Regional blood flow measurement with pulsed Doppler flowmeter in conscious rat JOSEPH R. HAYWOOD, RICHARD A. SHAFFER, CRAIG FASTENOW, GREGORY D. FINK, AND MICHAEL J. BRODY Department of Pharmacology, Bioengineering Resource Facility and The Cardiovascular Center, University of Iowa, Iowa City, Iowa 52242 HAYWOOD, JOSEPH R., RICHARD A. SHAFFER, CRAIG FAS- TENOW, GREGORY D. FINK, AND MICHAEL J. BRODY. RegionaL blood flow measurement with pulsed Doppler flotimeter in conscious rat. Am. J. Physiol. 241 (Heart Circ. Physiol. 10): H273-H278, 1981.-Development of techniques for the contin- uous measurement of regional blood flow and vascular resist- ance in intact small animals has been impeded primarily by the bulkiness of flow probes. The availability of an ultrasonic pulsed Doppler flowmeter system enabled us to construct miniaturized probes using l-mm-diameter piezoelectric crystals that emit a 20-mHz signal and receive the reflected sound waves from passing blood cells. The finished flow probe is approximately 2.5-4 mm long and 2 mm in cross-sectional diameter with lumen diameters appropriate for the rat, ranging from 0.7 to 1.2 mm. This report describes the materials and methods involved in constructing and implanting the probes in rats to monitor renal, mesenteric, and hindquarter blood flow velocity. The accuracy of the pulsed Doppler method in detecting changes in regional blood flow and vascular resistance was established by the demonstration of a highly significant correlation between veloc- ity recorded from the Doppler unit and volume flow recorded simultaneously. These data indicate that the ultrasonic pulsed Doppler flowmeter provides the opportunity to measure changes in regional blood flow and vascular resistance in a conscious freely moving rat. regional blood flow; miniature flow probes; vascular resistance; ultrasonic flow detection; chronic instrumentation MEASUREMENT OF REGIONAL blood flow in several vas- cular beds in conscious freely moving small animals has posed a difficult technological problem. Flow probes for square-wave electromagnetic flowmeters are bulky and are constructed with wire leads of large diameter. In addition, electromagnetic flowmetry ordinarily requires vascular occlusion for determination of zero flow. The dual-crystal configuration required for ultrasonic contin- uous-wave Doppler flowmeters also presents a size prob- lem, as well as difficulty in the proper orientation of the emitting and receiving crystals. These disadvantages make multiple implants of flow probes in small animals difficult and placement of the probes on small vessels, such as the rat renal artery, awkward. These problems may be circumvented by the development of the ultra- sonic pulsed Doppler flowmeter. Franklin et al. (4) first suggested that the pulsed Dop- pler flowmeter would be useful to estimate blood flow by using two piezoelectric crystals to determine upstream to downstream transit times. More recently, a multichannel directional pulsed Doppler system has been developed by Hartley and Cole (5), in which several probes are driven by a common oscillator. Like the continuous-wave Doppler, the pulsed Doppler flowmeter offers the advan- tage that zero flow can be determined electronically. However, in contrast to the continuous-wave flowmeter, the detection system of the pulsed Doppler consists of a single lightweight piezoelectric crystal that emits a 20- mHz ultrasonic signal. The same crystal receives the reflected signal from the passing blood cells in the inter- vals between ultrasonic pulses. Although the system mea- sures Doppler shift, a relative index of blood flow, the reduced size of the crystals permits the chronic implan- tation of multiple flow probes in small animals, such as the rat. This report describes detailed methods for the prepa- ration of the crystal and construction of the flow probe for implantation. Although Hartley et al. (6) have pre- sented preliminary observations on the use of this flow- meter system in the conscious rat, we report refined procedures for the acute or chronic placement of these probes around small- and large-diameter arteries in rats for application of the pulsed Doppler flowmeter in chron- ically instrumented rats. Finally, evidence is presented that changes in the velocity signals recorded from the flow probes are directly and reliably proportional to changes in true volume flow. METHODS The ultrasonic pulsed Doppler flowmeter used in this study was constructed by the Bioengineering Resource Facility of the University of Iowa from the plans gener- ously supplied by C. J. Hartley of Baylor College of Medicine. The system is basically the same as that de- scribed by Hartley and Cole (5); however, several modi- fications have been made to increase the sampling fre- quency of the echo signal to improve the accurate detec- tion of high-velocity signals.’ Modifications have also been made in the preparation of the crystal and the construction of the flow probe. Flow Probe Construction The pulsed Doppler flow probe consists of a Silastic cuff around a piezoelectric crystal with insulated copper wire leads. The four stages of construction are detailed ’ For details of modifications, contact C. Fastenow. 0363-6135/81/0000-0000$01.25 Copyright 0 1981 the American Physiological Society H273 by 10.220.33.5 on August 28, 2017 http://ajpheart.physiology.org/ Downloaded from

