rp-hplc method for the quantitative determination of fexofenidine hcl in coated tablets and human

Upload: chiralic

Post on 04-Apr-2018

222 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/31/2019 RP-HPLC Method for the Quantitative Determination of Fexofenidine HCl in Coated Tablets and Human

    1/13

    O R I G I N A L R E S E A R C H

    RP-HPLC method for the quantitative determination

    of fexofenadine hydrochloride in coated tabletsand human serum

    M. Saeed Arayne Najma Sultana Hina Shehnaz

    Amir Haider

    Received: 26 May 2009 / Accepted: 4 December 2009 Springer Science+Business Media, LLC 2009

    Abstract Fexofenadine is a non-sedative and selective peripheral H1 receptor

    antagonist prescribed for allergic rhinitis and chronic urticaria. This article deals

    with a simple, feasible, and sensitive isocratic reverse-phase high-performance

    liquid chromatographic method for the determination of fexofenadine hydrochloride

    in bulk drug, pharmaceutical dosage forms and in human serum. The chromatog-

    raphy was carried out at 20 2C using two different chromatographs and five

    different stationary phases. The isocratic mobile phase was phosphate buffer pH 7.4and methanol (methanolphosphate buffer, 35:65, v/v), detection was made at

    218 nm and the mobile phase flowed at 1 ml min-1. Validation parameters included

    linearity, accuracy, precision, specificity, limit of detection (LOD), limit of quan-

    tification (LOQ), and robustness over a linearity range 515 lg ml-1

    according to

    the ICH guidelines (r[0.9999), the inter- and intra-day precisions were relative

    standard deviation (RSD)\0.8%. The system suitability was scrutinized by

    capacity factor, tailing factor, and number of theoretical plates (capacity fac-

    tor[ 2.0, tailing factor B 2.0, and theoretical plates[2000). The retention time

    for five different stationary phases ranged from 3.78 to 4.15 min. The LOD andLOQ for the procedure were executed on samples containing very low concentra-

    tions of analytes on two different commercial brands of detectors.

    Keywords Fexofenadine Urticaria High-performance liquid chromatography

    Isocratic Robustness Five different stationary phases

    M. S. Arayne

    H. Shehnaz (&

    )

    A. HaiderDepartment of Chemistry, University of Karachi, Karachi 75270, Pakistan

    e-mail: [email protected]

    N. Sultana

    Department of Pharmaceutical Chemistry, Faculty of Pharmacy, University of Karachi,

    Karachi 75270, Pakistan

    Med Chem Res

    DOI 10.1007/s00044-009-9285-6

    MEDICINALCHEMISTRYRESEARCH

  • 7/31/2019 RP-HPLC Method for the Quantitative Determination of Fexofenidine HCl in Coated Tablets and Human

    2/13

    Introduction

    Fexofenadine, a,a-dimethyl-4-[1-hydroxy-4-[4-(hydroxydiphenyl-methyl)-1-pipe-

    ridinyl]butyl]-benzene acetic acid (Fig. 1), is used to relieve the allergy symptoms

    of seasonal allergic rhinitis (hay fever), including runny nose; sneezing; and red,itchy, or watery eyes; or itching of the nose, throat, or roof of the mouth in adults

    (Markham and Wagstaff, 1998; Simpson and Jarvis, 2000). It is carboxylic acid

    metabolite of terfenadine, a non-sedating selective histamine H1 receptor antagonist

    (Caballero et al., 1999). Unlike its precursor, fexofenadine lacks the cardiotoxic

    potential, effective in the management of allergic rhinitis, and chronic idiopathic

    urticaria for which it is a suitable option for first-line therapy (Inomata et al., 2009).

    Fexofenadine is a substrate of P-glycoprotein, used to explore activity in vivo

    because it is not metabolized in human body. Besides, no sedative or other central

    nervous system effects were observed and radiolabeled tissue distribution studies inrats indicated that fexofenadine does not cross the bloodbrain barrier (British

    Pharmacoepia, 2000; Barnes et al., 1993).

    On pharmaceutical dosage forms, very few methods from quality control are

    presented to determine fexofenadine in dosage form, in urine, serum, and

    pharmaceutical formulations (Drescher et al., 2002; Mattila and Paakkari, 1999).

