proton therapy coverage for prostate cancer treatment
TRANSCRIPT
![Page 1: Proton Therapy Coverage for Prostate Cancer Treatment](https://reader035.vdocuments.us/reader035/viewer/2022080101/57501def1a28ab877e8e1c4e/html5/thumbnails/1.jpg)
Int. J. Radiation Oncology Biol. Phys., Vol. 70, No. 5, pp. 1492–1501, 2008Copyright � 2008 Elsevier Inc.
Printed in the USA. All rights reserved0360-3016/08/$–see front matter
doi:10.1016/j.ijrobp.2007.09.001
CLINICAL INVESTIGATION Prostate
PROTON THERAPY COVERAGE FOR PROSTATE CANCER TREATMENT
CARLOS VARGAS, M.D.,*y MARCUS WAGNER, M.D.,* CHAITALI MAHAJAN, M.D.,*
DANIEL INDELICATO, M.D.,y AMBER FRYER, B.S.,* AARON FALCHOOK, B.S.,* DAVID HORNE, C.M.D.,*
ANGELA CHELLINI, B.S.,* CRAIG MCKENZIE, C.M.D.,* PAULA LAWLOR, C.M.D.,* ZUOFENG LI, D.SC.,*
LIYONG LIN, PH.D.,* AND SAMEER KEOLE, M.D.*
*University of Florida Proton Therapy Institute, Jacksonville, FL; and yDepartment of Radiation Oncology,University of Florida College of Medicine, Gainesville, FL
Purpose: To determine the impact of prostate motion on dose coverage in proton therapy.Methods and Materials: A total of 120 prostate positions were analyzed on 10 treatment plans for 10 prostatepatients treated using our low-risk proton therapy prostate protocol (University of Florida Proton Therapy Insti-tute 001). Computed tomography and magnetic resonance imaging T2-weighted turbo spin-echo scans wereregistered for all cases. The planning target volume included the prostate with a 5-mm axial and 8-mm superoin-ferior expansion. The prostate was repositioned using 5- and 10-mm one-dimensional vectors and 10-mm multidi-mensional vectors (Points A–D). The beam was realigned for the 5- and 10-mm displacements. The prescriptiondose was 78 Gy equivalent (GE).Results: The mean percentage of rectum receiving 70 Gy (V70) was 7.9%, the bladder V70 was 14.0%, and thefemoral head/neck V50 was 0.1%, and the mean pelvic dose was 4.6 GE. The percentage of prostate receiving78 Gy (V78) with the 5-mm movements changed by �0.2% (range, 0.006–0.5%, p > 0.7). However, the prostateV78 after a 10-mm displacement changed significantly (p < 0.003) with different movements: 3.4% (superior),�5.6% (inferior), and �10.2% (posterior). The corresponding minimal doses were also reduced: 4.5 GE,�4.7 GE, and �11.7 GE (p # 0.003). For displacement points A–D, the clinical target volume V78 coverage hada large and significant reduction of 17.4% (range, 13.5–17.4%, p < 0.001) in V78 coverage of the clinical targetvolume. The minimal prostate dose was reduced 33% (25.8 GE), on average, for Points A–D. The prostate minimaldose improved from 69.3 GE to 78.2 GE (p < 0.001) with realignment for 10-mm movements.Conclusion: The good dose coverage and low normal doses achieved for the initial plan was maintained with move-ments of #5 mm. Beam realignment improved coverage for 10-mm displacements. � 2008 Elsevier Inc.
Protons, Dose–volume, Prostate cancer, High dose, Motion.
INTRODUCTION
Interfraction prostate motion has been an area of major con-
cern for the past few years in prostate cancer therapy (1–6).
Large variations in daily prostate position have been de-
scribed in various studies (1, 2, 4, 6). Furthermore, studies
have found a poor correlation between the location of the
prostate and skin marks or bony anatomy (5, 7). Also, ultra-
sound systems cannot reliably define the daily prostate posi-
tion (4, 5). As a result, when daily treatment setup is based
on skin marks, bony anatomy, or ultrasound guidance, large
margins are necessary to ensure accurate delivery of the pre-
scribed dose to the prostate (2, 4, 5).
Several different approaches have been used in an attempt
to solve the problem of prostate motion and reduce the treat-
ment margins. Yan et al. (6) used multiple computed tomog-
raphy (CT) scans to define prostate motion and electronic
Reprint requests to: Carlos Vargas, M.D., University of FloridaProton Therapy Institute, 2015 N. Jefferson St., Jacksonville, FL32206. Tel: (904) 588-1800; Fax: (904) 588-1300; E-mail: [email protected]
14
daily images to quantify setup inaccuracies. They were able
to derive patient-specific margins through an off-line adap-
tive process. The planning target volume (PTV) was greatly
reduced compared with the margin definition using a class so-
lution. A second approach, on-line image guidance, relies on
daily information to derive the daily prostate position (2, 4, 5,
7, 8). Different approaches have been used, including cone-
beam CT, CT-on-rails, fiducial markers, and beacon tran-
sponders (1, 2, 4, 5, 7–11). The accuracy of these systems
relies on the accuracy of the measurement and the appropriate
patient translation adjustments in relation to the treatment
machine isocenter. None of these approaches eliminates the
problem of intrafraction motion, which must still be ad-
dressed.
Changes in prostate dose with defined changes in prostate
position are important in developing treatment strategies for
Conflict of interest: none.Received May 25, 2007, and in revised form Aug 15, 2007.
Accepted for publication Sept 8, 2007.
