1/12/2019 1 SECTION 03 REFERENCE IMAGES MIKE ENRIQUEZ, MPA, BSRT(R)(CT) AUTHOR/PRESENTER THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) The Beam Shaping Filter (3.) Pre-patient Collimator (4.) PATIENT Pre-Detector Collimator (5.) Detector Array (6.) Analog-to-Digital Converter (7.) CT Computer, aka, Array Processor (8.) Digital-to-Analog Converter (9.) Monitor (10.) [email protected] 209-617-4468 The Tuffest Stuff CT Registry Review 2 1 2

Upload: others

Post on 27-Sep-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

1

SECTION 03REFERENCE IMAGES

MIKE ENRIQUEZ, MPA, BSRT(R)(CT)

AUTHOR/PRESENTER

THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION

MAJOR COMPONENTS

ofThe

CT IMAGING CHAIN

X-ray Tube (1.)

The Generator (2.)

The Beam Shaping Filter (3.)

Pre-patient Collimator (4.)

PATIENT

Pre-Detector Collimator (5.)

Detector Array (6.)

Analog-to-Digital Converter (7.)

CT Computer, aka, Array Processor (8.)

Digital-to-Analog Converter (9.)

Monitor (10.)

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

2

1

2

Page 2: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

2

Scale of Contrast

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

Long Scale, Wide Window Short Scale, Narrow Window

THE WIDER THE WINDOW, THE LONGER THE SCALE, THE GRAYER THE IMAGE;WIDE WINDOWS RANGE FROM 500 – 2000;

WIDE WINDOWS ARE GOOD FOR BONY DETAIL

THE NARROWER THE WINDOW, THE SHORTER THE SCALE, THE MORE BLACK & WHITE THE IMAGE;

NARROW WINDOWS RANGE FROM 0 - 499;NARROW WINDOWS ARE GOOD FOR SOFT TISSUE DETAIL

3

Short vs. Long Scale“The wider the window, the longer the scale”

“The narrower the window, the shorter the scale”

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

4

3

4

Page 3: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

3

Which area of the spectrum represents higher ENERGY?

Short Wavelength, High Frequency = High EnergyLong Wavelength, Low Frequency = Low Energy

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

FAMILY MEMBER CHARACTERISTICS• NO MASS• NO CHARGE• TRAVEL @ C

5

ACQUISITION GEOMETRIES

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

WITH HELICAL, INTERPOLATION OF DATA IS NECESSARY; ALSO, OVER-RANGING IS AN ISSUE

6

5

6

Page 4: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

4

BEAM GEOMETRIES

•Cone Beam• Fan beam + expanded Z axis as it

exits from the patient and covers the length & width of the detectors

• MDCT SYSTEMS: 3rd & 4th

generation CT Systems

•Area Beam• Like that used in diagnostic to

obtain a KUB

• Includes X, Y & Z axes in CT

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

•Pencil Beam• Like that of a laser pointer

• 1st generation CT Systems only

•Fan Beam• The inverted V of beam

that covers the patient as in the SFOV

• 2nd, 3rd & 4th generation CT System

7

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

8

7

8

Page 5: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

5

THE DATA ACQUISITION SYSTEM

• PRIMARY COMPONENTS

• THE CT DETECTOR SYSTEM• Convert EM energy to Electrical

energy

• Xenon gas

• Scintillation photodiode

• THE ANALOG TO DIGITAL CONVERTER (ADC)• Convert electrical impulses to

Binary Data

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

9

THE 6 STEPS OF RECONSTRUCTION

• FROM THE CT DETECTORS TO THE ADC TO THE CT COMPUTER…

1. PREPROCESSING

• Filter SCAN DATA = MEASUREMENT DATA and correct for data set errors due to beam hardening, bad detector reading, scatter radiation

2. REFORMATTED RAW DATA

• This is scan data that has been corrected by preprocessing

3. CONVOLUTION WITH FILTER

4. IMAGE RECONSTRUCTION ALGORITHM

• Back projection of convolved data

5. RECONSTRUCTED IMAGES OF CT NUMBERS

6. IMAGE STORAGE, DISPLAY, RECORDING, ARCHIVING

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

10

9

10

Page 6: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

6

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

11

Important Cerebral Anatomy

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

12

11

12

Page 7: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

7

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

13

1. Superior sagittal sinus2. Frontal bone3. Lateral ventricle, anterior horn4. Head of the caudate nucleus5. Putamen6. Third ventricle7. Choroid plexus8. Parietal bone9. Inion10. Cerebellar vermis11. Pineal body12. Thalamus13. Internal capsule

