project operations plan savage well site ri/fs ...project operations plan savage well site ri/fs...
TRANSCRIPT
-
PROJECT OPERATIONS PLAN
SAVAGE WELL SITE RI/FS
MILFORD, NEW HAMPSHIRE
October, 1988
VOLUME II
Prepared by:
HMM ASSOCIATES, INC.336 Baker Avenue
Concord, MA 01742
2176/HAZ/1028
-
TABLE OF CONTENTS • "iK- - _ _> ...-, pg-ge"
1.0 INTRODUCTION 1-1
1.1 B ackground and Site History 1-1
1.2 Goals and Objectives of the Groundwater Sampling Program 1-1
2.0 ANALYTICAL PARAMETERS 2-1
2.1 Summary of Previous Analyses 2-1
2.2 Data Quality Objectives 2-6
2.3 Analytical Parameters of Concern 2-8
2.4 Scope of Groundwater Sampling Program 2-14
2.5 Deliverables 2-16
3.0 GROUNDWATER SAMPLING PROCEDURES 3-1
3.1 Water Level Measurement 3-1
3.2 Well Purging 3-3
3.3 Groundwater Sample Collection 3-7
3.4 Sample Containers and Preservatives 3-11
4.0 DECONTAMINATION PROCEDURES AND SAMPLING EQUIPMENT 4-1
4.1 Equipment Decontamination 4-1
4.2 Personnel Decontamination 4-3
5.0 SAMPLE RECORDS AND CHAIN OF CUSTODY 5-1
5.1 Labeling of Sample Containers 5-1
5.2 Recordkeeping and Chain of Custody Documentation 5-1
5.3 Handling and Transportation of Samples to the Laboratory 5-3
6.0 SITE SAFETY CONSIDERATIONS 6-1
6.1 Field Monitoring and Screening 6-1
7.0 REFERENCES 7-1
2176-027/HAZ/333 -i-
-
1.0 INTRODUCTION
1.1 Background and Site History
In February of 1983, volatile organic compounds were detected in the Savage Well during
routine water quality monitoring by the New Hampshire Water Supply and Pollution Control
Commission (NHWSPCC).
In response to the contamination, hydrogeological investigations were initiated at the O.K.
Tool Company and Hitchiner Manufacturing Company facilities which are located near th
Savage Well. The Hydrogeological Investigation Unit of the Water Supply and Pollution
Control Commission designed and implemented a study of the Savage Well area in the summer
and fall of 1984.
The study revealed that the area is underlain by an unconfined, high yield, overburden
aquifer. Volatile organic compounds have been detected in the groundwater and surface water
near Savage WTell.
HMM Associates Inc. has been tasked to conduct a Remedial Investigation/Feasibility
Study (RI/FS) at the Savage Well Site in Milford, NH. The RI/FS is to be performed in
accordance with the Technical Scope of Work prepared by U.S. EPA Region I (EPA, 1986a) and
be consistent with the National Contingency Plan effective February 18, 1986 (NCP), with the
EPA RI/FS Guidance dated June, 1985, to the extent the RI/FS Guidance is consistent with the
NCP, with EPA's "Interim Guidance on Superfund Selection of Remedy" and with the
Superfund Amendments, and Reauthorization Act of 1986 (SARA). To the extent that the
Technical Scope of work is inconsistent with the NCP, the NCP shall govern.
1.2 Goals and Objectives of the Groundwater Sampling Program
The goals of the groundwater sampling program are to produce a set of groundwater
samples which are representative of the medium under consideration and are suitable for
subsequent laboratory analysis. This groundwater sampling program is designed to retain the
integrity of samples through collection, transport and delivery to the analytical laboratory.
The objectives of the groundwater sampling program are to:
• determine groundwater elevations at selected observation wells which will be used to
construct a groundwater contour map;
2176-027/HAZ/333 1-1
-
• determine the seasonal fluctuations in the water table elevations and flow gradients at
the site from selected wells;
• determine the concentration of volatile organic compounds at selected monitoring
wells in both the unconsolidated and bedrock aquifers.
Limitations of the groundwater sampling program:
• Groundwater samples and measurements are to be collected and analyzed only from
the Phase I group of monitoring wells. Based on the results of this sampling program,
additional monitoring wells (Phase II) may be required to meet the objectives of the
investigation.
2176-027/HAZ/333 1-2
-
2.0 ANALYTICAL PARAMETERS
2.1 Summary of Previous Groundwater Analyses
There is a substantial quantity of hydrogeologic information available from previous
investigations of the Savage Well Site area. The purpose and scope of the previous
investigations has varied from water supply exploration to contamination investigations at
specific locations within the site. Volatile organic contamination of groundwater is distributed
throughout the site area. Contamination has been found in an area extending from O.K. Tool
Company to the west, Hitchiner Manufacturing Company and Hendrix Wire & Cable Company
to the south and to approximately 1,100 feet east of Savage Well. Three smaller areas of volatile
organic contamination in low concentrations were detected. Two of the areas were detected near
New England Steel Fabricators. The third area was located near the Hitchiner sludge disposal
site on Perry Road.
Existing monitoring well locations, industrial facilities and significant site features and
characteristics are exhibited on Figure 2.1.
2.1.1 Volatile Organic Contamination
Previous investigations at the O.K. Tool Company facility indicated that there are several
areas containing high levels of volatile organic compounds which may be acting as significant
sources of groundwater contamination (NHWSPCC, 1985). The highest levels of volatile
organic compounds in the groundwater were detected downgradient from the O.K. Tool
Company facility. Samples from the north side of Elm Street have generally contained
tetrachloroethylene in the largest concentration with lesser amounts of trichloroethylene,
1,2-trans-dichloroethylene and 1,1,1-trichloroethane. Samples from the south side of Elm Street
have generally contained 1,1,1-trichloroethane . in the highest concentrations with
tetrachloroethylene and 1,1-dichloroethane detected at lower concentrations.
Volatile organic compounds have been detected in the Hendrix-Hitchiner discharge
stream. The volatile organic compounds detected include 1,1,1-trichloroethane and methyl
isobutyl ketone in the greatest concentrations and relatively low concentrations of
tetrachloroethylene, 1,1-dichloroethane and 1,1-dichloroethylene.
2176-027/HAZ/333 2-1
-
LUH;
y> xO. _| Z
HICC
LU Q5 a:
OO !Z
CO
EE
2-2
-
The Hitchiner production well has contained similar volatile organic compounds such as
1,1,1-trichloroethane, tetrachloroethylene, and 1,1-dichloroethane. The eastern extent of volatile
organic contamination of the groundwater was found to extend at least 1,100 feet east of the
Savage Well.
Groundwater samples were collected for volatile organic compounds from 41 groundwater
sampling stations by NHWSPCC on September 11-13, 1984, and October 29, 1984, with the
analysis of these samples performed in accordance with the U.S. EPA Contract Lab Program.
Using this analytical data, the aerial distribution of volatile organic compounds in the
groundwater is shown in Figure 2.2. This figure indicates that an area of contamination extends
from O.K. Tool Company to the west, Hitchiner Manufacturing Company and Hendrix Wire and
Cable Company to the South, to approximately 1,100 feet east of the Savage Well.
2.1.2 Inorganic Analyses
Six water quality samples were collected for the analysis of inorganic parameters on
September 11-13, 1984 by NHWSPCC. The analytical results are presented in Table 2.1, and a
review of the data indicates several significant findings.
1) The levels of aluminum, arsenic, barium and manganese are observed at higher
concentrations at station MI-42 than at the other sample locations.
2) Nickel was detected only at Station MI-30.
3) In comparing the inorganic analyses from Station MI-49, which is the surface water
station on the Souhegan River at the most upgradient portion of the site (background)
to all of the other sampling stations, only the following parameters are observed to be
above these background conditions:
Aluminum Iron
Antimony Manganese
Arsenic Nickel
Barium Vanadium
Copper
2176-027/HAZ/333 2-3
-
2-4
-
sH
î0) F-
•>-> F—
rg QJ3 3
U Ol•F- (T.
cr 'OV. 00oc
2, SU 91« U
u .e wa u o •e • •£ O "x Ou ^ u a» oo ca 4 « « « *» ,si >ncMistm n m < D ( M a e > / t < M « o O '» -O X b « -1 -. » « v » O -• V (M V ̂ CM
— • V C M < * « V v v
£ « tj t1 » a «S 3 si S
e
»J t n e > o « O F < < « m n « < M i A O > / t ( < « « e a « nO * J U ^ ^4 X « V V f * k ^ ( M V M V F 4 C M V I N )ta * 9 v — CM v vv9> U» «I »
b
1,J3 • « « i » i ~ « C M » > « r > » « tn n CM 101
b i C M i n « O c 9 < M < « n n o < M O F 4 i n F 4 « o o <
-
2.1.3 Acid and Base/Neutral Extractable Organic Compounds
Ten water samples were collected for acid and base/neutral extractable organic analysis by
NHWSPCC on September 12-13, 1984. Six samples were collected from groundwater stations,
the remaining four samples were collected at surface water stations. A total often semi-volatile
compounds were tentatively identified and the results are presented in Table 2.2. The results
indicate that a lack of consistency exists among the compounds identified in samples collected
along the Hendrix-Hitchiner discharge stream. That is, of the five samples collected along the
Hendrix-Hitchiner discharge stream, for any given compound that is detected, that compound is
not detected at any of the other sampling stations on the Hendrix-Hitchiner discharge stream.
