platelet function analysis using multiple electrode aggregometry (multiplate ® )
TRANSCRIPT
Platelet Function analysis using Multiple Electrode Aggregometry (Multiplate®)
► Multiplate®
platelet function analysis in whole bloodbased on impedance aggregometry
► Multiplate® instrument
5 channels for parallel tests
easy to use Windows XP based software
automatic analysis and documentation
duplicate sensor forinternal quality control
electronic pipetting
► applications
Multiplate is already one of the most widely used platelet
function tests
put the test cell into the measuring position
attach the sensorcable
pipette 300 µl of saline+ 300 µl of blood*
allow 3 minutes for warming
and equilibration
add the activator
after 6 minutes: print the results discard the test cell
* usually hirudin or heparin blood
► Performing the test
► Multiplate® reagents
comprehensive line of dedicated
reagents
TRAP
platelet activation
GpIIb /IIIa receptor
exposure
degranulation
ADP
Arachidonic Acid
Collagen
ArA1
COX
TXA2
TXA2
COLtest
ASPItest
TRAPtest
ADPtestADPtest
PGE1 ADPtest HS (ADP + PGE1)
activation inhibition
GpIIb/IIIa antagonists:
Reopro ® (abciximab)Aggrastat ® (Tirofiban)Integrillin ® (Eptifibatid)
Aspirin ®NSAID
Clopidogrel PrasugrelCangrelor
1 release of arachidonic acid
► Multiplate tests 1/2
rest
ing
plat
elet
activated platelet
► Multiplate tests 2/2
test activation sensitivity not sensitive for
ASPItest arachidonic acid: is converted to TXA2 by platelet-own cyclooxygenase
aspirin, IIb/IIIa antagonists
clopidogrel, vWF
ADPtest ADP: binds onto platelet ADP receptors clopidogrel, IIb/IIIa antagonists
aspirin, vWF
ADPtest HS ADP + prostaglandin E1 (Prostaglandin is a natural inhibitor and enhances the sensitivity of the assay for clopidogrel)
clopidogrel, IIb/IIIa antagonists
aspirin, vWF
TRAPtest TRAP-6 (thrombin receptor activating peptide): TRAP-6 is a potent agonist which mimicks the platelet-activating action of thrombin
IIb/IIIa antagonists vWF, aspirin, clopidogrel (weak effect on TRAPtest)
COLtest collagen: collagen activates platelet and triggers a release of arachidonic acid from the platelet membrane, which is converted to TXA2 by the Cyclooxygenase
aspirin, IIb/IIIa antagonists
clopidogrel, vWF
RISTOtest Ristocetin: vWF dependent platelet activation via the GpIb receptor
Bernard-Soulier syndrome, severe vWD, aspirin
mild vWD
► multiple electrode aggregometry = MEA
test 1+2
- one Multiplate test cell incorporates two independent sensor units.
- the increase of impedance due to the attachment of platelets to the electrodes is detected for each sensor unit separately and transformed to arbitrary aggregation units (AU) that are plotted against time.
- the duplicate sensors serve as an internal control
- during each measurement Pearson´s correlation coefficient of single measurements of the curves assessed by the two electrode pairs and the difference of the two AUCs is calculated. The result is flagged if the values are outside of the acceptance range (correlation coefficient <0.98, difference to the mean curve >20%).
D. Sibbing et al, Thromb Haemost. 2008 Jan;99(1):121-6.
► blood sample for Multiplate analysis
- citrated blood is the typical sample anticoagulant used for platelet function analysis
- citrate complexes approx. 98% of the free calcium in the sample
- however calcium is an important second messenger of platelet activation
- therefore citrate has the potential to disturb platelet function tests
- for Multiplate analysis a tight attachment of platelets onto the impedance sensor is mandatory, which enhances the effect of calcium depletion on this method
- therefore the analysis of blood anticoagulated by hirudin or heparin is recommended
- hirudin blood collection tubes are commercially available from the manufacturer of Multiplate
„Since hirudin and other direct acting thrombin inhibitors do not reduce the concentration of divalent cations in plasma, these can be considered to provide a more physiological environment than the use of sodium citrate. On this basis, we would agree with the recommendation by Dynabyte that the use of hirudin or another direct acting thrombin inhibitor is preferable for measurements of platelet aggregation performed by MEA (multiple eletrode aggregometry = Multiplate).“ A. Johnson, Thromb Haemost. 2008 Jun;99(6):1127-9.
In PRP not all platelets may be present, the most active platelets and larger platelets may have been lost during centrifugation
In PRP platelets are studied in isolation which is NOT their natural situation
In whole blood other blood cells are also present - erythrocytes (a potential source of ADP and ATP and a means of removing adenosine) and leucocytes which also influence platelet aggregation.
S. Heptinstal et al
► Why study platelet function in whole blood rather than in PRP?
TRAPtest ASPItest ADPtest
no platelet inhibition
100 mg aspirin qd
75 mg clopidogrel qd
100 mg aspirin + 75 mg clopidogrel qd
tirofiban (Aggrastat® i.v.)
17 U
134 U139 U
98 U 89 U
31 U
8 U
88 U
17 U
113 U 102 U 89 U
7 U 3 U 3 U
► examples
A.B. *1971
very weak aggregation in all tests due to Glanzman Thrombasthenia
universityclinic
Frankurt / Main Hannover
medical school
Essen university clinic
► near-patient application of Multiplate
► Multiplate in interventional cardiology / neuroradiology
Aggrastat Aggrastat
2 patients before and
after aggrastat loading
dose
case report by Dr. med. Stefanie Müller-Schunk, Neuroradiologie, Klinikum der Universität München
aspirin 500 mg
i.v.
50 min after aspirin
application
case reports by Dr. med. Stefanie Müller-Schunk, Neuroradiologie, Klinikum der Universität München
aspirin 500 mg
i.v.
2 patients before and
after aspirin 500 mg i.v.
clopidogrclopidogrel el 2x75mg2x75mg
clopidogreclopidogrel 75mgl 75mg
Non-response
Response
► increase of clopidogrel dose
Stent thrombosis
0 6 min
AU
Clopidogrel 150mg/d
AU
0 6 min
Prasugrel 10mg/dF. Krötz, Kardiologie, Klinikum der Universität München
► change from clopidogrel to prasugrel
cartridge-basedplatelet
function tests
optical aggregometry
flow cytometry
Platelet Function
Analysis Market
• simple
• whole-blood-based
• 1-2 different tests available
• single chanel
• complicated
• low standardisation
• used in research and reference labs
• easy-to-use
• multi-channel
• several tests available
• open system
Multiplate ®
Multiplate® is one of the newer platelet function tests.
The system has a high sensitivity for anti-platelet drugs (Aspirin, Clopidogrel, Prasugrel, IIbIIIa antagonists).
It is widely used in laboratory and near-patient settings.
Studies have shown a good predictivity towards thromboembolism in platelet non-responsiveness (10 fold enhanced risk for stent thrombosis in a study including 1608 patients following PCI).
Studies have shown a predictivity for transfusion requirements during / after cardiac surgery.
Further applications include the assessment of platelet disorders, drug interactions and animal models.
► Summary