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

14

13

14

Page 8: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

8

ACQUISITION GEOMETRIES• 1st Generation

• Rotate-translate system

• Pencil beam scanning

• 2nd Generation

• Rotate-translate system

• Fan beam scanning

• 3rd Generation

• Rotate-rotate system

• Fan beam scanning

• 4th Generation

• Rotate only system

• Fan beam scanning

• 5th Generation• High speed CT scanners

• Obtain images in mS

• Electron beam CT Scanner

• Dynamic Spatial Reconstructor

• 6th Generation• Dual source CT scanner

• 2 tubes & 2 detector systems• Offset by 90 degrees

• Variable kVp

• 7th Generation• Flat panel CT Scanner

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

15

1. Maxillary bone2. Oral vestibule3. Masseter muscle4. Ramus, mandible5. Atlas6. Dens7. Spinal cord8. Internal jugular vein9. Mastoid tip10. Parotid gland11. Retromandibular vein12. Internal carotid artery13. pharynx14. Genioglossus muscle15. Vertebral artery16. Vertebra, spinous process17. Longus colli muscles18. Rectus/oblique capitus

muscle19. Splenius capitus muscle20. Pterygoid muscle21. External carotid artery

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

16

15

16

Page 9: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

9

Top of the Aortic Arch

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

1. Brachiocephalic Artery = Innominate Artery

A. Right Common Carotid ArteryB. Right Vertebral ArteryC. Right Subclavian Artery

2. Left Common Carotid Artery

3. Left Subclavian ArteryD. Left Vertebral

17

Circle of Willis

• ARTERIAL COMPOSITION• LEFT & RIGHT ANTERIOR CEREBRAL

ARTERIES

• ANTERIOR COMMUNICATING ARTERY

• LEFT & RIGHT INTERNAL CAROTID ARTERY

• POSTERIOR CEREBRAL ARTERY

• POSTERIOR COMMUNICATING ARTERY

• BASILAR ARTERY

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

18

17

18

Page 10: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

10

Circle of Willis

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

19

1. Frontal lobe2. Medial rectus muscle3. Superior rectus muscle4. Infraorbital fissure5. Nasal conchae6. Maxillary sinus7. Zygoma8. Maxillary bone9. Hard palate10. Mandible11. Tooth12. Nasal septum13. Inferior rectus muscle14. Lateral rectus muscle15. Optic nerve16. Sphenoid sinus17. Tongue18. Oral vestibule19. Masseter [email protected] 209-617-4468

The Tuffest Stuff CT Registry Review20

19

20

Page 11: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

11

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

21

CARDIAC ANATOMY & CIRCULATIONCORONARY ARTERY DETAILS1. Both the R & L main coronaries

arise from the base portion of the aorta

2. Both the R & L main coronaries lie on the epicardial surface & then branch

3. Venous return to the right atrium occurs via the coronary sinus

1. Important branches1. Great cardiac vein (Lt.

coronary vein)2. Middle cardiac vein3. Small cardiac vein

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

22

21

22

Page 12: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

12

1. Pectoralis major muscle2. Pectoralis minor muscle3. Ribs4. Aortic arch5. Vertebra6. Lung7. Pulmonary vessels8. Scapula9. Infraspinatus muscle10. Trachea11. Right brachiocephalic vein12. Left brachiocephalic vein13. Brachiocephalic artery14. esophagus

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

23

1. Sternum2. Pulmonary artery3. Ascending aorta4. Left atrium5. Descending aorta6. Vertebra7. Left inferior pulmonary vein8. Spinal cord9. Esophagus10. Right inferior pulmonary vein11. Superior vena cava12. Right atrial appendage13. lungs

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

24

23

24

Page 13: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

13

1. Right ventricle2. Aortic root3. Aortic valve4. Left anterior descending aorta5. Left ventricle6. Left atrium7. Descending aorta8. Left inferior pulmonary vein9. Spinal cord10. Vertebra11. Trapezius muscle12. Erector spinae muscle13. Azygos vein14. Esophagus15. Right inferior pulmonary vein16. Right lower lobe bronchus17. Right atrium18. Ribs19. Right coronary artery20. Left circumflex coronary artery

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

25

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

LEFT

26

25

26

Page 14: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

14

1. 1st cuneiform2. 2nd cuneiform3. 3rd cuneiform (lateral)4. Cuboid5. Navicular6. Calcaneus10. Achilles tendon

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

27

1. Extensor tendons2. Trapezium3. Scaphoid4. Capitate5. Hamate6. Triquetrum7. Pisiform8. Abductor &

opponensmuscles

9. Flexor tendons10.Radial artery11.Radial vein

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

28

27

28

Page 15: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

15

CERVICAL VERTEBRA (be able to label)

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

29

TYPICAL THORACIC VERTEBRA (be able to label)

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

30

29

30

Page 16: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

16

Typical LUMBAR (be able to label)

