![Page 1: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/1.jpg)
Review of PET Physics
Timothy Turkington, Ph.D. Radiology and Medical Physics
Duke University
Durham, North Carolina, USA
![Page 2: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/2.jpg)
Chart of Nuclides
N (number of neutrons)
Z (protons)
Nuclear Data Evaluation Lab.
Korea Atomic Energy Research Institute
![Page 3: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/3.jpg)
Nuclide half-life C-11 20.3 min N-13 10 min O-15 124 sec F-18 110 min Rb-82 75 sec
e.g., 18F 18O + e+ +
Positron Decay
ZAXNZ1
AYN1 e
+
![Page 4: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/4.jpg)
Positron Annihilation
e-e+
180 degrees
511 keV
511 keV
me=511 keV/c2
![Page 5: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/5.jpg)
Linear attenuation values for lead and water
Material (E) m (cm-1) HVT (cm) TVT (cm)
lead (140 keV) 22.7 0.031 0.10
lead (511 keV) 1.7 0.41 1.35
water (140 keV) 0.15 4.6 15.4
water (511 keV) 0.096 7.2 24.0
![Page 6: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/6.jpg)
Coincidence Event
![Page 7: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/7.jpg)
Projections
![Page 8: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/8.jpg)
Acquisition and Reconstruction
q
r
q
Data Acquisition Sinogram
(raw data)
Reconstructed Image
![Page 9: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/9.jpg)
PET Background Events
• Scatter
• Randoms
![Page 10: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/10.jpg)
Scattered Coincidence Event
In-Plane Out-of-Plane
Scatter Fraction S/(S+T)
With septa ~10-20%
w/o septa ~30-80%
![Page 11: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/11.jpg)
Random Coincidence Event
RR=2RaRb
a
b
![Page 12: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/12.jpg)
Correcting Background; Noise Equivalent Counts
)/?)2(/1(0;
0
0
2
TRTS
T
RP
TNEC
RP
T
T
TSNR
TPR
PRSPT
RSPT
RSTP randomsscattertruesprompts
S and R refer to scattered and random events on LORs that
subtend the imaged object.
More background more statistical image noise.
(Estimation of true events by subtracting S and R estimates)
(Variance propagation
and Poisson properties. 0
vs. R depends on randoms
correction method.)
![Page 13: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/13.jpg)
NEC Examples
TRTS
TNEC
1
Prompts Trues Scatter SF NEC
100 100 0 0 100
200 100 100 0.5 50
400 200 200 0.5 100
![Page 14: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/14.jpg)
Image Noise and Lesion Detection
600s 300s 150s 75s 38s 19s 10s
FBP
OS-EM
![Page 15: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/15.jpg)
Multiple Rings, 2D – 3D
direct
slices (n)
For n detector rings:
cross
slices (n-1)
total slices = 2n-1 Higher Sensitivity
(degraded axial resolution)
2D 2D 3D
![Page 16: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/16.jpg)
Time Of Flight PET: The influence of background
A B
D1 D2
How strong is source A?
Detectors measure counts CA+CB.
S.D. of measurement is
SNR = CA/
BA CC
BA CC
![Page 17: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/17.jpg)
SNR of measurement of A?
SNR = CA/
If all N sources are equal,
SNR = CA/
The influence of even more background
L DCBA CCCC
A B
D1 D2
C D E F
ANC
![Page 18: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/18.jpg)
If all N sources are equal,
SNR = CA/
where n is the number of sources within zone T1.
SNR improvement of sqrt(N/n); similar to counting longer by N/n.
Magical new detectors
A B
D1 D2
C D E F
T1 T2
Can distinguish between counts originating in segment T1
and counts originating in segment T2.
AnC
![Page 19: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/19.jpg)
Time-of-Flight PET
Detector midpoint
Annihilation location x
d1
d2
Dt=(d2-d1)/c=2x/c
x=Dt·c/2
![Page 20: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/20.jpg)
Fillable, Tapering
Phantom with 1 cm lesions
80 s
240 s
![Page 21: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/21.jpg)
Clinical Example 2
68 kg
male
TOF non-TOF
![Page 22: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/22.jpg)
Attenuated Event
![Page 23: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/23.jpg)
Coincidence Attenuation
d2
1d
AB C D
21
21
21
dde
de
de
PPPC
m
mm
Annihilation radiation emitted along a particular
line of response has the same attenuation
probability, regardless of where it originated on
the line.
