpg)
Scatter factor normalised to DAP (S)Scatter factor normalised to DAP (S)12
m2 )-1
125 kVp
8
10µG
y.(G
y.cm 100 kVp
85 kVp 70 kVp 50 kVp
4
6
r fac
tor,
S µ
0
2
0 30 60 90 120 150 180
Scat
ter
0 30 60 90 120 150 180
Angle of scatter
( ) ( )[ ]1f85kVdbS 234 θθθθ
JR Williams (1996) Br J Radiol, 69, 1032
( ) ( )[ ]1f85kVedcbaS 234 +−×+θ+θ+θ+θ=
![Page 28: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/28.jpg)
Scatter factor normalised to DAP (S)
12
-1
Scatter factor normalised to DAP (S)
8
10
Gy.
(Gy.
cm2 )-
125 kVp100 kVp 85 kVp 70 kVp 50 kVp
4
6
fact
or, S
µG
a -1.04E-07b 3 27E 05
0
2
0 30 60 90 120 150 180
Scat
ter f b 3.27E-05
c -2.75E-03d 8.37E-02e 1 58E+000 30 60 90 120 150 180
Angle of scatter
( ) ( )[ ]1f85kVdbS 234 θθθθ
e 1.58E+00f 5.99E-03
( ) ( )[ ]1f85kVedcbaS 234 +−×+θ+θ+θ+θ=
JR Williams (1996) Br J Radiol, 69, 1032
![Page 29: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/29.jpg)
Calculation vs measurement
Cli i l li tiMeasured/ Calc doseNo of DAP/ week Ave Calc Mean Min Max
Interventional Radiology (liver disease) 8 520 0.71 0.22 0.02 0.42Abdominal + lower limb angiograpy 8 1120 2.15 0.17 0.02 0.43
Clinical application positions Gy cm2 Dose mGy
Cardiac angiography 7 780 0.81 0.99 0.71 1.26Cerebral angiography 7 460 1.48 0.97 0.67 1.23Ba contrast studies (o/c tube) 9 430 0.66 0.74 0.50 1.13B t t t di ( / t b ) 10 960 1 62 0 28 0 05 0 86Ba contrast studies (u/c tube) 10 960 1.62 0.28 0.05 0.86General Radiography 10 460 0.96 0.64 0.30 1.06
![Page 30: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/30.jpg)
Validation of scatter data by Validation of scatter data by Monte-Carlo
Scatter fraction (S) @ 85 kV12
Measured
6
8
10
Gy
cm2
MeasuredMC
2
4
6
S µG
y/ G
00 30 60 90 120 150 180
Angle
![Page 31: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/31.jpg)
Comparison with NCRP 147
Scatter fraction (S) @ 85 kV12
UK
6
8
10
Gy
cm2 )-1
UK
NCRP 147
MonteCarlo
2
4
6
S µG
y (G
00 30 60 90 120 150 180
Angle
![Page 32: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/32.jpg)
Differences in scatter factors?
• Phantom• X-ray spectrum• X ray spectrum• Measurement methods
![Page 33: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/33.jpg)
Method in practicep12
m2 )-1
125 kVp
8
10µG
y.(G
y.cm 100 kVp
85 kVp 70 kVp 50 kVp
4
6
r fac
tor,
S µ
0
2
0 30 60 90 120 150 180
Scat
ter
( )
0 30 60 90 120 150 180
Angle of scatter
( ) ( )[ ]1f85kVedcbaS 234 +−×++++= θθθθ
![Page 34: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/34.jpg)
![Page 35: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/35.jpg)
![Page 36: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/36.jpg)
h d ( )Method in practice (1)
0 6
0.8
1.0
doseθ d
0 2
0.4
0.6
Rel
ativ
e
r
0.0
0.2
0 30 60 90 120 150 180
S tt i l
( )[ ] ( ) 12max cmGyµGy5.2kV031.0S
−+×=
Scattering angle
( )d = 1 m; θ = 117°
![Page 37: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/37.jpg)
![Page 38: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/38.jpg)
θ d = 1 m
∫∫ ⋅= θ
θθd
d)(SS15030
ave ∫ θd
![Page 39: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/39.jpg)
h d ( )Method in practice (2)
• Save over C-arm rotation – (30 to 150°)(30 to 150 )
• Smax for parallel beam
kV Smax Save
50 4.05 3.73
70 4.67 4.30
85 5.14 4.72
100 5 60 5 15100 5.60 5.15
125 6.38 5.86
![Page 40: Shielding for Diagnostic X-rays: UK Guidance - ITN for...Shielding for Diagnostic X-rays: UK Guidance ... • Lead shielding @ 90 kV 1.4 mm. Summary ... Calculation vs measurement](https://reader035.vdocuments.us/reader035/viewer/2022081800/5adc5f0f7f8b9ae1408b7cce/html5/thumbnails/40.jpg)
b dTo be continued…..
• Application examples–FluoroscopyFluoroscopy–Radiography
• Comparisons with NCRP 147• Comparisons with NCRP 147