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

31

ABDOMINAL VASCULATURE

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

32

31

32

Page 17: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

17

Abdomen Anatomy Review

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

33

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

34

33

34

Page 18: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

18

• FATTY LIVER INFILTRATION

If the spleen measures 10 or more HU greater than the liver, a fatty liver is indicated; or,If the liver measures 10 or more HU less than the spleen, a fatty liver is indicated

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

35

Pancreatic Anatomy

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

36

35

36

Page 19: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

19

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

37

CORTICO-MEDULLARY NEPHROGRAM EXCRETORY

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

30-70 sec. POST-Bolus 100-120 sec. POST-Bolus 3-16 min. POST-Bolus

38

37

38

Page 20: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

20

1. Falciform ligament3. Magenblase4. Barium-filled stomach5. Spleen8. Left adrenal gland11. Aorta12. Inferior vena cava13. Portal vein14. Liver17. Transverse colon18. Splenic vein19. Cortex, left kidney20. Medulla, left kidney21. Vertebra22. Right adrenal gland23. Pancreas

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

39

2. Duodenum4. colon, splenic flexure5. Spleen9. Aorta10. Inferior vena cava12. Liver15. Rectus abdominus muscle17. Pancreatic head18. Portal vein confluence19. Jejunum20. Superior mesenteric artery21. Left renal vein22. Right kidney23. Psoas muscle24. Erector spinae muscle25. Gall bladder

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

40

39

40

Page 21: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

21

UPPER ABDOMINAL ARTERIES

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

CELIAC AXIS ARTERIES• SPLENIC ARTERY• COMMON HEPATIC ARTERY

SUPERIOR MESENTERIC ARTERY

RIGHT & LEFT RENAL ARTERIES

INFERIOR MESENTERIC ARTERY

41

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

Thoracic & Abdominal Veins

42

41

42

Page 22: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

22

AZYGOS & HEMI-AZYGOS VEINS

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

FACTS:1. Innominate = Brachiocephalic2. Azygos connects the IVC & SVC

1. Courses the right vertebral column2. Azygos origin at pre-renal portion

of the IVC3. Azygos drains into the SVC

3. Hemi-azygos:1. Courses the left vertebral column2. Communicates with the Azygos at

the level of T-74. Accessory Hemi-azygos

1. Receives venous return from:1. Oblique vein of Lt. Atrium2. Coronary Sinus

43

Hepatic-Portal System

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

H-P SYSTEM CONTRIBUTING VEINS

• SUPERIOR MESENTERIC VEIN• INFERIOR MESENTERIC VEIN• GASTRO-SPLENIC VEIN

44

43

44

Page 23: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

23

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

45

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

46

45

46

Page 24: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

24

1. Bladder2. Left external iliac artery3. Left external iliac vein4. Obturator internus muscle5. Seminal vesicles6. Rectum7. Coccyx8. Ischium9. Femoral head10. Pubis11. Right external iliac artery12. Right external iliac vein13. Rectus abdominus muscle14. Quadriceps femoris muscle15. Prostate gland16. Iliopsoas muscle17. Sartorius muscle18. Femoral head19. Gluteus maximus muscle20. Gluteus minimus muscle21. Gluteus medius muscle22. Tensor fascia lata muscle

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

47

1. Bladder2. Sigmoid colon5. Gluteus minimus muscle6. Gluteus medius muscle7. Ovaries8. Uterus w/ IUD9. Ilium10. Gluteus maximus muscle11. Sacrum12. Piriformis muscle13. Rectum15. Iliacus muscle16. Right external iliac artery18. Rectus abdominusmuscle

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

48

47

48

Page 25: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

25

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

49

SYSTOLE is DEMONSTRATED- the AV VALVES are CLOSED; the SEMILUNAR VALVES are OPEN.Also, “LUB” occurs when the AV Valves CLOSE. “DUB” occurs when the SEMILUNAR VALVES CLOSE.

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

50

49

50

Page 26: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

26

• Blood travels to and from heart via the great vessels

• KNOW THESE:• Aorta

• Pulmonary Trunk/Arteries

• Pulmonary Veins

• Superior Vena Cava

• Inferior Vena Cava

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

51

Let’s Study This…

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review 52

51

52

Page 27: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

27

Tachy-, Brady- & the ECG

• Tachycardia- ECG

• Resting heart rate of greater than or equal to 100 beats per minute

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review 53

Tachy-, Brady- & the ECG

• Bradycardia- ECG

• Resting heart rate of under 60 beats per minute

• Usually not symptomatic until 50 beats per minute or less

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review 54

53

54

Page 28: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

28

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

55

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

56

55

56

Page 29: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

29

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

57

Cardiac Gating

Prospective

• A signal from the R-wave triggers scanning at a set point in the R-R interval

• In this way only a portion of the cardiac cycle is used

• Low dose procedure comparatively

• Used for calcium scoring

Retrospective

• Continuous acquisition throughout the cardiac cycle with simultaneous recording of the ECG.

