Q-SUNOperation and Maintenance
1
Q-SUN Xenon Arc TestersOperation and Maintenance
Andy Francis – Marketing Director
Bill Tobin – Senior Technical Marketing Specialist
Dave Duecker – Technical Marketing Specialist
Sean Fowler – Senior Technical Director
Q-Lab CorporationClick here to view the morning presentation.
Click here to view the afternoon presentation.
Q-SUNOperation and Maintenance
2
HousekeepingYou’ll receive a follow-up email from [email protected] with links to a survey, registration for future webinars, and to download the slides
• Our ongoing webinar series can be found at: q-lab.com/webinarseries
• Our archived webinars are hosted at:q-lab.com/webinars
• Use the Q&A feature in Zoom to ask us questions today!
Q-SUNOperation and Maintenance
3
Q-SUN Xenon Test Chambers
• Simulate light (outdoor direct sunlight, indoor filtered light); heat (elevated temperature), and water (humidity and water spray)
• Used to Evaluate durability and lightfastnessof a range of materials
Q-SUNOperation and Maintenance
4
Q-SUN Xenon Arc Models
Xe-3
Xe-1
Xe-2
Q-SUNOperation and Maintenance
5
• Safety
• Features and Functions
• Running a Test
• Calibration
• Maintenance
Q-SUN Topics
Q-SUNOperation and Maintenance
6
Safety InformationXenon Lamps
• Xenon Lamps give off UV light and should not be viewed directly
• Interlocks stop tester when door opened
Q-SUNOperation and Maintenance
7
Safety InformationInterlock Switches
Xe-1 Xe-2(overhead view)
Xe-3
Q-SUNOperation and Maintenance
8
Q-SUN Safety Hazards
Electrical Shock Hot Surface
Q-SUNOperation and Maintenance
9
• Safety
• Features and Functions
• Running a Test
• Calibration
• Maintenance
Topics
Q-SUNOperation and Maintenance
10
Q-SUN Overview1) Simple user interface
2) USB port for data transfer
3) Xenon lamps with irradiance control
4) Optical filters
5) Water spray
6) Onboard irradiance sensors
7) Black Panel Temp sensor
8) Specimen holders
9) Relative Humidity/CAT sensor
Q-SUNOperation and Maintenance
11
Q-SUN Optional Features
Q-SUNOperation and Maintenance
12
Q-SUN light delivery
Xenon arc lamps
Optical filters
Xe-2 lantern Xe-1 / Xe-3 flat filter
Q-SUNOperation and Maintenance
13
Optical Filters and Lamps
Xe-1 and Xe-3
Lamp Holder
Xenon Lamp
Filter Lantern
Xe-2
Q-SUNOperation and Maintenance
14
Xenon Arc Spectrum
0.0
0.5
1.0
1.5
2.0
2.5
250 300 350 400 450 500 550 600 650 700 750 800
Spectral Irradiance
(W/m2/nm)
Wavelength (nm)
Peak Sunlight (G177)
Q-SUN Daylight-Q
Q-SUNOperation and Maintenance
15
SOLAR EYE Irradiance Control System
• SOLAR EYE Irradiance Control maintains the same light output at all times
• Ballasts control lamp output
• Allows for excellent repeatability and reproducibility
Sensor
Lamp
Ballast
Q-SUNOperation and Maintenance
16
Irradiance Control SystemOnboard Sensors
Xe-1Xe-2
(overhead)
Xe-3
Q-SUNOperation and Maintenance
17
Temperature Control SystemBlack Panel (BP or IBP) Thermometer
Xe-1Xe-3
Q-SUNOperation and Maintenance
18
Temperature Control SystemBlack Panel Thermometer and CAT/RH Sensor
Black Panel Thermometer(IBP shown)
Chamber Air Temperature / Relative Humidity Sensor
Xe-2
Q-SUNOperation and Maintenance
19
Temperature SystemChamber Air Thermometer
• Optional in Xe-1
• Tester can control temperature by BP or CAT; the other is simply monitored
Q-SUNOperation and Maintenance
20
RH/Temperature Control SystemChamber Air Thermometer / RH Sensor
• One sensor monitors chamber temperature and RH in Xe-2 and Xe-3
• Disposable; replace annually
Q-SUNOperation and Maintenance
21
Chiller (Optional, Xe-1 / Xe-3)
Reduces minimum BPT by:
~10 °C (Xe-3)
