Page 1: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

Pressure Aging Vessel PAV3

Page 2: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

Contents

Preface    4Unpacking    4Warranty    4After-SaleSupport    4

Section1:Introduction    5Section2:Safety    6

2.1ForOwners,Operators,&Maintenance    62.2SafetyInstructions    72.3Warnings    72.4Cautions    8

Section3:SystemOverview    93.1GeneralDescription    93.2Specifications    103.3EnvironmentalConditions    11

Section4:UnpackingandInstallation    124.1Unpacking    124.2Installation    124.3PanHolderLeveling    124.4AirandPowerConnections    13

Section5:Operation    16

Page 3: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

Contents

Section6:Verification    436.1VerifyingTemperature    436.2VerifyingPressure    44

Section7:Troubleshooting    457.1TemperatureDoesNotRecovertoSetpoint    457.2TroubleshootingthePAVController    467.3ReattachingtheFlowAdjustValve    477.4UpdatingtheSoftware    49

Section8:Maintenance    518.1PreventiveMaintenance    518.2ChangingFuses    518.3SpareParts    52

AppendixA:Warranty    54

RevisedJanuary2015

Page 4: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

4

PrefaceUnpackingRetainallcartonsandpackingmaterialsuntiltheunitisoperatedandfoundtobeingoodcondition.Ifdamagehasoccurredduringshipping,notifyAppliedTestSystems(ATS)andthecarrierimmediately.Ifitisnecessarytofileadamageclaim,retainthepackingmaterialsforinspectionbythecarrier.

WarrantyAllnewATSsystemsareshippedwithawarranty.Unitshaveawarrantyagainstdefectivepartsandworkmanshipforonefullyearfromdateofshipment.PleaseseeAppendixA ofthismanualforcompletedetailsonthewarranty.

NOTE:Pleaseseethewarrantyinformationincludedwiththecomputer.ItdetailsinformationthatwillensurepropertransferofthecomputerwarrantyfromATStotheappropriatecompany.

After-Sale SupportIfthereareanyquestionsconcerningtheoperationoftheunitorsoftware,contacttheATSServiceDepartmentat+1-724-283-1212.Beforecalling,pleaseobtainthesoftwarerevisionnumberandtheserialnumberfromtheunit’sdatatag.Asampledatatagisshownbelowandcanbecompletedwiththeunit’sinformationforeasyreference.PleasebepreparedtogiveacompletedescriptionoftheproblemtoATSServiceEngineers.

Page 5: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

5

Section1:IntroductionTheAppliedTestSystems(ATS)PressureAgingVessel(PAV)simulatesin-serviceoxidativeagingofasphaltbinderaccordingtoproceduresdevelopedbytheStrategicHighwayResearchProgram(SHRP).ThePAVsystemisfullycompliantwiththemostrecentversionsofASTMstandardD6521andAASHTOstandardmethodR28.

Thisinformationisrelatedtothefunction,setup,safeoperation,andmaintenanceofthePAVanditsrelatedcomponents.

Thismanualisintendedforanyonewhowillbesettingupand/oroperatingthePAV.

Page 6: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

6

Section2:SafetyAllATSequipmentisdesignedtobeoperatedwiththehighestlevelofsafety.ThismanualandATSendeavortoeducatetheoperatoraboutsafetyissuessurroundingcertainpartsofthemachinerybyusingequipmentlabeling.

2.1 For Owners, Operators, and Maintenance PersonnelReadandunderstandallinstructionsandsafetyprecautionslistedinthismanualbeforeinstallingoroperatingtheunit.Ifthereareanyquestionsregardingoperationoftheunitortheinstructionsinthismanual,contacttheATSServiceDepartmentat+1-724-283-1212.

Inadditiontothesafetywarningslistedontheequipment,warningsarepostedthroughoutthismanual.Readandfollowtheseimportantinstructions.Failuretoobservetheseinstructionscanresultinpermanentdamagetotheunit,significantpropertydamage,personalinjury,ordeath.

ReadOperator’sManual

General Danger

ProtectiveEarth(Ground)

BurnHazard(HotSurface)

ElectricalShock/Electrocution

HandCrushForceFromAbove

EuropeanDirectiveCEMark

NoAccessforUnauthorizedPersons

Page 7: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

7

2.2 Safety Instructions1.Readandfollowallwarningandcautionstatementsinall relatedequipmentmanualsbeforeattemptingtooperate thismachine.Ifindoubtaboutanystatementorsequence,contactATSService.

2.Installationofelectricaldevicesmustbeaccomplishedby competentpersonnelanddoneinaccordancewithany currentlocalandnationalcodes.EquipmentgroundingisaMUSTforbothsafetyandproperoperation.

3.Beforesupplyingelectricalpowertotheunit,turnallpowerswitchesandcontrolstoanOFForNEUTRALposition.

2.3 WarningsThefollowingstatementsarewarningstatements.Unlikecautionstatements,warningstatementsalerttheoperatortoconditionsthatmayinjurepersonnel.Operatorsmustbeawareoftheseconditionsinordertopreventinjuriesthatmayoccurwhileoperatingthisequipment.

WARNING:Obeyelectricalcoderequirements.Theovenandcontrolsystemmustbewiredandgroundedinaccordancewithnationalandlocalelectricalcode

requirements.

WARNING:Becarefulwhenworkingwithequipmentatelevatedtemperatures.Inordertopreventburns,wearprotectiveclothing.

WARNING:Usecautionwhenopeningtheovenlid.Electrically-heatedequipmentcancausesevereburns.

