R&S®TSMA Release Notes Firmware Version 1.32.01.00
© 2017 Rohde & Schwarz GmbH & Co. KG
Muehldorfstr. 15, 81671 Munich, Germany
Phone: +49 89 41 29 - 0
Fax: +49 89 41 29 12 - 164
E-mail: mailto:[email protected]
Internet: http://www.rohde-schwarz.com
Subject to change
R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG. Trade names are trademarks of the owners.
The firmware of the instrument makes use of several valuable open source software packages. For information, see the "Open Source
Acknowledgement" document, which is available for download from the R&S TSMA product page at https://www.rohde-
schwarz.com/firmware/tsma/.
The following abbreviations are used throughout this document R&S®TSMA is abbreviated as R&S TSMA
Rel
ease
Not
es
1SP
6
R&S TSMA Information on the Current Version and History
Release Notes 1177.6081.02 - 10 2
Contents
1 Information on the Current Version and History .............................. 4
1.1 Version 1.32.01.00 ........................................................................................................ 4
1.2 Version 1.32.00.04 ........................................................................................................ 4
1.3 Version 1.31.00.07 ........................................................................................................ 4
1.4 Version 1.30.00.03 ........................................................................................................ 5
1.5 Version 1.28.00.06 ........................................................................................................ 6
1.6 Version 1.27.01.03 ........................................................................................................ 6
1.7 Version 1.20.03 ............................................................................................................. 7
1.8 Version 1.15.03 ............................................................................................................. 8
1.9 Version 1.07 .................................................................................................................. 9
1.10 Version 1.05 .................................................................................................................. 9
1.11 Version 1.04 ................................................................................................................10
2 Firmware Update ............................................................................... 11
2.1 Preparing Installation ................................................................................................11
2.1.1 Choosing the Way of Installation .................................................................................11
2.1.2 Preparation for Local Installation .................................................................................12
2.1.3 Preparation for Remote Installation .............................................................................12
2.1.4 Preparation for TSMA Web-GUI based Installation .....................................................13
2.1.5 Setting Mode of Operation ...........................................................................................13
2.2 Performing Firmware Update ...................................................................................14
2.2.1 Local Installation ..........................................................................................................14
2.2.2 Remote Installation ......................................................................................................15
2.2.3 Using the R&S TSMA Web-GUI ..................................................................................17
2.3 Installation of Remote ViCom Server.......................................................................19
2.3.1 Preparation ..................................................................................................................19
2.3.2 ViCom Server Installation ............................................................................................20
3 Software Installation ......................................................................... 22
3.1 R&S NESTOR Software .............................................................................................22
3.1.1 Preparation ..................................................................................................................22
3.1.2 NESTOR Installation ....................................................................................................22
