selective uv-filter detection with sensors based on stainless steel electrodes modified with...
TRANSCRIPT
PAPER www.rsc.org/analyst | Analyst
Publ
ishe
d on
28
Apr
il 20
09. D
ownl
oade
d by
Uni
vers
ity o
f M
inne
sota
- T
win
Citi
es o
n 01
/10/
2013
07:
34:5
8.
View Article Online / Journal Homepage / Table of Contents for this issue
Selective UV-filter detection with sensors based on stainless steel electrodesmodified with polyaniline doped with metal tetrasulfonated phthalocyaninefilms
Luiz Fernando Moreira,a Marcos Roberto de Vasconcelos Lanza,a Auro Atsushi Tanakab
and Maria Del Pilar Taboada Sotomayor*c
Received 3rd February 2009, Accepted 17th April 2009
First published as an Advance Article on the web 28th April 2009
DOI: 10.1039/b902273b
This work describes the construction and application of two amperometric sensors for sensitive
UV-filter determination. The sensors were prepared using stainless steel electrodes in which polyaniline
(PANI) was electrochemically polymerized in the presence of nickel (NiPcTS) or iron (FePcTS)
tetrasulfonated phthalocyanines. The sensor surface characterizations were carried out using atomic
force microscopy (AFM). The PANI/NiPcTS sensor was selective for the chemical UV-filter
p-aminobenzoic acid (PABA) and the PANI/FePcTS sensor was selective for octyldimethyl-PABA
(ODP), both in a mixture of tetrahydrofuran (THF) and 0.1 mol L�1 H2SO4 at a volume ratio of 30 : 70,
and with an applied potential of 0.0 mV vs. Ag|AgCl. A detailed investigation of the selectivity was
carried out for both sensors, in order to determine their responses for ten different UV filters. Finally,
each sensor was successfully applied to PABA or ODP quantification in sunscreen formulations and
water from swimming pools.
Introduction
The growing body of information concerning the damaging
effects of UV radiation, the hole in the ozone layer, and
protection against UV radiation, has been received with great
public interest, due to concerns surrounding skin cancer and skin
ageing. Protection against UV radiation essentially requires
a UV-filter to filter out UVB rays (290–320 nm), which are
responsible for sunburn, or UVA rays (320–400 nm), which cause
skin ageing.
Chemical UV-filters are lipophilic compounds that absorb
dangerous UV light, thereby decreasing the amount of solar
radiation reaching the skin. The efficacy of a cosmetic formula-
tion can be estimated using the sun protection factor (SPF),
which depends on the UV-filter’s content of the formulation.
However, although UV-filters have beneficial effects for the skin,
a number of dermatological reactions and estrogenic effects have
been described.1–3 On the other hand, the recent increase in the
number of products containing UV-filters, the extension of the
spectrum of their usage from sunscreens to cosmetics, and their
applications as additives in plastics, carpets and washing
powders, has resulted in the possibility of aquatic contamination.
However, the toxicology and estrogenic effects of these
compounds within the ecosphere are hitherto unknown or
unpublished.
Various analytical methods for UV-filter quantification have
been reported in the literature, most of which are based on the
aUniversidade Sao Francisco, Braganca Paulista, SP, BrazilbCentro de Ciencias Exatas e Tecnologia, Universidade Federal doMaranhao, Sao Luıs, MA, BrazilcDepartamento de Quımica Analıtica, Universidade Estadual Paulista‘‘Julio de Mesquita Filho’’, 14801-970 Araraquara, SP, Brazil. E-mail:[email protected]; Tel: +55-16-33016620
This journal is ª The Royal Society of Chemistry 2009
HPLC technique.4–10 However, the procedures for isolation of
sunscreen agents from cosmetic matrices prior to chromato-
graphic analysis require several sample manipulations, which are
laborious and time-consuming, and therefore not suitable for
routine analyses in the cosmetic industry or for environmental
monitoring. Moreover, large volumes of hazardous solvents
must normally be handled using chromatographic techniques,
which may subsequently be released into the environment.
Additionally, there are no official methods of analysis for many
of these UV-filters, chemicals that include octyldimethyl-PABA
(ODP) and 4-methylbenzylidene camphor (4-MBC).
The need for reliable analytical methods to determine the
active ingredients of pharmaceuticals and cosmetics in the envi-
ronment has recently been highlighted.1–3,11–13 The identification
and development of standard analytical procedures for these new
environmental contaminants will be extremely important for
their successful quantification. It is not only necessary to identify
more reliable, rapid, sensitive and selective methodologies, but
also to develop analytical methods that do not require expensive
equipment, and which at the same time are portable. The
chemical sensor is an analytical tool that incorporates all of these
characteristics.
In the last two decades, conducting polymers have attracted
the attention of researchers from different fields of chemistry and
physics.14–28 Of these polymers, polyaniline (PANI) stands out
due to its interesting characteristics such as easy chemical and
electrochemical synthesis, high electrical conductivity, and high
stability.29–33 For these reasons, it has been widely employed in
sensor and biosensor construction,34–42 amongst other applica-
tions. The possibility of preparing PANI films doped with metal
tetrasulfonated phthalocyanines (MPcTS), which can be effi-
ciently used as counter-anions to neutralize the polymer back-
bone, has been suggested by some researchers for use in battery
Analyst, 2009, 134, 1453–1461 | 1453
Publ
ishe
d on
28
Apr
il 20
09. D
ownl
oade
d by
Uni
vers
ity o
f M
inne
sota
- T
win
Citi
es o
n 01
/10/
2013
07:
34:5
8.
View Article Online
cathodes and other important industrial applications.43–46 The
use of PANI/MPcTS films in sensor construction offers advan-
tages, due to the known catalytic characteristics of the metal
phthalocyanines and analogous compounds,45,47–71 and provides
an alternative and novel method for the selective and sensitive
quantification of important analytes.