Upload: others

Post on 23-May-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Regional blood flow measurement with pulsed Doppler flowmeter … · 2017-08-28 · TENOW, GREGORY D. FINK, AND MICHAEL J. BRODY. RegionaL blood flow measurement with pulsed Doppler

Regional blood flow measurement with pulsed Doppler flowmeter in conscious rat

JOSEPH R. HAYWOOD, RICHARD A. SHAFFER, CRAIG FASTENOW, GREGORY D. FINK, AND MICHAEL J. BRODY Department of Pharmacology, Bioengineering Resource Facility and The Cardiovascular Center, University of Iowa, Iowa City, Iowa 52242

HAYWOOD, JOSEPH R., RICHARD A. SHAFFER, CRAIG FAS- TENOW, GREGORY D. FINK, AND MICHAEL J. BRODY. RegionaL blood flow measurement with pulsed Doppler flotimeter in conscious rat. Am. J. Physiol. 241 (Heart Circ. Physiol. 10): H273-H278, 1981.-Development of techniques for the contin- uous measurement of regional blood flow and vascular resist- ance in intact small animals has been impeded primarily by the bulkiness of flow probes. The availability of an ultrasonic pulsed Doppler flowmeter system enabled us to construct miniaturized probes using l-mm-diameter piezoelectric crystals that emit a 20-mHz signal and receive the reflected sound waves from passing blood cells. The finished flow probe is approximately 2.5-4 mm long and 2 mm in cross-sectional diameter with lumen diameters appropriate for the rat, ranging from 0.7 to 1.2 mm. This report describes the materials and methods involved in constructing and implanting the probes in rats to monitor renal, mesenteric, and hindquarter blood flow velocity. The accuracy of the pulsed Doppler method in detecting changes in regional blood flow and vascular resistance was established by the demonstration of a highly significant correlation between veloc- ity recorded from the Doppler unit and volume flow recorded simultaneously. These data indicate that the ultrasonic pulsed Doppler flowmeter provides the opportunity to measure changes in regional blood flow and vascular resistance in a conscious freely moving rat.

regional blood flow; miniature flow probes; vascular resistance; ultrasonic flow detection; chronic instrumentation

MEASUREMENT OF REGIONAL blood flow in several vas- cular beds in conscious freely moving small animals has posed a difficult technological problem. Flow probes for square-wave electromagnetic flowmeters are bulky and are constructed with wire leads of large diameter. In addition, electromagnetic flowmetry ordinarily requires vascular occlusion for determination of zero flow. The dual-crystal configuration required for ultrasonic contin- uous-wave Doppler flowmeters also presents a size prob- lem, as well as difficulty in the proper orientation of the emitting and receiving crystals. These disadvantages make multiple implants of flow probes in small animals difficult and placement of the probes on small vessels, such as the rat renal artery, awkward. These problems may be circumvented by the development of the ultra- sonic pulsed Doppler flowmeter.

Franklin et al. (4) first suggested that the pulsed Dop- pler flowmeter would be useful to estimate blood flow by using two piezoelectric crystals to determine upstream to downstream transit times. More recently, a multichannel

directional pulsed Doppler system has been developed by Hartley and Cole (5), in which several probes are driven by a common oscillator. Like the continuous-wave Doppler, the pulsed Doppler flowmeter offers the advan- tage that zero flow can be determined electronically. However, in contrast to the continuous-wave flowmeter, the detection system of the pulsed Doppler consists of a single lightweight piezoelectric crystal that emits a 20- mHz ultrasonic signal. The same crystal receives the reflected signal from the passing blood cells in the inter- vals between ultrasonic pulses. Although the system mea- sures Doppler shift, a relative index of blood flow, the reduced size of the crystals permits the chronic implan- tation of multiple flow probes in small animals, such as the rat.

