proficiency testing (external quality assessment) of drug detection in oral fluid

4
Proficiency testing (external quality assessment) of drug detection in oral fluid Joe Clarke a, * , John F. Wilson b a Altrix Healthcare plc, Birchwood Science Park, Warrington WA3 7BP, UK b Department of Pharmacology, Therapeutics and Toxicology, Wales College of Medicine, Cardiff University, Heath Park, Cardiff CF14 4XN, UK Received 20 September 2004; received in revised form 17 November 2004; accepted 17 November 2004 Available online 6 April 2005 Abstract Eighteen external quality assessment (proficiency testing) samples were prepared from client specimens collected with the Intercept 1 oral fluid collection device and by spiking drug-free oral fluid. Samples were circulated in pairs at quarterly intervals to 13 UK and USA based laboratories for analysis by a panel of OraSure micro-plate Intercept 1 enzyme immunoassay kits and hyphenated mass spectrophotometric techniques. During the survey, there was a single case of non-specificity in a false report for methadone. The major errors were of lack of sensitivity relative to the concentration thresholds specified for the immunoassays. The sensitivity for overall ‘present’/‘not found’ reports calculated as true positives/(true positives + false negatives) were for the amfetamine specific assay 50%, methyl-amfetamines 93%, barbiturates 64%, cannabinoids 73%, cocaine and metabolites 100%, benzodiazepines 69%, methadone 95%, opiates 79% (opiates excluding oxycodone 93%), phencyclidine 93% and human gamma-globulin 97%. A small number of the sensitivity errors were attributable to errors in chromatographic confirmation techniques. # 2005 Elsevier Ireland Ltd. All rights reserved. Keywords: Proficiency; Oral fluid; Amfetamines; Cannabinoids; Cocaine; Opiates; Human gamma-globulin 1. Introduction There is a growing interest and utilisation of alternate biological fluids than urine for drug testing. Oral fluid testing has seen a rapid increase in use over recent years and the publication of Proposed Revisions to Mandatory Guidelines for Federal Workplace Drug Testing Programs [1] highlight the growing acceptance of this testing medium. An increas- ing number of laboratories worldwide are now routinely analysing oral fluid samples collected from donors in a combination of workplace, criminal justice and rehabilita- tion settings. It is important that results provided by analy- tical laboratories testing the samples are error-free since, like urine samples from similar settings, they can be used to make clinical and/or legal decisions. Quality assurance must be a key principle of any analytical laboratory and the assessment of laboratory performance through external quality assessment (proficiency testing or EQA) schemes is an important element of a complete quality system [1–3]. In 2002, Cardiff Bioanalytical Services Ltd. (Cardiff, UK) was commissioned by Altrix Healthcare plc (Warring- ton, UK) to design and instigate an in-house EQA scheme. The scheme was to be offered to laboratories analysing samples collected using the Intercept 1 oral fluid collection device (OraSure Technologies, Inc., Bethlehem, PA, USA) that were subsequently analysed by enzyme immunoassays www.elsevier.com/locate/forsciint Forensic Science International 150 (2005) 161–164 * Corresponding author. Tel.: +44 1925 848900; fax: +44 1925 848949. E-mail address: [email protected] (J. Clarke). 0379-0738/$ – see front matter # 2005 Elsevier Ireland Ltd. All rights reserved. doi:10.1016/j.forsciint.2004.11.025

Upload: joe-clarke

Post on 26-Jun-2016

213 views

Category:

Documents


0 download

TRANSCRIPT

Proficiency testing (external quality assessment)

of drug detection in oral fluid

Joe Clarke a,*, John F. Wilson b

a Altrix Healthcare plc, Birchwood Science Park, Warrington WA3 7BP, UKb Department of Pharmacology, Therapeutics and Toxicology, Wales College of Medicine,

Cardiff University, Heath Park, Cardiff CF14 4XN, UK

Received 20 September 2004; received in revised form 17 November 2004; accepted 17 November 2004

Available online 6 April 2005

www.elsevier.com/locate/forsciint

Forensic Science International 150 (2005) 161–164

Abstract

Eighteen external quality assessment (proficiency testing) samples were prepared from client specimens collected with the

