round robin testing for mode i interlaminar …...j~rer, vol. 15, no. 4, winter 1993, pp. 269-281....

13
T. Kevin O'Brien ~ and Roderick H. Martin 2 Round Robin Testing for Mode I Interlaminar Toughness of Composite Materials Fracture REFERENCE: O'Brien, T. K. and Martin, R. H., "Round Robin Testing for Mode I Interlaminar Fracture Toughness of Com- posite Materials," Journal of Composites Technology & Research, J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round robin" test programs for measuring the Mode I interlaminar fracture toughness of advanced fiber-reinforced com- posite materials. Double cantilever beam (DCB) tests were con- ducted by participants in ASTM Committee D-30 on High Modulus Fibers and Their Composites and by representatives of the European Group on Fracture (EGF) and the Japanese Industrial Standards Group (JIS). DCB tests were performed on three AS4 carbon fiber- reinforced composite materials: AS4/3501-6 with a brittle epoxy ma- trix, AS4/BP907 with a tough epoxy matrix, and AS4/PEEK with a tough thermoplastic matrix. Difficulties encountered in manufactur- ing panels, as well as conducting the tests, are discussed. Critical issues that developed during the course of the testing are highlighted. Results of the round robin testing used to determine the precision of the ASTM DCB test standard are summarized. KEYWORDS: composite materials, double cantilever beam (DCB), interlaminar fracture toughness, delamination, standard test method The data contained herein were generated by voluntary par- ticipants using the double cantilever beam (DCB) test (Fig. 1). The DCB test consists of a unidirectional fiber-reinforced lam- inate, manufactured with a thin insert implanted at the mid-plane near one end to simulate a sharp crack, and loaded such that the delamination forms at the insert in an opening mode (Mode I). Specimens were cut from panels manufactured using prepreg voluntarily supplied by several material suppliers. A list of par- ticipants is included in the Appendix. A chronology of the ac- tivity is documented in Ref 1 in the form of excerpts from ASTM meeting minutes from 1986 to 1992. Early discussions (prior to 1985) resulted in limiting the DCB test to 0 ° unidirectional laminates to prevent the initial delami- nation from branching to interfaces away from the midplane [2]. The width-tapered DCB configuration [3] was abandoned be- cause of the added complexity of machining this configuration and the tendency for 0 ° unidirectional width tapered laminates to split at the juncture between the narrow and tapered regions. Furthermore, the phenomenon of fiber bridging between the two 0 ° plies on either side of the delamination [4--6] was first ob- served during this phase. Since 1986, five distinct rounds of testing were conducted. 1U.S. Army Research Laboratory, Vehicle Structures Directorate, NASA Langley Research Center, Hampton, VA 23681. 2Analytical Services and Materials, Inc., Hampton, VA 23666. The first round yielded useful data for AS4/BP907. However, little data were obtained for the other two materials because of problems that were experienced in obtaining sufficiently thin or completely disbonded inserts for starting the delaminations. The second round of testing yielded useful results for AS4/3501-6, although with fewer labs participating. However, problems were again encountered with the manufacture of AS4/PEEK panels with good quality inserts. The third round of testing was con- ducted in conjunction with the European Group on Fracture (EGF) and the Japanese Industrial Standards (JIS) group. Al- though sufficient AS4/PEEK panels were manufactured to con- duct a thorough test matrix, specimens obtained from these pan- els had problems with torn and folded aluminum inserts. The fourth round of testing, consisting of static tests from a DCB fatigue round robin, yielded more data on AS4~EEK specimens with the Kapton inserts. The fifth round of testing yielded suf- ficient data on AS4/PEEK specimens with thin Upilex inserts to determine the precision of the DCB standard test method. Background The DCB test consists of a unidirectional continuous fiber- reinforced laminate, manufactured with a thin insert at the mid- plane near one end, and loaded such that the delamination forms at the insert as a Mode I, or opening mode, fracture. The param- eters that were investigated in the round robins were: (1) the method of introducing the opening load, (2) specimen thickness, and (3) insert type and thickness. Figure 1 shows two configurations of the DCB where the load is introduced via piano hinges (Fig. la) or loading blocks (Fig. lb). A variation on the loading block configuratign designated "T-tabs" was also used (Fig. lc). In the first round of testing, load introduction was accomplished using either piano hinges or T-tabs. Piano hinges were used exclusively in the second round. By the third round, correction factors for loading blocks and tabs had been developed, with specific guidelines for when they were required. Hence, both piano hinges (Fig. la) and end load- ing blocks (Fig. lb) were used in rounds 3 through 5. Specimens in the first two rounds were 25-mm (1-in.) wide. In the third round, however, 20-mm-wide specimens were tested, but a limited number of 12.5, 25.0, and 37.5 mm (0.5, 1.0, and 1.5 in.) wide specimens were also tested. Because no significant width effect was discovered (see Ref 1), 20 to 25-mm wide specimens were tested in the fourth and fifth rounds. In the first and second rounds, specimens consisted of 24-ply laminates for the AS4/3501-6 and AS4/BP907 tests, and 36-ply © 1993 by the American Society for Testing and Materials 269 Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014 Downloaded/printed by NULL pursuant to License Agreement. No further reproductions authorized.

Upload: others

Post on 18-Jul-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Round Robin Testing for Mode I Interlaminar …...J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round

T. K e v i n O ' B r i e n ~ a n d R o d e r i c k H. M a r t i n 2

Round Robin Testing for Mode I Interlaminar Toughness of Composite Materials

Fracture

REFERENCE: O'Brien, T. K. and Martin, R. H., "Round Robin Testing for Mode I Interlaminar Fracture Toughness of Com- posite Materials," Journal of Composites Technology & Research, J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281.

ABSTRACt. This report summarizes the results of several inter- laboratory "round robin" test programs for measuring the Mode I interlaminar fracture toughness of advanced fiber-reinforced com- posite materials. Double cantilever beam (DCB) tests were con- ducted by participants in ASTM Committee D-30 on High Modulus Fibers and Their Composites and by representatives of the European Group on Fracture (EGF) and the Japanese Industrial Standards Group (JIS). DCB tests were performed on three AS4 carbon fiber- reinforced composite materials: AS4/3501-6 with a brittle epoxy ma- trix, AS4/BP907 with a tough epoxy matrix, and AS4/PEEK with a tough thermoplastic matrix. Difficulties encountered in manufactur- ing panels, as well as conducting the tests, are discussed. Critical issues that developed during the course of the testing are highlighted. Results of the round robin testing used to determine the precision of the ASTM DCB test standard are summarized.

KEYWORDS: composite materials, double cantilever beam (DCB), interlaminar fracture toughness, delamination, standard test method

The data contained herein were generated by voluntary par- ticipants using the double cantilever beam (DCB) test (Fig. 1). The DCB test consists of a unidirectional fiber-reinforced lam- inate, manufactured with a thin insert implanted at the mid-plane near one end to simulate a sharp crack, and loaded such that the delamination forms at the insert in an opening mode (Mode I). Specimens were cut from panels manufactured using prepreg voluntarily supplied by several material suppliers. A list of par- ticipants is included in the Appendix. A chronology of the ac- tivity is documented in Ref 1 in the form of excerpts from ASTM meeting minutes from 1986 to 1992.