92
![Page 2: Proton Therapy Coverage for Prostate Cancer Treatment](https://reader035.vdocuments.us/reader035/viewer/2022080101/57501def1a28ab877e8e1c4e/html5/thumbnails/2.jpg)
Proton therapy coverage for prostate cancer treatment d C. VARGAS et al. 1493
proton therapy. For our treatments, we prescribed a minimal
dose to the PTV of 74.1 Gy equivalent (GE) and directed that
78 GE cover 95% of the PTV. Thus, the doses within the PTV
were not completely homogeneous, and different clinical
target volume (CTV) positions even within the PTV had dif-
ferent minimal doses and prescribed dose coverage. To fully
evaluate the differences in dose distribution inside and out-
side the PTV, we repositioned the prostate using unidirec-
tional vectors of 5 and 10 mm and multidirectional vectors
of 10 mm.
In the present analysis, we simulated prostate displacement
in multiple dimensions and magnitudes in 10 different pa-
tients. By characterizing the variations in prostate dose cov-
erage of the prostate with displacement, we hoped to increase
the understanding of the effect of daily variations in prostate
location on the actual doses delivered to the target according
to historical setup techniques that use skin marks and/or bony
anatomy as a surrogate for prostate position. These findings
are critical to understanding the optimal margins or PTV
expansions and the role of image guidance.
METHODS AND MATERIALS
Between August and December 2006, the data from 10 patients
treated on our institutional review board–approved Phase II proton
protocol for low-risk prostate cancer were analyzed (University of
Florida Proton Therapy Institute 001). One of us (C.V.) was respon-
sible for reviewing all the plans and treating the patients. Subsequent
reviews were performed by our physics staff. The plans met all
protocol requirements, with no major deviations.
The patient characteristics were defined according to the National
Comprehensive Cancer Network low-risk definition: clinical Stage
T2a or less, prostate-specific antigen level of <10 ng/mL, and Glea-
son score #6. The staging workup for all patients included 10–12-
core biopsies, pelvic magnetic resonance imaging (MRI), bone scan,
chest X-ray, and blood work, including determination of prostate-
specific antigen, free prostate-specific antigen, alkaline phosphatase,
complete blood count, and testosterone levels.
SimulationAll patients underwent treatment simulation at the University
of Florida Proton Therapy Institute using a 16-slice large-bore
(40-cm) helical CT scanner and 0.23-T open MRI scanner (Phillips,
Eindhoven, The Netherlands). A custom-indexed vacuum-lock
bag was used for immobilization. A rectal balloon with 100 mL of
saline was used in all cases. Laser marks were used for visual inspec-
tion and to correct the alignment in all degrees of freedom. To mea-
sure the rotation of the patient at the level of the femoral heads, we
obtained a scout CT scan and created a horizontal guide line. A ver-
tical line was used to assess the vertical position and yaw. The pa-
tient was then scanned from 5 cm below the ischial tuberosities to
the L5–S1 interspace in 1-mm slices. For the MRI scan, T2-weighted
turbo spin-echo sequences were acquired with 3-mm spacing and
3-mm interpolation.
Volume definitionThe images were imported into Syntegra (Pinnacle, ADAC, Mil-
pitas, CA), and the two data sets were automatically fused. Land-
marks, including the symphysis pubis and femoral heads, were
used to define the appropriate bony alignment in manual correction.
Soft-tissue landmarks, including the rectal wall–prostate interface
and bladder–prostate interface were used to define the soft-tissue
alignment, and additional manual corrections were performed as
necessary to optimize the fusion of the MRI and CT data. The pros-
tate volume alone, using the sagittal-, coronal-, and transverse-reg-
istered information, defined the CTV and was used for contouring.
The seminal vesicles were not included in low-risk cases (12). The
CTV was expanded 5 mm in the axial and 8 mm in the craniocaudal
dimensions to create the PTV. A greater margin was used in the
superoinferior axis to counter large uncertainties when identifying
the prostate apex and base that affected changes in PTV coverage.
The rectum was contoured from the ischial tuberosities to the
sigmoid flexure for all cases (13). The rectal wall was defined with
a 3-mm internal wall extraction from the rectal contour to create
a rectal wall volume. The entire bladder was contoured, and the
bladder wall volume was also defined by a 3-mm extraction to create
an inner wall.
Proton planningThe treatments were planned with Eclipse software (Varian Med-
ical Systems, Palo Alto, CA), and all necessary beam information
from our gantries was downloaded into the planning software pro-
gram. We used double-scattered proton beams. The proton beam
angles were optimized for each patient. All cases were treated with
posterior oblique beams with angles between 4–12� and 168–176�.The energy for all cases was 232 MeV, and the range was 25.6–28.0
cm. The modulation varied from to 8.5 to 10.5 cm. Distal and prox-
imal margins were added to the PTV and were 5–8 mm minimum
from the edge of the PTV to the 98% isodose line. Smearing was
calculated according to the quadratic summation of the different
preparation and execution errors and was defined as 1.9 cm. The
plans were designed using two beams centered on the PTV. Each
beam was optimized independently with respect to PTV coverage
and to avoid the bladder and rectum. The beam angles were also
individually optimized to improve the PTV dose distribution and
minimize the rectal and bladder dose. The distance to the block
edge was 1 cm, except for around the rectum and bladder, where
the aperture was manually customized to minimize the dose to the
normal surrounding structures while maintaining the PTV coverage.
Specifically, the block edge distances were 7–8 mm in the posterior
aspect and 8–10 mm in the superior/bladder aspect. A dose of 78 GE
was prescribed in two GE daily fractions to the PTV. The target cov-
erage requirements were that 95% of the PTV would receive 100%
of the dose and that the minimal dose to the prostate (CTV) would
be 99% of the prescribed dose. The doses were calculated using a
relative biologic effectiveness of 1.1.
Position changesFor each case, a second image data set was created from the plan-
ning CT information. Using the bony anatomy and soft tissue as
guides, both identical data sets were fused to the MRI scan using
Eclipse software (Varian Medical Systems). Identical isocenter po-
sitions for the prostate were verified. In each case, the prostate vol-
ume was then repositioned relative to the original data set and to the
remainder of the pelvic anatomy. In total, 11 new prostate positions
were generated for each patient using each of the following unidirec-
tional shifts: 5 mm anterior, 5 mm inferior, 5 mm posterior, 5 mm
superior, 10 mm inferior, 10 mm posterior, and 10 mm superior.
