Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development i of iv Remisol Advance for Hematology Customers This Training Guide belongs to: _________________________________

Upload: vudat

Post on 24-Aug-2019

227 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

i of iv

Remisol Advance for

Hematology Customers

This Training Guide belongs to:

_________________________________

Page 2: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers Beckman Coulter Professional Development Rev. C (June 2015) ii of iv

WARNINGS AND PRECAUTIONS This training guide is not intended to replace the information in your Instrument Instructions for Use Manual (IFU) or User’s Guide. The information in the IFU or User’s Guide supersedes information in any other manual.

Content is for information and illustration purposes only. Each person assumes full responsibility and all risks arising from the use of the information. The information is presented “AS IS” and may include technical inaccuracies or typographical errors. Beckman Coulter reserves the right to make additions, deletions, or modifications to the information at any time without prior notice.

TRADEMARKS Beckman Coulter and the stylized logo are trademarks of Beckman Coulter, Inc. and are registered with the USPTO. All other trademarks, service marks, products or services are trademarks or registered trademarks of their respective holders.

Except as expressly provided above in the restrictions for use, nothing contained herein shall be construed as granting any license or right under any Beckman Coulter copyright. Nothing contained herein shall be construed as granting any license or right under any patent or trademark of Beckman Coulter or any third party.

COPYRIGHT Products, processes or technology in this document may be the subject of other intellectual property rights reserved by Beckman Coulter, and may not be licensed hereunder.

NOMENCLATURE

The Remisol Advance Data Manager is a registered trademark of Normand Info. Throughout the remainder of the training material, the Remisol Advance Data Manager will be referred to as the Remisol Advance.

Page 3: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

iii of iv

REVISION STATUS

Revision AA (October 2012)

Software release 1.5

Presented October 1, 2012, Beckman Coulter Professional Development Miami, Florida

Revision AA.1 (April 2013)

Software release 1.6

Presented April 1, 2013, Beckman Coulter Professional Development

Miami, Florida

Revision B (March 2014)

Software release 1.7

Presented March 31, 2014, Beckman Coulter Professional Development

Miami, Florida

Revision C (June 2015)

Software release 1.8

Presented June 22, 2015, Beckman Coulter Professional Development

Miami, Florida

Page 4: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers Beckman Coulter Professional Development Rev. C (June 2015) iv of iv

DAILY AGENDA

The following lists the discussions, information and practice exercises that take place in the classroom/laboratory.

Monday

General Information Learn about each other Safety: Personal Protective Equipment and Fire Evacuation Introduction to the course and materials Location of additional reference materials First day Registration Questionnaire Day 1 Basic Operations

Remisol Advance Introduction Samples List Request List Operations Shutdown/Startup RADV monitor (v1.7 and up)

Tuesday

Day 2 Rules and System Configuration Parameter and Profile Definition Comments and Tag Definition Rules Manual Diffpad Practice

Wednesday

Day 3 Basic Setup and Programming User Setup Report Setup Database Management Saving Configuration Programming Functions

Last day self-evaluation, course and instructor evaluations Graduation

Page 5: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

1 of 54

COURSE GUIDE DAY 1

BASIC OPERATIONS

This document is not intended to replace the information in your Instrument Instructions

for Use Manual (IFU). Information in the Instructions for Use Manual supersedes information in any other manual.

Day 2

Rules &

System Configuration

Day 3

Basic Setup &

Programming

Day 1

Basic Operations

Page 6: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

2 of 54

PERSONAL PROGRESS SUMMARY Basic Operations Modules Date Initials Facilitator Page#

Remisol Advance Introduction 3

Samples List 15

Request List 27

Shutdown/Startup 41

RADV Monitor (v1.7 and above) 47

Page 7: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

3 of 54

REMISOL ADVANCE INTRODUCTION

OBJECTIVES At the completion of this module, you will be able to:

Identify the main screen elements

Utilize the buttons and/or keyboard shortcuts

Arrange the Client Zone or Main Application Window for efficient use

Launch Remisol Advance

Access Help

Why Is It Important?

Learning how the Remisol Advance assists in managing laboratory processes can enable you to manage samples and results efficiently and improve laboratory workflow.

Identifying the software format and button functions will assist in using the Remisol Advance.

Resources

Remisol Advance Help

Page 8: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

4 of 54

Page 9: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

5 of 54

INFORMATION/PRACTICE SECTION

Software Overview

Remisol Advance Main Application Window

Elements and Functions

1. Title Bar – displays the control menu button, software name and laboratory name and minimize, maximize and close buttons

2. Menu Bar – contains the software menus 3. Toolbar – contains the software buttons 4. Client Zone – located between the toolbar and the status bar. Any

open window is displayed in this area 5. Status Bar – displays (from left to right) the current installed

version of software, the status of the software, connection status LEDs, the computer name, the name of the database, the current user name and the current date and time

12

3

4

5

Page 10: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

6 of 54

Menu Bar

The following table lists the Remisol Advance menus.

Menu Description

Environment Contains the submenus to setup peripherals (Instruments, host, etc.) and to carry out some system maintenance functions

Desk Contains submenus to create and work with patient requests and samples currently being processed

Archives Contains submenus to archive current patient requests and carry out other archiving functions

Configuration Contains submenus to configure parameters, profiles, comments, etc.

Quality Control Contains all submenus to carry out quality control functions

Audit Trail Contains the maintenance submenu allowing the manual entry of tracking events

Window Lists all the windows currently opened in Remisol Advance and displaying options

Keyboard Lists all edit options currently available and their shortcuts

Help Provides access to the integrated help and information on the software version, copyright, etc.

Page 11: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

7 of 54

Toolbar Functions

Toolbar buttons are shortcuts to frequently used operator functions. The table below identifies the buttons and their function.

Toolbar button

Button Name Function

Samples List

Displays a list of all current (non-archived) samples. May be filtered to user specifications

Request List

Displays a list of all current (non-archived) patient requests. May be filtered to user specifications

Custom List Use to query information in database

Messages List

Displays a list of all messages generated by rules for current (non-archived) requests

Previous Results

Displays a Previous Results Search window use to search for results in the archives

New Request Displays the Request Sheet window - use to program a new request at Remisol Advance

Search

Searches current (non-archived) requests for a specific patient request

Previous Request

Displays the previous request in the Request List

Next Request Displays the next request in the Request List

Comments Displays Comments window use to enter result

comments (technical notes/result interpretation)

Validate

Use to manually validate one or more requests or one or more results in selected request

Unvalidate

Use to manually unvalidate one or more requests or one or more results in selected request

Host Transmission

Use to manually upload selected results to the host computer

EQC

Displays the optional Extended Quality Control (EQC) program

EQC with blocked points

Print Displays the Requests Printing window use to print

selected requests

New

Creates a new sheet in the Request Sheet window, configuration screens and environment setup screens

Page 12: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

8 of 54

Toolbar button

Button Name Function

Save Use to save information

Delete Use to delete information

About

Displays the system software version, serial number and copyright information

Help Displays the User Guide

Stop Instrument blocking not activated

Instrument blocking activated

Manual Count (Diffpad)

Displays the manual diff window can be opened from a current request

Launching Remisol Advance

Step Action

1 Turn the power on at the Server console.

2

At the Log on screen, enter a User name and Password. This is a Windows log on.

User name: sysadm Password: session

Note: This is a generic administrator User name/Password.

3 Select Enter on the keyboard to continue.

4 At the Remisol Advance login window, enter the following:

Login: sysadm Password: session

Note In your laboratories, you will perform the same steps to launch the Client consoles.

Page 13: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

9 of 54

Accessing Remisol Advance Help

Step Action

1 Select the Help icon from the toolbar. This will display the Help topic corresponding to the window that is open.

2 Select (single click) the General Information chapter and then select Remisol Advance Software Overview. Read all four topics in this section.

3 From the General Information chapter, select Working with Windows and read Moving a Window.

4 Close the Remisol Advance Help by selecting the Exit button on the Help window. Note: Selecting the Exit button on the Title Bar will exit the Remisol Advance software.

Page 14: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

10 of 54

Communication Event Log

The Communication Event Log window opens automatically when the user logs on to Windows and anytime an instrument event occurs. Instrument events include downloading samples, uploading results, initializing messages, and line down messages.

Step Action

Note If the Communication Event Log has been closed, you may re-open the window by selecting Environment from the Menu bar then select Communication Event Log.

2

Using the reference from the topic Moving a Window, place the Communication Event Log window in the lower left quadrant of the Client Zone as shown below.

Communication Event Log

Page 15: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

11 of 54

Locking / Unlocking the Remisol Advance software

Step Action

1 Select Environment from the Menu Bar. Then select Lock Station.

2

Your system is locked when you see the Login/Password window.

You may also use the keyboard shortcut Ctrl+B to lock the station.

Note You may place a timeout to log a user off after xx minutes. This is an optional setting that can be done by your applications specialist.

Congratulations, you’ve finished! Continue to the Skill Check.