Early discussions (prior to 1985) resulted in limiting the DCB test to 0 ° unidirectional laminates to prevent the initial delami- nation from branching to interfaces away from the midplane [2]. The width-tapered DCB configuration [3] was abandoned be- cause of the added complexity of machining this configuration and the tendency for 0 ° unidirectional width tapered laminates to split at the juncture between the narrow and tapered regions. Furthermore, the phenomenon of fiber bridging between the two 0 ° plies on either side of the delamination [4--6] was first ob- served during this phase.

Since 1986, five distinct rounds of testing were conducted.

1U.S. Army Research Laboratory, Vehicle Structures Directorate, NASA Langley Research Center, Hampton, VA 23681.

2Analytical Services and Materials, Inc., Hampton, VA 23666.

The first round yielded useful data for AS4/BP907. However, little data were obtained for the other two materials because of problems that were experienced in obtaining sufficiently thin or completely disbonded inserts for starting the delaminations. The second round of testing yielded useful results for AS4/3501-6, although with fewer labs participating. However, problems were again encountered with the manufacture of AS4/PEEK panels with good quality inserts. The third round of testing was con- ducted in conjunction with the European Group on Fracture (EGF) and the Japanese Industrial Standards (JIS) group. Al- though sufficient AS4/PEEK panels were manufactured to con- duct a thorough test matrix, specimens obtained from these pan- els had problems with torn and folded aluminum inserts. The fourth round of testing, consisting of static tests from a DCB fatigue round robin, yielded more data on A S 4 ~ E E K specimens with the Kapton inserts. The fifth round of testing yielded suf- ficient data on AS4/PEEK specimens with thin Upilex inserts to determine the precision of the DCB standard test method.

Background

The DCB test consists of a unidirectional continuous fiber- reinforced laminate, manufactured with a thin insert at the mid- plane near one end, and loaded such that the delamination forms at the insert as a Mode I, or opening mode, fracture. The param- eters that were investigated in the round robins were: (1) the method of introducing the opening load, (2) specimen thickness, and (3) insert type and thickness.

Figure 1 shows two configurations of the DCB where the load is introduced via piano hinges (Fig. la) or loading blocks (Fig. lb). A variation on the loading block configuratign designated "T-tabs" was also used (Fig. lc). In the first round of testing, load introduction was accomplished using either piano hinges or T-tabs. Piano hinges were used exclusively in the second round. By the third round, correction factors for loading blocks and tabs had been developed, with specific guidelines for when they were required. Hence, both piano hinges (Fig. la) and end load- ing blocks (Fig. lb) were used in rounds 3 through 5.

Specimens in the first two rounds were 25-mm (1-in.) wide. In the third round, however, 20-mm-wide specimens were tested, but a limited number of 12.5, 25.0, and 37.5 mm (0.5, 1.0, and 1.5 in.) wide specimens were also tested. Because no significant width effect was discovered (see Ref 1), 20 to 25-mm wide specimens were tested in the fourth and fifth rounds.

In the first and second rounds, specimens consisted of 24-ply laminates for the AS4/3501-6 and AS4/BP907 tests, and 36-ply

© 1993 by the American Society for Testing and Materials 269

Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014Downloaded/printed byNULL pursuant to License Agreement. No further reproductions authorized.

Page 2: Round Robin Testing for Mode I Interlaminar …...J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round

270 JOURNAL OF COMPOSITES TECHNOLOGY & RESEARCH

Adhesive,

Insert

(a) with piano hinges (b) with loading blocks

Aluminum

1 extensometer

(c) with T-tabs

FIG. 1-DCB specimen configurations.

FIG. 2-Box plot showing G,, round robin results for AS4/BP907 DCB tests using piano hinges.

laminates for the AS4JPEEK tests, with a nominal ply thickness of 0.127 mm (0.005 in.). By the third round, guidelines for spec- imen thickness and correction factors for geometric nonlinearity were available. Hence, tests on 24-ply AS4/PEEK specimens were conducted in rounds 3 through 5.

The most sensitive parameter that was examined was the type and thickness of the insert used to start the delamination. Be- cause of the fiber bridging that develops in the unidirectional

FIG. 3-Box plot showing G,, round robin results for AS4iBP907 DCB tests using T-tabs.

DCB specimen after the delamination grows from the end of the insert, the value of GIc measured at the initiation of delamination from the end of the insert was considered the only measured value representative of the interlaminar fracture toughness of the material being tested (see Ref I). In the first round, 26-pm (1.0 mil = 0.001 in.) Kapton film inserts were used for the AS413501- 6 and AS4iBP907 specimens. However, data could only be ob- tained for the AS4jBP907, because the Kapton film layed up in

Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014Downloaded/printed byNULL pursuant to License Agreement. No further reproductions authorized.

Page 3: Round Robin Testing for Mode I Interlaminar …...J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round

O'BRIEN AND MARTIN ON INTERLAMINAR FRACTURE TOUGHNESS 271

TABLE 1---Summary of ASTM round robin data.

Load Avg. Mean (CV), (CV)R Round Material Intro # Labs Tests/Lab Insert G~, kJ/m z S, % SR %

1 AS4/BP907 Hinges 9 3 25 ~m Kapton 0.400* 0.028 7.0 0.077 19.3 AS4~P907 T-Tabs 7 3 25 txm Kapton 0.410" 0.041 10.0 0 .052 12.7

2 AS4/3501-6 Hinges 3 3 13 ixm Kapton 0.085* 0.015 17 .6 0.014 16.5 AS4/PEEK Hinges 2 4 13 Ixm Kapton 1.340" 0.139 10 .4 0.211 15.7

*Visual values using compliance calibration (Berry's Method).

the AS4/3501-6 specimens was not sprayed with a mold release agent before curing. The absence of release agent resulted in specimens that were intermittently bonded in the insert area and, hence, no useful data were obtained from these specimens. The AS4/PEEK specimens tested in the first round had 39-p,m (1.5- mil) thick folded aluminum foil inserts yielding a total insert thickness of 78 ~m (3.0 mil). These inserts proved to be too thick to measure a useful initiation value and, hence, the first round AS4~EEK data were of limited value.

In the second round, four distinct Kapton insert types were employed for AS4/3501-6 and AS4/PEEK specimens, resulting in three insert thicknesses. Inserts were either (1) 13-1xm (0.5- mil) single layers sprayed with a mold release agent; (2) 13-txm (0.5-mil) layers folded in two to achieve a 26-txm (1.0-mil) thickness, or 26-p,m (1.0-mil) single layers sprayed with a mold release agent; or (3) 26-1~m (1.0-rail) layers folded in two to achieve a 52-~m (2.0-mil) thickness. Data from this round in- dicated that the 13-1~m (0.5-mil) sprayed insert consistently yielded the most reasonable and conservative value of Gjc for all materials. However, the thinner inserts consistently yielded lower Gk values, without giving an indication that a minimum plateau had been obtained, as was observed in the literature for glass epoxy laminates [7].

In the third round, both 7-p,m (0.25-mil) and 13-1~m (0.5-mil) aluminum inserts were sprayed with a mold release agent and implanted in AS4/PEEK panels. The panels for the ASTM and JIS participants were X-rayed to examine the conditions of the inserts. Unfortunately, these radiographs indicated that many tears and folds were present in the aluminum inserts. Only spec- imens that appeared to be free of insert tears and folds in the radiographs were distributed to the ASTM and JIS participants, thereby limiting the number of specimens available from each panel. Unfortunately, even the specimens with inserts, which ap- peared straight and fiat in the radiographs, exhibited uncharac- teristic R-curves and yielded questionable initiation G~ values. Examination of the polished edge of an untested specimen in- dicated a tendency for the aluminum inserts to fold or crimp, resulting in the formation of resin pockets (see Ref 1). The spec- imens sent to the EGF participants were not X-rayed before they were tested, but yielded similar results. These data were sum- marized separately.