This report describes detailed methods for the prepa- ration of the crystal and construction of the flow probe for implantation. Although Hartley et al. (6) have pre- sented preliminary observations on the use of this flow- meter system in the conscious rat, we report refined procedures for the acute or chronic placement of these probes around small- and large-diameter arteries in rats for application of the pulsed Doppler flowmeter in chron- ically instrumented rats. Finally, evidence is presented that changes in the velocity signals recorded from the flow probes are directly and reliably proportional to changes in true volume flow.

METHODS

The ultrasonic pulsed Doppler flowmeter used in this study was constructed by the Bioengineering Resource Facility of the University of Iowa from the plans gener- ously supplied by C. J. Hartley of Baylor College of Medicine. The system is basically the same as that de- scribed by Hartley and Cole (5); however, several modi- fications have been made to increase the sampling fre- quency of the echo signal to improve the accurate detec- tion of high-velocity signals.’ Modifications have also been made in the preparation of the crystal and the construction of the flow probe.

Flow Probe Construction

The pulsed Doppler flow probe consists of a Silastic cuff around a piezoelectric crystal with insulated copper wire leads. The four stages of construction are detailed

’ For details of modifications, contact C. Fastenow.

0363-6135/81/0000-0000$01.25 Copyright 0 1981 the American Physiological Society H273

by 10.220.33.5 on August 28, 2017

http://ajpheart.physiology.org/D

ownloaded from

Page 2: Regional blood flow measurement with pulsed Doppler flowmeter … · 2017-08-28 · TENOW, GREGORY D. FINK, AND MICHAEL J. BRODY. RegionaL blood flow measurement with pulsed Doppler

H274 HAYWOOD ET AL.

in Fig. 1. A dissecting microscope should be used for each stage of construction.

Stage A. The 360AWG insulated silver-plated copper wire leads (CZ1174/SPC, Cooner Sales, Chatsworth, CA.) are prepared for soldering by first removing 1 mm of insulation from each end. Care must be exercised to prevent the multistranded wire from being nicked or cut. Noncorrosive 65/35 resin core solder is used to tin both ends of the wire. A miniature soldering iron (6 V, 6 W, pencil type) is used for all soldering steps. A very small bead of solder is placed on one end of the wire after tinning. The PZT-5A piezoelectric crystal (20 mHz, 1 mm diam x 0.004 mil, compress mode, fine-ground, chrome- gold plate; Valpey-Fisher, Hopkinton, MA) is placed on a firm flat surface. A thin film of TIX solder flux is applied to a small area near the edge with a 27.gauge needle. The flux will allow the solder to flow smoothly onto the chrome-gold plating. The beaded end of the wire is placed on the fluxed area, with the edge of the insulation flush with the edge of the crystal (do not apply pressure to the surface of the crystal). The solder bead is gently touched with the soldering iron until the solder flows onto the crystal surface, and the solder is kept to a small area of the crystal. If the solder is allowed to cover too much surface, the vibrating characteristics of the crystal will be diminished. The wire is then rotated carefully until the solder joint is face down. The other lead wire is soldered to the remaining surface (Fig. lA). The wire leads are taped down and very gently bent up to elevate the crystal into a horizontal position above the work surface.

Stage B. The piece is encapsulated with epoxy to strengthen and protect the crystal and solder joints (Fig. 1B). After being mixed, the epoxy (Plast no. 86/87, Fibre- Glast Developments, Dayton, OH) is degassed under vacuum for 15-20 min to remove all trapped air. If not removed, this trapped air will greatly attenuate the signal emitted and received by the crystal. The pot life of the

epoxy is about 1 h. The epoxy is applied with a 27-gauge needle to both surfaces and edges of the crystal, which should be covered by a very thin film of epoxy when finished. The epoxy is allowed to cure for 24 h, and the work is not moved during this time.

