results-and-discussion-chromatography_edited.docx
TRANSCRIPT
RESULTS AND DISCUSSION
In this experiment, two types of chromatography
based on solid-liquid chromatographic methods
were used for the purpose of separation and
purity determination. Column chromatography,
which runs a solvent that moves down slowly
through a glass column with packed adsorbent,
was used to separate the colored components of
malunggay leaves. On the other hand, thin layer
chromatography (TLC) which utilizes a stationary
phase composed of a thin layer of adsorbent
material on a glass plate or plastic sheet, was
used to separate the components of each
collected eluate to overall determine their purity
based on the retention values (Rf values)
obtained.
Starting with column chromatography for
separating the colored components of the
malunggay leaves, hexane-acetone was
determined to be the suitable solvent system to
be used as it run through the glass column with
packed adsorbent without dissolving the solid.
This reagent was continuously added up on the
column to prevent it from running dry, as the
pure extract of malunggay leaves was prepared
with the same reagent, to be added next to
perform the chromatography process.
As the pure extract slowly moved down the
column, 3.0 mL each of the solvents, specifically
hexane-acetone, acetone, and acetone-methanol
respectively, were added continously at the same
time while collecting different colors of eluates in
separate test tubes. For this, four different-
colored eluates were collected, of which two were
mixed to provide three distinct-colored eluates,
particularly, pale yellow, green, and yellow green.
The following table presents the data collected
under column chromatography.
Table 1. Characteristic Color and Volume of
Collected Eluate (Column Chromatography)
Color of ComponentVolume of Eluate
(Drops)
1 Pale yellow
2 Green
3 Yellow Green
These eluates were then used as individual
samples for spotting the TLC plates which were
then put inside the developing chamber for
completing the thin layer chromatography. The
movement of these eluates along the TLC plate
coated with silica-gel were possible through the
principle of capillary action.[1] Retention factors
were then determined by measuring the distance
of each component from the origin and by
dividing this to the solvent front, represented by
the formula:
Rf=Distanceof sample¿origin ¿Distanceof solvent ¿
origin¿
Table 2. Characteristic Color and Distance of
Component from Origin (X) with Determined
Rf Value
Color of
Component
Distance of
component
from origin (X)
in cm
Rf Value
1 Pale Yellow 4.89 cm 0.75
2 Green 4.79 cm 0.74
3 Yellow Green 0.42 cm 0.42
After the Rf values were calculated, several
deducations were made. The difference in the Rf
values of the components are based from their
solubility and the strength of their adsorption to
the adsorbent, in the case the TLC plate coated
with silica gel. The lower the Rf value of the
component, the more soluble it is to the solvent
system used, which was hexane-acetone. This is
based on the solubility principle ‘like dissolves
like’. The more soluble the component in the
solvent system is adsorbed more firmly to the
TLC plate.[1] With this principle, it was deduced
that the third component is non-polar, like the
polarity of the solvent sytem hexane-acetone.
From the Rf values calculated, it is also possible
to determine the identity the component by
comparing the experimental Rf values with the Rf
values of known components. The spots on the
TLC plate is also an indicator of purity although it
is not an absolute indicator of it. A pure
compound would appear as a single spot on the
plate and several spots are indications of
impurities.[2] It is also deduced that how close the
Rf values of a component to a known compound
is an indication of the purity of the component.
These explanations is now correlated with the
results obtained from the chromatogram shown
below. The third component has a polarity alike
with the solvent system, which was non-polar,
since it had the lowest Rf value out of the three
components. The several spots on the
Chromatogram, specifically in the second
component is also an indication that the
components we have separated still contain some
impurites. These impurites can be from the
different apparatus used and even from the
different solvents used in Column
Chromatography.
Figure 1.
Chromatogram under UV light
Figure 2. Dried Chromatogram with Markings
REFERENCEFrom the Internet[1] University of Massachusetts. (n.d.). Thin layer Chromatography [PDF]. Retrieved from http://www.chem.umass.edu/~samal/269/tlc.pdf
[2] Interpretation of experimental data [PDF]. (2011, August). Retrieved from http://chemistry.syr.edu/totah/che276/support/6a1.handouts/4.DataInterp.pdf