In the fourth round, several labs generated static DCB results on AS4/PEEK specimens with 13-t~m (0.5-mil) sprayed Kapton inserts as part of an ASTM fatigue round robin. Attempts to manufacture AS4/3501-6 graphite epoxy laminates with Kapton inserts were unsuccessful for the same reasons previously noted.

In the fifth round, tests were conducted on AS4/PEEK spec- imens with both 7.5-1zm (0.25-mil) and 13-p.m (0.5-mil) Upilex inserts. This fifth round of testing yielded sufficient data on AS4~ PEEK specimens with thin Upilex inserts to determine the pre- cision of the DCB standard test method.

Results for each round of testing were summarized first in the

form of a "box plot." These box plots were used simply to show trends in central tendency for groups of variables. A box plot represents each plotted variable as a separate box with a dark line drawn inside showing the median value of the variable and the top and bottom of the box representing the limits of +25% and -25% of the variable population. Lines extending from the top and bottom of the box mark the maximum and minimum for each variable. Typically, a maximum of 20 vari- ables can be plotted in a box plot. For consistency, however, a box plot was used to show trends in central tendency for test matrices with more than 20 variables. This often resulted in iso- lated data points being shown discretely on the plot if they fell outside of the box. Mean Gzc values and standard deviations for individual labs were then compared using bar charts. Finally, results for each round of testing were summarized in tables.

Results from Round 1

Nine labs each received three specimens to test where the load was introduced using piano hinges. Seven labs each received three specimens to test where the load was introduced using T- tabs. A first draft test procedure was sent to each lab. The data were reduced using a compliance calibration technique com- monly known as Berry's method [8].

Figures 2 and 3 show the visually observed initiation G~. val- ues, measured from a 26-t~m (1.0-mil) Kapton insert, for the AS4/BP907 DCB specimens. Figure 4 shows a comparison of mean G~ values for the six labs that performed tests with both configurations. Figure 5 shows the standard deviation in the data

Mean

~c' kJ/m 2

1 L I I I I I

L 0.8

[ ] TAB

0.6 ~ Hinge

0.4

0.2 ) 0 I ~ 1

m rr"

O) t -

q

FIG. 4--Comparison of mean Gtc results for AS4/BP907 DCB spec- imens with piano hinges and T-tabs.

Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014Downloaded/printed byNULL pursuant to License Agreement. No further reproductions authorized.

Page 4: Round Robin Testing for Mode I Interlaminar …...J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round

272 JOURNAL OF COMPOSITES TECHNOLOGY & RESEARCH

0 . t 2 I I I I I

s . g ,

kJ/m2

0.1

0.08

0.06

0.04

0.02

0 :s

#-

• TAB

[ ] Hinge

D o O0 0 q

FIG. 5---Comt~rison of standard deviation BP907 DCB specimens with piano hinges and

in Gk results for AS4/ T-tabs.

0 . 3 L i i i

Gic,

k Jim 2

0.2

DELAMINATION ONSET

0.1

13 ~mS I I

13 am F 26 },tm S

INSERT TYPE

I

1 26 t.tm F

FIG. 6---Box plot showing Gn¢ round robin results for AS4/3501-6 DCB tests using different insert types.

obtained for these six labs. The data for all the tabs that per- formed tests are summarized in Table 1. Similar mean G,c values were obtained for both configurations. Also summarized in Table 1 are the variability in mean G~c values for the piano hinge and T-tab configurations in the form of standard deviations within a given laboratory, S , (a measure of repeatability) and the standard deviations between laboratories, SR, (a measure of reproducibil- ity). These measures of repeatability and reproducibility are re- quired to obtain an estimate of the precision of the test method as specified by ASTM E 691, Practice for Conducting an Inter- laboratory Study to Determine the Precision of a Test Method. The data are also summarized in Table 1 as coefficients of var- iation (CV) calculated by dividing the standard deviation by the mean G,c value. Coefficients of variation are tabulated corre- sponding to the repeatability, (CV),, and reproducibility, (CV)R.

Results from Round 2

Three labs each received three specimens to test where the load was introduced using piano hinges. Figure 6 shows the G,c values measured by visually observing delamination onset at the edge of DCB speCimens of AS4/3501-6 with 13-1~m (0.5-mil) sprayed, 13-1~m (0.5-rail) folded, 26-p,m (1.0-mil) sprayed, and 26-1xm (1.0-rail) folded Kapton inserts. The results indicate that the 13-1xm (0.5-mil) sprayed inserts yield the lowest mean val- ues; the 13-p~m (0.5-mil) folded, 26-1xm (1.0-mil) sprayed in- serts, both of which result in a 26-p~m (1.0-mil) insert thickness, yield higher mean values; and the 26-1xm (1.0-mil) folded in- serts, which result in a 52-p~m (2.0-mil) insert thickness, yield the highest mean values and have the greatest scatter. Figures 7 and 8 show the mean Gtc values and standard deviation, respec- tively, for each of the three labs that performed the tests. Figure 9 compares the G~ values measured from the 13-txm (0.5-mil) sprayed insert for the three labs. The mean Glc values measured from the 13-1xm (0.5-mil) sprayed insert for the three labs, and the statistical measures of repeatability within a given laboratory, S , and the reproducibility between laboratories, SR, are sum- marized in Table 1. Also summarized in Table 1 are the coeffi- dents of variation corresponding to repeatability within a given

0.5

0.4

~c' 0.3

kJ/m 2 0.2

0.1

1 • 13 p, rnS

[ ] 13 u, mF

[ ] 26 txmS

[ ] 26 I.tm F

DREP N IST NASA

FIG. 7---Comparison of mean G~ results for AS4/3501-6 DCB spec- imens with different insert types.

S.D.p

kJ/m 2

0.1

0.08

0.06

0.04

0.02

0

I

• 13 ~mS

[ ] 13 .amF

B 26 ~mS

[ ] 26 i, tmF

DI::L=P NIST NASA

FIG. 8--Comparison of standard deviation in G~c results for AS4~ 3501-6 DCB specimens with different insert types.

Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014Downloaded/printed byNULL pursuant to License Agreement. No further reproductions authorized.

Page 5: Round Robin Testing for Mode I Interlaminar …...J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round

O'BRIEN AND MARTIN ON INTERLAMtNAR FRACTURE TOUGHNESS 273

~C '

kJ/m 2

0.2

0.16

0.12

0,08

0.04

0

! I

13 ~m SPRAYED KAPTON INSERT

DREP NIST NASA

kJ/m 2

3

2,5

2

1.5

1

0.5

0

/ NIST

I

• 13 pmS [] 13 ~mF

D 26 ~mS

[] 26 p, mF

,ii NASA

FIG. 9---Box plot showing GI~ round robin results for AS4/3501-6 DCB tests 13-t~m sprayed Kapton inserts.

FIG. 11--Comparison of mean Glc results for AS4/PEEK DCB spec- imens with different insert types.

laboratory, (CV),, and the reproducibility between laboratories, (CV)R.

