![Page 1: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/1.jpg)
Refresher Course 18Refresher Course 18
Shielding Of Medical Facilities. Shielding Design ConsiderationShielding Of Medical Facilities. Shielding Design Considerations For PETs For PET--CT CT FacilitiesFacilities
Juan Angel Cruzate and Adrián Pablo Discacciatti
![Page 2: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/2.jpg)
NonNon--invasive Medical Imaging Techniquesinvasive Medical Imaging Techniques
AnatomicalAnatomical
FunctionalFunctional
X-rays
CT ( Computed Tomography)
MRI (Magnetic Resonance Imaging )
Ultrasound
Nuclear Medicine
CT
PET
![Page 3: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/3.jpg)
Positron Emission Tomography
Positron Decay (β+)
Gamma coincidence detection
![Page 4: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/4.jpg)
Positron Emission Tomography
F-18 FDG Fluoro-2-deoxyglucose
Diagnostic technique ⇒ Radiotracer must have short half-life
Typical radiotracerused in PET
123
18F15O11C 13N
109.8 min
2 min
20.4 min
10 minthe most used
![Page 5: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/5.jpg)
Positron Emission Tomography
what about the images from PET scanner ?
so, we can see the tumour, but no anatomy details
![Page 6: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/6.jpg)
Positron Emission Tomography
then, how can a physician make a diagnosis ?
IMAGE FUSION
PETCT IMAGE FUSION
![Page 7: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/7.jpg)
here we have other examples of image fusion
PET
CT
IMAGEFUSION
![Page 8: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/8.jpg)
Positron Emission Tomography
IMAGE FUSION PET/CT units⇒
![Page 9: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/9.jpg)
PET/CT units
![Page 10: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/10.jpg)
PET/CT units
![Page 11: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/11.jpg)
PET/CT units
CONTROL CONSOLE
![Page 12: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/12.jpg)
Radiation Protection
typical radiotracer used in PET
123
18F15O11C 13N
109.8 min
2 min
20.4 min
10 min
the most used
the longesthalf-life
consequently, radiation protection measures are drew out assuming only studies with 18F
![Page 13: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/13.jpg)
Typical practice
The vial with the radionuclide is received and stored in the “hot lab”
radionuclide reception
![Page 14: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/14.jpg)
Typical practice
Prior to injection to the patient, the radioactive aliquot has to be transferred from the vial to the syringe.
“dose” preparation
![Page 15: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/15.jpg)
Typical practice
Then, syringe is transported up to the place of injection to patient (injection room)
syringe transportation
![Page 16: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/16.jpg)
Typical practice
patients are injected with an average 18F activity of about 555 MBq (15 mCi)
injection
![Page 17: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/17.jpg)
Typical practice
patients lie down in the uptake room for about 45-60 minutes (uptake time)
• to allow distribution of radionuclide throughout the body
• for reduction of uptake in the skeletal muscles
generally injection room and uptake room are the same
uptake
![Page 18: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/18.jpg)
Typical practicestretcher
![Page 19: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/19.jpg)
Typical practice
patients are instructed to void their bladder of urine accumulation (about 15 % of administered activity) before scanning
• to reduce gamma dose in the bladder
• to avoid signal interference due to gamma emission from bladder
uptake
![Page 20: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/20.jpg)
main gamma emission from annihilation photons(not from tumours)
![Page 21: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/21.jpg)
Typical practice
patients are given an approximately 20-30 minutes scan (acquisition time)
scanning
![Page 22: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/22.jpg)
but, what are the radiation sources ?
![Page 23: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/23.jpg)
Why is PET shielding different ?
• The 511 keV photons from 18F and the mobile nature of the source (source itself and patient) create some unique shielding design problems for a PET clinic.
![Page 24: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/24.jpg)
Why is PET shielding different ?
![Page 25: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/25.jpg)
Why is PET shielding different ?
• New clinics are commonly sandwiched into existing imaging centers that are densely populated. Areas above and below the clinics are routinely occupied by other offices.