Page 16: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

12 of 54

Page 17: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

13 of 54

REMISOL ADVANCE INTRODUCTION SKILL CHECK 1. Label the areas of the Home screen:

1. ____________________________

2. ____________________________

3. ____________________________

4. ____________________________

5. ____________________________

6. ____________________________

12

3

4

5

6

Page 18: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

14 of 54

2. Draw a line between the toolbar button and its function.

Button Function

Use to manually validate

Use to program a new request

Use to save information

Use to delete information

Displays the User Guide (Help)

Use to display the Samples List

Use to display the Request List

Use to upload selected results to the Host

3. In the Communication Event Log, what types of events are displayed?

___________________________________________________________

___________________________________________________________

4. What is the keyboard shortcut for the Samples List ? _____

Page 19: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

15 of 54

SAMPLES LIST

OBJECTIVES At the completion of this module, you will be able to:

Explain functions of the Samples List

Define Samples list window elements

Configure Samples List filters

Customize the Samples List

Open the Samples List window

Sort the Samples List by priority

Use the Samples List as a real time pending log

Why Is It Important?

The Samples List is a display of all current (non-archived) requests. The Samples List can be configured to display pending samples received into the laboratory and can be configured to display STATs at the top of the list. These features can be useful in the pre-analytical testing phase.

Resources

Remisol Advance Help

Page 20: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

16 of 54

Page 21: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

17 of 54

INFORMATION/PRACTICE SECTION

SAMPLES LIST The Samples List window gives an overview of all current (non-archived) samples. The Samples List window contains the following information:

Routine or STAT flags (green tubes are routine, red tubes are STAT)

Sample flags

Sample ID

Position (tube position)

Assigned Instrument

Patients last name and first name

Sample nature

Last run date and time

Page 22: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

18 of 54

Samples List Flags

In addition to the tube icon which is colored green for routine samples and red for STAT samples, the left side of the window (Flags Column) allows for 5 sample flags. The Samples List has five flag columns the operator can use to determine the status of each sample.

Flag

Column Flag Description

1

? T B E U

No response from instrument Downloaded Instrument is busy Transmission error occurred Sample ID was already used for a diluted sample

2

? T B E U

No response from automation line Downloaded Automation line is busy Transmission error occurred Sample ID was already used for a diluted sample

3 R Results received from the instrument but not yet reviewed

4 P R

Awaiting rerun results from the instrument Rerun results received from the instrument

5 M Sample ID changed manually

Samples List Filters

The system administrator can customize the samples displayed on the Samples List by using the “Filters” feature. The default filter is set to display all current (non-archived) requests. Each defined filter will have a tab with a name customized by the Remisol administrator. Filters can be further customized by changing tab colors, text backgrounds and colors, as well as text effects. Multiple filters can be defined. Some examples of Samples List Filters are:

Pending samples

STATs

Samples associated with a specific instrument

Samples requiring rerun

Tagged samples

Page 23: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

19 of 54

Page 24: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

20 of 54

Samples List

Step Action

1 Select to open the Samples List window.

2

Position the Samples List in the upper right hand quadrant of the Client Zone.

Create Samples List Filters

Step Action

1 Right click anywhere in the Samples List window to open the Samples List popup window.

2 Select Filters and then select New.

3

In the Sample Filters window fill out the following information: Filter Name: Pending Instruments: Select Heme only (only Heme will be backlit in blue). Display Samples: All checkboxes should be checked

Page 25: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

21 of 54

Step Action

4 Select the green check mark .

5 To modify the Default filter, single click on the Default filter and then right click and select Filters → Modify.

6

Change the filter name from Default to All Samples. Ensure the following information is selected: Instruments: Select all instruments so they are backlit in blue Display Samples: All checkboxes should be checked

7 Select the green check mark .

8 Select Filters and then select New.

9

In the Sample Filters window fill out the following information: Filter Name: Stat Instruments: Select Not assigned and Heme (these two selections will be backlit in blue) Display Samples: Select Stat only Custom colors: Select Stat only. Then click on Background and select red, then select OK. Click Background under Tab colors and select red, then select OK.

10 Select the green check mark .

Note For any Filter, Custom Colors, Tab colors, Default list colors and Alert limits can be applied to further customize filters.

11 How many tabs are displayed in the Samples List? _______

12

To simulate an LIS download, on the LIS & Instrument Simulator

desktop, double-click on the PowerSimu icon , then double-click on R2000 LIS DWNLD and then single-click on Hem Patient Batch.

13 How many samples are displayed in the All Samples filter? ______ How many samples are displayed in the Pending filter? ______ How many samples are displayed in the Stat filter? ______

Review

The Samples Filters created are: Pending - Samples received in the laboratory Stat - Pending Stat samples All Samples - All samples on the Remisol, no matter what the status (probably will not use this tab much)

Page 26: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

22 of 54

Displayed Fields

The Samples List can be configured to display user selected columns of data. The columns can be positioned for user preference. To change the position of a field in the Samples List, click and drag that field name in the Field bar to its new position.

To open the Displayed Fields window, right click on the Field bar on top of the Samples List, a popup window will display. This popup window contains the list of every field available for display for the current filter. Select or deselect a field name to either show or hide the field. The field names with a checkmark beside it are the ones currently displayed.

Step Action

1 With the Samples List open, choose the columns (display fields) of your choice and position the displayed field in the Samples List according to your preference.

2 Right click on the field bar to open the Samples List display field selections.

3 Deselect a displayed field of your choosing. Observe the changes in the Samples List.

4 Re-enable the field you deselected in step 3.

Congratulations, you’ve finished! Continue to the Skill Check.

Page 27: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

23 of 54

SAMPLES LIST SKILL CHECK

1. In the Samples List, a green graphic in the Flags columns indicates

the sample is a STAT.

True False

2. In the Samples List flag area, define the following flags:

T________________________________________________

R _______________________________________________

Page 28: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

24 of 54

Page 29: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

25 of 54

REQUEST LIST

OBJECTIVES At the completion of this module, you will be able to:

Define the functions of the Request List

Define Request List window elements

Configure Request List filters

Open a Request sheet

Manually validate results

Utilize Tube History for tracking

Why Is It Important?

The Request List gives a concise overview of all current (non-archived) patient requests. The Request List can be configured to display samples with results requiring operator review and/or validation.

The Request List is used to easily identify those samples that have results that do not meet the auto-validation criteria and require review. The Request List may be used as a critical or exceptions log.

Resources

Remisol Advance Help

Page 30: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

26 of 54

Page 31: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

27 of 54

INFORMATION/PRACTICE SECTION

REQUEST LIST The Request List displays the status of each sample as it is received in the lab. Information displayed includes:

Flags

Patient

Patient ID

Position

Sample Id

Department

Tag

Page 32: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

28 of 54

Request List Flags

The Request List has nine flag columns that the operator can use to determine the status of each sample.

Flag Column

Flag Description

1 S The request is a STAT 2 O At least one result is out of validation range

3 P C

The request is pending one or more results The request is complete

4 R Results have been received but not yet viewed 5 I The request has been printed 6 E A transmission error has occurred 7 V The request is complete and validated 8 T The request has been tagged

9 H Shown if at least one result for each parameter has been sent to the host. Removed if at least one parameter is pending

Request List Filters

The system administrator can customize the samples displayed on the Request List by using the “Filters” feature. The Default filter is set to display all current (non-archived) requests. Each defined filter will have a tab with a name customized by the Remisol administrator. Custom colors and icons can also be used as a tool to further distinguish requests/results within a filter.

Multiple filters can be defined. Filters may be used to view:

Results out of validation range (criticals) and require manual validation

Results requiring smear review

Requests that have been validated

Requests that need to be uploaded or have been uploaded

Tagged samples

Completed or Incomplete samples

STATs

Requests that have been downloaded and have one or more pending parameters

Requests with results that have been received but not yet viewed

Page 33: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

29 of 54

Results Review

Results may be reviewed directly from the Request List. The Request Sheet window displays:

Patient Demographics

Sample Information (i.e. draw date/time)

Results (parameter values)

Graphical data, histograms and dataplots may be sent for CBC,

CBC w/ Diff and/or RETIC

Comments (Sample, Instrument, Results, generated by rules, etc.)

Page 34: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

30 of 54

Results are highlighted with colors to indicate where they fall in relationship to the reference and/or validation range. Color codes include:

Code Color Description Out Low Green Out of validation range low Valid Low Yellow Out of reference range low, in validation range Normal White Inside reference and validation range Valid High Purple Out of reference range high, in validation range Out High Red Out of validation range high

Note 1: Alphanumeric results are color coded in light blue (results from the instrument

that have flags, e.g. R flag). Note 2: If you would like to have criticals (out of validation range high or low)

highlighted in red and actions (out of reference range high or low in validation range) highlighted in yellow, this option is done by your applications specialist.

Results Flags

Results may also be flagged. Result flags are listed in four columns to the right of the result. Result flags include:

Column Flag Description

1 V v

Result validated automatically Result validated manually

2

M C U D

Result manually entered or edited Calculated result UPC result Dilution factor applied to result

3 D X

Result passed delta check Result failed delta check

4 H Result sent to host

5

Optional EQC

QC QC QC QC

Flags as indicators of the EQC status for the parameter: White: no protocol defined/no data available Green: No mistake Yellow: Protocol in warning or Time-out Red: Protocol in error or blocked result

Page 35: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

31 of 54

Previous Results

When a request has been processed and validated, the results are stored in the database. Previous results for a Patient ID that have been archived will display next to the current results. The Previous Results columns will display the date and time the results were received by Remisol Advance.