Figure 10 shows the Glc values measured by visually observ- ing delamination onset at the edge of DCB specimens of AS4/ PEEK with 13-p,m (0.5-rail) sprayed, 13-~m (0.5-rail) folded, 26-tim (1.0-rail) sprayed, and 26-p,m (1.0-mil) folded Kapton inserts. Because of difficulties manufacturing these panels, there were only enough specimens for two labs, with only one lab testing all four insert types. Each lab tested four specimens per insert type. The data from the two labs that performed the tests are included in Fig. 10. The results indicate that the 13-tim (0.5- rail) sprayed inserts yield the lowest mean values; the 13-p~m (0.5-rail) folded, 26-p~m (1.0-mil) sprayed inserts, both of which result in a 26-p~m (1.0-mil) insert thickness, yield slightly higher mean values; and the 26-tim (1.0-rail) folded inserts, which re- sult in a 52-p,m (2.0-rail) insert thickness, yield the highest mean values.

Figure 11 shows the mean G/~ values for the four insert con- figurations for the two labs that performed the tests. The mean G~ values measured from the 13-txm (0.5-mil) sprayed insert for the two labs, and the standard deviations for repeatability and

2.5 I I 1 "r

1.5 Gic,

kJ/m 2 1

0.5

©

I DELAMINATION ONSET

0 J- ~ 13 p,m S 13 I.tm F 26 ttm S 26 ttm F

INSERT TYPE

FIG. lO---Box plot showing GI¢ round robin results for AS4/PEEK DCB tests using different insert types.

reproducibility are shown in Table 1. The coefficients of varia- tion corresponding to repeatability, (CV),, and reproducibility, (CV)R, are also shown in Table 1. However, a larger data set was needed to obtain an accurate estimate of the repeatability and reproducibility between laboratories.

Unlike the tests on AS4/3501-6 and AS4/BP907, the load de- flection curves for the AS4/PEEK DCB tests became nonlinear before the delamination was visually observed to initiate from the insert on the edge of the specimen (Fig. 12). Hence, several different initiation measurements, as well as several different data reduction methods, were proposed for reducing data from DCB tests on AS4/PEEK in round 3. As a prelude to the third round, these initiation measurements and data reduction methods were used to plot the data generated on 13-i.Lm (0.5-mil) sprayed Kapton insert tests from addition tests conducted by three labs, each testing four specimens, during round 2.

Figure 13 shows Gtc values measured using the load at onset of nonlinearity (NL), the load at visual observation of delami- nation onset at the edge (VIS), and the load corresponding 'to a 5% offset in the initial compliance of the DCB specimen (5%).

I P

% i Dead weight support

(a) Brittle matrix 5

(b) Tough matrix

FIG. 12 Load displacement trace from typical DCB tests.

Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014Downloaded/printed byNULL pursuant to License Agreement. No further reproductions authorized.

Page 6: Round Robin Testing for Mode I Interlaminar …...J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round

2 7 4 J O U R N A L O F C O M P O S I T E S T E C H N O L O G Y & R E S E A R C H

,.5t ~c , 1

k Jim 2

0.5

0 NL

FIG. 13--Box plot showing Gl~ round robin values for AS4/PEEK DCB tests with 13-tun sprayed Kapton inserts.

I I I 4 - I I I I I I I I I I I I I I I ' " ;

3.5 III MBT T ~ ~ ~ 3 - [ ] BRY 13#mAl. insert

2.5 - N M C C

Mean 2 NLGIc' 1.5

MO Y

0.5

t 0

< ~

These data were reduced using a modified beam theory (MBT) technique [9]. The MBT technique yielded slightly lower mean Gk values than Berry's method (BRY) for the same test data (see Ref 1). Also shown in Fig. 13 are the plateau values of Gt~ (PLAT) corresponding to stabilized delamination growth in the presence of fiber bridging. The mean NL G,c value for the three labs that tested the 13-p,m (0.5-mil) sprayed Kapton insert spec- imens is listed in Table 2. Also shown in Table 2 are the coef- ficients of variation for repeatability and reproducibility. How- ever, a larger data set was needed to obtain an accurate estimate of the repeatability and reproducibility between laboratories.

Results from Round 3

International Round Robin (ASTM and JIS)

Tests performed by ASTM and JIS participants using speci- mens that were X-rayed and appeared to have no tears or folds in the aluminum inserts are summarized first. The 13-1~m (0.5- mil) insert specimens were tested by 16 labs, whereas the 7-1xm (0.25-mil) specimens were tested by five labs. Each lab received four specimens to test per insert thickness.

Figure 14 shows the mean NL G~c values for the 16 labs that tested the 13-1~m aluminum insert specimens. Figure 15 shows the mean NL Gic values for the five labs that tested the 7-p,m aluminum insert specimens. The data were reduced using three data reduction methods: (1) MBT, (2) BRY, and (3) a Modified

FIG. 14 Comparison of mean NL G~c results for AS4/PEEK DCB specimens with 13-1~m aluminum inserts calculated using three different data reduction methods.

41 3.5 I1 MBT

3 ~ B R Y

2.5 ~] MCC

I l '"'1 "

7#m AI. insert

Mean NLGIc,

kJ/m 2

2

1.5

1

0.5

NASA BELL AFWAL CIBA G. NIP.OIL

FIG. 15--Comparison of mean NL G~ results for AS4/PEEK DCB specimens with 7-~m aluminum inserts calculated using three different data reduction methods.

Compliance Calibration (MCC) method [10]. For both insert thicknesses, the variation between the three data reduction meth- ods for any single lab was no greater than 3.1%. Hence, none of the three data reduction techniques was clearly superior to

TABLE 2---Summary of AS4/PEEK round robin data.

Round Tests/ Avg. Mean (CV), (CV)~ Round Robin # Labs Lab Insert Value G,c, kJ /m 2 Sv % SR %

ASTM 3 4 13 p,m Kapton NL 0,983* 0.132 13.4 0.178 18.1 ASTM/JIS 16 4 13 p,m AI NL 1.439" 0.187 13.4 0.261 18.1

Foil ASTM/JIS 16 4 13 Ixm AI VIS 1.724" 0.232 13.5 0.194 11.3

Foil ASTM/JIS 16 4 13 I~m At 5% 1.799" 0.213 11.8 0.146 8.1

Foil ASTM/JIS 5 4 7 tun AI Foil NL 1.727" 0.226 13.0 0.140 8.1 ASTM/JIS 5 4 7 p,m A1 Foil VIS 1.929" 0.257 13.3 0.201 10.4 ASTM/JIS 5 4 7 p,m At Foil 5% 2.059* 0.218 10.9 0.218 10.6

*All G~c values determined using Modified Beam Theory (MBT).

Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014Downloaded/printed byNULL pursuant to License Agreement. No further reproductions authorized.

Page 7: Round Robin Testing for Mode I Interlaminar …...J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round

O'BRIEN AND MARTIN ON INTERLAMINAR FRACTURE TOUGHNESS 275

the others. However, because the MBT method yielded the most conservative values of G~ for 80% of the specimens tested, the remaining data in this report are summarized using the MBT method only.