Four additional prostate (CTV) positions were defined by tridirec-
tional shifts: Point A, 10 mm superiorly, 10 mm posteriorly, and
10 mm left; Point B, 10 mm superiorly, 10 mm anteriorly, and 10
mm right; Point C, 10 mm inferiorly, 10 mm posteriorly, and 10
![Page 3: Proton Therapy Coverage for Prostate Cancer Treatment](https://reader035.vdocuments.us/reader035/viewer/2022080101/57501def1a28ab877e8e1c4e/html5/thumbnails/3.jpg)
1494 I. J. Radiation Oncology d Biology d Physics Volume 70, Number 5, 2008
mm right; and Point D, 10 mm inferiorly, 10 mm anteriorly, and 10
mm left. These 11 positions, determined according to the unidirec-
tional shifts, were surrogates for the prostate movements that occur
during and between treatment fractions. The beam was realigned in
the verification mode to correspond to prostate displacements of 5
and 10 mm.
Statistical analysisThe relative prostate coverage every 5 GE between 5 and 75 GE
and every 2 GE between 78 and 82 GE were used to completely de-
fine the dose–volume coverage for each prostate position. The min-
imal dose, maximal dose, and mean dose to the prostate for each
position was also analyzed. The rectal, bladder, femoral head, and
total pelvic doses were used to calculate the dose to the surrounding
normal structures. Analysis of variance was used to test the differ-
ences in coverage between the different treatment positions. Line
graph representations with standard errors were used for graphic
representation of the differences. Post hoc analyses were done using
least significant difference pair-wise multiple comparison tests.
Least significant difference pair-wise multiple comparison was
used to test the significance between the different prostate positions
and the initial prostate position. Beam realignment was done for all
5- and 10-mm displacements. The Student paired t test was used to
compare the prostate coverage with or without beam realignment. A
two-tailed p value of #0.05 was considered statistically significant.
Statistical analysis was performed with SYSTAT, version 11.0
(SPSS, Chicago, IL).
Table 1. Patient characteristics
Characteristic Value
Prostate volume (cm3)Mean volume 72.7Median 78.795% CI 60.9–84.5Range 45–95
Prostate length (cm)Mean 5.0Median 5.1
PTV (cm3)Mean 151.1Median 161.095% CI 131.6–168.4Range 106–181
PTV length (cm)Mean 7.0Median 7.0
Rectal volume (cm3)Mean 180Median 18895% CI 160–201Range 79–222
Rectal length (cm)Mean 13Median 13
Bladder volume (cm3)Mean 236.1Median 214.595% CI 188–284Range 88–442
Abbreviations: CI = confidence interval; PTV = planning targetvolume.
RESULTS
The patient characteristics are given in Table 1. The PTV
was, on average, twice the size of the prostate volume and
one-half the rectal length. Small changes in the PTV margins
significantly increased the PTV and, therefore, the irradiated
volume. For the initial prostate position, coverage to the pre-
scribed dose (percentage of prostate receiving 78 Gy [V78])
was >99.9% � 0.008%, the minimal dose was 78.17 �0.11 GE, the maximal dose was 81.30 � 0.96 GE, and the
mean dose was 79.60 � 0.35 GE.
Dose to normal structuresThe doses to the surrounding pelvic structures are given in
Table 2. Low doses were delivered to the rectum, bladder,
femoral heads, and pelvic tissue.
Prostate coverage with motionWithin the PTV. The dose–volume curves can be seen for
different prostate positions in Figs. 1–3. Excellent coverage
of the prostate was achieved when the prostate position
shifted only 5 mm in every direction (Fig. 1). The histogram
curve for the initial prostate location overlapped the curves
for the four 5-mm prostate shifts. The prostate (CTV) cover-
age only decreased after 78 GE, at which point the plot shoul-
ders began to curve downward. When the prostate target was
moved #5 mm, we observed >99% coverage to the pre-
scribed dose, and no statistically significant difference com-
pared with the initial prostate position (Table 3 and Fig. 4).
The minimal, maximal, and mean dose were similar to the
values obtained with the initial prostate position (Table 4
and Figs. 5 and 6).
Table 2. Percentage of volume of rectum and bladderreceiving doses of 10–80 GE and mean dose (n = 20 plans)
Relative volume SD
RectumV10 (%) 29.8 5.6V30 (%) 20.7 3.9V50 (%) 14.6 3.0V70 (%) 7.9 1.8V78 (%) 2.9 1.2V80 (%) 0.1 0.3Mean dose (GE) 14.2 3.7
BladderV10 (%) 36.4 13.2V30 (%) 27.7 11.1V50 (%) 21.5 9.1V70 (%) 14.0 7.2V78 (%) 7.6 5.1V80 (%) 2.1 3.0Mean dose (GE) 18.4 6.2
Femoral head (%)V35 11.1 10V50 0.1 0.3
Mean pelvic dose (GE) 4.6 0.9
Abbreviations: SD = standard deviation; GE = Gray equivalent;V10, V30, V35, V50, V70, V78, V80 = percentage of target volume re-ceiving 10, 30, 35, 35, 50, 70, 78, 80 Gy, respectively.
![Page 4: Proton Therapy Coverage for Prostate Cancer Treatment](https://reader035.vdocuments.us/reader035/viewer/2022080101/57501def1a28ab877e8e1c4e/html5/thumbnails/4.jpg)
Proton therapy coverage for prostate cancer treatment d C. VARGAS et al. 1495
Fig. 1. Dose–volume curve for initial prostate position and prostate positions at 5 mm for 1 case.