![Page 26: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/26.jpg)
Radiological evaluation
• to exposed workers• to members of the public
Assesment of the annual effective dose⇓
![Page 27: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/27.jpg)
Radiological evaluation• radionuclide involved• layout of the facility• shieldings• use of spaces in and out the facility • expected number of patients per year• working procedures
• administered activity
• uptake time
• acquisition time
![Page 28: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/28.jpg)
Radiological evaluation
• PET facility must comply with regulatory standards from its country
• Then, assessed annual effective dose must be compared with dose constrains from those standards
• And, obviously, must be less than those constrains
![Page 29: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/29.jpg)
effective dose
effective dose depends on
• incident photon energy
• irradiation geometry
radiationprotection ⇒
![Page 30: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/30.jpg)
common irradiation geometries
![Page 31: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/31.jpg)
but, in general, during a given practice we have a combination of geometries
![Page 32: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/32.jpg)
effective dose dependenceAP yields the highesteffective dose
![Page 33: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/33.jpg)
Effective dose rate constant (Γ)
the effective dose per unit time and per unit activity at 1m from an “unshielded” source
In radiation protection we are interested in antero-posterior geometry, because it shields
the most conservative value
![Page 34: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/34.jpg)
Effective dose rate constant (Γ)
1.39 10-4 mSv m2 / h MBq
![Page 35: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/35.jpg)
Administered activity decay
uptup eAtA λ−= 0)( 0Aα−
Adults ≈ 370-740 MBq (10-20 mCi)
children ≈ 4-5 MBq (108-135 μCi) per Kg
≈ 45 - 60 minutes≈ 10 - 15 %
![Page 36: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/36.jpg)
Transmission Factor
( )( )oo
o
rDrDk =
effective doseconsideringno shielding
effective doseconsideringshielding
![Page 37: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/37.jpg)
Materials used in PET shielding
• Lead (δ = 11.35 g / cm3)
• Steel (δ = 7.86 g / cm3)
• Ordinary concrete (δ ≈ 2.35 g / cm3)
• Masonry (mix of brick and mortar,δ ≈ 1.6 g / cm3)
• patient (δ ≈ 1 g / cm3)
![Page 38: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/38.jpg)
Materials used in PET shielding
![Page 39: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/39.jpg)
Materials used in PET shielding
but, what about k values for masonry ?
we can use k for ordinary concrete, but using
concrconcr
brickbrick XX
δδ
=
taking into account that δ ≈ 1.6 g/cm3
concrconcrconcrbrick XXXcmgcmgX 3/268.0
/35.2/6.1
3
3
≅==
![Page 40: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/40.jpg)
and what about the patient ?
studies suggest an effective body absorption factor of 0.36
so, for radiation protection patient may be considered as a point source with a transmission factor of k=0.64
![Page 41: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/41.jpg)
Dose Reduction Factor
( ) dteDtD tt
λ−•
∫=0
0( ) ( )teDtD λ
λ−•
−= 110
( ) ( )tet
tDtD λ
λ−•
−= 110
( )tt e
tR λ
λ−−⋅
⋅= 11
tRtDtD ⋅=•
0)(
⇒ ⇒
⇒
![Page 42: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/42.jpg)
Dose Reduction Factor
![Page 43: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/43.jpg)
Occupancy factor, T
the average fraction of the effective irradiation time that a maximally exposed individual remains in the area or point of interest
i.e the person who spends the most time there
![Page 44: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/44.jpg)
Occupancy factor, Tcontrolled areas
![Page 45: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/45.jpg)
Calculation model
point source model is used
• is a simple model
• is conservative for radiation protection purposes
![Page 46: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/46.jpg)
Calculation model
TRkd
NtAD t
op2Γ=
effective annual dose produced during operation
effective dose rate constant
activity when operation begins time duration
of operation
number of annual operations
distance from source to calculation point
transmission factor dose reduction factor
occupancy factor
![Page 47: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/47.jpg)
Radiation sources
• sources stored in hot lab
• patient in uptake room
•patient in PET/CT scanner
![Page 48: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/48.jpg)
The hot lab
lead container
typically 5 cm-thick k ≈ 5 10-4⇒
![Page 49: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/49.jpg)
Radiation sources
• sources stored in hot lab
• patient in uptake room
•patient in PET/CT scanner
![Page 50: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/50.jpg)
Calculation modelPUBLIC
WORKER
![Page 51: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/51.jpg)
Typical dose assessment
let’s suppose a typical PET study
• administered activity : 555 MBq (15 mCi)
• uptake time : 60 minutes
• acquisition time : 45 minutes
• 40 patients per week
![Page 52: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/52.jpg)
Typical dose assessment
dose constrains from Nuclear Regulatory Authority of Argentina will be supposed
• occupational exposure : 5 mSv per year
• members of public : 100 μSv per year
![Page 53: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/53.jpg)
Radiation protection considerations
usually, surrounding uncontrolled areas in the vicinity of the PET facility have an occupancy factor of 1
radiation safety must be guaranteed by the facility design itself and not by particular
procedures
![Page 54: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/54.jpg)
Radiation protection considerations
• Prior to injection to the patient, the radioactive aliquot has to be transferred from the vial to the syringe.
