radiation detection and measurement - university of …depts.washington.edu/imreslab/2011...

46
Fall 2011, Copyright UW Imaging Research Laboratory Radiation Detection Tom Lewellen, PhD [email protected] Nuclear Medicine Basic Science Lectures http://www.rad.washington.edu/research/our-research/groups/irl/education/ basic-science-resident-lectures September 2011 Tuesday, September 27, 11

Upload: dangkiet

Post on 22-Jul-2018

225 views

Category:

Documents


0 download

TRANSCRIPT

Fall 2011, Copyright UW Imaging Research Laboratory

Radiation Detection

Tom Lewellen, PhD

[email protected] Medicine Basic Science Lectures

http://www.rad.washington.edu/research/our-research/groups/irl/education/basic-science-resident-lectures

September 2011

Tuesday, September 27, 11

Fall 2011, Copyright UW Imaging Research Laboratory

Main points of last week’s lecture:

Charged particles• Short ranged in tissue

~ mm for betas (predictable, continuously slowing path)~ µm for alphas (more sporadic path)

• Interactions Types: Excitation, Ionization, Bremsstrahlung• Linear energy transfer (LET) - nearly continual energy transfer• Bragg ionization peak (LET peaks as particle slows down)

Photons• Relatively long ranged (~cm)• Local energy deposition - photon deposits much or all of their energy each interaction• Interactions Types: Rayleigh, Photoelectric, Compton, Pair Production• Compton - dominant process in tissue-equivalent materials for Nuc. Med. energies• Beam hardening - polychromatic photon beam• Buildup factors - narrow vs. wide beam attenuation• Secondary ionization - useful for photon detection

Now we shall discuss …How interaction of radiation can lead to detection

Tuesday, September 27, 11

Types of radiation relevant to Nuclear Medicine

Particle! Symbol! Mass (MeV/c2) ! Charge

Electron! e-, β -! ! 0.511 !! ! -1

Positron! e+, β+!! 0.511 !! ! +1

Alpha!! α! ! 3700 ! ! ! +2

Photon! γ! no rest mass! ! none

Tuesday, September 27, 11

• Loses energy in a more or less continuous slowing down process as it travels through matter.

• The distance it travels (range) depend only upon its initial energy and its average energy loss rate in the medium.

• The range for an α particle emitted in tissue is on the order of µm’s.

α Particle Range in Mattermono-energetic

- - - - - - - - - - - - - - - - - - -+ + + + + + + + + + + + + + + + +

µm’s

α

Tuesday, September 27, 11

• β particle ranges vary from one electron to the next, even for βs of the same energy in the same material.

• This is due to different types of scattering events the β encounters (i.e., scattering events, bremsstrahlung-producing collisions, etc.).

• The β range is often given as the maximum distance the most energetic β can travel in the medium.

• The range for β particles emitted in tissue is on the order of mm’s.

β Particle Range in Mattercontinuous energy spectrum

mm’s-

β±

Tuesday, September 27, 11

Interactions of Photons with MatterExponential Penetration: N=N0e-λx

Photoelectric effect! photon is absorbed!Compton scattering! part of the energy of the photon is absorbed! scattered photon continues on with lower energy

Pair production! positron-electron pair is created! requires photons above 1.022 MeV!

Coherent (Rayleigh) scattering! photon deflected with very little energy loss! only significant at low photon energies (<50 keV)

λ

cm’s

NN0

x

Tuesday, September 27, 11

Basic Radiation Detector System

Pulse or

Current storedto disk

incoming radiation

Analog-to-

digitalAmplify

& condition

Tuesday, September 27, 11

Basic Radiation Detector SystemsWhat do you want to know about the radiation?

Energy?Position (where did it come from)?How many / how much?