Outside the PTV. Inferior prostate coverage was seen
when the prostate position moved 10 mm (Fig. 2). Compared
with the initial prostate position, the volume receiving 78 GE
was moderately reduced (Table 3). Slightly larger changes
were seen for the prostate when it moved posteriorly rather
than superiorly or inferiorly. The minimal dose to the prostate
volume was also considerably decreased with these shifts
(Table 4). However, the maximal and mean doses were not
significantly altered, suggesting that these two parameters
might not be ideal to assess adequate CTV coverage (Table
4 and Fig. 6).
Combined motion. With multidirectional displacements
>10 mm, as defined by Points A–D, significant changes in
dose coverage were seen for the prostate (Tables 3 and 4
and Figs. 4–6). The best coverage achieved with such dis-
placement resulted in only 85% of the prostate receiving
the prescribed dose. Furthermore, the minimal dose de-
creased by 27 GE in the worst-case scenario (Fig. 5). As
seen with the 10-mm unidirectional shifts, the maximal and
mean doses poorly reflected the degree to which CTV cover-
age was compromised (Table 4 and Fig. 6).
Dose profilesThe dose profiles are illustrated in Fig. 7. The right–left
beam profile shows the beam dose distribution in the beam
direction. Given uncertainties in the correlation between the
Hounsfield units and stopping power (mean energy loss per
traveled path length [de/dx]), a correction curve was applied
to our planning system. Furthermore, the variability of the
correction de/dx curve and machine energy output were taken
into account when defining the range and modulation. As a
result, we extended our margins in the RL direction outside
the PTV by an additional 5–9 mm. The beam margins were
widest in this direction, which included the periprostatic
soft tissue (Fig. 7). The superoinferior profile shows the
changes with respect to the base and apex. A very sharp
dose fall off, only a few millimeters wide, was seen from
the 90% to 10% isodose lines. As the right side of the curve
Fig. 2. Dose–volume curves for initial prostate position and prostate positions at 10 mm for 1 case.
![Page 5: Proton Therapy Coverage for Prostate Cancer Treatment](https://reader035.vdocuments.us/reader035/viewer/2022080101/57501def1a28ab877e8e1c4e/html5/thumbnails/5.jpg)
1496 I. J. Radiation Oncology d Biology d Physics Volume 70, Number 5, 2008
Fig. 3. Dose–volume curves for initial prostate position and prostate positions A–D for 1 case.
suggests, we commonly saw a small increase of 3–4% over
the prescribed dose in the CTV (Fig. 7, arrow). The third
line represents the anteroposterior direction, with the right
side of the curve correlating with the posterior direction.
This illustrates how the aperture margin can be reduced in
the posterior aspect to achieve greater rectal sparing. The
beam fall off was 8–9 mm at the depth of the prostate poste-
riorly toward the rectum, and the distance between 95% and
50% was 9 mm.
Beam realignmentThe beam position was realigned with the prostate move-
ment for 5- and 10-mn displacements (Table 5). For the 5-
mm shifts, minimal differences were seen in coverage with
or without beam realignment. However, small statistically
significant improvements in the minimal target dose were
Table 3. Relative volume of prostate covered by prescriptiondose (78 GE) and absolute changes for different positions
Mean (Median) 95% CI Change (%) p*
Initial 100 (100.0) 99.99–100.0 NA5 mm
Anterior 99.6 (99.7) 99.2–100 �0.4 0.7Inferior 99.95 (100.0) 99.8–100 �0.05 0.9Posterior 99.5 (99.8) 98.9–100 �0.5 0.6Superior 99.99 (100.0) 99.9–100 �0.01 0.9
10 mmInferior 96.6 (97.0) 95.8–97.4 �3.4 0.003Posterior 89.8 (89.5) 87.1–91.4 �10.2 <0.001Superior 94.4 (94.9) 93.1–95.8 �5.6 <0.001
PointA 83.8 (84.8) 80.5–86.9 �16.2 <0.001B 85.7 (86.7) 83.4–87.9 �14.3 <0.001C 82.6 (82.0) 79.8–85.4 �17.4 <0.001D 86.5 (86.2) 83.9–89.2 �13.5 <0.001
Abbreviations: GE = Gray equivalent; CI = confidence interval.* Step-wise comparisons and Fisher’s least-significant-difference
test for different pairs between initial position and new prostateposition.
seen with realignment in the anterior (p = 0.001) or posterior
(p = 0.008) direction. For 10-mm shifts, larger, significant
improvements were seen for beam realignment. In the infe-
rior, posterior, and superior positions, the prostate minimal
V78 (p < 0.001) and the minimal dose (p < 0.001) were sig-
nificantly improved for all cases.
DISCUSSION
Interfraction changes in prostate position within the pelvis
are considered the major source of dosimetric misses for
prostate irradiation (1, 2, 4–8, 11, 14–19); therefore, margin
Fig. 4. Relative prostate volume receiving prescription dose (78 Gyequivalent) in several positions: 1, initial; 2, 5-mm anterior; 3, 5-mminferior; 4, 5-mm posterior; 5, 5-mm superior; 6, 10-mm inferior; 7,10-mm posterior; 8, 10-mm superior; 9, Point A; 10, Point B; 11,Point C; and 12, Point D.
