08/07/09 1
Performance Testing in Computer Tomography using
CATPHAN 600
Asim Gilani
Radiation Protection UnitMedical Physics DepartmentDubai Hospital
CATPHAN 500 & 600
CATPHAN 500 & 600
Mounting
CATPHAN 600
CATPHAN 600
CTP 404
Slice Geometry and Sensitometry Module
CTP 404
Diameter = 150 mm Thickness = 25 mm 4 wire Ramps at an angle of 23° 5 Acrylic spheres 8 sensitometry samples 4 rods
CTP 404
CTP 404
Phantom position verification Patient alignment system Circular symmetry Scan width Sensitometry
Phantom Alignment
Align phantom dots with the lasers
Phantom Alignment
Scan 6 or 8 axial slices with minimum slice widthStart 3 or 4 slices before the zero position
Phantom oriented to right side
Phantom oriented to left side
Phantom Tilted
Aligned Image
How to verify ?
Switch ON grid to check the Phantom Alignment
CUATION
Once the phantom is centered and aligned
DO NOT
touch or move the phantom by hand
Only move or align it by table movements
Test 1: LASER Accuracy
Align the phantom to lasers by moving the tableAlign all the dots to X, Y and Z LASERS as precisely as possibleScan axially with minimum slice thicknessIf the Image is same as previous one then LASERS are perfectly aligned
OTHERWISE !
Measurement of Z Axis Misalignment
Z-axis Misalignment
Measure the distance between center of the phantom and center of the slice Multiple by 0.42 to find the z-axis misalignment
4.99 x 0.42 = 2.1 mm
X and Y Misalignment
Switch ON grid and measure distance of the center of the grid to center of the image in X and Y direction and multiply by 0.42
X and Y Misalignment
X = 4.61 x 0.421.9 mm
Y = 3.26 x 0.421.4 mm
Test 2: Patient Alignment System
Take Scanogram or ScoutTake an axial Scan at the crosswireMeasure the accuracy of the scan
Test 3: Circular Symmetry
Take an axial Scan of the CTP 404Image must be circular not ellipticalMeasure diameter of the phantom in X & YMeasure distance between Teflon pinsAdditionally angle accuracy can be verified
Distance Accuracy
AngleAccuracy
Test 4: Slice Width Accuracy
Axially scan CTP 404 with different slice widthsMeasure slice with at (FWHM)M Full Width at Half MaximumSlice Thickness = FWHM * 0.42
6.00 * 0.42 = 2.52 mm
Test 5: Slice Sensitometry
Scan the phantom with different slice widthsMeasure CT Numbers of Different samples
CTP 404
Material CT #
Air
PMP
LDPE
Water
Polystyrene
Acrylic
Delrin
Teflon
- 1000
- 200
- 100
0
- 35
120
340
990
Table Movement Select axial scan with minimum slice widthMove table according to figure and scanVerify that you see the dot above
Water and kV Accuracy
Axially scan at different kV'sVerify that water CT number lies between -7 to + 7
Volume Verification
The volume of five spheres
????
CTP 591
Bead Geometry Module
CTP 591
Diameter = 150 mm Thickness = 40 mm 2 pairs of coarse ramps (39 beads each) 1 pair of precision ramp (25 Beads each) 2 Isolated beads for MTF and SSP 1 Steel wire of 50 µm diameter
Slice Thickness Verification
Make helical scan at different slice widths Verify that all slices are of same thickness
Pitch Verification
Make helical scan with different Pitches available
Verify the pitch by measuring the center to center difference in adjacent scan and subtracting by the slice width
Turn grid ON for help
MTF Measurement
The beads are for SSP and MTF measurements
We need software for these measurements
CTP 528
High Resolution Module
CTP 528
Diameter = 150 mm Thickness = 40 mm Radial gauge from 1 to 21 lines per cm 2 mm thick aluminum 2 beads of 0.28 mm 20 mm above and below
the center in Y-axis & 2.5 mm and 10 mm past the center in Z-axis
High Resolution Measurements
7 lp / cm visible
The beads are used for MTF and SSP measurements
HOW ????????????
CTP 515
Low Contrast Module
CTP 515
Diameter = 150 mm Thickness = 40 mm 39 cylindrical rods 3 contrasts, 0.3 %, 0.5 % & 1.0 % 3, 5, 7 & 40 mm length 2 to 9 & 15 mm diameter
Contrast Verification
(Measured Contrast) * (Smallest Diameter Discernible) ≅ Constant
(1.0) * (3.0) = 3.0
(0.5) * (5.0) = 2.5
(0.3) * (9.0) = 2.7
Contrast Measurements
Take ROI on 155 mm circles Note down the CT # Take ROI immediately around the circle Subtract the 2 ROI's It should be 10, 5 and 3
CTP 486
Uniformity Module
CTP 486
Diameter = 150 mm Thickness > 40 mm Uniform Module CT # between 0 to 2 % of water
Artifact Analysis
Change Window Width and level and observe any artifacts
Questions