To aid in the development of an alternative clean, rapid,
sensitive and selective methodology for UV-filter quantification,
the present work describes the development and application of
sensors constructed with stainless steel electrodes modified with
PANI/MPcTS films. For many of these materials, there is still no
official method of analysis, nor is there any regulation concern-
ing their maximum permitted concentrations in aquatic envi-
ronments.
Experimental
Reagents and solutions
All chemicals used were of analytical or HPLC grade when
necessary. Nickel(II) phthalocyanine-tetrasulfonic acid tetraso-
dium salt, iron(III) phthalocyanine-4,40,400,4%-tetrasulfonic
acid, 4-aminobenzoic acid (PABA), octyldimethyl-PABA (ODP)
[2-ethylhexyl-4-(dimethylamino)benzoate], 4-methylbenzylidene
camphor (4-MBC) [3-(4-methylbenzylidene)-bornan-2-one],
isoamyl p-methoxycinnamate (IMC) [isoamyl-4-(dimethylamino)-
benzoate], methyl anthranilate (MA) [methyl-2-aminobenzoate]
and 2-phenyl-5-benzimidazolesulfonic acid (PBS) were
acquired from Aldrich. Benzophenone-3 (BENZ-3) [2-hydroxy-
4-methoxybenzophenone], octyl-methoxycinnamate (OMC)
[2-ethylhexyl-4-methoxycinnamate], octyl salicylate (OSA)
[2-ethylhexyl salicylate] and avobenzone (AVB) [1-(4-methoxy-
phenyl)-3-(4-tert-butylphenyl)-1,3-propanedione] were pur-
chased from Fluka. Aniline was obtained from Acros Organics�,
and was vacuum distilled before use. Tetrahydrofuran (THF),
acetonitrile (MeCN), and sulfuric acid were obtained from Synth-
Brazil.
Buffer solutions were prepared using water purified in a Mil-
lipore Milli-Q system, and the pH was determined using a Met-
rohm� Model 781 pH/Ion meter.
The ODP and PABA standard solutions for the electroana-
lytical experiments were prepared in THF and 25 mmol L�1
H2SO4, respectively. For the HPLC experiments, these standard
solutions were prepared in the respective mobile phases.
Sensor construction
Purpose-built 304 stainless steel electrodes with a geometrical
area of 0.502 cm2 were used for sensor construction, since
stainless steel is a low cost electrodic material. Initially, the
surfaces of the electrodes were cleaned as previously described in
the literature,72–74 with some modifications. Briefly, the surface of
the stainless steel electrode was manually polished with 0.5 and
0.3 mm alumina suspensions, then washed with distilled water,
ultrasonicated for 30 s in deionized water, followed by a further
30 s ultrasonification in ethanol to eliminate any alumina parti-
cles from the surface. Verification of the electrode surface was
carried out by means of cyclic voltammetry in 0.1 mol L�1 H2SO4
under anaerobic conditions. A characteristic and reproducible
profile was obtained for the clean stainless steel electrode.
1454 | Analyst, 2009, 134, 1453–1461
After cleaning, the electrodes were modified following litera-
ture procedures43,75 with PANI, PANI/FePcTS or PANI/NiPcTS
films. For the PANI/FePcTS sensor, 10.0 mL of deoxygenated
0.5 mol L�1 aqueous sulfuric acid solution, containing 0.12 mol
L�1 aniline or aniline plus 5.0 � 10�3 mol L�1 MPcTS, was used
for the electrochemical synthesis of polyaniline and polyaniline
doped with the respective metal tetrasulfonated phthalocyanine
(MPcTS). The electropolymerization was performed potentio-
dynamically in the potential range of �0.2 to 1.6 V vs. Ag|AgCl,
at a scan rate of 50 mV s�1, over 10 cycles for PANI/FePcTS and
15 cycles for PANI/NiPcTS. The difference in the cycles used for
the sensors’ construction can be explained considering their
performance in the detection of the UV-filters. In addition to the
optimization of the film synthesis cycles, the aniline and MPcTS
concentrations, as well as the potential range for the electro-
synthesis, were also optimized to achieve the best performance
for each sensor in their respective UV-filter determinations. For
comparison, two stainless steel electrodes coated with PANI
films were prepared using 10 and 15 voltammetric cycles.
Sensor surface characterization
The AFM images were collected in contact mode using a Nano-
scope IIIa Multimode Microscope (Digital Instruments�, USA).
Film thickness estimation was carried out in a surface profiler
(Model Dektak 3, Veeco/Sloan Instruments Inc.).
Electrochemical measurements
The electrochemical experiments were performed using
a PGSTAT 30 potentiostat (EcoChemie–Autolab�), interfaced
with a personal computer for data acquisition and potential
control. All electrochemical measurements were carried out in
a conventional three-electrode cell at room temperature, with an
Ag|AgCl electrode as the reference, and a platinum wire as the
counter electrode. The sensors were the working electrodes.
All cyclic voltammetry and amperometric measurements for
ODP and PABA determination were carried out in 7.0 mL of
a mixture of 0.1 mol L�1 H2SO4 and THF in a volume ratio of 70
: 30 (the electrolyte solution). This mixture was previously opti-
mized to yield the best results. For the amperometric measure-
ments, the current obtained at 0.0 mV vs Ag|AgCl (previously
optimized) was, initially, continuously monitored until a steady
state was reached (after approximately three minutes), to obtain
the baseline. After that, 50 mL of 2.5 � 10�2 mol L�1 UV-filter
standard solution was added to the electrolyte, and the mixture
was stirred for 10 seconds to homogenize the solution. The
current was then monitored in the quiescent solution for 10 s,
until the current reached a new steady state. Successive additions
of the respective UV-filter standard solutions were made, in
order to obtain the analytical curve.