Note: Patient demographics are accessed by the Patient tab located next to the Graphics tab.

Page 36: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

32 of 54

Result Validation

Results within the laboratory established validation range may be automatically validated.

Manual validation occurs from the Request Sheet window by selecting the result(s) and then clicking on the Validate button from the toolbar. You may manually validate selected results or the entire request.

Request List

Step Action

1 From the toolbar, select to open the Request List.

2

Position the Request List in the Client Zone in the upper left quadrant.

Page 37: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

33 of 54

Create Request List Filters

Step Action

1 Right click anywhere in the Request List window to open the Request List popup window.

2

Select Filters and then select Modify and change the filter name from Default to New Results. Select the following: Display Requests: Select To Validate and To Upload (to host) only Instruments: Select DxH 800 only so it is backlit in blue Display only requests: Only Received (R) should be checked. Deselect all other checkboxes

3 Select the green check mark to confirm.

4 Right click anywhere in the Request List window to open the Request List popup window and select Filters → New.

4

Change the filter name from Default to Smear Review. Select the following information: Display Requests: Select To Validate and To Upload (to host) Instruments: Select DxH 800 so it is backlit in blue Display only requests: All checkboxes should be unchecked Custom colors area select Request Tagged.

5 Select the green check mark to confirm.

6 Each new filter that that is created has all the display columns selected. Right click on the Field Bar to open the Displayed Fields window.

7 Select the Flags and Sample Id to display, then select any other options you prefer to be viewed.

Note 1

To reposition display fields click on the Field name and drag it to its new location.

Note 2

Display columns may be resorted alphabetically or numerically by clicking on the display column.

Page 38: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

34 of 54

Step Action

8 Create the following Request List filters referring back to Step 1 if needed.

9

Select Filters → New Filter name: Stat Display Requests: Select To Validate and To Upload (to host) Instruments: DxH 800 only Display only requests: Select Stat only Custom Colors: Select Request Tagged

10

Select Filters → New Filter name: Path Review Display Requests: Select To Validate and To Upload (to host) Instruments: DxH 800 only Display only requests: Select Request Tagged only Custom Colors: Select Request Tagged

11 Select the green check mark to confirm.

Review

The Request Filters created are: New Results - Samples just run on the DxH 800 failing autovalidation Smear Review - Samples held for slide review Stat – Samples that are Stat and have results Path Review – Samples that have been manually tagged that are requiring review by the pathologist

12

To simulate an LIS upload, on the LIS & Instrument Simulator desktop,

double-click on the PowerSimu icon , then double-click on R2000 DXH 600/800 and then single-click on samples 1300 – 1307.

Page 39: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

35 of 54

Step Action

13

To view these results from the Request List, double-click on Patient ID 1300 to open the Request Sheet. To view the next sample, select the Next icon from the Toolbar. Use the chart on p.30 to answer the following questions: What color represents results out of validation range low? ________ What color represents results out of validation range high? ________ What color represents results that have a flag from the DxH? _______ What test result for Patient ID 1300 is out of validation range low?____ What color represents out of reference range high, in validation range? ______________

Note: For Hematology, typically the Validation range is used for critical ranges

and the Reference range is used for action limits not normal ranges.

Manual Validation of Requests

Step Action

Note If the Request List is not open, select the Request List icon from the Toolbar.

1 From the Request List, double click a request to open the request sheet to view the detailed sample results.

2 Manually validate the request by selecting the Validate button from the Toolbar.

3

To validate partial results, select the specific parameters. Note: Hold down the [Ctrl] key to select multiple parameters then

select the Validate button from the Toolbar.

4 Select Yes to the message “Verify that all results to be validated are highlighted. If no results are highlighted, ALL results will be validated. Do you confirm the manual validation?”

5 Use the Next button to access the next request in the Request List.

Page 40: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

36 of 54

Step Action

6 Review and manually validate all requests remaining in the Review tab.

7 Close the last request after manual validation is completed.

Tube History

Allows the operator to follow the processes completed for a sample (tracking the sample).

Identifies the date and time when there was communication about the tube between the host to the Remisol, Remisol to the instruments and the instruments to Remisol.

Filters within the tube history can be customized to search for samples based on a specific sample status.

Page 41: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

37 of 54

Step Action

1 Open a result from the Request List.

2 Right click anywhere within the Request Sheet to open the selections and single click on Tube History.

3 View the tracking for this sample.

Congratulations, you’ve finished! Continue to the Skill Check.

Page 42: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

38 of 54

Page 43: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

39 of 54

REQUEST LIST SKILL CHECK

1. In the Request List, what is the description of the following flags:

S______________________________________________

R______________________________________________

V______________________________________________

H______________________________________________

2. How many filters can be defined in the Request List?

________________________________________

3. In what list do you review sample results?

Samples List

Request List

4. A result highlighted in Red indicates

a.) Inside Reference Range and Validation Range

b.) Outside Validation Range Low

c.) Outside Validation Range High

d.) Outside Reference Range Low, in Validation Range

e.) Outside of Reference Range High, in Validation Range

5. Define the following Result Flags:

a.) V ___________________________________

b.) v ___________________________________

c.) H ___________________________________

d.) D ___________________________________

Page 44: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

40 of 54

Page 45: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

41 of 54

SHUTDOWN/STARTUP

OBJECTIVES At the completion of this module, you will be able to:

Shutdown Remisol Advance software

Startup Remisol Advance software

Why Is It Important?

There may be times when the Remisol loses communication with the LIS or the instruments. Communication can be restored by exiting and restarting the Remisol software but that process may not work every time. Shutting down the Remisol may be a helpful tool when troubleshooting and may be the only way to restore communication.

Resources

Remisol Advance Help

Page 46: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

42 of 54

Page 47: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

43 of 54

INFORMATION/PRACTICE SECTION

Shutdown Remisol Advance Client – DO NOT PERFORM THIS STEP

Note: We do not have both Clients in this laboratory. In your laboratories, you will have both Client(s) and a Server, the Client(s) must be shutdown first, then the Server.

Step Action

Note Do not perform these steps for a Client here in this classroom as there are no Clients. These steps would be performed in your laboratories.

1 Select the Close button on the Title bar on the Client.

2 Select Yes to the pop-up message: “This action will stop Remisol. Continue?”.

3 Select Start on the windows toolbar.

4 Click on the arrow beside the Log off button.

5 Select Shut down from the drop down list.

Shutdown Remisol Advance Server – PERFORM THIS STEP NOW

Step Action

1 Select the Close button. on the Title bar on the Server.

2 Select Yes to the pop up message: “This action will stop Remisol. Continue?”.

3 Select Start on the windows toolbar.

4 Click on the arrow located to the right-side of the Log off button.

5 Select Shut down from the drop down list.

6

On the ShutDown Windows window, enter a comment in the Comment: textbox as to the reason for shutting down the Remisol and then Select OK.

Page 48: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

44 of 54

Restarting the Remisol Advance Server / Client

If you have both a Client and Server, the Server must be restarted first.

Step Action

1 Turn on the power at the Server console.

2 Enter the User name and Password in the appropriate fields. User name: sysadm and Password: session Note: This is a Windows logon

3 Select the Enter key on the keyboard.

4 Enter the Login and Password in the appropriate fields. Login: sysadm and Password: session Note: This is the Remisol Advance logon

5 Select or the Enter key on the keyboard.

5 Repeat steps 1 – 4 for the Client(s).

Congratulations, you’ve finished! Continue to the Skill Check.

Page 49: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

45 of 54

SHUTDOWN/STARTUP SKILL CHECK

1. The Client and Server can be shutdown and restarted in any sequence.

True False

2. In what sequence should the Remisol Advance Client and Server be shutdown and restarted?

Shutdown

1st ______________________________

2nd ______________________________

Restarted

1st _______________________________

2nd _______________________________

Page 50: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

46 of 54

Page 51: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

47 of 54

RADV MONITOR

OBJECTIVES At the completion of this module, you will be able to:

Identify the status of the Remisol consoles, instruments and host(s)

Identify stopping, starting and resetting network connections and

interface

Identify when a reset of the system or a component of the system is

necessary

Why Is It Important?

There may be times when the Remisol Advance program may need to be closed and restarted as well as instrument and host interfaces reset

Resources

Remisol Advance Radv Monitor USER Guide v1.8

Page 52: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

48 of 54

Page 53: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

49 of 54

INFORMATION/PRACTICE SECTION

Radv monitor

Radv monitor is a monitoring & management console for devices and system

components across a Remisol Advance v1.7 and above network. A user with adequate rights will be able to monitor and control: Remisols Instruments Automations Host(s) Diagnostics, The database Services

Radv Launcher (RADVL) Radv Launcher (RADVL) is the service in charge of starting instruments, host(s), automations & diagnostics tasks and optionally Remisols themselves. It holds a central role in the Remisol Advance implementation. Remisol Advance will install the service with its Startup Type set to Automatic.

Launching Radv monitor The Radv monitor utility may be started in 3 different ways:

Start menu > All Programs > 10-Radv monitor Start menu > All Programs > Remisol Advance > Radv monitor C:/Remisol directory > Radv_Monitor

Page 54: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

50 of 54

Radv monitor interface By default the Radv monitor interface will open up on the Remisol tab.