Important properties of radiation detectors

(depends on application)Energy resolutionSpatial resolutionSensitivityCounting Speed

Tuesday, September 27, 11

Pulse Mode versus Current Mode

• Pulse mode– Detect individual photons– Required for NM imaging applications

• Current mode– Measures average rates of photon flux– Avoids dead-time losses

Tuesday, September 27, 11

Types of Radiation Detectorsdetection modes / functionality

• Counters– Number of interactions– Pulse mode

• Spectrometers– Number and energy of interactions– Pulse mode

• Dosimeters– Net amount of energy deposited– Current mode

• Imaging Systems– CT = current mode– NM = pulse mode

Tuesday, September 27, 11

Types of Radiation Detectorsphysical composition

• Gas-filled detectors

• Solid-state (semiconductor) detectors

• Organic scintillators (liquid & plastic)

• Inorganic scintillators

scintillators operate with a photo-sensor

(i.e. another detector)Tuesday, September 27, 11

Gas-filled Detectors

Ionizing event in airrequires about 34 eV

From: Physics in Nuclear Medicine (Sorenson and Phelps)

Tuesday, September 27, 11

Gas-filled detectors(operates in three ranges)

Geiger-Muller counters

Proportional counters

Ionization chambers– Radiation survey meters– Dosimeters (dose calibrator)

From: Radiation Detection and Measurement (Knoll, GF)

Tuesday, September 27, 11

Ionization Chambers

From: Physics in Nuclear Medicine (Sorenson and Phelps)

ATOMLAB 200 Dose Calibrator

No amplificationNo dead-timeSignal = liberated chargeSettings for different isotopesCalibrations

Ionization chamber region

Tuesday, September 27, 11

Geiger-Muller counters

From: Physics in Nuclear Medicine (Sorenson and Phelps)

No energy infoLong dead-timeThin window probe

Tuesday, September 27, 11

Dose calibratorFrom: The Essential Physics of Medical Imaging (Bushberg, et al)

A) Film packB) Black (opaque) envelopeC) FilmD) Plastic film badgeF) Teflon filterG) Lead filterH) Copper filterI) Aluminum filterJ) “Open window”

Dosimeter - Film Badge

Fall 2011, Copyright UW Imaging Research Laboratory

Tuesday, September 27, 11

Dose calibratorFrom: The Essential Physics of Medical Imaging (Bushberg, et al)

Pocket Dosimeter

Fall 2011, Copyright UW Imaging Research Laboratory

Tuesday, September 27, 11

Semiconductor Detectors

• Works on same principle as gas-filled detectors (i.e., production of electron-hole pairs in semiconductor material)

• Only ~3 eV required for ionization (~34 eV, air)• Usually needs to be cooled (thermal noise)• Usually requires very high purity materials or

introduction of “compensating” impurities that donate electrons to fill electron traps caused by other impurities

Tuesday, September 27, 11

Semiconductor Detectors

• CdZnTe detectors - can operate at room temperature - starting to show up in some dedicated cardiac gamma cameras

Tuesday, September 27, 11

Organic Liquid Scintillators(liquid scintillator cocktail)

• Organic solvent - must dissolve scintillator material and radioactive sample

• Primary scintillator (p-terphenyl and PPO)• Secondary solute (wave-shifter)• Additives (e.g., solubilizers)• Effective for measuring beta particles (e.g., H-3, C-14).

Tuesday, September 27, 11

Inorganic Scintillators(physical characteristics)

Absorption of radiation lifts electrons from valence to conduction band

Impurities (activators) create energy levels within the band gap permitting visible light scintillations

Tuesday, September 27, 11

Inorganic Scintillators(physical characteristics)

NaI(Tl) BGO LSO(Ce) GSO(Ce)

Density (gm/cm3) 3.67 7.13 7.4 6.71

EffectiveAtomic Number 51 75 66 59

AttenuationCoefficient(@ 511 keV, cm-1 ) 0.34 0.955 0.833 0.674

Light Output(photons/Mev) 40K ~8K ~30K ~20K

Decay Time 230 ns 300 ns 12 ns 60 ns40 ns

Wavelength 410 nm 480 nm 420 nm 430 nm

Index of Refraction 1.85 2.15 1.82 1.85

Hygroscopy yes no no no

Rugged no yes yes no

sensitivity

energy & spatial resol.counting speed

photo-sensor matchingmanufacturing / cost

relevant detector property

Tuesday, September 27, 11

Photomultiplier Tube (PMT) - most common photo-sensor currently in use for Nuclear medicine