WARNING:Allowadequatetimetorelieveallthepressurefromthevesselbeforeopeningtheovenlidandthepressurevessellid.Anyunrelievedpressurecan

causeequipmentdamageandpossiblepersonalinjury.

WARNING:Unpackandoperateonastablesurface.

Page 8: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

8

2.4 CautionsThefollowingstatementsarecautionstatements.Thesestatementsalerttheoperatortoconditionsthatmaydamageequipment.Operatorsmustbeawareoftheseconditionsinordertoensuresafeoperationofthisequipment.

CAUTION:Installationofelectricaldevicesmustbeaccomplishedbycompetentpersonnelanddoneinaccordancewithanycurrentlocalandnationalcodes.

CAUTION:ThePAVmustbegroundedandwiredinaccordancewithnationalandlocalelectricalcoderequirements.

CAUTION:BeforeenergizingtheelectricalpowertothePressureAgingVessel,placeallcontrolsinanOFFposition.

CAUTION:Donotexceedthemaximumoperatingtemperature.RefertothespecificationsinSection3ofthismanual.

CAUTION:ClosingtheflowadjustvalvewhenthePAVisinoperationwillcausedamagetothevalve.

CAUTION:Allsupportingandcontactingsurfacesmustbenon-flammable.Donotallowflammablematerialstocontacttheshell.

CAUTION:Ifanemergencyshutdownneedstobeperformed,placeON/OFFswitchinanOFFposition.Unitwilldepressurizeautomaticallywithin10-15minutes.

CAUTION:DonotoverflowPAVpans.Refertotestspecificationsforproperamountofmaterial.

Page 9: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

9

Section3:SystemOverview

3.1 General DescriptionTheATSPressureAgingVessel(PAV)isdesignedspecificallytopreparespecimensfortestsdevelopedbytheStrategicHighwayResearchProgram(SHRP)tosimulatein-serviceoxidativeagingthatoccursinasphaltbindersduringservice.

APAVsystemconsistsofthefollowing:ASME/CEcodestainlesssteelpressurevesselenclosedinastainlesssteeloven,pressureregulator,meteringvalve,pressuretransducerandindicator,RTD,temperaturecontrollerwithdatalogging,andpanholder.

3.1.1 Vessel Certification

ASMECode:Unlessotherwisenoted,thePressureAgingVessel(PAV)isdesigned,fabricated,inspected,tested,andstampedinaccordancewiththeASME Boiler and Pressure Code,SectionVIII,Divison1,forserviceatamaximumpressureof2.45MPa/350psigandamaximumtemperatureof150°C/300°F.The ASME Boiler and Pressure Vessel CodeisanInternationallyRecognizedStandard.

ContactATSforthevesselcertificate/U-IAReportandCEDeclarationofConformity.

Page 10: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

10

3.2 SpecificationsBenchtopUnit IntegralVessel/OvenDesign

VerticalLoadingwithFixture

SpecimenCapacity 10Pans

SpecimenPans PerAASHTOT179

PhysicalDimensions 30”high,18”deep,27½”wide (760mmhigh,460mmdeep,700mmwide)

Weight 290lbs(130kg)

OperatingPressure 304.5psi(2.1±0.05MPa)

TemperatureAccuracy Setpoint±0.5°C

TemperatureRange 80°Cto115°C

RTDTemperatureMeasurement 0.1°CResolutionand±0.1°Accuracy MicroprocessorTemperatureControl

OvertemperatureProtection

PowerRequirements

ASMEOvenHighLimitSwitch293°F(145°C)

240VAC,10amps,50/60Hz

VesselPerASMECodeSectionVIII Division1;1992A93ContactATS

VesselDesignConditions

PressureSafetyReleaseValve

TemperatureUniformity

PressureDisplayAccuracy

“TIME”TimerAccuracy

withthePAV’sserialnumberforacopyofthecertification.

325psig(2.24MPa)at250°F(121°C)

340 psi

±0.5°C

±1%

0.1%

“OUTOFRANGE”TimerAccuracy 0.1%

Page 11: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

11

AirInlet Male¼”NPTFitting

SolenoidValve Closedwhenpowerisapplied.In theeventofpowerloss,solenoid valvewillopenandrelease anypressureinthePAVvessel.

3.3 Environmental ConditionsOperationoftheATSPressureAgingVesselinvolvesuseofhightemperaturesandcontentsbeingcontainedathighpressure.ThelocationandplacementofthePAVshouldalwaystaketheseelementsintoconsiderationtoensureasafeworkingenvironment.ThePAVismeantforuseinlaboratory/factorysettingsinadryandcleanworkenvironment.TheworksurfacethatthePAVisplacedonshouldbeacleanandsturdyworksurfaceatareasonableworkingheightandawayfromanywater,gas,orelectricalhazards.Theworksurfaceshouldbeabletosupportmorethantheweightoftheunititself.TheindoorareathePAVisplacedinshouldbewellventilatedandcontainingnoopenflamesormaterialsthatmayconstituteafirehazard.

Page 12: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

12

Section4:UnpackingandInstallation4.1 UnpackingCarefullyunpacktheequipmentandinspectitfordamageduringshipment.Retainallcartonsandpackagingmaterialsuntiltheunitisoperatedandfoundtobeingoodcondition.Ifdamagehasoccurredduringshipping,notifythecarrierandATSimmediately.Ifitisnecessarytofileadamageclaim,retainthepackingmaterialsforinspectionbythecarrier.