    Three methods were presented for determination of fexofenadine in pure form and

    its R(1) and S(-) enantiomers were analyzed in plasma and urine by validated

    performance liquid chromatographic (HPLC) methods in commercial dosage forms

    (Gazy et al., 2002). Another bioanalytical method was reported using solid phaseextraction and liquid chromatography with electrospray (Naidong et al., 2002).

    Mass spectrometry (LC/MS/MS), RPLC, ionspray tandem mass spectrometry

    detection (Gergov et al., 2001; Fu et al., 2004), and fluorescence detection (Uno

    et al., 2004) are used in the pharmaceutical dosage form using ultraviolet

    spectrophotometry (Pratt et al., 1999; Radhakrishna and Om Reddy, 2002; Milne

    et al., 2000; Hamman et al., 2001; Russell et al., 1998; Pathaket al., 2008; Karakus

    et al., 2008).

    HPLC is the most widely used technique in pharmaceutical companies, clinical

    laboratories, and research and development laboratories. Fexofenadine is the most

    frequently prescribed H1 receptor antagonist manufactured by many pharmaceutical

    companies world wide. The methods reported in the literature for the determination

    of fexofenadine are time-consuming, difficult, and expensive making them

    unpractical for everyday clinical trials. Also, there is no specific method for the

    Fig. 1 Fexofenadine

    Med Chem Res

  • 7/31/2019 RP-HPLC Method for the Quantitative Determination of Fexofenidine HCl in Coated Tablets and Human

    3/13

    determination of only fexofenadine in pharmaceutical dosage forms as well as in

    human serum. Therefore, there was a need for developing an HPLC method for the

    determination of fexofenadine in reference drug material, pharmaceutical formula-

    tions, and in human serum, which should have adequate sensitivity and short elution

    time making it suitable as a regular method for pharmaceutical and clinical labs.Various HPLC methods have been reported for the quantitation of H1-receptor

    antagonists like cetirizine, buclizine, and levocetirizine in dosage formulations and

    human serum (Sultana et al., 2009; Arayne et al., 2005; Arayne et al., 2008;

    Gowekar et al., 2007). The aim of this study was to present the method which

    should be rapid, selective, linear, precise and sensitive, and should be less time-

    consuming (analysis time 3.784.15 min). The method presented in this article was

    validated according to the ICH guidelines (ICH guidelines Topic Q2 (R1)

    Validation of Analytical Procedures), and the low limit of quantification (LOQ)

    and limit of detection (LOD) values make it a better choice for the estimation offexofenadine in human serum. This validated method was also applied on different

    brands of fexofenadine available in Pakistan (60 mg tablets) which supports the

    analysis of fexofenadine in bulk, in dosage formulations, and in human serum for

    therapeutic purpose using HPLC.

    Experimental

    Reagents and chemicals

    Fexofenadine hydrochloride reference substance (99.6%) was obtained from Aventis

    Pharma (Pvt.) Limited, Pakistan. Monobasic potassium phosphate, potassium

    hydroxide, HPLC, and analytical grade solvent (methanol) were purchased from

    Merck (Germany), water for HPLC was prepared by double distillation and filtration

    through Millipore 0.45 lm membrane filter (Millipore, Milford, MA, USA), and

    degassed with Branson 3200 ultrasonic bath. Three commercial preparations, Fexet

    tablet (Getz Pharma Pakistan (Pvt.) Ltd., Fexofast tablet (Platinum Pharmaceuticals

    (Pvt.) Ltd., and Telfast tablet (Sanofi Aventis Pharma (Pvt.) Ltd., Pakistan, all

    containing 60 mg fexofenadine/tablet were assayed.

    Chromatographic system and conditions

    Two chromatographs were used (a) Shimadzu liquid chromatograph equipped

    with a model LC-10AVP isocratic pump, and model SPD-10AVP UV detector.

