08/22/08 1 Performance Testing and Dosimetry in Computer Tomography Asim Gilani Radiation Protection Unit Medical Physics Department Dubai Hospital

Upload: asim-gilani

Post on 01-Jun-2015

4.545 views

Category:

Technology


3 download

DESCRIPTION

Performance Testing of CT (Computer Tomography) using ACR (American College of Radiology) Phantom

TRANSCRIPT

Page 1: Qaulity Control of CT

08/22/08 1

Performance Testing and Dosimetry in Computer

Tomography

Asim Gilani

Radiation Protection Unit

Medical Physics Department

Dubai Hospital

Page 2: Qaulity Control of CT

08/22/08 2

Annual Performance Evaluation

•Alignment light accuracy

•Alignment of table to gantry

•Table/gantry tilt

•Slice localization from localization image (SPR)

•Table incrementation accuracy

•Slice thickness

•Image quality

– High contrast (spatial) resolution

– Low contrast resolution

– Image uniformity

– Noise

– Artefact evaluation

•CT number accuracy and constancy

Page 3: Qaulity Control of CT

08/22/08 3

Annual Performance Evaluation

•Display devices– Video display

– Hard-copy display

•Dosimetry-Computed Tomography dosimetry index (CTDI)

-Patient radiation dose for representative examinations

•Safety evaluation– Visual inspection

– Audible/visual signals

– Posting requirements

– Scattered radiation measurements

Page 4: Qaulity Control of CT

08/22/08 4

ACR Performance Phantom

Page 5: Qaulity Control of CT

08/22/08 5

Positioning Accuracy

CT # Accuracy

Slice Width Accuracy

Low Contrast Resolution

High Contrast (Spatial) Resolution

Image Uniformity and Noise

Image Artifacts

Distance Measurement Accuracy

The ACR CT phantom was designed to examine a broad range of scanner parameters

Page 6: Qaulity Control of CT

08/22/08 6

4 3 2 1

20 c

m

4 cm

Head Foot

AlignmentAlignment

CT # CT #

Slice widthSlice width

Low Low contrast contrast

resolutionresolution

Uniformity Uniformity & noise & noise

Distance Distance accuracy accuracy

& SSP& SSP

High High contrast contrast

resolutionresolution

Page 7: Qaulity Control of CT

08/22/08 7

•Align Module 1 to lasers

•Scan with Hi Resolution Chest protocol

Test 1: Phantom and Scanner Test 1: Phantom and Scanner AlignmentAlignment

Page 8: Qaulity Control of CT

08/22/08 8

Page 9: Qaulity Control of CT

08/22/08 9

ACR Performance Phantom

Page 10: Qaulity Control of CT

08/22/08 10

Air ≈ -1000 HU

Water ≈ 0 HU

Polyethylene ≈ -97 HU

“Bone” ≈ +910 HU

Acrylic ≈ +120 HU

MODULE 1

Page 11: Qaulity Control of CT

08/22/08 11

MODULE 1

Page 12: Qaulity Control of CT

08/22/08 12

High-resolution chest technique

Must see all four BBs

Longer wire must have same number of lines above and below (±1)

Wires are 0.5 mm apart in z-direction

WW = 1000WL = 0

Page 13: Qaulity Control of CT

08/22/08 13

•Prescribe a scan at center of Module 4 i.e., 120 mm superior

•Scan with Hi Resolution Chest protocol

Test 2: Table Distance Travel Test 2: Table Distance Travel TestTest

Page 14: Qaulity Control of CT

08/22/08 14

12

5 9

4

6

7

8

10

Bar patterns:lp/cm MODULE 4

Page 15: Qaulity Control of CT

08/22/08

DATA FORMDATA FORM

Page 16: Qaulity Control of CT

08/22/08 16

Test 3: CT Number Accuracy and Test 3: CT Number Accuracy and Slice ThicknessSlice Thickness

•Align Module 1 to lasers

•Scan with Adult Abdomen protocol & all kVp’s available

•Record all HU and slice thickness

•Also scan with 3, 5, 7 mm & hi resolution chest slice thickness

•Record water HU and slice thickness for each scan.

Page 17: Qaulity Control of CT

08/22/08 17

Adult abdomen technique

CT number measured in 5 materials

CT number of water measured at several scan widths and kVps

WW = 400WL = 0

“Bone”

Water

Poly

Acrylic Air

Page 18: Qaulity Control of CT

08/22/08 18

Adult abdomen technique

Scan width measured at several scan widths

Page 19: Qaulity Control of CT

08/22/08

DA

TA

FO

RM

DA

TA

FO

RM

Page 20: Qaulity Control of CT

08/22/08 20

Test 4: Low Contrast ResolutionTest 4: Low Contrast Resolution

• Scan Module 2 with Abdomen and Head protocols

• Record the smallest rod seen

Page 21: Qaulity Control of CT

08/22/08 21

25 mm

6 mm

5 mm

4 mm

3 mm

2 mm

Low contrast = 6 HU ± 0.5 HU

MODULE 2

Page 22: Qaulity Control of CT

08/22/08 22

Adult abdomen and head techniques

Record diameter of the smallest set of rods for which all 4 rods can be seen

ROI check of absolute contrast using large rod WW = 100

WL = 100

Page 23: Qaulity Control of CT

08/22/08

DATA FORMDATA FORM

Page 24: Qaulity Control of CT

08/22/08 24

Test 5: Uniformity, Noise, Test 5: Uniformity, Noise, Artifacts, and Distance AccuracyArtifacts, and Distance Accuracy

• Scan Module 3 with Abdomen protocol• Record HU & SD in center• Record HU at 12, 3, 6, 9 o’clock• Examine for artifacts and record• Check distance accuracy (optional)

Page 25: Qaulity Control of CT

08/22/08 25

MODULE 3

100 mm

Page 26: Qaulity Control of CT

08/22/08 26

Adult abdomen and head techniques

Measure uniformity and noise

Assess for artifacts

Measure distance accuracy(optional)

WW = 100WL = 0

Page 27: Qaulity Control of CT

08/22/08

DATA FORMDATA FORM

Page 28: Qaulity Control of CT

08/22/08 28

• Scan Module 4: – Hi Resolution Chest protocol– Abdomen protocol

• Record limiting lp/mm

Test 6: High Contrast Test 6: High Contrast ResolutionResolution

Page 29: Qaulity Control of CT

08/22/08 29

12

5 9

4

6

7

8

10

Bar patterns:lp/cm MODULE 4

Page 30: Qaulity Control of CT

08/22/08 30

Adult abdomen, adult head and high-resolution chest techniques

Record the first highest frequency bar pattern for which the bars and spaces merge

WW = 100WL 1100

Page 31: Qaulity Control of CT

08/22/08 31

WW = 400WL = 0

CT ART?

At soft-tissue window settings these streaks are visible and not of concern

Page 32: Qaulity Control of CT

08/22/08

DATA FORMDATA FORM

Page 33: Qaulity Control of CT

Questions

Page 34: Qaulity Control of CT

08/22/08 34

• ACR Manual

ReferencesReferences