4.2 Installation1.UseanoverheadcraneorforklifttoremovethePAVfromthecrateandpositionitinthedesiredlocation.

2.PositionthePAVinthetestingposition,suchasonthePAVheavy-dutysteelstandoronaworkbenchorliftcart.

CAUTION:Allsupportingandcontactingsurfacesmustbenonflammable.Donotallowflammablematerialstocontacttheshell.

3.LevelthePAVbyadjustingthescrewsonthefourfeet.

4.InspecttheO-ringseal.Makesureitiscleanandlightlycoatedwithsiliconegrease.IftheO-ringappearstobewornordamaged,contactATS.

4.3 Pan Holder Leveling1.OncethePAVisinplace,verifythattheringlocatedinsidethebottomofthepressurevesselislevelbyplacingtheparallelbarandalevelacrossthetopofthelevelingring.

WARNING:Whenplacingthelevelandtheparallelbarontothering,donotinterferewiththeRTDlocatedinthebottomofthepressurevessel.TheRTDisfactory-set

andshouldnotbemoved.

2.Ifthelevelindicatesthattheringisnotevenlybalanced,makethenecessarycorrectionsbyrotatingthethreevertical-setscrewstolowerorraisethering.Usea1/8”hexwrenchtomakethisadjustment.

Page 13: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

13

3.Afterlevelingthering,handrotatethehorizontal-setscrewsinordertosecuretheringinthebottomofthevessel.

4.4 Air and Power Connections1.Installtheairpressurelinetothe1/4”malepipe-threadedfittingprovidedonthebackofthecontrolcabinet.

CAUTION:Usearegulatedairsupplyof340to350psi(2.34to2.41MPa).Donotexceed350psi(2.41MPa)orthePAV’sperformancemaybeimpairedbyexcessive

pressure.

2.Aconnectorisprovidedwiththeunit.Thisconnectormaybechangedifitdoesnotmatchtheavailablereceptacle.Powerrequirementsare:230-240VAC,±5V,10amp,single-phase,50/60Hz.

CAUTION:ThePAVrequiresaminimumof230VACtooperatecorrectly.Checkthesupplyvoltagetoensureitis230-240VAC.Ifthesupplyvoltageislessthan

230VAC,aStep-UporBuck-Boosttransformercanbe suppliedtoprovidetheproperoperatingvoltages. ContactATS.BepreparedtoprovidetheATSService representativewithinformationabouttheavailableACvoltagesothecorrecttransformerwiringconfiguration canbeprovided.

CAUTION:ThePAVandcontrolsystemmustbegroundedaccordingtonationalandlocalcoderequirements.

Page 14: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

14

3.Ifequipped,connecttheoptionalUPSbatterybackupsupply.

a.UsethesuppliedstandardpowercordtoplugtheUPS intothewall.Thesystemshouldchargefor24hours beforeconnectingittothePAV.

b.AttachthesuppliedcordfromtheUPStothePAV.

c.Seemanufacturer’sliteratureforfurthersetup instructions.

4.ChecktheflowadjustvalveonthePAV(markedFlowControl)toensureitissetatthecorrectvalues.ThisvalvecontrolstherateatwhichthePAVdepressurizes.Whentheflowadjustvalveissetattherecommendedsetting,thePAVwilldepressurizeatoptimalrate.Theunitwillhaveasticker(seebelow)withthecorrectvalues.UsethevaluemarkedFlowAdj.

NOTE:SeeSection7forinformationonwhattodoiftheflowadjustvalvedetaches.

Page 15: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

15

4.4.1 Regulator Connections 1.Securetheairconnectiontotheairtankusingthesuppliedmaleandfemaleconnections.

2.Connectthehosetotheregulator.

NOTE:DoNOTconnectthehosetothepartmarkedFLOW.Thisisthepressurereliefvalve,andnothingshouldbeconnectedtoit.

NOTE:OutsideoftheUnitedStates,adifferentadaptermayberequired.ACGAadapterisincluded.

3.AttachthehosetothePAVusingthequickconnections.ConnectthemaleendofthehosetothefemaleconnectiononthePAVbyslidingthefemaleconnectiontowardsthePAV,connectingthehose,andreleasingtheslide.

4.Verifythatallconnectionsaretightlysealedbyapplyingsoapywaterontheconnectionsandturningontheair.Ifsudsappear,applymorecompoundortapetosealtheleak.

Page 16: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

16

Section5:Operation

PoweronthePAVbyturningtheMainPowerswitch1. locatedonthefrontofthePAVclockwisetothe“ON”position.

NOTE:Makesuretheairisconnectedandsetat340-350psiasdescribedinSection4.4.

Page 17: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

17

Software/SettingupUserLanguage2.

Select“View.”ThiswilldisplaytheInformationa. screen.ThecurrentversionofthePAVsoftwarewilldisplay.

Page 18: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

18

b. ThePAVwilldisplayitemsinEnglishasthedefaultlanguage.TochangethelanguagefromEnglishtooneofthepreprogrammedlanguages,accesstheoptionsfromthetouch-screendisplay.

Select“View”i.

Select“Language”ii.

Touchthedesiredlanguageforthedisplay.iii.

c. Select“Done”toreturntothepreheatstartupscreen.

Page 19: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

19

SettingupaBasicTest3.

Temperaturea.

Setthechambersetpointbytouchingthei. greenboxabove“ChamberSetpoint.”

Enterthesetpointtemperatureintheii. keypadthatpopsup.

