quality in transfusion transmitted infections (tti) testing

43
Quality in Transfusion Quality in Transfusion Transmitted Infections Transmitted Infections (TTI) Testing (TTI) Testing

Upload: santino-boardley

Post on 15-Dec-2015

232 views

Category:

Documents


3 download

TRANSCRIPT

Quality in Transfusion Transmitted Quality in Transfusion Transmitted Infections (TTI) Testing Infections (TTI) Testing

Teaching Aim

To familiarize participants with essential quality elements that govern TTI tests such as validation of test run, types of control used and maintenance of equipments.

Why we require Quality in TTIs testing

Blood is a source of TTIs (14 new viruses in the past 28 years- one new virus every two years!! )

Quality is Essential to prevent TTIsLack of quality can result either in TTI or

wastage of bloodTTI testing is dependent on number of

variables which need to be controlled

Essential elements governing quality in TTI

Quality of the specimen used for testingQuality of kits used for testingQuality & Calibration of equipment used Use of SOPs for testingType of Controls used while testing Interpretation of resultsValidation of resultsRecord keepingTraining of staff in SOPs

The quality of specimen used for testing

Sample should be properly labeledSample should be clear and sterileLipemic, hemolysed and contaminated

specimens do not yield reliable resultsBio-safety measures are very crucial to prevent

laboratory infections - handling and disposal.Avoid adding preservatives - some interfere

with test results

Quality of kits used for testing

In general kits with highest sensitivity and specificity should be used in a BTS for TTI testing

All kits and reagents should be used within the expiration date

kits which are used should have approval of certifying authority (DCGI,NACO)

Never interchange reagents from one kit to another or one lot to another

Quality and calibration of equipments used

Always use standard equipment in a BTS for TTI testing - ELISA reader & washer, Micro-pipettes, incubators, shakers etc.

Periodic calibration of equipment is vital to maintain quality -

Periodic servicing of equipment is crucial for optimal use of equipment- ELISA washers

Proper documentation on equipment check and its performance is essential- maintain records

Controls used in the assay for testing

Internal kit controls : Include the positive control, Negative control. At times may include a calibrator provided by the manufacturer

External controls: Include positive samples from the laboratory either pooled or single, diluted or undiluted. Essential to incorporate this to monitor quality in testing procedures

Intra-run and Inter-run reproducibility : three slots/run and on three consecutive days

ValidationValidation

It is a way of assuring that a system, process or

equipment is performing the way it is supposed to do so.

Validation toolsValidation tools

• Include positive and negative controls in every test run

• Include additional validation measures where possible e.g.

- rapid test with internal control- use of mechanical readers if available: to reduce subjectivity

Interpretation of test Interpretation of test resultsresults

As rule all readings (both quantitative and qualitative) and calculations should be checked by two individuals

Validation of every run is essential for proper interpretation of results

Proper records - print outs of results, calculations of cut off values, graphs, etc. should be maintained

Any errors detected should be brought to the notice of the concerned staff and corrective measures instituted promptly

ControlsControls

Internal controls Set of controls (Positive & Negative)

provided along with the kit

To be used only in those batches of kit from which they originate

Do not detect minor deterioration of kits

Controls (contd..)External controls

Set of controls included from outside Positive (Borderline Reactive) & Negative Detect minor error in the assay performance

Sources of External ControlsSources of External Controls

- National reference laboratories

- Commercial control panels

- In-house prepared external controls

- Pooled test kit controls

- Samples collected from other laboratories

Preparation of In-house External Controls

Select sero-positive serum/plasma

Retest the sample with another kit

Heat inactivate the sample @ 560C X 30

minIf plasma taken, re-calcify it to obtain

serumMake serial dilution of the sample with a

sero-negative serum

NACO Guidelines

Making Suitable Dilutions

100 ul serumin tube

100ul diluent in each tube

Mix and Transfer

Each tube is a 1:2 dilution

of the previous tube

Discard

200 ul serumin tube 1

Preparation of In-house External Controls (contd…)

Select the dilution with ER b/w 1.5 to 2

Prepare external control aliquots of dilution selected above

Store at – 200C or below @ 1 year

After test run, calculate ER for each dilution ELISA ratio = sample OD / cut off OD

Once thawed, control can be kept @ 2-80C, 1 wk

0.00

1.00

2.00

3.00

4.00

5.00

6.00

7.00

8.00

U.N

o.9

607

1:0

2

1:0

4

1:0

8

1:1

6

1:3

2

1:6

4

1:1

28

1:2

56

1:5

12

1:1

02

4

1:2

04

8

1:4

09

6

1:8

19

2

1:1

63

84

1:3

27

68

E. R

AT

IO

DILUTION

E.RATIO

oPOS. CONTROL

CUT OFFE.P.C.

Need for E ratioE Ratio =Cut off values differ depending on the

principle of the test , manufacturer guidelines and recommended protocol for the calculation.

Some degree of variation in internal controls results in the variation in the cut off values due to ◦Variation in incubation condition◦Preparation of reagents◦Plate to plate and well to well variation (antigen

coating)

Sample OD Cut off OD

• OD of the controls would expectedly influence the OD value of the test samples in similar directions.

• However, the relative reactivity of the given sample and the cut off would not vary.