![Page 24: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/24.jpg)
Attenuation losses - PET and SPECT
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
0 2 4 6 8 10 12 14 16 18 20
PET
140 keV SPECT
cylinder radius (cm)
Fra
ctio
n e
mitte
d
Events surviving attenuation in cylinder
Fra
ctio
n E
mitte
d fro
m c
ylin
de
r
![Page 25: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/25.jpg)
Obese Patient
![Page 26: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/26.jpg)
Technologist Size Variations
a
b mb~2.5ma
![Page 27: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/27.jpg)
Attenuation Effects
AC NAC x-ray CT
![Page 28: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/28.jpg)
Calculated Attenuation Correction
d I = I0e-md
![Page 29: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/29.jpg)
Transmission Attenuation Measurement
positron source
![Page 30: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/30.jpg)
Using CT image for AC
![Page 31: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/31.jpg)
Attenuation Correction Accuracy
• CT-based attenuation correction is performed on almost all PET studies.
• Is it being done well?
– Is the CT accurate (e.g., water = 0)?
– Is the CT accurate under all relevant conditions?
– Is the translation between CT# and 511 keV μ appropriate?
– Patient motion between CT and PET?
– …
![Page 32: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/32.jpg)
Attenuation Correction
Photons emitted along this
line will be attenuated by a
factor that can be determined
from the corresponding CT
scan.
![Page 33: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/33.jpg)
Attenuation Correction
The biggest source of error in
PET AC is patient motion
between the CT and the PET
scans.
This particular PET photon
trajectory will be
undercorrected. The intensity
on this side of the body will be
artificially low.
CT
![Page 34: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/34.jpg)
Spatial Resolution
2222 bRRRR rangeacoldetsys
Rdet = resolution of detectors ( d )
Racol = resolution from photon acollinearity (=0.0022D)
Rrange = resolution from positron range
b = block effect
![Page 35: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/35.jpg)
Depth of Interaction Uncertainty
• Uncertainty in the origin of radiation when measured obliquely in a detector.
• Resolution in radial direction worsens with increasing radius.
• High stopping power helps: Interactions more likely in front of detector
• Some high resolution systems sacrifice sensitivity by shortening detectors (to mitigate DOI effects.)
• Some systems (HRRT) use two layers of detectors to lessen effect. Others propose a measurement of the DOI.
![Page 36: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/36.jpg)
Block Detector (GE, Siemens)
Photomultiplier(s) Scintillation Crystals
![Page 37: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/37.jpg)
PET Ring with Block Detectors
![Page 38: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/38.jpg)
Curved Plate Pixelated Camera (Philips)
![Page 39: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/39.jpg)
Image Reconstruction
![Page 40: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/40.jpg)
Image Reconstruction Methods
FBP ML-EM 10 ML-EM 30 ML-EM 50
OS-EM 1 OS-EM 2 OS-EM 3 OS-EM 4
(28 Subsets)
Filtered Back-Projection Maximum Likelihood Expectation Maximization
Ordered Subsets Expectation Maximization
![Page 41: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/41.jpg)
Detection Process
ii
i
1
3
2
4
j
j
j
npix
j
jijii pbm1
Image reconstruction: λ=p-1(mi-bi) What is p-1?
i
j i = projection line (line of response)
j = source pixel
λj= radioactivity at voxel j
pij = probability of emission from j
being measure in i
bi = background contributing to i
mi = expected counts on projection i
![Page 42: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/42.jpg)
Extended Distribution Example
smoothed
![Page 43: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/43.jpg)
Maximum Likelihood Expectation Maximization (ML-EM)
npix
j
jijii pbm1
ii
i
1
3
2
4
j
j
j
i
j
i
nbin
invox
k
nkiki
njij
nbin
i
ij
nj m
pb
p
p
1
1
)(
1
)1( 1
j(n) is the estimated activity in
voxel j at iteration n.
Shepp LA, Vardi Y., IEEE Trans Med Imag 1:113-121, 1982.
Lange K, Carson R., J Comput Assist Tomo 8:306-316, 1984.
