Resting ECG System
Cardiospy® User manual
for EC-12R; EC-12R/S; EC-12S recorder
LABTECH LTD.
Address 4031 Debrecen Vág utca 4., HUNGARY
E-mail [email protected]
Tel +36-52-310-128
Fax +36-52-412-023
INTRODUCTION RESTING ECG SYSTEM
2 V5.03_2014_11 LABTECH LTD.
1 Introduction
Dear customer!
Thank you for purchasing our Cardiospy Resting ECG System. We sincerely hope that
our product will help you with your diagnostic and therapeutic work. To be able to use
Resting ECG System and computer programs appropriately, please read the manuals
carefully. A basic knowledge of computers and Windows based applications is required.
We appreciate any comment you might have on the use of the System. Please, turn to us
with confidence: our colleagues and representatives will be pleased to help you, even in
person if need be.
What is a Resting ECG System?
Connected to a PC, the small sized and compact ECG device makes a complete 12-
Channel Resting ECG system with the software attached.
How to use a Resting ECG System?
During an examination made by the Resting ECG system, a record of 5 to 60 seconds is
made of a patient and this recording is analysed by the software.
The data evaluation is aided by the interpretation system of the software. The devices
can be operated in network and all the patient data is managed in a common database.
Quality certifications
The CE mark indicates that it meets the essential
requirements of European Council Directive 93/42/EEC
concerning medical devices.
Labtech operates a quality system certified to ISO 9001:2008
and ISO 13485:2003.
RESTING ECG SYSTEM INTRODUCTION
LABTECH LTD. V5.03_2014_11 3
We wish you success in the use of our resting ECG system.
Should you have any questions, ideas, partnership suggestions please contact us:
LABTECH LTD.
Address 4031 Debrecen Vág utca 4., HUNGARY
E-mail [email protected]
Tel +36-52-310-128
Fax +36-52-412-023
TABLE OF CONTENTS RESTING ECG SYSTEM
4 V5.03_2014_11 LABTECH LTD.
2 Table of contents
1 INTRODUCTION ................................................................................................... 2
2 TABLE OF CONTENTS ........................................................................................ 4
3 PREPARATIONS ................................................................................................... 6
3.1 List of accessories .......................................................................................... 6
3.2 Principles of Operation ................................................................................. 10
3.3 Warnings and Contraindications: ................................................................. 11
4 INSTRUCTIONS .................................................................................................. 13
4.1 EC-12R (Bluetooth) Recorder Components ................................................. 13
4.2 EC-12R/S (Bluetooth) Recorder Components ............................................. 14
4.3 EC-12S (Bluetooth) Recorder Components ................................................ 15
4.4 Database management interface ................................................................... 16
4.4.1 Toolbar ..................................................................................................... 16
4.4.2 List of patients .......................................................................................... 20
4.4.3 Records toolbar ........................................................................................ 21
4.4.4 List of records .......................................................................................... 22
4.4.5 Selected System ....................................................................................... 22
4.4.6 Filters ....................................................................................................... 23
4.4.7 Search ....................................................................................................... 23
4.4.8 Database information ............................................................................... 23
4.5 How to start a new recording? ...................................................................... 24
4.5.1 Connecting the ECG device (USB type) .................................................. 24
4.5.2 Connecting the ECG device (Bluetooth type) .......................................... 24
4.5.3 Patient preparation ................................................................................... 24
4.5.4 New recording in resting mode ................................................................ 28
4.6 Resting ECG mode display .......................................................................... 29
4.7 Toolbar ......................................................................................................... 32
4.7.1 Start monitoring ....................................................................................... 32
RESTING ECG SYSTEM TABLE OF CONTENTS
LABTECH LTD. V5.03_2014_11 5
4.7.2 Mode ......................................................................................................... 32
4.7.3 ECG leads ................................................................................................. 32
4.7.4 Filters ........................................................................................................ 34
4.7.5 Settings ..................................................................................................... 35
4.7.6 Print Report ............................................................................................... 36
4.7.7 Print current sample .................................................................................. 36
4.7.8 Export the Report ...................................................................................... 36
................................................................................................................... 36
4.7.9 Automatic report ....................................................................................... 37
4.7.10 Print and export settings ....................................................................... 38
4.7.11 Help ...................................................................................................... 40
4.7.12 Paper speed ........................................................................................... 40
4.7.13 Amplitude ............................................................................................. 40
4.7.14 Record length ....................................................................................... 40
5 EDITING AND MEASUREMENT POSSIBILITIES ......................................... 42
5.1 Arrhythmia analysis in resting and stress test ECG systems ......................... 42
5.2 Edit and delete ectopic events ....................................................................... 43
5.3 Ruler .............................................................................................................. 44
6 ERROR CORRECTION, MAINTENANCE ........................................................ 45
6.1 USB-02 signs and error messages ................................................................. 45
6.2 ECG Recorder signs and error messages ...................................................... 46
6.3 General and communication failure .............................................................. 46
7 TECHNICAL SPECIFICATION ......................................................................... 51
PREPARATIONS RESTING ECG SYSTEM
6 V5.03_2014_11 LABTECH LTD.
3 Preparations
The table below shows which recorders can be used with each system.
EC-12R EC-12R/S EC-12S
Resting ECG X X X
Continuous ECG X X
Stress ECG X X
3.1 List of accessories
Please check the listed units and accessories before installing the Resting ECG System.
New
Code No. Code No. Specification
EC
-12
R
EC
-12
R
(Blu
eto
oth
)
EC
-12
R/S
EC
-12
R/S
(Blu
eto
oth
)
EC
-12
S
(Blu
eto
oth
)
Picture
CAS-
012RS-01
12RS-V5-
01
Case for
holding the rest
and stress
system
1 1 1 1 1
CAS-
00IRD-01
12RS-V1-
01
Case for
holding the
IRDA and Old
type recorders
1 - 1 - -
ACC-
0SWCD-
01
12RS-V5-
02
CD with
installation
software
1 1 1 1 1
NOTE: BEFORE STARTING TO USE THE RESTING ECG SYSTEM YOU ARE EXPECTED TO
FOLLOW LABTECH LTD.’S INSTALLATION AND UPDATE GUIDE, WHICH PROVIDES
GUIDANCE ON THE INSTALLATION OF THE SYSTEM.