Intercept1 oral fluid collection device and by spiking drug-free oral fluid. Samples were circulated in pairs at quarterly intervals

to 13 UK and USA based laboratories for analysis by a panel of OraSure micro-plate Intercept1 enzyme immunoassay kits and

hyphenated mass spectrophotometric techniques. During the survey, there was a single case of non-specificity in a false report

for methadone. The major errors were of lack of sensitivity relative to the concentration thresholds specified for the

immunoassays. The sensitivity for overall ‘present’/‘not found’ reports calculated as true positives/(true positives + false

negatives) were for the amfetamine specific assay 50%, methyl-amfetamines 93%, barbiturates 64%, cannabinoids 73%, cocaine

and metabolites 100%, benzodiazepines 69%, methadone 95%, opiates 79% (opiates excluding oxycodone 93%), phencyclidine

93% and human gamma-globulin 97%. A small number of the sensitivity errors were attributable to errors in chromatographic

confirmation techniques.

# 2005 Elsevier Ireland Ltd. All rights reserved.

Keywords: Proficiency; Oral fluid; Amfetamines; Cannabinoids; Cocaine; Opiates; Human gamma-globulin

1. Introduction

There is a growing interest and utilisation of alternate

biological fluids than urine for drug testing. Oral fluid testing

has seen a rapid increase in use over recent years and the

publication of Proposed Revisions to Mandatory Guidelines

for Federal Workplace Drug Testing Programs [1] highlight

the growing acceptance of this testing medium. An increas-

ing number of laboratories worldwide are now routinely

analysing oral fluid samples collected from donors in a

combination of workplace, criminal justice and rehabilita-

* Corresponding author. Tel.: +44 1925 848900;

fax: +44 1925 848949.

E-mail address: [email protected] (J. Clarke).

0379-0738/$ – see front matter # 2005 Elsevier Ireland Ltd. All rights r

doi:10.1016/j.forsciint.2004.11.025

tion settings. It is important that results provided by analy-

tical laboratories testing the samples are error-free since, like

urine samples from similar settings, they can be used to

make clinical and/or legal decisions. Quality assurance must

be a key principle of any analytical laboratory and the

assessment of laboratory performance through external

quality assessment (proficiency testing or EQA) schemes

is an important element of a complete quality system [1–3].

In 2002, Cardiff Bioanalytical Services Ltd. (Cardiff,

UK) was commissioned by Altrix Healthcare plc (Warring-

ton, UK) to design and instigate an in-house EQA scheme.

The scheme was to be offered to laboratories analysing

samples collected using the Intercept1 oral fluid collection

device (OraSure Technologies, Inc., Bethlehem, PA, USA)

that were subsequently analysed by enzyme immunoassays

eserved.

J. Clarke, J.F. Wilson / Forensic Science International 150 (2005) 161–164162

Table 1

Analyte concentrations distributed by the oral fluid EQA scheme

between May 2002 and May 2004 and mean coefficient of variation

(CV) of measurements by scheme participants

Analyte Concentrations

(ng/ml)

CV% (range)

d-Amfetamine 10, 56, 283 16 (14–17)

Methylamfetamine 60, 61, 764 28 (6–54)

Pseudoephedrine 197 –

manufactured by the same organisation. Two samples would

be sent in a blind fashion every calendar quarter and

participants would report either qualitative results only or

a combination of qualitative and quantitative results, the

latter from a combined method of chromatography and mass

spectrometry. Thirteen laboratories from the USA and UK

have participated in the scheme and the results reported for

the 18 samples circulated between May 2002 and May 2004

are detailed and discussed in this paper.

Secobarbital 30, 31, 50 2

Phenobarbital 61 –

D9-Tetrahydrocannabinol 0.6, 1.1, 4.0 42 (35–50)

Cocaine 3, 4, 6, 16, 45 34 (11–52)

Benzoylecgonine 6, 6, 9, 15, 17, 24 35 (14–84)

Diazepam 149 23

Temazepam 2 30

Methadone 6, 11, 22, 48 28 (14–48)

(d)Propoxyphene 405 –

6-Acetylmorphine 11, 13, 18, 57, 258 54 (41-61)

Morphine 15, 16, 16, 21, 63 60 (17–161)

Codeine 2, 3, 3, 7 49 (19–79)