From: Physics in Nuclear Medicine (Sorenson and Phelps)

photo-sensor needed with scintillators

Tuesday, September 27, 11

Sample Spectroscopy SystemHardware

From: The Essential Physics of Medical Imaging (Bushberg, et al)

incoming high-energy gamma ray

converted to 1000s of visible photons

~20% converted to electrons

electron multiplicationbecomes electric signal larger current or

voltage

more electrons

more scintillation photons

higher gamma energy deposited in

crystal

Tuesday, September 27, 11

Silicon Photomultipliers(Geiger-mode APDs)

Fall 2011, Copyright UW Imaging Research Laboratory

Tuesday, September 27, 11

GM-APDs Arrays at the UW

Zecotek PhotonicsType 3-N 8x8 array with 3.3 mm square elements

SensL 4x4 array with 3 mm square elements

Fall 2011, Copyright UW Imaging Research Laboratory

Tuesday, September 27, 11

Interactions of Photons with a Spectrometer

A. PhotoelectricB. Compton + PhotoelectricC. ComptonD. Photoelectric with characteristic

x-ray escapeE. Compton scattered photon from

lead shieldF. Characteristic x-ray from lead

shield

From: The Essential Physics of Medical Imaging (Bushberg, et al)

Tuesday, September 27, 11

Sample Spectroscopy SystemOutput

From: Physics in Nuclear Medicine (Sorenson and Phelps)

From: The Essential Physics of Medical Imaging (Bushberg, et al)

counting mode

Ideal Energy Spectrum

Remember - you are looking at the energy deposited in the detector!

Tuesday, September 27, 11

Energy Resolution

From: Physics in Nuclear Medicine (Sorenson and Phelps)

Realistic Energy Spectrum

Tuesday, September 27, 11

Sample Spectrum (Cs-137)

A. PhotopeakB. Compton continuumC. Compton edge

D.! Backscatter peakE.! Barium x-ray photopeakF.! Lead x-rays

Detection efficiency (32 keV vs. 662 keV)

From: The Essential Physics of Medical Imaging (Bushberg, et al)

Tuesday, September 27, 11

Sample Spectrum (Tc-99m)

A. PhotopeakB. Photoelectric with

iodine K-shell x-ray escape

C. Absorption of lead x-rays from shield

From: The Essential Physics of Medical Imaging (Bushberg, et al)

Tuesday, September 27, 11

Sample Spectrum (In-111)

source detector

From: Physics in Nuclear Medicine (Sorenson and Phelps)

Tuesday, September 27, 11

Effects of Pulse Pileup (count rate)

From: Physics in Nuclear Medicine (Sorenson and Phelps)

Tuesday, September 27, 11

Interaction Rate and Dead-time

paralyzable non-paralyzable

From: The Essential Physics of Medical Imaging (Bushberg, et al)

True rate

Mea

sure

d ra

te

time

deadtime

= recorded events

Tuesday, September 27, 11

Calibrations• Energy calibration (imaging systems/spectroscopy)

– Adjust energy windows around a known photopeak– Often done with long-lives isotopes for convenience ! Cs-137:

Eγ= 662 keV (close to PET 511 keV), T1/2=30yr! Co-57: Eγ= 122 keV (close to Tc99m 140 keV ), T1/2=272d

• Dose calibration (dose calibrator)– Measure activity of know reference samples (e.g., Cs-137 and

Co-57)– Linearity measured by repeated measurements of a decaying

source (e.g., Tc-99m)

Tuesday, September 27, 11

Raphex Question

D58. The window setting used for Tc-99m is set with the center at 140 keV with a width of +/-14 keV i.e., 20%. The reason for this is: A. The energy spread is a consequence of the statistical broadening when

amplifying the initial energy deposition event in the NaI(Tl) crystal. B. The 140 keV gamma ray emission of Tc-99m is not truly monoenergetic but

the center of a spectrum of emissions. C. The higher and lower Gaussian tails are a consequence of compton scattering

within the patient. D. The result of additional scattered photons generated in the collimator. E. A consequence of patient motion during scanning.

