protoDUNE – single-phase Instrumentation - Gas Analyzers Stephen Pordes, Fermilab 4/25/17 S. Pordes - protoDUNE single-phase Instrumentation review - gas analzyzers 1

Upload: others

Post on 07-Apr-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: protoDUNE – single-phase Instrumentation - Gas Analyzers ... · Expanded region around Pump Restart after Power Outage 3/23/17 15 PrM1 &4 NanoTrace O2 Gas Analyzer Pump Turn On

protoDUNE – single-phaseInstrumentation- GasAnalyzers

StephenPordes,Fermilab

4/25/17 S.Pordes- protoDUNEsingle-phaseInstrumentationreview- gasanalzyzers 1

Page 2: protoDUNE – single-phase Instrumentation - Gas Analyzers ... · Expanded region around Pump Restart after Power Outage 3/23/17 15 PrM1 &4 NanoTrace O2 Gas Analyzer Pump Turn On

protoDUNE – single-phaseInstrumentation- GasAnalyzers

Motivation:

• Checkthequalityoftheargondeliveredforwater,oxygenandnitrogencontent

• Monitortheprogressofthepurgingofthecryostatatmosphere

• Lookfortheexistence/appearanceofleaksduringthegasphaseofrecirculationandpurification

• Ensurethattheultimatelevelofatmosphericcontaminationislowenoughnottobeasignificantloadonthepurificationsystem

• Monitortheeffectivenessoftheliquidpurificationinitsearlystagesbeforethepuritymonitorscomeonline.

• Helpdiagnoseproblemsshouldtheyoccurduringoperation,includingassessingcontaminationintroducedincaseofemergencyremoval/replacementofequipment.

ThislistmaynotbecompletebutisbasedonourexperienceatFermilab.Thefirstusewasmotivatedbytryingtounderstandthepurgingprocessforremovingtheatmospherefromacryostat.

4/25/17 S.Pordes- protoDUNEsingle-phaseInstrumentationreview- gasanalzyzers 2

Page 3: protoDUNE – single-phase Instrumentation - Gas Analyzers ... · Expanded region around Pump Restart after Power Outage 3/23/17 15 PrM1 &4 NanoTrace O2 Gas Analyzer Pump Turn On

protoDUNE – single-phaseInstrumentation- GasAnalyzers

Experience

• Have(1in~10)rejectedtrailer-loadofargonforhighoxygen(7ppmvs5ofspec;companyinitiallydeniedandthenagreed)

• Havedetectedleaksduringpurgingandrecirculationprocess

• Haveusedtounderstandmajorsourcesthataffectelectrondrift-lifetime

• Haveusedtomonitorpurificationprocessincryostatbeforepuritymonitorsaresensitive

• Haveusedtocomparecontaminantsinullagewithelectrondrift-lifetimetounderstandcontaminantflowinullage.

• Haveusedasdiagnosticwhenincidents(e.g.,pumprestart)haveaffectedelectrondrift-lifetime.

4/25/17 S.Pordes- protoDUNEsingle-phaseInstrumentationreview- gasanalzyzers 3

Page 4: protoDUNE – single-phase Instrumentation - Gas Analyzers ... · Expanded region around Pump Restart after Power Outage 3/23/17 15 PrM1 &4 NanoTrace O2 Gas Analyzer Pump Turn On

protoDUNE – single-phaseInstrumentation- GasAnalyzers

Example

4/25/17 S.Pordes- protoDUNEsingle-phaseInstrumentationreview- gasanalzyzers 4

Expanded region around Pump Restart after Power Outage

3/23/17 15

PrM1 &4

NanoTrace O2 Gas Analyzer

Pump Turn On

O2 Spikes (2.5 ppb) Then falls as LAr is filtered

180 us Lifetime (~1.7 ppb O2 equiv)

Pump Off

Pump output pressure

2days

O2analyzer(p

pb)

2

1

3

PumpturnonproducesaspikeinOxygen

Page 5: protoDUNE – single-phase Instrumentation - Gas Analyzers ... · Expanded region around Pump Restart after Power Outage 3/23/17 15 PrM1 &4 NanoTrace O2 Gas Analyzer Pump Turn On

protoDUNE – single-phaseInstrumentation- GasAnalyzers

Performance:

• Measureonetypeofcontaminantonlyandoverlimitedrange

• Sensitivitiesforoxygenandwaterrangedowntofractionsofappb

• Nitrogendevicesarelesssensitiveandinourexperiencemuchlessstable– theyrequireweeklycalibrationtooperateatthe~0.3ppmaccuracyandshouldbecalibratedbeforeanycrucialtest