![Page 6: Proton Therapy Coverage for Prostate Cancer Treatment](https://reader035.vdocuments.us/reader035/viewer/2022080101/57501def1a28ab877e8e1c4e/html5/thumbnails/6.jpg)
Proton therapy coverage for prostate cancer treatment d C. VARGAS et al. 1497
formulas have been heavily weighted to account for this error
(2, 6, 16). For proton therapy, dosimetric variation with pros-
tate displacement has not been adequately described. It is crit-
ical to understand how prostate dose coverage is altered by
motion along clearly defined vectors for (1) patient-specific
intra- and inter-fraction prostate motion profiles; and (2)the impact of the actual dose distribution in the CTV with
prostate motion to quantify the potential benefit of image-
Table 4. Minimal, maximal, and average dose to prostatewith absolute changes for different positions
Dose Mean (Median) 95% CI Change p*
MinimalInitial 78.2 (78.1) 77.9–78.4 0 NA5 mm
Anterior 76.3 (76.4) 75.3–77.2 �1.88 0.7Inferior 78.4 (78.1) 77.5–79.4 0.26 0.9Posterior 77.3 (77.3) 75.7–78.8 �0.89 0.6Superior 78.5 (78.3) 77.6–79.4 0.31 0.9
10 mmInferior 73.4 (72.6) 71.2–75.7 �4.74 0.003Posterior 66.4 (66.3) 60.9–72.0 �11.72 <0.001Superior 73.6 (72.9) 71.6–75.7 �4.53 <0.001
PointA 52.5 (53.9) 49.0–55.9 �25.72 <0.001B 53.6 (52.9) 50.0–57.‘ �24.61 <0.001C 50.7 (52.4) 45.7–55.7 �27.47 <0.001D 53.9 (53.0) 50.1–57.7 �24.22 <0.001
MaximumInitial 81.3 (81.0) 80.6–82.0 0 NA5 mm
Anterior 81.3 (81.0) 80.6–81.9 �0.020 0.9Inferior 81.2 (81.0) 80.5–81.9 �0.11 0.8Posterior 81.1 (80.9) 80.4–81.7 �0.245 0.5Superior 81.0 (80.9) 80.3–81.7 �0.29 0.5
10 mmInferior 81.2 (81.0) 80.5–81.9 �0.106 0.8Posterior 80.8 (80.8) 80.1–81.4 �0.525 0.2Superior 80.9 (80.8) 80.2–81.5 �0.431 0.3
PointA 80.6 (80.7) 80.2–81.1 �0.649 0.1B 81.1 (80.9) 80.5–81.6 �0.23 0.6C 81.2 (80.9) 80.6–81.8 �0.103 0.8D 81.3 (81.2) 80.7–82.0 0.022 0.9
MeanInitial 79.6 (79.6) 79.4–79.9 0 NA5 mm
Anterior 79.6 (79.6) 79.4–79.9 0.010 0.9Inferior 79.0 (79.5) 77.8–80.2 �0.58 0.6Posterior 79.2 (79.5) 78.5–79.8 �0.43 0.7Superior 78.8 (79.5) 77.1–80.4 �0.84 0.5
10 mmInferior 76.9 (79.4) 71.2–82.6 �2.68 0.02Posterior 78.1 (78.8) 76.3–80.0 �1.48 0.2Superior 79.2 (79.3) 79.1–79.4 �0.35 0.8
PointA 78.5 (78.6) 78.2–78,8 �1.07 0.4B 78.7 (78.6) 78.4–78.9 �0.95 0.4C 78.4 (78.3) 78.1–78.7 �1.22 0.3D 78.7 (78.9) 78.5–79.0 �0.86 0.5
Abbreviations: CI = confidence interval; NA = not applicable.* Step-wise comparisons and Fisher’s least-significant-difference
test for different pairs.
guided therapy and the minimal PTV expansion necessary
to maximally benefit from the dose distribution advantages
of proton therapy.
As a result of the use of two fields and lateral oblique beam
angles, the dose distribution shapes for our treatments vary
Fig. 5. Minimal prostate dose in several positions: 1, initial; 2, 5-mm anterior; 3, 5-mm inferior; 4, 5-mm posterior; 5, 5-mm superior;6, 10-mm inferior; 7, 10-mm posterior; 8, 10-mm superior; 9, PointA; 10, Point B; 11, Point C; and 12, Point D.
Fig. 6. Mean prostate dose in several positions: 1, initial; 2, 5-mmanterior; 3, 5-mm inferior; 4, 5-mm posterior; 5, 5-mm superior; 6,10-mm inferior; 7, 10-mm posterior; 8, 10-mm superior; 9, Point A;10, Point B; 11, Point C; and 12, Point D.
![Page 7: Proton Therapy Coverage for Prostate Cancer Treatment](https://reader035.vdocuments.us/reader035/viewer/2022080101/57501def1a28ab877e8e1c4e/html5/thumbnails/7.jpg)
1498 I. J. Radiation Oncology d Biology d Physics Volume 70, Number 5, 2008
Fig. 7. Dose profiles in different beam directions at beam isocenter: 1, superoinferior (SI); 2, anteroposterior (AP); and3, right–left (RL).
from the traditionally seen isodose lines for intensity-modu-
lated radiotherapy. For proton therapy, the dose line distribu-
tions will cover the PTV with a more square-shaped dose.
Within the PTV, dose homogeneity will be great, with
a dose variation of 95–104% at any point. However, a 9%
dose variation seen in the PTV, as well as a sharp fall off
from the 80% to 20% isodose line, can affect CTV coverage
in different positions. To fully evaluate the coverage of the
CTV for different positions inside and outside the PTV, it
is insufficient to evaluate the isodose lines within the PTV.
Thus, the CTV needs to be repositioned and the coverage an-
alyzed for the different motion vectors.