RESTING ECG SYSTEM PREPARATIONS
LABTECH LTD. V5.03_2014_11 7
New
Code No. Code No. Specification
EC
-12R
EC
-12R
(Blu
eto
oth
)
EC
-12R
/S
EC
-12 R
/S
(Blu
eto
oth
)
EC
-12S
(Blu
eto
oth
)
Picture
DEV-
0USBA-
02
12RS-V5-
03-A
USB-02/A
Bluetooth
Interface Unit
(rest and stress;
holter
monitoring)
- 1 - 1 1
DEV-
012RS-04
12RS-V5-
04-A
ECG recorder
unit (EC-12R/S) - - 1 - -
DEV-
012RS-05
12RS-V5-
04-B
ECG recorder
unit (EC-12R/S
Bluetooth)
- - - 1 -
DEV-
0012S-05
12RS-V5-
04-C
ECG recorder
unit (EC-12S
Bluetooth)
- - - - 1
DEV-
012RS-04
12RS-V5-
04-A
EC-12R ECG
recorder unit
1 - - - -
DEV-
012RS-05
12RS-V5-
04-D
EC-12R ECG
recorder unit
(BT)
- 1 - - -
CUF-
00LAC-
31
LAC-3.1
Stress BP cuff,
Adult (25-
35cm)
- - - - 1
PREPARATIONS RESTING ECG SYSTEM
8 V5.03_2014_11 LABTECH LTD.
New
Code No. Code No. Specification
EC
-12R
EC
-12R
(Blu
eto
oth
)
EC
-12R
/S
EC
-12 R
/S
(Blu
eto
oth
)
EC
-12S
(Blu
eto
oth
)
Picture
BAG-
012RS-01
12RS-V5-
05-A
Recorder bag
(EC-12R, EC-
12RS, non-
wireless)
1 - 1 - -
BAG-
012RS-02
12RS-V5-
05-B
Recorder bag
(EC-12R, EC-
12RS, EC-12S,
wireless)
- 1 - 1 1
CHA-
0004X-01
12RS-V5-
07
Battery charger
MAXI - 1 - 1 1
BAT-
00000-02
12RS-V5-
08
1.2V
Rechargeable
batteries (AA)
- 4 - 4 4
CAB-
00ECG-
03
12RS-V5-
09-A
12 CH standard
patient cable
(D-SUB -
Banana
connector)
1 - - - -
CAB-
00L12-06
12RS-V5-
09-B
12 CH standard
patient cable
(HDMI-Banana
connector)
- 1 - - -
CAB-
00L12-05
12RS-V5-
09-C
12 CH standard
patient cable
(HDMI-Snap
connector)
- - - 1 1
CAB-
00ECG-
01
12RS-V5-
09-D
12 CH standard
patient cable
(D-SUB - Snap
connector)
- - 1 - -
RESTING ECG SYSTEM PREPARATIONS
LABTECH LTD. V5.03_2014_11 9
New
Code No. Code No. Specification
EC
-12R
EC
-12R
(Blu
eto
oth
)
EC
-12R
/S
EC
-12 R
/S
(Blu
eto
oth
)
EC
-12S
(Blu
eto
oth
)
Picture
ELE-
CLAMP-
01
12RS-V5-
10-A
Adult clamp
electrodes (1
pack / 4 pcs) 1 1 - - -
ELE-
CHEST-
01
12RS-V5-
10-B
Adult chest
electrodes (1
pack / 6 pcs)
ELE-
00000-01
12RS-V5-
11
Single-use ECG
electrodes 1 1 1 1 1
CAB-
00USB-
03
12RS-V5-
12-B
PC USB cable,
5m (USB A -
USB B)
1 - 1 1 1
CAB-
00USB-
06
12RS-V5-
16
USB-SIO cable
(USB A - D-
SUB, for
ergometer,
treadmill)
- - 1 1 1
BEL-
00000-01
12RS-V5-
06 Waist belt - - 1 1 1
CAB-
00USB-
01
12RS-V5-
15
Extension cable
for USB-02 - 1 - 1 1
USR-
00000-01
12RS-V5-
15-A
Installation and
Update Guide
1 1 1 1 1
USR-
00000-04
12RS-V5-
15-B
User manual for
rest and stress
systems
PREPARATIONS RESTING ECG SYSTEM
10 V5.03_2014_11 LABTECH LTD.
New
Code No. Code No. Specification
EC
-12R
EC
-12R
(Blu
eto
oth
)
EC
-12R
/S
EC
-12 R
/S
(Blu
eto
oth
)
EC
-12S
(Blu
eto
oth
)
Picture
BAG-
012RS-04
12RS-V5-
17
Patient cable
bag - - 1 1 1
PAD-
00000-01
12RS-V5-
15-A Mouse pad 1 1 1 1 1
*The manufacturer maintains the right to change at any time.
3.2 Principles of Operation
Resting ECG examinations are useful in the case of patients known to have a
cardiovascular disease or who are at increased risk of developing one. The examination
may provide useful information regarding diagnosis and prognosis and it also supports
making rational decisions, including risk analysis before a medical and surgical
intervention.
It also gives useful information regarding the presence or absence of arrhythmias,
conduction defects, chamber enlargement, myocardial hypertrophy, myocardial
ischemia, myocardial necrosis, pericardial inflammation and electrolyte
disturbance.Figyelmeztetések és ellenjavallatok a Nyugalmi EKG rendszerhez
NOTE: PLEASE REFER TO THE CODE NUMBER ABOVE WHEN PLACING AN ORDER.
RESTING ECG SYSTEM PREPARATIONS
LABTECH LTD. V5.03_2014_11 11
3.3 Warnings and Contraindications:
Never use the recorder:
- in the presence of flammable anaesthetics or explosive surroundings
- in strong heat
- in strong electrostatic field
- in wet conditions (the recorder is not waterproof)
- Do NOT remove recorder covers - the recorder does not contain any parts which could
be serviced by users
- Do NOT immerse the recorder in any fluid
- Always use non-aggressive fluids to clear the patient cables and the recorder.
To achieve good and safe connection, to take care of the patient’s skin and to avoid
infection use only good quality single-use electrodes that are marked with the CE sign.
Do not use electrodes with expired warranty or dry or used ones from old, previously
opened packages because it will result in bad recording quality.
Usage with other devices:
The system may be used safely in patients with pacemakers, Implanted Cardiac
Defibrillator (ICD) or other implanted devices. The implanted devices will not disrupt
or be influenced by the operation of the recorder, as well as the recorder will not be
influenced by the operation of the above devices.
The recorder may be used simultaneously with a high frequency surgical device, in this
case make sure that the ECG electrodes are placed as far as possible away from the
surgical field.
The ECG recorders are not defibrillator-protected! Never use withnon-implanted
defibrillators!