Stage C. After 24 h, the crystal is placed in a Silastic mold designed to hold the crystal at an approximately 45’ angle to the vessel, the optimum angle for detection of flow velocity.. The mold is a S-mm-length of Silastic tubing (1.0 x 1.0 mm; Dow Corning, Midland, MI). One end is cut perpendicular to the lumen, -the other end is cut at a 45’ angle to the lumen. A squared-off 22-gauge needle with sharpened edges is used to punch a hole in the side of the mold. The paired probe wires are pulled through this hole until the crystal is inside the mold. The crystal should now be perpendicular to the lumen, facing the 45” angle cut (Fig. 1B). The beveled end of the Silastic mold is filled with sticky dental wax (Caulk Red Sticky Wax; R. D. Caulk, Milford, DE). A 27.gauge needle is used to fill the unbeveled end of the mold with Silastic medical adhesive, which is allowed to cure for 24 h. This material stabilizes the crystal in the mold. A thin film of polyurethane foam (Plast 24/25A, Fibre-Glast Developments) is spread with a 27-gauge needle over the surface of the cured Silastic. The polyurethane foam, which acoustically baffles the back side of the crystal preventing the detection of flow from any surrounding areas (Fig. lC), also is allowed to cure 24 h.

Stage D. The final stage of probe construction (Figs. 1, C and D) is the formation of the Silastic cuff that is sutured around the vessel. The crystal assembly is then attached to a stainless steel hypodermic needle, appro- priate in size to the diameter of the vessel, by warming the wax slightly and pressing the mold flush with the needle (Fig. 1C). When the wax cools, it will hold the mold firmly to the needle. A small amount of adhesive (Silastic medical adhesive type A 891, Dow Corning) is placed on the end of a 27.gauge needle and painted on

Chrome/Gold Plate

20mHz Piezoelectric Crysta!

Acoustic Baffle

S/S Tubing f

FIG. 1. Stages in construction of di- rectional pulsed Doppler flow probe. A: initial procedure of attaching wire leads to piezoelectric crystal; B: placement of crystal in depressed epoxy; C relation- ship of probe to mold, placement of acoustical buffer, and orientation of mold to needle with dental wax; D: encapsu- lation of probe with medical-grade Silas- tic.

by 10.220.33.5 on August 28, 2017

http://ajpheart.physiology.org/D

ownloaded from

Page 3: Regional blood flow measurement with pulsed Doppler flowmeter … · 2017-08-28 · TENOW, GREGORY D. FINK, AND MICHAEL J. BRODY. RegionaL blood flow measurement with pulsed Doppler

REGIONAL BLOOD FLOW IN CONSCIOUS RAT

the crystal mold complex (Fig. 1D) by rotating the needle to cover all sides. A large cork may be used to hold the needle to facilitate rotation of the mold during encapsu- lation. Care should be taken to prevent a bulky buildup of Silastic on the mold. The adhesive should be rather thin on the upper part of the mold but thicker on the underside opposite the probe to add strength for suturing. After 24-h curing, the encapsulated probe is removed from the needle by cutting parallel to the needle on the side opposite the crystal. The cut should be straight to ensure even flaps for suturing. A warm soldering iron is placed on the needle to melt the dental wax, and the probe is very gently peeled off the needle. The remaining wax is removed from the probe cavity with small forceps. The pencil soldering iron is then inserted into the cavity to melt away any remaining wax from the surface of the crystal.

Chronic Implantation of Flow Probes

Measurement of mesenteric, r.enal, and hindlimb blood flow velocity in the consciqus animal requires chronic placement of the Doppler flow probes on the major arteries. To accomplish this, rats are anesthetized with 50 mg/kg pentobarbital sodium, and a midline abdominal incision is made, exposing the superior mesenteric and left renal arteries and the abdominal aorta. After the lumen of the probe is filled with coupling gel (Ultrasonic Gel, Tucker Laboratories, Orange, NJ), the probe is placed around the blood vessel and loosely sutured closed with 6-O ophthalmic silk (Fig. 2). The insulated wire leads are anchored to the abdominal wall to prevent a change in the orientation of the probe to the vessel. The wires are then tunneled out the back of the abdominal cavity and led subcutaneously to the back of the neck. The wire leads are soldered to an ultraminiature receptacle (GF-6, Microtech, Boothwyn, PA). Three holes are then drilled in the skull (no. 68 drill bit) and small stainless steel screws are inserted (no. 52-10, Lomat Watch, Montreal, Canada). The receptacle is placed over the screws, and the entire unit is attached to the skull with cranionlastic