HPLC analysis
The chromatographic analyses were performed using a Shi-
madzu� Model 20A liquid chromatograph fitted with an UV/Vis
detector (SPD-20A), an autosampler (SIL–20A) and a degasser
(DGU–20A5), and coupled to a personal computer. A C18
column (250 � 4.6 mm, Shim–Pack CLC–ODS) was positioned
This journal is ª The Royal Society of Chemistry 2009
Publ
ishe
d on
28
Apr
il 20
09. D
ownl
oade
d by
Uni
vers
ity o
f M
inne
sota
- T
win
Citi
es o
n 01
/10/
2013
07:
34:5
8.
View Article Online
inside an oven (Shimadzu� CTO–10AS) to maintain a constant
temperature.
For the ODP quantification, the method proposed by
Ribeiro76 was used, due to the lack of any official analytical
method. For this, the mobile phase was prepared by mixing
acetonitrile and water (pH ¼ 2.7) with H3PO4 in a ratio of 93 : 7
(v/v). The flow rate was 1.0 mL min�1 and the sample injection
volume was 20 mL. The column temperature was maintained at
30 �C. The measurement wavelength was 309 nm and the analysis
time was 11 minutes for each standard and sample.
For the PABA quantification, the method proposed by
Richter et al. was used.77 The mobile phase was phosphate buffer
(0.0075 mol L�1, pH 7.3)/acetonitrile in a ratio of 90 : 10 (v/v).
The flow rate was 0.8 mL min�1 and the sample injection volume
was 50 mL. The column temperature was maintained at 40 �C.
The measurement wavelength was 280 nm and the analysis time
was 6 minutes for each standard and sample.
Sunscreen analysis
Sunscreens based on two different compositions containing ODP
(8% (w/w), declared value) or PABA (15% (w/w), declared value)
were tested. Anionic gel-cream samples were dissolved in the
same solvents in which the standard solutions of ODP and
PABA were prepared, while for the HPLC quantification they
were dissolved in their respective mobile phases. Non-ionic
samples were prepared for sensor application in the same way as
the anionic samples, and were then filtered in order to separate
out the remainder of the solid. For the HPLC application, non-
ionic samples were pre-treated following a procedure established
by ANVISA (Agencia Nacional de Vigilancia Sanit�aria do Bra-
zil-National Agency of Sanitary Monitoring).78 For this, 2.0 g of
the gel-cream containing ODP or PABA was mixed with 4.0 mL
of methanol and 250 mL of 1.0 mmol L�1 H2SO4, the mixture was
then warmed to 40 �C for 10 minutes, and finally ultrasonicated
for 3 minutes. After this pre-treatment, the solution was filtered
and injected into the chromatograph for analysis.
Aquatic sample analysis
Two water samples from private swimming pools were enriched
with ODP or PABA and then analyzed using the proposed
sensors in order to evaluate the matrix effect, using the recovery
values.
In addition, solid phase extractions (SPE) were optimized for
ODP and PABA recovery, using these aquatic samples, for later
application with the sensors using the TOC (total organic
carbon) technique. The TOC technique was preferred to the
HPLC methodology because the sample compositions were well
known, so a more rapid analysis could be carried out without the
generation of residues or the consumption of reagents, making
this a ‘‘clean’’ procedure (although it is not selective for the
monitoring of the optimization of the SPE technique). The TOC
equipment used was a Shimadzu Model TOC-VCP analyzer.
To obtain the best conditions for the analyte extraction using
the SPE technique, the cartridge type and the pre-concentration
flow were previously optimized for each of the analytes. For this,
the organic matter in the aqueous sample containing a known
amount of UV-filter was analyzed before and after having passed
This journal is ª The Royal Society of Chemistry 2009
through the cartridge. The extent of the removal of organic
material (corresponding to the analyte in the aquatic sample)
after passage through the cartridge indicated the efficiency of the
extraction procedures. The flow rate allowing total retention
(extraction) in the cartridges was 1 mL min�1 for both analytes.
The most efficient ODP retention was obtained using an ODS-
C18 cartridge, while a cyan-modified silica cartridge provided the
optimum retention for PABA. Both cartridges were obtained
from Agilent Technologies (AccuBOND�, 6 mL, 1000 mg, 50 mm
particle size).
The analyte pre-concentration steps used in this work were as
follows: (1) the cartridge was conditioned (50 mL methanol, 50
mL Milli-Q water), and the pre-concentration flow rate adjusted
(1 mL min�1); (2) the analyte was pre-concentrated (200.0 mL
sample); (3) interferents were removed (50 mL Milli-Q water); (4)
analytes were eluted. For the electroanalytical measurements, the
elution employed 10.0 mL of either THF (for ODP), or 0.025 mol
L�1 H2SO4 (for PABA). For the comparative method (HPLC),
the elution was carried out using the respective mobile phases for
each analyte; (5) the cartridge was cleaned (50 mL methanol).
