quality assurance in stereotactic quality assurance in...

12
1 Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic Radiotherapy David Shepard, Ph.D. Swedish Cancer Institute Seattle, WA Timothy Solberg, Ph.D. University of Nebraska Medical Center Omaha, NE Outline Mechanical aspects Linac Frames Beam data acquisition Commissioning of TP system End-to-end evaluation Imaging and Image Fusion Frameless Radiosurgery References and Guidelines Quality Assurance in Linac SRS/SBRT Podgorsak EB, Olivier A, Pla M, et al, Dynamic rotation stereotactic radiosurgery, Int J Radiat Oncol Biol Phys. 11: 1185-1192, 1988 Betti OO, Derechinsky VE, Hyperselective encephalic irradiation with linear accelerator, Acta Neurochir. 33: 385, 1984 Lutz W, Winston KR, Maleki N, A system for stereotactic radiosurgery with a linear accelerator, Int J Radiat Oncol Biol Phys. 14(2):373-81, 1988 Friedman WA, Bova FJ, The University of Florida radiosurgery system, Surg Neurol. 32(5):334-42, 1989 LOTS of variation in Linac SRS technologies … Siddon RL, Barth NH, Stereotaxic localization of intracranial targets, Int J Radiat Oncol Biol Phys. 1987 13(8):1241-6, 1987 … and in radiation delivery Isocentric Accuracy: The Winston-Lutz Test Isocentric Accuracy: The Winston-Lutz Test

Upload: tranmien

Post on 03-Jul-2018

222 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Quality Assurance in Stereotactic Quality Assurance in ...aapm.org/meetings/amos2/pdf/29-7875-86879-766.pdf · the first 160 IMRT patients System Commissioning End-to-end dosimetry

1

Quality Assurance in Stereotactic Radiosurgery and Fractionated Stereotactic

Radiotherapy

David Shepard, Ph.D.Swedish Cancer Institute

Seattle, WA

Timothy Solberg, Ph.D.University of Nebraska Medical Center

Omaha, NE

Outline

• Mechanical aspects• Linac• Frames

• Beam data acquisition• Commissioning of TP system• End-to-end evaluation• Imaging and Image Fusion• Frameless Radiosurgery• References and Guidelines

Quality Assurance in Linac SRS/SBRT

Podgorsak EB, Olivier A, Pla M, et al, Dynamic rotation stereotactic radiosurgery, Int J RadiatOncol Biol Phys. 11: 1185-1192, 1988

Betti OO, Derechinsky VE, Hyperselectiveencephalic irradiation with linear accelerator, Acta Neurochir. 33: 385, 1984

Lutz W, Winston KR, Maleki N, A system for stereotactic radiosurgery with a linear accelerator, Int J Radiat Oncol Biol Phys. 14(2):373-81, 1988

Friedman WA, Bova FJ, The University of Florida radiosurgery system, Surg Neurol. 32(5):334-42, 1989

LOTS of variation in Linac SRS technologies …

Siddon RL, Barth NH, Stereotaxic localization of intracranial targets, Int J Radiat Oncol BiolPhys. 1987 13(8):1241-6, 1987

… and in radiation delivery

Isocentric Accuracy: The Winston-Lutz Test

Isocentric Accuracy: The Winston-Lutz Test

Page 2: Quality Assurance in Stereotactic Quality Assurance in ...aapm.org/meetings/amos2/pdf/29-7875-86879-766.pdf · the first 160 IMRT patients System Commissioning End-to-end dosimetry

2

Isocentric Accuracy: The Winston-Lutz

Test

Is the projection of the ball centered within the

field?

Good results: x ≤ 0.5 mm

What can you expect and what can you do about it?

Before After

Maciunas et al, Neurosurgery 35:682-695, 1995

Localization Accuracy of Stereotactic Frames For SRT, it is essential to ensure daily reproducibility of frames

Kooy et al, IJROBP 30:685-691, 1994

σ = 0.4 mm (n=2000), but “The frame is not appropriate for single fraction radiosurgery, as a large setup error (> 2 mm) for a single treatment cannot be excluded.