8 Mesenteric

Flow

Renal Flow ww

Aortic Flow (kW

Arterial Pressure mmw

H275

cement (Plastic Products). An ultraminiature plug (GM- 6, Microtech) with shielded miniature twisted pair cables (NMUF=2/30=4046SJ, Cooner Sales) with a flexible spring covering connects the Doppler flowmeter to the receptacle containing the flow-probe wires. The signals are then recorded either as pulsatile or mean flow velocity of a Doppler shift (kHz) on a Beckman Dynograph. In addition, polyethylene catheters (PE-50) are placed in the femoral artery and jugular vein and exteriorized through the back of the neck. Femoral artery pressure is monitored with a Century CP-01 pressure transducer, permitting calculation of regional vascular resistances from pressure and mean flow recordings. Drug injections are made via the venous catheter.

Comparison of Pulsed Doppler and Electromagnetic Flowmete&

To ascertan the validity of the pulsed Doppler flow- meter for use in determining acute changes in regional blood flow, a simultaneous comparison was made of responses measured with Doppler and electromagnetic flowmeters. Blood flow to the left kidney was measured with an electromagnetic flowmeter using the autoperfu- sion method developed by Fink and Brody (2). This technique involves the diversion of blood from the left carotid artery through an extracorporeal flow probe ret- rograde into the lower aorta. Blood flow is supplied only to the left kidney by occluding the aorta between the superior mesenteric artery and the left renal artery, thus creating a cul-de-sac in the aorta and leaving only the kidney through which blood may pass. By cannulating the lower aorta, a section of the vessel was free for placement of the Doppler flow probe with an 18.gauge lumen while leaving the kidney and renal artery un- touched. All other blood vessels between the lower aorta and left renal artery were ligated. Alterations in renal blood flow were induced by injection of graded doses of norepinephrine (10, 20, and 40 ng) and acetylcholine (3, 10, and 30 ng) intraarterially into the extracorporeal tubing.

8

FIG. 2. Pulsatile flow tracings of mes- enteric, abdominal aortic, and renal blood flows in conscious rat as measured by Doppler shift with pulsed Doppler flowmeter. Femoral artery catheter per- mits simultaneous monitoring of arterial pressure.

by 10.220.33.5 on August 28, 2017

http://ajpheart.physiology.org/D

ownloaded from

Page 4: Regional blood flow measurement with pulsed Doppler flowmeter … · 2017-08-28 · TENOW, GREGORY D. FINK, AND MICHAEL J. BRODY. RegionaL blood flow measurement with pulsed Doppler

H276 HAYWOOD ET AL.

Measurement of Doppler Shift

Velocity signals recorded from flow probes around vessels are made optimal for each probe separately. The range adjustment of the flowmeter is altered until the peak voltage displacement (analogous to the Doppler shift) is achieved. This corresponds to focusing the ultra- sound on the midstream of the vessel. Zero flow is estab- lished by turning the ultrasound signal off. (In all exper- imental trials electronic zero was equal to zero flow obtained by arterial occlusion distal to the probe.)

Data Analysis

Changes in renal blood flow (ml/min) and flow velocity in Doppler shift (kHz) were compared by least-squares regression analysis. A similar correlation was made for percent changes in vascular resistance calculated for the two methods. Significance of the correlation was deter- mined by testing the significance of the Pearson product- moment correlation. Significance was assumed at the P < 0.05 level.

RESULTS

The flow probes constructed in our laboratory and the pulsed Doppler flowmeter have been used repeatedly in both acute and chronic rat preparations. As shown in Fig. 2, excellent pulsatile velocity profiles that coincide with arterial pressure pulses were recorded from con- scious rats. During short-term chronic experiments (2-6 days), rats prepared with Doppler flow probes have re- mained healthy. No experiments are performed the day after surgery. In chronic experiments, the incidence of probe failure has been very low.