    Detection was made at 218 nm. CLASS-GC chromatography software. (b) The HPLC

    Chromatograph Series 200 was produced by Perkin Elmer, USA, and equipped

    with autosampler, pump, UV/VIS detector, Totalchrom Navigator Version 6.3.1.0504

    software, interface 600 series LINK were used. Five different brands of station-

    ary phase were used: (i) 250 9 4.6 i.d. mm KROMASIL 100-5 C-18 column

    (5 lm particle size) HICHROM (UK), (ii) 250 mm 9 4.6 mm i.d. 9 5 lm parti-

    cle size GEMINI C18 Phenomenex (USA), (iii) 250 mm 9 4.6 mm i.d. 9

    5 lm particle size NUCLEOSIL 100-5 C18 MachereyNagel (Germany),

    Med Chem Res

  • 7/31/2019 RP-HPLC Method for the Quantitative Determination of Fexofenidine HCl in Coated Tablets and Human

    4/13

    (iv) 250 mm 9 4.6 mm i.d. 9 5 lm Discovery C18 SUPELCO (USA), (v)

    250 mm 9 4.6 mm i.d. 9 5 lm HypersilTM ODS column Thermo Electron Corpo-

    ration (UK). On both systems, mobile phase run isocratically, the mobile phase was

    prepared by mixing methanol and 6.8 g monobasic potassium phosphate in 1000-ml

    water and pH 7.4 (adjusted with potassium hydroxide), 35:65 (v/v), respectively. Theinjection volume was 20 ll and the run time was 10 min. The mobile phase was

    filtered using a 0.45-lm membrane filter (Millipore) and degassed with Branson 3200

    ultrasonic bath. The mobile phase flow rate was 1.0 ml min-1. Injections were carried

    out using a 20-ll loop at room temperature (20 2C).

    Preparation of stock standard solution

    A 10 mg amount of fexofenadine reference substance was accurately weighed,

    dissolved in mobile phase, and diluted to volume in a 100-ml volumetric flask.Standard solution was obtained by diluting the above solution with mobile phase to

    a concentration of 10 lg ml-1.

    Preparation of sample solutions

    All the three commercially available brands of fexofenadine were analyzed

    separately by preparing a composite of 20 tablets by grinding them to a fine,

    uniform size powder, using mortar and pestle. After calculating the average tablet

    weight, amount corresponding to 10 mg of fexofenadine was accurately weighedand quantitatively transferred into a 100-ml volumetric flask. Approximately, 60 ml

    mobile phase was added and the solution was shaken mechanically for 15 min, then

    flask was made up to volume with mobile phase, and mixed. After filtration through

    Millipore 0.45 lm membrane, the solution was diluted with mobile phase to a

    concentration of 10 lg ml-1.

    Extraction and storage of blood samples

    Fresh blood samples from healthy volunteers were collected daily, centrifuged, and

    separated, and then 10 ml of acetonitrile was added in 1.0 ml of plasma and

    vortexes for 1 min, centrifuged for 10 min at 10,000 rpm. After that, supernatant

    was filtered by 0.45-lm pore size membrane filter. An aliquot serum sample was

    prepared with fexofenadine hydrochloride to achieve final concentration; the serum

    was kept at -20C until analyzed. Human plasma samples were spiked with

    fexofenadine with mobile phase to 15 lg ml-1 final solution concentrations.

    Result and discussion

    Method validation

    The method was validated according to the ICH guidelines (ICH guidelines

    Topic Q2 (R1) Validation of Analytical Procedures). The following validation

    Med Chem Res

  • 7/31/2019 RP-HPLC Method for the Quantitative Determination of Fexofenidine HCl in Coated Tablets and Human

    5/13

    characteristics were addressed: linearity, accuracy, precision, and specificity, LOD,

    LOQ, robustness, and their chromatograms are illustrated in Figs. 2, 3, 4, 5, 6, 7, 8,

    9, and 10.

    Fig. 2 Chromatogram of 10 lg ml-1 of fexofenadine reference standard

    Fig. 3 Chromatogram of Fexet tablet placebo

    Fig. 4 Chromatogram of Fexet tablet contains 10 lg ml-1 of fexofenadine

    Med Chem Res

  • 7/31/2019 RP-HPLC Method for the Quantitative Determination of Fexofenidine HCl in Coated Tablets and Human

    6/13

    Fig. 5 Chromatogram of Fexofast tablet placebo

    Fig. 6 Chromatogram of Fexofast tablet contains 10 lg ml-1 of fexofenadine

    Fig. 7 Chromatogram of Telfast tablet placebo

    Med Chem Res

  • 7/31/2019 RP-HPLC Method for the Quantitative Determination of Fexofenidine HCl in Coated Tablets and Human