End-to-End Localization Accuracy

Embed a hidden target Perform a CT scanFix the phantom in a frame

Create a treatment plan

End-to-End Localization Accuracy

Set up in treatment room Evaluate

Page 3: Quality Assurance in Stereotactic Quality Assurance in ...aapm.org/meetings/amos2/pdf/29-7875-86879-766.pdf · the first 160 IMRT patients System Commissioning End-to-end dosimetry

3

End-to-End Localization Accuracy(with verification of MLC shapes, orientation, etc.)

End-to-End Localization Accuracy(Surely my vendor has checked this)

Dose = TMR • OAR • Output • MU

Beam Data Acquisition and Dose Computation

Traditional radiosurgery dose algorithms are quite simple, with dose specified by the product of measured beam parameters:

Corrections for tissue heterogeneities, surface irregularities, etc. are generally neglected.

But accurate measurement of beam data is difficult

Beam Data

Measurement

Depth Dose (PDD or TMR)

Off Axis Ratios

Relative Output

Absolute Dose

Device

Diode / Ion Chamber

Diode / Ion Chamber

Diode / Ion Chamber

Ion Chamber

Beam characterization is best performed in water. Diodes are well suited to small field dosimetry though other dosimeters can certainly be used.

Dosimetry EquipmentScanning Diode

Pinpoint Chamber .015 cc

Stereotactic Diode

DIODE WARNING: Diodes exhibit enough energy dependence that ratios between large and small field measurements are inaccurate at the level required for radiosurgery

What to do?

Measure output factors ≤ 2x2 cm2 using diode≥ 2x2 and ≤ 4x4 cm2 using diode & chamber≥ 4x4 cm2 using chamber

Page 4: Quality Assurance in Stereotactic Quality Assurance in ...aapm.org/meetings/amos2/pdf/29-7875-86879-766.pdf · the first 160 IMRT patients System Commissioning End-to-end dosimetry

4

0.00

0.20

0.40

0.60

0.80

1.00

1.20

0 50 100 150 200 250 300Depth (cm)

TMR

4.0 mm5.0 mm6.0 mm7.5 mm10.0 mm12.5 mm15.0 mm

Depth Dose Data: Circular Cones

0

0.2

0.4

0.6

0.8

1

1.2

0 2 4 6 8 10 12 14Distance (mm)

Off

Axi

s R

atio

4.0 mm5.0 mm6.0 mm7.5 mm10.0 mm12.5 mm15.0 mm

Profiles: Circular Cones

0

10

20

30

40

50

60

70

80

90

100

110

0 50 100 150 200 250 300 350 400 450

Depth (mm)

Perc

ent D

epth

Dos

e

6x612x1218x1824x2430x3036x3642x4260x6080x8098x98

Depth Dose Data: µMLC Penumbra: Cones versus µMLC

0

0.5

1

1.5

2

2.5

3

3.5

0 20 40 60 80 100

Nominal Field Size (mm)

Penu

mbr

a, 8

0%-2

0% (m

m)

ConesMMLC

Radiosurgery beams exhibit a sharp decrease in output with decreasing field size

This means that with small collimators, treatment times can be long

Output Factors: Circular Collimators

0.00

0.20

0.40

0.60

0.80

1.00

1.20

0.0 20.0 40.0 60.0 80.0 100.0Field Diameter (mm)

Rel

ativ

e O

utpu

t Fac

tor

0.50

0.60

0.70

0.80

0.90

1.00

0.0 5.0 10.0 15.0 20.0Field Diameter (mm)

Rel

ativ

e O

utpu

t Fac

tor

Page 5: Quality Assurance in Stereotactic Quality Assurance in ...aapm.org/meetings/amos2/pdf/29-7875-86879-766.pdf · the first 160 IMRT patients System Commissioning End-to-end dosimetry

5

0

0.2

0.4

0.6

0.8

1

1.2

0 20 40 60 80 100 120Field Size (mm)

Scat

ter F

acto

r

Scatter Factors: µMLC

Surveyed beam data from 40 identical Linac SRS Units:

Percent Depth Dose

Relative Scatter Factors

Absolute Dose-to-Monitor Unit CF

Reference Condition

Applied statistical methods to compare data

Seems Simple Enough ….

