Sample Analysis Design – Step 2 – Calibration/Standard Preparation Choice of calibration method dependent upon several factors: 1. potential matrix effects 2. number of samples 3. consistency of matrix across samples

Upload: ledieu

Post on 13-May-2018

220 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Step 2 – Calibration/Standard Preparation

• Choice of calibration method dependent upon several factors:

• 1. potential matrix effects

• 2. number of samples

• 3. consistency of matrix across samples

Page 2: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Step 2 – Calibration/Standard Preparation

• EXTERNAL CALIBRATION:• Prepare a set of standard solutions to cover the

expected range of analyte concentrations

• Fit a least squares regression line

• y = mx + b

and calculate analyte concentration in unknowns

Page 3: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Step 2 – Calibration/Standard Preparation

23Na calib curve (Medium resolution)

y = 17557xR2 = 0.9992

0

100000

200000

300000

400000

500000

600000

700000

800000

900000

1000000

0.00 10.00 20.00 30.00 40.00 50.00 60.00

conc ppb

cps

Page 4: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Step 2 – Calibration/Standard Preparation

44Ca calib curve (Medium resolution)

y = 676.92xR2 = 0.9961

0

5000

10000

15000

20000

25000

30000

35000

40000

0.00 10.00 20.00 30.00 40.00 50.00 60.00

conc ppb

cps

Page 5: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Step 2 – Calibration/Standard Preparation

• Advantages of External Calibration

– Easy to prepare

– Quick

– Widely used technique

Page 6: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Step 2 – Calibration/Standard Preparation• Disadvantages of External Calibration:

• Need to matrix match calibration solutions and samples

• If standards containing <2000 ug/ml (ppm) are being used, then preparing the standards as simple aqueous solutions using the acid matrix (5% HNO3) employed for the samples is sufficient

• HOWEVER, if the samples contain a very high concentration of one (or more) elements, then this may not be adequate

Page 7: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Step 2 – Calibration/Standard Preparation• Preparation of External Calibration Solutions:

• Need to evenly space calibration concentrations

• If the highest concentration is much higher than the rest, linear regression introduces bias favoring the high point

• X = independent variable = concentration

• Y = dependent variable = counts/second

Page 8: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Standard Addition Method

• Aliquots of spike are added to unknown samples to increase the ion signal intensities for elements of interest

• Typically use at least three aliquots of sample spiked with evenly spaced amounts of analyte

• These spiked aliquots of sample are used to generate a calibration line and calculate the concentration in the sample

Page 9: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Standard Addition Method

• S0 = unspiked sample

• S1 = sample spiked with analyte at concentration x

• S2 = sample spiked with analyte at concentration 2x

• S3 = sample spiked with analyte at concentration 3x

• S4 = so on and so on

Page 10: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Standard Addition Method

AMT

y = 29387x + 279235R2 = 0.9992

0

50000

100000

150000

200000

250000

300000

350000

400000

450000

500000

0 1 2 3 4 5 6 7

Concentration (ppb)

Cps

Page 11: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Standard Addition Method

• The concentration of the unknown solution is then determined by dividing the y-intercept value by the slope of the sample-spike mixing line.

– From example on previous slide,

– Conc. sol’n = 279235 / 29387 = 9.5 ppb

– If the original sample was a solution, then this is the concentration of the analyte in question in the solution

Page 12: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Standard Addition Method

• If the original sample was in solid form that you digested and subsequently converted into a solution;

• then in order to determine the concentration of the analyte in question, you must factor in the amount of total analyte in the solution and the dry weight of the sample powder

Page 13: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Standard Addition Method

• If we continue with the same example, the solution has a concentration of 9.5 ppb, and the original volume of the unknown solution was 10 ml (g) prior to aspirating some of it into the plasma for analysis, then the total amount analyte in the solution is:

• = 10 g x 9.5 ng/g (ppb)• = 95 ng, or• = 0.095 µg

• If the amount of powder weighed out was 0.1 g, then the concentration of the element in question is:

• Conc. = 0.095 µg/0.1 g– = 0.95 µg/g or ppm

Page 14: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Standard Addition Method

