prueba de disolutor
DESCRIPTION
farmacia, test de cualificacionTRANSCRIPT
-
dx.doi.org/10.14227/DT060199P16
Hollow ShaftTM Sampling Method for Dissolution Testing
Richard Hengst and Rolf Rolli, Sotax Corporation, Furlong, PA
Introduction
With the increased testing demand in the laboratories of pharmaceutical
companies, governmental institutions and contract service companies,
automation becomes an important issue. Automated procedures can guarantee a
higher throughput of samples with increased accuracy and a decreased variation
due to human interference. With pressure to reduce the average cost per analysis
automation is a viable solution. This paper compares Hollow ShaftTM sampling
versus manual sampling in apparatus 1 and 2 dissolution testing.
Background
Sampling procedures in dissolution testing apparatus generate a great interest in
the scientific community as they can influence test results significantly.
Therefore, the regulatory authorities have taken this into consideration and have
issued guidelines. The USP 23 [1] describes exactly how samples should be
taken; "Withdraw a specimen from a zone midway between the surface of the
Dissolution Medium and the top of the rotating basket or blade, not less than 1
cm from the vessel wall" (see figure 1). With the increased tendency to automate
dissolution procedures this becomes even more important. The USP 23 requires
"No part of the assembly, including the environment in which the assembly is
placed, contributes significant motion, agitation, or vibration beyond that due to
the smoothly rotating stirring element [1]. In USP apparatus 1 and 2 dissolution
tests the samples are traditionally taken with a probe placed into the vessel. The
probe adds turbulence if it remains in the liquid phase. Lift devices can be
installed to move the probes in and out of the vessel at the required sampling
times. In addition filtration with a probe sampling system may not meet the lab
requirements. To overcome these problems, the Hollow ShaftTM sampling system
was introduced several years ago and installed in numerous pharmaceutical
laboratories throughout the world (figure 2). The Hollow ShaftTM sampling
system allows for short sampling intervals, down to 20 seconds per sample and
therefore the analysis of fast release drugs is possible.
-
Figure 1: Pink rectangle is
the sampling zone according
to USP 23
Figure 2: Hollow Shaft
Sampling System
The scientific community represented by the International Pharmaceutical
Federation (FIP) issued new guidelines to reflect recent developments in testing,
methodology and equipment [2]. In regards to automation, the guidelines state,
"Validation of automated systems, concerning the sampling and analytical
procedure, including media preparation and test performance, has to consider
accuracy, precision and avoid contamination by any dilutions, transfers, cleaning
or sample or solvent preparation procedures." It should be proven that there is no
significant difference between data obtained with the manual dissolution
equipment and the automated system, including manipulations such as permanent
sampling probes, additional valves, hollow shaft, etc. In addition, the USP has
issued guidelines regarding automated sampling and states[3] "If automated
equipment is used for sampling and the apparatus is modified, validation of
modified apparatus is needed to show that there is no change .."
Comparisons between manual and Hollow ShaftTM sampling have already been
presented to show that no significant differences exist between the results of
samples taken manually or by means of Hollow ShaftTM [4,5]. This study presents
an extended comparison of manual (probe) sampling versus Hollow
ShaftTM sampling using semi automatic on-line and off-line systems and a fully
automated dissolution system, the SOTAX AT70. The official USP calibrator
tablets were used as the test samples. All results have been taken from dissolution
systems located in pharmaceutical labs to show a realistic comparison of
equipment in routine use at different locations around the world.
Equipment
The following equipment was used in the study:
-
On-line: SOTAX AT7 dissolution
tester with Hollow ShaftTM sampling
system and CY7-50 piston pump
connected to a Perkin Elmer
Lambda 20 spectrophotometer
Off-line: SOTAX AT7 dissolution
tester with Hollow ShaftTM sampling system and CY7-50 piston pump connected
to a C 613 fraction collector
Filters: Whatman GF/D 2.7 m microfiber filters
Fully automated system :
SOTAX AT70 with Hollow ShaftTM sampling system and
CY7-50 piston pump connected to a Perkin Elmer
Lambda 20 and/or to a fraction collector C 613 for
combined on/off-line operation
Filters: Gelman 0.45 m glass fiber filters
Sampling:
In all equipment, manual sampling through the SOTAX
pipette guide ensured that the manual sample was consistently taken from the
same position. The automated sampling was done through the Hollow ShaftTM.