3.2 QualiPoc© Software ...................................................................................................24
3.2.1 Preparation ..................................................................................................................24
3.2.2 QualiPoc© Installation ...................................................................................................25
R&S TSMA Information on the Current Version and History
Release Notes 1177.6081.02 - 10 3
3.3 R&S ROMES Software ...............................................................................................27
3.3.1 Preparation ..................................................................................................................27
3.3.2 ROMES Installation......................................................................................................27
4 Customer Support ............................................................................ 30
R&S TSMA Information on the Current Version and History
Release Notes 1177.6081.02 - 10 4
1 Information on the Current Version and
History
1.1 Version 1.32.01.00
Improvements
Version Improvements
1.32.01.00 Restart via web-GUI triggers power cycle of scanner sub. component 1.32.01.00 New version of TsmeTools (TsmeTools v1.4.0.2) 1.32.01.00 Watchdog time out triggers power cycle
Solved Problems
Version Solved Problems
1.32.01.00 TSMA6-BP STATE LEDs not operated in stand-by mode
1.2 Version 1.32.00.04
Improvements
Version Improvements
1.32.00.04 .NET 4.6..2 is part of the firmware setup 1.32.00.04 Successful ROMES license file installation is confirmed in the web-GUI
Solved Problems
Version Solved Problems
1.32.00.04 Hang-up in battery communication when battery pack is connected 1.32.00.04 Power on via power button sometimes not possible
1.3 Version 1.31.00.07
Firmware package contents
Version Contents
1.31.00.07 Bluetooth configuration in Web-GUI in Win10
1.31.00.07 Additional mode “Smart Monitor” supported in Web-GUI
1.31.00.07 Auto-Power On mode with TSMA-BP power supply
1.31.00.07 Charge indication with TSMA-BP in STB mode
1.31.00.07 ROMES “Probe Mode” removed
R&S TSMA Information on the Current Version and History
Release Notes 1177.6081.02 - 10 5
1.31.00.07 Backup / Recovery Win10
1.31.00.07 Updated online help
Improvements
Version Improvements
1.31.00.07 Auto-Refresh off all dynamic Web-GUI contents
1.31.00.07 Remote Shutdown with power on delay 30 s
Solved Problems
Version Solved Problems
1.31.00.07 NESTOR mode not loaded after restart when activated in Web-GUI
1.4 Version 1.30.00.03
Firmware package contents
Version Contents
1.30.00.03 Reboot button in web-GUI
1.30.00.03 TSME service adaptations for Win10
1.30.00.03 Support Challenger
1.30.00.03 TSME Tools v1.3.5.3
1.30.00.03 ViCom Server v15.80
Improvements
Version Improvements
1.30.00.03 Only installed SW packages are offered in mode selection of the web-GUI
Solved Problems
Version Solved Problems
1.30.00.03 Device hang-ups (power button and battery pack communication)
1.30.00.03 ViCom Server does not start after power up
1.30.00.03 Unsynchronized launch of QualiPoc and ROMES Probe processes at device startup
1.30.00.03 Bluetooth adapter disabled after firmware update
1.30.00.03 ViCom SampleApp not selectable in web-GUI for download
1.30.00.03 Failed user backups due to small partition size
1.30.00.03 Wrong device identity with connected TSME in a MIMO setup
Known Issues
Version Known-Issues
R&S TSMA Information on the Current Version and History
Release Notes 1177.6081.02 - 10 6
1.5 Version 1.28.00.06
Firmware package contents
Version Contents
1.28.00.06 Support for QualiPoc v16.2 (Windows service)
Improvements
Version Improvements
1.28.00.06 Online Help updated
1.28.00.06 Reboot requests are indicated in web-GUI status text.
Solved Problems
Version Solved Problems
1.28.00.06 NESTOR mode – service is not started after restart
1.28.00.06 MeasurementConfigVicom.xml not evaluated
1.28.00.06 Web-GUI password protection does not work
1.28.00.06 Firmware not downward compatible with TSMA Access-DLL.
1.28.00.06 Mode changes from “ROMES Probe” mode don’t end ROMES satellite application.
1.28.00.06 Watchdog initiated restarts when not initialized.
Known Issues
Version Known-Issues
1.28.00.06 Bluetooth adapter disabled after firmware update / initial activation via web-GUI requires a restart.
1.28.00.06 ViCom Sample App setup for Android devices is not selectable in File Transfer/Update menu of the web-GUI.
This is only valid for the TSMA ViCom setup v15.75. Will be solved in a higher version
of the TSMA rViCom setup (TSMA-rViCom-Setup-<Version>.exe)..