A typical response to an intravenous bolus of norepi- nephrine administered to a conscious rat 3 days after implantation is shown in Fig. 3. The predictable de- creases in flow to the gut and kidney occurred. Little

4 Mesenteric Blood Flow 2 - v

(KHZ) 0 8

Hindquarter Blood Flow 4 E - w -

w-w 0 8

Renal

Blood Flow 4

(KHz) 7

0 E

240 Arterial

Pressure 120 mmw F

OL t t t

Norepinephrine Hexamethonium Norepinephrine 1.0 pg/kg 25 mg/kg 1.0 pg/kg

FIG. 3. Typical responses in conscious rat of mesenteric, hind- quarter, and renal mean blood flow velocity measured with pulsed Doppler flowmeter and arterial pressure to 1.0 pg/kg norepinephrine before and after ganglionic blockade with 25 mg/kg hexamethonium. Vascular resistance determinations immediately preceding and during peak response to each intervention are printed on each flow tracing.

change in hindquarter flow, presumably resulting from nearly simultaneous direct vasoconstriction and reflex vasodilation, was seen. All three vascular beds were calculated to exhibit an increase in vascular resistance. Ganglionic blockade with hexamethonium caused a large decrease in arterial’pressure, leading to a passive decrease in mesenteric and hindquarter blood flows. Renal flow remained relatively constant despite the large decrease in arterial pressure, presumably due to autoregulation. Following hexamethonium, the same dose of norepineph- rine again caused decreases in mesenteric and renal blood flow and a slight increase in hindquarter flow. In this example, the increase in hindquarter blood flow is pre- sumably due to the larger increase in arterial pressure redistributing cardiac output to the muscle bed. The calculated change in vascular resistance to all three vas- cular beds during norepinephrine injection after gangli- onic blockade was potentiated.

A comparison of instantaneous responses detected by the pulsed Doppler systemrand the electromagnetic flow- meter (Fig. 4) indicates that changes in Doppler shift elicited by renal nerve stimulation correlate very well, both qualitatively and quantitatively, with alterations in renal blood flow. When changes in renal blood flow produced by graded doses of acetylcholine and norepi- nephrine were compared by both methods of blood flow measurement, a linear relationship was observed (Fig. 5)) with a highly significant correlation coefficient (r = 0.969, P < 0.001). Individual regression lines for the five animals were very consistent, as evidenced by minimal scatter of the data points (Fig. 5). When the responses to norepi- nephrine and acetycholine were compared in terms of percent change in initial vascular resistance (changes in Doppler shift/initial s Doppler shift), the two methods again showed a highly significant correlation (r = 0.973, P c 0.001).

DISCUSSION

The pulsed Doppler flowmeter affords an unique op-

300 r Arterial

Pressure (mmHg)

150

0 t-

Renal Blood Flow Electromagnetic

Flow (ml/min)

Renal Blood Flow

-ft ## ++ +* 2Hz 4Hz 8Hz 16Hz

Renal Nerve Stimulation

FIG. 4. Simultaneous measurement of arterial pressure and renal blood flow monitored with an electromagnetic flowmeter and pulsed Doppler flowmeter. Vasoconstriction induced by graded frequencies of renal nerve stimulation showing proportional responses between 2 methods of flow measurement.

by 10.220.33.5 on August 28, 2017

http://ajpheart.physiology.org/D

ownloaded from

Page 5: Regional blood flow measurement with pulsed Doppler flowmeter … · 2017-08-28 · TENOW, GREGORY D. FINK, AND MICHAEL J. BRODY. RegionaL blood flow measurement with pulsed Doppler

REGIONAL BLOOD FLOW IN CONSCIOUS RAT HZ77

y = 0.358x - r = 0.969

A Doppler Shift (KHz)

+0.8

1.0 +2.0

FIG. 5. Comparison of changes in renal blood flow measured with square-wave electromagnetic flowmeter (Carolina Medical Electronics) and pulsed Doppler flowmeter in 5 rats. Changes in flow were caused by graded doses of acetylcholine and norepinephrine. Least-squares regression analysis yielded linear relationship with correlation coeffi- cient of r = 0.969.