• Needaswitchyardtoallowsamplingfrommultiplesources

4/25/17 S.Pordes- protoDUNEsingle-phaseInstrumentationreview- gasanalzyzers 5

PC4 Gas Analyzers

Manufacturer Delta-F Delta-F Delta-F Servomex/Kontrol Analytik Tiger OpticsModel NanoTrace II DF-560 LaserTrace O2 LaserTrace H20 DF-310 DF-310E K2001ns HALO H20 Sum

Budget price $35k $5k $10k $25k $25k $180kSpecies analyzed Oxygen Oxygen Water Oxygen Oxygen Nitrogen Water

Ranges of operation 0-20 ppm 0-1.25 ppm 0-2.5 ppm 0-5,000 ppm 0-50 ppm 0 - 100 ppm 0-20 ppmLDL (experience based) 1 ppb 1 ppb 1 ppb 1 ppm 100 ppb 100 ppb 4 ppb

Comments

The only high sensitivity oxygen meter in this set that is reliable. Used to monitor tank liquid purification and oxygen filter saturation. Long recovery time from oxygen upsets, typically use DF-310E above 1 ppm to protect it and prevent long recovery times.

High range of this oxygen meter is necessary for tank purge from air and to monitor tank vapor space during any tank extraction or insertion type repairs that introduce gross amounts of contamination.

Mid range oxygen analyzer. Used during the purge from air and gas recirculation phases until the NanoTrace can be brought online or any time the contamination is above 1 ppm to protect the NanoTrace.

Only N2 analyzer, necessary for light collection

Typically this analyzer monitors the tank vapor space so the water outgassing can be integrated over the entire run. The other water analyzer is then used to sample from other points in the system.

Tiger Optics - one unit

$80k

These two analyzers are one unit that can't be separated. The oxygen analyzer is difficult to operate

and at the moment is compromised. The water analyzer is the most sensitive water analyzer in the

system. The water analyzer is used to check for water filter saturation and other careful measurements. The

water analyzers take a long time to come to equilibrium because the water in the gas stream must come to

equilibrium with the tubing along its entire length. Thus a water analyzer can't be switched between sources on a short time scale. It can often take days to reach a

stable reading

Moderateandhighsensitivityanalyzersavailabletothefirstatmosphere-to-purityexperienceatFermilab

Page 6: protoDUNE – single-phase Instrumentation - Gas Analyzers ... · Expanded region around Pump Restart after Power Outage 3/23/17 15 PrM1 &4 NanoTrace O2 Gas Analyzer Pump Turn On

protoDUNE – single-phaseInstrumentation- GasAnalyzers

4/25/17 S.Pordes- protoDUNEsingle-phaseInstrumentationreview- gasanalzyzers 6

Various H2O, N2 and O2 Gas Analyzers

Vaisala Dew Point Meter

N2

H2O & O2

O2

H2O

O2 (0->25%) Monitor

Menorah built by PAB techs.MicroBooNE equivalent cost $30k

TheMenorahswitchyard(7in,8out)

Page 7: protoDUNE – single-phase Instrumentation - Gas Analyzers ... · Expanded region around Pump Restart after Power Outage 3/23/17 15 PrM1 &4 NanoTrace O2 Gas Analyzer Pump Turn On

protoDUNE – single-phaseInstrumentation- GasAnalyzers

Proposal– topofrange:• Vaisala orequivalentDewPointMonitor• Mediumrange(DF310E)andhighsensitivity(DF550)oxygenanalyzers• Nitrogenanalyzer(Servomex Servopro orLDetek8000)• Highsensitivity(0.6ppb)(butnothighestsensitivity)H20analyzer(TigerOpticsHalo3)Modestproposal• OmittheH20analyzer

4/25/17 S.Pordes- protoDUNEsingle-phaseInstrumentationreview- gasanalzyzers 7

Input Output toReceived liquid Halo, DF310E, Servopro N2, fast purgeCryostat during purge DF310E, Servopro N2, Vaisala, Halo, fast purgeCryostat Liquid Halo, DF 550, Servopro N2, fast purgeCryostat Ullage Halo, DF 550, Servopro N2, fast purgeFeedthrough flow Halo, fast purgeNitrogen Calibration Servopro N2, fast purge

Table 1: Gas switchyard

3 Interfaces

The gas analyzer system involves interfaces with the cryogenics and with the controlssystem. The active equipment is outside the main cryostat and its operation is notexpected to interfere with the apparatus inside the cryostat.

Cryogenics

The cryogenics system provides the access points to the argon, and the piping fromthe access points to the switchyard just before the analytic instrumentation. Access isproposed to the line receiving argon from the vendor, to the liquid and to the ullagein the cryostat, to the main boil-o↵ line, and to the line returning the argon to thecrysotat. The total flow of argon through the analyzers will be of the order of 4 l/m(gas at room temperature) which translates to about 1.6 kg/day.

