ph analzyer

28
pH Analyzer

Upload: mohamed-sarhan

Post on 19-Mar-2017

23 views

Category:

Engineering


0 download

TRANSCRIPT

Page 1: pH Analzyer

pH Analyzer

Page 2: pH Analzyer

WHAT IS pH ?

pH is a measure of the acidity or alkalinity of a water solution. The acidity or alkalinity of a water solution is determined by the relative number of hydrogen ions (H+) or hydroxyl ions (OH-) present. Acidic solutions have a higher relative number of hydrogen ions, while alkaline (also called basic) solutions have a higher relative number of hydroxyl ions.

Acids are substances which either dissociate (split apart) to release hydrogen ions or react with water to form hydrogen ions.

Bases are substances that dissociate to release hydroxyl ions or react with water to form hydroxyl ions.

Page 3: pH Analzyer

In practice, pH is often assumed to be the negative logarithm of the hydrogen ion concentration:

pH = - log10 [H+]

Page 4: pH Analzyer

At 25°C, a neutral solution has a pH of 7.0, while solutions with pH < 7 are acidic and solutions with pH > 7 are alkaline. The normal overall pH range is 0 to 14 pH, although solutions containing non-water solvents can have pH values outside this range.

Page 5: pH Analzyer

The cell consists of a measuring electrode, a reference electrode, atemperature sensing element, and the liquid being measured. The voltage of the cell is directly proportional to the pH of the liquid. The pH meter measures the voltage and uses a temperature-dependent factor to convert the voltage to pH. Because the cell has high internal resistance, the pH meter must have a very high input impedance.

THE MEASUREMENT OF pH

Page 6: pH Analzyer

The pH electrode uses a specially formulated, pH sensitive glass in contact with the solution, which develops a potential (voltage) proportional to the pH of the solution.

A glass pH electrode consists of an inert glass tube with a pH sensitive glass tip, either hemispherical (bulb) or flat in shape, blown onto it. The tip contains a fill solution with a known pH,

Page 7: pH Analzyer

The reference electrode is designed to maintain Constant potential at any given temperature Serves to complete the pH measuring circuit within the solution It provides a known reference potential for the pH electrode. The difference in the potentials of the pH and reference electrodes provides a millivolt signal proportional to pH..

Page 8: pH Analzyer

Most pH sensors are designed to produce a 0 mV signal at 7.0 pH, with a (theoretically ideal) slope (sensitivity) of -59.16 mV / pH at 25°C

Page 9: pH Analzyer

The millivolt signals produced by the pH and reference electrodes are temperature dependent. However, the pH and reference electrode combination exhibits an isopotential point, which is a pH and millivolt potential at which the potential is constant with temperature changes. The isopotential point is most frequently designed to be at 7.0 pH and 0 mV. Using the isopotential point with a theoretical knowledge of electrode behavior makes it possible to compensate (correct) the pH measurement at any temperature to a reference temperature (usually 25°C),

Page 10: pH Analzyer

Compensation (correct) the pH measurement

Compensation (correct) the pH measurement means to remove the temperature dependence in the intercept and to shift the origin of the axes to pH 7. The result is plotted in the Figure. Two lines appear on the graph. One line shows how cell voltage changes with pH at 25°C, and the other line shows the relationship at 50°C.

The lines, which are commonly called isotherms, intersect at the point (pH 7, 0 mV). An entire family of curves, each having a slope determined by the temperature and all passing through the point (pH 7, 0 mV) can be drawn on the graph.

Shows why temperature is important in making pH measurements. When temperature changes, the slope of the isotherm changes. Therefore, a given cell voltage corresponds to a different pH value, depending on the temperature.

Page 11: pH Analzyer

Reference Poisoning

The mechanism of reference poisoning is a conversion of the reference from a silver-silver chloride based electrode to an electrode based on a different silver compound.

These ions include bromide, iodide, and sulfide ions. When these ions enter the fill solution, they form insoluble precipitates with the silver ions in the fill solution.

But there is no initial effect on the potential of the reference, because the silver ions lost to precipitation are replenished by silver ions dissolving off the silver chloride coating of the silver wire. It is not until the silver chloride coating is completely lost that a large change in the potential so the reference electrode must be replaced.