Experimental Methods
All tests were executed according to the USP apparatus calibrator test using
current lots of USP tablets and standards.
-
The dissolution medium was prepared with deionized water and the specified
chemicals. The medium was deareated by helium sparging for 30 sec while
maintaining a temperature of 40C. After the test of 30 minutes the samples were
withdrawn simultaneously. All samples have been measured with a Perkin-Elmer
Lambda spectrophotometer except on the AT70 #97.1.001 where a Shimadzu
UV-1601 was used. The spectrophotometers were equipped with either 1mm or
10mm flow through cuvettes . Standard readings were done before and after the
sample measurements. The mean of both standard values was used to calculate %
dissolved.
Test conditions
Salicylic acid Prednisone
Standard:
Lot I, Weigh approx.
50mg and dissolve in
1000ml phosphate
buffer pH 7.4 +/- 0,05
(to increase solubility
1% Ethanol was
added)
Lot K, Weigh approx.
50mg and dissolve in
1000ml deionized
water (to increase
solubility 2% Ethanol
was added)
Tablets: Lot N, 300mg active
substance
Lot L, 50mg active
substance
Dissolution Medium:
900ml deaerated
phosphate buffer pH
7.40 +/- 0.05, 37.0 +/-
0.3C
900ml deaerated
water, 37.0 +/- 0.3C
Sampling Interval: 30 minutes 30 minutes
Measurement: 296nm 242nm
Paddle: 100 rpm 50 rpm
Basket: 100 rpm 100 rpm
Results
The apparatus involved in the study are described by the serial numbers. The
sampling method is coded as follows:
M: Manual
HS on-line: Hollow shaft sampling on-line with spectrophotometer (see
Equipment).
-
HS off-line: Hollow shaft sampling off-line with fraction collector and
subsequent measurement with spectrophotometer (see Equipment).
The % dissolved results from each vessel, the mean and the difference of manual
sampling minus hollow shaft sampling is shown in figure 3.
Figure 3.
Table 1
AT7 on-line systems
The following systems were used in the study: AT7 #96.4.162 , AT7 #96.4.145.
Table 2
AT7 off-line systems
The following systems were used in the study: AT7 #91.1.074 two studies, AT7
#91.1.056, AT7 #94.4.098, AT7 #94.4.097, AT7 #91.1.056.
Table 3
AT70 on-line and combined on/off-line
The following systems were used in the study: AT70 #97.1.001, AT70
#97.1.002, AT70 #97.1.005, AT70 #97.1.006, AT70 #97.1.007, AT70 #97.1.008
Statistical Evaluation
All data for the paddle method were combined to recalculate mean and standard
deviation to have a clear statistical comparison.
-
Conclusion
The data show that samples taken by means of the Hollow ShaftTM method are
comparable to the results taken by a manual method. The system of Hollow
ShaftTM sampling is fully compliant with all current FIP,USP and FDA guidelines
and is an efficient sampling method. The Hollow ShaftTM sampling system can be
utilized in a range of configurations from fully automated to on- or off-line
operation for reliable and consistent data.
References
[1] USP 23, January 1, 1995 page 1791
[2] Pharmacopeial Forum, Volume 21, Number 5, 1371-1382
[3] Pharmacopeial Forum 23 Number 6 (Nov-Dec 1997), page 5257
[4] Dissolution Technologies, Vol. 3, issue 2, May 1996, 11-15
[5] Rolf Rolli oral presentation at FIP Spring Meeting, 15th May 1997, Allschwil