• Data can be reconstructed at any point in the R-R interval.

• Cine loops can be generated

• Higher radiation dose

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

58

57

58

Page 30: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

30

A Convolution “kernel”The averaging of 9 values to create 1 value

As applied to raw data

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

0.70

59

“PITCH” Exists because Slice “PITCH” is Real!

“PITCH” is the term that describes the relationship between tube rotation, table indexing and beam width at the detectors

As SLICE “PITCH” increases, the value of “PITCH” increases

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

60

59

60

Page 31: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

31

“PITCH” Exists because Slice “PITCH” is Real!

“PITCH” is the term that describes the relationship between tube rotation, table indexing and beam width at the detectors

As SLICE “PITCH” increases, the value of “PITCH” increases

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

61

HELIX Expansion, HELIX Contraction

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

AS THE HELIX EXPANDS, PITCH INCREASES

AS THE HELIX CONTRACTS,

PITCH DECREASES62

61

62

Page 32: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

32

“PITCH” relationships…

• INTERPRETING the diagram…• AS “P” increases:

• “PITCH” increases,

• dose decreases,

• slice pitch increases, • anatomy covered

per tube rotation increases; and,

• resolution decreases

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

63

“PITCH” relationships…

• INTERPRETING the diagrams…• AS “P” decreases:

• “PITCH” decreases,

• dose increases,

• slice pitch decreases, • anatomy covered

per tube rotation decreases; and,

• resolution increases

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

64

63

64

Page 33: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

33

“Pitch”, the Helix, Dose & Resolution

• As the helix expands, “pitch” Increases• Dose & Resolution

DEcrease

• As the helix contracts, “pitch” Decreases

• Dose & Resolution INcrease

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

65

CT specific patient dose measures• DLP takes into account the entire scanned volume and

tissue irradiated

• Can multiply by weighting factor to get effective dose

• DLP represents an improved method for estimating patient dose (e.g., effective dose)

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

66

65

66

Page 34: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

34

CT Specific MeasuresEstimating Effective Dose in mSvApplying k values for Adults

• Effective dose = DLP x k (where k is the weighting factor)• Representative adult weighting factors for k are:

»Head/Neck .0031»Head .0021»Neck .0059» Chest .014»Abdomen .015

–IF, DLP = 251 mGy-cm for a CHEST CT–THEN, Effective dose = DLP or 251 mGy-cm x 0.014 (k) = 3.5 mSv

PEDIATRIC CHART:AAPM Report No. 96These are k factor charts for kids aged 0 -1 -5 -10

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

67

Dose uniformity decreases as SFOV and patient thickness increases

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

100 rad = 1.0 Gy = 1,000 mGy10 rad = 0.10 Gy = 100 mGy1 rad = 0.01 Gy = 10 mGy1.2 rad = 0.012 Gy = 12 mGy3.1 rad = 0.031 Gy = 31 mGy3.3 rad = 0.033 Gy = 33 mGy3.5 rad = 0.035 Gy = 35 mGy4.0 rad = 0.040 Gy = 40 mGy4.1 rad = 0.041 Gy = 41 mGy4.4 rad = 0.044 Gy = 44 mGy4.5 rad = 0.045 Gy = 45 mGy5.0 rad = 0.050 Gy = 50 mGy

68

67

68

Page 35: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

35

From Achieving Ideal Computed Tomographic Scan Length in Patient With Suspected Urolithiasis

Overranging/Z-Axis Overscanning

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

69

Acquisition Parameter Settings

Detector Configuration

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

70

69

70

Page 36: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

36

Soft tissue ww=400, wl=40

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

Bone window ww=2000, wl=150Brain window ww=100, wl=35

Window Samples(Brain)

71

ARTIFACT REVIEW- PART 1

MOTION METALSTREAK

OUT-OF-FIELD

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

72

71

72

Page 37: SECTION 03 REFERENCE IMAGES · THE TUFFEST STUFF CT REGISTRY REVIEW SOLUTION MAJOR COMPONENTS of The CT IMAGING CHAIN X-ray Tube (1.) The Generator (2.) ... THE DATA ACQUISITION SYSTEM

1/12/2019

37

ARTIFACT REVIEW- PART 2

BEAM HARDENING

METALSTREAK

WINDMILL RING

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

73

Injection rate controls the time to peak enhancement

As injection rate increases, the time to peak enhancement decreases; and/or,

As injection rate decreases, the time to peak enhancement increases.

[email protected] 209-617-4468 The Tuffest Stuff CT Registry Review

74

73

74