2.2 Data Quality Objectives
The stated purpose of the Data Quality Objectives (DQOs) are to ensure that data of
sufficient type and quality are obtained to support remedial response decisions and accelerate
project planning and implementation. The specific DQOs of the Groundwater Sampling Plan are
to collect groundwater quality samples which are representative of the medium under
consideration and suitable for subsequent analysis. Achieving the DQOs will enable HMM
Associates and EPA to be in agreement on the basis for evaluating data sufficiency and
adequacy which result in decision making in other tasks of the Savage Well Site RI/FS.
2.2.1 Representativeness of Existing Groundwater Data
As stated previously, there is a substantial quantity of hydrogeologic information available
from previous investigations of the Savage Well Site area. A significant amount of this
hydrogeologic data has been derived from samples collected from, or measurements made in,
previously installed wells. Many of these previously installed wells may not be suitable for
groundwater monitoring purposes. The general construction requirements for acceptable
groundwater monitoring wells (including filter pack, placement of annular sealant, well intake
design, etc.) are described in U.S. EPA - TEGD, 1986b, U.S. EPA "Practical Guide for
Ground-Water Sampling," 1985 and U.S. EPA "Manual of Ground-Water Quality Sampling
Procedures", 1981. Following these guidelines and reviewing the construction specifications of
the existing wells, very few, if any, meet the necessary criteria which would allow samples
collected from such wells to be considered representative of the medium under consideration.
Furthermore, considering the high transmissivity values that have been calculated for the aquifer
and the pumping rates which are sustained by the Hitchiner production well and the
2176-027/HAZ/333 2-6
-
TABLE 2.2
Acid and Base/Neutral Extractable Organic CompoundsSavage Well Site, Mil ford
Results in mlcrograms per literMppb)
Site 'lumber/
Descript ion
Ml-5/Savage WellObservat ion Well *4
MI-7/Savage WellObservation We l l *7
MI-20/O.K. Tool Co.,M-lb
MI-26/O.K. Tool Co. ,M-5b
MI-30/Hitchiner Mfg. Co.,H-3
MI-42/H1tChiner Mfg. Co.,Stream Well "C"
Ml-49/Soughegan River -North Purgatory Road
Ml-51/Soughegan River - d/s ofconfluence with Hendrlx-Hitchiner discharge stream
Ml-52/Hitchiner dischargeat weir
Ml-53/Hendrix-Hitchinerdischarge stream d/s oftheater access
1 Tentatively identified coBlanks indicate compound
Date
9/13/84
9/13/84
9/13/84
9/13/84
9/12/84
9/12/84
9/13/84
9/13/84
9/12/84
9/>2/84
•pounds,not dttcc
hexa
hydr
o-2H
-aze
pin
2-on
e (c
ap
ro)a
cta
*)31
39
34
21
Tabulatad.
00I/I
i.3
3
72
68
tid r«
2-c
yclo
he
xen
1-on
e10
40
19
suits
benz
oic
aci
d
-
Hendrix production well, the groundwater quality conditions in the aquifer under investigation
are very dynamic. With these considerations in mind, groundwater quality data obtained by
HMM Associates during the conduct of the Savage Well Site RI/FS may not necessarily
correlate well with data collected during previous investigations.
2.3 Analytical Parameters of Concern
As part of the Remedial Investigation, this groundwater sampling program is designed to
characterize the nature and extent of contamination such that HMM and U.S. EPA will be able
to agree on the basis for conducting the Feasibility Study and for evaluating remedial
alternatives.
The parameters suggested for analysis in the Savage Well Site EPA Technical Scope of
Work include the following groups:
o Volatile Organic Compounds (Table 2.3)
o Hazardous Substance List (HSL) Compounds (Table 2.4)
• Volatile organic compounds
• Acid and base neutral extractable compounds
• Metals
Pesticides/PCBs
• Cyanide
o Secondary Drinking Water Regulations (Table 2.5)
Site specific information suggests that analysis of some of the parameters or groups of
parameters is not warranted. The rationale for including or deleting specific groups of
parameters for analysis is described below.
2.3.1 Volatile Organic Compounds
Based on previous investigations completed to date, the primary contaminants detected in
the aquifer include 1,1,1-trichloroethane, tetrachloroethylene, 1,1-dichloroethane and
1,2-trans-dichloroethylene, as well as other volatile organic compounds present at lower
concentrations (NHWSPCC, 1985). This groundwater sampling program will involve sampling
many of the proposed (Phase I) monitoring wells for volatile organics analysis by EPA Method
624 and/or the Hazardous Substance List (HSL) volatile organics by the Contract Lab Program
(CLP).
2176-027/HAZ/333 2-8
-
H M : ' . o o i
VOLATILE ORGANIC COMPOUNDS
(by EPA Method 8240)
ChloromethaneBromomethaueVinyl chlorideChloroe thaneMethylene chlorideTrichlorofluoromethane1,1 -Dichloroethene1,1 -Dichloroethanetrans-1,2-DichioroetheneChloroform1,2-Dichloroethane1,1,1 -TrichloroethaneCarbon tetrachlorideBrornodichloromethane1,2-Dichloropropanetrans-1,3 -DichloropropeneTrichloroetheneBenzeneDibromochlorotnethane1,1,2-Trichloroethanecis-J ,3-Dichloropropene2-Chloroethylvinyl etherBromofonn1,1,2,2-TetrachloroethaneTetrachloroetheneTolueneChlorobenzeneEthyl benzene1,3-Dichlorobenzene1,2-Dichloioberi2ene1,4-DichlorobenzeneXylenes - total
EPA 1986: SW-846 Test Methods for Evaluating Solid Waste
2176-027/HAZy333 2-9
-
TABLE 2.4
EPA CONTRACT LABORATORY PROGRAMHAZARDOUS SUBSTANCE LIST fHSL) COMPOUNDS
tolittttt
1.2.3.4.S.6.7.8.9.
10.11.12.13.14.IS.1(.17.18.19.20.21.22.23.it.25.2«.27.28.29.30.31.32.33.M.35.
St«i*0ltt11t« (CMtlMMd)
Chlorwwttun*Ntt«1t
Vinyl chlorideCMoro«th«neNethylene chlorideAcrtoneCtrbon dl»ulf1d»1,1-OlchloroetHenel.l-D1ehlorotth«neTrins-l,2-d1ch1eroetneneCMoroforn1.2-D1ch1oroetiMne2-lutine«e1.1.1-TrlchloroethineC«rbon mrachlerldtVinyl ictUteIroMOdlchloroMthtneI.l,2.?-Tetr»er1oroet»wne1.2-OlchloropropineTrtnt-1.3-d1ch1oropropeneTrUhloroethent01»roMch1oroMth«iwI,l.2-Tr1chloroethii»eBenzeneC1»-l,J-41ehloropropene2-Cnloroethyl vinyl tthtrtroriofom2-Heitnone4-MthyM-pwiunw*TetrichloroctnenfTolueneQilerobtnttotCthyl btnitntStyrtn*Totil
Swlvolitllts
36.37.38.39.40.41.42.43.44.45.46.47.48.49.SO.SI.52.53.54.55.56.$7.58.59.60.61.61.63.
8U(2-cMoro*ttty1 )itMr2-CMoropn«nol1 , 3-01 chl erobtnsent1. 4*01 chl oreb*«uint8tfisy1 ilcenel1 .2-01 ehl erebcflziiw2-Nethjrlphtnol9 1 1 ( 2-Clll erol lOprflpyl )iti«r
IMi1trmo»41prapytMliitHtuchlor«ttli«MIl1tr«b*fiien«Itopnerow2>N1tro«lit««ol2.4.01wthylplWMl
IU(2-ch10rwtHny)MtlMM2.4-01ehlorop*ne1l.2.4>Tr1cMorot»«tztAtKtphthiltnt
Htitchl oreb«U41 tn«4-Chl«ro-)H«thy1pK«2-mthyln«phth»lit»«
t»-ertiel)
65.6*.67.68.69.70.71.72.73.74.75.76.77.78.79.80.81.82.83.84.84.86.87.88.89.90.91.92.93.94.9$.96.97.9C.99.
100.
2-Ch1oren*pMHtlt«««2-H1tre«nnin»Dtiwthyl phthtltttActMphthyl*n«3-NUrointlliwActntphthtnt2,4.D1nitreph«ne14.N1tropn«nolD1b«niefuran2.4-01n1trotOlutfl*2.6>0(il1ti>0tplvtn«OltthylpktlMUte4-Chlor«ph«fiyl phtnyl tthcrFluercn*
4.6-01n1tro-2-«ttlivlpn««iolIUn1trpfod1pht«ylMrin*4.8rawp«wnyl ptwtiyl ttlwrHti*eti1erabt«iin«NntiChlorophtnol
AHthrtcint01-n-butylj*th«l»tfFluor»nthf««>yr«ntButyl btntyl phth«l«tt3.3'-01chlero6«flXld1nt8tnzo(a)inthricti««!1l(Z-«thylh«iyl)p»«th»l«tt
127. Alwlnw128.129.130. krlM131. 8trynii132. C«d*1uB133.134.135. Cob*It136. Copper137. Iron138. U«d139. N«fn«ilMi140. H«ng«nt*t141. Mercury142.143.144.145. S11v«r146. Sotflw*147. Th«l11uB14*. VlMdlUH149. Z1nc150. Cy»n1d»
01-n-ottyl phth«1»tt8ttizo(b)fVuowithtfl«8tnte(k)f1u«r«Ptn«flt8tftZoU)pyr*nt
01b»«zo(»,h)«nthrie»n«8«ize(9,h.1}ptrylti«t
N«t1c1tfn/FC8
Ut. *IpM-8HC102. BtU-IHC103. OtlU-MC1M. SMM-IHC105. Htpttchlar106. »ldr1n107. Htpuchler tpoildt101. Cndoittif*" 1109. Oltldrln110. 4.4--OOC111. Cndrln112. tnde»ulf»n 11113. 4, 4 '-000II*'. Endetulftn s«lf«tcus. 4.4-.OOT116. Endrln ketont117. Mttneiychler111. CMaradtnt119.120.121. MOCL08-1221122. MQCLW-123:123. AWXLOI-1242124. NUKLOi.1248in. MOCLOi-UM121. MOCLO^UM
Federal Register 50: 26632, July 24,1986.