The screen is split into 3 parts: Button bar – tab buttons, provide access to the various tabs; each is

dedicated to monitoring / controlling a specific type of device or system software component

List of devices (left panel) – depending on the tab selected, will display all corresponding devices on the network Details (right panel) – will provide details on the specific item selected in the

left panel list; it will also allow interaction, through buttons and options, with the device

Page 55: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

51 of 54

Remisols tab

Instruments tab

Automations tab Tab buttons

Hosts tab

Database tab

Services tab

Reset All button - The Reset All button is used to restart all Remisols (software) and Radv launchers on the network.

.

Page 56: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

52 of 54

SERVICES TAB – FACILITATOR DEMO

Page 57: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Rev. C (June 2015) Beckman Coulter Professional Development

53 of 54

RADV MONITOR SKILL CHECK

1. Who is allowed to launch Radv Monitor? _______________________________________________

2. Match the Radv Monitor tab icon to its title.

A. B. C. D. E. F. ___ Database ___ Services ___ Instrument(s) ___ Automation ___ Remisol(s) ___ Host(s)

3. What is the difference between the following Start types?

Manual Start _____________________________________________________ _____________________________________________________ Auto Start ______________________________________________________

______________________________________________________

4. If production cannot be stopped and it is not possible to Reset All, it is possible to:

Restart __________________________________________________________

Page 58: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 1 Beckman Coulter Professional Development Rev. C (June 2015)

54 of 54

Page 59: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

1 of 38

COURSE GUIDE DAY 2

RULES AND SYSTEM CONFIGURATION

Day 2

Rules &

System Configuration

Day 3

Basic Setup &

Programming

Day 1

Basic Operations

Page 60: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 2 of 38

PERSONAL PROGRESS SUMMARY

Rules & System Configuration Modules

Date Initials Facilitator Page#

Parameter &

Profile Definition

3

Comments &Tag Definition

9

Rules 15

Manual Diffpad 29

Practice 33

Page 61: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

3 of 38

PARAMETER AND PROFILE DEFINITION

OBJECTIVES At the completion of this module, you will be able to:

Define a new Parameter

Define Profiles

Define Priorities and Instrument codes

Define Host codes

Why is it important?

System Administrators for the Remisol may need to define a new test. In addition, determining the laboratory’s requirements for parameters are a vital part of the installation process. Identifying parameter requirements will be a factor in the smooth installation of the Remisol Advance.

Resourses

Remisol Advance Help

Page 62: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 4 of 38

Page 63: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

5 of 38

INFORMATION/PRACTICE SECTION - FACILITATOR DEMO

Parameter definition

A test is identified as a Parameter in the Remisol Advance. To define a new parameter in the Remisol Advance, the system administrator must ensure that all test information is provided as well as communication parameters for the test. This would include:

Parameter Definition

Parameter code

Parameter Abbreviation

Parameter Name

Measuring Units

Decimal Places

Sample Nature

Sample Type

Delta Check Formulas

Reference and Validation Rules

Rules (Used to trigger actions based on conditions)

Parameters can be defined for manual tests or calculated tests.

Parameter Communications

Successful communication requires:

Host Setup identifies Interface Type with the host

Instrument Setup assigns Interface Type to the specific instrument(s)

Instrument Code

Use to identify communication of parameter(s) between the Host and Remisol Advance

Assign code to same interface type identified in Host Setup

Use to identify communication of parameter between Remisol Advance and the Instrument

Assign code to interface type identified in Instrument Setup used between the Remisol Advance and the Instrument

Parameter codes are supplied with the instrument

Page 64: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 6 of 38

Priorities – Priority is used to define how Remisol Advance manages the feasibility, the assignment and download of a parameter or a profile to a specific instrument or group of instruments

Priority 1: The parameter is compatible with the instrument or group; Remisol Advance automatically assigns and downloads this parameter to the instrument if possible

Priority 2: The parameter is compatible with the instrument or group; the parameter can be downloaded after a host query, or manually (assigning and downloading have to be done manually or by rules)

Priority 9: The parameter is NOT compatible with the instrument or group; it cannot be downloaded, but results can be received from the instrument (if the instrument code exists)

Parameter Communication Host Remisol Instrument

Note: To view this data select Environment → Priorities and Instrument Codes

1. Host Setup identifies Interface Type between Host and Remisol

2. Host Code (defined in Instrument Codes Setup Menu) identifies parameter transmission code

1. Instrument Setup identifies Interface Type between Remisol and Instrument.

2. Instrument Code (defined in Instrument Codes Setup Menu) identifies parameter code.

3. Requires Priority assignment to determine which instrument(s) to download and download type

Page 65: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

7 of 38

Changing Parameter and Profile Display Order – Facilitator Demo

The Parameter and Profile Display Order window is used to change the order in which parameters and profiles are displayed on screen.

Step Action

1 From the menu bar, select Environment then select Display Order to open the Parameter and Profile Display window.

2 Select a parameter or profile in the Initial Order box to move it to the Final Order box.

3

Repeat step 2 to continue moving parameter over.

Note: If you moved the wrong parameter over select it again to move it back.

4 Using a patient Sample Report from the DxH 800, move parameters over to correspond with the layout.

5 Repeat steps 2 and 3 for the profiles in any order you choose.

6

Select the save icon from the Toolbar to save your changes.

7 From the menu bar, select Configuration → Profile Definition .

8 From the Profile Configuration window, select the drop down arrow to the right of Profile Code: and select HEM_PROFILE.

9 View the parameters that are included in a HEM_PROFILE.

Page 66: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 8 of 38

Page 67: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

9 of 38

COMMENTS AND TAG DEFINITION

OBJECTIVES: At the completion of this module you should be able to:

Define Comments

Define Tags

Apply comments and tags to requests

Why is it important

Customizing comments and tags will allow you to easily identify samples or results that require operator attention.

Resources

Remisol Advance Help

Page 68: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 10 of 38

Page 69: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

11 of 38

INFORMATION/PRACTICE SECTION

Tag Management

The management of tags was improved in Remisol Advance. In the previous versions, the user could merely tag or untag requests, samples or patients. Now the user can define custom tags which can be used with requests and samples.

Step Action

1 Select Configuration from the Menu Bar and then select Tag Definition.

2 In the Tag name field type Path review.

3 In the Comment Field type Further review of results required.

Note: this is an optional field.

4 In the Colors in the lists box choose Custom colors radio button.

5 Change the Background color to the color of your choice.

6 If needed, change the Text color to black.

7 Select the check box for Bold and deselect any other options.

8 Select the save Icon from the Toolbar.

9 Close the Tag Configuration window.

10 Select Configuration from the Menu Bar and then select Tag Definition.

11 In the Tag name field type ER.

12 Select the save Icon from the Toolbar.

13 Close the Tag Configuration window.

14 Exit and Restart the Remisol Advance software.

Page 70: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 12 of 38

Defining Comments

This option allows the user to predefine technical comments or interpretations. Also, these may be categorized so that they are displayed only where required. The comment abbreviation can be used as a shortcut to enter one or more technical comments or interpretation in the Comments window for a patient. Note: These are known as Result level (interpretation) comments and Run level (technical) comments

Step Action

1 From the menu bar select Configuration then select Comments Definition to open the comments configuration window.

2 In the Abbrev. column, type CLT.

3 In the Name column type in Sample clotted.

4 Select the category in which you want the abbreviated comment to appear. For this exercise choose Technical (Tech.).

5 Select the save icon from the toolbar to save the changes you made.

6 In the Abbrev. column, type QNS, then in the Name column type Quantity not sufficient.

7 Select the Tech checkbox and then select save from the toolbar.

8

To simulate an LIS download, on the LIS & Instrument Simulator

desktop, double-click on the PowerSimu icon , then double-click on R2000 LIS DWNLD and then single-click on samples 1305 and 1306.

9

To simulate an LIS upload, on the LIS & Instrument Simulator

desktop, double-click on the PowerSimu icon , then double-click on R2000 DXH 600/800 and then single-click on samples 1305 and 1306.

10 From the Request List, double-click on sample 1305 and then select the Comments icon from the toolbar.

Page 71: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

13 of 38

Step Action

11 Under Technical Comments, select Choose and double-click on the technical comment of your choice.

12 Save the comment by selecting the save icon and on the next screen select the save icon again.

13

Below the tabular area on the left side of the result screen, check the Tagged checkbox and click on the drop down arrow and choose Path review.

14 Select the save icon from the toolbar and close the request.

Page 72: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 14 of 38

Page 73: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

15 of 38

RULES

OBJECTIVES At the completion of this module you will be able to:

Identify Rule Definitions

Explain common uses for rules

Differentiate between types of rules

Define Rule Validation

Identify conditions and actions for predefined rules

Modify a rule

Create a Rule for critical samples utilizing the Messages List

Utilizing Played Rules

Why is it important?

Rules help you to manage results and increase efficiency in the laboratory. New rules must be validated to ensure quality before the rule can be applied in the laboratory.

Utilizing the rule features will enable you to configure the Remisol Advance to autovalidate results, add messages and comments, utilize delta checking, utilize reflex testing, and perform other automatic additions/ checks.