Dihydrocodeine 12, 20, 29 –

Oxycodone 25 –

Phencyclidine 2.0, 4.0 32

Thioridazine (mesoridazine) 650 (500) –

2. Materials and methods

2.1. Preparation of oral fluid samples

Samples were prepared from oral fluid collected from

two sources. The first source of testing matrix was samples

that had been collected from clients by one of a number of

UK testing centres. The samples were collected using the

Intercept1 oral fluid collection device. The residual sample

remaining after routine laboratory analysis was selected for

use from those cases that had tested positive for the required

drug compounds. The sampling process involves placing a

collection pad impregnated with a mixture of common salts

between the sample donor’s lower gum and cheek and

leaving in place for at least 2 min. The collection pad

encourages movement by osmotic means of oral fluids

including oral mucosal transudate (OMT) from the surfaces

inside the mouth into the pad. The pad is then removed from

the mouth and inserted into a dedicated transport vial

containing 0.8 ml of a preservative (chlorhexidine digluco-

nate, Tween 20, deionised water) which minimises any

degradation of substances that might have been collected

[4,5].

Drug concentrations were determined in the donor sam-

ples at the test centre by GC–MS and samples stored at

�20 8C. The second source of matrix was oral fluid collected

by non-stimulated expectoration into a sterile tube by a drug-

free volunteer. The fluid was centrifuged and supernatant

diluted 1 + 3 (v/v) by Intercept1 negative calibrator for oral

fluid testing. EQA samples were prepared by dilution of the

client specimens with drug-free matrix to give the desired

concentration of drugs, by augmentation and addition of

certain analytes to the client specimens by spiking weighed-

in concentrations of drugs, and by spiking drugs into drug-

free matrix. All samples were heat treated at 60 8C for 1.5 h

and 1.7 ml volumes distributed in liquid form in plastic

screw-topped tubes (Cat no. 72.609 Sarstedt Ltd., Leicester,

UK) to participating laboratories by courier service.

The analytes included during the survey and their con-

centration is presented in Table 1. D-Amfetamine was spiked

into two samples and the third high concentration was a

client specimen. Methylamfetamine, pseudoephedrine,

secobarbital, phenobarbital, temazepam, (d)propoxyphene,

oxycodone, phencyclidine and thioridazine with its meta-

bolite mesoridazine were all spiked into samples. Cannabi-

noids were present in two client specimens and D9-

tetrahydrocannabinol was spiked at 4 ng/ml into drug-free

matrix in a third sample. Metabolised cocaine was present

in five client samples and benzoylecgonine alone was

spiked into a further sample at the highest concentration

circulated. Metabolites from diazepam use were present in

one client sample and from methadone treatment in four

samples. Five samples from heroin users were included.

Two of the latter additionally contained dihydrocodeine,

and four samples contained low concentrations of codeine.

The concentration of morphine in one sample was aug-

mented by spiking to obtain a concentration above the

opiate cut-off. A further sample was positive for opiates

from use of dihydrocodeine alone and contained 20 ng/ml

dihydrocodeine.

2.2. Target compounds

Laboratories participating in the EQA programme

reported the drugs detected in the samples within a spe-

cified 4-week deadline using some or all of a common

panel of OraSure micro-plate Intercept1 enzyme immu-

noassay (EIA) kits. The micro-plate assays are competitive

immunoassays intended for use in the qualitative determi-

nation of drugs and metabolites in oral fluid collected with

the Intercept1 oral fluid collection device. The drug

groups and the cut-off concentrations applied are given

in Table 2. One laboratory used immunoassays from an

alternative supplier (Cozart Bioscience Ltd., Abingdon,

UK). The Intercept1 oral fluid EIA cut-off concentration

is, in several cases, 25% of that specified by the US

J. Clarke, J.F. Wilson / Forensic Science International 150 (2005) 161–164 163

Table 2

Cut-off concentrations and target analytes applied in OraSure

Technologies, Inc., micro-plate Intercept1 enzyme immunoassay

kits

Test group Cut-off

concentration

(ng/ml)