2176-027/WPPHAZ/333 2-10
-
TABLE 2.5
NATIONAL SECONDARY DRINKING WATER
REGULATION PARAMETERS
Parameter
Chloride (Cl)
Color
Copper (Cu)
Corrosivity
Foaming agents (Surfactants)
Iron (Fe)
Manganese (Mn)
Odor
PH
Sulfate (SO4)
Total Dissolved Solids (TDS)
Zinc (Zn)
Federal Register 42:17144, March 31, 1977.
2176-027/HAZ/333 2-11
-
2.3.2 Acid and Base Neutral Extractahle Organic Compounds
Potential sources of ABN compounds exist in the plastics and/or vulcanization processes
utilized in the operations of Hitchiner Manufacturing Company and Hendrix Wire and Cable
Company. Low levels of ABN compounds including acetophenone were tentatively identified in
selected wells and surface water stations by NHWSPCC in 1984. Selected proposed (Phase I)
monitoring wells will be sampled and analyzed for acid and base neutral extractable organic
compounds and acetophenone following HSL methods under the Contract Lab Program.
A previous investigation conducted by NHWSPCC in 1984 included sampling several
surface water stations, both on the Souhegan River and the Hitchiner-Hendrix discharge stream,
for 20 metals. The results indicate that only a few of these metals are detected above
background conditions (i.e., upgradient of the site on the Souhegan River). These metals are:
aluminum, antimony, arsenic, barium, copper, iron, manganese, nickel, and vanadium.
The NPDES permits for both Hendrix Wire and Cable Company and Hitchiner
Manufacturing Company require that several constituents be analyzed on a routine basis to
monitor the quality of wastewaters generated at each facility. The metals which may be present
in wastewater generated at each facility which has been included in the NPDES monitoring
programs include:
Cadmium IronChromium LeadHexavalent Chromium NickelCopper Zinc
Based on the previous metals analyses and the NPDES permit requirements, the following
group of metals will be included for analysis in this Groundwater Sampling Plan and for the
purposes of this plan will be referred to as the "Savage Well Site Metals".
Aluminum IronAntimony LeadArsenic ManganeseBarium NickelCadmium VanadiumChromium ZincCopper
2176-027/HAZ/333 2-12
-
2.3.4 Pesticides. PCBs and Cyanide
Based on the operational activities and site history information compiled to date, there is
no evidence which would suggest that pesticides, PCBs or cyanide would be components of
wastes generated by Hendrix Wire and Cable Company, Hitchiner Manufacturing Company,
O.K. Tool Company, or New England Steel Fabricators. It is anticipated that any detection of
pesticide or PCB compounds may be spurious and noncorrelatable to the volatile organiccompound contaminant plume that is known to exist. Because no rational basis exists forsuspecting pesticide, PCB or cyanide contamination at the four identified PRP facilities, these
parameters will not be included for analysis in the Groundwater Sampling Plan.
2.3.5 Secondary Drinking Water Regulation Parameters
The secondary drinking water regulation contaminant levels established by the Safe
Drinking Water Act are set for aesthetic purposes. The group of parameters included in these
regulations are those which may adversely affect the aesthetic qualities of drinking water, such
as taste, odor, color and appearance, and which thereby may deter public acceptance of drinking
water provided by public water systems. At considerably higher concentrations, these
contaminants may also be associated with adverse health implications. These secondary levelsrepresent reasonable goals for drinking water quality, but they are not federally enforceable. All
of the Secondary Drinking Water Regulation compounds are naturally occurring parameters, the
levels of which, in groundwater, are dependent on a variety of geological, hydrological, and
meteorological conditions. The levels of these parameters may have little correlation with the
hazardous or nonhazardous wastes generated and disposed of at any industrial facility.
The Savage Well Site EPA Technical Scope of Work requests that this group of
parameters be sampled and analyzed from selected monitoring wells. There does not appear to
be a rational basis for requesting this group of parameters to be analyzed, however, a few of the
metals (copper, iron, manganese) from this group have been previously identified in surface
waters samples and are included in the Metals Analysis (Section 2.3.3) described previously.None of the other Secondary Drinking Water Regulation Parameters will be included for
analysis in this Groundwater Sampling Plan.
2176-027/HAZ/333 2-13
-
2.4 Scope of Groundwater Sampling Program
Fifteen (Phase I) groundwater monitoring well clusters are scheduled to be installed as
part of the groundwater monitoring well installation program (see Figure 2.3). A minimum
two-week well stabilization period must occur prior to collecting groundwater quality samples.
All groundwater samples will be collected in accordance with the procedures contained in this
plan.
• All Phase I well couplets (MW-1 through MW-15) will be sampled for volatile
organic compounds by EPA Method 624. This is anticipated to be approximately 30
samples (15 Phase I well couplets) unless more than two wells are installed at any
particular Phase I well location.
• The results and interpretation of the Phase I well sampling and analysis will be
summarized in a letter report and submitted to EPA for review. Based on the
sampling results and interpretations made by HMM Associates, up to twenty Phase I
monitoring wells will be selected in concurrence with EPA for sampling of: HSL
volatile organic compounds; ABN extractable compounds (five of which will include
analysis of acetophenone); and the Savage Well Site metals.
• The results and interpretation of the above described volatile organics, ABNs and
metals analyses will be summarized in a letter report which will be submitted to the
EPA for review. Based on the sampling results and interpretations made by HMM
Associates, the scope of the Phase n monitoring well installation program will be
negotiated with EPA. Upon completion of the Phase n well installation program, all
of the Phase n monitoring wells will be sampled for volatile organic compounds by
• EPA Method 624.
• The Phase II volatile organic compound well sampling results and interpretation will
be summarized in a letter report and submitted to EPA for review. Based on all
groundwater data collected by HMM Associates in conjunction with the RI/FS,
fifteen of the Phase I and/or Phase n monitoring wells will be selected, in
concurrence with EPA, for additional sampling. To monitor seasonal variations in
groundwater level, groundwater quality, and to characterize and predict how
contaminant migration will respond in the future, the fifteen selected monitoring
wells will be sampled and analyzed for volatile organic compounds by EPA Method
624 on a quarterly basis for one year. This will involve four sampling rounds of the
fifteen selected monitoring wells. The results and interpretations will be summarized
in letter reports and submitted to the EPA.
2176-027/HAZ/333 2-14
-
2-15
-
2.5 Deliverables
The above described letter reports, which include the groundwater sampling results,
interpretation of the data, a map of water level elevation data and a map exhibiting the
concentration of volatile organic compounds in the overburden and bedrock aquifers, will be
submitted to EPA at the completion of each sampling round.
2176-027/HAZ/333 2-16
-
3.0 GROUNDWATER SAMPLING PROCEDURES
The objectives of the groundwater sampling program are to obtain accurate groundwater
level measurements and to collect samples which are representative of the groundwater at the
screened interval of the subject monitoring well. Special precautions must be taken to ensure
that the samples collected are representative of the groundwater at that screened interval and that
the sample is neither altered or contaminated by the sampling, handling, storage, and
transportation procedures.
3.1 Water Level Measurement
Obtaining accurate water level measurements from each of the proposed groundwater
monitoring wells is necessary to facilitate construction of a groundwater contour map of the site
which will exhibit approximate direction of groundwater flow, hydraulic gradients and
surface/groundwater interactions. It is important to obtain water level measurements before
installing purge or sampling pumps in the well.
Accurate groundwater level measurements will be obtained by one of the following
methods:
3.1.1 Chalked Steel Tape
A narrow stainless steel tape which is graduated with raised lettering throughout its entire
length in feet, tenths, and hundredths of a foot is among the most accurate of methods for
obtaining groundwater level measurements (USGS, 1980). The bottom few feet of the tape is
chalked and lowered into the well to the anticipated water level depth so that the chalked portion
of the steel tape intercepts the groundwater level. A bar is to be placed across the top of the
protective casing so that the steel tape can be held at,an even foot mark which can be accurately
referenced to the measuring point which is to be clearly marked on the northern side of the
protective casing. The tape is then withdrawn and the demarcation on the chalked tape indicates
the water table surface. The length of the wetted portion of the tape is subtracted from the foot
reading held at the measuring point. Measurements taken in this manner are generally accurate
to the nearest one hundredth of a foot. Three readings to within 0.01 feet will be recorded for
each measurement to ensure that reproducible results are obtained.
2176-027/WPPHAZ/333 3-1
-
3.1.2 Fiberglass Tape with Popper
A fiberglass tape which is graduated throughout its entire length in feet, tenths, and
hundredths of a foot with a stainless steel popper attached to the bottom can be an easy method
for obtaining accurate water level measurements depending on the depth to groundwater and the
position of the water table with respect to the top of well screen. A bar is to be placed across the
top of the protective casing so that the fiberglass tape can be referenced to the measuring point
on the protective steel casing. The fiberglass tape should be lowered into the well until the
stainless steel popper contacts the water table. The water table surface is detected acoustically
by the popping sound created by trapping air within the concave popper. The fiberglass tape is
to be lowered and raised repeatedly in smaller increments until the tape can be held and
referenced to the measuring point on the protective casing. The depth to the water table surface
is then determined by adding the length of the attached popper to the recorded value on the
fiberglass tape. Three readings to within 0.01 foot will be recorded to ensure that reproducible
results are obtained.