    7/13

    Fig. 8 Chromatogram of Telfast tablet contains 10 lg ml-1 of fexofenadine

    Fig. 9 Chromatogram of blank plasma sample from healthy volunteer

    Fig. 10 Chromatogram of plasma sample spiked with 2.0 lg ml-1 of the fexofenadine

    Med Chem Res

  • 7/31/2019 RP-HPLC Method for the Quantitative Determination of Fexofenidine HCl in Coated Tablets and Human

    8/13

    System suitability

    Having optimized the efficiency of a chromatographic separation, the quality of

    chromatography was monitored by applying the following system suitability tests:

    capacity factor, tailing factor, and of theoretical plates. The system suitabilitymethod acceptance criteria set in each validation run were capacity factor[2.0,

    tailing factor B 2.0, and theoretical plates[2000 (CDER: Center for Drug

    Evaluation and Research, 1994). In all cases, the RSD for the analyte peak area

    for two consecutive injections was\2.0%.

    LOD and LOQ

    The LOD and LOQ studies for the developed procedure were performed on samples

    containing very low concentrations of analytes on two different commercial brandsof detectors according to the ICH guidelines as shown in Table 1. According to the

    visual evaluation method, LOD was uttered by establishing the minimum level at

    which the analyte can be reliably detected. LOQ was considered as the lowest

    concentration of analytes in standards that can be reproducibly measured with

    acceptable accuracy and precision.

    Linearity

    Standard curves were constructed daily, for three consecutive days, using fivestandard concentrations in a range 515 lg ml-1 for fexofenadine. This concen-

    tration range corresponds to 50150% w/w levels of the nominal analytical

    concentration (Table 2). The linearity of peak area responses versus concentrations

    was demonstrated by linear least square regression analysis. The regression equation

    and correlation coefficient (r) were y = 24340x ? 120 (y peak area, x concentra-

    tion) and 0.9999, respectively.

    Accuracy

    The accuracy of the method was evaluated by determination of the recovery of

    fexofenadine on 3 days at three levels of concentrations. Commercial preparation C

    Telfast tablet (60 mg) was spiked with fexofenadine standard solution, corre-

    sponding to 50150% of the nominal analytical concentration (10 lg ml-1). The

    results showed good recoveries ranging from 98.77 to 101.45%. The mean recovery

    data obtained for each level as well as for all levels combined (Table 2) were within

    Table 1 LOD and LOQ on different brands of detectors

    Shimadzu SPD-10AVP

    UV detector (ng ml-1

    )

    Perkin Elmer UV/VIS

    detector Series 200 (ng ml-1

    )

    LOD 20 10

    LOQ 35 20

    Med Chem Res

  • 7/31/2019 RP-HPLC Method for the Quantitative Determination of Fexofenidine HCl in Coated Tablets and Human

    9/13

    2.0% of the label claim for the active substance with an RSD\2.0%, which

    satisfied the acceptance criteria set for the study.

    Precision

    Precision of the method was determined by measuring the repeatability (intra-day

    precision) and intermediate precision (inter-day precision), both expressed as%RSD. The repeatability was evaluated by assay of six samples from each

    pharmaceutical commercial preparation (Table 3), at the same concentration

    (10 lg ml-1), on the same day. The intermediate precision was calculated from

    the results obtained on three different days.

    Specificity

    Fexet tablet contains 60 mg of the drug and the following excipients: pregela-

    tinized starch, lactose, croscarmellose sodium, and microcrystalline cellulose.

    Fexofast tablet contains 60 mg of the drug and the following excipients: lactose,

    croscarmellose sodium, and microcrystalline cellulose. Telfast tablet contains

    Table 2 Accuracy and linearity of method determined by recovery of fexofenadine from tablets

    solutions spiked with standard solution

    5 lg ml-1

    50%

    7.5 lg ml-1

    75%

    10 lg ml-1

    100%

    12.5 lg ml-1

    125%

    15 lg ml-1

    150%

    Day 1 99.98 98.77 100.21 101.45 99.88

    Day 2 99.96 98.88 100.08 100.33 100.60

    Day 3 100.24 98.78 100.78 100.44 99.96

    Mean (n = 3) 100.06 99.14 100.36 100.74 100.15

    %RSD 0.15 0.54 0.37 0.67 0.39

    Total mean

    (n = 15)