portunity to record changes in regional blood flow in conscious freely moving rats. Other blood flow measure- ment methods require the use of large flow probes and bulky lead wires that preclude their use on several small blood vessels in the same animal. The advantages of the pulsed Doppler system, on the other hand, are numerous. The use of a single piezoelectric crystal and 36-AWG wire leads permits miniaturization of the flow probe so that it is approximately 2.5-4 mm diam. The probes easily fit around the renal and mesenteric arteries and the lower abdominal aorta of the rat without interfering with the function of other organs. Furthermore, the fine-gauge wires tunneled subcutaneously do not impede normal body movements. Zero flow is determined without an occluding cuff. Finally, although volume flow is not de- rived quantitatively, percentage changes in flow, and thus vascular resistance, can be calculated accurately.

Detection of flow with the pulsed Doppler system is dependent on changes in the emitted ultrasonic fre- quency caused by the reflection of the signal off moving blood cells. The change in frequency, referred to as Doppler shift, is proportional to the velocity of the blood cells in the vessel, given that the angle of the crystal to the vessel and the velocity of sound through the fluids and surrounding tissues remain constant. Then, assuming the cross-sectional area of the vessel adjacent to the probe remains constant, the Doppler shift is also propor- tional to the actual blood flow to the organ. This linear relationship between volume flow and Doppler shift ex- ists for both the continuous-wave Doppler flowmeter (13)

REFERENCES

and the pulsed Doppler flowmeter (5). However, differ- ences may exist from animal to animal in the actual Doppler shift for a given blood flow. This may be related to the orientation of the probe to the vessel, the coupling medium between the probe and vessel, or the intrinsic characteristics of each probe.

Although decreases in Doppler shift change linearly with blood flow during vessel occlusion, it is of consider- able interest to determine the relationship between blood flow alterations and Doppler shift during pharmacologic interventions that both increase and decrease organ blood flow. In this study, comparison of the responses measured with the pulsed Doppler system and an elec- tromagnetic flowmeter indicates that changes in Doppler shifts induced by graded doses of norepinephrine or acetylcholine are proportional to alterations in electro- magnetically measured renal blood flow. Furthermore, when percent changes in renal vascular resistance to the vasoactive agents were calculated, a significant correla- tion continued to exist for the two methods of blood flow measurement.

The proportionality of the changes in blood flow de- tected by the ultrasonic pulsed Doppler and the electro- magnetic flowmeter systems indicate the Doppler method is a valid technique for measuring alterations in regional blood flow or changes in vascular resistance in the awake rat. The observations described in this paper demonstrate the feasibility of simultaneous measurement in the rat of renal, mesenteric, and hindquarter blood flow and arterial pressure. The development of refined techniques for Doppler flow-probe construction and im- plantation in the rat permit the opportunity to gain further understanding of the regulation of the cardiovas- cular system during experimental or pharmacologic in- terventions in the conscious freely moving rat.

Although we have now had considerable experience with these techniques in rats (3, 7, 9, 10, 12), this report describing flow-probe construction and animal prepara- tion make the pulsed Doppler flowmeter system available for extensive application. In fact, investigators at our institution are now using modifications of the basic tech- nology described here to record cerebral blood flow ve- locity from stereotaxically placed probes mounted on needles (8), coronary flow velocity in human subjects (10A) and in rats (11) using probes attached to the cardiac surface with suction (lOA), pial artery flow veloc- ity recorded through a pial window (l), and regional hemodynamics in the conscious cat. It is anticipated that many other investigators with special needs for blood flow measurement will be attracted to the advantages of the pulsed Doppler approach.

Present addresses: J. R. Haywood, Dept. of Pharmacology, Univer- sity of Texas Health Science Center, San Antonio, TX 78284; G. D. Fink, Dept. of Pharmacology, Michigan State University, East Lansing, MI 48824.

Received 3 March 1980; accepted in final form 16 March 1981.