3.1.3 Interface Probe
An electrical interface probe can be used to obtain accurate water level measurements
where Non-Aqueous Phase Liquids (NAPL's) are known or suspected to exist. The interface
probe is capable of detecting either polar or nonpolar liquids and, therefore, can be used both to
measure the depth to groundwater and the thickness of NAPL's present in a monitoring well.
The interface probe to be used is an Oil Recovery Systems (ORS) Interface Probe. The
ORS probe utilizes an intrinsically safe, electrically operated probe which, when submerged in
liquid, distinguishes water from NAPL's by measuring the conductivity of the liquid. The
interface probe is attached to a teflon coated tape which is graduated in feet, tenths and
hundredths of a foot. With the instrument in the on position, the probe is to be lowered inside
the well casing until a pulsating tone is audible on the tape reel assembly. The pulsating tone is
indicative of a polar liquid (i.e., groundwater). The tape should be raised and lowered repeatedly
at smaller increments until the level of the water table surface is identified by the audible
pulsating tone. A bar placed across the top of the protective well casing will provide a reference
to the measuring point on the well casing. A continuous tone is audible when the probe is
submerged in NAPL's. By raising and lowering the interface probe, the depth and thickness of
NAPL's can easily be obtained. Three readings to within 0.02 foot should be recorded to ensure
that reproducible results are obtained.
2176-027 /WPPHAZ/333 3-2
-
3.1.4 Recording Water Level Measurements
Any of the above described methods are capable of yielding accurate data, however, one
method may be preferred under various circumstances. If NAPL's are known or suspected to
exist, the interface probe method should be used, If, in the judgement of the sampler making
measurements, the fiberglass tape method does not produce an adequate popping sound, the
chalked steel tape or interface probe methods are acceptable alternatives. All measurements will
be recorded on an HMM Groundwater Sampling Report form (Figure 3.1).
3.2 Well Purging
Before collecting samples from groundwater monitoring wells, each well will be
adequately purged to remove stagnant water from the well which may not be representative of
the overall groundwater quality at that sampling site. The length of the water column and the
volume of water contained in the monitoring well will be calculated and recorded on the HMM
Groundwater Sampling Report form.
The types of instruments which will be used to purge stagnant well water may include a
316 stainless steel bailer, a peristaltic pump fitted with teflon tubing, a stainless steel steel/teflon
bladder pump, a positive displacement purge pump, a gasoline powered 1.5" centrifugal pump,
or a hand operated diaphram pump. The selection of well purging instrument will be based on
several criteria: 1) the depth to groundwater; 2) the length of the water column; 3) the volume
of water in the well; 4) the anticipated volume of water to be purged prior to sampling; and 5)
the recharge characteristics of the monitoring well.
The flow rate of the peristaltic pump can generally be adjusted more easily, and to lower
flow rates, than the bladder or purge pumps and pumps efficiently with very little submergence
of the teflon tubing. Therefore, the peristaltic pump would be desired for purging wells which
are relatively shallow, have shorter water columns, and have very low recharge rates. The
stainless/teflon bladder pump, the purge pump apparatus, the centrifugal pump and the diaphram
pumps are better suited for deeper wells with high recharge rates where large volumes of water
are anticipated to be purged.
In higher yielding wells, it will be more practical to utilize a more efficient well purging
device. The 1.5" centrifugal pump (up to 30 gpm), positive displacement purge pump (up to 5
gpm), the hand operated diaphram pump (up to 2 gpm) or the positive displacement bladder
pump (up to 0.5 gpm) are all acceptable devices for purging stagnant water from a monitoring
well prior to collecting water quality samples. The basis for selecting one purging device over
2176-027/WPPHAZ/333 3-3
-
HMM ASSOCIATES INC.
MONITORING WELL NO..
SAMPLE I.D.
GROUNDWATER SAMPLING REPORT
DATE.
SHEET OF
DIAMETER OF WELL: (FT) RADIUS OF WELL (R) :. .(FT)
WATER LEVEL MEASURING DEVICE:
DECONTAMINATION PROCEDURES OF DEVICE:
DEPTH TO GW BELOW MEASURING POINT (d) :.
TOTAL DEPTH OF WELL BELOW MEASURING POINT (D):.
LENGTH OF WATER COLUMN (L): (D-d)=
.(FT)
.(FT)
.(FT)
VOLUME OF WATER COLUMN (V): (3.14xRxRxL).
WELL VOLUME: (7.48xV)=
TYPE OF PURGE PUMP:
.(CUBIC FT)
(GAL)
TYPE OF SAMPLE PUMP:
DECONTAMINATION PROCEDURES OF PUMP:
TIME pH TEMP. (deg.C) Sp. COND. (umhos/cm) VOLUME (GAL)
(PURGE UNTIL pH, TEMPERATURE AND CONDUCTIVITY STABILIZE)
TOTAL VOLUME PURGED: .(GAL)
ANALYTICAL PARAMETERS:
COMMENTS:
2176-027/WPPHAZ/333
FIGURE 3.1
3-4
-
F
If
c
IMM ASSOCIATES
IONITORING WELL
JAMPLE I.D.
TIME
INC.
GROUNDWATER SAMPLING REPORT
NO . DATE
SHEET OP
pH TEMP, (deg C) Sp. COND. (umhos/cm) VOLUME (GAL)
2176-027/WPPHAZ/333
FIGURE 3.1 (continued)
3-5
-
another is dependent on the hydrologic criteria of the monitoring well described above. Recent
studies reveal that purging a well over a wide range of pumping rates will yield the same results
(NCASI 1982). The on-site hydrogeologist responsible for supervising sample collection
activities will make a judgment on the most efficient and practical well purging device. The 316
stainless steel bailer is among the most versatile purging instrument because it is controlled
entirely by the operator. That is, it can be used to purge all types of wells, however, the operator
must be very careful not to agitate the water table when lowering the bailer down the well.
Excessive agitation of the water in the well with the bailer may increase the oxidation potential
and cause aeration of volatile organic compounds.
At monitoring wells completed in high yield formations, generally three to five well
volumes will be purged prior to sampling the well. Field measurement of groundwater quality
will be conducted to ensure that stagnant water has been purged from the well and that water
representative of the groundwater conditions in the aquifer at that sampling location is being
drawn into the well for sampling. In practical terms, monitoring wells are considered adequately
purged when the indicator parameters, temperature, pH, and specific conductance, have
stabilized (USGS 1980). These measurements may be affected by exposure to the atmosphere,
but still provide the best means of determining when representative formation water has reached
the well (EPA, 1987). Once these parameters have stabilized (less than 10 percent fluctuation
for three successive readings), the water drawn into the well should be representative of the
groundwater conditions and the well is ready to be sampled.
The increment between successive measurements of the indicator parameters will be
dependent on the volume of water in the monitoring well. Using the information on the HMM
Groundwater Sampling Report form, the volume of water in the well should be multiplied by
three to obtain the volume of water which, in general, would be the minimum volume to be
purged. This value should be divided by six to obtain the approximate volume increment for
measuring the indicator parameters. Measurement of the indicator parameters should be made at
commencement of purging activities. At least six measurements of the indicator parameters
should be made in order to determine when the well has been adequately purged. The purged
well water should be contained in a bucket or some other apparatus to determine approximate
total volume purged from the well. Purged water should be disposed of on the ground surface
such that the water can infiltrate (EPA, 1987). Purge water should not be disposed into surface
waters or the well (EPA, 1987).
2176-027/WPPHAZ/333 3-6
-
The indicator parameters will be measured in the field with an Orion SA-230 temperature
compensating pH meter and a YSI Model 33 Salinity-Conductivity-Temperature (SCT) Meter.
The Orion pH meter and the YSI SCT meter are to be calibrated in accordance with the
manufacturer's specifications.
At monitoring wells completed in low yield formations which recharge very slowly, the
procedures for measurement of indicator parameters should be initiated. If the well is purged
completely, the purging activity should be ceased and the well should be allowed to recharge. If
recovery time is sufficiently short, the well should be allowed to recover completely, then
purging and measurement of indicator parameters should be continued, if possible, until
stabilization has been achieved. If recovery time is long, the well should be allowed to recharge
until a sufficient volume of water exists in the well that the sampling apparatus can be
completely purged prior to collecting the samples.
All field measurements of the indicator parameters, well recharge characteristics (i.e., very
slow recharge which precludes continuous purging) and the approximate total volume of water
purged from the well should be recorded on the HMM Groundwater Sampling Report form. The
well purging apparatus will be removed and decontaminated in accordance with the
decontamination procedures in Section 4.0 of this plan unless the purging apparatus is to be used
for sample collection.
3.3 Groundwater Sample Collection
Water samples should be collected when the chemistry of the groundwater being pumped
has stabilized as indicated by pH, specific conductivity, and temperature (EPA, 1985). The
sampling apparatus to be used to collect groundwater samples will be either a positive
displacement stainless steel/teflon bladder pump (QED Environmental Systems Well Wizard
Model T-1200), a peristaltic pump fitted with teflon tubing and a 1000 ml teflon sample trap
(Geotech Environmental Equipment Series I GeoPump) or a stainless steel/teflon bailer. In
general, groundwater samples should be withdrawn with the positive displacement bladder pump
or bailer unless the length of the water column in the well is short (i.e., less than three feet) and
the monitoring well recharges very slowly. If these conditions exist, see Section 3.3.2 of this
plan for sample withdrawal with the peristaltic pump. During all phases of the groundwater
sample collection procedures, all sampling team members are to wear clean (decontaminated)
nitrile gloves. All sampling equipment must be decontaminated in accordance with the
procedures in Section 4.0 of this plan before they are lowered into the monitoring well.