~20 °C (Xe-1)
Q-SUNOperation and Maintenance
22
Q-SUN Water Delivery
• Water Spray
– Front
– Back
– Dual (Auxiliary)
– Immersion
• Relative Humidity control (Xe-2 and Xe-3)
Q-SUNOperation and Maintenance
23
Q-SUN Water Quality
• Purified (RO/DI) water is required
• Spray water requires even higher purity and also low silica
• Q-Lab follows ASTM G151 recommendations
Water Delivery
Resistivity (Ω·cm)
Conductivity(μS/cm)
Silica(ppm)
Total Dissolved Solids (ppm)
pH
Spray > 5 M < 0.2 < 0.1 < 0.1 6-8
Humidity > 200 k < 5.0 No requirement < 2.5 6-8
Q-SUNOperation and Maintenance
24
Spray Systems (Optional)
Front
Back
Dual
Q-SUNOperation and Maintenance
25
Water Spray System
• Two nozzles, used for:
– Mist
– Thermal Shock
– Erosion
Xe-1 Xe-3
• Features
– Pulse Rate Control
– Automatic Fault Detector
Q-SUNOperation and Maintenance
26
Water Spray System
• One nozzle, used for:
– Mist
– Thermal Shock
• Features
– Pulse Rate Control
– Automatic Fault Detector
• Second Nozzle for optional Back Spray
Xe-2
Q-SUNOperation and Maintenance
27
Relative Humidity Control System
• Feedback Loop System
• RH/CAT Sensor
• Main Controller
• Humidity Generator
– Nebulizer in Xe-2
– Boiler in Xe-3
Heater
Chamber
Recirc. Air Blower
Fresh Air blower
Q-SUNOperation and Maintenance
28
Q-SUN Specimen Exposure Areas
Xe-1Xe-2
Xe-3
Q-SUNOperation and Maintenance
29
• Safety
• Functions of the Tester
• Running a Test
• Calibration
• Maintenance
Q-SUN Topics
Q-SUNOperation and Maintenance
30
Step 1: Select an Optical Filter
• Check your test method
– Almost all test methods describe the filter type
• If you don’t test to a specific method:
– Outdoor applications: Daylight - Q
– Indoor applications: Window - Q
Q-SUNOperation and Maintenance
31
Q-SUN Optical Filters
0.0
0.5
1.0
1.5
2.0
2.5
3.0
250 300 350 400 450 500 550 600 650 700 750 800
Spectral Irradiance
(W/m2/nm)
Wavelength (nm)
ASTM G177 Daylight
Daylight - Q
Extended UV - Q/B
Window - Q
ExtendedUV
Window filter
Natural sunlight
Daylight filter
Q-SUNOperation and Maintenance
32
Step 2: Select Irradiance Settings
• 340 nm Control Point
– For outdoor service environments
• 420 nm Control Point
– For indoor service environments
• TUV Control Point
– 300-400 nm (wideband, general service environments)
– Typically used for European standards
Q-SUNOperation and Maintenance
33
Select Irradiance Values
Achievable irradiance values vary by optical filter, tester type, and control point
Q-SUNOperation and Maintenance
34
Step 3: Select Black Panel Type
Panel ConstructionASTM Designation
ISO Designation
Temp Range (°C)
Black painted stainless steel
Uninsulated Black Panel
Black Panel 45-110
Black painted stainless steel mounted on0.6 cm white PVDF
InsulatedBlack Panel
Black Standard 50-120
Q-SUNOperation and Maintenance
35
Black Panel Temperature Control
• BP/IBP is typically mounted in holder (sometimes directly on tray)
• Irradiance, color, and thickness all affect specimen temperature
Q-SUNOperation and Maintenance
36
Step 4: Programming
• Light
• Dark
• Light + Spray
• Dark + Spray
• Dark + Spray Front and Back
• Dual Spray
• Light + Dual Spray
• Light + Immersion
• Dark + Immersion
Q-SUNOperation and Maintenance
37
Programming TestsLegacy Controller
Xe-3
Xe-1
Xe-2
Program key
Q-SUNOperation and Maintenance
38
Programming TestsNew Gen 4 Main Controller
Xe-3
Xe-1
Xe-2
USB/LED
Q-SUNOperation and Maintenance
39
Status and Menu Screens
Status screenSetpoint and actual controls
Test timers
Menu screenCycle management
CalibrationSettings
Q-SUNOperation and Maintenance
40
Status and Menu Screens