1.6 Version 1.27.01.03
Firmware package contents
Version Contents
1.27.01.03 Support for various ROMES modes in web-GUI
1.27.01.03 Configurable start-up delay in web-GUI
1.27.01.03 Auto power off feature in case of DC IN absence – configuration via web-GUI
R&S TSMA Information on the Current Version and History
Release Notes 1177.6081.02 - 10 7
1.27.01.03 User image creation from web-GUI / requires image version 1.03 / 1.09
1.27.01.03 Tool chain to create & deploy a master image
1.27.01.03 Windows tray icon for battery charge
1.27.01.03 TsmeTools update to v1.3.5.1
1.27.01.03 System restart button via web-GUI
1.27.01.03 Remote shut-down support
1.27.01.03 Version info table of all installed SW applications / web-GUI
1.27.01.03 WLAN permanent off configurable via web-GUI
Improvements
Version Improvements
1.27.01.03 Auto-refresh of dynamic web-GUI contents (battery power, temperatures)
1.27.01.03 Charging indication off battery pack via LEDs
1.27.01.03 Battery charge indication of battery pack when TSMA powered off / STB
1.27.01.03 WLAN AP is now indicated via the STATE LED
1.27.01.03 Power on hysteresis in case off low power battery state
1.27.01.03 File size limitation for web-GUI based downloads increased to 4 GB
Solved Problems
Version Solved Problems
1.27.01.03 Firmware installation blockings when TsmeTools were updated manually
1.27.01.03 Hang-ups when changing the operation mode in web-GUI
1.27.01.03 Changing band options via Web-GUI
1.27.01.03 Installation / query of NESTOR licenses via Web-GUI
Known Issues
Version Known-Issues
1.7 Version 1.20.03
Firmware package contents
Version Contents
1.20.03 Update TsmeTools v1.3.4.5
1.20.03 Updated Online Help v05
1.20.03 Default keyboard layout English US
1.20.03 Indication of failed scanner link via STATE LED
1.20.03 Indication of “Scanner Only Mode” via STATE LED
1.20.03 Progress of software installation is indicated by blinking STATE LED
Improvements
Version Improvements
R&S TSMA Information on the Current Version and History
Release Notes 1177.6081.02 - 10 8
1.20.03 Auto-Shutdown with low battery power
1.20.03 Status feedback in Web-GUI while firmware / software installation
1.20.03 Minor changes in Web-GUI usability
1.20.03 Selected NESTOR setup is displayed in the Web-GUI
Solved Problems
Version Solved Problems
1.20.03 Instable / missing WLAN access point after power up
1.20.03 Diversity start in “PC mode” when executing setup
Known Issues
Version Known-Issues
1.20.03 Web-GUI does not support installation and query of NESTOR licenses
1.8 Version 1.15.03
Firmware package contents
Version Contents
1.15.03 Support of NESTOR Measurement Probe
1.15.03 Installation of NESTOR option keys via Web-GUI added
1.15.03 Update TsmeTools v1.3.4.3
Improvements
Version Improvements
1.15.03 LAN port setting via Web-GUI
1.15.03 Bluetooth binding via Web-GUI
1.15.03 SCPI Support (Date, Time)
1.15.03 Reliability of WLAN access point if enabled via Web-GUI
Solved Problems
Version Solved Problems
1.15.03 Incomplete firmware setups due to hardware watchdog
Known Issues
Version Known-Issues
1.15.03 Diversity setup 15.1.0.27 alters default WLAN key / manual correction necessary
1.15.03 NESTOR setup 1.4 requires manual setting of Windows firewall permissions
1.15.03 Sporadic absence of WLAN access point after power up
R&S TSMA Information on the Current Version and History
Release Notes 1177.6081.02 - 10 9
1.9 Version 1.07
Firmware package contents
Version Contents
1.07 Configurable Power On Mode ( Auto Power On / Remember Last State)
1.07 Default paths for measurement applications (NESTOR, rVicom Server, Diversity) are configured with a MeasurementConfig.xml file
1.07 Configuration of TSMA as WLAN client via Web-GUI
1.07 Configuration of NESTOR auto-start mode with selectable workspace via Web-GUI
Improvements
Version Improvements
1.07 Matching of LAN/WLAN settings in Web-GUI and in Windows OS
1.07 WLAN Access Point: default on
1.07 Web-GUI now reports absolute state of charge
Known Issues
Version Known-Issues
1.07 NESTOR slave mode configuration in Web-GUI not supported yet
1.07 Applying fixed IP address und subnet mask of LAN port via Web-GUI initializes of LAN Gateway and DNS server address to 10.0.0.4