![Page 44: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/44.jpg)
Extended Distribution Example
1 2 3 4 5
10 20 30 40 50
![Page 45: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/45.jpg)
What is p?
ii
i
1
3
2
4
j
j
j
npix
j
jijii pbm1
i
j Increasing levels of complexity:
1) p is 1’s and 0’s - very sparse
2) p is fractional values, depending on how
column intersects with voxel - still pretty
sparse
3) p includes attenuation (lower values) -
still pretty sparse
4) p includes resolution effects (collimator
blurring, etc) - somewhat sparse
5) p includes background - not sparse at all
![Page 46: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/46.jpg)
Compensations/Corrections
Putting physical effects into p allows the reconstruction
to compensate for those effects.
For example, including attenuation in the model leads to
attenuation correction in the final image. Putting system
blurring (imperfect spatial resolution) can improve the final
image resolution (and/or noise).
![Page 47: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/47.jpg)
ML-EM Characteristics
• Not Fast
• Non-negative pixel values - can lead to
biases
• Noise is greatest in areas of high activity
• Can compensate for physical effects
• Allows image reconstruction from limited
projections.
![Page 48: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/48.jpg)
Ordered Subsets Expectation Maximization (OS-EM)
Hudson HM, Larkin RS . IEEE Trans Med Imag 13:601-609, 1994.
i
nbin
invox
k
nkiki
njij
nbin
i
ij
nj m
pb
p
p
1
1
)(
1
)1( 1
Instead of processing all projection for each image
update, projections are divided into subsets, and the image
is updated after processing each subset.
![Page 49: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/49.jpg)
1 2 3 4 5
10 20 30 40 50
ML-EM
OS-EM 10 subsets
2 ss 4 ss 6 ss 8 ss 1 iter 2 iter
![Page 50: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/50.jpg)
Results – 4:1 Hot Spheres
![Page 51: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/51.jpg)
Regularized Reon (Q.Clear)
![Page 52: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/52.jpg)
Regularized Recon (“Q.Clear”)
4min 2min 1min 30s
![Page 53: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/53.jpg)
Image Quantitation
![Page 54: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/54.jpg)
What is Image Quantitation?
• Generally - Deriving numbers from images
• Volume measurement
• Motion measurements (e.g., ejection fraction)
• Distributions of radiotracers, and, under the right circumstances, the underlying physiological processes.
![Page 55: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/55.jpg)
What Factors Affect Quantitation of Radionuclide Distributions?
• Successful calibration of scanning system (counts/s to activity)
• Accurate corrections for
– attenuation
– scatter
– randoms
– dead time (the bigger the effect, the more accurate the correction must be)
• Quantitative reconstruction algorithm
• Resolution effects (degradation of small structures)
• ROI (Region Of Interest) Analysis
![Page 56: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/56.jpg)
Standardized Uptake Value (SUV)
• The SUV radioactivity concentration (what the scanner measures) normalized to injected dose and body mass:
• This can also be thought of as the local concentration divided by the body mean concentration
• Dimensions are mass/volume (e.g. g/ml). Since this is the body (mostly water; 1ml=1g), it is almost dimensionless.
• This is sometimes referred to as a semi-quantitative measure (compared to kinetic modeling.)
massbodydoseinjected
ionconcentratityradioactivSUV
/
![Page 57: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/57.jpg)
Sphere Size vs. Spatial Resolution Sphere diameters:
“1”, 2, 3, 4, 5, 6, 7, 8 pixels
no smoothing
2 pixel FWHM
3D smoothing
3 pixel FWHM
3D smoothing
“Full Recovery”:
When at least some pixels in a region
retain full intensity
“Recovery coefficient”:
Ratio between measured and actual
values (due to resolution effects only)
RC 1
Sphere diameter 3X FWHM
resolution full recovery (for 3D
blurring)
(Not so bad if non-spherical, or non-
3D blurring
![Page 58: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/58.jpg)
3D Sensitivity vs. axial FOV
Sens (FOV)2
![Page 59: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/59.jpg)
GE Discovery IQ
![Page 60: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/60.jpg)
Solid State Photodetectors
• Avalanche Photodiodes – Compact, work in B field
– Siemens PET/MR
• SSPM, SiPM, etc. – Compact, work in B field, excellent timing
– GE PET/MR, Philips PET/CT (Vereos)
![Page 61: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/61.jpg)
Image Quality vs. Size
Patient Size
Imag
e Q
ual
ity
![Page 62: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/62.jpg)
NEMA NU-2 Performance Tests
• Spatial Resolution
• Sensitivity
• Count Rate and Scatter Fraction
• Image Quality
![Page 63: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/63.jpg)
PET Performance - Sensitivity
Five Concentric 70 cm
long aluminum tubes
surrounding source
Scanner Axis
Why long line source? Whole body sensitivity.