Resources

Remisol Advance Help

Page 74: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 16 of 38

Page 75: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

17 of 38

INFORMATION/PRACTICE SECTION

Rules Overview

Rules are formulas that are designed to trigger specific actions. Some examples of how rules can be used are listed below:

Automatically validate a result

Convert results using a formula

Calculate a parameter

Trigger a message

Order a test

Add a comment to a request/result

The Remisol Advance generally uses the following format for rule writing: IF this condition exists, THEN this action will happen or ELSE this action will happen. The condition builds the rule for the IF portion and the action builds the rule for the THEN or ELSE portions. Using “AND” or “OR” within a rule can be used for multiple conditions. If “AND” is used, both conditions must be met for the action to be performed. Using “OR” means at least one of the conditions will result in an action.

Rules fit into three categories:

Download Rules - Requests received from the LIS

Parameter Rules - Rules based on single parameters

Upload Rules - Actions based on a single parameter or group of parameter results

Download Rules

Download Rules generate actions based on requests received from the host. Download rules may be used to automatically add new tests to a request, delete requested tests, or notify the user to review something on the request. Some examples of download rules are:

Add a test to specific type of order: Add CBC when a Retic only is

ordered

Change sample priority: If sample received from ER, make priority STAT

Page 76: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 18 of 38

Parameter Rules

Parameter Rules generate actions for an individual parameter (test). Parameter rules may be used to auto validate results, check a result against a delta check formula, add a comment, or modify a result. Some examples of parameter rules are:

Generate a message if a delta check fails: IF HGB delta check fails, add comment: Failed Delta Check

Modify a result: IF manual comment contains ”Anisocytosis Slight”, Then report “SLT”

Upload Rules

Upload Rules generate actions based upon results received from instruments. Upload rules may be used to add a reflex test, add a calculated parameter, display a message, validate a result, stop rules processing or change tube status to a STAT. Some examples of Upload Rules are:

Validate a result if it falls within the validation range: If HGB is within validation range, validate result

Order a reflex test if result greater than critical limit and reflex test not previously ordered: If only CBC was ordered and WBC <=2.0 or >=30.0, then reflex DIFF

Important Information about Rules

A Rule may involve a parameter, reference limit, validation limit, delta check, patient age, sex, doctor, location, comment, etc.

Rules are applied from the first Rule to the last, in the order that the rules are listed

The rules are played in the following order:

1. Download Rules

2. Parameter Rules

3. Upload Rules

Note: Use the server when accessing the test database. All communication between systems (LIS, Remisol Advance/EQC and Instruments) is stopped when using the test database.

Page 77: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

19 of 38

Rules Validation

Each laboratory is responsible for validating all rules for their laboratory setting, patient population and requirements

Rules are evaluated in a test database to avoid alteration of the Remisol production database

Rules should not be activated in the live environment until tested in the testing environment

Once rules are defined they are validated using the following process:

Production Database (Live Environment):

1. Rules are written but not activated

2. Patient requests that will be used to test the rules are manually requested

3. Representative data to test the rules at the decision points are input into the manually programmed requests

4. Export results to a file accessible in the testing mode

Test Database:

1. Switch to the Test Database

2. Activate the rule(s) for testing

3. Import manually entered results into the test database

4. Print rules report and review to verify appropriate actions

5. Modify the rule(s) if necessary and repeat testing

Note: The Test Database configuration can be save and then loaded into the Production Database

Production Database (Live Environment):

1. Return to the Production Database

2. Activate the new rule(s) or load the test database configuration

3. Save the new configuration

4. Delete manually entered requests

Page 78: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 20 of 38

Creating a Rule – Do on the Server

Step Action

1 Select Configuration from the Menu Bar then select Parameter Definition.

2 Select WBC from the Param. Code drop down box.

3 Select the Rules button to open the Rule for WBC box.

4 Select the New button.

5 Enter “Perform manual diff” in the Name field for the Rule name.

6 From the drop down list in the Condition box, select “Into the validation range”.

7 Choose the check box for reverse the condition.

8

Select the Add button.

Note: The rule should look like this:

(NOT(InValRange(WBC))) in the IF box.

9

From the Actions drop down box choose Display Message then type “Perform manual diff” in the text box.

Note: the text box will not appear until the action Display A Message is chosen.

10

Select the Add button.

Note: The message should look like this “MSG(Perform Manual Diff)”in the Then box.

11 Select the green check mark .

Note Important: For v1.8, DO NOT activate the Rule at this time.

12 Select the save button on the toolbar.

Note 1: Before the rule can be activated, it needs to be tested in a test environment. This is applicable to v1.6 and v1.8.

Note 2: Only for v1.7: The rule must be activated before testing.

Page 79: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

21 of 38

Testing Rule

Step Action

1 Select Configuration → Parameter Definition → Param Code: WBC → Reference & Validation Ranges → Default Validation Range to verify the lower and upper validation range for WBC.

Note Use values outside the ranges above to test the rule.

2 Manually program a new request by selecting New Request

from the toolbar.

3

Enter 0001 and select the Enter key on the keyboard. Select Yes to the popup window, “No such person exists. Create a new demography?”.

4 Put the cursor in the Last Name box and type RULES TEST and then select the Enter key on the keyboard.

5 Double click on the parameter WBC. You will see WBC is now in the Summary box on the right side of that screen.

6 Select the green check mark .

7 In the Sample ID Assignment box, enter LOW and then select

the green check mark .

8 Select save from the toolbar.

9

Type 0001, for the next sample, select Enter and then select the

green check mark as this demographic has been created in step 4.

Repeat steps 5 to 8 to enter the rest of the information.

Patient ID

Last Name Sample ID Test

0001 Rules Test LOW VALID WBC

0001 Rules Test HIGH VALID WBC

0001 Rules Test HIGH WBC After completing these 4 entries, close the Request sheet window.

10 From the Request List, select the first Sample ID Low to input results.

11 Right click on the test name WBC.

Page 80: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 22 of 38

Step Action

12 Select Add Manual Run from the pop-up window.

13 Select the result cell to the right of “Downloaded”.

14 Manually enter the numeric result 2.9 for the WBC result and then press Enter on the keyboard.

15 Select the Save icon from the toolbar .

16 Select Yes to the prompt “Do you confirm that entered values are correct”.

17 Select the Next Request button on the Toolbar.

18

Repeat steps 5-11 for each Sample ID you requisitioned.

Use the table below for the rest of the requests:

WBC

Sample ID Numeric Value Low Valid 3.0 High Valid 20.0

High 20.1

19 Close the last request.

Exporting Results

In order to test results in the testing mode, requests (sample results) need to be in a file that is accessible in the testing mode.

Step Action

1 Using the Request created in the previous exercise open the first request, (Sample ID LOW) to be exported

2 Right click the mouse.

3` Select Export Selected Sample to File.

4 On the left side of the Save As window, select Libraries – Documents.

Page 81: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

23 of 38

Step Action

5 Enter the Sample ID Low as the file name.

6 Select Save.

7 Select the Next Request button on the Toolbar.

8 Repeat steps 2-7 for each Sample ID you requisitioned.

9 Close the last request.

Switch to Testing Mode

Testing a rule must be done on the Server

Step Action

1 Select Environment from the Menu Bar.

2 Select Mode: Rules Validation.

3 Select Switch in Test Mode….

4 Enter (TestDB) in the Switch to: field.

Note You can assign a new name or select a test database name of a previously saved database on the drop down list.

5 Leave the Database files destination folder area with the default selection.

6 Optional area: Select Export current dictionary (overwrite test database) option to create a dictionary.

7 Select OK.

8

Select Yes to the pop-up message:

This function can take few minutes. Instrument should be in standby. Do you want to continue?.

9 Select Yes to the pop-up message:

Remisol must be restarted….Do you want to stop Remisol now?

10 Select the Remisol icon on the desktop to enter the Test Mode.

Page 82: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 24 of 38

Step Action

11 On the Top of the screen it will state you are in the Remisol Advance™(Rules Testing).

12

Activate the rule created earlier refer to the exercise Creating a Rule page 20 if needed.

Note: You do not need to exit and restart this time.

Import Results in Test Mode

Before beginning the actual testing of rules created, results need to be imported into the testing environment. For Classroom purposes we will use the results we exported earlier using the steps below:

Step Action

1 Select Environment from the Menu Bar.

2 Select Mode: Rules Validation.

3 Select Import Data.

4 Select Import from files…button from the Rules Testing window.

5

Select the file name(s) from the saved locations.

Note: to select multiple files hold down the [Ctrl] key while selecting the file names.

6 Select Open.

7

In the Select Steps window select the following checkboxes:

Play rules on imported samples

Print requests after rules was played

Export request after rules was played

Export/print rule trace

Note: You can select the Delete all data from the database checkbox to purge the database.

8 Select OK.

9 Select OK to the pop-up message:

“Data exportation done”.

Page 83: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

25 of 38

Step Action

10 Close the Rules Testing window.

11 Review the reports generated and determine if the rules passed or failed.

12 Correct and retest your rule if it did not demonstrate correct behavior. Ask the Facilitator to discuss possible solutions.

13 When finished with all testing, exit the Testing Mode by exiting and restarting the Remisol Advance software.

14 Modify the rule if necessary and then activate the rule.

15

Delete all of the requests manually created to test rules, by selecting the sample and selecting the Trashcan (delete) button from the Toolbar.