• This method works best if the slope of the calibration line is not too shallow

– This will create more uncertainty in the location of the intersection between the cps of your unknown and the calibration line

Page 15: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Standard Addition Method

• For maximum precision it’s necessary that the amount of sample be the same in each aliquot

• Also want the amount of spike added to be the same for each aliquot

• Amount of spike added should be as small as possible (usually 0.1 ml to 10 ml total volume)

Page 16: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Standard Addition Method

• Ideally, the highest spike concentration should be approximately equal to the concentration of analyte in the unknown

• Need to have some idea of the concentration in the sample prior to analysis

Page 17: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Standard Addition Method

• Advantages:

– Overcomes matrix differences– More precise and accurate than external calibration

• Disadvantages:

– Requires at least three aliquots for each sample– Run lengths become much longer and more

preparation time is required

Page 18: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Isotope Dilution

• Most accurate and precise calibration method available

• Requires analyte with two stable isotopes

• Monoisotopic elements cannot be determined via isotope dilution

• Spike natural sample with enriched isotope spike of analyte

Page 19: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:
Page 20: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:
Page 21: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Isotope Dilution

• The amount of spike is selected so that the resulting ratio between spiked isotope and unspiked isotope is near unity –maximizes precision

• Typically use the most abundant isotope as the reference -- maximizes sensitivity

Page 22: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Isotope Dilution

• Check isotope ratio in unspiked sample to determine if the “natural ratio” in the sample matches with the predicted ratio

• If not -- interference in acting on one or both of the isotopes

• Always attempt to use interference free isotopes

Page 23: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Isotope Dilution

• Prepare the spike to desired concentration

• Add spike as early as possible – after equilibration of spike and sample you don’t have to have complete sample recovery

• During any stage of the process complete equilibration is absolutely necessary

Page 24: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Isotope Dilution

• Analyze the solution on the ICP using many repetitive scans (to maximize precision)

• Need to measure isotopic ratios on standards of a known ratio in order to correct for machine mass discrimination

• Use previous equation to calculate concentrations!

Page 25: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Isotope Dilution

• Advantages:

– Most accurate and precise method for quantitative elemental concentrations

– Partial loss of analyte during preparation is compensated for since physical and chemical interferences are not an issue -- will cancel out as they will affect each isotope identically

– Ideal form of internal standardization since another isotope of the same element is used in this capacity

Page 26: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Isotope Dilution

• Disadvantages:

– Generally only applicable to multiple-isotopic elements

– Need an enriched isotope spike for the analyte of interest - not always available or sometimes at very high cost

– Need two interference free isotopes– VERY time consuming

Page 27: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design

STEP 3 – INTERNAL STANDARDIZATION & INSTRUMENT

DRIFT CORRECTION

Page 28: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Internal Standard

• Every sample should be analyzed with an internal standard (IS)

• What is an internal standard (IS)?

– element that is added to EVERY sample/ blank/calibration standard/QA sample/etc., that is not expected to be in the sample in appreciable quantities and is not an element of interest

– use IS to monitor machine drift (both short and long term) and matrix effects

Page 29: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Internal Standard

• Choice of IS depends upon which elements you are quantifying

• The IS should have similar properties in the plasma as element(s) of interest

• ICP-MS: similar in mass/ionization potential

Page 30: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Internal Standard

• Example:

– attempting to quantify U - use Th

– attempting to quantify most transition metals - use As

– attempting to quantify REEs - use Re

– 115In and 103Rh are common IS for general use

– alternatively, you can add several IS to each sample

Page 31: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Internal Standard

• From previous slide, we assume that samples have little or no Th, As, or Re

• It’s important to have an idea of what’s in your sample prior to quantitative analysis

• Solid samples can use a naturally occurring element as IS, provided that you know the concentration in each sample

Page 32: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Internal Standard

• Procedure for IS use:

• Calculate the concentration of the IS in each centrifuge tube – the latter will contain an aliquot of your sample and an aliquot of the IS