Results and discussion
Stainless steel electrodes modified with PANI/MPcTS
Cyclic voltammetry was used for PANI and PANI/MPcTS film
preparation by the electro-polymerization of aniline in the
absence or presence of the corresponding metal tetrasulfonated
phthalocyanine. Fig. 1 shows the cyclic voltammetry profiles
recorded for the stainless steel electrode and the PANI and
PANI/MPcTS films on stainless steel electrodes, in a de-aerated
0.1 mol L�1 H2SO4 solution. These profiles were always repro-
ducible for the unmodified and for the PANI and PANI/MPcTS
modified electrodes. It can be observed that for the unmodified
stainless steel electrode, there was one oxidation peak at 1.20 V
vs. Ag|AgCl whose current decreased when PANI was deposited
on the surface, probably due to the presence of the non-
conductive pernigraniline form of the PANI,36 implying a lower
electroactivity of the film at this potential.79 In the presence of
PANI and PANI/MPcTS, three peaks were observed. For the
PANI/MPcTS film, the oxidation at 0.60 V appeared as a weak
shoulder, and another oxidation peak appeared at 1.2 V. At this
last potential, the current was higher for the PANI and the
stainless steel electrodes, meaning that the presence of MPcTS,
the PANI counter-anion, increased the conductivity of the films.
This phenomenon could be explained by an increase in protonic
transport during the redox processes compared with undoped
polyaniline films, as has been previously described by Damos
et al.43
It is worth emphasizing that the electro-polymerization of the
PANI/MPcTS was carried out in different potential ranges.
Some of these were: �0.2 to 0.8; �0.2 to 1.0; 0.0 to 1.0; �0.2 to
1.6 V vs. Ag|AgCl (always with a scan rate of 50 mV s�1). The
PANI/MPcTS electrosynthesis was carried out in solutions of 0.1
and 0.5 mol L�1 H2SO4, HClO4, HNO3 or HCl, and in mixtures
of concentrated H2SO4/dimethyl formamide (DMFA) and 0.1
mol L�1 H2SO4/DMFA in a volume ratio of 95 : 5. All different
conditions were based on procedures described in the literature.
Only the sensor prepared using a �0.2 to 1.6 V potential range in
Analyst, 2009, 134, 1453–1461 | 1455
Fig. 1 Cyclic voltammograms for unmodified stainless steel electrode
(dashed line); and for stainless steel electrode modified with PANI
(dotted line) or with PANI/MPcTS (solid line) in de-aerated 0.1 mol L�1
H2SO4 at a scan rate of 50 mV s�1. (a) PANI/FePcTS and (b) PANI/
NiPcTS.
Fig. 2 AFM images of: (a) 304 stainless steel surface; (b) stainless steel
modified with a PANI film; (c) stainless steel modified with a PANI/
FePcTS film; (d) stainless steel modified with a PANI/NiPcTS film.
Publ
ishe
d on
28
Apr
il 20
09. D
ownl
oade
d by
Uni
vers
ity o
f M
inne
sota
- T
win
Citi
es o
n 01
/10/
2013
07:
34:5
8.
View Article Online
0.5 mol L�1 aqueous sulfuric acid solution gave promising results
for UV-filter detection, so that these conditions were selected for
further experiments. For the other conditions, the sensors did not
show any variation in the cyclic voltammograms or the chro-
noamperometry signals for any of the UV-filters studied.
Sensor surface characterization
AFM images were used to obtain a visual understanding of the
modification of the stainless steel electrodes with PANI and
PANI/MPcTS films. Fig. 2 shows the AFM images obtained for
(a) stainless steel, (b) PANI, (c) PANI/FePcTs and (d) PANI/
NiPcTS film surfaces. In these images, small imperfections on the
stainless steel surface can be observed, despite its previous pol-
ishing (Fig. 2a). However, Fig. 2b clearly shows the change in the
electrode surface following aniline polymerization, with the
apparent formation of polymeric networks covering the stainless
steel surface.
In the third and fourth images (Fig. 2c and 2d), abrupt changes
in the morphology of the stainless steel surfaces can be observed.
In the case of the PANI/FePcTS electrode (Fig. 2c), a rough
surface appears due to the formation of FePcTS agglomerates.
In the case of the PANI/NiPcTS film (Fig. 2d), a more
1456 | Analyst, 2009, 134, 1453–1461
homogeneous distribution can be observed, compared to the
PANI/FePcTS film. Energy dispersive X-ray emission graphs
were used to determine the amount of the metal complexes
immobilized on the films, and indicated that in each film the mass
percentages were 73% and 82% for iron and nickel, respectively.
The PANI/FePcTS and PANI/NiPcTS films showed thick-
nesses of 10 mm and 20 mm, respectively. This difference in the
thicknesses obtained for the two films may be simply explained
by the different cycles used for the sensors construction, which
were based on the optimum performances observed for each
sensor. In addition, it can be seen from Fig. 3 that the greater
thickness of PANI/NiPcTS resulted in surfaces that are more
conductive.
Analytical characteristics of the sensors
The voltammetry profiles of the sensors in the electrolyte solu-
tion consisting of a mixture of H2SO4 and tetrahydrofuran
(THF), in a previously optimized volume ratio of 70 : 30, are
shown in Fig. 3. Profiles typical of doped conductive polymer
films in aqueous/organic media80 can be observed. On the other
hand, two well-defined peaks can be seen for the sensors in the
electrolyte solution. In addition, the anodic and cathodic
currents for PANI/NiPcTS were higher than those for the PANI/
FePcTS sensor, as was suggested by the energy dispersive X-ray
and thickness studies.
The experiments showed that in 7.0 mL of the 0.1 mol L�1
H2SO4/THF 70 : 30 (v/v) solution, at least ten successive addi-
tions of 500 mL of the UV-filter standard solution (the majority
of which was soluble in organic media) could be carried out
without phase separation. Hence, this solution was chosen for
use in all analyses with the proposed sensors.
In a first stage, cyclic voltammetry experiments used the
stainless steel electrode and the PANI sensor in order to explore
the electrochemical behavior of the UV-filters. No changes in the
voltammogram profiles (data not shown) were observed for the
compounds evaluated using these two electrodes.