2176-027/WPPHAZ/333 3-7
-
3.3.1 Sample Withdrawal with the Positive Displacement Pump
The positive displacement pump should be lowered to the screened interval of the
monitoring well. If the sampling pump was not used to purge the well, well water should be
pumped through the apparatus until the positive displacement pump has been completely
purged. With the 1.5" O.D. stainless steel pump fitted with 100 feet of 0.375" I.D. teflon tubing,
the volume necessary to completely purge the apparatus is approximately 1.7 gallons.
3.3.1.1 Sampling for Volatile Organic Compounds
Once the sampling pump has been completely purged, the discharge rate of the pump
should be adjusted to achieve a flow rate low enough to fill a 40 ml volatile organic analysis
(VGA) vial without aerating the sample. Every effort should be made to obtain a slow, steady,
nonaerated stream of water so that aeration and volatilization of the water will be minimized.
The sample will be rejected and collected with a bailer if the discharge from the bladder pump is
non-continuous or turbulent. Collect the samples by holding the vial at an angle so that aeration
is minimized. Avoid touching the lip of the vial or the teflon liner in order to reduce potential
contamination of the sample. If the sample cannot be collected directly in the vial, a stainless
steel or teflon trap should be filled and directly transferred to the 40 ml vial (Holden, 1984).
Fill the vial until a positive meniscus forms at the top of the vial. Holding the vial upright
with one hand, use the other hand to replace the teflon lined cap without trapping air between the
sample and the septum. The cap should be screwed on tightly, but not so tightly that the glass
threads on the sample vial may fracture or break. Three vials will be collected at each sampling
location.
Examine the vial carefully for air bubbles by turning the vial upside down and tapping it
gently. If small air bubbles are present in the sample, or if the glass threads have been fractured
or damaged, discard the sample and begin the procedure again. Continue until no air bubbles
(zero headspace) are present in any volatile organic samples. This process is very important
because if volatile organic compounds are present in the sample, they have the potential to
volatilize into the headspace and subsequent analyses may provide erroneous and misleading
results.
When certain that zero headspace has been achieved, identify the sample by completing
the Sample I.D. label with an indelible ink marker. Place the three VOA vials into a "ziploc"
storage bag and immediately place the samples in a storage chest with sufficient ice to maintain
a temperature of 4°C or less. Pertinent information and comments are to be recorded on the
HMM Groundwater Sampling Report form.
2176-027/WPPHAZ/333 3-8
-
Z - ' - - I L : _ 1 F F ' i ' H f r 1 M -• - • ' ' - F H G E . 0 0 .-,
3.3,1.2 Sampling for Other Required_Paramet.er.S
The general sampling procedures for other required parameters are similar to that of
volatile organic compounds except that such extreme care is not required to minimize aeration
and volatilization of the sample. Therefore, the flow rate of the positive displacement bladder
pump can be adjusted to achieve the maximum discharge per cycle (i.e., approximately 250 ml
to 300 ml). Samples should be collected in the appropriate precleaned and preserved sample
containers (see Section 3.4). The sample containers will have a label affixed by laboratory
which states the types of analyses to be performed and the preservative, if any, contained in the
sample bottle.
The sample should be collected in the appropriate sample container, taking care to avoid
touching the lip of the sample bottle to the teflon discharge tube of the bladder pump to reduce
the potential for contaminating the sample. Sample bottles and containers will be filled nearly to
the top using care not to overflow the sample bottle where loss of the required preservative could
occur. Samples collected for metals analysis will be filtered on site as soon as practically
possible after collection and prior to the addition of preservative. The sample bottle is to be
sealed tightly with the appropriate cap and required information provided on the sample ID label
using an indelible marker. Pertinent information and comments are to be recorded on the HMM
Groundwater Sampling Report form. At the completion of sample collection, all sample
collection apparatus must be decontaminated in accordance with Section 4.0 of this Groundwater
Sampling Plan.
3.3.2 Sample Withdrawal wiih the Peristaltic Pump
The positive displacement stainless steel/teflon bladder pump or bailer should be used to
withdraw groundwater samples from all monitoring wells unless the length of the water column
in the weH is short (i.e., less than three feet) and the recharge rate of the monitoring well is very
slow. If use of the positive displacement pump is impractical, a peristaltic pump fitted with 1/4"
O.D. teflon tubing and a 1000 ml teflon sample trap will be used to withdraw groundwater
samples. All sampling equipment must be decontaminated in accordance with the procedures in
Section 4.0 of this plan before they are lowered into the monitoring well.
3.3.2.1 Samplejyilhdrawal for Volatile Organic Compounds
A clean length (approximately 2 feet) of flexible silicon tubing is to be fitted inside the
QED Bladder pomp head. The silicon tubing only needs to be long enough to fit a piece of 1/4"O.D. teflon tubing into the intake end and Jong enough to direct the discharge into an
2I76-027/WPPHAZ/333 3-9
-
: j 0 F P 0 P h 11 r! H S I-, 0 r p H G E . 0 0 2'
appropriate backet to measure discharge if required. An approximately 18" length of teflon
tubing is connected from the bladder pump to one side of the 1000 ml teflon sample trap and a
25 foot length of 1/4" O.D. teflon tubing is connected to the other side of the 1000 ml teflon
sample trap. The 25 foot length of teflon tubing is lowered down the monitoring well until it is
fully extended or until the end of the tube coincides with the top of the monitoring well screen.
Excess teflon tubing should be coiled and placed aside so that it does not interfere with the
sample collection process. The bladder pump should be turned on in the forward direction so
that groundwater is drawn up the 25 foot length of teflon tubing, filling the 1000 ml teflon trap,
and discharging out the flexible silicon tubing on the other end of the bladder pump head.
Once the 1000 teflon trap has completely filled, adjust the flow rate of the pump to the
lowest setting. Turn the pump to the off position and remove the 25 foot length of teflon tubing
from the well, coiling it as it is removed to keep the tubing orderly and manageable. Hold the
cofl in a comfortable position in one hand so that the end of the tubing can be easily controlled
for the sample collection process. With the other hand, reverse the flow of the bladder pump so
that the sample will now flow out the end of the teflon tubing. Place a precleaned 40 ml volatile
organic analysis (VOA) vial, with the cap removed, in an upright position oil the control panel of
the bladder pump. Turn the pump on and off so that a few milliliters of water are discharged
onto die ground. This will purge excessive air that may exist in the end of the teflon tubing.
Every effort should be made to obtain a slow, steady, nonaerated stream of water so that aeration
and volatilization of the water will be minimized. Collect the samples by holding the teflon
tubing at an angle to the vial. Avoid touching the lip or sides of the vial to reduce potential
contamination of the sample.
Fifl the vial until a positive meniscus forms at the top of the vial. Holding the vial upright
with one hand, use the other hand to replace the teflon lined cap without trapping air between the
sample and the septum. The cap should be screwed on tightly, but not so tightly that the glass
threads on the sample vial may fracture or break. Three vials will be collected at each sampling
location.Examine the vial carefully for air bubbles by turning the vial upside down and tapping it
gently. If small air bubbles are present in the sample, or if die glass threads have been fractured
or damaged, discard the sample and begin the procedure again. Continue until no air bubbles
(zero headspace) are present in any volatile organic samples. Tlu's process is very important
because if volatile organic compounds are present in the sample, they have the potential to
volatilize into the headspace and subsequent analyses may provide erroneous and misleading
results.
2176-027/WPPHAZ/333 3-10
-
12 ' oc i^ :.-.-! f - 'PL'M HUM1 HbbOV. P f t b E . O Q l
When certain that zero hcadspace has been achieved, identify the sample by completingthe sample I.D. label with an indelible ink marker. Place the three VOA vials into a "zlploc"storage bag and immediately place the samples in a storage chest with sufficient ice to maintaina temperature of 4DC or less. Pertinent information and comments will be recorded on the HMMGroundwater Sampling Report form.
3.3.2,2 Sample Withdrawal for Other Parameters
The general sampling procedures for other required parameters involve collection ofsample into the 1000 ml teflon sample trap and transfer of the sample to the appropriate samplecontainer. Lower the 25 foot length of teflon tubing into the monitoring well to the same depthas that for the volatile organic analysis samples. Turn the pump on in the forward position andadjust the flow rate to the maximum discharge position. Once the 1000 ml teflon trap is full,torn the pump off. Unscrew the base of the trap from the lid and pour the contents into theappropriate precleaned and preserved sample containers (see Section 3.4). The samplecontainers will have a label affixed by the laboratory which states the type of analyses to beperformed and the preservative, if any, contained in the sample bottle.
The sample should be collected in the appropriate sample container, taking care to avoidtouching the teflon sample trap to the lip of the sample bottle to reduce potential contaminationof the sample. Sample bottles and containers will be filled nearly to the top, using care not tooverflow the sample bottle where loss of the required preservative could occur. The samplebottite is to be sealed tightly with the appropriate cap and required information provided on thesample I.D. label using an indelible marker.
Screw the teflon sample trap back onto its lid and turn the peristaltic pump on until the1000 ml trap is filled whh sample. Repeat the above process until all required samples havebeen collected- Pertinent information and comments will be recorded on the HMM GroundwaterSampling Report form. At the completion of sample collection, all sample collection apparatusmust be decontaminated in accordance with Section 4.0 of this Groundwater Sampling Plan.
2176-027/WPPHAZ/333 3-11
-
3.3.3 Sample Withdrawal With a Bailer
Bailers used for collecting groundwater quality samples shall be constructed of stainless
steel and/or teflon materials. Selection of bailer wire should take into account both material
(teflon coated wire is preferred) and length (sufficient to reach interval of concern). The bailer
and bailer wire should be decontaminated prior to use and between all sampling locations, and
care should be exercised to ensure that'the wire is kept away from the ground or other sources of
contamination.
