NREL is a national laboratory of the U.S. Department of Energy, Office of Energy Efficiency and Renewable Energy, operated by the Alliance for Sustainable Energy, LLC. Results of first outdoor comparison between Absolute Cavity Pyrgeometer (ACP) and Infrared Integrating Sphere (IRIS) Radiometer at PMOD Atmospheric System Research Science Team Meeting (March 1821, 2013) by Ibrahim Reda*, Julian Gröbner**, Stefan Wacker**, and Tom Stoffel* * = NREL , ** = PMOD/WRC NREL/PR3B1058121

Upload: others

Post on 30-Jul-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Results of first comparison between Cavity Pyrgeometer and … · 2019-08-21 · ACP‐W BB 0 0.5 1 1.5 17.66 17.74 17.82 UTC W/m 2 Residualsof fittingNetirradiancevs V tp ‐1.5

NREL is a national laboratory of the U.S. Department of Energy, Office of Energy Efficiency and Renewable Energy, operated by the Alliance for Sustainable Energy, LLC.

Results of first outdoor comparison between Absolute Cavity Pyrgeometer (ACP) and Infrared Integrating Sphere (IRIS) Radiometer at PMOD

Atmospheric System Research

Science Team Meeting (March 18‐21, 2013)by

Ibrahim Reda*, Julian Gröbner**, Stefan Wacker**, and Tom Stoffel*

* = NREL , ** = PMOD/WRC

NREL/PR‐3B10‐58121

Page 2: Results of first comparison between Cavity Pyrgeometer and … · 2019-08-21 · ACP‐W BB 0 0.5 1 1.5 17.66 17.74 17.82 UTC W/m 2 Residualsof fittingNetirradiancevs V tp ‐1.5

2

Outline

The ACP and IRIS are developed to establish a world reference for calibrating pyrgeometers with traceability to SI units. The two radiometers are unwindowed with negligible spectral dependence, and traceable to SI units through the temperature scale (ITS‐90). The first outdoor comparison between the two designs was held from January 28 to February 8, 2013 at the Physikalisch‐Metorologisches Observatorium Davos (PMOD).  The difference between the irradiance measured by ACP and that of IRIS was within 1 W/m2. A difference of 5 W/m2 was observed between the irradiance measured by ACP&IRIS and that of the interim World Infrared Standard Group (WISG).

2

Page 3: Results of first comparison between Cavity Pyrgeometer and … · 2019-08-21 · ACP‐W BB 0 0.5 1 1.5 17.66 17.74 17.82 UTC W/m 2 Residualsof fittingNetirradiancevs V tp ‐1.5

3

Absolute Cavity Pyrgeometer (ACP)‐ ACP Net irradiance:

‐ By cooling the ACP casetemperature, and since Watm

is stable, then,

‐ Then the atmospheric longwave irradiance is, 

Where,K1, Vtp, ϵ , K2, Wr , Wc and τ are the reciprocal of ACP’s responsivity, thermopile voltage, gold emittance, detector’semittance, receiver irradiance, CPC irradiance, and throughput (NIST characterization), consecutively.

K V W W K Wtp atm c r1 21 2* * ( ) * ( ) * *

WK V K W W

atmtp r c

1 22 1* ( ) * * ( ) *

KW K W

Vc r

tp1

21 2

( ) * ( ) * *

Reference: Reda, I.; Zeng, J.; Schulch, J.; Hanssen, L.; Wilthen, B.; Myers, D.; Stoffel, T. Dec. 2011. “An absolute cavity pyrgeometer to measure the absolute outdoor longwave irradiance with traceability to International System of Units, SI”. Journal of Atmospheric and Solar-Terrestrial Physics, 77 (2012) 132-143. http://dx.doi.org/10.1016/j.jastp.2011.12.011

Page 4: Results of first comparison between Cavity Pyrgeometer and … · 2019-08-21 · ACP‐W BB 0 0.5 1 1.5 17.66 17.74 17.82 UTC W/m 2 Residualsof fittingNetirradiancevs V tp ‐1.5