1.10 Version 1.05
Firmware package contents
Version Contents
1.05 No changes to previous version
Improvements
Version Improvements
1.05 LAN/WLAN configuration in the Web-GUI improved
Known Issues
Version Known-Issues
1.05 NESTOR mode selection / configuration via Web-GUI not supported
1.05 Web interface displays wrong software version
1.05 TSMA configuration as WLAN client via Web-GUI not supported
1.05 DHCP setting in Web-GUI does not set DNS Server to "obtain DNS server automatically
1.05 Fix IP LAN configuration in Web-GUI sets Gateway and DNS server to 10.0.0.4
1.05 Web interface displays ASOC for batteries in overview page
R&S TSMA Information on the Current Version and History
Release Notes 1177.6081.02 - 10 10
1.11 Version 1.04
Firmware package contents
Version Contents
1.04 TSMA Firmware
1.04 ViCom Server v15.51
1.04 Microcontroller v1.01
1.04 Installation Assistant Service v2.52.18
1.04 Compass Platform v3.1.15.206
1.04 Compass Platform LXI v3.1.15.206
1.04 Visa Runtime v2.53.2.0
1.04 MDNS Service v2.0.320.12
1.04 TSME Tools v1.3.3.6
New Functionality
Version Functions
1.04 Initial release
Known Issues
Version Known-Issues
1.04 LAN/WLAN configuration via Web-GUI might sometimes work not properly
1.04 NESTOR mode configuration in Web-GUI not yet supported
1.04 IP address settings in Web-GUI sometimes not applied in Windows
R&S TSMA Firmware Update
Release Notes 1177.6081.02 - 10 11
2 Firmware Update
Firmware update packages for the R&S TSMA could be downloaded from the Rohde &
Schwarz web page https://www.rohde-schwarz.com/en/firmware/tsma/.
The firmware setup file is named “TSMA-Setup-<Version>.exe”.
Remote ViCom Server Update
“Remote ViCom Server (rViCom server)” is a supplementary firmware package.
This package is only necessary when the R&S TSMA is operated in “rViCom Server”
mode. After firmware installation, please proceed with rViCom server installation, see
Installation of Remote ViCom Server.
For Firmware Version 1.07 and Lower
The local installation, see chapter 2.2.1 Local Installation, is the recommended way to
upgrade devices with firmware 1.07 and lower.
2.1 Preparing Installation
2.1.1 Choosing the Way of Installation
There are several ways how to upgrade R&S TSMA firmware / software. Select the
preferred way.
2.1.1.1 Local execution of the setup file on the R&S TSMA Requires to connect mouse, keyboard and monitor to the R&S TSMA und to execute
the setup file on the device, see Chapter 2.1.2, “Preparation for Local Installation”.
2.1.1.2 Remote execution of the setup file from a distant PC The R&S TSMA need to be connected via LAN / WLAN to the remote PC. The setup is
executed and controlled from the remote PC (Windows 7 / 10). There is no need to
interface the R&S TSMA, see Chapter 2.1.3, “Preparation for Remote Installation”.
2.1.1.3 Web-GUI based upgrade from a handheld device using an USB
stick Requires the setup file to be available on a USB memory stick. The memory stick need
to be connected to an USB socket of the R&S TSMA. The web-GUI could be called
locally on the R&S TSMA or from a remote device (Windows PC) or handheld device
(Android), see Chapter 2.1.4, “Preparation for TSMA Web-GUI based Installation”.
R&S TSMA Firmware Update
Release Notes 1177.6081.02 - 10 12
2.1.2 Preparation for Local Installation
1. Connect mouse, keyboard to the USB ports and a monitor to the mini display port.
Display Port Adapter DVI2HDMI: R&S TSPC-DPDH (R&S No. 3592.4060.02)
2. Switch on the R&S TSMA via power button.
3. Copy the setup file:
a) For firmware installation: Copy the firmware setup file “TSMA-Setup-
<Version>.exe” into a temporary directory on the local disk of the R&S
TSMA.
b) For rViCom server / software installation: Copy the corresponding setup file
into a temporary directory on the local disk of the R&S TSMA.
4. Proceed with section 2.1.5 Setting Mode of Operation.
2.1.3 Preparation for Remote Installation
1. Copy the setup file:
a) For firmware installation: Copy the firmware setup file “TSMA-Setup-
<Version>.exe” into a temporary directory on the remote PC.
b) For rViCom / software installation: Copy the corresponding setup into a
temporary directory on the remote PC.