Motion within our PTVZhang et al. (20) was able to show inferior CTV coverage
without an image-guided approach for proton therapy that re-
sulted from prostate movement. Reductions of $10% in the
volume treated to the prescribed dose were seen in 30% of
cases. Quantification of the reduction in terms of absolute
changes in position, however, was not available. In our study,
we were able to show that prostate motion of #5 mm, or
within our PTV margin, allowed coverage of our prescription
dose to 99.8% of the volume (Table 3). Furthermore, as dem-
onstrated in Fig. 1, our PTV dose–volume histogram curves
for prostate motion overlapped with the curve for our original
position CTV. As a result, the minimal doses to the CTV
were, on average, decreased only 0.6 GE. Thus, systematic
motion of #5 mm will have a small effect on the dose deliv-
ered. This was because of the shape of the isodose lines as
they conform to the PTV. As seen in Fig. 1, the CTV dose
curve was very steep with a small shoulder. As a result, the
minimal dose for the initial CTV was $78 GE, which wais
maintained with motion #5 mm. However, it is important
to realize that movements of the CTV of 8 mm in the super-
oinferior dimension—still within the PTV—will result in
larger reductions in coverage than the isodose lines falling in-
side the PTV in this axis. This information could also be use-
ful to quantify the coverage of the prostate after daily image
guidance in which the intrafraction error will likely be within
Table 5. Prostate coverage with and without beamrealignment for different prostate positions
Beam realignment (SD)
Coverage No Yes
5-mm AnteriorProstate V78 (%) 99.6 (0.5) 100 (0.03)Prostate minimal dose (GE) 76.52 (1.17) 78.15 (0.27)
5-mm InferiorProstate V78 (%) 99.6 (0.5) 100 (0.03)Prostate minimal dose (GE) 78.03 (0.34) 78.19 (0.23)
5-mm PosteriorProstate V78 (%) 99.4 (0.8) 100 (0.007)Prostate minimal dose (GE) 76.75 (1.49) 78.29 (0.30)
5-mm SuperiorProstate V78 (%) 99.9 (0.02) 100 (0.004)Prostate minimal dose (GE) 78.10 (0.30) 78.27 (0.32)
10-mm InferiorProstate V78 (%) 96.5 (1.2) 100 (0.1)Prostate minimal dose (GE) 72.47 (0.90) 78.07 (0.27)
10-mm PosteriorProstate V78 (%) 89.8 (3.9%) 100 (0.1)Prostate minimal dose (GE) 64.75 (5.90) 78.31(0.53)
10-mm SuperiorProstate V78 (%) 94.4 (2.0) 100 (0.3)Prostate minimal dose (GE) 72.78 (0.70) 78.28 (0.41)
Abbreviations as in Table 2.
![Page 8: Proton Therapy Coverage for Prostate Cancer Treatment](https://reader035.vdocuments.us/reader035/viewer/2022080101/57501def1a28ab877e8e1c4e/html5/thumbnails/8.jpg)
Proton therapy coverage for prostate cancer treatment d C. VARGAS et al. 1499
5 mm. However, as mentioned by Litzenberg et al. (2), mar-
gins of $5 mm are needed to account for inter-setup error.
An alternate option to deliver the prescribed dose for
interfraction motion >5 mm would be to increase the PTV
margins. However, increasing the PTV margins would com-
promise the low dose delivered to the rectum and bladder
(Table 2). Expanded margins will increase the normal tissue
dose and the risk of toxicity (13, 21, 22). As seen in Table 5,
the beam realignment can have small, but significant, bene-
fits, even for 5-mm movements.
Motion outside our PTVMargins of approximately 10 mm are necessary for the
treatment of prostate cancer without off-line or on-line image
guidance (2, 4–6). This distance served as the basis of our
theoretical prostate shifts. After a 10-mm displacement of
the CTV, 93.6% of the volume received the target dose (Ta-
ble 2). The largest changes were seen with posterior move-
ment, likely attributable to our reduced (7–8-mm) posterior
margin (Fig. 7). Because tissue densities vary across the
PTV, and we had a relatively broader region uncovered in
the posterior aspect, we could selectively decrease our poste-
rior aperture margin distance. This allowed us to achieve very
good coverage with a small (#5 mm) posterior shift toward
the edge of the PTV. However, posterior displacements
>5 mm would place a large percentage of the CTV in the
low-dose region, substantially decreasing the coverage. As
a result, when we moved the CTV 10 mm in the superior
and inferior directions, the minimal dose was only decreased
by 4.6 GE or 5.9% of the prescribed dose (78 GE), but the
CTV coverage was reduced by 11.7 GE or 15% with poste-
rior shifts of 10 mm.
When a proton target is displaced in the beam direction, the
effect is different than that seen with intensity-modulated ra-
diotherapy. Theoretically, prostate motion into the beam will
not compromise the coverage in the same manner as would
setup errors. However, changes in the position of the prostate
can affect target coverage by altering the amount of tissue
a proton beam must transverse. As a result, motion in-line
with the beam direction was only defined when we analyzed
the combined inter- and intrafraction prostate motions (23).
Because movement along the anteroposterior vector is seen
in prostate cancer, it is likely that the underdosing found in
our study with posterior displacements would be reflected
in decreased coverage during a course of proton radiotherapy
(24). A large benefit could be realized by using image guid-
ance for movements outside the PTV. However, it is neces-
sary to account for variations in stopping power in beam
realignment to optimize prostate coverage in an image-
guided environment.
Inter- and intrafraction motionPoints A–D reflect motion due to interfraction error com-
pounded by intrafraction motion. All four points were used
to evaluate the results of the 10-mm movements in the three
axes. This displacement distance was chosen on the basis of
the prostate motion analysis performed by Litzenberg et al.
(2). Kupelian et al. (9) demonstrated that intrafraction dis-
placement >5 mm and lasting $30 s occurred during the
course of therapy in 15% of fractions. As a result, the treated
volume covered by the prescription dose dramatically de-
creased (range, 82.6–85.7%). By accounting for potentially
large interfraction errors with on-line corrections, we can
minimize the PTV expansion required to account for intra-
fraction motion. Therefore, prostate coverage is influenced
largely by the degree of the displacement. The minimal
dose to the CTV was also greatly reduced (range, 24.2–
27.4 GE).
The average and maximal CTV doses are poor reflections
of the varying target coverage. The minimal change in either
descriptive measure corresponded to dramatic changes in our
dose–volume histograms. This raises questions about the
utility of average or maximal dose reporting when attempting
to describe the relationship between target motion and CTV
coverage.