13-~m Aluminum Insert Results----Figure 16 summarizes Gk values measured using the load at onset of nonlinearity (NL), the load at visual observation of delamination onset at the edge (VIS), and the load corresponding to a 5% offset in the initial compliance of the DCB specimen (5%) for the 16 labs that con- ducted tests on 13-t~m (0.5-mil) aluminum insert specimens. Each lab received four specimens to test. Figure 17 shows the mean G~ values for the 16 labs that tested the 13-~m aluminum insert specimens. Figure 18 shows the standard deviation in the data reported by each of these 16 labs.

The mean NL, VIS, and 5% offset G~ values measured from the 13-1~m (0.5-mil) aluminum insert for the 16 labs, and the standard deviations for repeatability and reproducibility, are listed in Table 2. Also summarized in Table 2 are the coefficients of variation for repeatability and reproducibility.

As noted in Table 2, the variability between laboratories was greater for the NL onset measurements than for the VIS or 5%

qC' kJ /m 2

3

2 .5

2

1.5

1

0 .5

0

I I I

o o

O 16 LABS O 60 TESTS

NL VlS 5 %

FIG. 16--Box plot showing G~ round robin values .for AS4/PEEK DCB tests with 13-t~m aluminum inserts.

Mean

C~c, kJ/m 2

4

3 .5

3

2 .5

2

1.5

1

i I I I I I I I I I I I I I I I ~[

[ ] NL VI,q 13 p_m AI.insert

0.5

< _ ~ < Z m

r , 5 _ < - -

_

S,O.~ kJ/m 2

0.8

0.6

0 .4

0.2

0

• NL [ ] VIS

[ ] 5°/o

13 lim AI. insert

-r < ~

FIG. 18~Comparison of standard deviation in Gzc values for AS4~ PEEK DCB specimens with 13-~m aluminum inserts.

offset measurements. However, the average mean NL Glc value was significantly lower than the VIS and 5% offset G/c values. Figure 19 shows the percentage difference between the NL and 5% Gk values. The average difference in G~c was 20.2%.

In 73% of the tests with the 13-t~m (0.5-mil) aluminum in- serts, propagation values of G~, corresponding to delamination growth in the presence of fiber bridging, were lower than NL or VIS, or both, onset values. In 70% of the individual tests, the VIS G~ values were lower than the 5% offset values.

7-tJ~n Aluminum Insert Results----Figure 20 summarizes Glc values measured using the load at onset of nonlinearity (NL), the load at visual observation of delamination onset at the edge (VIS), and the load corresponding to a 5% offset in the initial compliance of the DCB specimen (5%) for the five labs that conducted tests on specimens with the 7-1xm (0.25-mil) alumi- num inserts. Each lab tested four specimens. Figure 21 shows the mean GI~ values for the five labs. Figure 22 shows the stan- dard deviation in the data reported by each of these five labs.

The mean NL, VIS, and 5% offset G/c values measured from

1 0 0 I I I I l i l l l l i i l l

80

60

13 #m AI. insert avg A~c = 20.2%

o~

Jill | ! j < ~

FIG. 17--Comparison of mean Glc values for AS4/PEEK DCB spec- FIG. 19--Percentage difference in NL and 5% offset G[c values for imens with 13-1xm aluminum inserts. AS4/PEEK DCB specimens with 13-txm aluminum inserts.

Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014Downloaded/printed byNULL pursuant to License Agreement. No further reproductions authorized.

Page 8: Round Robin Testing for Mode I Interlaminar …...J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round

276 JOURNAL OF COMPOSITES TECHNOLOGY & RESEARCH

q C ~ kJ/m 2

3

2.5

2

1.5

1

0.5

0 t

I

5 LABS 20 TESTS

I

VlS I

NL 5%

FIG. 20---Box plot showing G~ round robin values for AS4/PEEK DCB tests with 7-p~m aluminum inserts.

Mean ~c'

kJ/m 2

4

3.5

3

2.5

2

1.5

1

0.5

0

I I [ ] NL

[ ] V lS

[ ] 50/0

I I

7 l~m AI. insert

I I

1 NASA BELL AFWAL CIBAG. NIP.OIL

FIG. 21--Comparison of mean G~ values for AS4/PEEK DCB spec- imens with 7-txm aluminum inserts.

I I I I

S.O.~ kJ/m 2

0.8

0.6

0.4

0.2

• NL

[ ] VlS

[ ] 5%

7 ~tm AI. insert

NASA BELL AFWAL CIBAG. NIP.OIL

FIG. 22--Comparison of standard deviation in G,c values for AS4~ PEEK DCB specimens with 7-1xm aluminum inserts.

the 7-1xm (0.25-mil) aluminum insert for the five labs, and the standard deviation corresponding to the repeatability and repro- ducibility, are listed in Table 2. Also summarized in Table 2 are the coefficients of variation for repeatability and reproducibility.

The reproducibility and repeatability were similar for all three onset measurements. However, as was noted for the 13-1xm (1.0- mil) insert specimens, the NL Gzc values were significantly lower than the VIS and 5% offset G~c values. In 75% of the tests with the 7-p,m (0.25-mil) aluminum inserts, propagation values of G~c, corresponding to delamination growth in the presence of fiber bridging, were lower than NL or VIS, or both, onset values. The VIS G~c values were lower than the 5% offset values in 100% of the individual tests.

In nearly 75% of the tests with aluminum inserts, PLAT val- ues of Glc corresponding to delamination growth in the presence of fiber bridging were lower than NL or VIS, or both, onset values. This resulted in an R-curve, a plot of G~c as a function of delamination length, that rose and then decreased below VIS or NL G~c values, or both (Fig. 23). In contrast, R-curves for specimens with Kapton inserts always achieved PLAT Gxc values above the NL and VIS G1c values (Fig. 24). Microscopy studies performed at ICI on untested specimens indicated that localized yielding (crimping) may have occurred during cutting of the aluminum foil inserts (See Ref 1). These crimps, which were not evident in the original panel radiographs, were responsible

I I I

2.5

GIc' 2 kJ/m 2

1.5

O

" VIS O

"X 13 #m AI j

0 0 0 NL O O O O

PLAT

1 I ................ I I

40 55 70 85 100 Delamination Length, mm

FIG. 23--R-curve for AS4/PEEK DCB specimens with 13-txm alu- minum inserts.

2.5 I I

kJ/m 2

1.5

-13 p.m Kapton )

0.5 I I I ........... 3O 45 6O 75 9O

Delamination Length, mm

FIG. 24--R-curve for AS4/PEEK DCB specimens with 13-~m Kap- ton inserts.

NL 0

SOjo@' o o o o o pLAT

VlS O

Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014Downloaded/printed byNULL pursuant to License Agreement. No further reproductions authorized.

Page 9: Round Robin Testing for Mode I Interlaminar …...J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round

O'BRIEN AND MARTIN ON INTERI_AMINAR FRACTURE TOUGHNESS 277

TABLE 3---Summary of AS4/PEEK round robin data.