![Page 64: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/64.jpg)
Sensitivity - 3D, r=0
![Page 65: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/65.jpg)
Scatter Fraction and Count Rate
![Page 66: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/66.jpg)
Scatter Fraction
• S/(S+T) --- low is good
• reflects energy resolution and geometry
• Method: – Axial low-intensity line source in phantom
– Mask out LORs not subtending phantom
– Line source makes sinusoid in each slice’s sinogram
– Pixels off the sinusoid measure background (scatter and random)
– At very low rates, no randoms.
![Page 67: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/67.jpg)
Spatial Resolution
• Small (< 1 mm) point sources placed in several locations within FOV
• Scan, reconstruct with very small (< 1/10 expected FWHM) pixels.
• Measure FWHM of resulting profiles in all three directions
Generally, pixels should be 1/3 the expected FWHM or smaller, for any
NM application. Combination of recon FOV and image matrix.
![Page 68: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/68.jpg)
Spatial Resolution
![Page 69: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/69.jpg)
Which count-based PET metric is the best
predictor of image quality?
10%
71%
0%
19%
0% 1. Total counts (prompts)
2. True counts
3. Random counts
4. Noise-equivalent counts (NEC)
5. Scatter fraction
![Page 70: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/70.jpg)
4. Noise-equivalent counts (NEC)
Which count-based PET metric is the best
predictor of image quality?
Cherry, Sorenson, and Phelps “Physics in Nuclear Medicine”
![Page 71: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/71.jpg)
Why is photon attenuation a large effect in
PET (compared to SPECT)?
14%
0%
5%
81%
0% 1. There’s no way to correct it.
2. Both photons must be detected.
3. The photon energy is high.
4. The photon energy is low.
5. Timing resolution is imperfect.
![Page 72: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/72.jpg)
2. Both photons must be detected
Why is photon attenuation a large effect in
PET (compared to SPECT)??
Cherry, Sorenson, and Phelps “Physics in Nuclear Medicine”
![Page 73: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/73.jpg)
What factor is essential to localizing the PET
annihilation locations?
5%
18%
64%
0%
14% 1. The 511 keV energy.
2. The F-18 half-life.
3. The anti-parallel paths.
4. The scanner sensitivity.
5. The electron mass.
![Page 74: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/74.jpg)
3. The anti-parallel paths
What factor is essential to localizing the PET
annihilation locations?
Cherry, Sorenson, and Phelps “Physics in Nuclear Medicine”
![Page 75: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/75.jpg)
What effect does increasing iterations in ML-
EM image reconstruction have?
19%
76%
5%
0%
0% 1. Increased patient comfort.
2. Less CPU time.
3. Lower resolution.
4. Increased noise.
5. Shorter scan time.
![Page 76: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/76.jpg)
4. Increased noise.
What effect does increasing iterations in ML-
EM image reconstruction have?
Cherry, Sorenson, and Phelps “Physics in Nuclear Medicine”
![Page 77: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/77.jpg)
Which of the following is directly measured
in current NEMA NU-2 PET tests?
0%
5%
0%
14%
82% 1. Spatial resolution.
2. Timing resolution.
3. Energy resolution.
4. PET-CT alignment accuracy.
5. Image reconstruction speed.
![Page 78: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/78.jpg)
1. Spatial Resolution.
Which of the following is directly measured
in current NEMA NU-2 PET tests?
“Performance Measurements of Positron Emission Tomographs
(PETs)”, NEMA-NU2 2012, National Electrical Manufacturer’s
Association
![Page 79: Radiology and Medical Physics Duke University …amos3.aapm.org/abstracts/pdf/99-30891-359478-114804.pdfRadiology and Medical Physics Duke University Durham, North Carolina, USA Chart](https://reader033.vdocuments.us/reader033/viewer/2022060207/5f03ea057e708231d40b6483/html5/thumbnails/79.jpg)
Thank you.