This journal is ª The Royal Society of Chemistry 2009
Fig. 4 Relative response (%) obtained with PANI/FePcTS (a) and
PANI/NiPcTS (b) sensors for ten different UV-filters (ODP, PABA,
OMC, IMC, AVB, 4-MBC, PBS, BENZ-3, MA and OSA). The
parameter was calculated considering the sensor response of PANI/
FePcTS for ODP or PANI/NiPcTS for PABA as 100%, respectively.
Fig. 3 Cyclic voltammograms for stainless steel electrode modified with
PANI (dotted line) or with PANI/MPcTS (solid line) in the electrolyte
solution (0.1 mol L�1 H2SO4/THF 70 : 30 (v/v) ratio) at a scan rate of 50
mV s�1. (a) PANI/FePcTS and (b) PANI/NiPcTS.
Publ
ishe
d on
28
Apr
il 20
09. D
ownl
oade
d by
Uni
vers
ity o
f M
inne
sota
- T
win
Citi
es o
n 01
/10/
2013
07:
34:5
8.
View Article Online
Next, in order to investigate for which of the UV-filters the
sensors gave better results, cyclic voltammetry was performed
using the electrolyte solution in the presence of 1.8 � 10�4 mol
L�1 of each of the ten UV-filters studied for the PANI/FePcTS
and PANI/NiPcTS sensors. The results obtained in these
experiments are shown in Fig. 4 as the response (as current
variation), at the adequate potential of 0.0 V vs. Ag|AgCl, of each
proposed sensor for the ten compounds evaluated. This potential
was used in these experiments because it is a potential free of
other electrochemical interferences. The PANI/FePcTS sensor
showed a higher response for ODP and the PANI/NiPcTS sensor
showed a higher response for PABA. Therefore, each of the UV-
filters was analyzed by the corresponding PANI/MPcTS sensor.
Fig. 5 and 6 show the voltammetric and chronoamperometric
profiles and their corresponding analytical curves, obtained at
0.0 V vs. Ag|AgCl, for ODP and PABA, respectively. This
applied potential was previously optimized (data not shown) for
each analyte and sensor. It is important to emphasize that, in the
sensors field, potentials close to 0.0 V are preferred in order to
minimize possible interferences from other electrochemically
active compounds.
All of the analytical parameters obtained for the ODP and
PABA quantifications are summarized in Table 1. For the ODP
determination, the PANI/FePcTS sensor showed a linear range
between 1.8 and 10.5 � 10�4 mol L�1, a very high sensitivity and
low detection and quantification limits, which were calculated as
This journal is ª The Royal Society of Chemistry 2009
suggested by the IUPAC recommendations.81 The PANI/
NiPcTS sensor possessed a linear range for PABA between 0.2
and 1.4 � 10�3 mol L�1, a lower sensitivity and higher detection
and quantification limits than the PANI/FePcTS sensor.
Parameters such as response time, repeatability and repro-
ducibility were very similar and can be attributed to the film
preparation procedures, which were practically the same in each
case. The repeatability of the measurements was estimated by
plotting seven successive analytical curves in the corresponding
linear range for each analyte and sensor. In this case, the
repeatability was defined in terms of the relative standard devi-
ation (RSD) of the seven sensitivities obtained. The reproduc-
ibility of each PANI/MPcTS sensor was estimated as the RSD of
the sensitivities obtained using four different sensors constructed
on different days.
Application
Firstly, the proposed sensors were used for analyses of the UV-
filters in different sunscreen formulations (A, B, C and D). The
concentrations of the UV-filters were calculated using the
external calibration method, and the results compared to the
chromatographic method using the paired t-test, which showed
Analyst, 2009, 134, 1453–1461 | 1457
Fig. 6 Typical profile of the PANI/NiPcTS sensor for PABA under
optimized conditions. (a) Cyclic voltammetry plot. (b) Chronoampero-
metric measurements for successive additions of PABA. (c) Analytical
curve.
Fig. 5 Typical profile of the PANI/FePcTS sensor for ODP under
optimized conditions. (a) Cyclic voltammetry plot. (b) Chronoampero-
metric measurements for successive additions of the analyte. (c)
Analytical curve.
Publ
ishe
d on
28
Apr
il 20
09. D
ownl
oade
d by
Uni
vers
ity o
f M
inne
sota
- T
win
Citi
es o
n 01
/10/
2013
07:
34:5
8.
View Article Online
that the results were statistically the same at a confidence level of
95% (Table 2). It should be stressed that no sample pre-treatment
had to be carried out using the sensors for sunscreen analyses in
order to obtain results similar to those obtained by the HPLC
method. Both methods gave similar relative standard deviation
values, suggesting an excellent precision of the proposed sensors.
All these results suggest that the proposed methodology is a more
efficient and rapid alternative for sunscreen analysis.
In order to confirm that the sensors could be applied to aquatic
environments, ODP and PABA recovery studies were performed
using additions of known quantities of the analytes to samples of
1458 | Analyst, 2009, 134, 1453–1461
water from swimming pools (Table 3). Good recoveries were
obtained, with values close to 100% for both UV-filters, indi-
cating that the sensors can be successfully applied to the analysis
of these kinds of samples without influence from the matrix.