2. Establish a LAN respectively WLAN connection between the remote PC and the
R&S TSMA, see following section.
2.1.3.1 Using LAN Connection
1. Connect the LAN port (R&S TSMA) with the host PC LAN port. The default LAN
setting of the R&S TSMA is DHCP client. Ask your network administrator for
details on how to configure the remote PC.
2. Switch on the R&S TSMA via power button.
3. Proceed with section 2.1.5 Setting Mode of Operation.
2.1.3.2 Using WLAN Connection
1. Switch on the R&S TSMA via power button.
2. On the remote PC search for the TSMA WLAN and connect to the network. The
necessary access parameters can be found on a label at the bottom of the device.
WLAN SSID: TSMA-<xxxxxx> xxxxxx: individual serial number
WLAN key: instrument
Per default the WLAN access point is enabled. See in the
R&S TSMA Firmware Update
Release Notes 1177.6081.02 - 10 13
R&S TSMA User Manual how the enable the WLAN access point when disabled.
3. Proceed with section 2.1.5 Setting Mode of Operation.
2.1.4 Preparation for TSMA Web-GUI based Installation
1. Switch on the R&S TSMA via power button.
2. Copy the setup file onto a USB memory stick
a) For firmware installation: Copy the firmware setup file “TSMA-Setup-
<Version>.exe” into the root directory of the stick.
b) For rViCom / software installation: Copy the corresponding setup file into the
root directory of the stick.
3. Proceed with section 2.1.5 Setting Mode of Operation.
2.1.5 Setting Mode of Operation
1. The R&S TSMA Web-GUI has to be loaded to check the mode of operation
a) Remote installation from a WLAN connected device
Start the web browser and enter the following URL:
http://192.168.1.10/
b) Remote installation from a LAN connected device
Enter the URL http://TSMA-xxxxxx.local into the browser.
TSMA-xxxxxx = R&S TSMA host name. This information can be found on a
label at the bottom of the device.
c) Local installation
Open the Internet Explorer. The Web-GUI should be loaded automatically. If
not, enter the URL http://localhost into the browser.
2. The mode of operation is displayed in the overview page. If the mode is already
“PC Mode” you can proceed with step 4.
3. If a different mode is activated, go to the “System” page and select “PC Mode”.
Press the “Submit” button to activate this setting. You are asked if you want to
change to mode. Acknowledge with “OK”. The R&S TSMA will reboot.
R&S TSMA Firmware Update
Release Notes 1177.6081.02 - 10 14
A status text is displayed at the bottom of the Web-GUI. The reboot is finished
when “No error” is displayed again.
Double check that “PC Mode” is now activated.
4. The R&S TSMA is ready for firmware update / software installation. Proceed with
2.2, “Performing Firmware Update” respectively 2.3.2, “ViCom Server Installation”
or 3, “Software Installation”.
2.2 Performing Firmware Update
2.2.1 Local Installation
1. Consider the instruction in the chapter 2.1.2, “Preparation for Local Installation”.
2. Open the Windows Explorer and execute the setup file
“TSMA-Setup-<Version>.exe”.
3. The “R&S Software Distributor” window comes up. Select "Local Installation" and
press “Next”.
4. The package dialog lists all the available packages in the setup. Normally, this
dialog can be skipped without any changes. “Automatic Reboot” should be
selected.
R&S TSMA Firmware Update
Release Notes 1177.6081.02 - 10 15
5. Press “Install” to start setup.
The STATE LED starts blinking blue.
Important Note
Do not switch off or unplug from power while running firmware update!
Several reboot cycles may be executed automatically while running the update.
6. Setup is finished when the R&S TSMA comes up after a reboot. The R&S
distributor dialog is closed automatically.
The STATE LED changes into constant green or amber.
7. Reload the R&S TSMA Web-GUI and double check the displayed firmware
version in the “Overview” menu. The correct version number need to be indicated.
The R&S TSMA is ready for operation.
2.2.2 Remote Installation
1. Consider the instruction in chapter 2.1.3, “Preparation for Remote Installation”.
2. Execute the firmware setup file on the connected remote PC. The “R&S Software
Distributor” dialog box is coming up.