Environmental conditions of storage and operation:
Temperature-range: -10-50oC
Relative humidity: 10-95% non-condensing
Air pressure-range: 700-1060 hPa
Opening the enclosure by unauthorized person may invalidate warranty. Modification of
the ME equipment can only be performed by the manufacturer. In case of problems
contact a professional repair shop.
NOTE: IF ANY ABNORMALITY OCCURS IN THE UNIT, SUSPEND THE OPERATION IMMEDIATELY
AND DISCONNECT THE UNIT FROM THE PATIENT.
PREPARATIONS RESTING ECG SYSTEM
12 V5.03_2014_11 LABTECH LTD.
Symbols
This equipment meets the requirements of Directive 93/42/EEC Annex II
(excluding section 4).
Do not dispose of this product and batteries as unsorted municipal waste.
Prepare this product for reuse or separate collection as specified by
Directive 2002/96/EC of the European Parliament and the Council of the
European Union on Waste Electronic and ElectricalEquipment (WEEE).
MDD classification IIa. EMC class B. EMC group 1.
Bluetooth Wireless Communication Technology
Pay special attention to the part, marked with the exclamation mark
CF type on patient‟s side
Date of manufacturing
Manufacturer
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 13
4 Instructions
4.1 EC-12R (Bluetooth) Recorder Components
FRONT VIEW
REAR VIEW
ECG recorder
Patient cable
- easy to replace -
Recorder status
indicating lamp
Symbols
Recorder ’s
configuration type
Serial
number
Battery cover Battery
EC-12R
INSTRUCTIONS RESTING ECG SYSTEM
14 V5.03_2014_11 LABTECH LTD.
4.2 EC-12R/S (Bluetooth) Recorder Components
FRONT VIEW
REAR VIEW
Recorder status
indicating lamp Patient cable
- easy to replace -
- easy to replace -
Symbols
Recorder ’s
configuration type
Serial
number
Battery cover Battery
EC-12R/S
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 15
4.3 EC-12S (Bluetooth) Recorder Components
FRONT VIEW
REAR VIEW
Recorder status
indicating lamp Patient cable
- easy to replace -
- easy to replace -
BP measurement
cuff with KO
microfone
Symbols
Recorder ’s
configuration type
Serial
number
Battery cover Battery
INSTRUCTIONS RESTING ECG SYSTEM
16 V5.03_2014_11 LABTECH LTD.
4.4 Database management interface
The database management interface enables us to add physicians and patients by name
and ID number (e.g., social security number), which can be sorted and filtered
according to different criteria. The interface also offers a search option by name or ID.
The database management interface displays the list of patients and the pertaining
number of records. Here we can view previous records and start new recordings as
well.
4.4.1 Toolbar
The toolbar contains buttons which provide access to the most often used functions in
the software.
New patient: Here you can add new patients to the list. Patient’s first and last names,
birth dates and patient IDs are fields which need to be filled in. The
software allows only unique patient IDs to be entered into the database,
which is the reason why it is recommended to use this field for entering the social
security number of each patient (or any other code which uniquely identifies a person).
Entering address and height is optional when registering a patient.
Modify: Patient data can be edited by clicking on the Modify button. The interface
allows every field to be edited except patient ID (social security number)
2.2.2
2.2.8
2.2.5
2.2.6
2.2.1
2.2.4
2.2.7 2.2.3
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 17
Medical record: This function gives a summary about the medical record. This
summary can be edited and we can also add explanations and
observations.
Delete: Click on the Delete button for deleting patients. When patient is deleted, all of
its corresponding records are deleted as well. After deleting, there is no
way to restore data.
Settings:
- Select physician: In this menu the physician performing
the current test can login to the software, and also select to
view the records of other physicians whose records are not
password protected.
- Change password: By selecting this option, the physician
performing the current test can change his/her password.
- Administration: Selecting this menu point, we can see
names of Test Physicians. Other names can be added to the
list in this menu point by clicking on “New” button; data of
existing physicians can be edited by clicking on
“Modify”,and “Delete” serves to delete the physician (a
physician can only be removed if there are no records
relegated to his/her name). Security settings and administrator rights can be set under
this menu point. Further information on modifying user authorization can be found in
Cardiospy Installation and Update Guide.
- Select database: In this menu, you can choose to create a new database or to use a
different one. These databases operate independently of each other, so a patient created
in one does not appear in another. The software always uses the database set for the
previous operation.
TO INSERT YOUR OWN INSTITUTION’S LOGO INTO THE PRINTED REPORT, ADD THE DESIRED
LOGO UNDER SETTINGS / ADMINISTRATION TO THE DOCTOR YOU WISH TO USE IT WITH.
EACH THE LOGO HAS TO BE ADDED TO EACH DOCTOR WE WISH TO USE THE LOGO WITH,
ONE BY ONE.
INSTRUCTIONS RESTING ECG SYSTEM
18 V5.03_2014_11 LABTECH LTD.
Clicking on the Add button creates a new
database. If you need to work with a
different one, select it and click on OK.
There are three possible ways to create a
database:
Local database folder: The database is
stored in a local directory on your
computer. Network drives are not
supported. Optimal for single computer
usage. Select the directory for the
database and it is ready.
- Network mode- Microsoft SQL Server:
SEE DESCRIPTION IN INSTALLATION AND
UPDATE GUIDE’S CHAPTER 4.3.
SQLite database: SEE DESCRIPTION IN
INSTALLATION AND UPDATE GUIDE’S
CHAPTER 4.2.
- File names as: The order of first and last name can be set in this menu.
- Input devices: Barcode readers and magnetic card readers can be set here.
- FTP settings: SEE DESCRIPTION IN INSTALLATION AND UPDATE GUIDE’S CHAPTER 5.1.
- Communication: HL7 communication protocol can be started here.
- Select systems:
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 19
Here you can select the ECG systems you
wish to use. Inactive systems will not
appear in the field of Selected Systems. A
system is inactivated by removing the
green tick from the box to its left.
Help:
- User manual: Choose the system in the Select systems
panel (further description in point 2.2.5). Once you click on
User manual button, the user manual of the currently
selected system will appear.
- Installation and update guide: The guide on software
installation and system configuration also contains a
description of the software update process.
- What’s new: The novelties of the current software version can be found in this option.
- Languages: The software offers a choice of several languages, from which you can
select the one you wish to work in.
- Multilanguage translator: This function enables the user to translate each term
appearing in the software into the desired target language. A password is necessary to
enter this menu point. In case you would like to use this function, turn to your
distributor for a password.
- About: Information on software version, the operational system and copyrights can be
found here.
