Recent Innovations in the field of traceable calibration of liquid

milli-flow rates with liquids other than water

Hugo Bissig, Martin Tschannen, Marc de Huu

Page 2: Recent Innovations in the field of traceable calibration ... · Recent Innovations in the field of traceable calibration of liquid milli-flow rates with liquids other than water Hugo

Agenda

1. Motivation

2. Milli-Flow Facility

3. Liquids other than water

4. Conclusion

Micro Nano 2016, Amsterdam NL, 13 – 14.12.2016 2

Page 3: Recent Innovations in the field of traceable calibration ... · Recent Innovations in the field of traceable calibration of liquid milli-flow rates with liquids other than water Hugo

3

Motivation

Accuracy: ± 5 %

Benefits of internationally

traceable calibrations

High-end

calibration lab

0.05 %

0.10 - 0.15 %

0.15 - 0.30 %

~ 1 %

2 ~ 5 %

Less

Un

ce

rtain

ty

More

NMI

Primary reference values

End-User

International traceability

Micro Nano 2016, Amsterdam NL, 13 – 14.12.2016

Page 4: Recent Innovations in the field of traceable calibration ... · Recent Innovations in the field of traceable calibration of liquid milli-flow rates with liquids other than water Hugo

4

Micro-Flow facility covers already 100 nl/min to 1 ml/min.

Motivation

But calibration times are long for flow rates below 1ml/min !

Micro Nano 2016, Amsterdam NL, 13 – 14.12.2016

Page 5: Recent Innovations in the field of traceable calibration ... · Recent Innovations in the field of traceable calibration of liquid milli-flow rates with liquids other than water Hugo

5

Micro-Flow and Milli-Flow Facilities

Motivation

Accurate and fast calibration method

for flow rates from 100 nl/min to 400 ml/min !

Water

Liquids

Micro Nano 2016, Amsterdam NL, 13 – 14.12.2016

Page 6: Recent Innovations in the field of traceable calibration ... · Recent Innovations in the field of traceable calibration of liquid milli-flow rates with liquids other than water Hugo

Milli-Flow Facility – working principle

6

Gravimetric and volumetric standard

Flow rate range: 0.2 ml/min – 400 ml/min

Pressure range: 0 bar – 8 bar (upstream DUT)

Temperature: 22 °C (ambient)

Uncertainties: 0.07 %

Micro Nano 2016, Amsterdam NL, 13 – 14.12.2016

Page 7: Recent Innovations in the field of traceable calibration ... · Recent Innovations in the field of traceable calibration of liquid milli-flow rates with liquids other than water Hugo

Milli-Flow Facility – working principle

7

(A) syringe pump, (B) beaker on balance,

(C, D, E) pressure sensors, (F, G, H) temperature sensors,

(I) pressure security valve, (J) water reservoir

Micro Nano 2016, Amsterdam NL, 13 – 14.12.2016

Page 8: Recent Innovations in the field of traceable calibration ... · Recent Innovations in the field of traceable calibration of liquid milli-flow rates with liquids other than water Hugo

Milli-Flow Facility – homemade syringe pump

8

(A) high precision linear stage, (B) linear measuring system,

(C) syringe, (D) mounting syringe body, (E) mounting and

positioning for syringe plunger.

Micro Nano 2016, Amsterdam NL, 13 – 14.12.2016

Page 9: Recent Innovations in the field of traceable calibration ... · Recent Innovations in the field of traceable calibration of liquid milli-flow rates with liquids other than water Hugo

Milli-Flow Facility– control of evaporation rate

(A) outlet needle, (B) beaker with cover, (C) glass filter,

(D) water in evaporation trap, (E) mount for T and rH sensor,

(F) balance, (G) tubing for humidity exchanger

9 Micro Nano 2016, Amsterdam NL, 13 – 14.12.2016

Page 10: Recent Innovations in the field of traceable calibration ... · Recent Innovations in the field of traceable calibration of liquid milli-flow rates with liquids other than water Hugo

Flow rate determination Least square Linear Regression

10

Page 11: Recent Innovations in the field of traceable calibration ... · Recent Innovations in the field of traceable calibration of liquid milli-flow rates with liquids other than water Hugo

Micro Nano 2016, Amsterdam NL, 13 – 14.12.2016

Repeatability and Reproducibility of measurements with different syringes and water

100 ml syringe

11

10 ml syringe

Page 12: Recent Innovations in the field of traceable calibration ... · Recent Innovations in the field of traceable calibration of liquid milli-flow rates with liquids other than water Hugo

Liquids other than water

12 Micro Nano 2016, Amsterdam NL, 13 – 14.12.2016

Page 13: Recent Innovations in the field of traceable calibration ... · Recent Innovations in the field of traceable calibration of liquid milli-flow rates with liquids other than water Hugo

Liquids other than water

13

The flow rate range of the milli-flow facility is limited by:

• pressure drop at the DUT

• liquid properties (viscosity)

Advantages

use any liquid

faster calibration procedure

Micro Nano 2016, Amsterdam NL, 13 – 14.12.2016

Page 14: Recent Innovations in the field of traceable calibration ... · Recent Innovations in the field of traceable calibration of liquid milli-flow rates with liquids other than water Hugo

Conclusion

14

Syringe Pump Micro-Flow Milli-Flow

Flow rates: 100 nl/min – 5 ml/min 0.2 ml/min – 400 ml/min

Pressure range

(upstream DUT):

0 – 8 bar (2.5 bar) 0 – 8 bar

Temperature: 22 °C (ambient) 22 °C (ambient)

Uncertainty (k=2): 0.70 % – 0.15 % 0.07 %

portable flow generator

any liquid

Capability

to calibrate

Method

Flow

sensors

or

Flow

generators

Micro Nano 2016, Amsterdam NL, 13 – 14.12.2016