Calibrator

compound

Amfetamine specific 100 d-Amfetamine

Methyl-amfetamines 40 d-Methamfetamine

Barbiturates 20 Secobarbital

Cannabinoids 1 D9-Tetrahydrocannabinol

Cocaine and metabolites 5 Benzoylecgonine

Benzodiazepines 1 Nordazepam

Methadone 5 Methadone

Opiates 10 Morphine

Phencyclidine 1 Phencyclidine

Human IgG 500 Human IgG

Table 3

Sensitivity and specificity of reports from the oral fluid EQA scheme

Test group Sensitivity (%) Specificity (%)

Amfetamine specific 50 100

Methyl-amfetamines 93 100

Barbiturates 64 100

Cannabinoids 73 100

Cocaine and metabolites 100 100

Benzodiazepines 69 100

Methadone 95 99

Opiates 79 100

Phencyclidine 93 100

Human IgG 97 No data

Department of Health and Human Services (HHS) [1] to

account for the approximate 1 + 3 dilution that occurs on

extraction of the Intercept1 collection pad with the pre-

servative in the transport vial [5,6]. Additional compounds

to those specified by HHS are included in the panel to more

closely resemble the existing urine testing programmes

and the requirements for clinical testing and monitoring.

The panel of tests also included a specific test for the

presence of human gamma-globulin (IgG) at a concentra-

tion above a threshold of 500 ng/ml in order to monitor

sample integrity and to exclude non-human and other

substituted fluids [7].

Where positive immunoassay results occurred, a num-

ber of laboratories used GC–MS, GC–MS–MS and LC–

MS–MS techniques to identify and quantify the drug

compounds concerned. In addition to reporting the indi-

vidual test results, laboratories combined their immu-

noassay and chromatographic data to determine an over-

all ‘present’ or ‘not found’ report for the 10 specified

analyte groups (Table 1) and for individual compounds

identified.

2.3. Data selection and statistical analysis

The overall ‘present’/’not found’ reports were analysed

for the 10 target groups listed in Table 1. Data were included

in the analysis where either an above threshold concentration

of drug group or no drug was present. Data for drug groups

present in samples at below threshold concentrations were

excluded. Individual results from the 18 samples were

classified as being true positive (TP), true negative (TN),

false positive (FP) or false negative (FN) for the 10 target

groups. In eight cases, where a laboratory had reported an

uncertain result for a test compound present in a sample, the

result was classed as FN. Six uncertain results for analytes

not present were excluded. Sensitivity was calculated

as 100 � TP/(TP + FN) and specificity as 100 � TN/

(TN + FP).

3. Results

Between 6 and 13 (mean 11) laboratories reported on

each distribution of samples. The sensitivity and specificity

for the nine analyte groups and for the human IgG integrity

test are presented in Table 3. The mean and range of

coefficients of variation for the chromatographic measure-

ments reported by scheme participants are given in Table 1.

A clear relationship between precision and drug concentra-

tion was not evident.

There were three cases of false negative reports for

specific compounds as a result of failures in confirmation

techniques. On two occasions, a laboratory failed to detect

morphine with a morphine-specific immunoassay following

a positive opiate group test. In a similar situation, a labora-

tory reported zero measured concentrations for 6-acetyl

morphine (6-AM), morphine and codeine by GC–MS fol-

lowing a positive opiate immunoassay test when all three

compounds were present in the sample.

4. Discussion

A single false positive report was made during the survey.

It was an unconfirmed immunoassay report for methadone in

the sample spiked with (d)propoxyphene. No lack of spe-

cificity was observed in samples containing pseudoephe-

drine or thioridazine and its metabolite. All other reporting

errors were false negative or uncertain reports where an

above threshold concentration of analyte was present.