INSTRUCTIONS RESTING ECG SYSTEM
20 V5.03_2014_11 LABTECH LTD.
ECG Device Connection Status Indicator: This indicator shows the current
connection status of the devices (the USB-02 interface, the ECG recorder, the blood
pressure monitor and the stress device).
The ECG Device Connection Status Indicator has three types: the first one is used with
the Holter, ABP and Resting Test systems. The second type is used with the Stress Test
systems, and the third type with the Continuous ECG systems.
1
2
1
2 3 4 5
1
2
.
3 5
1. Status of Bluetooth or direct connection
2. Status of connection with Holter ECG recorder /ABP recorder / Resting ECG
recorder / Test ECG recorder
3. Status of connection with blood pressure monitor
4. Status of connection with loading machine
5. Status of connection with Pulse Oximeter (this section is only available when Pulse
Oximeter is activated under System Configuration menu)
Colour codes for the ECG Devices Connection Status Window:
green - The device is properly connected
yellow - Manual operation, the device is not connected to the computer
red – No connection with the device
grey – No connection is needed for recording
4.4.2 List of patients
This menu point lists the patients in the database. Clicking on the appropriate tab of the
header sorts the available data by patient’s first or last name, date of birth, ID number,
or date of the last record.
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 21
4.4.3 Records toolbar
This is a toolbar for managing existing records.
Load record: Opens the selected record from the list. Once the record
is open we can edit, analyze or print it.
Import / export: In this menu point we can save the current record to the directory of
our choice. This directory can be on a network drive or on a
portable drive (e.g., USB stick). The name of the exported
record will include the first and last name of the patient, the
date of recording and patient ID, separated by low lines. We
can also import any previously saved (exported) record from
any directory. The imported record is automatically saved to
the current database.
- Import selected records: With this function we can import
records from a directory of our choice. Should you need to
import more records at a time, select records while holding
the CTRL key, then press Import selected records.
- Import records by date: Imports all the records of a selected directory which were
recorded in a given interval.
- Download from FTP: We can download records from the given FTP server. We can
see all the available directories and Cardiospy files in the appearing window, the left
side column is for navigation among folders, the right one is for Cardiospy files. The
files can be arranged according to their Names, IDs, Record dates and Upload dates, and
a search bar is also available to help find the necessary record. Holding the Ctrl button
enables highlighting and downloading multiple records at a time. Records can be
deleted by pressing the Delete button, and triggering the “Delete downloaded records”
automatically erases all records after downloading them.
INSTRUCTIONS RESTING ECG SYSTEM
22 V5.03_2014_11 LABTECH LTD.
- Export selected records: With this function we can export records selected in the list
of records. Should you need to export more records at a time, select records while
holding the CTRL key, then press Export selected records.
- Export records by date: Exports all the records which were recorded in a given
interval.
- Export records of selected patients: Holding CTRL while clicking on several patients
enables the multiple selections of patients. This option saves all the records of the
selected patients.
- Upload to FTP: We can select multiple records from a single patient and upload them
to the given FTP server.
- Upload to FTP by date: Every record is uploaded which fits to all the criteria given
and was recorded within a given period.
- Upload to Labtech: In case you have any problem with the interpretation of a record,
or you face any difficulties you can upload the record to the Labtech server.
- Print rest records by date: Prints every rest record which was made with the given
conditions and recorded in a given interval.
- Import sample data: In case a CD was used for installing the Cardiospy software and
there are no sample records in the database, you can import them by pressing this
button.
Delete selected record: This buttons deletes every selected record from the database.
After deleting, there is no way to restore data. Should you need to
delete more records at a time, select records while holding the
CTRL key, then press Delete selected records.
Delete records by date: Deletes every record recorded in the
given interval according to the criteria selected.
4.4.4 List of records
Displays the records of the patient selected in List of patients (point 2.2.2). Here you
can search among the records. Should you need to select more records at a time, you can
do so while holding the CTRL key.
4.4.5 Selected System
Here you can select the ECG systems you wish to use, configure systems, add new
recorders before you start recording. You can also start Demo recording in this field.
Please keep in mind that a demo recording can only be started if the USB-02 device is
not connected to the PC.
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 23
4.4.6 Filters
This field provides a number of filters which enable the filtering of records by different
criteria. Once the database is filtered, List of patients displays only those who fit the
criteria.
You can filter records by:
- Physician: When this filter is used, List of patients displays those patients whose
records have been made by the doctor selected.
- Date: When this filter is used, List of patients displays those patients who have
records made today / in the last 2 days / in the last 7 days / this month / this year.
- Status: When this filter is used, List of patients displays those patients who have
temporary / pre-evaluated / evaluated / approved records.
- Record type: When this filter is used, we can view all types of records or only those
pertaining to the system currently selected in Select system field.
- Pacemaker: When this filter is used, we can view patients who have and those who do
not have a pacemaker in separate lists.
4.4.7 Search
The Search (name or ID) field is used for narrowing down the list of patients
displayed. Once we enter a letter or a number in the field, List of patients will only
display patients whose first or last name, or the ID begins with the pertinent letter or
number. The more letters we type in the field, the more we narrow down the number of
patients appearing in the list. Clicking on the Reset button cancels the filtration and the
default List of patients will appear, where every patient and every record is displayed.
4.4.8 Database information
This field contains general information about the database, such as access path, number
and size of records contained, etc.
INSTRUCTIONS RESTING ECG SYSTEM
24 V5.03_2014_11 LABTECH LTD.
4.5 How to start a new recording?
4.5.1 Connecting the ECG device (USB type)
After the installation of the Cardiospy software, connect the ECG device to the PC by
connecting one end of the USB cable to the device and its other end to an available USB
slot on the computer. A flashing green light indicates proper connection between the
computer and the ECG device.
4.5.2 Connecting the ECG device (Bluetooth type)
After the installation of the Cardiospy software, connect the USB-02 interface into an
available USB slot on the computer (you can also use the extension cord for the USB-
02). A flashing green light indicates proper connection between the computer and the
ECG device
After connecting the ECG device to the computer, attach the electrodes to the patient
and connect the patient cable to the device. Please proceed to read the guide of electrode
placement.
4.5.3 Patient preparation
Step 1: Place the case of the recorder unit on the right side of the patient and fix it on
the belt on his/her waist. Although the case is washable, it is recommended to avoid
skin-contact.
Step 2: Clean the surface of the skin thoroughly using:
- razor /in case it is necessary/
- abrasive preparation pad and
- alcohol swab
The ECG System is designed for external skin-surface ECG recording. The electrodes
can be applied only on injury-free skin.
Step 3: Attach the single-use electrodes as it is indicated below, and by keeping to the
instructions shown on the electrode’s package.