Tuesday, September 27, 11

Raphex Answer

D58. The window setting used for Tc-99m is set with the center at 140 keV with a width of +/-14 keV i.e., 20%. The reason for this is: A . Photons, which impinge upon the crystal, lose energy by Compton

scattering and the photoelectric effect. Both processes convert the gamma ray energy into electron energy. On average approximately one electron hole pair is produced per 30 eV of g amma ray e nergy deposited in the crystal. These electrons result in the release of visible ligh t when trapped in the crystal. These light quanta are collected and amplified by photomultiplier tubes. The statistical fluctuation in the number of light quanta collected and their amplification is what causes the spread in the detected energy peak, even when most of the Tc-99m photons deposit exactly 140 keV in the NaI(Tl) crystal.

Tuesday, September 27, 11

The count rate for a 1 µCi source is measured as 25 kcps by a well counter. Assuming no corrections are applied, the measured count rate for a 10 µCi source will be:

a. 250 kcpsb. Less than 250 kcpsc. Greater than 250 kcps

Because of deadtime effects

Question

Fall 2011, Copyright UW Imaging Research Laboratory

Tuesday, September 27, 11

How many peaks would you expect for a99m-Tc sample placed outside a well counter?

1

1

At higher doses you will get distortion of the photopeak and a high end tail on the energy spectra due to pileup. See slide 31 from lecture.

What about inside a well counter?

Is your answer dose dependent?

Question

Fall 2011, Copyright UW Imaging Research Laboratory

Tuesday, September 27, 11

How many peaks would you expect for a 68-Ge sample placed outside a well counter? What about inside a well counter?

Outside, 1Inside, 2

68-Ge is a positron emitter (e.g., PET)

Question

Fall 2011, Copyright UW Imaging Research Laboratory

Tuesday, September 27, 11

Of the following, the most efficient detector for x-rays is:

From: The Essential Physics of Medical Imaging (Bushberg, et al)

NaI(Tl) is an inorganic scintillator and is much more efficient at detecting x-rays

than gas filled detectors.

a. Geiger counterb. NaI(Tl) detectorc. Single channel analyzerd. Ionization chambere. Pocket (self-reading) dosimeter

Question

Fall 2011, Copyright UW Imaging Research Laboratory

Tuesday, September 27, 11

Gas multiplication occurs in:

From: The Essential Physics of Medical Imaging (Bushberg, et al)

a. Geiger-Mueller countersb. Scintillation detectorsc. Semiconductor detectorsd. Ionization chamberse. Dose calibrators

Question

Fall 2011, Copyright UW Imaging Research Laboratory

Tuesday, September 27, 11

In a pho tomu l t i p l i e r t ube , t he photocathode is at a positive voltage with respect to the first dynode.

From: The Essential Physics of Medical Imaging (Bushberg, et al)

False

False

Small changes to the voltage applied to an ionization chamber have a large effect upon the charge collected from each interaction with ionizing radiation.

Question (True or False)

Fall 2011, Copyright UW Imaging Research Laboratory

Tuesday, September 27, 11

A 1 MeV beta particle produces a pulse of the same amplitude in a G-M detector as a 200 keV beta particle.

From: The Essential Physics of Medical Imaging (Bushberg, et al)

True

Question (True or False)

Fall 2011, Copyright UW Imaging Research Laboratory

Tuesday, September 27, 11

Which detector system is most appropriate and accurate for the measurement of a pure beta source:

From: Raphex

a. Ionization chamberb. Geiger Muller tubec. NaI(Tl) well scintillation counterd. Thermoluminescent dosimetere. Liquid scintillation counter

Question

Fall 2011, Copyright UW Imaging Research Laboratory

Tuesday, September 27, 11

a. Identify the energy of a radionuclideb. Reject Compton scattered photonsc. Separate a mixture of radionuclidesd. Alter the sensitivity or resolution of the

systeme. All of the above

From: Raphex

A pulse height analyzer (PHA) window can be used to:

Question

Fall 2011, Copyright UW Imaging Research Laboratory

Tuesday, September 27, 11