The lowest sensitivity was observed for the specific

amfetamine assay where 50% of laboratories failed to detect

amfetamine at over twice the threshold concentration. The

93% sensitivity for methyl-amfetamine was the result of two

laboratory errors. One was an immunoassay negative at

61 ng/ml whilst, in the second case, a laboratory detected

methyl-amfetamine in the same sample by immunoassay but

quantified the level by GC–MS as being below threshold at

30 ng/ml and therefore returned a report of ‘not found’. The

target barbiturate compound, secobarbital, was missed by

more than half the participants on two occasions when

spiked at 150% of the barbiturate group cut-off. On the

J. Clarke, J.F. Wilson / Forensic Science International 150 (2005) 161–164164

second occasion, the concentration was verified by chroma-

tography. Sensitivity to phenobarbital was 100% at a con-

centration of 61 ng/ml. The 73% sensitivity for cannabinoids

was largely the result half the participants returning negative

reports for the sample containing the close to cut-off con-

centration of 1.1 ng/ml. The laboratory employing the

Cozart immunoassay missed spiked D9-tetrahydrocannabi-

nol at 4 ng/ml. In the two benzodiazepine-containing sam-

ples, insensitivity was caused by temazepam spiked at 2 ng/

ml. However, the cross-reactivity of the benzodiazepine

Intercept1 micro-plate EIA to this compound is 55%, giving

an effective concentration of only 1.1 ng/ml relative to the

cut-off of 1 ng/ml. Insensitivity to opiates was the result of

negative reports in samples containing dihydrocodeine

metabolites alone or spiked with oxycodone. The reported

cross-reactivity of the Opiates Intercept1 micro-plate EIA to

dihydrocodeine is 185% (10 ng/ml) but that of oxycodone is

only 0.8% (1000 ng/ml) making this compound effectively

invisible.

A small number of sensitivity errors for the reports of

analyte groups were attributable to errors with the chroma-

tographic confirmation techniques. On one occasion, a

laboratory failed to confirm and hence report the presence

by GC–MS of four analytes, secobarbital, diazepam, metha-

done and phencyclidine, despite the respective immunoassay

tests being reported positive.

The major source of error observed was thus of sensi-

tivity where immunoassays failed to achieve their specified

cut-off. The issue of variable cross-reactivity of group assays

to different compounds in the group is potentially relevant to

the barbiturate, benzodiazepine and opiate immunoassays

but is only represented in the current data by the failure to

detect oxycodone. If these data are omitted, the sensitivity

for the opiate group increases to 93%. The analytical per-

formance of laboratories participating was comparable to

that observed for some commercial immunoassays used in

urine drug testing [8]. The increased difficulty that might

result from the concentration of drugs being typically lower

in oral fluids than in urine appears not to be a significant

factor in poly drug identification. A lack of precision in

quantitative measurements for some analytes when mea-

sured in oral fluid is evident in the scale of the coefficients of

variation observed. A low level of unexplained errors can

affect mass-spectrometric techniques applied to either

matrix.

References

[1] Department of Health and Human Services. Proposed Revisions

to Mandatory Guidelines for Federal Workplace Drug Testing

Programs, Federal Register 69 (71) (2004) 19673–19732.

[2] United Kingdom Laboratory Guidelines for Legally Defensible

Workplace Drug Testing, Version 1.0, March 2001. http://

www.wdtforum.org.uk/pdfs/wdtgde�1.pdf.

[3] International Organization for Standardization, General require-

ments for the competence of testing and calibration laboratories,

ISO/IEC 17025, 1999.

[4] R.S. Niedbala, K.W. Kardos, D.F. Fritch, S. Kardos, T. Fries, J.

Waga, Detection of marijuana use by oral fluid and urine

analysis following single-dose administration of smoked and

oral marijuana, J. Anal. Toxicol. 25 (5) (2001) 289–303.

[5] E.J. Cone, L. Presley, M. Lehrer, W. Seiter, M. Smith, K.W.

Kardos, D. Fritch, S. Salamone, R.S Niedbala, Oral fluid testing

for drugs of abuse: positive prevalence rates by InterceptTM

immunoassay screening and GC–MS–MS confirmation and

suggested cutoff concentrations, J. Anal. Toxicol. 26 (8)

(2002) 541–546.

[6] E.J. Cone, Saliva testing for drugs of abuse, Ann. NYAcad. Sci.

694 (1993) 91–127.

[7] P.P. Mortimer, J.V. Parry, Detection of antibody to HIV in saliva:

a brief review, Clin. Diagn. Virol. 2 (1994) 231–243.

[8] J.F. Wilson, B.L. Smith, Evaluation of detection techniques and

laboratory proficiency in testing for drugs of abuse in urine: an

external quality assessment scheme using clinically realistic

urine samples, Ann. Clin. Biochem. 36 (5) (1999) 592–600.