It is assumed that DUNE will provide the switchyard, pressure regulators, meteringvalves and any pressure-booster pump needed to satisfy the input pressure requirementson the analyzers. It is hoped that the Neutrino platform group will provide a smalldry pump for evacuation of lines as needed.

Controls

We will wish to record the outputs from the analyzers every minute or so in somedatabase. The standard readout provided is 4 to 20 mA but there are options for eth-ernet data transfer if desired. The actual implementation will be decided by discussionwith the DCS group. It is proposed that the state of the switchyard (which source isconnected to which analyzer) will be recorded by hand on an electronic form.

The analyzers plus switchyard will take about 1 standard relay rack of space whoselocation will be decided by discussion with DCS and the cryogenics group

4 Charge Status

In response to the charge:

Page 8: protoDUNE – single-phase Instrumentation - Gas Analyzers ... · Expanded region around Pump Restart after Power Outage 3/23/17 15 PrM1 &4 NanoTrace O2 Gas Analyzer Pump Turn On

protoDUNE – single-phaseInstrumentation- GasAnalyzers

Responsetocharge:Inresponsetothecharge:1.Thesetofinstrumentationproposedwillcovertheneedsforgasanalysisfromdeliverytopurgetooperations2.ItisenvisagedthatthesametypeofinstrumentationisappropriateforDUNE.ThelengthofDUNEmayrequireseveralsetsofinstrumentationdistributedalongthelengthofthecryostat.3.Themajoritemthatisnotastockcommercialdeviceisthevalveswitchyard.Thereisexperiencebothinbuildingsuchadevicein-houseandinpurchasingonesothetimestoprocureorconstructareknown.TheanalyzerscomewithcommercialwarrantiesandtheswitchyardcanbeconstructedandqualifiedatFermilaband checkedatCERN.TheitemscanbeshippedtoCERNasordinarycommercialobjects.Calibration/spangaseswillbeorderedandusedforinitialcommissioning.ItisassumedthatinstallationofthepipingtotheanalyzerrackwillbedonebytheNeutrinoPlatform.4.Thereisnoinstallationplanasyet.(April2017).ThisclearlyneedstoberesolvedwiththeNeutrinoplatform.5.TheconnectionstothecryogenicssystemaredefinedinthePipingandInstrumentationDiagram(EDMS).Thelocationoftheanalyzerrackandswitchyardareyettobedetermined.6.Theslowcontrolsystemneedstorecordthereadingsfromthesedevicesandpossiblyexerciseacalibrationcycle.Itisexpectedthatthemeasurementsfromtheanalyzerswillberecordedinadata-baseandaccessibleinsomehistorian.Thereadoutinterfaceshavenotbeendiscussedwiththeslowcontrolsystemandthisneedstobeunderstoodandagreedinthenext6months.

4/25/17 S.Pordes- protoDUNEsingle-phaseInstrumentationreview- gasanalzyzers 8

Page 9: protoDUNE – single-phase Instrumentation - Gas Analyzers ... · Expanded region around Pump Restart after Power Outage 3/23/17 15 PrM1 &4 NanoTrace O2 Gas Analyzer Pump Turn On

protoDUNE – single-phaseInstrumentation- GasAnalyzers

Responsetocharge(2)7.Weexpectthattheelectricalgroundingissuespresentedbytheexternalpipingusedforsamplingwillberesolvedbytheelectricalengineerswhentheydefinethegroundingfortheexternalpipingingeneral.8.Theequipmentisallexternaltothecryostatandisnotexpectedtointerferewithdetectoroperations.Wewillcheckifthepumpwhichmaybeusedtoboostthesamplelinepressuresproducesanyelectricalnoisethataffectsthedetectorreadout.Thesystemneedstobeoperatingparticularlyduringthegaspurgeandrecirculationphases.9.safetyofinstallationandoperationTheanalyticinstrumentationcarrieselectricalcertification.Installationoftheswitchyardandtheinstrumentationrequirescarefulleak-checkingbutisotherwisestraightforward.Onceinstalledandcommissioned,theoperationsburdenislimitedtosettingtheappropriatesourcesontheappropriatedevicesandsorecording,andperiodiccalibrationsoftheN2sensor.Whenvacuumpumpsareused,procedureswillbefollowedtoensurethatwedonotpullavacuumwhereitisnotwanted.10.TheproposalisbasedonexperienceatFermilab wheretheseinstrumentshavebeenusedextensively.WehaveconsultedthepeoplewhobuiltthesystemsatFermilab andintendtocontinuethisconsultationandtoincorporatetheresultsoftheirandlocal(CERN)experience.

4/25/17 S.Pordes- protoDUNEsingle-phaseInstrumentationreview- gasanalzyzers 9