Finally, to increase sensitivity, the solid phase extraction (SPE)
method was optimized for each analyte using the swimming pool
water samples. The results obtained after SPE extraction were
compared with the HPLC method (Table 4). The results obtained
using the sensors were statistically the same, using the paired
t-test at a confidence level of 99%, as the results obtained by the
chromatographic method. Such increase in sensitivity may be
This journal is ª The Royal Society of Chemistry 2009
Table 1 Analytical parameters obtained for ODP and PABA determination using the proposed sensors based on the PANI/MPcTS films
Parameter
Sensor
PANI/FePcTS for ODP PANI/NiPcTS for PABA
Linear ranges/mol L�1 1.8 to 10.5 � 10�4 0.2 to 1.4 � 10�3
Sensitivity/mA L mol�1 24093 � 405 6487 � 189Correlation coefficients 0.997 0.9992Detection limits/mmol L�1 8.0 37.5Quantification limits/mmol L�1 25 125Response times/s 0.5 0.5Measurement repeatability (RSD, n ¼ 7) 1.7 2.1Sensor reproducibility (RSD, n ¼ 4) 3.2 2.9
Table 2 Determination of ODP and PABA in sunscreen samples using the sensors based on PANI/MPcTS films
Sensor Sample (composition)
UV-filter concentration in % (w/w)
Student t-valuescDeclared value Comparative method (HPLC)a Proposed sensorb
PANI/FePcTS for ODP A (non-ionic gel) 8 7.94 � 0.06 8.00 � 0.03 3.4B (anionic gel) 8 8.01 � 0.01 8.04 � 0.04 1.3
PANI/NiPcTS for PABA C (non-ionic gel) 15 14.5 � 0.5 14.9 � 0.3 2.3D (anionic gel) 15 14.9 � 0.1 15.0 � 0.3 0.6
a Standard deviation of two duplicates. b Standard deviation for three replicates. tcrit ¼ 4.30 for 2 degrees of freedom. c Statistical values at a confidencelevel of 95%.
Table 3 Recovery values in water from swimming pools determined by the proposed sensors for each of their corresponding analytes
Sensor Sample Added UV-filter/mol L�1 Found UV-filter/mg L�1 Recoverya (%)
PANI/FePcTS for ODP E 1.80 � 10�4 1.77 � 10�4 98.3 � 1.4PANI/NiPcTS for PABA F 7.30 � 10�4 7.18 � 10�4 98.3 � 0.4
a Standard deviation for three replicates.
Table 4 Determination of ODP and PABA in aquatic samples applying the optimized SPE method using the proposed sensors, and the comparativemethod
UV-filter Sample
UV-filter concentration/mg L�1 Student t-valuee
Real value Comparative method (HPLC)a Proposed sensorb HPLC Sensor
ODPc E 50.0 47.6 � 1.3 48.9 � 0.7 2.6 2.7PABAd F 100.0 98.4 � 0.14 98.3 � 0.4 16.2 7.4
a Standard deviation of two duplicates. b Standard deviation for three replicates. tcrit ¼ 63.7 and 9.92 for 1 and 2 degrees of freedom, respectively.c PANI/FePcTS sensor. d PANI/NiPcTS sensor. e Statistical values at a confidence level of 99%.
Publ
ishe
d on
28
Apr
il 20
09. D
ownl
oade
d by
Uni
vers
ity o
f M
inne
sota
- T
win
Citi
es o
n 01
/10/
2013
07:
34:5
8.
View Article Online
useful in order to extend the application of the sensors to a wider
range of environmental media (such as wastewater from treat-
ment plants, surface waters, or sewage effluents).
Conclusions
This work describes an alternative and novel methodology for
the determination of chemical UV-filters based on two highly
sensitive and selective amperometric sensors constructed using
stainless steel electrodes electrochemically modified with PANI/
MPhTS. The sensors showed high sensitivity and selectivity for
the determination of PABA and ODP, and were satisfactorily
applied to the analysis of sunscreens in water from swimming
This journal is ª The Royal Society of Chemistry 2009
pools, thus, opening up the possibility of the quantification of
these chemicals in other aquatic environments. The use of
stainless steel electrodes modified with PANI/FePcTS or PANI/
NiPcTS films is a promising alternative procedure which is clean,
rapid, sensitive and robust, for UV-filter determinations in
sunscreen formulations and aquatic samples.
Acknowledgements
The authors gratefully acknowledge financial support from
CNPq (Proc. 470025/2006-9), ABN/Banco Real and FAPESP.
LFM is indebted to FAPESP for a fellowship. We thank Prof. Dr
Gilberto Medeiros Ribeiro and Mr Vinicius do Lago
Analyst, 2009, 134, 1453–1461 | 1459
Publ
ishe
d on
28
Apr
il 20
09. D
ownl
oade
d by
Uni
vers
ity o
f M
inne
sota
- T
win
Citi
es o
n 01
/10/
2013
07:
34:5
8.
View Article Online
Pimentel (LNLS, Campinas, SP) for assistance in the use of the
AFM microscope.
References
1 C. Schelecht, H. Klammer, W. Wuttke and H. Jarry, Arch. Toxicol,2006, 80, 656.
2 M. Schlumpf, H. Jarry, W. Wuttke, R. Ma and W. Lichtensteiger,Toxicology, 2004, 199, 109.
3 H. Klammer, C. Schelecht, W. Wuttke and H. Jarry, Toxicology,2005, 215, 90.
4 G. Potard, C. Laugel, A. Baillet, H. Schaefer and J. P. Marty, Int.J. Pharm, 1999, 189, 249.
5 S. Scalia, J. Chromatogr., 2000, A 870, 199.6 A. Chisvert and M. C. Pascual-Martı, Fresenius’ J. Anal. Chem., 2001,
369, 638.7 A. Townshend, R. A. Wheatley, A. Chisvert and A. Salvador, Anal.
Chim. Acta, 2002, 462, 209.8 S. C. Rastogi, Contact Dermatitis, 2002, 46, 348.9 D. J. Schakel, D. Kalsbeek and K. Boer, J. Chromatogr., 2004, A
1049, 127.10 G. Lunn, HPLC – Methods for Pharmaceutical Analysis, J. Wiley,
New York, 2000, vol. 1, p. 306.11 G. J. Smith and I. J. Miller, J. Photochem. Photobiol., 1998, A 118, 93.12 E. P. Santos, S. Garcia, Z. M. F. Freitas and A. L. Barth, Rev. Vis.