0

10

20

30

40

50

60

70

80

90

100

110

0 50 100 150 200 250 300 350 400 450Depth (mm)

Per

cent

Dep

th D

ose

Institution A

0

10

20

30

40

50

60

70

80

90

100

110

0 50 100 150 200 250 300 350 400 450Depth (mm)

Per

cent

Dep

th D

ose

Institution AInstitution BInstitution C

Observations of some treatment units: 6 mm x 6 mm MLC

0

0.2

0.4

0.6

0.8

1

1.2

0 50 100 150 200 250 300

Depth (mm)

Institution A

0

0.2

0.4

0.6

0.8

1

1.2

0 50 100 150 200 250 300

Depth (mm)

Institution A

Institution B

Observations of some treatment units: 5 mm collimator

0

0.2

0.4

0.6

0.8

1

1.2

0 50 100 150 200 250 300

Depth (mm)

Tiss

ue M

axim

um R

atio

Institution A

0

0.2

0.4

0.6

0.8

1

1.2

0 50 100 150 200 250 300

Depth (mm)

Institution A

Institution B

Observations of some treatment units: 15 mm collimator

0

10

20

30

40

50

60

70

80

90

100

110

0 50 100 150 200 250 300 350 400 450Depth (mm)

Perc

ent D

epth

Dos

e

Institution AInstitution B

6.0 %

10.3 %

Page 6: Quality Assurance in Stereotactic Quality Assurance in ...aapm.org/meetings/amos2/pdf/29-7875-86879-766.pdf · the first 160 IMRT patients System Commissioning End-to-end dosimetry

6

Relative Output Factor: 6 mm x 6 mm MLC

Field Size (mm)

Ou

tpu

t Fact

or

~45%

Relative Output Factor: 42 mm x 42 mm MLC

Field Size (mm)

Ou

tpu

t Fact

or

~6%

Commissioning your system: Does calculation agree with measurement?

Calculation arc-step = 10o

Directly verify the delivered distribution

Calculation arc-step = 2o

80, 50, 20% isodose lines of calculation and measurement

Relative Dosimetry for individual fieldsRelative Dosimetry for composite plan

80, 50, 20% isodose lines of calculation and measurement

Page 7: Quality Assurance in Stereotactic Quality Assurance in ...aapm.org/meetings/amos2/pdf/29-7875-86879-766.pdf · the first 160 IMRT patients System Commissioning End-to-end dosimetry

7

Use an appropriate phantom Absolute dosimetry

x = 0.26 1σ = 1.75

Difference between calculation and measurement: the first 160 IMRT patients System Commissioning

End-to-end dosimetry is essential

Directly verify the delivered dose

What about independent verification of monitor units?

Circular Collimators

Micro-MLC

Dynamic Conformal Arc

Page 8: Quality Assurance in Stereotactic Quality Assurance in ...aapm.org/meetings/amos2/pdf/29-7875-86879-766.pdf · the first 160 IMRT patients System Commissioning End-to-end dosimetry

8

Imaging: We need MR

“MRI contains distortions which impede direct correlation with CT data at the level required for SRS”

Stereotactic Radiosurgery – AAPM Report No. 54

TS Sumanaweera, JR Adler, S Napel, et al., Characterization of spatial distortion in magnet resonance imaging and its implications for stereotactic surgery,” Neurosurgery 35: 696-704, 1994.

Other References

Why do we need CT?

• Use CT for geometric accuracy

• Use MR for target delineation

Y Watanabe, GM Perera, RB Mooij, “Image distortion in MRI-based polymer gel dosimetry of Gamma Knife stereotactic radiosurgery systems,” Med. Phys. 29: 797-802, 2001.

1.8 ± 0.5 mm shift of MR images relative to CT and delivered dose

Shifts occur in the frequency encoding direction

Due to susceptibility artifacts between the phantom and fiducialmarkers of the Leksell localization box

Frequency Encoding = L/R Frequency Encoding = A/P

Y Watanabe, GM Perera, RB Mooij, “Image distortion in MRI-based polymer gel dosimetry of Gamma Knife stereotactic radiosurgery systems,” Med. Phys. 29: 797-802, 2001.

What do we do about MR spatial distortion?

Image Fusion

Automatic Image Fusion for Extracranial Anatomy

DOESNT WORK!

Page 9: Quality Assurance in Stereotactic Quality Assurance in ...aapm.org/meetings/amos2/pdf/29-7875-86879-766.pdf · the first 160 IMRT patients System Commissioning End-to-end dosimetry

9

Image Fusion for Extracranial Anatomy

Other sources not evaluated:

MRI, relocatable frames, …

Stereotactic Radiosurgery, AAPM Report No. 54, 1995

So, how accurate is (frame-based) linac radiosurgery?