NOTE: A NOISE-FREE RECORD CAN BE OBTAINED ONLY IF ALL THE STEPS OF PATIENT
PREPARATION HAVE BEEN FOLLOWED ACCORDING TO THE INSTRUCTIONS. DO NOT START THE
RECORDING IF ECG SIGN IS NOISY.
FOR MORE INSTRUCTIONS PLEASE CHECK THE PATIENT’S PREPARATION INSTRUCTIONS.
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 25
Standard 1
IEC label AHA
label Electrode placement
C1 V1 Fourth intercostal space at the right sternal border.
C2 V2 Fourth intercostal space at the left sternal border.
C3 V3 Midway between C4 and C2.
C4 V4 Mid-clavicular line in the fifth intercostal space.
C5 V5 Anterior axillary line on the same horizontal level as C4.
C6 V6 Mid-axillary line on the same horizontal level as C4 and C5.
L LA Slightly below left clavicle, far from main muscles.
R RA Slightly below the right clavicle, far from main muscles.
F LL Lower edge of the rib cage, or at the level of the umbilicus at
the left mid-clavicular lines.
N RL Lower edge of the rib cage, or at the level of the umbilicus at
the right mid-clavicular lines.
INSTRUCTIONS RESTING ECG SYSTEM
26 V5.03_2014_11 LABTECH LTD.
Standard 2
IEC label AHA
label Electrode placement
C1 V1 Fourth intercostal space at the right sternal border.
C2 V2 Fourth intercostal space at the left sternal border.
C3 V3 Midway between C4 and C2.
C4 V4 Mid-clavicular line in the fifth intercostal space.
C5 V5 Anterior axillary line on the same horizontal level as C4.
C6 V6 Mid-axillary line on the same horizontal level as C4 and C5.
L LA Left arm.
R RA Right arm.
F LL Left leg.
N RL Right leg.
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 27
5 leads
IEC label AHA
label Electrode placement
C1 V1 Fourth intercostal space at the right sternal border.
C2 V2 Fourth intercostal space at the left sternal border.
C3 V3 Midway between C4 and C2.
C4 V4 Mid-clavicular line in the fifth intercostal space.
C5 V5 Anterior axillary line on the same horizontal level as C4.
C6 V6 Mid-axillary line on the same horizontal level as C4 and C5.
L LA Slightly below left clavicle, far from main muscles.
R RA Slightly below the right clavicle, far from main muscles.
F LL Lower edge of the rib cage, or at the level of the umbilicus at
the left mid-clavicular lines.
N RL Lower edge of the rib cage, or at the level of the umbilicus at
the right mid-clavicular lines.
* The only white chest electrode can be placed anywhere between C1 and C6
positions.
INSTRUCTIONS RESTING ECG SYSTEM
28 V5.03_2014_11 LABTECH LTD.
4.5.4 New recording in resting mode
Create a new patient or select one from the database. Click on Start Recording.
Select the number of record: you can set the suitable number of records here: Settings
menu – Number of rest records
Set the lenght of record: you can choose to create 5, 10, 20 seconds or 60 minutes
long rest records in the upper menu.
Start monitoring by clicking on the start monitoring icon at the top left corner.
Start the examination by clicking on the button on the top left corner of the display.
In case you chose to make several records, click on buttons , , and to save each
record.
The automatic interpretation can be modified by clicking on the Edit Interpretation
button.
In case you chose to make a Long Rest record, clicking on Start rec button starts the
recording. The records is stopped and saved automatically after 60 minutes, in case you
would like to stop it earlier click on the End Rec button.
The software automatically saves the record to the database, then makes interpretation
and calculates some parameters - all of these are displayed on the screen.
Placing Freeze and Mark signals for arrhythmia events (only in case of 60 m,inute
long records)
In case an arrhythmia is discovered during the stress measurement, it can be indicated
by the space button. In this case you will be able to check these marked deviations after
finishing the test.
You can also add markers after the resting test or while you are examining the record
(Replay).
Making an emergency recording
In emergency cases, you can make recordings without giving any specific
data about the patient. Click on the “Emergency Patient” icon below the
Selected systems menu. In this case, the patient is named Emergency
000x in the database, where the 000x indicates a sequence number. Such
patient data can be edited later.
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 29
4.6 Resting ECG mode display
R1 Toolbar: The toolbar contains buttons that give you fast access to the most
common tasks in the program.
R2 Select patient: Select patient to load previous records or make new ones,
and also you can add new patients to the database here.
R3 Select patient / Database page: Manage patients, export and import ECG
records and archive data.
R4 Patient’s Name: Name of the current patient.
R5 Click to choose the previous or next record of the patient.
R6 Click to delete the currently selected record.
R7 ECG Devices Connection Status Window: Move the mouse here to check
the right operation between the ECG devices and the PC.
R8 Current Heart Rate: Displays current value for Heart Rate of the patient.
INSTRUCTIONS RESTING ECG SYSTEM
30 V5.03_2014_11 LABTECH LTD.
R9 Global Parameters/Intervals: Displays Intervals and Axes.
R10 Displays the interpretation.
R11
Edit interpretation: You can add information to the resting ECG, like a
description of any arrhytmia that has occured. This comment will be
saved with the record and also appear in print. The length of the comment
cannot exceed 50 characters.
R12 Medians: ECG Averages.
R13 Amplitude: Set to 5, 10 or 20 mm/mV.
R14 ECG record leads: Displays the ECG leads.
R15 Show/hide medians: Click to show/hide the ECG averages.
R16 Scrollbar: Move it to see the requested part of the Record.
R17 Continuous Playback: Click on for continuous playback.
R18 Comparison: Compare a patient's two rest ECG's.
R19 ST Calculation: the actual value of ST calculation is shown
R20 Time Scale: Long-resting records’ time and marks of the events are
shown
R21 BMI / BMI Prime: indicates the BMI index of the patient
R22 Test time: Shows the elapsed time.
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 31
During Stress/Continuous/Stress recording a whole view displays all previously taken
record divided to one minute intervals, under ECG displaying. This view enables
checking of arrhythmic events detected previously. It displays in 5 minutes intervals and
the scrollbar in the right ensures the navigation within the record. The view of the
record can be selected in the drop-down menu.
INSTRUCTIONS RESTING ECG SYSTEM
32 V5.03_2014_11 LABTECH LTD.
4.7 Toolbar
4.7.1 Start monitoring
Pressing this button to start monitoring the patient before the recording is
started.