Round Tests/ Avg. Mean (CV), (CV)R Round Robin # Labs Lab Insert Value Gk, kJ/m 2 S, % SR %

3 EGF 6 4 13 ~m A! NL 1.574" 0.230 14.6 0.283 18.0 Foil

EGF 6 4 13 IJ, m Al V-IS 1.953" 0.209 10.7 0.286 14.6 Foil

EGF 6 4 13 I~m Al 5% 1.938" 0.202 10.4 0.220 11.4 Foil

EGF 4 4 7 ~m Al Foil NL 1.575" 0.160 10.2 0.121 7.7 EGF 4 4 7 Ixm Al Foil VIS 1.883" 0.150 8.0 0.154 8.2

4 ASTM 10 3 13 I~m Kapton NL 1.303" 0.180 13.8 0.207 15.9 ASTM 10 3 13 l~m Kapton VIS 1.549" 0.198 12.8 0.151 9.7 ASTM 10 3 13 ;~m Kapton 5% 1.713" 0.171 10.0 0.139 8.1

*All Gk values determined using MBT.

for the formation of resin pockets at the end of the inserts re- suiting in elevated G~c values. The tendency of aluminum inserts to crimp when cut may be worse in the thinner 7-tLm foils, which yield higher apparent G~c values than the 13-t~m foils (Table 2).

International Round Robin (EGF)

The 13-1~m (0.5-mil) insert specimens were tested by six labs, whereas the 7-1~m (0.25-mil) insert specimens were tested by four labs. Each lab received four specimens to test. The data were reduced using the MBT method. These tests, performed by EGF participants using specimens that were not X-rayed to iso- late specimens with tears or folds in the aluminum inserts, yielded results similar to those obtain by ASTM and JIS partic- ipants. Hence, only a summary of these results are presented in this report. A detailed description of the results may be found in Ref 1.

The mean NL, VIS, and 5% offset G~c values measured from the 13-lxm (0.5-mil) aluminum insert for the six labs, and the standard deviations for repeatability and reproducibility, are listed in Table 3. Also summarized in Table 3 are the coefficients of variation for repeatability and reproducibility. The variability between laboratories was greater for the NL onset measurements than for the VIS or 5% offset measurements. However, as noted for the ASTM/JIS results, the average mean NL Glc value was significantly lower than the VIS and 5% values.

The mean NL, VIS, and 5% offset Gxc values measured from the 7-t~m (0.25-mil) aluminum insert for the four labs, and the standard deviation corresponding to the repeatability and repro- ducibility, are listed in Table 3. Also summarized in Table 3 are the coefficients of variation for repeatability and reproducibility. The reproducibility and repeatability between laboratories were similar for the NL and VIS onset measurements. However, as noted for the ASTM/JIS results, the average mean NL Gxc value was significantly lower than the VIS values.

Results from Round 5

For this round, 13-1~m Kapton polyimide film inserts were sprayed with a mold release agent and were implanted before consolidation of the AS4/PEEK panels. Three static DCB tests were performed by each of ten labs as part of an ASTM DCB fatigue round robin. For all these tests, R-curves achieved PLAT G~c values above the NL and VIS G~c values (Fig. 24).

Gic,

kJ/m 2

3

2 .5

2

1.5

1

0 .5

I ! !

_ L

©

, T ~ ±

• I

NL VlS 5%

FIG. 25--Box plot showing G~c fatigue round robin values for AS4~ PEEK static DCB tests with 13-t~n Kapton inserts.

Figure 25 summarizes Gtc values measured using the load at onset of nonlinearity (NL), the load at visual observation of delamination onset at the edge (VIS), and the load corresponding to a 5% offset in the initial compliance of the DCB specimen (5%) for the ten labs that each conducted tests on three speci- mens with the 13-~m (0.5-mil) Kapton inserts as part of the DCB fatigue round robin. Figure 26 shows the mean Gtc values for the ten labs. Figure 27 shows the standard deviation in the data reported by each of the ten labs.

The mean NL, VIS, and 5% offset Glc values measured from the 13-1~m (0.5-mil) Kapton insert for the ten labs, and the stan- dard deviations for repeatability and reproducibility, are listed in Table 3. Also summarized in Table 3 are the coefficients of variation for repeatability and reproducibility. The variability be- tween laboratories was greater for the NL onset measurements than for the VIS or 5% offset measurements. However, the av- erage mean NL Gk value was significantly lower than the VIS and 5% values.

The reproducibility and repeatability of data from this round robin were similar to earlier round robins conducted with 13- I~m inserts. However, the mean NL Gxc values for this round robin were lower than the mean NL Gx,. values obtained with the aluminum inserts, but higher than those obtained from speci- mens with Kapton inserts in the original ASTM round robin. However, none of the two round robins conducted on specimens with Kapton inserts satisfied the requirements for a database to justify the precision statement for an ASTM standard (see

Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014Downloaded/printed byNULL pursuant to License Agreement. No further reproductions authorized.

Page 10: Round Robin Testing for Mode I Interlaminar …...J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round

278 JOURNAL OF COMPOSITES TECHNOLOGY & RESEARCH

Mean

kJ/m 2

3.5

3

2.5

2

1.5

1

0.5

0

1 , , ~ J i J ~ , i

E O

t~

t3 E

"O ¢.- ¢5

[ ] t~L 0.8 13 ~trn Kapton insert

[ ] VIS Static Data

[ ] 5% 0.6

0.4

o: _ _ _

FIG. 26--Comparison of mean G~c values for AS4/PEEK DCB spec- imens with 13-~xm Kapton inserts.

FIG. 27--Comparison of standard deviation in Glc values for AS4~ PEEK DCB specimens with 13-1~m Kapton inserts.

ASTM E 691). The required database includes a minimum of five tests conducted by at least six different laboratories. In order to generate the required database, and to quantify the sensitivity of G~c to insert thickness, a second international round robin was conducted.

Results from Round 5

For this round robin, both 7.5- and 13-1xm Upilex polyimide film inserts were sprayed with a mold release agent and were implanted before consolidation of the AS4/PEEK panels. Five specimens of each thickness insert were tested by nine labs. Each lab conducted the tests according to a draft ASTM DCB standard. For all DCB tests with Upilex inserts, R-curves achieved PLAT G~c values above the NL and VIS G~c values similar to results for specimens with Kapton inserts Gig. 24).

Figures 28 and 29 summarize G1c values measured using the load at onset of nonlinearity (NL), the load at visual observation of delamination onset at the edge (VIS), and the load corre- sponding to a 5% offset in the initial compliance of the DCB specimen (5%) for the nine labs that each conducted tests on five specimens with the 13-1xm and 7.5-p.m Upilex inserts. Fig- ures 30 and 31 show the mean G~c values for the nine labs. Figures 32 and 33 show the standard deviation in the data re- ported by each of the nine labs.

The mean NL, VIS, and 5% offset G~c values measured from the 13-p,m and 7.5-1xm Upilex inserts for the nine labs, and the standard deviations for repeatability and reproducibility, are listed in Table 4. Also summarized in Table 4 are the coefficients of variation for repeatability and reproducibility. The variability between laboratories was greater for the NL onset measurements than for the VIS or 5% offset measurements. However, the av- erage mean NL G~c values was significantly lower than the VIS and 5% values.

The reproducibility and repeatability of data from this round was as good as, and in many cases better than, the earlier rounds. Mean NL Gk values for this round robin were lower than ob- tained from all the previous round robins except for the original ASTM round robin with Kapton inserts. The average NL Gk values for the 7.5-1xm Upilex insert specimens were 6.3% lower

kJ/m 2

3

2.52 I

1.5

0.5

I I

9 LABS 44 TESTS

t I

NL VIS 5 % 1

FIG. 28--Box plot showing Gnc round robin values for AS4/PEEK DCB tests with 13-1~m Upilex inserts.

kJ/m 2

3

2.5

2

1.5

1

0.5

0

! I

8

9 LABS 45 TESTS

| | ..............

