prueba de disolutor

Upload: myriam

Post on 06-Jan-2016

3 views

Category:

Documents


0 download

DESCRIPTION

farmacia, test de cualificacion

TRANSCRIPT

  • dx.doi.org/10.14227/DT060199P16

    Hollow ShaftTM Sampling Method for Dissolution Testing

    Richard Hengst and Rolf Rolli, Sotax Corporation, Furlong, PA

    Introduction

    With the increased testing demand in the laboratories of pharmaceutical

    companies, governmental institutions and contract service companies,

    automation becomes an important issue. Automated procedures can guarantee a

    higher throughput of samples with increased accuracy and a decreased variation

    due to human interference. With pressure to reduce the average cost per analysis

    automation is a viable solution. This paper compares Hollow ShaftTM sampling

    versus manual sampling in apparatus 1 and 2 dissolution testing.

    Background

    Sampling procedures in dissolution testing apparatus generate a great interest in

    the scientific community as they can influence test results significantly.

    Therefore, the regulatory authorities have taken this into consideration and have

    issued guidelines. The USP 23 [1] describes exactly how samples should be

    taken; "Withdraw a specimen from a zone midway between the surface of the

    Dissolution Medium and the top of the rotating basket or blade, not less than 1

    cm from the vessel wall" (see figure 1). With the increased tendency to automate

    dissolution procedures this becomes even more important. The USP 23 requires

    "No part of the assembly, including the environment in which the assembly is

    placed, contributes significant motion, agitation, or vibration beyond that due to

    the smoothly rotating stirring element [1]. In USP apparatus 1 and 2 dissolution

    tests the samples are traditionally taken with a probe placed into the vessel. The

    probe adds turbulence if it remains in the liquid phase. Lift devices can be

    installed to move the probes in and out of the vessel at the required sampling

    times. In addition filtration with a probe sampling system may not meet the lab

    requirements. To overcome these problems, the Hollow ShaftTM sampling system

    was introduced several years ago and installed in numerous pharmaceutical

    laboratories throughout the world (figure 2). The Hollow ShaftTM sampling

    system allows for short sampling intervals, down to 20 seconds per sample and

    therefore the analysis of fast release drugs is possible.

  • Figure 1: Pink rectangle is

    the sampling zone according

    to USP 23

    Figure 2: Hollow Shaft

    Sampling System

    The scientific community represented by the International Pharmaceutical

    Federation (FIP) issued new guidelines to reflect recent developments in testing,

    methodology and equipment [2]. In regards to automation, the guidelines state,

    "Validation of automated systems, concerning the sampling and analytical

    procedure, including media preparation and test performance, has to consider

    accuracy, precision and avoid contamination by any dilutions, transfers, cleaning

    or sample or solvent preparation procedures." It should be proven that there is no

    significant difference between data obtained with the manual dissolution

    equipment and the automated system, including manipulations such as permanent

    sampling probes, additional valves, hollow shaft, etc. In addition, the USP has

    issued guidelines regarding automated sampling and states[3] "If automated

    equipment is used for sampling and the apparatus is modified, validation of

    modified apparatus is needed to show that there is no change .."

    Comparisons between manual and Hollow ShaftTM sampling have already been

    presented to show that no significant differences exist between the results of

    samples taken manually or by means of Hollow ShaftTM [4,5]. This study presents

    an extended comparison of manual (probe) sampling versus Hollow

    ShaftTM sampling using semi automatic on-line and off-line systems and a fully

    automated dissolution system, the SOTAX AT70. The official USP calibrator

    tablets were used as the test samples. All results have been taken from dissolution

    systems located in pharmaceutical labs to show a realistic comparison of

    equipment in routine use at different locations around the world.