Acad., 2001, 2, 71.13 M. Balmer, H. R. Buser, M. D. Muller and T. Poiger, Environ. Sci.
Technol., 2005, 39, 953.14 M. Nikpour, H. Chaouk, A. Mau, D. J. Chung and G. Wallace,
Synth. Met., 1999, 99, 121.15 V. Misoska, W. Price, S. Ralph and G. Wallace, Synth. Met., 2001,
121, 1501.16 J. Ding, D. Zhou, G. Spinks, G. Wallace, S. Forsyth, M. Forsyth and
D. MacFarlane, Chem. Mater., 2003, 15, 2392.17 B. Winthe-Jensen, J. Chen, K. West and G. Wallace, Macromolecules,
2004, 37, 5930.18 J. Causley, S. Stitzel, S. Brady, D. Diamond and G. Wallace, Synth.
Met., 2005, 151, 60.19 B. Winthe-Jensen, J. Chen, K. West and G. Wallace, Polymer, 2005,
46, 4664.20 B. Winthe-Jensen, M. Forsyth, K. West, J. W. Andreasen, G. Wallace
and D. R. MacFarlane, Org. Electron., 2007, 8, 796.21 M. C. Gallazzi, L. Tassoni, C. Bertarelli, G. Pioggia, F. Di Francesco
and E. Montoneri, Sens. Actuators, 2003, B 88, 178.22 A. Fort, S. Rocchi, M. B. Serrano-Santos, N. Ulivieri, V. Vignoli,
G. Pioggia and F. Di Francesco, Sens. Actuators, 2005, B 111–112,193.
23 A. Arena, N. Donato, G. Pioggia, G. Rizzo and G. Saitta,Microelectron. J., 2006, 37, 1384.
24 G. Pioggia, F. Di Francesco, A. Marchetti, M. Ferro, R. Leardi andA. Ahluwalia, Biosens. Bioelectron., 2007, 22, 2624.
25 G. Pioggia, F. Di Francesco, A. Marchetti, M. Ferro andA. Ahluwalia, Biosens. Bioelectron., 2007, 22, 2618.
26 A. Arena, N. Donato, G. Saitta, G. Rizzo, G. Neri and G. Pioggia,J. Sol-Gel Sci. Technol., 2007, 43, 41.
27 A. Arena, N. Donato, G. Saitta, G. Pioggia and G. Rizzo, Solid-StateElectron., 2007, 51, 639.
28 A. Mazzoldi, M. Tesconi, A. Tognetti, W. Rocchia, G. Vozzi,G. Pioggia, A. Ahluwalia and D. De Rossi, Mater. Sci. Eng., 2008,C 28, 1057.
29 P. T. Sotomayor, I. M. Raimundo Jr., A. J. G. Zarbin,J. J. R. Rohwedder, G. O. Neto and O. L. Alves, Sens. Actuators,B, 2001, 74, 157–162.
30 E. M. Geni�es, A. Boyle, M. Lapkowski and C. Tsintavis, Synth. Met.,1990, 36, 139–182.
31 A. Watanabe, K. Mori, A. Iwabuchi, I. Iwasaki and Y. Nakamura,Macromolecules, 1989, 22, 3521–3525.
32 R. Bodalia, R. Stern, C. Batich and R. Duran, J. Polym. Sci., Part A:Polym. Chem., 1993, 31, 2123–2127.
33 L. H. C. Mattoso, Quim. Nova, 1996, 19, 388–398.34 S. C. Canobre, S. R. Biaggio, R. C. Rocha-Filho and N. Bocchi,
J. Braz. Chem. Soc., 2003, 14, 621–627.
1460 | Analyst, 2009, 134, 1453–1461
35 Y. Xiujuan, W. Jie and Z. Ding, Russ. J. Electrochem., 2003, 39,894–897.
36 P. Santhosh, K. M. Manesh, A. Gopalan and K. P. Lee, Anal. Chim.Acta, 2006, 575, 32–38.
37 X. Luo, A. J. Killard, A. Morrin and M. R. Smyth, Electrochim. Acta,2007, 52, 1865–1870.
38 Z. Wang, J. Yuan, M. Li, D. Han, Y. Zhang, Y. Shen, L. Niu andA. Ivaska, J. Electroanal. Chem., 2007, 599, 121–126.
39 K. Arora, N. Prabhakar, S. Chand and B. D. Malhotra, Biosens.Bioelectron., 2007, 23, 613–620.
40 J. Yuan, D. Han, Y. Zhang, Y. Shen, Z. Wang, Q. Zhang and L. Niu,J. Electroanal. Chem., 2007, 599, 127–135.
41 K. Arora, N. Prabhakar, S. Chand and B. D. Malhotra, Anal. Chem.,2007, 79, 6152–6158.
42 Z. Q. Zu, S. Jeedigunta and A. Kumar, J. Nanosci. Nanotechnol.,2007, 7, 2092–2095.
43 F. S. Damos, R. C. S. Luz, A. A. Tanaka and L. T. Kubota,J. Electroanal. Chem., 2006, 589, 70–81.
44 C. Coutanceau, A. E. Hourch, P. Crouigneau, J. M. L�eger andC. Lamy, Electrochim. Acta, 1995, 40, 2739–2748.
45 O. E. Mouahid, C. Coutanceau, E. M. Belgsir, P. Crouigneau,J. M. L�eger and C. Lamy, J. Electroanal. Chem., 1997, 426,117–123.