Page 84: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 26 of 38

Messages List

The Messages List window gives a concise list of all messages created by extended rules for all current (non-archived) requests. One of the primary uses of the Messages List is to notify the operator of critical results. Other types of messages may also be triggered by rules. Since critical results appear in the Messages List, it is the first list you would use to review results. This will ensure that critical results are acted upon first.

Messages List Filter

The message filter window is used to modify each display filter. The filters allow the user to specify which messages to display in the message list window. You can select one or several of the follow filters:

Display the messages for requests that have not been reviewed (message will then disappear after the requests are reviewed)

One or several of the instruments that generate the results

Step Action

1 From the toolbar, select the Messages List icon and move that box to the lower right quadrant of the Client Zone.

2 Right click anywhere in the Messages list box and select Filters → Modify and change the name from Default to Critical results.

3 Check the checkbox for Show only messages not reviewed and select the DxH 800 only so it is backlit in blue.

4 Under Tab color change the background color to red.

Note The (*) beside the Message generated by rules header on the title bar allows the user to know a filter is currently set. This will appear for any filters (Samples and Request Lists).

4

To simulate an LIS download, on the LIS & Instrument Simulator

desktop, double-click on the PowerSimu icon , then double-click on R2000 LIS DWNLD and then single-click on Hem Patient Batch.

5

To simulate an LIS upload, on the LIS & Instrument Simulator desktop,

double-click on the PowerSimu icon , then double-click on R2000 DXH 600/800 and then single-click on samples 1306.

6 What message is posted in the Messages List?____________________

Page 85: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

27 of 38

Played Rules – Facilitator Demo When entering results, a quick check to determine the rules that were evaluated and if they passed or failed (condition requirements) can be performed by clicking the right mouse button on the result and selecting Played Rules from the contextual menu.

Rules that display green with a status of Passed indicate that the rule was

evaluated and all conditions were present and the action was performed.

Rules that display in red with a status of Failed indicate that the rule was evaluated and the conditions were NOT met, so no action was taken. This does NOT indicate the rule is configured incorrectly, rather the conditions of the rule are not present.

Rules that display in purple indicate the rule does not have enough

information to evaluate. This indicates a rule that may not be configured correctly.

Page 86: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 28 of 38

Page 87: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

29 of 38

MANUAL DIFFPAD

OBJECTIVES At the completion of this module you will be able to:

Configure users in the Diffpad Configure setup for the Diffpad Utilize comments as morphology

Why is it important?

The manual DiffPad can be used when a manual differential is required directly from the Remisol Advance console.

Resourses

Remisol Advance Help

Page 88: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 30 of 38

Page 89: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

31 of 38

Information/Practice Section If you are using a hematology instrument, you can carry out a manual diff directly

from the Remisol Advance console. You can access this screen by selecting the Desk / Manual count menu, or pressing CTRL+K from the request sheet, or by clicking on the microscope toolbar button.

Diffpad Configuration – Facilitator Demo The Hematology Standard Dictionary contains a DiffPad with cell types

already defined. This can be used as is or edited for the cell types you want to use.

Before you can add a user to the DiffPad configuration, the user must be

defined in Remisol.

The Diffpad is setup through the following path: Start → All Programs → Remisol Advance → Intranet Advance → Remisol Advance Login: and

Password. Performing a manual diff using Remisol Diffpad

Step Action

1

To simulate an LIS download, on the LIS & Instrument Simulator desktop,

double-click on the PowerSimu icon , then double-click on R2000 LIS DWNLD and then single-click on Hem Patient Batch.

2

To simulate an LIS upload, on the LIS & Instrument Simulator desktop,

double-click on the PowerSimu icon , then double-click on R2000 DXH 600/800 and then single-click on samples 1300 and 1303.

Page 90: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 32 of 38

Step Action

3

Open the request for Patient ID 1300 from the Request List and select the Manual Count (Diffpad) Button from the toolbar.

IMPORTANT NOTE: Single click the manual diff icon one time from the toolbar.

4

Perform a manual diff entering the cells identified in the table below:

Cell Number Segs 65

Bands 1 Lymph 14 Mono 10 Eos 5

Atypical Lymphs 5

Or, you may enter the 100 cell diff of your choice.

5

To enter morphology, you may select either the drop-down arrow for Comments or select the Choose button. If the Choose button is selected, double-click on the morphology of your choice.

6 When the diff is complete, select the save icon.

Page 91: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

33 of 38

PRACTICE

OBJECTIVES At the completion of this module, you will be able to:

Create a Parameter Rule

Create a rule for critical samples

Create a delta check

Why is it important?

System Administrators for the Remisol may need to define rules. In addition, determining the laboratory’s requirements for rule validation is a vital part of the not only the installation but for the future.

Resourses

Remisol Advance Help

Page 92: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 34 of 38

Page 93: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

35 of 38

PRACTICE EXERCISES

EXERCISE 1: CREATE A PARAMETER RULE Step Action

1 Select Configuration → Parameter Definition, click on the down arrow in Param. Code and select WBC.

2 Select Rules → New and in the Name field type WBC Critical.

3 Select the down arrow next to Condition, scroll down and select Test result. WBC will be populated in the next box.

4 In the next box select < and then in the next box type 2.0. Click Add.

5 Select the Or radio button and then repeat step 3 and select > and then type in 50.0. Click Add.

6 In the IF condition text box, manually type .Value after the word WBC for both conditions. The condition will now read: (WBC.Value<2.0) OR (WBC.Value>50.0)

7 Select the drop arrow under Actions and select Add a comment 2 and type Critical into the text box. Select Add.

8 Add another Action, select Add a comment 2, type CNV and select Add.

9 Add another Action, select Display a message, type Critical result and select Add.

10 Enable the rule by checking the This rule is active checkbox.

11 Select the green check mark and then select the save icon to save the rule.

12 The Rule will read: IF (WBC <= 2.0) OR (WBC >= 50.0) THEN COMMENT2(Critical); COMMENT2(CNV); MSG(Critical result)

Page 94: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 36 of 38

EXERCISE 2: CREATE A PARAMETER RULE Step Action

1 Select Configuration → Parameter Definition, click on the down arrow in Param. Code and select PLT.

2 Select Rules → New and in the Name field type PLT Non-numeric.

3 Select the down arrow next to Condition, scroll down and select Test result. PLT will be populated in the next box.

4 In the next box select = and then in the next box type NULL. Click Add.

5 In the IF condition text box, manually type .Value after the word PLT for the condition. It will read (PLT.Value = NULL)

6 Select the drop arrow under Actions and select Add a comment 2 and type CNV into the text box. Select Add.

7 To add another Action select Add a comment 2, type Integrity check and select Add.

9 Enable the rule by checking the This rule is active checkbox.

10 Select the green check mark and then select the save icon to save the rule.

11 The Rule will read: IF (PLT.Value = NULL) THEN COMMENT2(CNV); COMMENT2(Integrity check).

EXERCISE 3: CREATE A COMMENT RULE

Step Action

1 Select Configuration → Parameter Definition, click on the down arrow in Param. Code and select NE%.

2 Select Rules → New and in the Name field type Manual Diff. 3 Select the down arrow next to Condition, select Search in comment. 4 In the next box select Suspect Flags. In the next box select Contains.

5 In the next box select Left Shift. Then select Add. Note: You may select Or and select other Suspect Flags, for example Imm. NE 1. Remember to erase the number 1 in the IF box.

Page 95: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Rev. C (June 2015) Beckman Coulter Professional Development

37 of 38

Step Action

6 Select the drop arrow under Actions and select Add a comment 2 and type DNV into the text box. Select Add.

7 To add another Comment2, type Perform manual diff and select Add. 8 Enable the rule by checking the This rule is active checkbox.

9 Select the green check mark and then select the save icon to save the rule.

10

The Rule will read: IF (HEM_SUSPECT Contains ‘Left Shift’) OR (HEM_SUSPECT Contains ‘Imm.NE’) THEN COMMENT2(DNV); COMMENT2(Perform manual diff)

EXERCISE 4: CREATE A DELTA CHECK FORMULA

Step Action

1 Refer to User Guide → Configuration → Parameter Definition, scroll down to Delta Check and read this section.

2 Select Configuration → Parameter Definition, click on the down arrow in Param. Code and select HGB.

3 In the Delta Check box, and using the formula from Example 1 in the User Guide as a reference, write a formula whereby R1-R2 is not greater than 2.0 with a Validity of 2 days.

4 Select Test and in the Formula Testing box, enter 13.0 for R1 and 11.0 for R2.

5 Select Test and the Result should be 0.00000 which indicates a Fail. Note: A Pass is indicated by 1.00000.

Page 96: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 2 Beckman Coulter Professional Development Rev. C (June 2015) 38 of 38

NOTES FOR RULE WRITING

1. When a rule contains a result#, add .Value after the parameter HGB.Value<12.0

2. Do not use punctuation in comments; 9( ( . , “ () ) 3. Use only specific short comments and do not get too wordy 4. Comment goes to the LIS and Comment2 stays in Remisol for Actions 5. For Comment2 use Review Slide, Scan Slide or Manual Diff. A rule will be

written later that plays off these comments 6. When using OR and AND in a Rule, must use double parantheses, for example

((HGB.Value,12.0) OR (HGB.Value>20)) AND (Age=365) Note: If using the Rules Wizard, it will provide the correct syntax, i.e. the parentheses ( ).