4.7.2 Mode
The software can be used in three operating modes:
- Resting ECG mode
- Continous ECG mode
- Stress test mode
4.7.3 ECG leads
In this menu point: you can modify the leads displayed on the screen, the
ST calculation and measuring methods; view displaying from 1column to 2
columns or make display the mean curve and setup the ECG contrast and
millimetre paper.
The renewed Leads menu gives more possibilities and the ECG leads wanted to be
appeared on the screen during recording, and analyzing are more flexible. From now on
seven users can setup and rename the views. Moreover Cabrera and Cabrera +
Pericordial arrangements become default.
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 33
You can select if you would like to display the lead in Full view or in Freeze. You can
invert any of the leads.
Contrast modification of ECG curves
The modification of waves on the monitor is possible. Modify separately the line
stickiness of VES, SVES, normal beat curves. The program enables default settings to
distinguish each event. Modify the settings in Setting menu, ECG contrast tab.
INSTRUCTIONS RESTING ECG SYSTEM
34 V5.03_2014_11 LABTECH LTD.
Millimeter paper setting
This window is active in stress and holter system, by clicking view menu, Millimetre
paper menu. Adjustment of millimetre paper size correlating to the monitor is possible.
It is possible to setup real, 1:1 scale appearance of the millimetre paper on screen.
4.7.4 Filters
Here you can switch on/off the filters.
• Smoothing
• Baseline
• Mains (50/60 Hz) –available only at resting mode
Switch on all filters!
Manual setting: setting manually
how many pixels match for 1 mm.
Automatic setting: the cube
underneath is symbolising the
monitor, showing the resolution
and display size. The resolution of
the monitor is defined by the
system setup, but the display size
must be set by rolling the cursor
above the cube. Finalise the new
setting by pressing "Set to
recommended" button. Then press
the set button, and the program
sets the size on 1:1 scale, if the
monitor shows 10mm it will be the
same size in reality too.
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 35
4.7.5 Settings
You can set the various parameters of the software influencing the analysis
in this menu.
Settings / Interpretation of the Resting ECG:
- All: The whole text of the automatic diagnosis is displayed on the screen.
- Presumptive: Only the sure part and presumptive diagnosis is displayed.
- Sure: Only those parts of the diagnosis are displayed which contain correct,
certain statements.
- None: No text of the diagnosis is displayed at all.
The automatic interpretation can be modified by clicking on the Edit Interpretation
button.
You can modify the text of the interpretation by pressing the Edit interpretation button.
Please note that once you modify the interpretation, Sure option will not be available
anymore.
Colour settings…:
You can find, modify and delete the colouring combinations displayed on the screen of
the program here. Aside from the recommended colour designs you can adjust personal,
user colour combinations from Custom1 to Custom5 and save them by clicking the Save
button.
Lead-off warning:
The displaying methods can be chosen:
Lead-off warning >>
Opportunities for display:
- Always on toolbar: the program gives a warning sign on the toolbar in the
moment of the electrodes’ lead-off.
- •Always in window: the program gives a warning sign in a window appearing
in the middle of the screen in the moment of the electrodes’ lead-off.
- Once in window: the program gives a warning sign of the electrode’s lead-off
once and if you close the appearing window it won’t pop up again.
- Never: the program does not display the electrode’s lead-off.
New analysis
Used to analyse the previous records with new settings (for example: J+ … changes.
Service… :
Here you can find the information about the type of software used during the
measurements, the serial number of the ECG recorder, the actual Firmware version and
the currently chosen record’s settings. Here you can follow the stressing records’
diagnosis settings as well.
INSTRUCTIONS RESTING ECG SYSTEM
36 V5.03_2014_11 LABTECH LTD.
4.7.6 Print Report
You can print the record on display by clicking on the printer icon, it will
display a print preview window. In this window you can set the parameters
of printing, like the leads to be printed, printing mode, printing colour,
number of reports and pages to be printed, and the printer itself can be
selected also.
4.7.7 Print current sample
Prints the ECG you actually see on your PC screen.
4.7.8 Export the Report
Export the record into PDF, image formats, DICOM or GDT.
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 37
4.7.9 Automatic report
Print out/export your reports automatically from print out/export settings
menu/automatic report tab. Select the function which would be necessary for you and if
required select the export library as well. The reports can be printed per sheets, or per
groups. It is possible to arrange into libraries, using patient ID and recording time.
It is possible to print out automatically, PDF, JPEG and DICOM export.
You can modify the output format under the Export tab. You can modify here the
resolution of output picture.
Here you can setup that the program arranges the stress and rest records in a different
file.
INSTRUCTIONS RESTING ECG SYSTEM
38 V5.03_2014_11 LABTECH LTD.
4.7.10 Print and export settings
ECG drawing: Here we can set the graph
paper, the EKG curve, the VES and the
SVES curve’s thickness on the print
findings.
Cutting of noisy ECG: In the whole ECG
view the program completely cuts the noisy
sections and they do not appear on the
finding.
Page layout: It is allowed to put a logo in
the finding’s headpiece; you can attach the
expert’s report to the Summary II finding or
you can display it as a separate finding. You
can take off the automatic expert’s report
from the finding Summary I.
Quality: Printer-specific settings. You can select high-resolution or low resolution
quality. You can send the findings to the printer in vector graphic or raster graphic
modes. The colouring method can be set as black and white or coloured. Furthermore
you can change the line smoothing and the contrast here as well for the print findings.
Printing possibilities:
ECG samples: In the case of 5, 10 and 20 second long records it means printing the
whole record, while with 60 minute long records it means printing only the samples of
the intervals which were given previously
Marked ECG samples: Prints the events related to the Marker.
Parameters of marked ECG samples: Prints the parameters of the events related to
the Marker.
ECG for selected lead: Prints the ECG of the selected lead.
Print current sample: This function can be reached in the Toolbar. Press button and
the currently displayed ECG section will be printed.
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 39
Print settings:
Leads: You can choose which leads you would like to print (Standard > I, II, III;
Unipolar > aVR, aVL, aVF; Standard and Unipolar; Precordial > V1-V6 and 12 leads.)
Selected Printer: You can select the default printer by clicking on Setup. You can
select local and network printers as well. The software will always save your last
selection.
Printing Colour: Colour or Black & White.
If the amplitude of the ECG is too large and does not have enough space on the report
display, you can modify it by ECG vertical offset scroll bar in the printing setup.
Each curve will appear correctly:
INSTRUCTIONS RESTING ECG SYSTEM
40 V5.03_2014_11 LABTECH LTD.
4.7.11 Help
In this menu it is possible to invoke the built-in user manual, change the
language of the software or have a look at the Keyboard Usage window.