In order to minimize disturbance and potential volatization of the sample, the bailer should
be lowered slowly, using a hand-over hand motion or a bailer wire reel to the screened interval
of the well. Once the bailer has been immersed at the screened interval, allow several seconds
for the bailer to fill. Then slowly pull the bailer up, again using a hand-over-hand motion or a
bailer wire reel. As the bailer nears the top of the well casing, caution should be exercised to
prevent the bailer from being withdrawn too rapidly.
3.3.3.1 Sampling for Volatile Organic Compounds
As the bailer is withdrawn, the VOA vials should be opened and held near the top of the
casing (or other convenient location) so that the sample can be decanted into the vials as soon as
possible to minimize volatilization of sample constituents. The sample should be transferred
from the bailer to the vial carefully so that aeration or turbulence of the sample does not occur.
If turbulent flow occurs during the transfer process, the sample will be rejected and the sampling
procedure will be repeated. After ensuring that zero headspace remains in the vials (as outlined
in Section 3.3.1.1), the vials should be tightly capped, labelled, and placed in a storage chest
with ice.
3.3.3.2 Sample Withdrawal for Other Parameters
The general sampling procedures for other required parameters involve collection of
sample with the bailer as described above and transferring the sample into the appropriate
sample container, taking care to avoid touching the bailer to the lip of the sample bottle to
reduce potential contamination of the sample. Sample bottles and containers will be filled
nearly to the top, using care not to overflow the sample bottle where loss of the required
preservative could occur. The sample bottle is to be sealed tightly with the appropriate cap and
required information provided on the sample I.D. label using an indelible marker.
2176-027/WPPHAZ/333 3-12
-
Pertinent information and comments will be recorded on the HMM Groundwater
Sampling Report form. At the completion of sample collection, all sample collection apparatus
must be decontaminated in accordance with Section 4.0 of this Groundwater Sampling Plan.
3.4 Sample Containers and Preservatives
All samples will be collected in bottles, vials and containers prepared and supplied by
Alliance Technologies Corporation. Where appropriate, necessary preservatives will be added
to the sample containers by Alliance. The recommended sampling container, preservative,
maximum holding time, and minimum volume required for analysis for each group of
parameters is listed on Table 3.1.
2176-027/WPPHAZ/333 3-13
-
OSWER-9950.1
TABLE 3.1
SAMPLING AND PRESERVATION PROCEDURES FOR DETECTION MONITORING*
ParameterRecommended
Container13Maximum
PreservativeHolding Time
Minimum VolumeRequired for
Analysis
PH
Specific conductance
TOC
TOX
Indicators of Ground-Water Contamination*-
T, P. G Field determined None 25 ml
T, P. G Field determined None 100 ml
G. amber. T-l1ned Cool 4ac,d 28 days 4 x 15 mlcap" HC1 to pH
-
TABLE 3.1 (continued)
SAMPLING AND PRESERVATION PROCEDURES FOR OETECTION MONITORING
Parameter
EndrlnLindaneMethoxychlorToxaphene2.4 0
2.4.5 TP SHvex
RadiumGross AlphaGross Beta
. . . „ , Minimum VolumeReconroended _ .4 MaximumPreservative Required forContainer" Holding Time Analysis
T. G Cool. 4°C 7 days 2,000 ml
•
P. G Field acidified to 6 months 1 gallonpH 12. 0.6 gascorbic addr
G only
T, G
G, T-l1ned
Cool. 48C H2S04 to 28 dayspH
-
OSWZR-9950.1
TABLE 3.1 (continued)
SAMPLING AND PRESERVATION PROCEDURES FOR DETECTION MONITORING
C8ased on the requirements for detection monitoring (§265.93). the owner/operator mustcollect a sufficient volume of ground water to allow for tht analysis of four separatereplicates.
^Shipping containers (cooling chest with 1ce or ice pack) should be certified as to the 4actemperature at time of sample placement Into these containers. Preservation of samplesrequires that the temperature of collected samples be adjusted to tht 4°c Immediately aftercollection. Shipping coolers must be at 4°C and maintained at 4°C upon placement of sampleand during shipment. Maximum-minimum thermometers art to be placed Into the shipping chestto record temperature history. Chaln-of-custody forms v*111 have Shipping/Receiving andIn-transit (max/mm) temperature boxes for recording data and verification.
80o not allow any head space m the container.
'use ascorbic acid only in the presence of oxidizing agents.
^Maximum holding time 1s 24 hours wnen sulflde 1s present. Optionally, all samples may betested with lead acetate paper before the pH adjustment In order to determine if sulflde ispresent. If sulflde 1s present. H can be removed by addition of cadmium nitrate powderuntil a negative spot test is obtained. The sample 1s filtered and then NaOH 1s added topH 12.
3-16
-
4.0 DECONTAMINATION
4.1 Equipment Decontamination
To meet the objectives of RI/FS, it is imperative that every effort be made to collect
samples which are representative of .the medium under consideration. Decontamination of
sampling equipment is of utmost importance in preventing cross contamination and in
maintaining the integrity of this Groundwater Sampling Plan.
This discussion on decontamination applies to portable sampling equipment that is used at
more than one well or sampling location. This equipment must be thoroughly cleaned prior to
collecting each sample to minimize the potential for the sampling equipment to be a source of
contamination.
Decontamination requirements are reduced for dedicated or single-use disposable
sampling equipment. Dedicated or disposable equipment should be handled in such a way that it
does not come into contact with dirty and potentially contaminated surfaces. All sampling team
members are required to wear nitrile gloves during all aspects of the decontamination
procedures. Gloves are to be decontaminated utilizing hand held spray bottles with 1) an
alconox detergent solution, 2) 20 percent methanol solution, and finally 3) deionized water.
Wipe gloves dry with a clean disposable absorbant towel.
4.1.1 Decontamination of Water Level Measurement Devices
Remove the water level measurement device from the monitoring well, decontaminating
the length of the tape which was lowered into the well. The decontamination procedure is as
follows:
• Saturate a clean paper towel with 20 percent methanol solution. All decontamination
solution solvents should be reagent quality or higher. Water used to make up the
solution should be deionized.}
• Wipe along the length of the tape, discarding and replacing the towel as it becomes
soiled.
• Utilizing a hand-held spray bottle, douche the bottom few feet of the stainless steel
tape, popper, or interface probe with 1) an alconox detergent solution, 2) 20 percent
methanol solution, and finally 3) deionized water.
2176-027/WPPHAZ/333 4-1
-
• Place the measurement apparatus in a clean container (storage case, ziploc bag, etc.)
The water level measurement device is now ready for use at the next well.
Decontamination of Pumps
• While withdrawing the tubing and pump from the well, decontaminate the exterior of
tubing by spraying with alconox detergent solution and wiping with a clean paper
towel saturated with 20 percent methanol solution. Once all of the tubing has been
removed from the well, cleaned, and wound onto a spool, rinse the tubing with clean
water from a sprayer.
• Decontaminate the exterior of the pump by spraying first with alconox detergent
solution, followed by a 20 percent methanol solution.
• Place the pump or the suction line of the pump (with the exterior now cleaned) into a
decontamination tube consisting of a section of PVC pipe with a water-tight cap on
the bottom, or a decontamination tub.
• Fill the decontamination tube or tub with an alconox detergent solution and turn on
the pump. While pumping, add 2 gallons of 20 percent methanol solution into the
tube, followed by 2 gallons of clean water. Discharge water should be allowed to
infiltrate into the soil away from the well. When the pump has been completely
purged, turn the pump off. Store the pump in the decontamination tube until it is
ready to be placed in the next well. The tube should be occasionally rinsed to keep it
clean.
Decontamination of Bailers
For portable bailers, the bailer should be filled and drained twice with an alconox
detergent solution; then filled and drained twice with a 20 percent methanol solution. The
exterior and interior of the bailer should be cleaned with the methanol solution, wiped with a
clean towel, then rinsed with clean water. The bailer should be placed in a clean storage case or
a disposable type storage bag and is now ready to be used at the next well.
2176-027/WPPHAZ/333 4-2
-
4.2 Personnel Decontamination
All decontamination of personnel protective equipment is to be conducted in accordance
with "Site Safety Plan, Savage Well Site RI/FS, Milford, NH."
Personnel decontamination areas will be established at each sampling location. All
personal protective equipment will be disposed of, or decontaminated at the conclusion of each
work day. A designated container for tyvek suits and other disposables will be located on the
site. Tyvek suits, respirator cartridges, and other disposables (inner gloves) will be doffed at the
conclusion of each work day and replaced with new equipment prior to commencing work on the
following work day. Respiratory equipment, boots, outer gloves, and foul weather gear will be
washed and rinsed, then placed in a designated personal protective equipment storage area.
2176-027/WPPHAZ/333 4-3
-
5.0 SAMPLE RECORDS AND CHAIN OF CUSTODY
5.1 Labeling of Sample Containers
To prevent misidentification of samples, the samplers will complete labels affixed to each
sample container by Alliance Technologies laboratory. The labels will be sufficiently durable to
remain legible even when wet. The lab will provide precleaned sample containers with labels
which state the types of analysis to be performed and the preservatives, if any, contained in the
sample bottle. The following types of information will be recorded on the label using an
indelible marker only:
• Identification Number
• Date and Time of Collection
• Analysis to be Performed
• Preservatives (if any)
• Name of Sampler
5.2 Recordkeeping and Chain of Custody Documentation
To establish the documentation of water level measurements, well purging procedures,
sample collection procedures, and decontamintion procedures, all of which are necessary for
data validation, all pertinent information will be recorded on an HMM Groundwater Sample
Report form (see Figure 3.1).