To counter this effect, multiple junction reference electrodes are used, which consist of two or more liquid junctions and fill solutions to slow the progress of the poisoning ions. Gelling of the reference fill solution is also used to prevent the transport of poisoning ions by convection.

Page 12: pH Analzyer

Liquid Junction Potential

Potassium chloride is chosen for the fill solution not only for its ability to solubilize silver ion, but also because it is equitransferent. Equitransferent means that the positive potassium ion and the negative chloride ion diffuse through a water solution at nearly the same rate. When diffusing through the liquid junction of a reference electrode, both the positive and negative ions tend to move together and maintain a net zero charge at all points within the liquid junction.However, not all solutions are equitransferent, and when the process solution is not, a liquid junction potential can result. When a positive ion diffuses through the liquid junction faster than a negative ion, or vice versa, a charge imbalance will result. This gives rise to an opposing potential, which is liquid junction potential. This potential gets added to the potentials of the pH and reference electrodes and causes an offset to the pH measurement of typically a few tenths of a pH.

Page 13: pH Analzyer

pH IN AQUEOUS SOLUTIONSPractical pH Measurement RangeAlthough the range of pH measurements is defined to be 0 to 14 pH, solutions with pH values near the extremes of this range are often better measured using conductivity.

Page 14: pH Analzyer

Tx: Ph202 YOKOGAWASensor: PH8EHPPlace :DEMI WATER

Page 15: pH Analzyer

The Model PH8EHP pH Sensor is used to configure a pure water pH transmissionsystem. • Measurement principle: Glass electrode method• Sample temperature range : 0 to 508C• KCl solution consumption: Maximum 3 ml/day• Sample pressure : Under atmospheric pressure• Sample flow rate : 30 to 600 ml/minNote : The maximum flow rate is restricted by the measurement solution conductivity.

Process application:why put at this phase in the process?

Page 16: pH Analzyer

XMT-PCALIBRATING TEMPERATURE

Page 17: pH Analzyer

XMT-PCALIBRATING pH

For pH sensors, two-point buffer calibration is standard. Both automatic calibration and manual calibration are available

Auto calibration avoids common pitfalls and reduces errors. Its use is recommended. In auto calibration the Solu Comp Xmt calculates the actual pH of the buffer from the nominal value entered by the user and does not accept calibration data until readings are stable

Manual calibration the user enters buffer values and judges when readings are stable. The pH reading can also be standardized, that is, forced to match the reading from a referee instrument. Finally, if the user knows the electrode slope (at 25°C), he can enter it directly.

Page 18: pH Analzyer

XMT-P Auto CALIBRATING pH

Page 19: pH Analzyer

XMT-P Auto CALIBRATING pH

Page 20: pH Analzyer

XMT-P Auto CALIBRATING pH

Page 21: pH Analzyer

XMT-P Manual CALIBRATING pH

Page 22: pH Analzyer

XMT-P Manual CALIBRATING pH

Page 23: pH Analzyer

XMT-P ENTERING A KNOWN SLOPE VALUE CALIBRATING pH

If the electrode slope is known from other measurements, it can be entered directly into the transmitter. The slope must be entered as the slope at 25°C. To calculate the slope at 25°C from the slope at temperature t°C, use the equation:

Page 24: pH Analzyer

XMT-P TROUBLESHOOTING

Page 25: pH Analzyer

Broken pH Glass and pH Glass Z High

These messages mean that the pH sensor glass impedance is outside the programmed limits. To read the impedance go to the main display and press until Glass Imp appears in the display. The default lower limit is 10 MΩ. The default upperlimit is 1000 MΩ. Low glass impedance means the glass membrane — the sensing element in a pH sensor — is cracked or broken. High glass impedance means the membrane is aging and nearing the end of its useful life. High impedance can also mean the pH sensor is not completely submerged in the process liquid.

1. Check the sensor wiring, including connections in a junction box.2. Verify that the sensor is completely submerged in the process liquid.3. Check the sensor response in buffers. If the sensor can be calibrated, it is in satisfactory condition. To disable the fault message, reprogram the glass impedance limits

XMT-P TROUBLESHOOTING

Page 26: pH Analzyer

PH202G TROUBLESHOOTING

Page 27: pH Analzyer

PH202G TROUBLESHOOTING

Page 28: pH Analzyer