Preparation of QC charts/Levy Jennings charts

Include at least 30 runs on the same graph Mean and ± 2SD is calculated plotted on

the graph.E ratios are plotted on the Y axis in chart

and consecutive dates of runs are plotted on X axis.

Change of operator and batch of assay should be recorded

Levy-J ennings Control Chart

MeanMean+2SD+2SD

MeanMean

Outlier

Outlier

Mean-2SD

Outlier

(New operator)

E Ratio of External control sampleobtained on each day of testing

Mean

+2SD

-2SD

Calculation of Cut off

The cut off value for ELISA is calculated according to the formula given in kit insert of test.

For eg: (The mean of negative control (OD) reading at 405 nm) + 0.2

                                  = 0.005 + 0.2                                               = 0.205 

• Less than 0.205 is Negative

• More than 0.205 is Positive  

Calculation of Mean: ELISA Tests

• Collect optical density (OD) values for Controls

for each assay run.

• Collect cutoff (CO) value for each run.

• Calculate ratio of OD to CO (OD/CO) for each

• Use these ratio values to calculate the Mean, SD

and CV%

Mean is calculated as

X - ∑X Sum total of E ratio n = Number of reactivity

•Standard deviation = Each of t he individual values (E ratio) are compared with the mean (X) to find out the deviations from the mean. •If the E ration is x1 then the deviation will be X1 ~ X which is expressed as ‘d’. The deviations are then squared.

These squared deviations are added and expressed as ∑d2 or ∑(Xn - X)2 or The result is then divided by the number of readings.

•The square root of the above value is taken to find out

SD = ∑ (Xn – x )2

n

Coefficient of Variation –

- it is expressed as the percentage and the following formula is used.

SD x 100

CV (%) = Mean (X)

- CV less than 10% is considered as an indication oflittle variation

L J chart –Scope and application

Detection of the following

•Systematic variation

•Random variation

•Lot to lot variation

•Day to day variation

Applications of control charts

Highlight the outliers (values outside +/- 2 SD)

Reveal batch to batch variation

Reveal operator to operator variation

Changes in assay performance even when

test runs are valid

Systematic Variation

Trend-Results change gradually in either direction indicating slowly changing parameters-deteriorating reagents, equipment failing

Shift-Results fall sharply on one side of the mean indicating a major change has occurred

Chart showing shift and trend

Interpretation of aberrant resultsControl values of six consecutive runs fall on

one side of mean(SHIFT)◦Switching to new lot of kits◦New reagents◦Changes in incubation temperature◦New technical hand

Six consecutive points distributed in on general direction (TREND)◦Deterioration of reagents◦Slowly faltering equipment.

Random Variation

Observance of one result significantly different from other results without any pattern

Causeso Transcription errorso Sample mix-upoPoor pipette precisionoPoor mixing of samplesoReader not calibratedoWashing inconsistent

Common source of errors in TTI testing

Transcription errors- mislabeling, data entryErrors in addition of samples to the plate -

interchanging specimensLack of equipment maintenance & calibration -

Fungus on filters, volumesPoor techniqueSudden change of staff / kit used in the BTSLack of periodic training and updating of SOPs/

manuals

Error management in TTI testingTranscription errors- mislabeling, data entry -

VigilanceErrors in addition of samples to the plate -

interchanging specimens - Vacant slot in rackLack of equipment maintenance & calibration -

Fungus on filters, volumes - Fixing responsibility & supervision

Poor technique - Training of staff, SupervisionSudden change of staff / kit used in the BTS - Training Lack of periodic training and updating of SOPs/

manuals - Vigilance and Supervision

ELISA Reader

Quality Control of Equipment

ELISA Reader….contd.Photometric instrument filter should be protected from moisture and

fungal growthkeep silica gel packs in the filter boxCalibration is done every six months (supplier) OD of special plates & standard colour solution

are recordedResults of OD should be within 10% of expected Daily check – negative & positive controls added

to each run

ELISA Washer

After Use-Fill the rinse bottle with about 500 ml of distilled water.Dispose off the unused wash bufffer. Rinse with distilled water, a couple of times and leave about 500 ml in the wash bottle.Fix the cap tightly.

Water baths & IncubatorsDaily recording of temp. using a calibrated

thermometerAcceptable results are the expected temp.

± a narrow range (± 0.5°C) predetermined by the laboratory

Pipette

Maintenance of Pipette Do not dispense volatile / corrosive materials as

it will disturb the vaccumVolume should be increased or decreased

graduallyPipette must always be returned to zero position

after useAlways store pipettes in vertical or erect position

in a holderPipette snout must cleaned regularly with moist

filter paper after use

Calibration of PipetteAll items at ambient room temp. Record the weight of empty beaker Record the temp. of tube with distilled waterPipette a known volume of distilled water

(expected volume) Record: [wt of beaker + D.W.] – wt of empty beaker

= weight of D.W.

Calibration ….contd. weight of water

Delivered vol. = temp. factor x sp. gravity of water

Repeat this 10 times, changing the tip Calculate mean, SD and CV

expected vol. – delivered vol. x 100 % Deviation =

expected vol. % deviation < 1.5% CV < 1%

Learning outcome

• To understand the essential quality elements in TTI testing

• Interpretation of test results• Importance of validation • Maintenance of equipments in TTI testing lab