Although real benefits have been obtained through dose
escalation, we have continued to see target underdosage
when patients were treated without image guidance. This
highlights two different issues: (1) image guidance is pivotal
for external radiotherapy delivery, including proton therapy,
if truly high doses are to be delivered; and (2) the true dose
necessary to treat prostate cancer patients with different
risk factors remains unknown for image-guided treatment.
The Radiation Therapy Oncology Group recently opened
a key study comparing 75.6 Gy in 1.8-Gy fractions and 70
Gy in 2.5-Gy fractions (biologically effective dose, 84.5
Gy [a/b = 1.5 in 1.8 Gy]) with image guidance. At our insti-
tution, we have adopted an image guidance protocol that
includes intraprostatic fiducial markers and daily X-ray local-
ization in the true beam direction before treating each field.
Strategies for proton therapyOrgan motion has been a major concern in the clinical ap-
plication of proton therapy. Because the high-dose area shape
and depth are determined by the calculation of the cumulative
stopping power through the tissue in the beam path, the var-
iations in stopping power with beam realignment is of con-
cern. Two major treatment philosophies can be used. First,
if changes in the stopping power are to be avoided to maintain
conformality in a static image, the margins should be in-
creased to account for all prostate motion, at the cost of
greater normal tissue doses. However, if the goal is to reduce
the normal tissue doses by using tighter margins and daily
confirmation of pretreatment prostate position, the dosimetry
should be optimized for coverage of the different prostate po-
sitions through dynamic position representation. In this situ-
ation, changes in the stopping power relative to the beam path
should be accounted for. We have used the latter philosophy.
The dose distribution in proton therapy is determined by
the different tissues along the beam path. Thus, a beam actu-
ally traversing a longer path in bone than modeled will de-
posit its energy short of the target. Furthermore, a beam in
air will travel longer and deposit its energy beyond the target.
Given this uncertainty in range, the beam orientation must be
![Page 9: Proton Therapy Coverage for Prostate Cancer Treatment](https://reader035.vdocuments.us/reader035/viewer/2022080101/57501def1a28ab877e8e1c4e/html5/thumbnails/9.jpg)
1500 I. J. Radiation Oncology d Biology d Physics Volume 70, Number 5, 2008
optimized to limit delivering the dose to normal structures be-
fore or after the target. Therefore, the ideal beam orientation
will be determined by a path to the target that minimizes
interaction with dose-limiting normal structures.
Image guidance for proton therapy will only determine the
x and y axes, and the depth at which the dose will be delivered
is determined by the range calculated during the treatment
planning process. More specifically, the depth of the high-
dose regions is determined at planning by the range and
modulation. Three different options are then possible: (1) de-
termine the depth of the target on a daily basis and change the
range and modulation accordingly; (2) reduce prostate mo-
tion; and (3) determine the motion in the beam direction and
quantify the changes in radiologic path length and incorporate
this information into the treatment plan. The first option will
require smaller proximal and distal margins around the PTV
that are not feasible. Daily cone-beam CT scans will be nec-
essary to determine the changes in the radiologic path length,
in addition to the target position. The second option, although
ideal, has certain limitations. Rectal balloons have not been
found to decrease interfraction motion, the greatest source of
error (25). Intrafraction motion with a rectal balloon needs
further study, preferably incorporating cine-MRI informa-
tion. The third option is feasible as long as the variation in
depth of the target is relatively small. The prostate has rela-
tively large vectors of anteroposterior and superoinferior
motion but very little motion in the right–left direction cor-
responding to the beam orientation (7, 26, 27). At our institu-
tion, we have implemented the third strategy.
The second problem for image guidance results from
changes in the prostate position in relation to the bony anat-
omy and other pelvic structures. Differences in prostate
position will determine different radiologic path lengths for
the beam (28). Thus, the changes in density of the surround-
ing structures also must be determined and incorporated
into the compensator design (28, 29). A relatively simple
approach consists of analyzing all sources of preparation
and execution error in a defined plane perpendicular to
the beam direction. Thus, a quadratic summation of all
sources of error in a given direction should determine the
smearing. We have been using daily image guidance before
each beam with an orthogonal pair of X-rays, including
a true beam’s eye view, which decreases the sources of
execution error. A smearing of 1.9 cm was determined to
be necessary. This decreased plan conformality in a single
static image. However, coverage was maintained with this
strategy, with prostate positional changes of 5 and 10 mm
(Table 5).
CONCLUSION
Excellent PTV coverage with minimal normal structure
exposure can be achieved with planning using on a single
prostate position. Optimal dose coverage was maintained
for small prostate shifts; however, larger movements dramat-
ically decreased the prostate dose. Off-line or on-line image-
guided approaches are necessary for proton treatment of
prostate cancer if genuinely greater doses are to be delivered
to the tumor, as illustrated by our data.
REFERENCES
1. Letourneau D, Martinez AA, Lockman D, et al. Assessment of
residual error for online cone-beam CT-guided treatment of
prostate cancer patients. Int J Radiat Oncol Biol Phys 2005;
62:1239–1246.2. Litzenberg DW, Balter JM, Hadley SW, et al. Influence of intra-
fraction motion on margins for prostate radiotherapy. Int JRadiat Oncol Biol Phys 2006;65:548–553.
3. Rudat V, Schraube P, Oetzel D, et al. Combined error of patient
positioning variability and prostate motion uncertainty in 3D
conformal radiotherapy of localized prostate cancer. Int J RadiatOncol Biol Phys 1996;35:1027–1034.
4. Scarbrough TJ, Golden NM, Ting JY, et al. Comparison of
ultrasound and implanted seed marker prostate localization
methods: Implications for image-guided radiotherapy. Int JRadiat Oncol Biol Phys 2006;65:378–387.