    100.022

    %RSD 0.743

    Telfast

    tablet 60 mg (Sanofi Aventis Pharma (Pvt) Ltd.) were spiked with fexofenadine standardsolution, corresponding to 50150% of the nominal analytical concentration (10 lg ml-1)

    Table 3 Intra-day (n = 6) and inter-day (n = 18) precision results from each commercial preparation

    tablets assay on three consecutive days

    Day Fexet tablet (60 mg) Fexofast tablet (60 mg) Telfast tablet (60 mg)

    Intra-day Inter-day Intra-day Inter-day Intra-day Inter-day

    Mean %RSD Mean %RSD Mean %RSD Mean %RSD Mean %RSD Mean %RSD

    1 99.94 0.49 99.96 0.45 99.91 0.65 99.71 0.66 99.89 0.52 99.81 0.77

    2 99.97 0.44 99.93 0.56 99.82 0.55

    3 99.87 0.24 99.89 0.22 99.80 0.67

    Med Chem Res

  • 7/31/2019 RP-HPLC Method for the Quantitative Determination of Fexofenidine HCl in Coated Tablets and Human

    10/13

    60 mg of the drug and the following excipients: lactose, croscarmellose sodium, and

    microcrystalline cellulose. Specificity was evaluated by preparing placebo samples

    of each placebo of the commercial formulation of tablets containing the same

    excipients and specificity was performed by preparing placebo tablets of the

    commercial products. The solutions were prepared using the same proceduredescribed for the sample solutions and injected thrice (Table 3) and the charac-

    teristic chromatograms are shown in Figs. 28.

    Robustness

    International variations in liquid chromatographic conditions were used to evaluate

    the robustness of the assay method. In this study, the chromatographic parameters

    monitored were retention time, area, capacity factor, tailing factor, and theoretical

    plates. The robustness acceptance criteria set in the validation was the sameestablished on system suitability test described earlier.

    Different brand of chromatographic columns

    Three injections of Telfast tablet (Sanofi Aventis Pharma (Pvt.) Ltd., containing

    60 mg fexofenadine/tablet) solutions having a concentration of 10 lg ml-1 were

    injected on each brand of column to evaluate robustness (Table 4).

    Different instruments

    Three injections of each commercial preparation were injected on two different

    brands of liquid chromatograph by two different analysts data established in

    Table 5.

    Different personnel

    Variation in results due to different analysts data is established by taking the %RSD

    of recovered % as shown in Table 5, which is found 0.28 and 0.45 for analysts A

    and B, respectively.

    Table 4 Chromatographic parameters of robustness evaluation

    Different columns brands

    (n = 3)

    Retention

    time

    Peak

    area

    Capacity

    factor

    Resolutiona Tailing

    factor

    Theoretical

    plates

    HICHROM KROMASIL 3.78 242798 3.15 3.12 1.21 2366

    Phenomenex GEMINI 3.80 243375 1.68 3.48 1.08 3365

    MachereyNagel

    NUCLEOSIL4.15 241624 2.39 3.54 1.02 3687

    SUPELCO Discovery 4.12 239963 2.17 3.02 1.34 2489

    Thermo Electron Corporation

    HypersilTM3.98 241840 2.36 3.21 1.22 2562

    a In relation to the nearest peak

    Med Chem Res

  • 7/31/2019 RP-HPLC Method for the Quantitative Determination of Fexofenidine HCl in Coated Tablets and Human

    11/13

    Recovery and regression characteristics of fexofenadine, in human serum

    Different concentrations of human plasma samples (15 lg ml-1) were linear,

    accurate, precise, and selective by running three replicates of each concentration

    measured for 5 days; typical chromatogram for the concentration of 2 lg ml-1 is

    shown in Fig. 10. The mean recoveries, coefficient of variation, LOD, and

    quantification values are summarized in Table 6.