7. Download Rules: Received from the LIS. These rules are used once they are received from the LIS. Download Rules can be used for adding a test, deleting a test or warn the operator that something needs to be reviewed

Parameter Rules: Stops results from going to the LIS. They can be used to add/suppress instrument parameter automatically or warn the operator of something abnormal by displaying messages Upload Rules: Go to the LIS. These rules are used once results are received

from the instrument. Upload Rules can be used for reflexes, reruns, add other instrument tests or add a calculated parameter.

Page 97: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

1 of 34

COURSE GUIDE DAY 3

BASIC SETUP AND PROGRAMMING

Day 2

Rules &

System Configuration

Day 3

Basic Setup &

Programming

Day 1

Basic Operations

Page 98: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 2 of 34

PERSONAL PROGRESS SUMMARY

Basic Setup & Programming

Modules Date Initials Facilitator Page#

Users Setup 4

Report Setup 5

Database Management 7

Saving Configuration 21

Programming Functions 25

Page 99: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

3 of 34

SYSTEM CONFIGURATION

OBJECTIVES At the completion of this module, you will be able to:

Create new users

Create a Report Header

Why is it important

The system configuration allows laboratories to define user access and a report header if it is your laboratory protocol to print reports.

Resources

Remisol Advance Help

Page 100: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 4 of 34

INFORMATION/PRACTICE SECTION

Users Setup

Different levels of access/privileges may be defined for different individual users. Only users with system administrative privileges can define new users.

Step Action

1 Select Environment from the Menu Bar then select Users Setup.

2 From the Users Setup submenu, select Access Rights.

3 From the Users pop-up window select New.

4 In the Your password field type in session. This is the password for the administrator currently logged in.

6

Create the following users and privileges. Note: Passwords can be up to 25 characters.

User Name Password Allow Access to:

User 1 Dorothy Gale Oz Input manual results Manual validation

User 2 Clark Kent Superman Administrator

7 Move the desired access privileges from the Access Denied box to Access Allowed box by single clicking the privilege.

8 Select the green check mark to save the settings.

9 Select End to exit the Users window.

10 Select Ctrl+B or Environment → Lock Station to log off.

11 Log back on using sysadm/session.

Note For the Manual Diffpad, users need to setup in Remisol first, then in the Diffpad.

Page 101: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

5 of 34

Changing Report Header

The report header allows for laboratories to define header information for patient and quality reports to be printed. Please note that Remisol Advance also prints the laboratory name as defined in the PMF license file at the bottom of the page.

Step Action

1 From the menu bar select Environment → Report Setup → Report Header to open the Report Header Setup window.

2

Create a Report Header using example below:

Line One: Hospital Name

Line Two: Hospital Address

Line Three: City, State

Line Four: Contact Name

Line Five: Contact Number

3

Select the green check mark to save your changes and close the window.

Note: The Cancel button will close the window without saving changes.

4 Close the Remisol Advance software by selecting the on the title bar.

Note When changes are made to the Environment, the Remisol Advance software needs to be restarted.

Page 102: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 6 of 34

Page 103: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

7 of 34

DATABASE MANAGEMENT

OBJECTIVE At the completion of this module you will be able to:

Manually archive results

Set up the Remisol Advance to automatically archive

Back Up archive to external media

Restore archive from external media

Configure Automatic Export

Why is it important?

The Remisol Advance will operate more efficiently and quickly when the database is properly maintained. When the database contains too many current requests the Remisol may exhibit performance issues. Maintaining archived records will ensure the patient records are available for delta checking and easy retrieval. Backing up archives allows for long term storage and result traceability to an instrument. Resources

Remisol Advance Help

Page 104: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 8 of 34

Page 105: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

9 of 34

INFORMATION/PRACTICE SECTION

Database

The Remisol Advance database consists of both Current Requests (non-archived) and Archived requests. The amount of data stored in the database should be maintained by deleting and/or archiving requests per the laboratories protocol. LIS functionality may contribute to the decision on how to manage data.

Current Requests

Current requests should be archived or deleted at least once a day

Current requests can be manually archived or deleted

Deleting current requests removes them from the database and CANNOT be restored.

The system can also be set up to either automatically archive or delete requests based on:

After a specified number of days (deleting) or time (for archiving)

Upload to host

Archived Results

Archiving current request allows for:

Delta Checking – current results are compared to the last archived result

Easy retrieval to review previous results

Multiple runs of the same Sample ID can be archived (reruns)

Previous results traceability, i.e. who validated, QC Status

Reruns can be requested from an archived request

Page 106: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 10 of 34

Archive Information

Archived requests have additional information stored with each result. This information can be used to identify result information as well as validation, instrument, and reagent traceability. The information is accessed from the result cell of the request. The following table describes each of the items in the Archive Information window.

Field Flag Description

Sample ID _ Sample ID of result

Sample Modification (sample flag)

M Sample ID was manually changed

Stat (patient flag) S Stat request Validation

(result flag) V v

Result validated automatically Result validated manually

Result Flag (result flag)

M C U D

Result manually entered or edited Calculated result UPC result Dilution factor applied to result

Delta check (result flag)

D X

Result passed delta-check Result failed delta-check

Send to host (result flag)

H Result sent to host

Number of result _ Number of results Source _ Instrument the result was run on

Dilution factor _ Remisol Advance dilution factor Reagent(s) _ Serial number of the reagent used Reagent lot

number _

Lot number of the reagent used

Validated by _ Person who validated this result manually Received on _ Date and time of reception

EQC Status (Optional Module)

_

EQC status flag: White: No protocol defined no data available

Green: No Mistakes

Yellow: Blocked results or Time-Out

Red: Protocol in error

Page 107: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

11 of 34

Archived Patient Results

Archived Request Management

Archived requests should be either deleted or stored offline to maintain

database efficiency.

Archived requests can be manually deleted or set up to automatically delete after a specified time. Deleting archived requests removes them from the database. Deleted requests/results cannot be restored.

Backup archives to external media (flash drive or CD) for offline storage. Data can be restored to the archive database from the external media for traceability purposes.

Page 108: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 12 of 34

Archive Maintenance

Remisol Advance V1.5 Database Capacity NOTE: Assumes data base is on the T310 Server

Number of Archive Days to Keep: Hematology (No Graphics Stored)

#Tubes/day* Tests/day Maximum Number of Archive days 100 2600 365 500 13000 365 1000 26000 250 1500 39000 200 2000 52000 120

*Assumes 26 tests/tube, 2 technical comments per request and 13 results Comments or run comments.

Number of Archive Days to Keep: Hematology WITH Graphics

#Tubes/day* Tests/day Number of Archive days to keep 100 1500 365 500 1300 365 1000 26000 250 1500 39000 200 2000 52000 120

*Assumes 26 tests/tube with graphics, 2 technical comments per request and 13 result comments or run comments, 3 histograms and 2 scatterplots

Page 109: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

13 of 34

General Setup

For automatic archiving and deletion, from the Menu Bar select Environment then General Setup.

Automatic Export

From Remisol Advance 1.8, a new functionality was added to automatically back up the latest archives, those that have not been backed up already (via the General setup option). This can be set up via the Archives → Automatic export menu, this will bring up the Automatic Archive Export Setup window.

Page 110: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 14 of 34

Manually Archive Requests

Step Action

1 Select Archives from the Menu Bar then Archive Current Request(s).

2 Select Yes on the message window to open the Requests Archiving window.

3 Select All radio button in the Archive Requests area.

4 In the From box, type in the date from 7 days ago.

5 In the To box, type in yesterday’s date.

6

Choose the following Checkboxes: Archive Graphs Archive only validated requests

Leave all other checkboxes deselected.

7 In the Instrument box, select DxH 800.

8 Select the green check mark to save changes.

9

At the message: “This action will destroy all patient demographics which have no results. Do you want to continue? Note: Keep check box selected for Keep Monitor-It patients and data and then select Yes.

10 At the message “Database Purged” Select OK.

11 Close the Request Archiving window.

Page 111: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

15 of 34

Creating a Backup of the Archives

Step Action

1 Ask your facilitator for external media for performing the backup.

2 Select Archives from the Menu Bar then select Save to open the Save Archive window.

3 In the Patient ID field put in a * (wild card) and leave all other fields corresponding to patient information blank.

4 In the From field type in the date from 7 days ago.

5 In the To field type in yesterday’s date.

6 Select Extract.

7 Select Select All.

8 Select Save.

9 On the left hand side of the Save As window under “Computer” select the appropriate drive.

10 For the File Name: type Practice Backup and today’s date

11 Select Save.

13

Select No at the prompt asking if you want to delete archives after performing backup. Selecting No is for class room purposes only. Laboratories may choose to delete the archives once they have been backed up.

Note Selecting Yes to the prompt gives the message “This action will destroy all patient demographics which have no results. Do you want to continue?”.

14 Close the Save Archives window.

15 Remove the media and save it for another exercise later.

Page 112: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 16 of 34

Configure Auto Archive

Step Action

1 Select Environment from the Menu Bar then Host Setup to open the Host Setup window.

2 In the Host Upload area choose the check box for Archive After Upload.

3 In the field for Defer the archiving or the deletion of, enter 0.

4 Select the save icon from the Toolbar.

5 Close the Host Setup window.

6 Exit and Restart the Remisol Advance software.

Restoring Archived Data

Step Action

1 Place the external media you used in the Backup Archived exercise into the appropriate drive.

2 Select Archives from the Menu Bar.

3 Select Restore.

4 Select Extract.

Note You can restore a specific patient or date range by filling in the appropriate fields.

