Page 1: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

SPARC Workshop

Human Use - Cross-Cutting Technologies – Part II

Kevin L. Kilgore, Ph.D.

MetroHealth Medical CenterLouis Stokes Cleveland VAMC

Case Western Reserve University

Page 2: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Nerve Block/Modulation• Temporary Block: reversible (within

seconds to minutes) arrest conduction of action potential along a neuron

• Modulation: increase or decrease the excitability of a neuron (or group of neurons)

• These features are particularly relevant to SPARC applications, because down-regulation is often as important as up-regulation (in contrast to most neuromuscular applications)

Page 3: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Nerve Block/Modulation• Technologies available

• Kilohertz Frequency Alternating Current (KHFAC)

• Direct Current• Infrared (thermal) block• Optogenetics• Magnetic• Others…?

Page 4: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

KHFAC – Brief Tutorial• Illustrates the difficulties that we face in

applying “new” technologies to “not fully characterized” physiology• In contrast to motorneuron->skeletal

muscle• Illustrates the critical need for clinicians

and engineers to work directly together if SPARC initiatives are to be successful

Page 5: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

KHFAC – Back to the friendly confines of the somatic nervous system

Nerve Conduction Block Experimental Setup

Page 6: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

KHFAC Instant Reversibility

Proximal Prox Stim Block Distal Stim Distal

Page 7: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

KHFAC Graded Block

0

10

20

30

40

50

60

70

80

90

100

0 1 2 3 4 5 6 7 8 9 10

Sine Wave Amplitude (Vpp)

Blo

ck E

ffec

tiven

ess

(%)

Page 8: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Block in Sensory NerveAv

erag

ed C

AP

(uV)

30kHz HFAC Block

Time (ms)

8Vpp

6 Vpp

4Vpp

2 Vpp

0 Vpp

10 Vpp

Aβ Aγ Aδ

5uV

Page 9: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Direct Effect of KHFAC on Neural Structures

• KHFAC can block conduction of action potentials (within 10ms in some cases)

• KHFAC block can be quickly reversible (within 1s)

• Block effect is strongly dependent on amplitude• At block threshold and above, activation always

occurs at startup (“onset response”)• There is a lot of unexplored territory in this field• Neural response to KHFAC is complex• Almost all of the basic work to date is in

myelinated nerves

Page 10: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Typical stimulation frequency –neural response is simple and very well characterized

Amplitude (Vpp)

subthreshold activation NotSafe

20 Hz Stimulation

Dur

atio

n of

Del

iver

y

Page 11: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Neural Response at ~5kHz and higher

The neural response to KHFAC is extremely complex.

This map only considers the amplitude-time domain!

Dur

atio

n of

Del

iver

y

Page 12: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

KHFAC Block

• You must know the dose you are delivering– What is the real output of the generator?

• You must know what the KHFAC is doing to the nerve fibers– Activated? Blocked? Nothing? Cooked?

• Requires a methodical approach– See Canning/Kollarik approach – start with the

simplest system possible and work up to the complete in-vivo system

Page 13: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

DC Block with No Onset ResponseWell known in the literature

Blocking Stimulus 50µA Amplitude100 ms

Single Fiber Recording

Mechanostimulator

Petruska, Hubscher, Johnson : Anodally focused polarization of peripheral nerve allows discrimination of myelinated and unmyelinated fiber input to brainstem nuclei. Exp Brain Res 1998.

Page 14: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Considering Direct Current Nerve Block

CBDC Waveform

The current controlled waveform is a charge-balanced direct current waveform (CBDC) consisting of a cathodic blocking phase followed by a anodic recharge phase.

The total charge delivered was less than the Q value for Pt-Black and IrO2 electrodes. The length of the recharge was selected to return 100% of the charge at a current level of

10% of the cathodic DC.

Page 15: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Nerve Block and Neuromodulation

• Features Relevant to SPARC• Downregulation of neural activity

• Availability and Human Use Status• KHFAC – human use• Other modalities – animal only

• Areas for Research and Development• Consistent and repeatable block• Chronic safety• Small fiber block (esp. C-fiber)• Targeted block of small fibers

Page 16: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Nerve Cuff Electrodes• Features Relevant to SPARC

• Generally – activate whole nerve/fascicles• Generally >1mm diameter nerves

• Availability and Human Use Status• 10+ years of human use• Have been used on vagus

• Areas for Research and Development• Target <1mm diameter nerve branches• Activation of specific fibers sizes• Surgical techniques for electrode placement

Page 17: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Implanted Electrodes in Proximity to Nerve

• Features Relevant to SPARC• Can target nerves of any size• Easy to implant• Poorly controlled selectivity

• Availability and Human Use Status• Widely used in humans and widely

available• Areas for Research and Development

• Usefulness and practicality for targeting small nerve branches

Page 18: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Optrodes for Neural Activation

• Features Relevant to SPARC• High degree of specificity

• Availability and Human Use Status• Not yet available for human use

• Areas for Research and Development• Use as tool in animal models?• Groundwork for human use (safety,

practical designs)• Possible unique features, such as

activation and block; fiber type selectivity

Page 19: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Non-invasive Neural Activation

• Features Relevant to SPARC• Non-invasive• Poor specificity

• Availability and Human Use Status• Low threshold for human use

• Areas for Research and Development• Possibility of activating autonomic neural

structures from skin surface

Page 20: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Sensor Modalities• Blood Pressure/Bladder Pressure• Biomarkers of various sorts• pH• Ions• Cardiac output• Airway resistance• Glucose• More to come…

Page 21: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Sensor Systems

• Features Relevant to SPARC• Most systems will require regulation to a

mean, not to an extreme• Availability and Human Use Status

• Few have extensive human use• Areas for Research and Development

• Identify important modalities• Chronic efficacy of sensors• Accuracy of sensor data• Incorporation into complete system

Page 22: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Implant Systems• Stimulators• Stimulator/sensor systems• Modular systems

Page 23: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Implant Systems

• Features Relevant to SPARC• In many cases, will require stimulation,

block, sensing, and signal processing• Availability and Human Use Status

• Some have extensive human use• Areas for Research and Development

• Identify system requirements• Regulatory requirements

Page 24: Restoration of Hand and Arm by Functional Neuromuscular ... · neuromuscular applications) Nerve Block/Modulation ... complex. This map only considers the amplitude-time domain! Duration

Cross-Cutting TechnologySummary

• Technologies for Activation• Technologies for Block/Modulation• Sensor Technologies• System Technologies• Signal Processing• Computational Modeling