4

The Infrared Integrating Sphere (IRIS) Radiometer

Key features of the IRIS Radiometer

• Windowless

• Irradiance measurement by using a 60 mm gold‐plated integrating sphere as input optic

• High sensitivity from a windowless pyroelectric detector

• Flat spectral response

• Measurement frequency 0.1 Hz

• Automatic unattended operation

• Nighttime measurements only

Reference surface

DetectorAperture

Shutter

IRIS UncertaintyU95%= 1.8 Wm‐2 summer (+15°C)

2.4 Wm‐2 winter (‐15°C)

Reference: Gröbner, J., A Transfer Standard Radiometer for atmospheric longwave irradiance measurements, Metrologia, 49, S105-S111,2012.

Page 5: Results of first comparison between Cavity Pyrgeometer and … · 2019-08-21 · ACP‐W BB 0 0.5 1 1.5 17.66 17.74 17.82 UTC W/m 2 Residualsof fittingNetirradiancevs V tp ‐1.5

5

ACP versus PMOD‐BB on Jan 29 to Feb 2, 2013 ACP Net irradinace vs Vtp during cooling

y = 0.0842x ‐ 255.2R2 = 0.9978

‐285

‐280

‐275

‐270

‐265‐350 ‐300 ‐250 ‐200 ‐150 ‐100 ‐50 0

Vtp (uV)

W/m

2

WACP‐WBB

0

1

2

3

1.00 11.00UTC

W/m

2

Residuals of fitting Net irradiance vs Vtp

‐0.4

‐0.2

0

0.2

0.4

1.00 11.00

UTC

W/m

2

ACP Net irradinace vs Vtp during cooling

y = 0.0834x ‐ 254.62R2 = 1 ‐400

‐350

‐300

‐250‐1400 ‐1200 ‐1000 ‐800 ‐600 ‐400 ‐200 0

Vtp (uV)

W/m

2

WACP‐WBB

0

1

2

3

13.16 13.24 13.33 13.41 13.49UTC

W/m

2

Residuals of fitting Net irradiance vs Vtp

‐0.3

‐0.1

0.1

0.3

13.16 13.24 13.33 13.41 13.49

UTC

W/m

2

ACP Net irradinace vs Vtp during cooling

y = 0.0827x ‐ 254.7R2 = 0.9992

‐360

‐340

‐320

‐300

‐280‐1400 ‐1200 ‐1000 ‐800 ‐600 ‐400 ‐200 0

Vtp (uV)

W/m

2

WACP‐WBB

0

1

2

3

11.62 11.71 11.79 11.87UTC

W/m

2

Residuals of fitting Net irradiance vs Vtp

‐1.5

‐0.5

0.5

1.5

11.62 11.71 11.79 11.87

UTC

W/m

2

ACP Net irradinace vs Vtp during cooling

y = 0.0822x ‐ 273.94R2 = 0.9996

‐380

‐360

‐340

‐320‐1200 ‐1000 ‐800 ‐600 ‐400 ‐200 0

Vtp (uV)

W/m

2

WACP‐WBB

0

0.5

1

1.5

17.66 17.74 17.82UTC

W/m

2Residuals of fitting Net irradiance vs Vtp

‐1.5

‐0.5

0.5

1.5

17.66 17.74 17.82

UTC

W/m

2

Page 6: Results of first comparison between Cavity Pyrgeometer and … · 2019-08-21 · ACP‐W BB 0 0.5 1 1.5 17.66 17.74 17.82 UTC W/m 2 Residualsof fittingNetirradiancevs V tp ‐1.5