Can we do better?

Spatial Resolution: 0.05 mm

Temporal Resolution: 0.03 s

Localization Accuracy: 0.2 mm

Optical Photogrammetry

Stereophotogrammetry in Radiotherapy

(Active) Infrared Photogrammetry

Stereophotogrammetry in Radiotherapy

Stereophotogrammetry in Radiotherapy Frameless Radiosurgery an UNMC

Page 10: Quality Assurance in Stereotactic Quality Assurance in ...aapm.org/meetings/amos2/pdf/29-7875-86879-766.pdf · the first 160 IMRT patients System Commissioning End-to-end dosimetry

10

How do we know the system is targeting properly?End-to-end evaluation that mimics a patient procedure

Embed a hidden target Perform a CT scanFix the phantom in a frame

Create a treatment plan Reposition & Irradiate Evaluate

Set up in treatment room

X-ray

Obtain Stereo X-rays

DRR

Fuse with DRRs

Results of Phantom Data

• Sample size = 50 trials (justified to 95% confidence level, +/- 0.12mm)

0.32

-0.01

Long.

0.82

0.05

Vert.

0.420.56Standard Deviation

1.11-0.06Average

3D vectorLat.(mm)

Comparison in 45 SRS patients and 565 SRT fractions

Single Fraction

0

0.2

0.4

0.6

0.8

1

1.2

-8.0 -6.0 -4.0 -2.0 0.0 2.0 4.0 6.0 8.0

LateralSup/InfAnt/Post

Single Fraction 0.55

0.04

Long.

0.72

0.25

Vert.

0.550.66Standard Deviation

1.00-0.02Average

3D vectorLat.(mm)

Multiple Fractions

0

0.2

0.4

0.6

0.8

1

1.2

-8.0 -6.0 -4.0 -2.0 0.0 2.0 4.0 6.0 8.0

LateralSup/InfAnt/Post

1.321.032.111.24Standard Deviation

2.360.170.470.17AverageMultiple Fraction

Long. Vert. 3D vectorLat.(mm)

Page 11: Quality Assurance in Stereotactic Quality Assurance in ...aapm.org/meetings/amos2/pdf/29-7875-86879-766.pdf · the first 160 IMRT patients System Commissioning End-to-end dosimetry

11

Superior / Inferior

0

0.2

0.4

0.6

0.8

1

1.2

-8.0 -6.0 -4.0 -2.0 0.0 2.0 4.0 6.0 8.0

Multiple FractionSingle Fraction

1.321.032.111.24Standard Deviation

2.360.170.470.17AverageMultiple Fraction

Single Fraction

0.55

0.04

Long.

0.72

0.25

Vert.

0.550.66Standard Deviation

1.00-0.02Average

3D vectorLat.(mm)

Comparative Patient Shift Data for Cranial Cases

Frameless SRS on a

conventional linac

End-to-end evaluation: Extracranial

Lots of redundancy

• Patient is marked

• Immobilization device is marked

• Location of IR markers is recorded

• Table heights and shifts are provided

Extracranial Applications - Spine

Page 12: Quality Assurance in Stereotactic Quality Assurance in ...aapm.org/meetings/amos2/pdf/29-7875-86879-766.pdf · the first 160 IMRT patients System Commissioning End-to-end dosimetry

12

Extracranial Applications - Spine Extracranial Applications - Lung

Extracranial Applications - Lung Radiosurgery Guidelines

ASTRO/AANS Consensus Statement on stereotactic radiosurgery quality improvement, 1993

RTOG Radiosurgery QA Guidelines, 1993

AAPM Task Group Report 54, 1995

European Quality Assurance Program on Stereotactic Radiosurgery, 1995

DIN 6875-1 (Germany) Quality Assurance in Stereotactic Radiosurgery/Radiotherapy, 2004

AAPM Task Group 68 on Intracranial stereotactic positioning systems, 2005

ACR Practice Guidelines for the Performance of Stereotactic Radiosurgery, 2006

ACR Practice Guidelines for the Performance of Stereotactic Body Radiation Therapy, 2006