To establish the documentation necessary to trace sample possession from time of
collection, an HMM Chain-of-Custody record (see Figure 5.1) will be completed and accompany
every sample. Chain of Custody refers to the sequence of sample collection, handling, storage
and transport to the analytical laboratory. The chain of custody document is a record of who
handled the sample, and when it was passed on to the next person's custody. A sample is
considered to be under an authority's custody if:
• It is in the person's possession
• It is in view of the person after being in their possession
• It is locked up in a secure storage area after being in that person's physical possession
2176-027/WPPHAZ/333 5-1
-
"" o- u
-
A proper chain of custody procedure will be maintained during the groundwater sampling
program to protect the legal integrity of the data produced by the laboratory. Legally, the sample
must be traceable from the time of collection to the time of analysis in the laboratory.
The HMM Chain of Custody record includes information regarding:
• The project identification
• The samplers names
• The station number
• The date and time of collection
• The sample identification
• The type and number of containers
• The analysis to be performed
• Preservatives added to the sample containers
• Pertinent remarks
The chain of custody record will be completed by the sampler at the time of sample
collection and will remain with the samples at all times through delivery at the analytical
laboratory. When the custody of the samples is transferred from ones' physical possession to
anothers' physical possession, the chain of custody form must record the signature, time and
date by the person relinquishing the samples and the signature must be recorded by the person
receiving possession of the samples.
The original (top copy) chain of custody document is to stay with the samples at all times.
Duplicate copies may be kept with each person relinquishing custody of the samples. The
original chain of custody form is to be returned by Alliance Technologies to HMM Associates
with the final analytical data sheets.
5.3 Handling and Transportation of Samples to the Laboratory
Samples will be packed with adequate ice or blue ice to keep them below 4°C from the
time of collection until delivery to Alliance Technologies. HMM Associates samplers will
deliver all collected samples from the Savage Well Site to HMM Associates corporate offices in
Concord, MA. The custody of the samples will be transferred to HMM Associates Quality
Assurance Manager, or a designated alternative, and either 1) placed in temporary storage in the
HMM Associates environmental sample refrigerator where they will await pick-up by Alliance
Technologies, or 2) they will be immediately transported, and custody transferred, to Alliance
Technologies. In all cases, transfer of custody must be kept to a minimum.
2176-027/WPPHAZ/333 5-3
-
6.0 SITE SAFETY CONSIDERATIONS
6.1 Field Monitoring and Screening
Air monitoring shall be performed within the work area on-site in order to detect the
presence and the relative levels of toxic substances. Monitoring may also be conducted to
identify other dangerous situations such as the presence of flammable or explosive atmospheres
and/or oxygen deficient environments. The data collected during field monitoring shall be used
to determine the appropriate levels of personal protective equipment. Monitoring shall be
conducted in order to determine baseline data on potential hazards prior to entry in the work
area, and periodically while conducting work on-site to evaluate any changes in conditions of the
specific work area. Each work area must be screened for ambient levels of contamination prior
to initiating work activities, and during any change in site conditions (i.e., opening of a
monitoring well).
Air monitoring of the sampling workspace should be conducted when the monitoring well
cap is removed and vented, and during well purging activities. Personal protective equipment
should be selected in accordance with the guidelines and action levels contained in the "Site
Safety Plan". Air monitoring and field screening equipment will consist of an HNU
photoionizer, a Thermo Environmental Instruments (OVM) photoionizer, or a Foxboro organic
vapor analyzer (OVA). The HNU, OVM, and OVA will be used for the determination of
organic vapor activity in the work environment. The HNU and OVM have the ability to detect
from 1 ppm to 2000 ppm. The OVA has the ability to detect from 1 ppm to 1000 ppm. All
groundwater sampling activities must be conducted in accordance with "Site Safety Plan, Savage
Well Site RI/FS, Milford, NH".
2176-027/WPPHAZ/333 6-1
-
7.0 REFERENCES
1. U. S. Environmental Protection Agency, "Work Plan for The Savage Municipal Well Site,
RI/FS, Milford, NH, Volume I: Technical Scope of Work", 1986(a).
2. New Hampshire Water Supply and Pollution Control Commission (NHWSPCC),
"Hydrogeological Investigation of the Savage Well Site, Milford, New Hampshire, Vol. I
&H", 1985.
3. U.S. Environmental Protection Agency "RCRA Ground-Water Monitoring Technical
Enforcement Guidance Document", OSWER-9950.1, 1986b.
4. U.S. Environmental Protection Agency, "Practical Guide for Ground-Water Sampling",
EPA-600/2-85/104, 1985.
5. U.S. Environmental Protection Agency, "Manual of Ground-Water Quality Sampling
Procedures", EPA-600/2-81-160, 1981.
6. U.S. Environmental Protection Agency "Test Methods for Evaluating Solid Waste",
EPA/SW-846, 2nd edition, 1982.
7. U.S. Geological Survey, "National Handbook of Recommended Methods for Water-Data
Acquisition", Reston, VA, Revised 1980.
8. National Council of the Paper Industry for Air and Stream Improvement, "A Guide to
Groundwater Sampling", NCASI Technical Bulletin No. 362, 1982.
9. U.S. Environmental Protection Agency "Ground-Water Monitoring Series - Technical
Papers", CERI-87-7,1987.
10: Holden P.W., "Primer on Well Water Sampling for Volatile Organic Compounds",
University of Arizona, Water Resources Research Center, 1984.
2176-027/WPPHAZ/333
-
SAVAGE WELL SITE RI/FSMILFORD, NEW HAMPSHIRE
SAMPLING AND ANALYSIS/WORK PLANGROUNDWATER SAMPLING
SUBTASK 2G
HMM Document No. 2176-027/HAZ/333
October, 1988
Prepared by:
HMM ASSOCIATES, INC.336 Baker Avenue
Concord, MA 01742
-
TABLE OF CONTENTS
Page
1.0 INTRODUCTION 1-1
1.1 B ackground and Site History 1-1
1.2 Objective 1-1
1.3 Overview 1-2
1.4 Approach to Work 1 -5
2.0 TEST BORING PROCEDURES (AUGERS) 2-1
2.1 Hollow Stem Auger Method 2-1
2.2 Drilling Method 2-1
2.3 Advantages/Disadvantages 2-4
3.0 TEST BORING PROCEDURES (WASH AND DRIVE) 3-1
3.1 Wash and Drive Method 3 -1
3.2 Drilling Method 3-1
3.3 Advantages/Disadvantages 3-4
4.0 CASING ADVANCEMENT (ALTERNATE METHOD) 4-1
4.1 Spinning Casing 4-1
4.2 Drilling Method 4-1
4.3 Advantages/Disadvantages 4-3
5.0 SPECIAL DRILLING PROCEDURES OF BORINGS IN CONTAMINATED 5-1
AREAS
5.1 Objective 5-1
5.2 Telescoping Casing 5-2
5.3 Hollow-Stem Auger/Flash Joint Casing 5-2
5.4 Temporary Casing/Hollow-Stein Auger 5-4
5.5 Drilling Spoils 5-6
6.0 DECONTAMINATION 6-1
6.1 Overview 6-1
6.2 Procedure 6-1
2176-022/HAZ/739 -i-
-
TABLE OF CONTENTS (Continued)
Page
7.0 SOIL SAMPLING 7-1
7.1 Overview 7-1
7.2 General Procedures 7-1
8.0 ROCK CORING 8-1
8.1 Overview 8-1
8.2 Procedure 8-1
8.3 Rock Core Sampling 8-2
8.4 Specialized Rock Coring 8-5
8.5 Rock Core Handling 8-6
9.0 MONITORING WELLS 9-1
9.1 Monitoring Well Construction 9-1
9.2 Well Development 9-3
10. PIEZOMETERS 10-1
10.1 Piezometer Construction 10-1
11.0 AQUIFER TESTING 11-1
11.1 Overview 11-1
12.0 DELIVERABLES 12-1
13.0 ATTACHMENTS 13-1
1) HMM Associates - Boring Log
2) HMM Associates - Monitoring Well Log
3) HMM Associates - Permeability Test Data Sheets
4) HMM Associates - Bedrock Packer Test Data Sheet
5) Soil Classification Criteria
6) Rock Classification Criteria
7) Guild Drilling Co. Brochure
2176-022/HAZ/739 -ii-
-
1.0 INTRODUCTION
1.1 Background and Site History
In February of 1983, volatile organic compounds were detected in the Savage Well during
routine water quality monitoring by the New Hampshire Water Supply and Pollution ControlCommission (NHWSPCC).
In response to the contamination, hydrogeological investigations were initiated at the O.K.Tool Company and Hitchiner Manufacturing Company facilities which are located near the
Savage Well. The Hydrogeological Investigation Unit of the Water Supply and PollutionControl Commission designed and implemented a study of the Savage Well area in the summerand fall of 1984.
The study revealed that the area is underlain by an unconfined, high yield, overburdenaquifer. Volatile organic compounds have been detected in the ground water and surface waternear Savage Well.
HMM Associates Inc. has been tasked to conduct a Remedial Investigation/FeasibilityStudy (RI/FS) at the Savage Well Site in Milford, NH. The RI/FS is to be performed inaccordance with the Technical Scope of Work prepared by U.S. EPA Region I (EPA, 1986a) and
be consistent with the National Contingency Plan effective February 18, 1986 (NCP), with the
EPA RI/FS Guidance dated June, 1985, to the extent the RI/FS Guidance is consistent with the
NCP, with EPA's "Interim Guidance on Superfund Selection of Remedy" and with theSuperfund Amendments, and Reauthorization Act of 1986 (SARA). To the extent that theTechnical Scope of work is inconsistent with the NCP, the NCP shall govern.