6

Transient vs steady state in BB, Jan 29‐Feb 2, 2013

WACP‐WBB

0

0.5

1

1.5

2

2.5

3

3.5

12.85 12.94 13.02 13.10 13.19 13.27 13.35 13.44

UTC

W/m

2

Transient during coolingSteady State before cooling

WACP‐WBB

0

0.5

1

1.5

2

2.5

3

3.5

23.38 23.47 23.55 23.63 23.72 23.80

UTC

W/m

2

Transient during coolingSteady State before cooling

WACP‐WBB

0

0.5

1

1.5

2

2.5

3

3.5

11.32 11.41 11.49 11.57 11.66 11.74 11.82 11.91

UTC

W/m

2

Transient during coolingSteady State before cooling

WACP‐WBB

00.20.40.60.81

1.21.41.61.82

17.35 17.43 17.51 17.60 17.68 17.76 17.85

UTC

W/m

2

Transient during coolingSteady State before cooling

Page 7: Results of first comparison between Cavity Pyrgeometer and … · 2019-08-21 · ACP‐W BB 0 0.5 1 1.5 17.66 17.74 17.82 UTC W/m 2 Residualsof fittingNetirradiancevs V tp ‐1.5

7

Outdoor ACP&IRIS at night Feb. 5, 2013

Page 8: Results of first comparison between Cavity Pyrgeometer and … · 2019-08-21 · ACP‐W BB 0 0.5 1 1.5 17.66 17.74 17.82 UTC W/m 2 Residualsof fittingNetirradiancevs V tp ‐1.5

8

Outdoor ACP&IRIS at night Feb. 5, 2013

Page 9: Results of first comparison between Cavity Pyrgeometer and … · 2019-08-21 · ACP‐W BB 0 0.5 1 1.5 17.66 17.74 17.82 UTC W/m 2 Residualsof fittingNetirradiancevs V tp ‐1.5

9

Outdoor ACP, IRIS & WISG at night Feb. 5, 2013

Page 10: Results of first comparison between Cavity Pyrgeometer and … · 2019-08-21 · ACP‐W BB 0 0.5 1 1.5 17.66 17.74 17.82 UTC W/m 2 Residualsof fittingNetirradiancevs V tp ‐1.5

10

Irradiance difference (WISG minus IRIS) at PMOD

Results

Average Offset (IWV>10) ‐4.1 ± 1.5 Wm‐2

Gradient (IWV<10)‐0.45 ± 0.1 Wm‐2mm‐1IWV

WISG 1 WISG 2 WISG 3

WISG 4

Night averages

IRIS u95

From Julian’s presentation, IRS2012-Germany(Data from 180 nights)

Page 11: Results of first comparison between Cavity Pyrgeometer and … · 2019-08-21 · ACP‐W BB 0 0.5 1 1.5 17.66 17.74 17.82 UTC W/m 2 Residualsof fittingNetirradiancevs V tp ‐1.5

11

Irradiance difference (ACP minus WISG) at NREL

Three cooling cycles on November 18 and 21, 2012 with 40% RH at SRRLConsistent with Julian’s observation with high water vapor*

* Algebra is reversed for consistency with NREL’s historical files

Page 12: Results of first comparison between Cavity Pyrgeometer and … · 2019-08-21 · ACP‐W BB 0 0.5 1 1.5 17.66 17.74 17.82 UTC W/m 2 Residualsof fittingNetirradiancevs V tp ‐1.5

12

Preliminary Conclusions• Special set‐up of ACP in BB due to unknown gradient in CPC• Outdoor agreement between ACP & IRIS to within 1 W/m2

• Irradiance measured by WISG is ~4 W/m2 lower than that measured by ACP&IRIS. Is Consistent with a Water Vapor Column of 8 mm. This was also observed at NREL/SRRL at RH = 40% (on November 18, 2012 at NREL/SRRL: Water Vapor Column from 7 mm to 9 mm during cooling cycles)

• Future comparison with higher/lower water vapor to resolve observed spectral effect on outdoor pyrgeometer calibrations

• A 3rd design might increase confidence in establishing a consensus reference with traceability to SI units.