R&S TSMA Firmware Update
Release Notes 1177.6081.02 - 10 16
3. Select "Remote Installation" and click "Next >".
4. The package dialog lists all the available packages to install.
All packages, which you need to install are already pre-selected.
5. Click “Next >”.
6. After a while any detected R&S TSMA in the network is listed in the “Device List”
dialog. Tick the device(s) to update. If the column “serial number” is not filled
move the mouse pointer over the listed items to get related device information.
7. Click "Install".
Do not switch off or remove power from the R&S TSMA while executing this
update!
8. The state of the installation process is displayed in the “Status” column. The
process is finished when the status indicates “Ready…”.
9. Press “Exit”.
10. Reload the R&S TSMA Web-GUI and verify the installed firmware version in the
“Overview”
The R&S TSMA is ready for operation.
R&S TSMA Firmware Update
Release Notes 1177.6081.02 - 10 17
2.2.3 Using the R&S TSMA Web-GUI
1. Consider the instructions in chapter 2.1.4, “Preparation for TSMA Web-GUI based
Installation”.
2. In the R&S TSMA Web-GUI navigate to the “File Transfer / Update” menu. Open
the “Select Firmware Setup File” select box.
3. Select the appropriate firmware setup file TSMA-Setup-<Version>.exe and
press the “Execute Setup” next to the select box.
4. A confirmation dialog box appears. Confirm with “OK” to start the update.
5. The firmware installation starts. The Web-GUI displays “Installation in progress…”.
The STATE LED is blinking blue for the whole update process.
During firmware upgrade the device may run several reboot cycles.
R&S TSMA Firmware Update
Release Notes 1177.6081.02 - 10 18
Do not switch off or remove power from the R&S TSMA while executing this
update!
6. At the end of the installation the status text in the R&S TSMA Web-GUI changes
to “Installation finished – system is rebooting”.
7. After the reboot the R&S TSMA Web-GUI is refreshed automatically. The upgrade
is finished when the status text changes to “No error” and the STATE LED
changes into constant green or amber.
8. The “Overview” page in the Web-GUI need to display the installed firmware
version.
R&S TSMA Firmware Update
Release Notes 1177.6081.02 - 10 19
The R&S TSMA is ready for operation.
2.3 Installation of Remote ViCom Server
The “Remote ViCom Server” package is supplementary for the R&S TSMA. This
package is pre-installed on any shipped R&S TSMA.
An upgrade of the Remote ViCom package is only necessary when operating the R&S
TSMA in “rViCom Server” mode.
This installation could be neglected when running the R&S TSMA in any other mode of
operation (NESTOR, ROMES, QualiPoc).
The installed version on the R&S TSMA need to match with the applied version of the
rViCom client API on the remote Android device. So be careful to update.
In cases of a client / server version mismatch there are the following possibilities:
a) Downgrade the rViCom server version on the R&S TSMA to match with the
applied rViCom client version on the Android device.
b) Upgrade the rViCom client version on the connected tablet / smartphone to
comply with the installed rViCom server version on the R&S TSMA.
2.3.1 Preparation
1. Download the firmware setup file TSMA-rViCom-Setup-<Version>.exe from
the product support database https://www.rohde-schwarz.com/en/firmware/tsma/.
2. Choose the way of installation and follow the instructions how to prepare, see
Chapter 2.1.1, “Choosing the Way of Installation”.
R&S TSMA Firmware Update
Release Notes 1177.6081.02 - 10 20
2.3.2 ViCom Server Installation
2.3.2.1 ViCom Server Local Installation
1. Open the Windows Explorer and execute the setup file
“TSMA-rViCom-Setup-<Version>.exe”.
2. The “R&S Software Distributor” window comes up. Select "Local Installation" and
press “Next”.
3. All subsequent steps are similar to local firmware installation, see Chapter 2.2.1,
“Local Installation” step 4.
Do not switch off or remove power from the R&S TSMA while executing this
update!
4. Setup is finished when the STATE LED changes from blinking blue to constant
green or amber (firmware version ≥ 1.20.03).