NL VIS 5 %

FIG. 29--Box plot showing Gic round robin values for AS4/PEEK DCB tests with 7.5-1xm Upilex inserts.

Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014Downloaded/printed byNULL pursuant to License Agreement. No further reproductions authorized.

Page 11: Round Robin Testing for Mode I Interlaminar …...J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round

O'BRIEN AND MARTIN ON INTERLAMINAR FRACTURE TOUGHNESS 279

M e a n

GIc, kJ/m 2

4

3.5

3

2.5

2

1.5

1

0.5

0

,< z co <

- J <

r-- o o I - - r r

u_

FIG. 30---Comparison of mean G~¢ values for AS4/PEEK DCB spec- imens with 13-p.m Upilex inserts.

I I I I I I I I

0.8

0.6 S.D.,

kJ/m 2 0.4

0.2

• NL [] VIS

[ ] 5°1o

13 Ixm Upilex insert

o

< N u__

FIG. 32--Comparison of standard deviation in G~ values for AS4~ PEEK DCB specimens with 13-1xm Upilex inserts.

I 4

3.5

3

2.5

2

1.5

1

0.5

0

I I I I I I I

[] NL 7.5 t~m Upitex insert

[ ] V lS

[ ] 5%

M e a n

kJ/m 2 ~ ;~ ~ ;

I I I I I 1 I I

W > , " 0 03 cO c 2~ -~ < N < ..J o LU ~ z co ~ ,,a:_

FIG. 31---Comparison of mean G~¢ values for AS4/PEEK DCB spec- imens with ZS-t~m Upilex inserts.

s .g .~

kJ/m 2

I I ....... I I I I I ............. I I

0.8

0.6

0.4

0.2

• NL [] VlS

[ ] 5o/0

7.5 pm Upilex insert

L o 8 o

FIG. 33--Comparison of standard deviation in G~,. values for AS4~ PEEK DCB specimens with Z5-~m Upilex inserts.

than the average NL Glc values for the 13-1xm Upilex insert specimens.

Summary

As a result of round 5, the draft DCB standard was updated and submitted for ballot within ASTM subcommittee D30.06 on Intedaminar Properties in June of 1992 and then, following res- olution of negative votes, was resubmitted for concurrent sub- committee and D30 main committee ballot in December of 1992. Guidelines were included in the standard for choosing piano hinges, blocks, or t-tabs for load introduction. The stan- dard includes generation of the complete R-curve for each test and reporting of G~c values measured using the load at onset of nonlinearity in the load versus displacement plot (NL), the load at visual observation of delamination onset at the edge (VIS), and the load corresponding to a 5% offset in the initial compli- ance of the DCB specimen (5%). However, the standard makes several recommendations.

First, because of the difficulty initiating delaminations in brit- tle epoxy matrix composites from polyimide (Kapton) films sprayed with a mold release agent, PTFE (Teflon ®) film inserts were recommended for these materials. Polyimide films are rec- ommended only for materials with high cure (or consolidation) temperatures.

Second, because specimens with insert thicknesses greater than 13 i~m yield unrealistically high G~c values, and because the difference in average NL Gk values for 7.5- and 13-txm Upilex inserts was relatively small (6.3%), an insert thickness requirement of 13 Ixm or less was adopted for the ASTM DCB standard. The 7.0 to 7.5-1~m inserts were optional because they represent minimum polyimide film thicknesses that are presently commercially available. Furthermore, these ultra-thin films are typically more difficult to obtain, and are considerably more dif- ficult to handle than the 13-1xm films. The polyimide films were recommended over the aluminum films for use as inserts in the DCB test because of the problems with crimping, tears, and folds in aluminum inserts noted in the first international round robin.

Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014Downloaded/printed byNULL pursuant to License Agreement. No further reproductions authorized.

Page 12: Round Robin Testing for Mode I Interlaminar …...J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round

280 J O U R N A L O F C O M P O S I T E S T E C H N O L O G Y & R E S E A R C H

TABLE 4----Summary of AS4/PEEK round robin data.

Round Tests/ Avg. Mean (CV), (CV)R Round Robin # Labs Lab Insert Value Gt,, kJ/m 2 S, % SR %

5 ASTM/JIS~GF 9 5 13 I.un Upilex NL 1.262" 0.132 10.5 0.I10 8.7 ASTM/JIS/EGF 9 5 13 I~m Upilex VIS 1.532" 0.167 10.9 0.075 4.9 ASTM/JIS/EGF 9 5 13 ~m Upilex 5% 1.549" 0.144 9.3 0.080 5.1 ASTM/JIS/EGF 9 5 7.5 Ixm Upilex NL 1.182" 0.126 10.7 0.111 9.4 ASTM/JIS/EGF 9 5 7.5 ~m Upilex VIS 1.447" 0.126 8.7 0.075 5.2 ASTM/JIS/EGF 9 5 7.5 Itm Upilex 5% 1.451" 0.130 9.0 0.096 6.6

*All Gtc values determined using MBT.

Third, three data reduction methods for calculating G1c values were included. These consisted of a Modified Beam Theory (MBT), a Compliance Calibration method (CC), and a Modified Compliance Calibration method (MCC). Because Glc values de- termined by the three different data reduction methods differed by no more than 3.1%, none of the three was dearly superior to the others. However, because of the MBT method yielded the most conservative values of G~c for 80% of the specimens tested, this method was recommended as the preferred data reduction technique. The area method was not recommended because it did not yield an initiation value of Gk or a delamination resis- tance curve.

Fourth, the NL Gxc value was recommended as the preferred measure of Mode I intedaminar fracture toughness for generat- ing delamination failure criteria in durability and damage toler- ance analyses of laminated composite structures. However, all three initiation values may be used for the other purposes cited in the scope, such as identifying the effects of fiber surface treat- ment, local variations in fiber volume fraction, and processing and environmental variables on G~c and comparing quantitatively the relative values of G~c for composite materials with different constituents.

The preference for NL Gzc values for durability and damage tolerance analyses is based on physical observations, made using video based in-situ dye penetrant enhanced X-radiography, that the delaminations initiate in the DCB specimen at the end of the insert, in the interior of the specimen width, when the load de- flection curve becomes nonlinear [11-14]. The difference in NL and VIS Gz, values is negligible for brittle epoxy matrix com- posites, but the difference is significant for tough thermoplastic

AGic, %

50 : '" j

4O

30 - I:16' 0

1 3gm AI

5 labs

10 labs

i i

NL-5%

[] NL-VIS

[] VLS-5%

9 labs 9 labs

7p.rn 1 3p.m 1 3/~m 7.5p.m At Kapton Upilex Upilex

FIG. 34--Percentage variation in mean Gl¢ initiation values for AS4/ PEEK laminates with different insert types.

C.V., %

40

30

20

10

0

I I ............. |

[ ] NL

[] VlS

E] 5%

6 labs 10 labs

5 labs ~

3p.m 7~tm 1 3~tm AI AI Kapton

9 labs ~ l

1 3~m 7.51sin Upilex Upilex

FIG. 35--Coefficient of variation in Glc initiation values for AS4~ PEEK laminates with different insert types.

matrix composites. As shown in Figure 34, mean VIS and mean 5% offset G~ values were typically 18 to 22% higher than mean NL Gtc values even though VIS and 5% measurements were more repeatable (Fig. 35). Hence, The NL G~c values are con- servative values corresponding to the first onset of delamination.