Temperaturerangesare1. programmablefrom50⁰C-150⁰C;however,onlytemperaturesbetween80⁰C-115⁰Carecalibratedtobewithin±0.5⁰Cofthesetpoint.

Page 20: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

20

DoNOTenterasetpoint2. temperaturehigherthantheactualtemperaturevaluewhichwillbeusedfortheAgingProcedure.ThePAVisdesignedtooperatemostefficientlywhenthePressureVesselandSampleRackarepreheatedtothetemperatureatwhichAgingwillbedone. NOTE:AttemptingtopreheatthePressureVesseland/orSampleRacktoahighertemperaturethantheAgingproceduretemperaturemayresultineitheralongerperiodoftimerequiredforthePAVtorecovertemperatureafterPressureisappliedand/orapoorercontrolofthechambertemperatureafterpressureisappliedtothechamber.

Whenfinished,touch“ENT.” iii.

Page 21: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

21

AgingDurationb.

Setthedesireddurationtoagethei. samplesbytouching“View”andthen“Time.”Touchthegreenboxabove“DurationofTestTime.”

Entertheamountoftimedesiredforii. samplepressurizationinthekeypadthatpopsup.

Timeisprogrammablefrom1to1. 99hours.

ASTMD6521,AASHTOR28,and2. EN14769havespecifictimerequirements.Checkthelateststandardforspecifictimesiftestingtotheserequirements.

Page 22: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

22

Select“Done”toreturntothe“PreheatStartup”c. screen.

Tochangethesoaktime,selectSettingsfromthed.Informationscreen.Thesoaktimecanbesetatanytimefrom5-20minutes.

Ifusingpreheatedsamples,thesoaki. timecanbereducedtofiveminutes.Ifthesamplesareatroomtemperature,select20minutesforthesoaktime.Thiswillpreheatthesamplestotheagetemperaturepriortoaging.

TheheatersonthePAVcanbesettoautomaticallye. shutoffonceatestiscompleted.Select“On”forthisoption.Selecting“Off”willcausetheheaterstocontinuerunningafteratestiscomplete.

Page 23: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

23

Insertthef. emptypanholderintothePAVandclosethevessellid. NOTE:DoNOTloadanysamplesintothePAVatthistime. NOTE:Bolt1inthediagramisapivotboltandshouldonlybeloosenedtoallowthevessellidtoslideopenandrestonthesupportblockandtoslideclosed.Thisboltshouldnotberemoved.Forcingthisboltmaydamagetheboltassembly.

Slidethevessellidclosedandinsertbolts.Theg. boltsonlyneedtobeengagedaturnortwo.Itisnotnecessaryatthistimetocompletelytightenthebolts.Closetheovenlidandsecureusingthelatches.

PivotBolt.

DONOTREMOVE.

Page 24: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

24

Select“Off”fromthe“PreheatStartup”screenh.tostartthepreheatingprocess.WhenthePAVisheating,theHeatModewilldisplay“Heating”andaflamesymbolwillappear.

Page 25: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

25

OncethePAVhasreachedthesetpoint,thedisplayi. willprompttheoperatortoinsertthespecimens,securethelid,andselect“Age.”

Carefullyremovethespecimenholderi. fromthePAVasitwillbeveryhot. NOTE:ThePAVneedstoexperienceheatlosssoitknowswhattodoafter“Age”isselected.Ifitdoesnotexperienceheatloss,thePAVgetsconfusedanddoesnotknowwhattodonext.ThisheatlossshouldhappenwhenthespecimensareloadedintothePAV.

Leavethevessellidopen.ThePAVisii. designedtorecoverfromtheheatlossduringthisstage.Ifthevessellidisclosedandthetemperaturedoesn’tdropenough,thePAVmaygiveerraticresultsandhaveadifficulttimerecoveringtothesetpoint.

Loadspecimensintorack.iii.

Page 26: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

26

Loadalltenpansintothesample1. holdereveniftherearenottensamplestoage.

Quicklyandcarefullyreturnthesampleiv. holdertothePAV.Thevesseltemperaturewilldecreasewhentheovenandvessellidsareopen.Toavoidalongtemperaturerecovery,attempttoloadthePAVandsecurethevesselandovenlidsasquicklyandassafelyaspossible. NOTE:DonotloadsamplesintothePAVuntilthePAVisheatedtothesetpointandthe“PreheatReady”screenisdisplayed.

Closethevessellidandsecuretheboltsv. accordingtothediagram.

Tighteninsequentialorderwiththevi. wrenchprovided.Goaroundtwicetotightenboltsto55inch/pounds.

PivotBolt.

DONOTREMOVE.

Page 27: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

27

Oncethevessellidissecured,closetheovenlidj.andsecurewiththelatches.

Immediatelyk. select“Age”fromthetouchscreen.DonotwaitforthetemperatureinthePAVtorecovertothesetpointpriortoselecting“Age.”

Thedisplaywillthenresembletheimagebelowl. andwillattempttorecovertheheatlostduringspecimenloading.

Page 28: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

28

Withintwohoursofselectingthe“Age”function,m.thePAVtemperaturewillrecoverthetemperaturelostwhileloadingthesamples.Whenthetemperaturerecoverstowithin0.3°Cofthesetpointtemperature,anapproximate20-minutedelaywillbegin.Thisisbuiltintothesoftwaretoallowthechambertemperaturetostabilize. Followingthis,thePAVwillbegintopressurize.ThetimerontheAgingHeatUpscreenwilldisplaythetotaltimefromwhenthe“Age”buttonwaspresseduntilthePAVstartstopressurize.WhenthePAVhaspressurizedtoslightlyabove2.00MPa,theagingcyclewillautomaticallybegin.Thescreenshownatthistimewillresemblethe“AgingPressurized”screen.