Status screenSetpoint and actual controls
Test timers
Menu screenCycle management
CalibrationSettings
Q-SUNOperation and Maintenance
41
Programming Tests
Q-SUNOperation and Maintenance
42
Step 5: Calibration
Q-SUNOperation and Maintenance
43
Step 6: Specimen Mounting
• Specimen Capacity– Xe-1: 17 (51 × 102 mm)
– Xe-2: 31 (45 × 132 mm)
– Xe-3: 55 (51 × 102 mm)
• Type of holder– Open-Backed (thick, rigid specimens)
– Solid-Backed (flexible specimens)
– Three-dimensional
• Specimen Tray– Solid
– Mesh (open)
• Masking– Commonly used in textile testing
Q-SUNOperation and Maintenance
44
Specimen Holders (Xe-1, Xe-3)
3D Specimen HoldersPanel Holders
Q-SUNOperation and Maintenance
45
Specimen Mounting Trays (Xe-1, Xe-3)
Solid Tray Open Mesh Tray
Q-SUNOperation and Maintenance
46
Q-SUN Xe-2 Specimen Holders (Xe-2)
Open-backed Close-backed Textile masks
Q-SUNOperation and Maintenance
47
Mounting Flexible Specimens
Textiles Thin Films
Q-SUNOperation and Maintenance
48
Step 7: Running the Test
• Specimen Repositioning
• Tester status and monitoring
• Data Logging via VIRTUAL STRIPCHART (optional)
• Calibration
Q-SUNOperation and Maintenance
49
Specimen Repositioning• Ensures best repeatability and reproducibility
• Perform at least 4 times per test
• Important for both rotating rack and flat array testers
Q-SUNOperation and Maintenance
50
LED Status Indicator
Q-SUNOperation and Maintenance
51
Tester Status
Q-SUNOperation and Maintenance
52
Monitoring the Test
Q-SUNOperation and Maintenance
53
• Safety
• Functions of the Tester
• Running a Test
• Calibration
• Maintenance
Q-SUN Topics
Q-SUNOperation and Maintenance
54
Q-SUN Calibration
• Irradiance sensor
– Every 500 light hours
• Black Panel temp sensor
– Every 6 months
• Chamber air temp sensor
– Every 12 months
• Calibration best practices
– Calibrate whenever test conditions are changed
– Calibrate at operating conditions
– Always calibrate the lamps before calibrating the Black Panel
Q-SUNOperation and Maintenance
55
Irradiance Calibration Devices
CR20 UC1 + UC20
Q-SUNOperation and Maintenance
56
Irradiance calibration ports
Xe-1Xe-2
Xe-3
Q-SUNOperation and Maintenance
57
Irradiance CalibrationNew Q-SUN Testers
Plug UC20 Smart Sensor directly into tester USB port
Select optical filter and calibrate
Q-SUNOperation and Maintenance
58
Black Panel Calibration Devices
CT202 UC1 + UC202
BP IBP
Q-SUNOperation and Maintenance
59
Black Panel temperature sensor calibrationTester black panels and calibrated black panel are placed
adjacent to each other in the chamber
Tester Black Panel
Calibration black panel
Calibrationdevice
Q-SUNOperation and Maintenance
60
Black Panel Temp Sensor Calibration
In holders On tray
Q-SUNOperation and Maintenance
61
Black Panel Thermometer Calibration
New Q-SUN:Let temperature stabilize
and press Calibrate
Legacy Q-SUN; enter value from UC202 or CT202
Q-SUNOperation and Maintenance
62
RH/CAT Sensor ReplacementYearly (Xe-2 and Xe-3)
Xe-3Xe-2
Q-SUNOperation and Maintenance
63
Chamber Air Temp Sensor Calibration
Q-SUNOperation and Maintenance
64
Yearly Reference Device Calibration
Replace/return reference calibration devices to Q-Lab or local representative for recalibration/replacement
– UC20 Irradiance Smart Sensors (replace/recalibrate)
– UC202 Temperature Smart Sensors (replace/recalibrate)
– Chamber Air Temp / RH sensor (replace)
– CR20 (recalibrate)
– CT202 (recalibrate)
Q-SUNOperation and Maintenance
65
• Safety
• Functions of the Tester
• Running a Test
• Calibration
• Maintenance
Q-SUN Topics
Q-SUNOperation and Maintenance
66
Xenon Lamp Replacement Schedule