User Manual
The User Manual is available in different languages. By changing the language of the
software, the user manual (help) will be shown in the same language, as well. However,
if the user manual is not available in the chosen language yet, the English manual will
be shown.
You need to install Adobe
Reader
application to see the User Manual, this is
automatically installed along with the Cardiospy software.
Languages
Here you can select the language of the software. However if the program is not
available in the chosen language yet, the English program version will be worked.
Multilanguage Translator…
The Multilanguage Translator menu offers a great help to translate the text of the
software into any language. This menu can be used only with the password provided by
the manufacturer.
4.7.12 Paper speed
Set here the speed of the paper (mm/sec) to the measurement. The paper
speed can be set for 12.5, 25 or 50 mm/sec.
4.7.13 Amplitude
You can change the amplitudes (mm/mV) here.
4.7.14 Record length
The length of the record can be set for 5, 10, or 20 seconds, or Long record
(max. 1 hour). This function is available only in Resting mode.
RESTING ECG SYSTEM INSTRUCTIONS
LABTECH LTD. V5.03_2014_11 41
List of parameters
The list of parameters does not only appear on the printed report, but we can also
examine it on the PC’s screen. Once the resting record is finished, Details button
becomes active on the right side. Clicking on it displays the list of parameters.
EDITING AND MEASUREMENT POSSIBILITIES RESTING ECG SYSTEM
42 V5.03_2014_11 LABTECH LTD.
5 Editing and measurement possibilities
5.1 Arrhythmia analysis in resting and stress test ECG systems
Real-time arrhythmia analysis is available in the software, this means that the VES and
SVES.. events are detected during the resting and stress test monitoring. We can also
add various stop and alarms criteria for such events.
Re-analysing older ECG records (Settings/New analysis) detects and displays V and SV
events.
We can set the Arrhythmia and V ectopic rate among the Alarm and stop criteria in
Settings / Parameters.
RESTING ECG SYSTEM EDITING AND MEASUREMENT POSSIBILITIES
LABTECH LTD. V5.03_2014_11 43
5.2 Edit and delete ectopic events
Arrhythmia analysis first appeared in Cardiospy version v5.02.00., now this function
has been extended with various additional functions. VES and SVES beats are not only
detected but can be edited during and after recording. It is possible to delete groups of
events in case it was a false event. The summary table displays the number of ectopic
beats and their frequency in percentage, in comparison to all the beats during the record.
Editing and modifying events can be conducted in the ECG view as well. Use the left
button of the mouse to select beats, and the right button to modify their type.
EDITING AND MEASUREMENT POSSIBILITIES RESTING ECG SYSTEM
44 V5.03_2014_11 LABTECH LTD.
5.3 Ruler
We can measure time and amplitude by holding the left mouse button while moving the
cursor, this way you select a certain section of the ECG curve. Time: Represents the
elapsed time of the record. Selected intervall: Represents the intervall since the mouse
was pressed in the blue box provided in ms. Indicated RR intervall: Here we can see
the number of beats, the length of the section between the first and last QRS and the
average HR measured in the intervall from the selected section. Amplitude is measured
by calculating the difference of the two points where the first and second ruler intersects
the ECG curve.
RESTING ECG SYSTEM ERROR CORRECTION, MAINTENANCE
LABTECH LTD. V5.03_2014_11 45
6 Error correction, maintenance
Before going through this chapter please read the Installation and Update Guide
carefully in which we detail how you should install and operate the program and the
devices. If you cannot solve the problem based on these manuals, then read the
information below.
6.1 USB-02 signs and error messages
Light Description Countermeasure
red Power LED
USB-02connected to
the PC but it does not
operate
Terminate the USB connection. After 5-
10 ms reconnect it to the PC. If the USB -
02 is still red, follow the below
instruction:
Install the USB-02 drivers again by
deleting and reinstalling the Cardiospy
program. Go to Start menu / Programs /
Labtech Ltd / Uninstall Cardiospy –
Wizard: put checkmark in the Remove
USB-02 drive switch and dele the
program. Terminate the USB-02 device
during the reinstallation process.
green Power LED
USB-02 connected to
the PC and operates
properly
-----
Wireless LED in
orange flashing
USB-02 sign offline
status.
The recorder is not connected to the
device. Switch on the (insert full charged
batteries) EC-12R / EC-12R/S / EC-12S
for the communication connection.
Wireless LED in
green flashing
Bluetooth
communication
between ECG
recorder.
You can start recording.
red Wireless LED
Bluetooth module has
failure in USB-02
interface.
Please contact the service support.
ERROR CORRECTION, MAINTENANCE RESTING ECG SYSTEM
46 V5.03_2014_11 LABTECH LTD.
6.2 ECG Recorder signs and error messages
Light Description Countermeasure
NO light
sign ----
Please check the status of the batteries or change
them for new charged ones. If the problem is still
exists despite of using new charged batteries, please
contact service support.
Orange
flashing
Three pieces of
short beep after
inserting charged
batteries
Offline operation. Start the Cardiospy program to
enable the connection to USB-02 interface. The
light of the recorder becomes green.
Green
flashing
Two pieces of
short beep
Normal operation, the recorder and the USB-02
interface are connected with each other, you can
start the recording.
Continuous
red
Continuous
beeping
See the General communication failure: NO 1. The
ECG led lights in red, and beeping
* The light sign of the recorder means the Led sign on the front side of the device.
6.3 General and communication failure
NO Error / Symptom Countermeasure
1.
THE ECG LED
LIGHTS IN RED
AND BEEPING
Discharged battery charged status. Please use properly
charged batteries or use new fully charged batteries.
Please insert the batteries into the charger and wait until
the charging time finished. If beeping does not finish
after batteries changing, the recorder has been damaged.
POWER LED
WIRELESS LED
RESTING ECG SYSTEM ERROR CORRECTION, MAINTENANCE
LABTECH LTD. V5.03_2014_11 47
NO Error / Symptom Countermeasure
Please contact service support.
2. NO OR BROKEN
ECG DRAWING
Distance between USB-02 interface and recorder should
not be more than 6 meters. For proper communication
between USB-02 interface and recorder (in linear
distance) should not be an object that makes radio
frequency interfere. Using the lengthened cable put the
USB-02 interface a well visible place. For better
communication you can stick it to the monitor or to a
stand.
3. *
CANNOT START
THE BLOOD
PREASSURE
MEASUREMENT.
PROGRAM SHOWS
THE
MEASUREMENT
STARTED BUT
THE DEVICE DOES
NOT MEASURE
Check the battery status (Status display or
Setup/service/battery status menu). Open the record and
navigate to the Service menu.
Properly charged batteries: marked with green line.