IfthePAVdoesnotreachthesetpointandi. pressurizewithinthetwo-hourtimeframeorhasdifficultyreturningtothesetpointafterpressurizing,checkthatthepowertothePAVis230-240volts.230voltselectricityistheminimumrequiredforthePAVtooperatecorrectly.

Ifthepowersupplyis230-240voltsand thetemperaturehasnotrecovered, verifythefollowing:

Besurethatthevessela. lidremainsopenwhenloadingsamplepansintothesampleholder.

Verifythatthesamplesareb.notbeingheatedabovethePAVsetpointpriortoinsertingthemintothePAV.

Page 29: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

29

Verify“Age”wasc. selectedfromthetouchscreenimmediatelyafterloadingthesamples,securingthevessel,andlatchingthelidasdescribedintheprevioussteps.

IfthePAVreturnstothesetpointbutii. doesnotpressurizecompletely,checkthattheairlineisconnectedandopentoallow340-350psiofairflow.

Page 30: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

30

Oncethisscreenappears,theCurrentErrorTimen. counterwillbegintocount.TheCurrentErrorTimecounterwillcontinuetocountuntilthechambertemperatureiswithin±0.5°CoftheChamberSetpoint.TheAccumulatedErrorTimecounterdoesnotincrementuntiltheCurrentErrorTimecounterstopscounting.Atthattime,theCurrentErrorTimecountervaluewillbeloadedintotheAccumulatedErrorTimecounter.IfthetemperatureorpressureinsidethePAVgoesoutsideoftheallowablelimits,theCurrentErrorTimecounterwillresetandbegincountinguntilthetemperatureand/orpressurehasreturnedtowithinallowablelimits,andtheAccumulatedErrorTimecounterwillagainbeupdated.TheAccumulatedErrorTimecounterwillonlyupdateaftertheCurrentErrorTimecounterhasquitcounting. NOTE:ItispossiblethattheCurrentErrorTimecounterwillcontinuetocountforseveralsecondsormorewhentheChamberTemperaturedisplayediswithin±0.5°CoftheChamberSetpoint.ThisisnormalandisduetothefactthattheChamberTemperaturedisplayis1decimalpointofprecisionandisroundedup,whiletheerrortimeisbasedon3decimalpointsofprecision.

Page 31: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

31

OncethePAVispressurized,itispossibletoselecto.“Graph”toseeagraphofreal-timedatabeingcollected.

Whenfinished,select“Done”toreturntothep. previousscreen.

Oncethetestiscomplete,thedisplaywillq.changetoresembletheonebelow.Pressurewillautomaticallyreleaseinanapproximatelinearstate andthePAVwillremainatthepreheatedtemperatureuntilthe“Reset”buttonispressed.

Ifthepressuredoesnotreleasewithinthei. allowabletimespecified,checkthattheflowadjustvalvelocatedonthefrontofthePAVisopentothenumberstatedonthecalibrationsticker.

Todownloaddata,select“Done”fromtheii. AgingCompletescreenandthen“Data”fromtheinformationscreen.

Page 32: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

32

Page 33: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

33

InsertastoragedeviceintotheUSBportiii. onthePAV. NOTE:AwindowwillappearanytimeastoragedeviceininsertedintotheUSBport.Donotpressanythingatthistime.Thewindowwillcloseaftertenseconds.

Afterthewindowcloses,select“Backupiv. DataUSB1.”

Datawillbesavedasa.CSVfile1. thatcanbeimportedtoMicrosoftExcel®. NOTE:The.csvfilescreatedusetheUSsystemofacommaasthedelineator.

Page 34: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

34

SettingupAutomaticStart4.

FromthePreheatStartUpscreen,select“View”a. andthen“Time”toreachthescreenthatwillallowoperatorstosetupthePAVtostartautomatically.

Setthecurrenttimeanddatebyselectingtheb.greenboxesandsettingthecurrentvaluesinthe“SetClockCurrentTime”box.

SetthetimeanddateforthePAVtostartc. automaticallyinthe“SetHeatStartTime”box.

Select“TimerDeactivated”toturnthetimerond.andenableautomaticpreheatatthetimeanddateimputed.

Page 35: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

35

SettingupRemoteCommunication5.

ThePAV3isnetwork-readyandcanbeviewedanda. controlledwithVNCViewersoftware.ATSstronglyrecommendscontactinganetworkadministratortosetupthisfeature.

UsingVNCViewersoftwarerequirestwoIPb.addresses(within1/100ofeachother),asubnetmask,andagatewayforeachPAV3onthenetwork.

HMIIPAddress:____.____.____.____i.

RMIPAddress:____.____.____.____ii.

SubnetMask:____.____.____.____iii.

Gateway:____.____.____.____iv.

TheHMIIPAddressistheaddressofthetouchc. screencontroller.

TheRMIPAddressistheaddressoftheRMinsided.ofthePAV.Thesetwoaddressesneedtobewithin1/100ofeachothertocommunicate.

Defaultsettingsare:i.

HMIIP:172.16.1.991.

RMIP:172.16.1.1002.

Sub:255.255.255.03.

CheckIPaddressesandsettingsine. three places.

i. OnthePAVdisplay,pressView/Languagesmenu.