• This operation is carried out interposing a lead shield between the vial and the operator.
• In order to allow visual shadowing of the operation, lead glass has to be used.
• This operation may last only a few seconds
![Page 55: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/55.jpg)
Radiation protection considerations
lead glass
![Page 56: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/56.jpg)
Radiation protection considerations
• after the syringe was loaded with the radioactive solution, it has to be translated to the uptake room, where the patient is injected.
lead
![Page 57: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/57.jpg)
Radiation protection considerations
• Once in the uptake room, radioactive aliquot is injected to the patient.• Because of the high effective dose constant associated, hand doses for individuals drawing up and administering PET radiopharmaceuticals can be substantial.
• For the typical case described, the dose rate at 5 cm from an unshielded syringe is 30.86 mSv/h
• This operation may last between up to 1 min,
• In this case the annual dose to a technician should be about 1070 mSv
![Page 58: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/58.jpg)
Radiation protection considerations
Tungsten syringe shields can reduce the hand dose, but the additional weight can make injections difficult.
![Page 59: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/59.jpg)
Radiation protection considerations
• The injection may be administered to the patient using a catheter, with the syringe located within the mentioned lead shielding.
![Page 60: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/60.jpg)
Radiation protection considerations
• Other ways to reduce hand dose
• Use automatic dispensing systems
• Divide the injection responsibilities among the staff.
![Page 61: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/61.jpg)
Radiation protection considerations
• If a single member of the staff would remain at 1 m from the patient during uptake time, then the annual dose to this technician should be 85 mSv
• Consequently, the staff should develop procedures to minimise the time spent near the radioactive patient.
In other words, the practice should be “optimised”.
![Page 62: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/62.jpg)
Radiation protection considerations
• Information collection, explanations, blood collection or other tests should be performed as much as possible before radioactivity has been administered.
• Remote monitoring of the patients using video cameras could also be used.
![Page 63: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/63.jpg)
Radiation protection considerations• Portable lead shields can be used effectively to shield patients in uptake rooms, where they are required to remain stationary.
![Page 64: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/64.jpg)
Radiation protection considerations
• The gantry of the PET tomograph can provide a substantial reduction of the dose rate.
• This depends on the actual geometry and placement of the tomograph in the room as well as the type of scanning procedures.
• If information on the tomograph shielding characteristics is available from the vendor, it can be incorporated into the calculation.
but normally the most conservative approach is taken, i.e. no shielding from the tomograph is
assumed.
![Page 65: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/65.jpg)
Radiation protection considerations
• During image acquisition, at least one technologist is located at the PET system console where both, the patient and the progress of the imaging study, can be monitored.
• Ideally, the console area should be located more than 2 m away from the scanner to reduce the operator dose below ALARA levels.
• In this case, the annual dose results to be 11.4 mSv
• which correspond to about 7 mm of lead or equivalent
![Page 66: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/66.jpg)
wallslead + masonry
CONTROL CONSOLE
viewing windowglass lead
![Page 67: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/67.jpg)
Radiation protection considerations
• New facilities can efficiently use concrete to achieve required shielding factors, while in existing facilities lead is often the best resort.
• Uncontrolled areas with high occupancy should be located as far as possible from the uptake and imaging rooms.
![Page 68: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/68.jpg)
Radiation protection considerations
• Portable lead shields may be used effectively to shield patients in uptake rooms, where they are required to remain stationary.
![Page 69: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/69.jpg)
Radiation protection considerations
• The placement of the door must be carefully considered to avoid the expense with installing one with substantial lead shielding.
• It is a good idea to have the hot bathroom, reserved for PET patients, within the immediate imaging area, so that they do not alter the background counts of other detection devices as they pass through the clinic.
![Page 70: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/70.jpg)
Radiation protection considerations
• If uncontrolled areas are located above and below the uptake and tomograph rooms, the spacing between floors may need to be greater than normal.
• If that is not feasible, the floors need to be able to bear the weight associated with additional shielding.
![Page 71: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/71.jpg)
Radiation protection considerations
lead layers
![Page 72: Refresher Course 18 Shielding Of Medical Facilities ...EF74BEF9-6C7E-4DFC-9B4E... · Shielding Of Medical Facilities. Shielding Design Considerations For PET-CT ... Positron Emission](https://reader031.vdocuments.us/reader031/viewer/2022011802/5b3f55347f8b9a3a138bf8c1/html5/thumbnails/72.jpg)
Thanks for your attention…