2.3.2.2 ViCom Server Remote Installation
1. Execute the setup file TSMA-rViCom-Setup-<Version>.exe on the remote
PC.
2. The “R&S Software Distributor” comes up. Select "Remote Installation" and press
“Next >”.
R&S TSMA Firmware Update
Release Notes 1177.6081.02 - 10 21
3. All subsequent steps are similar to remote firmware installation, see Chapter
2.2.2, “Remote Installation” step 4.
2.3.2.3 ViCom Server Installation via Web-GUI
1. Select the appropriate firmware setup file TSMA-rViCom-Setup-
<Version>.exe and press the “Execute Setup” next to the select box.
2. All subsequent steps are similar to firmware installation via web-GUI, see Chapter
2.2.3, “Using the R&S TSMA Web-GUI” step 4.
R&S TSMA Software Installation
Release Notes 1177.6081.02 - 10 22
3 Software Installation
Only use the dedicated TSMA setups for software installation.
3.1 R&S NESTOR Software
Only execute the dedicated NESTOR setup for TSMA. The setup file is named:
TSMA-NESTOR-Setup-<Version>.exe.
3.1.1 Preparation
1. Download the software setup file TSMA_NESTOR_Setup-<Version>.exe from
the Rohde & Schwarz FTP server respectively NESTOR CD-ROM.
2. Choose the way of installation and follow the instructions how to prepare, see
Chapter 2.1.1, “Choosing the Way of Installation”.
3.1.2 NESTOR Installation
3.1.2.1 NESTOR Local Installation
1. Open the Windows Explorer and execute the NESTOR setup file
“TSMA_NESTOR_Setup-<Version>.exe”.
2. The “R&S Software Distributor” window comes up. Select "Local Installation" and
press “Next”.
R&S TSMA Software Installation
Release Notes 1177.6081.02 - 10 23
3. All subsequent steps are similar to local firmware installation, see Chapter 2.2.1,
“Local Installation” step 4.
Do not switch off or remove power from the R&S TSMA while executing this
update!
4. Setup is finished when the STATE LED changes from blinking blue to constant
green or amber (firmware version ≥ 1.20.03).
3.1.2.2 NESTOR Remote Installation
1. Execute the setup file “TSMA_NESTOR_Setup-<Version>.exe” on the remote
PC.
2. The “R&S Software Distributor” comes up. Select "Remote Installation" and press
“Next >”.
3. All subsequent steps are similar to remote firmware installation, see Chapter
2.2.2, “Remote Installation” step 4.
R&S TSMA Software Installation
Release Notes 1177.6081.02 - 10 24
Do not switch off or remove power from the R&S TSMA while executing this
update!
3.1.2.3 NESTOR Installation via Web-GUI
1. Select the appropriate NESTOR setup file “TSMA_NESTOR_Setup-
<Version>.exe” and press the “Execute Setup” next to the select box.
2. All subsequent steps are similar to firmware installation via Web-GUI, see Chapter
2.2.3, “Using the R&S TSMA Web-GUI” step 4.
3.2 QualiPoc© Software
Only execute the dedicated QualiPoc© setup for TSMA. The setup file is named:
DiversityProbe4QP-DeploymentKit-<Version>.exe.
3.2.1 Preparation
1. Download the QualiPoc© setup file “DiversityProbe4QP-DeploymentKit-
<Version>.exe” from the Rohde & Schwarz FTP server respectively QualiPoc©
CD-ROM.
R&S TSMA Software Installation
Release Notes 1177.6081.02 - 10 25
2. Choose the way of installation and follow the instructions how to prepare, see
Chapter 2.1.1, “Choosing the Way of Installation”.
3.2.2 QualiPoc© Installation
3.2.2.1 QualiPoc© Local Installation
1. Open the Windows Explorer and execute the QualiPoc© setup file
“DiversityProbe4QP-DeploymentKit-<Version>.exe”.
2. The “R&S Software Distributor” window comes up. Select "Local Installation" and
press “Next”.
3. All subsequent steps are similar to local firmware installation, see Chapter 2.2.1,
“Local Installation” step 4.