1. Checkthedifferentlanguages andtheIPaddresses: HMI-172.16.1.99and RM-172.16.1.100.

Page 36: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

36

2. Thesesettingsmayneedtobe entered.

3. Check/setthedateandtime.

ii. OnthePAVdisplayscreen,usethe“Settings”icon.Thisisageariconthatappearsinthelowerrightcornerafterthearrowispressed.

Page 37: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

37

1.Enter“1”sixtimesonthe“virtual keyboard.”

Page 38: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

38

2.FindtheNetworkTabandmakesurethe spotfor“IPaddressgetfrombelow”is selected.

3.FillinthecorrectnumbersfortheIP address.172.16.1.99isthedefaultIP address.Thesubnetmaskshouldbeleft as 255.255.255.0.

4.GototheVNCServertab,itisthe secondtothelasttabandensure “StartVNC”isgreyedout.Ifitis not,selectittoenabletheVNCViewer toconnect.

Page 39: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

39

iii.OntheWatlowEZ_ZoneController: 1.OpenthesidepanelofthePAV.

2.FindtheRUIinside;itwilllooklikethe picturebelow:

NOTE:TheRUIhasmanysettings.Settings thatarenotmentionedintheinstructions shouldnotbechanged.Changesto anyothersettingsmaycausethemachine tofunctionimproperly.

Page 40: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

40

3.Pressthebottomtwobuttons simultaneously(theAdvancekeyand theDownkey)forthreesecondsuntil thedigitaldisplayshowsCOMin theupperdisplayandrUiinthelower display.Thecommunicationmenu shouldappear.

NOTE:UsetheUpandDownkeystomove throughthemenus.TheInfinitykeygoes backonemenulevel,theAdvancekey selectsamenu.

4.PresstheAdvancekeyonlyuntilCom1 appears.

5.Usetheuparrowtochangethisto Com2andusetheAdvancekeyto selectthismenu.

a.iP.F1isthefirstsetofnumbers intheIPaddress.UsetheUp andDownkeystochangethis valuetothedesiredvalue. Thedefaultvalueis172.Select theAdvancekeywhen complete.

Page 41: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

41

b.iP.F2isthesecondsetof numbersintheIPaddress.Use theUpandDownkeystochange thisvaluetothedesiredvalue. Thedefaultvalueis16.Select theAdvancekeywhen complete.

c.iP.F3isthethirdsetofnumbers intheIPaddress.UsetheUp andDownkeystochangethis valuetothedesiredvalue.The defaultvalueis1.Selectthe Advancekeywhencomplete.

d.iP.F4isthefourthsetof numbersintheIPaddress. UsetheUpandDownkeysto changethisvaluetothe desiredvalue.Thedefaultvalue is100.SelecttheAdvancekey whencomplete.

Page 42: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

42

e. Thesubnetcanbechangedthe samewayastheIPaddress. iP.FS1,iP.FS2,iP.FS3,andiP.FS4 arethevaluestochangethe subnet.Thedefaultvaluesare 255.255.255.0.Usethe Advancekeytocyclethrough theseoptionsandusetheUp andDownkeystochangetheir values.

f.Onceallchangeshavebeen made,holdintheInfinitykey fortwosecondstoreturnto thehomepage.

g.Onceallthevalueshavebeen enteredinthePAVunderthe languagessection,underthe settingssection,andintheRUI, poweroffthePAVandthen turnthePAVonagain.Ifa yellowerrorboxdisplays, checkthattheIPaddressesare within1/100ofeachotherand thesamevalueshavebeen enteredinallthreeplacesin thePAV.

Page 43: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

43

Section6:Verification6.1 Verifying TemperatureToverifytemperature:

1.Placethetemperatureprobeinsidethecalibrationblock.NotetheinsideofthePAVafterthepanholderisremoved.

NOTE:ThiscalibrationblockisfromtheATSPAVVerificationKit.SeeSection8.3.

2.PlacetheotherholeinthecalibrationblockovertheRTDon thebottomofthevessel.OncetheblockissecureontheRTDandthetemperatureprobe,slidethePAVlidclosed.Useasmallwrenchorpopsiclesticktokeepthelidfrompinching thewireonthetemperatureprobe.DONOTBOLTTHE LIDDOWN.

3.Verifythetemperature.

a.SetupthesetpointandstartthePAV.

Page 44: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

44

b.Verifythatthetemperaturedisplayonthethermometer matchesthetemperaturedisplayon thePAV.Allowtemperaturetoreachsetpointand stabilizeforonehour.

c.NoteanydifferenceandcallATSforcalibrationorusethedifferencetooffsetfuturetests.

NOTE:Donotverifytemperatureandpressureatthesametime. 6.2 Verifying Pressure

Toverifypressure:

1.Removethescrewintopofthevessellid.

2.Screwthepressuregageintothatspace.Makesurethelidis securedaccordingtothedirectionsinSection5.

3.ApplythepressuretothePAV.Allowpressuretoreach2.10.1MPaandstabilizeforfifteenminutes.

4.Verifythatthedisplayonthepressuregagematchesthe displayedpressureonthePAVdisplay.

5.NoteanydifferenceandcallATSforcalibrationifthedifferencefallsoutofspecification.

NOTE:ItisbesttohavethePAVsetpointatalowtemperatureandtheSoakTimesetforfiveminutes.

NOTE:Donotverifytemperatureandpressureatthesametime.