5. Serago CF, Buskirk SJ, Igel TC, et al. Comparison of daily meg-
avoltage electronic portal imaging or kilovoltage imaging with
marker seeds to ultrasound imaging or skin marks for prostate
localization and treatment positioning in patients with prostate
cancer. Int J Radiat Oncol Biol Phys 2006;65:1585–1592.6. Yan D, Lockman D, Brabbins D, et al. An off-line strategy for
constructing a patient-specific planning target volume in adap-
tive treatment process for prostate cancer. Int J Radiat OncolBiol Phys 2000;48:289–302.
7. Schallenkamp JM, Herman MG, Kruse JJ, et al. Prostate posi-
tion relative to pelvic bony anatomy based on intraprostatic
gold markers and electronic portal imaging. Int J Radiat OncolBiol Phys 2005;63:800–811.
8. de Boer HC, van Os MJ, Jansen PP, et al. Application of the no
action level (NAL) protocol to correct for prostate motion based
on electronic portal imaging of implanted markers. Int J RadiatOncol Biol Phys 2005;61:969–983.
9. Kupelian P, Willoughby T, Mahadevan A, et al. Multi-institu-
tional clinical experience with the Calypso system in localiza-
tion and continuous, real-time monitoring of the prostate
gland during external radiotherapy. Int J Radiat Oncol BiolPhys 2007;67:1088–1098.
10. Madsen BL, Hsi RA, Pham HT, et al. Intrafractional stability of
the prostate using a stereotactic radiotherapy technique. Int JRadiat Oncol Biol Phys 2003;57:1285–1291.
11. Serago CF, Chungbin SJ, Buskirk SJ, et al. Initial experience
with ultrasound localization for positioning prostate cancer pa-
tients for external beam radiotherapy. Int J Radiat Oncol BiolPhys 2002;53:1130–1138.
12. Kestin L, Goldstein N, Vicini F, et al. Treatment of prostate can-
cer with radiotherapy: Should the entire seminal vesicles be
included in the clinical target volume? Int J Radiat Oncol BiolPhys 2002;54:686–697.
13. Vargas C, Yan D, Kestin LL, et al. Phase II dose escalation
study of image-guided adaptive radiotherapy for prostate can-
cer: use of dose–volume constraints to achieve rectal isotoxic-
ity. Int J Radiat Oncol Biol Phys 2005;63:141–149.14. Wu J, Haycocks T, Alasti H, et al. Positioning errors and pros-
tate motion during conformal prostate radiotherapy using
on-line isocentre set-up verification and implanted prostate
markers. Radiother Oncol 2001;61:127–133.
![Page 10: Proton Therapy Coverage for Prostate Cancer Treatment](https://reader035.vdocuments.us/reader035/viewer/2022080101/57501def1a28ab877e8e1c4e/html5/thumbnails/10.jpg)
Proton therapy coverage for prostate cancer treatment d C. VARGAS et al. 1501
15. Willoughby TR, Kupelian PA, Pouliot J, et al. Target localiza-tion and real-time tracking using the Calypso 4D localizationsystem in patients with localized prostate cancer. Int J RadiatOncol Biol Phys 2006;65:528–534.
16. Van Herk M, Remeijer P, Rasch C, et al. The probability of cor-rect target dosage: Dose–population histograms for derivingtreatment margins in radiotherapy. Int J Radiat Oncol BiolPhys 2000;47:1121–1135.
17. van der Heide UA, Kotte AN, Dehnad H, et al. Analysis of fi-ducial marker-based position verification in the external beamradiotherapy of patients with prostate cancer. Radiother Oncol2007;82:38–45.
18. Van den HF, Powell T, Seppi E, et al. Independent verificationof ultrasound based image-guided radiation treatment, usingelectronic portal imaging and implanted gold markers. MedPhys 2003;30:2878–2887.
19. Trichter F, Ennis RD. Prostate localization using transabdomi-nal ultrasound imaging. Int J Radiat Oncol Biol Phys 2003;56:1225–1233.
20. Zhang X, Dong L, Lee AK, et al. Effect of anatomic motion onproton therapy dose distributions in prostate cancer treatment.Int J Radiat Oncol Biol Phys 2007;67:620–629.
21. Vargas C, Martinez A, Kestin LL, et al. Dose–volume analysisof predictors for chronic rectal toxicity after treatment of pros-tate cancer with adaptive image-guided radiotherapy. Int JRadiat Oncol Biol Phys 2005;62:1297–1308.
22. Sohn M, Yan D, Liang J, et al. Incidence of late rectal bleedingin high-dose conformal radiotherapy of prostate cancer using
equivalent uniform dose-based and dose–volume-based normaltissue complication probability models. Int J Radiat Oncol BiolPhys 2007;67:1066–1073.
23. Thomas SJ. Margins for treatment planning of proton therapy.Phys Med Biol 2006;51:1491–1501.
24. Padhani AR, Khoo VS, Suckling J, et al. Evaluating the effectof rectal distension and rectal movement on prostate gland po-sition using cine MRI. Int J Radiat Oncol Biol Phys 1999;44:525–533.
25. van Lin ENJT, van der Vight LP, Witjes JA, et al. The effect ofan endorectal balloon and off-line correction on the interfractionsystematic and random prostate position variations: A compar-ative study. Int J Radiat Oncol Biol Phys 2005;61:278–288.
26. Nederveen AJ, Dehnad H, van der Heide UA, et al. Comparisonof megavoltage position verification for prostate irradiationbased on bony anatomy and implanted fiducials. RadiotherOncol 2003;68:81–88.
27. Van Herk M, Bruce A, Kroes AP, et al. Quantification of organmotion during conformal radiotherapy of the prostate by threedimensional image registration. Int J Radiat Oncol Biol Phys1995;33:1311–1320.
28. Moyers MF, Miller DW, Bush DA, et al. Methodologies andtools for proton beam design for lung tumors. Int J Radiat OncolBiol Phys 2001;49:1429–1438.
29. Moyers MF, Miller DW. Range, range modulation, and field ra-dius requirements for proton therapy of prostate cancer. TechnolCancer Res Treat 2003;2:445–447.