46 E. C. Venancio, A. J. Motheo, F. A. Amaral and N. Bocchi, J. PowerSources, 2001, 94, 36–39.
47 L. T. Kubota, Y. Gushikem, J. Perez and A. A. Tanaka, Langmuir,1995, 11, 1009–1013.
48 D. J. Dobson and S. Saini, Anal. Chem., 1997, 69, 3532–3538.49 S. S. Huang, H. Tang and B. F. Li, Mikrochim. Acta, 1998, 128,
37–42.50 E. F. Perez, L. T. Kubota, A. A. Tanaka and G. O. Neto, Electrochim.
Acta, 1998, 43, 1665–1673.51 J. Manriquez, J. L. Bravo, S. Gutierrez-Granados, S. S. Succar,
C. Bied-Charreton, A. A. Ordaz and F. Bedioui, Anal. Chim. Acta,1999, 378, 159–168.
52 J. A. P. Chaves, M. F. A. Ara�ujo, J. J. G. Varela Jr. and A. A. Tanaka,Ecl. Quim., 2003, 28, 9–20.
53 A. C. Serra, E. C. Marcalo and A. M. d’A. Rocha Gonsalves, J. Mol.Catal. A: Chem., 2004, 215, 17–21.
54 K. I. Ozoemena and T. Nyokong, Electrochim. Acta, 2006, 51,2669–2677.
55 W. J. R. Santos, A. L. Sousa, R. C. S. Luz, F. S. Damos,L. T. Kubota, A. A. Tanaka and S. M. C. N. Tanaka, Talanta,2006, 70, 588–594.
56 R. C. S. Luz, F. S. Damos, A. A. Tanaka and L. T. Kubota, Sens.Actuators, B, 2006, 114, 1019–1027.
57 B. Agboola, K. I. Ozoemena and T. Nyokong, Electrochim. Acta,2006, 51, 4379–4387.
58 T. Abe, K. Nagai, H. Ichinoche, T. Shibata, A. Tajiri andT. Norimatsu, J. Electroanal. Chem., 2007, 599, 65–71.
59 F. Matemadombo, P. Westbroek, T. Nyokong, K. I. Ozoemena,K. D. Clerck and P. Kiekens, Electrochim. Acta, 2007, 52, 2024–2031.
60 R. C. S. Luz, A. B. Moreira, F. S. Damos, A. A. Tanaka,L. T. Kubota and J. Pharm, Biomed. Anal., 2006, 42, 184–191.
61 C. Berrios, M. S. Ureta-Za~nartu and C. Gutierrez, Electrochim. Acta,2007, 53, 792–802.
62 R. Fogel, P. Mashazi, T. Nyokong and J. Limson, Biosens.Bioelectron., 2007, 23, 95.
63 B. Agboola and T. Nyokong, Electrochim. Acta, 2007, 52, 5039.64 P. N. Mashazi, K. I. Ozoemena and T. Nyokong, Electrochim. Acta,
2006, 52, 177.65 K. I. Ozoemena and T. Nyokong, Electrochim. Acta, 2006, 51, 2669.66 F. Matemadombo and T. Nyokong, Electrochim. Acta, 2007, 52,
6856.67 G. Mbambisa, P. Tau, E. Antunes and T. Nyokong, Polyhedron,
2007, 26, 5355.68 P. Tau and T. Nyokong, J. Electroanal. Chem., 2007, 611, 10.69 P. N. Mazashi, P. Westbroek, K. I. Ozoemena and T. Nyokong,
Electrochim. Acta, 2007, 53, 1858.70 N. Nombona, P. Tau, N. Sehlotho and T. Nyokong, Electrochim.
Acta, 2008, 53, 3139.71 N. Sehlotho, S. Griveau, N. Ruill�e, M. Boujtita, T. Nyokong and
F. Bedioui, Mater. Sci. Eng., 2008, C 28, 606.72 P. Calvo-Marzal, S. S. Rosatto, P. A. Granjeiro, H. Aoyama and
L. T. Kubota, Anal. Chim. Acta, 2001, 441, 207–214.
This journal is ª The Royal Society of Chemistry 2009
Publ
ishe
d on
28
Apr
il 20
09. D
ownl
oade
d by
Uni
vers
ity o
f M
inne
sota
- T
win
Citi
es o
n 01
/10/
2013
07:
34:5
8.
View Article Online
73 M. D. P. T. Sotomayor, A. A. Tanaka and L. T. Kubota,J. Electroanal. Chem., 2002, 536, 71–81.
74 M. D. P. T. Sotomayor, A. A. Tanaka and L. T. Kubota, Electrochim.Acta, 2003, 48, 855–865.
75 A. T. Ozyilmaz, M. Erbil and B. Yazici, Prog. Org. Coat., 2004, 51,47–54.
76 R. P. Ribeiro, Desenvolvimento e validacao de metodologia deanalise do teor de filtros solares em determinacao do FPS in vitroem formulacoes foto–protetoras comerciais, PhD Thesis, UFRJ,Rio de Janeiro, 2004.
This journal is ª The Royal Society of Chemistry 2009
77 K. Richter, R. Oertel and W. Kirch, J. Chromatogr., A, 1996, 729,293–296.
78 P. Barcellos, D. M. Bergmann and M. Medeiros, Guia de Controle deQualidade de Produtos Cosm�eticos: Uma Abordagem sobre os EnsaiosFısicos e Quımicos, ANVISA, Brasılia, 2007, p. 84.
79 T. C. Wen, Y. H. Chen and A. Golapan, Mater. Chem. Phys., 2002,77, 559–570.
80 A. G. B. Cruz, J. L. Wandell, R. A. Simao and A. M. Rocco,Electrochim. Acta, 2007, 52, 1899–1909.
81 A. Currie, Anal. Chim. Acta, 1999, 391, 105–126.
Analyst, 2009, 134, 1453–1461 | 1461