    Conclusions

    The LOD and LOQ were carried out on two different commercial brands of detectors

    according to the ICH guidelines. The intra- and inter-day precision studies showed

    good reproducibility with coefficients of variation\0.77%. The ruggedness of the

    method was determined by carrying out the experiment on different instruments like

    Shimadzu HPLC (LC-10), Perkin Elmer Series 200 HPLC by different operators

    using different columns. Robustness of the method was determined by making slight

    changes in the chromatographic conditions. No marked changes in the chromato-

    grams were observed.

    Thus, the proposed RP-HPLC method for the estimation of fexofenadine in

    dosage forms is accurate, precise, linear, rugged, robust, simple, and rapid. Hence,

    Table 5 Different instrument and different personnel variation

    Commercial

    preparation

    (n = 3)

    Shimadzu liquid chromatograph Perkin Elmer Series 200 liquid

    chromatograph

    %

    Difference

    in

    instrumentsAnalyst A Analyst B

    Found (mg) %RSD %Recovery Found (mg) %RSD %Recovery

    Fexet 59.79 1.1245 1.12 99.65 60.10 1.0132 1.01 100.16 0.51

    Fexofast 59.89 1.0325 1.03 99.81 59.60 1.2261 1.22 99.33 0.48

    Telfast 60.12 1.0576 1.05 100.20 60.05 1.0463 1.04 100.08 0.12

    Table 6 Recovery of fexofenadine from human serum (n = 3)

    S. no Concentration(lg ml-1

    ) Mean(lg ml-1

    ) (%) Recovery Relativeerror (%) CV (%)

    1 1 0.995 99.5 0.5 0.9

    2 2 2.0 100 -0.15 1.21

    3 3 3.009 100.3 -0.3 0.8

    4 4 3.998 99.95 0.05 0.68

    5 5 5.085 101.7 -1.7 0.16

    Correlation coefficient (r2) 0.9997

    LOD 0.03

    LOQ 0.1

    Med Chem Res

  • 7/31/2019 RP-HPLC Method for the Quantitative Determination of Fexofenidine HCl in Coated Tablets and Human

    12/13

    the RP-HPLC method is suitable for the quality control and routine testing of the

    raw materials, different dosage forms, different formulations, and in human serum.

    References

    Arayne MS, Sultana N, Siddiqui FA (2005) Determination and quantification of cetirizine HCl in dosage

    formulations by RP-HPLC. Pak J Pharm Sci 18:711

    Arayne MS, Sultana N, Nawaz M (2008) Simultaneous quantification of cefpirome and cetirizine or

    levocetirizine in pharmaceutical formulations and human plasma by RP-HPLC. J Anal Chem

    63:881887

    Barnes CL, McKenzie CA, Webster KD, Poinsett-Holmes K (1993) Cetirizine: a new, non-sedating

    antihistamine. Ann Pharmacother 27:464470

    British Pharmacoepia (2000) Her Majesty Stationary Office, pp 342343

    Caballero E, Ocana I, Azanza JR, Sadaba B (1999) Fexofenadine: a antihistaminic review of its practical

    characteristics. Rev Med Univ Navarra 43:9397

    CDER: Center for Drug Evaluation and Research (1994) Reviewer guidance for validation ofchromatographic methods. Disponvelem: http://www.fda.gov/cder/guidance/cmc3.pdf

    Drescher S, Schaeffeler E, Hitzl M, Hofmann U, Schwab M, Brinkmann U, Eichelbaum M, Fromm MF

    (2002) MDR1 gene polymorphisms and disposition of the P-glycoprotein substrate fexofenadine. Br

    J Clin Pharmacol 53:526534

    Fu I, Woolf EJ, Matuszewski BK (2004) Determination of fexofenadine in human plasma using 96-well

    solid phase extraction and HPLC with tandem mass spectrometric detection. J Pharm Biomed Anal

    35:837846

    Gazy AA, Mahgoub H, El-Yazbi FA, El-Sayed MA, Youssef RM (2002) Determination of some

    histamine H1 receptor antagonists in dosage forms. J Pharm Biomed Anal 30:859867

    Gergov M, Robson JN, Ojanpera I, Heinonen OP, Vuori E (2001) Simultaneous screening and

    quantification of 18 antihistamine drugs in blood by liquid chromatography ionspray tandem massspectrometry. Forensic Sci Int 121:108115

    Gowekar NM, Pande VV, Kasture AV, Tekade AR, Chandorkar JG (2007) Spectrophotometric

    estimation of ambroxol and cetirizine hydrochloride from tablet dosage form. Pak J Pharm Sci

    20:250251

    Hamman MA, Bruce MA, Haehner-Daniels BD, Hall SD (2001) The effect of rifampin administration on

    the disposition of fexofenadine. Clin Pharmacol Ther 69:114121

    ICH guidelines Topic Q2 (R1) Validation of analytical procedures: methodology. www.emea.europa.eu/

    htms/human/ich/ichquality.htm.