Q-SUN tester IrradianceWarrantedLamp Life (hrs)
Legacy Typical 1500
NEW “E” Model Typical 3000
NEW “E” Model Maximum 1000
• Xenon lamps need to be replaced because they “age” (spectral shift to less UV)
• Q-SUN Optical Filters do not age
Q-SUNOperation and Maintenance
67
Q-SUN Xe-1, Xe-3 Lamp Replacement
(2) Remove lamp and clean housing
(1) Remove lamp housing from tester
(3) Clean UV filters
Q-SUNOperation and Maintenance
68
Q-SUN Xe-1, Xe-3 Lamp Replacement
(4) Replace lamp
(7) Install replacement lamp
(6) Final lamp check
(5) Verify trigger finger contact
Q-SUNOperation and Maintenance
69
Q-SUN Xe-2 Lamp Replacement
(1) Remove old lamp
(2) Remove and clean filter lantern with ammonia
Q-SUNOperation and Maintenance
70
Q-SUN Xe-2 Lamp Replacement
(4) Verify trigger finger contact
(3) Install replacement lamp
Q-SUNOperation and Maintenance
71
Routine Maintenance
• Clean / Replace Air Filters (Monthly)
• Clean humidifier
• Inspect Water Filter
• Clean Spray Nozzles
• Inspect Chamber Wall Reflectors
• Clean slip ring track (Xe-2)
Every 6 months
Q-SUNOperation and Maintenance
72
Clean or Replace Air Filters
Disposable Washable
Q-SUNOperation and Maintenance
73
Clean or Replace Air Filters
Xe-3 chamber air filter
Xe-3 Ballast Blower air filter
Xe-1 air filter
Chiller air filter
Q-SUNOperation and Maintenance
74
Clean or Replace Air Filters
Xe-2 Chamber Air Filter
Xe-2 LampCoolingAir Filter
Q-SUNOperation and Maintenance
75
Clean humidifier
Water pan Reservoir
• Disassemble humidifier water pan and reservoir
• Clean with alcohol or mild detergent
Q-SUNOperation and Maintenance
76
Inspect Water Filter
Xe-3Xe-2
Q-SUNOperation and Maintenance
77
Clean Spray Nozzles
• Remove and disassemble for cleaning
• Clean in ultra-sonic cleaner or …
• Clean with anti-scaling cleaner (to remove calcium and magnesium deposits) or…
• Thorough washing and rinsing in detergent
Q-SUNOperation and Maintenance
78
Inspect Chamber Reflectors
Clean - OK Cloudy - Replace
Q-SUNOperation and Maintenance
79
Clean slip ring track (Xe-2)
• Wipe down with scouring pads
• Clean with isopropyl alcoholNew contactless system
for BP Temp Sensor!
Q-SUNOperation and Maintenance
80
Q-Lab Tester Education Program• This presentation is a condensed version of our full
Q-SUN operator training program
• Q-Lab offers on-site and virtual training, including hands-on sessions and more in-depth review of tester construction and operation!
• Q-Lab offers customized training plans including:– Hands-on operator training for your existing tester.
– Lecture-style presentations
– Customizable subject matter
– Content for all experience levels
– Affordable, full day sessions with all-in-one pricing
– Private or group sessions at your site or a shared site
• Contact [email protected] for more info
Smrithi KumarCustomer Education Specialist
Q-SUNOperation and Maintenance
81
Q-Lab Repair and Tester Support• Free email/phone support for North America and remote offices worldwide
– Call +1 440 835 8700 (Mon-Fri 8-5 EST)
– Email [email protected]
• On-site emergency repair
• ISO 17025 Accredited on-site calibration audits for all models (current and legacy)– Includes a comprehensive tester calibration
– Thorough inspection / data collection for functionality and on-site training
– Endorsed by A2LA/ILAC for temperature, humidity, irradiance, pressure and orifice (MTG)
• New tester commissioning and re-commissioning
• Accessory installation– Add Water Repur, Chiller, or Space Saver; verify commissioning of relocated tester
• Call Q-Lab Repair or email [email protected] for a quote today!