Do not switch off or remove power from the R&S TSMA while executing this
update!
4. Setup is finished when the STATE LED changes from blinking blue to constant
green or amber (firmware version ≥ 1.20.03).
3.2.2.2 QualiPoc© Remote Installation
1. Execute the QualiPoc© setup file “DiversityProbe4QP-DeploymentKit-
<Version>.exe” on the remote PC.
2. The “R&S Software Distributor” comes up. Select "Remote Installation" and press
“Next >”.
R&S TSMA Software Installation
Release Notes 1177.6081.02 - 10 26
3. All subsequent steps are similar to remote firmware installation, see Chapter
2.2.2, “Remote Installation” step 4.
Do not switch off or remove power from the R&S TSMA while executing this
update!
3.2.2.3 QualiPoc© Installation via Web-GUI
1. Select the appropriate QualiPoc© setup file “DiversityProbe4QP-
DeploymentKit-<Version>.exe” and press the “Execute Setup” next to the
select box.
2. All subsequent steps are similar to firmware installation via Web-GUI, see Chapter
2.2.3, “Using the R&S TSMA Web-GUI” step 4.
R&S TSMA Software Installation
Release Notes 1177.6081.02 - 10 27
3.3 R&S ROMES Software
3.3.1 Preparation
1. Download the ROMES setup file “TSMA_ROMES_Setup-<Version>.exe” from
the Rohde & Schwarz FTP server respectively ROMES CD-ROM.
2. Choose the way of installation and follow the instructions how to prepare, see
Chapter 2.1.1, “Choosing the Way of Installation”.
3.3.2 ROMES Installation
3.3.2.1 ROMES Local Installation
1. Open the Windows Explorer and execute the ROMES setup file
“TSMA_ROMES_Setup-<Version>.exe”.
2. The “R&S Software Distributor” window comes up. Select "Local Installation" and
press “Next”.
3. All subsequent steps are similar to local firmware installation, see Chapter 2.2.1,
“Local Installation” step 4.
Do not switch off or remove power from the R&S TSMA while executing this
update!
4. Setup is finished when the STATE LED changes from blinking blue to constant
green or amber (firmware version ≥ 1.20.03).
3.3.2.2 ROMES Remote Installation
R&S TSMA Software Installation
Release Notes 1177.6081.02 - 10 28
1. Execute the ROMES setup file “TSMA_ROMES_Setup-<Version>.exe” on the
remote PC.
2. The “R&S Software Distributor” comes up. Select "Remote Installation" and press
“Next >”.
3. All subsequent steps are similar to remote firmware installation, see Chapter
2.2.2, “Remote Installation” step 4.
Do not switch off or remove power from the R&S TSMA while executing this
update!
3.3.2.3 ROMES Installation via Web-GUI
1. Select the appropriate ROMES setup file “TSMA_ROMES_Setup-
<Version>.exe” and press the “Execute Setup” next to the select box.
R&S TSMA Software Installation
Release Notes 1177.6081.02 - 10 29
2. All subsequent steps are similar to firmware installation via Web-GUI, see Chapter
2.2.3, “Using the R&S TSMA Web-GUI” step 4.
R&S TSMA Customer Support
Release Notes 1177.6081.02 - 10 30
4 Customer Support
Technical support – where and when you need it
For quick, expert help with any Rohde & Schwarz equipment, contact one of our
Customer Support Centers. A team of highly qualified engineers provides telephone
support and will work with you to find a solution to your query on any aspect of the
operation, programming or applications of Rohde & Schwarz equipment.
Up-to-date information and upgrades
To keep your instrument up-to-date and to be informed about new application notes
related to your instrument, please send an e-mail to the Customer Support Center
stating your instrument and your wish. We will take care that you will get the right
information.
Europe, Africa, Middle East Phone +49 89 4129 12345 [email protected]
North America Phone 1-888-TEST-RSA (1-888-837-8772) [email protected]
Latin America Phone +1-410-910-7988 [email protected]
Asia/Pacific Phone +65 65 13 04 88 [email protected]
China Phone +86-800-810-8828 / +86-400-650-5896 [email protected]