5 Select Extract from files.

6 Select the green check mark .

Note If prompted select Cancel if not using the A drive (floppy disk).

7 Locate the file name to restore. Note: Should be Practice Backup and today’s date.

8 Select Open to open the Restore Archives window.

9 Choose Select All.

Page 113: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

17 of 34

Step Action

10 Select Restore.

11 Select Close.

12 Remove the external media.

Result Search from Archive

Step Action

1 Select the Previous Results button from the Toolbar.

2 Type in the Patient ID or Last Name in the appropriate fields use Patient Name: Margie Smith Patient ID: 1305.

3 Press Enter on the keyboard.

4 Select the Patient from the list in the Demographics Search window.

5 Select the green check mark .

6 In the Previous results window double click on a result to open the Archive Information window to view additional information stored with the result.

Automatic Export

Step Action

1 From the Menu bar, select Archives → Automatic export

2 Select the drop down arrow to the right of Computer dedicated to the export and select the name of the Server R2000.

3 Select the frequency of your choice.

4 Select the green check mark .

Congratulations, you’ve finished! Continue to the Skill Check

Page 114: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 18 of 34

Page 115: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

19 of 34

DATABASE MANAGEMENT SKILL CHECK

1. Where do you go to set up automatic archiving or deleting of requests after a specified time frame?

________________________________________________

2. Archives cannot be saved to external media therefore it is important to maintain the database?

TRUE FALSE

3. Results can be manually archived.

TRUE FALSE

4. How do you setup Automatic Export?

_______________________________________________

Page 116: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 20 of 34

Page 117: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

21 of 34

SAVING CONFIGURATION

OBJECTIVE At the completion of this module you will be able to:

Save the configuration

Explain the importance of saving the configuration

Why is it important

This option allows the user to save the current dictionary configuration. The configuration and environment settings saved by this procedure are parameters, profiles, departments, control definitions, instrument setup, host setup, etc.

Resources

Remisol Advance Help

Page 118: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 22 of 34

Page 119: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

23 of 34

Saving the Configuration

Step Action

1 Ask your instructor for an external media device.

2 Select Configuration from the menu bar then select Save.

3 Insert external media.

4

Choose the appropriate drive list under “Computer” on the left side of the “Save As” box.

5 Leave the file name as the default (yyyymmdd) which should be today’s date with a .dic extension.

6 Select Save, when save complete remove external media device.

Page 120: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 24 of 34

Page 121: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

25 of 34

PROGRAMMING FUNCTIONS

OBJECTIVES At the completion of this module you will be able to:

Manually program a request

Locate previously run results

Add a test to a previous request

Why is it important?

Programming features such as manual program, adding a test, and requesting a rerun will allow you to utilize the Remisol Advance when the laboratory LIS is down.

Resources

Remisol Advance Help

Page 122: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 26 of 34

Page 123: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

27 of 34

INFORMATION/PRACTICE SECTION

Program a New Request with no Previous Demographics

Step Action

1 Select the New Request button from the Toolbar.

2 Type a Patient ID of your choice.

3 Press Enter on the keyboard.

4 Select Yes to the popup window message “No such person exists. Create a new demography”.

5 In the Last Name field type in Thomas.

6 In the First Name field type in Maureen.

7 In the Date of Birth Field put in 09/23/1954

8 In the Sex: drop down box choose Female.

9 Choose the STAT check box. Note: The STAT check box is located by the date and time.

10 Select Add parameter or select the Enter key.

11 Under Default sort choose CBC or Individually select each parameter.

12 Select the green check mark to save changes.

13 Using a downtime barcode label in the Sample ID Assignment window change the barcode in the Sample ID field. Note: You can also designate the sample is a STAT here.

14 Select the green check mark and select from the Toolbar.

15 Close the New Request Window.

Page 124: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 28 of 34

Program a New Request with Previous Demographics

Step Action

1 Select the New Request button from the Toolbar

2 Type in the Patient ID 1311 (Yoda Rie) and press Enter on the keyboard.

3 Choose Yoda Rie from the Popup box and press enter or select the green

check mark .

4 Select Retic from the individual listed parameters.

5 Select the green check mark .

6 In the Sample ID field change the ID to 4321012.

7 Select the green check mark .

8 Select the save button from the Toolbar.

Add Parameter to a Request

Step Action

1 Select the Search button from the Toolbar.

2 Type in the sample ID for Maureen Thomas that was used in Program a New Request with no Previous Demographics.

3 Select the green check mark to open the Request sheet.

Note If more than one run is available a window will appear with the various runs. Double click on the desired run.

4 Right click on any result to open the Request Sheet popup menu.

5 Select Create a request from this run. Note: this is available if the results are archived.

Note If you chose the Patient from the Request List when you right click choose Add parameters.

6 From the Parameters List under Profile choose HEM_DIFF.

Page 125: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

29 of 34

SYSTEM MANAGEMENT

OBJECTIVES The System Management will be a chance to:

Review material

Identify Remisol Advance Operator duties

Identify requirements for a smooth installation

Review training co-worker requirements

Why is it important?

A smooth installation of the Remisol Advance requires some forethought on how to utilize the system. Most installations are customized to individual laboratory needs. Installation worksheets should be filled out and used by the specialist configuring the system to meet the laboratories needs.

The questions on the following pages may help you think about some features of the Remisol Advance that you will use in your lab.

Security

How will you protect validation of results by unauthorized users?

How many System Administrators will you have in your facility?

Who will be responsible for assigning passwords?

With Remisol Advance you may define a logoff time-out to lock the software after a predefined amount of time inactivity.

Page 126: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 30 of 34

Page 127: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

31 of 34

INFORMATION / PRACTICE SECTION

Samples List

What Samples List Filters will you use to help you manage samples? How will you customize the filters to be as effective as possible?

Pending Samples

STAT Samples

Pending Rerun

Tagged

Other? _____________________________

Request List

What Request List Filters will you use to help you manage reviewing results? How will you customize the filters using colors and completeness icons to be as effective as possible?

Samples that need Review

Completed STATs

Critical Results

Tagged Samples

Other? ____________________________

Rules

How will your lab utilize rules to help manage results and increase efficiency? Will you use rules to tag or un-tag requests?

Download Rules: actions based on request received from the Host

Add a test when received from a specific doctor

Add a comment

Parameter Rules: actions generated for a specific parameter

Generate a message if delta check fails

Add a comment

Upload Rules: actions based on results received from the instruments

Validate a results

Generate a message

Reflex test ordered

Page 128: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 32 of 34

Messages List

What messages would you like to be displayed on your message list? Customized Rules determine what message displays

Sample results that require the physician to be notified

Critical Results

Samples that require dilutions

Delta Check Failed

Scan Slide

Manual Diff Required

Pathologist Review Required

Other? ____________________________

Database Management

Archive or Delete Daily

Archive

Automatically

After Upload

Delay ___ hours?

Manually

Back Up Archives

Media Type

How Often?

Delete Archives

Automatically after _____ days

Automatically deletes patient demographics with no results (unused patient, if there is no current request related to that patient and when the archived data is deleted because of its age)

Feature is activated by un-checking the option “Keep unused demographics” under ‘Number of archived days to keep’ from the General Setup menu.

Manually “Delete patients with no records” from the Environment menu.

Page 129: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Rev. C (June 2015) Beckman Coulter Professional Development

33 of 34

Quality Control

Integrated Quality Control module

Define/Manage QC

Extended Quality Control

Define/Manage QC

Shut off Auto validation when QC is out of range

Patient Means as an additional tool

Sample ID Management

Never re-use Sample IDs. There is a potential for incorrect sample identification if this recommendation is not followed.

Each patient sample should have a unique sample ID

STOP Button

Blocking

Block downloads

Block validations

Training Co-Workers

Training Plan

Who will be responsible for training staff?

How many people will be trained during each training session?

Does your lab have a current training plan in place?

Training Environment

Time, Interruptions and Attitude

What other items will impact training co-workers?

Training Documentation

Does your lab have training documentation?

Page 130: Remisol Advance for Hematology Customers · Remisol Advance for Hematology Customers Remisol Advance for Hematology Customers Rev. C (June 2015) Beckman Coulter Professional Development

Remisol Advance for Hematology Customers

Remisol Advance for Hematology Customers, Day 3 Beckman Coulter Professional Development Rev. C (June 2015) 34 of 34

Resources for Training

What references will you use to assist in training?

Job Aids

Remisol Advance User Guide

Personnel

Who will be trained to operate the Remisol Advance?

Will all techs be trained on all procedures?

Training and Education Webpage

www.beckmancoulter.com → Support →Training and Education → Diagnostics → Select the Discipline → Select the Instrument/System → Select the appropriate tab

Tools and Aids

Competency Checklists

Job-Aids

Video Learning Clips

Training Guides

Instructor Led

Description of Instructor Led Training classes

www.beckmancoulter.com → Support →Training and Education → Diagnostics → Webinars & Conferences

Webinars and Conferences

Live Webinars

Local conferences

On-Demand

Pre-training

E-Learning

Recorded webinars