1. BUSIJA, D. W., D. D. HEISTAD, AND M. L. MARCUS. Effect of blood flow changes to renal nerve stimulation and intraarterial sympathetic nerves on cerebral vessels during acute, moderate drug administration in the rat. Am. J. Physiol. 234 (Heart Circ. increases in arterial pressure in dogs and cats. Circ. Res. 46: 696- Physiol. 3): H219-H222, 1978. 702, 1980. 3. FINK, G. D., J. BUGGY, J. R. HAYWOOD, A. K. JOHNSON, AND M. J.

2. FINK, G. D., AND M. J. BRODY. Continuous measurement of renal BRODY. Hemodynamic effects of electrical stimulation of forebrain

by 10.220.33.5 on August 28, 2017

http://ajpheart.physiology.org/D

ownloaded from

Page 6: Regional blood flow measurement with pulsed Doppler flowmeter … · 2017-08-28 · TENOW, GREGORY D. FINK, AND MICHAEL J. BRODY. RegionaL blood flow measurement with pulsed Doppler

H278 HAYWOOD ET AL.

angiotensin and osmosensitive sites. Am. J. Physiol. 235 (Heart Circ. Physiol. 4): H445-H45 1, 1978.

4. FRANKLIN, D. L., D. W. BAKER, R. M. ELLIS, AND R. F. RUSHMER. A pulsed ultrasonic flowmeter. IRE Trans. Med. EZectron. 6: 204- 206, 1959.

5. HARTLEY, C. J., AND J. S. COLE. An ultrasonic pulsed Doppler system for measuring blood flow in small vessels. J. Appl. Physiol. 37: 626-629, 1974.

6. HARTLEY, C. J., R. M. LEWIS, C. L. SEIDEL, C. GREEN, AND F. A. DUNN. Multichannel pulsed Doppler blood flow sensing in con- scious rats (Abstract). Annu. Conf. Eng. Med. BioZ. 30th Loj: Angeles, CA. p. 354, 1977.

7. HAYWOOD, J. R., R. A. SHAFFER, AND M. J. BRODY. Measurement of regional blood flow and arterial pressure in the conscious spon- taneously hypertensive rat. Jpn. Heart J. 20: 93-94,1979.

8. KNUEPFER, M., D. D. HEISTAD, M. L. MARCUS, AND M. J. BRODY. Continuous monitoring of cerebral arterial circulation using a pulsed Doppler flowmeter (Abstract). Federation Proc. 37: 733, 1978.

9. KNUEPFER, M. M., F. J. GORDON, A. K. JOHNSON, AND M. J. BRODY. Identification of descending cardiovascular pathways from

the anteroventral third ventricle (AV3V) region (Abstract). Fed- eration Proc. 38: 263,1979.

10. MAHONEY, L. T., J. R. HAYWOOD, W. J. PACKWOOD, A. K. JOHN- SON, AND M. J. BRODY. Effect of anteroventral third ventricle (AV3V) lesions on regional hemodynamic responses to renal affer- ent nerve stimulation (Abstract). Circulation 58, Sup&. II: 68,1978.

~OA.MARCUS, M., C. WRIGHT, D. DOTY, C. EASTXIAM, D. LAUGHLIN, P. KRUMM, C. FASTENOW, AND M. J. BRODY. Measurements of coro- nary velocity and reactive hyperemia in the coronary circulation of man. Circ. Res. In Press.

11. PETERS, K. G., R. D. WANGLER, D. LAUGHLIN, P. KRUMM, R. 3. TOMANEK, AND M. L. MARCUS. Quantitative characterization of myocardial reactive hyperemia in the rat (Abstract). Federation Proc. 39: 397, 1980.

12. Touw, K. B., J. R. HAYWOOD, R. A. SHAFFER, AND M. 3. BRODY. Contribution of the sympathetic nervous system to vascular resist- ance in the conscious young and adult spontaneously hypertensive rats. Hypertension 2: 408-418, 1980.

13. VATNER, S. F., D. FRANKLIN, AND R. L. VAN CITTERS. Simultane- ous comparison and calibration of the Doppler and electromagnetic flowmeters. J. Appl. Physiol. 29: 907-910, 1970.

by 10.220.33.5 on August 28, 2017

http://ajpheart.physiology.org/D

ownloaded from