DPG Frühjahrstagung 2013, Dresden
PMT series tests for the XENON1T Water Cherenkov Muon Veto
Christopher W. Geis
Johannes-Gutenberg Universität MainzInstitut für Physik
2013/03/05
on behalf of the XENON collaboration
![Page 2: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/2.jpg)
DPG Frühjahrstagung 2013, Dresden Chr. Geis 1 / 15
Outline
● The XENON1T Muon Veto● PMT test facility at JGU Mainz
● Gain● Darkrate● Afterpulses
● Summary
![Page 3: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/3.jpg)
DPG Frühjahrstagung 2013, Dresden Chr. Geis 2 / 15
Goal of XENON1T:
Reach SI WIMP-Nucleus cross section of
σ ≈ 2 ∙ 10-47 cm²
The XENON1T Experiment
Increase of two orders of magnitude in sensivity requires a similar reduction in
background
2010-201x:XENON1T
![Page 4: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/4.jpg)
DPG Frühjahrstagung 2013, Dresden Chr. Geis 3 / 15
The XENON1T Muon Veto
Tagging of muons via their Cherenkov radiation emitted during their transit through the water tank.
Hamamatsu R5912 HQE8" PMTs
- 750m³ Volume- Walls cladded with highly reflective foil
84 PMTs located in three arrays: - Top: Ring of 24 PMTs at 4.5m radius, looking down - Lateral: 3 times 12 PMTs in equaly spaced rings - Bottom: Ring of 24 PMTs at 4.5m radius, looking up
![Page 5: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/5.jpg)
Chr. Geis 4 / 15
Muon Veto PMTs
→Performance of each PMT has to be tested before final mounting in the Muon Veto!
PMT holding construction(Stainless steel and POM)
Photocathode with high sensitivy to Cherenkov-UV light
- Cherenkov radiation is UV.- Nearly 100% reflection of foil for λ > 380nm
→Use of PMT with high QE for 350-450nm
![Page 6: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/6.jpg)
Chr. Geis 5 / 15
Test facility in Mainz
Tests take place in 4 Aluminium boxes
- 60 x 60 x 60 cm - painted black inside - feedthroughs for HV, Signal and LED
![Page 7: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/7.jpg)
DPG Frühjahrstagung 2013, Dresden Chr. Geis 6 / 15
LED calibration
PMMA sheathed fiber bundle with 4 tails - Litetec, United Kingdom - 2.2mm diameter, 1mm core - polished flat ends
Fiber bundle
470nm LED BNC input connector
![Page 8: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/8.jpg)
DPG Frühjahrstagung 2013, Dresden Chr. Geis 7 / 15
Data acquisition logic
Pulse generator LED Box
ADC
PMTTest
Boxes
HV PowersupplyDAQ PC
Trigger Signal
CAEN V1724 fADC4 channels
![Page 9: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/9.jpg)
DPG Frühjahrstagung 2013, Dresden Chr. Geis 8 / 15
Test protocol
Tests done for each PMT:
- Gain for nominal voltages ± 50, ± 100, ± 150 V
- Dark rate for nominal voltages ± 50, ± 100, ± 150 V
- Afterpulses for nominal voltage
![Page 10: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/10.jpg)
DPG Frühjahrstagung 2013, Dresden Chr. Geis 9 / 15
Gain calculation
Used fit formula:
![Page 11: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/11.jpg)
DPG Frühjahrstagung 2013, Dresden Chr. Geis 10 / 15
Gain vs. different HV
![Page 12: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/12.jpg)
Chr. Geis 11 / 15
Dark rate
1 s.p.e = 20.8 ADC channels
![Page 13: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/13.jpg)
DPG Frühjahrstagung 2013, Dresden Chr. Geis 12 / 15
Afterpulses
We define an afterpulse as a pulse with two consecutive samples over a certain threshold.
Afterpulses occur, when photoelectrons ionizing residual gas molecules inside the photomulti-plier tube.
Baseline window: 0 to 0.8 µsPrimary peak window: Pos - 0.1µs to Pos + 0.4µsAfterpulses window: Pos + 0.4µs to Pos + 20µs
![Page 14: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/14.jpg)
DPG Frühjahrstagung 2013, Dresden Chr. Geis 13 / 15
Afterpulses charge
Charges of all afterpulses in one event are summed up.
Fraction afterpulses/primary pulse is here 8%
![Page 15: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/15.jpg)
DPG Frühjahrstagung 2013, Dresden Chr. Geis 14 / 15
Afterpulses charge
Charge fraction Afterpulse/Primary pulse calculated for each PMT:
![Page 16: PMT series tests for the XENON1T Water …xenon.physik.uni-mainz.de/presentations/20130305_Talk...2013/03/05 · DPG Frühjahrstagung 2013, Dresden PMT series tests for the XENON1T](https://reader034.vdocuments.us/reader034/viewer/2022042306/5ed27d73995ce412b22babc8/html5/thumbnails/16.jpg)
DPG Frühjahrstagung 2013, Dresden Chr. Geis 15 / 15
Summary and Outlook
Almost all 90 PMTs tested so far (in Mainz and Bologna)
Test results will lead to decisions which PMT will be mounted in the Muon Veto and which will be used as spare
Just one single PMT with unexpected behaviour observed so far.
→Additional long term test with this PMT started
→ Construction of water tank starting in April 2013. Mounting PMTs during autumn!
→ First dark matter data taking by XENON1T expected in 2014