s j doran, k k koerkamp*,
DESCRIPTION
S. Department of Physics, University of Surrey, Guildford, GU2 7XH, UK. Dept of Applied Physics, University of Twente, Enschede, NL. S J Doran, K K Koerkamp. Technical development of a high resolution CCD-based scanner for 3-D gel dosimetry: (II) Problems encountered. - PowerPoint PPT PresentationTRANSCRIPT
Technical development of a high resolution CCD-based scanner for 3-D gel dosimetry:
(II) Problems encountered
S J Doran, K K Koerkamp*,
S Department of Physics,University of Surrey,Guildford, GU2 7XH, UK
Department of PhysicsUniversity of Surrey
Department of Applied PhysicsUniversity of Twente, NL
*
S J Doran, K K Koerkamp Dept of Applied Physics,University of Twente,Enschede, NL
Structure of talk
• Factors determining signal detected
• Detector and projection screen characteristics
• The “ring artifact” and how to remove it
• The “correction scan” procedure
• Sample containers
• The dynamic range problem
• Conclusions
Signal measured in CCD tomography
Hglamp
Cylindrical lens, pinholeand filter pseudo
point-source
Lens parallel beamScanning tank withmatching medium
Exposed gel
Unexposed gel Diffuser screen on which realshadow image forms
CCDdetector
Standard 50mmcamera lens
PC with frame-grabber card
Turntable controlled by acquisitioncomputer via stepper motors • Light field L(x,y)
• Projection screenPS(x,y)
• Detector responseD(x,y,S)
• Gel absorptionG(x,y,)
• Reflection and refraction• None of these quantities is known a priori.
• We can estimate L(x,y) relatively easily.
• PS(x,y) and D(x,y,S) can be a problem.
CCD detector characteristics: (1) Dark response
• We started out using a cheap CCD detector (~£120).
• The “noise” from the detector has a clear structure.
• This has serious consequences for improvement in signal by averaging frames.
(naverages)1/2
0 5 10 15 20
1 /
SNR
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
for pixels in detector column
for pixels in detector row
-5
+5
-5
+5
Image from detector with lens cap on!
CCD detector characteristics: (2) Light response
• Response measured by exposing CCD to different light levels, obtained using two polaroids rotated w.r.t. each other.
• No need for parallel beam here. Collimating optics and projection screen not used.
• Does the response vary with pixel position?
0
200
400
600
800
1000
1200
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1
Relative intensity
Pix
el v
alu
e
Measured data points6th order polynomial fit
Open light field and projection screen
• Relatively easy to separate slowly varying L(x,y) from PS(x,y) and D(x,y).
• Oscillating the projection screen up and down “smears out” some of the granularity.
• How much of the residual noise comes from PS(x,y) and how much from D(x,y)?
• Replacing the projection screen shows the extent of the problem.
P ix e l n u m b e r in x -d ire c t io n0 5 0 1 0 0 1 5 0 2 0 0 2 5 0 3 0 0 3 5 0 4 0 0 4 5 0 5 0 0
Pix
el v
alu
e
6 0 0
7 0 0
8 0 0
9 0 0
1 0 0 0
1 1 0 0
P ix e l n u m b e r in y -d ire c t io n
0 5 0 1 0 0 1 5 0 2 0 0
Without oscillating projection screen
Oscillating projection screen
Pros and cons of two different projection screens
• Screen 1 (engineering tracing film) is granular, but produces sharp images.
• Screen 2 (opal white perspex) has much less granularity, but the projection images are blurred.
Pixel number in x-direction0 50 100 150 200 250 300 350 400 450 500
Pix
el v
alu
e
600
700
800
900
1000
1100
Consequences of PS(x,y) and D(x,y)
• The “noise” generated is coherent between projections.
• This gives rise to a characteristic ring artifact.
Ideal object Simulated artifact Experimental artifact Artifact removed by “wobble”
“Correction” of the ring artifact via “wobble”
• At each projection step, the detector moves randomly relative to the tank by a few pixels.
• Can be achieved either by moving tank or camera.
• This allows us to sample the response functions of different pixels over the course of the acquisition.
• Coherent noise turns into random noise!
Hglamp
CCDdetector
PC with frame-grabber card
Containers and the correction scan
• Sample container has refractive index different to that of the sample and the matching medium.
• This causes partial reflection and refraction.
• Containers are imperfect, leading to artifacts in the projections.
• Problems can be partly overcome by taking the ratio of images before (“correction scan”) and after irradiation.
Before irradiation After irradiation Processed sinogram
Scratch mark
Artifacts due to imperfect containers
• Minute scratches on the container wall cause spurious reflection and refraction.
• These are easily seen as parallel tracks in the sinograms.
• They lead to characteristic artifacts at the edge of the field-of-view
- 6 Gy
6 Gy
Dynamic range problems
• Video capture card has limited dynamic range (10-bit).
• Light travelling through low-dose region saturates ADC.
• Light travelling through high-dose region registers a very low signal that is strongly affected by noise.
• Extremely important to make absorption of matching medium same as that of unirradiated gel.
Conclusions
• We now understand many of the causes of artifacts in the OCT images.
• Most of the artifacts can be removed by investment in higher quality components (particularly the CCD, projection screen and sample container).