    Equipment

    The following equipment was used in the study:

  • On-line: SOTAX AT7 dissolution

    tester with Hollow ShaftTM sampling

    system and CY7-50 piston pump

    connected to a Perkin Elmer

    Lambda 20 spectrophotometer

    Off-line: SOTAX AT7 dissolution

    tester with Hollow ShaftTM sampling system and CY7-50 piston pump connected

    to a C 613 fraction collector

    Filters: Whatman GF/D 2.7 m microfiber filters

    Fully automated system :

    SOTAX AT70 with Hollow ShaftTM sampling system and

    CY7-50 piston pump connected to a Perkin Elmer

    Lambda 20 and/or to a fraction collector C 613 for

    combined on/off-line operation

    Filters: Gelman 0.45 m glass fiber filters

    Sampling:

    In all equipment, manual sampling through the SOTAX

    pipette guide ensured that the manual sample was consistently taken from the

    same position. The automated sampling was done through the Hollow ShaftTM.

    Experimental Methods

    All tests were executed according to the USP apparatus calibrator test using

    current lots of USP tablets and standards.

  • The dissolution medium was prepared with deionized water and the specified

    chemicals. The medium was deareated by helium sparging for 30 sec while

    maintaining a temperature of 40C. After the test of 30 minutes the samples were

    withdrawn simultaneously. All samples have been measured with a Perkin-Elmer

    Lambda spectrophotometer except on the AT70 #97.1.001 where a Shimadzu

    UV-1601 was used. The spectrophotometers were equipped with either 1mm or

    10mm flow through cuvettes . Standard readings were done before and after the

    sample measurements. The mean of both standard values was used to calculate %

    dissolved.

    Test conditions

    Salicylic acid Prednisone

    Standard:

    Lot I, Weigh approx.

    50mg and dissolve in

    1000ml phosphate

    buffer pH 7.4 +/- 0,05

    (to increase solubility

    1% Ethanol was

    added)

    Lot K, Weigh approx.

    50mg and dissolve in

    1000ml deionized

    water (to increase

    solubility 2% Ethanol

    was added)

    Tablets: Lot N, 300mg active

    substance

    Lot L, 50mg active

    substance

    Dissolution Medium:

    900ml deaerated

    phosphate buffer pH

    7.40 +/- 0.05, 37.0 +/-

    0.3C

    900ml deaerated

    water, 37.0 +/- 0.3C

    Sampling Interval: 30 minutes 30 minutes

    Measurement: 296nm 242nm

    Paddle: 100 rpm 50 rpm

    Basket: 100 rpm 100 rpm

    Results

    The apparatus involved in the study are described by the serial numbers. The

    sampling method is coded as follows:

    M: Manual

    HS on-line: Hollow shaft sampling on-line with spectrophotometer (see

    Equipment).

  • HS off-line: Hollow shaft sampling off-line with fraction collector and

    subsequent measurement with spectrophotometer (see Equipment).

    The % dissolved results from each vessel, the mean and the difference of manual

    sampling minus hollow shaft sampling is shown in figure 3.

    Figure 3.

    Table 1

    AT7 on-line systems

    The following systems were used in the study: AT7 #96.4.162 , AT7 #96.4.145.

    Table 2

    AT7 off-line systems

    The following systems were used in the study: AT7 #91.1.074 two studies, AT7

    #91.1.056, AT7 #94.4.098, AT7 #94.4.097, AT7 #91.1.056.

    Table 3

    AT70 on-line and combined on/off-line

    The following systems were used in the study: AT70 #97.1.001, AT70

    #97.1.002, AT70 #97.1.005, AT70 #97.1.006, AT70 #97.1.007, AT70 #97.1.008

    Statistical Evaluation

    All data for the paddle method were combined to recalculate mean and standard

    deviation to have a clear statistical comparison.

  • Conclusion

    The data show that samples taken by means of the Hollow ShaftTM method are

    comparable to the results taken by a manual method. The system of Hollow

    ShaftTM sampling is fully compliant with all current FIP,USP and FDA guidelines

    and is an efficient sampling method. The Hollow ShaftTM sampling system can be

    utilized in a range of configurations from fully automated to on- or off-line

    operation for reliable and consistent data.

    References

    [1] USP 23, January 1, 1995 page 1791

    [2] Pharmacopeial Forum, Volume 21, Number 5, 1371-1382

    [3] Pharmacopeial Forum 23 Number 6 (Nov-Dec 1997), page 5257

    [4] Dissolution Technologies, Vol. 3, issue 2, May 1996, 11-15

    [5] Rolf Rolli oral presentation at FIP Spring Meeting, 15th May 1997, Allschwil