Page 45: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

45

Section7:Troubleshooting7.1 Temperature Does Not Recover to Setpoint

Check that the PAV is receiving 230V-240V

of power

NO YES

You will need a transformer. Call ATS

for a quote.

Did you wait till the PAV was at setpoint to load samples into

the PAV?

YES NO

Do not load samples into the PAV until

setpoint is reached and PAV prompts it

Are you preheating your samples above

the aging temperature?

YES NO

Only preheat your samples to the set-

point of the PAV using your lab oven

Are you leaving the vessel lid open while loading specimens in

pan rack?

NO YES

The PAV is designed to recover the heat

lost. Leave the vessel lid open

See notes on following page

Page 46: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

46

7.2 Troubleshooting the PAV ControllerThePAV3wasdesignedtoallowformaximumuserflexibilitywhilealsomeetingthecurrentASTM,AASHTO,andENspecifications.Thecontrollerisprogrammedtorespondtospecificsituationstoallowforallofthesecriteriatobemet.

OptimumAgingPerformanceisachievedbyfollowingthesesteps:

1.PreheatthePAVandemptythesampleracktothedesired conditioningtemperature.

2.WhenPAVshowsthePreheatReadyscreen:

a.Removethehotsamplerack.

b.Leavevessellidopen.

c.Loadsamplesintothesamplerack.

d.Placesamplerackinsidethepressurevessel.

e.Closelidandsecureboltsasdescribed.

f.Press“Age”immediatelyafterlidissecuredtobeginthe automaticagingprocess.DoNOTwaitfortemperature ofthePAVtorecovertothesetpointbeforepressing “Age”button.

3.DoNOTpreheatsamplespriortoplacinginPAVtoa temperaturehigherthanthetemperaturetobeusedtoagethesamplesinthePAV.Itisrecommendedtouseroomtemperaturesamples.Thebuilt-insoaktimewillpreheatthesamples accordingtospecification.

Page 47: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

47

7.3 Reattaching the Flow Adjust ValveIntheeventtheflowadjustvalvedetaches,pleaseusethefollowingstepstocorrectit.

1.Theflowadjustvalvehasthreeparts:theshaft,thebrasscollar,theadjustmentknob.

2.Thebrasscollarfitsovertheshaft,andtheadjustmentknobfitsoverthebrasscollar.Ifthebrasscollarisnotvisible,checktheinsideoftheadjustmentknob.Ifitisinsidetheadjustmentknob,usea1/16in.hexwrenchtoloosenthescrewinthe adjustmentvalvesothebrasscollarcanberemoved.

3.Rotatetheshaftfullyclockwisetoclosetheflowadjustvalve.

4.Placethebrasscollarbackontheshaftandensureitisagainstthevernierbarrel.Usea1/16in.hexwrenchinsertedintotheholeshownbelowandtightenthebrasscollarontotheshaft,ensuringitstaystightlyagainstthevernierbarrel.

Page 48: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

48

5.Lineuptheadjustmentknobwithzeroat12o’clock,andinsertoverthebrasscollarandtothelastlineofthevalvebody.Usingthe1/16in.hexwrench,insertintotheholeasshownbelowandtighten.

6.Turnthevalveslowlycounter-clockwisetoopen.Eachfullrotationshouldequalonelineonthebarrel.Rotatetheknobtothevaluerecordedonthecalibrationsticker.Ifthereisnovalueonthecalibrationsticker,thedefaultvalueis10.

Page 49: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

49

7.4 Updating the Software1.PowerupPAV3fromthefrontpanel.

2.UnscrewUSBportcoverandinstallflashdrivecontainingtheupdatedversionsoftwareintotheUSBport.

3.Inafewseconds,amenuwillpopup.ChooseDownload.

4.Anotherpop-upmenuwillappearpromptingforaPassword.Usingthekeyboarddisplayedonthescreenbelow,type 111111andselectOK.

5.Thenextscreenthatappearswillhave3options.

a.√ChecktheboxthatindicatesDownloadProjectFiles.

b.Unchecktheboxes:

i.ClearHistoryFiles.

ii.DownloadHistoryFiles.

6.Twosubdirectorieswillappear:pccardandusbdisk.

a.Selectusbdisk.

b.Selectdisk_a_1.

c.SelectOK.

7.Thenecessaryfileswillnowbedownloaded.

8.SelectView.

9.Atthetopofthescreen,thecurrentversionofsoftware shouldbeindicated.

10.SelectDone.

Page 50: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

50

11.VerifyIPaddresssettingsbyselectingView,thenLanguage.

12.Alllanguagesshouldhavethefirstrowindicating172,16,1,99andthesecondrowindicating 172,16,1,100unlesschangesweremadetosetupremotecommunication.Ifchangesweremadetosetupremote communication,verifythesettingsarecorrect.Ifnot correct,selecttheappropriatefieldandenterthecorrectsetting.

13.Thesoftwareupdateisnowcompleted.

Page 51: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

51

Section8:Maintenance8.1 Preventive Maintenance1.Keepthepressurevessellidboltslubricatedusingananti-seizeagent.Repeatedinsertionandextractionofstainlesssteelboltsinastainlesssteelvesselwithoutusinganti-seizepastecan causegalling.Lowerspeedinsertionandextractionofboltscanalsohelptopreventgalling.Use¼to½thespeedfor steel in steel.

2.PeriodicallylubricatetheO-ringwithsiliconelubricant.The O-ringismadeofsyntheticrubberandislocatedinthepressurevesselflange.DamagedorwornO-ringsshouldbereplacedwiththesamematerial.