Battery voltage only slightly decreases during blood
pressure measurement.
Improperly charged or old batteries: give high voltage at
the first measurement. Please charge the batteries and
check their status. If the symptom is the same please
change the batteries to new ones.
ERROR CORRECTION, MAINTENANCE RESTING ECG SYSTEM
48 V5.03_2014_11 LABTECH LTD.
NO Error / Symptom Countermeasure
If the batteries has not got problem, but not able to
measure the blood pressure then please check the
drawing. If it is broken that might indicate a Bluetooth
connection failure. Distance between USB-02 interface
and recorder should not be more than 6 meters. For
proper communication between USB-02 interface and
recorder (in linear distance) should not be an object that
makes radio frequency interfere. Using the lengthened
cable put the USB-02 interface a well visible place.
4. *
NO CONNECTION
WITH THE BLOOD
PREASSURE DEVICE
OR NOT SELECTED
Check if the proper blood reassure device is selected in
System Configuration menu and if the device
application is switched.
Check the connection of the devices.
If you use different type blood pressure monitor, you can
enter the data manually into the SW by the manual blood
pressure.
5. *
SOFTWARE
MESSAGE:
MICROPHONE IS
NOT CONNECTED
Check the connection of the KO microphone’s
connector to the ECG device. It connects from the BP
cuff with a black cable to the ECG device.
RESTING ECG SYSTEM ERROR CORRECTION, MAINTENANCE
LABTECH LTD. V5.03_2014_11 49
NO Error / Symptom Countermeasure
6.
**
STRESS TEST
DEVICE
CONNECTION
FAILURE
Check the connection between the computer and the
stress test device (connection of communication cable),
and the charging status (if the stress test device’s main
switch is on), and if under the System Configuration
menu/Loading device menu is the right loading device
was selected. In case of treadmill: check if the
Emergency Stop button is ejected. Perform chapter 4.1.3
in the Installation and Update Guide.
7. NOISY ECG SIGN
Examine if all the ECG electrodes were attached to the
patient appropriately. To produce high quality record the
electrodes should be attached carefully by qualified
staff.
8. PRINTOUT DOES
NOT START
Check if the selected printer is switched on. If you have
selected a local printer, then examine if the connection
to the PC is appropriate (connection of communication
cable).
* in case of EC-12S built-in blood pressure ECG device
** EC-12R/S, EC-12S Stress ECG systems
ERROR CORRECTION, MAINTENANCE RESTING ECG SYSTEM
50 V5.03_2014_11 LABTECH LTD.
Further steps
If the above description does not solve the problem, please send an e-mail with the
following
PLEASE DESCRIBE THE FAULT:what has happened.
- SEND A BLACKBOX: Cardiospy Blackbox: after the installation you can find a
Generate black box shortcut under the Start Menu / Labtech / Cardiospy/ . Please send
the blackbox file to our e-mail address.
- SEND A REPORT ABOUT THE FAULT: Export the report from Cardiospy program
(Patient menu – Import Export – export record), and send us the record. If the record is
too big (more than 2-4mb) ask for our ftp details via our e-mail address
-SEND SCREENSHOT ABOUT THE FAULT: You can make it with Print Screen
button, then import it to Paint with CTRL+Vand save the file.
Maintenance and Calibration
If the ECG device is used according to the preset parameters as prescribed, there is no
need for calibration or maintenance.
The case of the recorder unit may be cleaned, if necessary.
The accuracy of controls is assured by using long lasting parts and techniques that are
able to preserve their accuracy during the life cycle of the product.
Cleaning of device and patient cable
Do not dip the device in liquid, for device-cleaning use moistened wipes with a
moderate acting (not agressive) disinfectant or cleaning agent. Liquid getting inside of
the device can cause electric failures! Length of patient cable: 65-85 cm
BP cuff cleaning
Before cleaning of textile cuff please take off the rubber hose and Korotkoff-
microphone.
Wipe or wash the cuff off with moderate effect (not aggressive) and temperature (max.
30 °C) antiseptic or cleaning agent. After its dry on room temperature put back the
rubber hose and Korotkoff-microphone.
Validation
Blood Pressure device validation in accordance with the requirements of the applicable
standards and as the validation term expired.
RESTING ECG SYSTEM TECHNICAL SPECIFICATION
LABTECH LTD. V5.03_2014_11 51
7 Technical specification
Technical specification EC-12R EC-12R/S EC-12R
(Bluetooth)
EC-12R/S
(Bluetooth)
EC-12S (Bluetooth)
Leads I, II, II, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
BP measurement
method -
Auscultational
BP interval of
measurement -
30-260 mmHg;
Device’s
measuring limit:
20-280 mmHg
BP
measurement’s
accurary
-
± 3 mmHg
PC connection USB cable Bluetooth (wireless type)
Patient protection EN 60601 (IEC-601-1), CF type
Power supply /
Can be exchanged
during recording!
/
USB cable
2 x 1,2 V rechargeable batteries
(or 2 x 1.5 V AA alkaline battery)
Dynamic interval ±20 mV
DC offszet range ±1000 mV
Frequency
response 0.05 Hz … 150Hz
Sampling rate 1000 Hz
A/D resolution 16 bit
Input impedance ≥100 MΩ
Size 125 x 70 x 25 mm 125 x 70 x 33 mm
Weight ~ 100 g ~ 110 g ~ 200 g
TECHNICAL SPECIFICATION RESTING ECG SYSTEM
52 V5.03_2014_11 LABTECH LTD.
Software specification
Simple, user friendly software with multiple functions
Precise QRS classification
Built-in expert system,
Short and long resting ECG record
SAECG
Various printable reports
Customizable alarm and stop criteria
Comparing measurements
DICOM, GDT, MFER, HL7
Local and network operation
Export and import functions
Arrhythmia analysis
Pacemaker detection
RESTING ECG SYSTEM TECHNICAL SPECIFICATION
LABTECH LTD. V5.03_2014_11 53
Software minimum configuration
Processor: Min. Core 2 Duo (2 Ghz)
RAM: Min. 1,5 GB (Windows XP) recommended: 2 GB
Min. 2 GB (Windows Vista, Windows 7)
HDD: 200 GB
Display: min: 19” display 1024X768 resolution, Color XGA
recommended 22” display 1680X1050 resolution
Interface: 1 USB port for EC-ABP
2 USB ports for EC-1-12H
Printer: Laser or LaserJet (Min 15-20 pages/minutes)
Operating
systems:
Windows XP ( Service Pack 2 )
Windows Vista
Windows 7
( 32 and 64 bit ) with Windows Update enabled