APPENDIX

List of Round Robin Participants

Round 1

1. NASA Langley Research Center 2. Texas A&M University 3. Defense Research Establishment Pacific (DREP) Canada 4. University of Compiegne, France 5. Royal Aerospace Establishment (RAE) England 6. Shell Development Company 7. Boeing Commercial Airplane Co. 8. Imperial Chemicals Industries (ICI) England 9. Rohr Industries, Inc.

10. University of Delaware

Round 2

1. NASA Langley Research Center 2. Texas A&M University 3. Defense Research Establishment Pacific 09REP) Canada 4. National Institute for Standards Technology (NIST)

Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014Downloaded/printed byNULL pursuant to License Agreement. No further reproductions authorized.

Page 13: Round Robin Testing for Mode I Interlaminar …...J~RER, Vol. 15, No. 4, Winter 1993, pp. 269-281. ABSTRACt. This report summarizes the results of several inter- laboratory "round

O'BRIEN AND MARTIN ON INTERLAMINAR FRACTURE TOUGHNESS 281

Round 3

ASTM/JIS

1. NASA Langley Research Center 2. Bell Helicopter Co. 3. Hamilton Standard (HAM S.) 4. University of Dayton Research Institute (UDRI) 5. McDonnell Douglas Helicopter Co. (MDHC) 6. BASF, Charlotte, NC 7. Lockheed Aeronautical Systems (LOCK.) 8. Industrial Products Research Institute (IPRI) Japan 9. Israel Aircraft Industries, Ltd. (IAI)

10. Air Force Wright Aeronautical Laboratories (AFWAL) 11. Ciba Geigy Corp. (CIBA G.), Anaheim, CA 12. Nippon Steel Co. (NIP.S.) Japan 13. 3M Corporation 14. Sikorsky Aircraft Co. (SIKOR.) 15. University of Tokyo, Japan 16. Nippon Oil Co. ~IP.OIL) Japan

EGF

1. Imperial College, England 2. University of Portugal 3. FFA, Sweden 4. The Welding Institute, England 5. F_cole Polytechnic Federale de Lausanne

Switzerland 6. Imperial Chemicals Industries (ICI), England 7. University of Cranfield, England

(EPFL),

Round 4

1. NASA Langley Research Center 2. Bell Helicopter Co. 3. Integrated Technologies (INTEC) 4. Royal Aerospace Establishment (RAE), England 5. The Welding Institute (TWI), England 6. Wichita State University (WSU) 7. Ecole Polytechnic Federale de Lausanne (EPFL),

Switzerland 8. Industrial Products Research Institute (IPRI), Japan 9. University of Missouri (U.Mo.)

10. Israel Aircraft Industries, Ltd. (IAI)

Round 5

1. NASA Langley Research Center 2. Bell Helicopter Co. 3. Rohr Industries, Inc. 4. McDonnell Douglas Helicopter Co. (MDHC) 5. BASF, Charlotte, NC 6. Air Force Wright Aeronautical Laboratories (AFWAL) 7. Imperial College (U.London), England

8. IFREMER, France 9. University of Tokyo, Japan

References

[1] O'Brien, 1". K. and Martin, R. H., "Results of ASTM Round Robin Testing for Mode I Interlaminar Fracture Toughness of Composite Materials," NASA TM 104222, August 1992.

[2] Wilkins, D. J., Eisenmann, J. R., Camin, R. A., et al., "Charac- terizing Delamination Growth in Graphite Epoxy," Damage in Composite Materials: Basic Mechanisms, Accumulation, Toler- ance, and Characterization, ASTM STP 775, K. L Reifsnider, Ed., American Society for Testing and Materials, Philadelphia, 1982, pp. 168-183.

[3] Bascom, W. D., Bitner, R. J., Moulton, R. J., and Siebert, A. R., "The Interiaminar Fracture of Organic-Matrix Woven Reinforced Composites," Composites, Vol. 11, No. 9, January 1980, pp. 9- 18.

[4] de Charentenay, E X., Harry, J. M., Prel, Y. J., and Benzeggagh, M. L., "Characterizing the Effect of Delamination Defect by Mode I Delamination Test," The Effect of Defects in Composite Materials, ASTM STP 836, D. J. Wilkins, Ed., American Society for Testing and Materials, Philadelphia, 1984, pp. 84-103.

[5] Russell, A. J., "Factors Affecting the Opening Mode Delamination of Graphite Epoxy Laminates," DREP Materials Report 82-Q, Defense Research Establishment Pacific, Victoria, B. C., Canada, 1982.

[6] Johnson, W. S. and Mangalgari, P. D., "Investigation of Fiber Bridging in Double Cantilever Beam Specimens," Journal of Composites Technology and Research, Vol. 9, Spring 1987, pp. 10-13.

[7] Martin, R. H., "Effect of Initial Delamination on G~c and G~ Val- ues from Glass Epoxy Double Cantilever Beam Tests," Proceed- ings, 3rd American Society for Composites (ASC) Conference, Se- attle, WA, Technomic Publishers, September 1988, pp. 688-700.

[8] Berry, J. P., "Determination of Fracture Energies by the Cleavage Technique," Journal of Applied Physics, Vol. 34, No. 1, January 1963, pp. 62-68.

[9] Hashemi, S., Kinloch, A. J., and Williams, J. G., "The Analysis of Interlaminar Fracture in Uniaxial Fibre-polymer Composites," Proceedings of the Royal Society of London, Vol. A427, No. 1827, 1990, pp. 173-199.

[10] Kageyama, K. and Hojo, M., "Proposed Methods for Interlaminar Fracture Toughness Tests of Composite Laminates," Proceedings of the 5th U.S./Japan Conference on Composite Materials, Tokyo, June 1990, pp. 227-234.

[11] de Kalbermatten, T., Jaggi, R., and Flueler, P., "Intedaminares Bruchverhalten von CFK-Epoxy/PEEK-Laminaten in Mode I an DCB-Proben (Intedaminar Fracture of CFK-Epoxy/PEEK Lami- nates in Mode I and DCB-Test)," EMPA Report No. 116'598/ZEP, Swiss Federal Laboratories for Materials Testing and Research, Dubendorf, Switzerland, September 1990.

[12] Brunner, A. J., "Real Time X-Ray Radiography and Simultaneous Acoustic Emission Monitoring During Mode I Crack Propagation in Double Cantilever Beam (DCB) Specimens from Carbon-Fiber Reinforced PEEK, EMPA Report No. 116'598/9 Zft~, Swiss Federal Laboratories for Material Testing and Research, Dubendoff, Swit- zerland, December 1992.

[13] Flueler, P., "Crack Propagation in Fiber Reinforced Composite Materials Analysed with Insitu Microfocal X-ray Radiography and Simultaneous Acoustic Emission Monitoring," Proceedings of the ECCM Conference on Composites Testing and Standardization, Amsterdam, 1992.

[14] de Kalbermatten, T., Jaggi, R., Flueler, P., and Kausch, H. H., Journal of Materials Science Letters, Vol. 11, 1992, pp. 543-546.

Copyright by ASTM Int'l (all rights reserved); Sun Sep 7 04:50:56 EDT 2014Downloaded/printed byNULL pursuant to License Agreement. No further reproductions authorized.