    Inomata N, Tatewaki S, Ikezawa Z (2009) Multiple H1-antihistamine-induced urticaria. J Dermatol

    36:224227

    Karakus S, Kucukguzel I, Kucukguzel SG (2008) Development and validation of a rapid RP-HPLC

    method for the determination of cetirizine or fexofenadine with pseudoephedrine in binarypharmaceutical dosage forms. J Pharm Biomed Anal 46:295302 [Epub 2007 October 22]

    Markham A, Wagstaff AJ (1998) Fexofenadine. Drugs 55:269274 discussion 275276

    Mattila MJ, Paakkari I (1999) Variations among non-sedating antihistamines: are there real differences?

    Eur J Clin Pharmacol 55:8593

    Milne RW, Larsen LA, Argensen KLJ, Bastlund J, Stretch GR, Evans AM (2000) Hepatic disposition of

    fexofenadine: influence of the transport inhibitors erythromycin and dibromosulphothalein. Pharm

    Res 17:15111515

    Naidong W, Shou WZ, Addison T, Maleki S, Jiang X (2002) Liquid chromatography/tandem mass

    spectrometric bioanalysis using normal-phase columns with aqueous/organic mobile phasesa

    novel approach of eliminating evaporation and reconstitution steps in 96-well SPE. Rapid Commun

    Mass Spectrom 16:19651975

    Pathak SM, Kumar AR, Musmade P, Udupa N (2008) A simple and rapid high performance liquid

    chromatographic method with fluorescence detection for the estimation of fexofenadine in rat

    plasmaapplication to preclinical pharmacokinetics. Talanta 76:338346 [Epub 2008 March 13]

    Pratt C, Brown AM, Rampe D, Mason J, Russell T, Reynolds R, Ahlbrandt R (1999) Cardiovascular

    safety of fexofenadine HCl. Clin Exp Allergy 29:212216

    Med Chem Res

    http://www.fda.gov/cder/guidance/cmc3.pdfhttp://www.emea.europa.eu/htms/human/ich/ichquality.htmhttp://www.emea.europa.eu/htms/human/ich/ichquality.htmhttp://www.emea.europa.eu/htms/human/ich/ichquality.htmhttp://www.emea.europa.eu/htms/human/ich/ichquality.htmhttp://www.fda.gov/cder/guidance/cmc3.pdf
  • 7/31/2019 RP-HPLC Method for the Quantitative Determination of Fexofenidine HCl in Coated Tablets and Human

    13/13

    Radhakrishna T, Om Reddy G (2002) Simultaneous determination of fexofenadine and its related

    compounds by HPLC. J Pharm Biomed Anal 29:681690

    Russell T, Stoltz M, Weir S (1998) Pharmacokinetics, pharmacodynamics, and tolerance of single- and

    multiple-dose fexofenadine hydrochloride in healthy male volunteers. Clin Pharmacol Ther 64:612621

    Simpson K, Jarvis B (2000) Fexofenadine: a review of its use in the management of seasonal allergic

    rhinitis and chronic idiopathic urticaria. Drugs 59:301321Sultana N, Arayne MS, Shamshad H (2009) In vitro studies of the interaction between cetirizine and H2

    receptor antagonists using spectrophotometry and reversed-phase high-performance liquid chroma-

    tography. Med Chem Res. doi:10.1007/s00044-009-9204-x

    Uno T, Yasui-Furukori N, Takahata T, Sugawara K, Tateishi T (2004) Liquid chromatographic

    determination of fexofenadine in human plasma with fluorescence detection. J Pharm Biomed Anal

    35:937942

    Med Chem Res

    http://dx.doi.org/10.1007/s00044-009-9204-xhttp://dx.doi.org/10.1007/s00044-009-9204-x