1-2 Objective
The objective of the groundwater monitoring well installation program is to determine thehydrologic properties of the aquifer in order to characterize the transport of contaminants ingroundwater by:
• Determining the nature and extent of contamination sufficient to define the
boundaries of contaminant plumes and quantify the plumes;
Visually determining the subsurface stratigraphy and structure includinglithologies, grain sizes, sorting, fracturing, homogeneity/heterogeneity, etc., for
each different rock and soil type;
2176-022/HAZ/739 1-1
-
• Determining the concentration, environmental fate, transport mechanisms andother significant characteristics of each contaminant;
• Evaluating the waste mixtures and contaminants between groundwater and soil orrock;
• Quantifying the hydrogeological factors (e.g., in-situ permeability and conductivityof each soil and rock type and depth of saturated zone;
• Quantifying the routes of groundwater migration transport rates and receptors;
• Determining the seasonal fluctuation(s) in the water table elevations, flowgradients and contaminant concentration^) simultaneously with factors such asprecipitation, and runoff;
• Evaluating the contribution to contaminant loading of the aquifer and surfacewaterbodies;
• Reviewing and illustrating groundwater and surface water classifications andassessing the need for institutional controls on groundwater use;
• Assessing the extent to which the hazardous substances will migrate once the
limits of the plume are determined;
• Evaluating all pertinent physical and chemical waste characteristics that may affectthe type of treatment possible; and
• Assessing the potential, characteristics, extent and risk of future releases, if any, ofresiduals remaining onsite.
1.3 Overview
The installation of monitoring wells will be conducted in two phases: In Phase I, up to 15well clusters consisting of up to two wells, one overburden approximately 35 feet deep and oneshallow bedrock well screened in the bedrock fracture zone at approximately 75 feet, 4 deep
bedrock wells cored at a minimum of 50 ft into bedrock, and 5 piezometers will be located basedon geophysical information and past sampling data. Phase II will involve locating a number ofadditional wells (up to 20 clusters with two wells) to supplement Phase I sampling and boringinformation.
2176-022/HAZ/739 1-2
-
The Phase II wells will be used to fine tune the monitoring well system after a complete review
of Phase I information. No attempt will be made to predict Phase II well cluster locations until
the Phase I groundwater quality data is available.
HMM Associates understands that the Phase I well clusters will consist of two individual
wells depending on the saturated thickness of the overburden aquifer and then these clusters
could possibly consist of three wells under special conditions supplemented by Phase II
drillings. The wells will have a 10-foot screened interval which will allow for measurement of
the piezometric head and groundwater quality within a selected zone of the aquifer. The
clustered wells are necessary to determine the vertical hydraulic gradients and the groundwaterquality in the various zones. Screen placement will be contingent upon the results of fieldscreening of split-spoon samples and the nature of overburden materials encountered in each testboring. The monitoring well locations depicted on the enclosed figure (Figure 1) were selectedbased upon the following criteria.
• Well cluster MW-1 will provide a monitoring location for background waterquality to the west of the site as the generalized flow direction of groundwater isfrom west to east.
• Well cluster MW-2, which, includes a bedrock well installation, will provide
groundwater quality, groundwater elevation data, and bedrock elevation data on the
north side of the river, to assess potential for contaminant migration across theSouhegan River in both the overburden and bedrock strata.
• Well clusters MW-3 and MW-4 will provide information on stratigraphy, soil
conditions, and groundwater quality and elevations downgradient from the
Hitchiner Landfill and the New England Steel Fabricators, Inc. (NEST FAB)facility, respectively. Well cluster MW-4 includes a bedrock well, which willprovide information on bedrock conditions and potential contaminant migration in
bedrock.
• Well clusters MW-5 and MW-7 will provide information on areas likely to bedowngradient of NEST FAB and Hitchiner Landfill and upgradient of Hitchiner
and Hendrix facilities.
Well clusters MW-6, MW-8, MW-9, and MW-10 will provide information ongroundwater quality and groundwater elevation in the vicinity of several industries
located to the north and south of Route 101.
2176-022/HAZf739 1-3
-
O O(3 m mo g g
33D Dm mZ 2
-
Well clusters MW-11, MW-12, MW-13, and MW-14 will provide groundwater
quality information in the vicinity of the Hitchiner-Hendrix discharge stream and
on contaminant migration in a general easterly direction.
• Bedrock wells MW-11 and MW-14 will provide water quality information and
information on bedrock conditions east of the Drive-In Theater and in the vicinity
of the discharge stream, to assess the potential for contaminant migration in
bedrock.
• Cluster well MW-15 will be placed to help characterize the easterly extent of
contamination which may be attributed to the site.
Each of the deep cluster wells will be drilled 10 feet in rock to conform bedrock.
• Piezometers P1-P2 will be set in the Souhegan River to evaluate whether the
stream is recharging the groundwater or vice versa. P3 and P4 will be set in the
discharge stream. P5 will be used to help determine the extent of contamination to
the east. The piezometers will be hydraulically isolated by a cement/bentonite seal.
1.4 Approach to Work
HMM Associates will conduct the hydrogeologic investigation of the site as follows; the
investigation will include the installation of a total of 15 overburden well clusters, 4 bedrock
wells and 5 piezometers as part of Phase 1 program. A well cluster will consist of two
individual wells depending on the saturated thickness of the overburden aquifer. The wells will
have a 10-foot screened interval which will allow for measurement of the piezometric head and
groundwater quality within a selected zone of the aquifer.
Screen placement is to be contingent upon results of field screening with the OVA of
split-spoon samples and the nature of overburden materials encountered in each test boring. The
setting of the screen will be determined by EPA staff and HMM Associates geologist on site,
after reviewing the data from the soil cores and evaluating stratigraphic changes.
HMM will subcontract all of the drilling and monitoring well installations to Guild
Drilling Co. of East Providence, Rhode Island. Established in 1953, Guild is a recognized leader
in the test boring business. A brochure provided by Guild follows this section.
Guild will mobilize two drilling rigs with hollow stem augering, drive and wash and rock
coring capabilities to install the monitoring wells. One rig will be a truck mounted B-40 (or
equivalent). The second rig will be an all terrain vehicle mounted B-40 or equivalent.
2176-022/HAZ/739 1-5
-
Ten of the Phase 1 overburden wells will be cored (10 feet) into rock to confirm depth to
rock and the bedrock characteristics. The exact locations of all wells will be determined based
on the outcome of the results of the geophysical program. Based on field data the Phase I
drilling may be supplemented with wells from Phase n prior to the first round of sampling to
better define the limits of groundwater contamination, and scoping of subsequent Phase II if
necessary.
2176-022/HAZ/739 1-6
-
2.0 TEST BORING PROCEDURES fAUGERS)
2.1 Hollow Stem Auger Method
This method is suitable for unconsolidated deposits that do not have large cobbles or
boulders. Hollow-stem augers are continuous flight augers equipped with a hollow core. During
advancement, a removable center plug is placed in the hollow core to prevent soil materials from
entering this hole which serves as an inner casing. The augers are advanced by a combination of
rotation and downward pressure. The hole cuttings are carried upwards to the surface along the
outside of the augers on the screw-shaped auger flights. When the desired depth is reached, the
center plug can be removed and undisturbed or representative samples may be obtained by
passing the sampling tools through the center of the auger and out the bottom. Monitoring wells
can also be installed through the center of the augers. Figures 2 and 2A are examples of typical
hollow stem auger equipment.
Commonly, the inside diameter of the hollow stem is 4 to 6 inches, and the augers produce
a hole 8 to 12 inches in diameter. Borings up to 150 ft. are possible using this method. Auger
rigs are skid, truck and track mounted, giving them excellent mobility. This drilling method is
relatively fast in medium and fine-grained soils such as clays and outwash sands.
Hollow-stem auger methods have significant limitations in investigations in contaminated
areas due to the potential for cross-contamination in the borehole. Contaminated cuttings
moving up an auger flight may cross contaminate overlying clean zones, or contaminated auger
flights penetrating to greater depths may carry the contaminants down with them. Additionally,
the rotating action of the augers causes a smearing in fine-grained soils. This smearing may
significantly reduce the local permeability in the borehole, resulting in erroneous estimates of
in-situ permeability from field tests. Additionally, this smearing may effectively seal off a zone
where a monitoring well may be installed, and well development may not be adequate to
remediate this effect.
2.2 Drilling Method
1. The hollow stem plug is placed inside the lead auger and the lead auger is attached
to the drill head. Auger teeth which are sharp, protruding metal tabs, are located at
the tip of the lead auger to assist in cutting the hole. The auger is drilled or
"screwed" into the ground by a combination of rotation and downward pressure. If
obstructions are encountered, the auger may "walk" or become crooked. If this
occurs at a shallow depth it is advisable to move the boring over slightly and start
the hole again.
2176-022/HAZ/739 2-1
-
* • ' ^— * • •' • • • • ' «
.••.'. ! V- V. . •• • ' • • • • ' • *. . * . • ,1» • • • . • • • • « . . '• • • •» - * • « •
• •
• - •.C-" ̂ . •.'"•/5• • - • *̂
Auf«r drilling t«eboiqu«.(Trcm "Soil Stopliaff Method*and Iquipo«ntLoaff7«ar Co.)
2176-026/HAZ/309
FIGURE 2
2-2
-
Tht continuous-flight auger bores Into the soil and rotates thecuttings upward along the flights. Tht uppermost cuttings artdischarged at the surface to aakt roo» for the spact of tho augeras 1t penetrates additional soils.
2176-026/HAZ/309
FIGURE 2A
2-3
-
2. During drilling, the cuttings from the borehole will travel up the auger flights to
the ground surface where they are usually shoveled to the side. The auger cuttings
can be inspected and described. Oftentimes, stratigraphic changes can be observed
due to changes in the cuttings, or the