recessed diss style frontall™ outlets - class 1 inc.€¦ · 13-03-2014  · diss recessed...

2
Updated: March 13, 2014 © 1997- 2014 Class 1 Inc. | 565 Boxwood Drive | Cambridge, ON | N3E 1A5 | T: 1 (800) 242-9723 | class1inc.com Recessed DISS Style FRONTALL™ Outlets Oxygen | Vacuum | Medical Air | Nitrous Oxide | Nitrogen | Carbon Dioxide WAGD | Instrument Air | Slide Patented Cartridge Valve Design Ease of Maintenance Compensates for variation in wall thickness from ½" to 1¼" Bilingual Labelling ed CRN Registered cations 5" centres 360° swivel inlet tube The Class 1 Inc. Frontall™ medical gas recessed outlets represent the most technically advances means of providing medical gas connections to the central piping system. The outlets incorporate a unique patented front-loaded feature which makes the outlet easy to repair, saving valuable time and money. The outlets shall be modular singles, and shall incorporate nameplate colour coding in accordance with CSA Z7396.1 / NFPA99. The medical gas outlets shall be designed so that, once installed, routine service of both the primary and secondary check valves can be accomplished without removing the c portions. The primary check valve shall be utilized and of the cartridge type. So that the primary the gas supply to the outlet, the secondary check valve shall ow of the positive pressure gas services. There shall be no secondary check for vacuum or WAGD services. Medical gas outlets which require c components for routine service shall not be acceptable. The seal between the nameplate and the rough-in shall be a double O-ring which is the gas service and is capable of supporting 10 pounds @ 2" from the nameplate without leaking. The outlet nameplate shall be permanently colour coded. The outlet trim plate shall be durable plastic, scratch resistant, have a protective label, attached with the nameplate to the rough-in assembly, and provide automatic plaster adjustment from 1/2 to 1-1/4 of an inch (1.3 cm to 3 cm). The name of the gas service shall be permanently marked on the outlet bracket and the chrome plated brass outlet body. The outlet's rough-in supply tube shall be a 7 inch (18 cm) length of 1/2 OD copper Type K, and labelled with the name of the gas service. Medical gas outlets shall be cleaned for oxygen service in accordance with the current Compressed Gas Association (CGA) Pamphlet G-4.1, capped and placed in a protective container for shipment. The outlets shall be installed in strict accordance with manufacturer's instructions, and tested before use in conformance with local and federal codes. Features

Upload: others

Post on 02-Aug-2020

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Recessed DISS Style FRONTALL™ Outlets - Class 1 Inc.€¦ · 13-03-2014  · DISS Recessed Outlets - Complete DISS Ceiling Outlets - Complete Medical Air Nitrous Oxide Oxygen Medical

Updated: March 13, 2014 © 1997- 2014

Class 1 Inc. | 565 Boxwood Drive | Cambridge, ON | N3E 1A5 | T: 1 (800) 242-9723 | class1inc.com

Recessed DISS Style FRONTALL™ OutletsOxygen | Vacuum | Medical Air | Nitrous Oxide | Nitrogen | Carbon Dioxide

WAGD | Instrument Air | Slide

Patented Cartridge Valve Design

Ease of Maintenance

Compensates for variation in wall thickness from ½" to 1¼"

Bilingual Labelling

ed

CRN Registered

cations

5" centres

360° swivel inlet tube

The Class 1 Inc. Frontall™ medical gas recessed outlets represent the most technically advances means of providing medical gas connections to the central piping system. The outlets incorporate a unique patented front-loaded feature which makes the outlet easy to repair, saving valuable time and money.

The outlets shall be modular singles, and shall incorporate nameplate colour coding in accordance with CSA Z7396.1/ NFPA99.

The medical gas outlets shall be designed so that, once installed, routine service of both the primary and secondary check valves can be accomplished without removing the

c portions. The primary check valve shall be utilized and of the cartridge type. So that the primary

the gas supply to the outlet, the secondary check valve shall

ow of the positive pressure gas services. There shall be no secondary check for vacuum or WAGD services. Medical gas outlets which require

c components for routine service shall not be acceptable. The seal between the nameplate and the rough-in shall be a double O-ring which is

the gas service and is capable of supporting 10 pounds @ 2" from the nameplate without leaking.

The outlet nameplate shall be permanently colour coded. The outlet trim plate shall be durable plastic, scratch resistant, have a protective label, attached with the nameplate to the rough-in assembly, and provide automatic plaster adjustment from 1/2 to 1-1/4 of an inch (1.3 cm to 3 cm). The name of the gas service shall be permanently marked on the outlet bracket and the chrome plated brass outlet body. The outlet's rough-in supply tube shall be a 7 inch (18 cm) length of 1/2 OD copper Type K, and labelled with the name of the gas service.

Medical gas outlets shall be cleaned for oxygen service in accordance with the current Compressed Gas Association (CGA) Pamphlet G-4.1, capped and placed in a protective container for shipment. The outlets shall be installed in strict accordance with manufacturer's instructions, and tested before use in conformance with local and federal codes.

Features

Page 2: Recessed DISS Style FRONTALL™ Outlets - Class 1 Inc.€¦ · 13-03-2014  · DISS Recessed Outlets - Complete DISS Ceiling Outlets - Complete Medical Air Nitrous Oxide Oxygen Medical

Updated: March 13, 2014 © 1997- 2014

Class 1 Inc. | 565 Boxwood Drive | Cambridge, ON | N3E 1A5 | T: 1 (800) 242-9723 | class1inc.com

Recessed DISS Style FRONTALL™ OutletsOxygen | Vacuum | Medical Air | Nitrous Oxide | Nitrogen | Carbon Dioxide

WAGD | Instrument Air

Ordering Information

Layout

DISS Recessed Outlets - Complete

DISS Ceiling Outlets - Complete

Medical Air Nitrous Oxide Oxygen Medical Vacuum

Nitrogen WAGD Carbon Dioxide

Instrument Air

XM1116 XM1104 XM1124A XM1122 XM1112 XM1132 XM1108 XM112080

Medical Air Nitrous Oxide Oxygen Medical Vacuum

Nitrogen WAGD Carbon Dioxide

Instrument Air

XM1216 XM1204 XM1224A XM1222 XM1212 XM1232 XM1208 XM122080