• Divide the measured ion signal (CPS) by the concentration of your IS to derive the factor = CPS/ppb

• Divide CPS/ppb of each tube by the CPS/ppb for those measured for the blanks since these are not influenced by possible effects due to sample matrices

• The latter yields a dimensionless correction factor (I refer to it as a normalization factor)

• Use correction factor to adjust analyte counts for drift or matrix effects

Page 33: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:
Page 34: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:
Page 35: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Internal Standard

• Advantages:

• Fluctuations are monitored in each sample/ calibration / blank

• Disadvantages:

• Assume that behavior of IS is the same as the analyte

Page 36: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Instrumental Drift

• Correct for instrument drift with:

• Internal standardization is a common procedure

• Use of drift corrector solutions (DCS)

Page 37: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Instrumental Drift

• Drift Corrector Solutions (DCS):

• Measure the same solution intermittently throughout the course of the analytical session

• Change in ion signal is assumed to be linear between each DCS measurement

Page 38: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Instrumental Drift

• The DCS should contain all elements of interest and can be matrix matched to samples

– Example: use standard reference materials (SRMs) for DCS

Page 39: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Instrumental Drift

• Apply a linear correction to samples between DCS solutions

• DCS1 + ((DCS2 - DCS1)*F)

• F = position dependent fraction

Page 40: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Instrumental Drift

• Advantages of DCS correction:

– all analytes are monitored for drift

– nothing added to sample solutions

• Disadvantages of DCS correction:

– assume change is linear

– cannot easily monitor matrix effects

Page 41: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Background & blanks

• Standard blank - blank used to monitor polyatomic ion interferences, gas peaks, and contamination from reagents; used for background subtraction

• Procedural blank - blank used to monitor contamination acquired during all stages of sample preparation; grinding, digestion, acidification, powdering, etc

Page 42: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Background & blanks

• Use of blanks during an analytical session:

• ALWAYS begin an analytical session with at least one standard blank

• Analyze standard blanks periodically throughout the course of the session in particular to monitor memory effects

• Process and analyze at least one procedural blank at some point during your research study; for its analysis, it’s preferable to measure it early in order to avoid any potential memory effects

Page 43: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Background & blanks

• The more standard blanks that are run during an analytical session, the more information you will have with regards to monitoring change(s) in background levels throughout the entire session

Page 44: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Background & blanks

• How to determine “the background”:

• 1. just use the first standard blank

• 2. average all standard blanks

• 3. take median of all standard blanks

• 4. apply statistical analysis to standard blanks• and select some of them

Page 45: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Background & blanks

• Outlier tests:

• 1. I know the truth

• 2. Looks different

• 3. Statistical “proof”

Page 46: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Background & blanks

• Option 1 should be avoided - unscientific and invalid

• Option 2 is better but only if the measurement is repeated

• Option 3 is the best approach, but needs to be carried out carefully in order to avoid false negatives and positives

Page 47: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Background & blanks

• Huber Outlier Test

• take median of all values

• calculate absolute deviation |xi - xm|

• take mean of absolute deviations (MAD)

• multiply MAD by coefficient (k = 3-5)

• anything higher than k*MAD is rejected as outlier

Page 48: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –Background & blanks

• Calculation of Limit of Detection (DL) and Limit of Quantification (QL)

• Easy way: • LOD = 3*STDEVblank;• LOQ = 10*STDEVblank

Page 49: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Sample Analysis Design –SUMMARY

• A ‘good’ analytical method will:

• 1. provide the means to calculate an accurate background level

• 2. allow for correction of instrument drift

• 3. use Internal standardization to monitor matrix effects

• 4. provide some method for monitoring/ correcting interferences

• 5. Use a proper calibration strategy

Page 50: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:
Page 51: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:
Page 52: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:
Page 53: Sample Analysis Design – Step 2 – …asimonet/ENGV60500/Lecture_8_10_11_2011.pdfSample Analysis Design – Step 2 – Calibration/Standard Preparation • EXTERNAL CALIBRATION:

Example Calculation –Determination of Ca and Na in

beetle blood

Using External Calibration Method