8.2 Changing Fuses1.ControlCircuitFuse:Thecontrolcircuitfuseisa2amp,250volt5mmx20mmBUssmannGMA(orequivalent)fuse.Itislocatedattheleftrearofthesystem.

2.HeaterCircuitFuse:Theheatercircuitfuseisa10amp,250voltsemiconductor-typeBUssmanGMA2A(orequivalent)fuse.Itislocatedinsidetheleftcontrolcabinet.Iftheheatercircuitfuseshouldmalfunction,contacttheATSServiceDepartmentat+1-724-283-1212forinstructionsonhowtoservicethisfuse.

Page 52: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

52

8.3 Spare PartsSomepartsaresuppliedfromoutsideATS.Refertomanufacturer’sliterature.

ConsumablesPartNumberPAV1045PAV-PAN-KITPAV-LUBEPAV1038PAV1044PAV1068PAV1070PAV1072

AccessoriesPartNumberPAV-115-SU-TRANS

DescriptionPAVO-RingSetOf(10)SpecimenPansAnti-SeizeLubricantFuse,2A5X20MMSlowBlowF2 Fuse,10Amp250VoltLed,Tri-ColorDiffusedLed,RedDiffused4Led,GreenDiffused3

DescriptionStep-UpTransformer-Enablescustom-erswithpowerof115VAC,1Ph,60Hz,10AMPtooperatePAV(powerrequire-mentsof240VAC,1Ph,50/60Hz,10AMP.)

PAV-BUCK-TRANS BuckBoostTransformer-Increasesexisting208voltsto230volts

PAV-UPS-BU UPSBatteryBackUpSystem-PreventsPowerFailures,Sags,Surges,UnderandOverVoltages.

PAV-VER-KIT PAVVerificationKit-Includescalibrationblock,thermometricdevice,andpressuregage

Page 53: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

53

AccessoriesPartNumber DescriptionPAV-VER-KIT-NIST NISTPAVVerificationKit-Includescali-

brationblock,NISTcertifiedthermometricdevice,andNISTcertifiedpressuregage

PAV-STAND PAVSteelStand-StandforPAVPAV-CART HydrolicLiftCart-foreasyheightadjust-

mentofthePAVPAV-LEVEL Levelwithparalellbar101521 RTDThermometerW/Probe,Nist100068 DigitalPressureGage-NIST100065 Thermometer100067 ImmersionProbeRTD-usewith100065100069 Digital Pressure GagePAV-CAL-BLOCK CalibrationblockforPAV

SparePartsPartNumber DescriptionPAV-SS-REG-Europe SingleStageRegulatorforuseinEuropePAV-SS-REGULATOR SingleStageRegulatorPAV-WRENCH Wrench,HexSocket1/2InDrivePAV1148 LevelingRing,4-8310,Rv.2,It1PAV9012 Tool,PanHandlingPAV9015 Adapter,CGAPAV9016 Wrench,HexSocket1/2InDrivePAV9017 Wrench,Allen1/8InchPAV9020 Hose,HighPressureAssyPAV-PAN-HOLDER PAVSpecimenRack

Page 54: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed

54

AppendixA:Warranty

WarrantyStatementYourAppliedTestSystemsproducthasbeenmanufacturedandinspectedbyexperiencedcraftsmen.AppliedTestSystemswarrants,fortheoriginalpurchaser,eachproducttobefreefromdefectsinmaterialandworkmanshipforaperiodofthirteen(13)monthsfromdateofshipmentortwelve(12)monthsfromdateofinstallationwhichevercomesfirst.Thiswarrantydoesnotapplytofailurescausedbynormalusage,misuse,orrepairorservicebyunauthorizedpersonnel,nordoesitcoverlimitedlifeelectricalcomponentswhichdeterioratewithagesuchastubes,lamps,fuses,andheaters.ThewarrantydoesnotextendtoproductsnotmanufacturedorassembledbyAppliedTestSystems.

Thiswarrantyisexpresslylimitedtotherepair,replacement,oradjustmentoftheproductatAppliedTestSystems’option.TheproductmustbereturnedtotheAppliedTestSystemsfactoryoranauthorizedrepaircenter.AppliedTestSystemsshallnotbeliableforanylabor,transportation,orinstallationcoststhatmayariseinconnectionwiththeproductorreturn.

Toobtainwarrantyservice:

1.AppliedTestSystemsmustbepromptlynotifiedinwritingof thedefect.

2.Uponreceiptofwrittenauthorization,saiddefective equipmentisreturnedasdirected,withtransportation chargesprepaidbythebuyerand–

3.AppliedTestSystemsexaminationofsuchequipmentdiscloses toitssatisfactionthatthedefectexistsandwasnotcaused bynegligence,misuse,improperinstallation,accident,or unauthorizedrepairoralteration.

Thiswarrantyisinlieuofallotherwarranties,expressedorimplied,includingtheimpliedwarrantyofmerchantabilityorfitnessforparticularpurpose.InnoeventshallAppliedTestSystemsbeliablefordirect,indirect,special,incidental,collateralorconsequentialdamages.

TheaforementionedprovisionsdonotextendtheoriginalwarrantyperiodofanyarticlethathasbeeneitherrepairedorreplacedbyAppliedTestSystems.

AppliedTestSystemsreservestherighttochangepublishedspecifications.

Page 55: Pressure Aging Vessel PAV3 · Section 3: System Overview 3.1 General Description The ATS Pressure Aging Vessel (PAV) is designed specifically to prepare specimens for tests developed