An Orthopaedic Surgeon’s

Perspective on Implant Design

Neil Rushton

Emeritus Professor of Orthopaedics

University of Cambridge

Page 2: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

The Acetabular Component Remains the Weak Link

Owen TD, Moran CG, Smith SR, Pinder IM. Results of uncemented porous-coated anatomic total hip replacement. J Bone Joint Surg Br. 1994 Mar;76(2):258-62.

The Swedish National Arthroplasty Register

Annual Report 2003 (published in May 2004) Website: www.jru.orthop.gu.se

r

Page 3: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside
Page 4: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Particle disease

Page 5: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

All articulating materials wear and

therefore produce particles

• Polyethylene

• Bone cement

• Metals

• Ceramics

• “Biological” materials

• Suture & ligament

materials

• Ectopic tissue

Page 6: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

The effect of persistent particles

Surprising extent of the

biological effect of wear

particles

Bone defects heal if

movement is reduced

(Per Aspenberg)

Particles inhibit healing

Page 7: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Bone Atrophy

Page 8: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Bone wasting

Page 9: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Optifix Prosthesis

% BMD Difference at 6 months

-20

-15

-10

-5

0

5

10

R1 R2 R3 R4 R5 R6 R7

% BMD

Difference

Page 10: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside
Page 11: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Particles and

Stress Shielding

are Additive

Page 12: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Aims for a New Hip

• Modulus-matched acetabular component

• Bone is subject to normal forces

• Biological fixation

• Large heads

• Conserve bone

Page 13: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

The Original

Page 14: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Hole for bifurcated peg

Compare cut-out size

Page 15: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Materials used for Cambridge Cup

• Carbon Fibre Reinforced

Polybutylene

Terephthalate (CF-PBT)

• Ultra High Density

Polyethylene (UHDPE)

• Hydroxyapatite (HA)

coating

Page 16: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

100º arc

Recessed arms

2 part snap fit

Page 17: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Early implantation

Page 18: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Retrieval Analysis

Cambridge Hip

Page 19: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Retrieved Specimen

Page 20: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Embedded Specimen

Page 21: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Specimen in Accutom

Page 22: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Three top cuts

Page 23: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Sections from face, spikes and cut-out

Page 24: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

HA

CFRPBT

Bone

Marrow

1.5mm

1.1mm

Image Analysis

Page 25: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Histology comparison

HA coated Uncoated

Page 26: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

HA Resorption

Page 27: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

13 Year Survivor

Page 28: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Clinical trial :

50 female patients

10+ year prospective study

Displaced subcapital femoral neck fracture

Addenbrooke’s Hospital, Cambridge and

St Helier Hospital, Carshalton

>70yrs of age

All walking with no more than one cane prior to fracture

All of sound mind, able to give informed consent

and agreeable to follow-up

Page 29: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

DEXA

Page 30: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Comparison With Other Results

Overall Mean %

change of BMD at 6

months

Overall Mean %

change of BMD at

24 months

Cambridge cup -13% -6%

Females with

Mecron threaded

titanium cup

(Sabo et al10 )

-19% -20%

N.B. surgical

bone loss

Page 31: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Improvements in

Design

Page 32: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

FEA

Page 33: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Finite Element Analysis

• Partners: Exponent

Philadelphia

• Sponsored by Stryker

• At least 500,000

elements

• FE studies agree with

our clinical study

Page 34: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Cadaver Trials

Page 35: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Cadaveric Trials

Page 36: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Prevention of Binding

• Precision reaming

• Relief of arms

Page 37: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

MITCH PCR™ Cup

Page 38: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Cup Deformation Analysis

Interference 0mm

Polar contact

Interference 0.1mm

Polar contact

Interference 0.3mm

Polar contact

Polar contact

Interference 0.5mm Interference 0.7mm

Polar contact Annular contact

Interference 0.9mm

Page 39: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Spherical head

Profile of cup

Transition Zone

Bearing shape

Polar bearing

* Care required with reaming

Page 40: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Wear

Page 41: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Wear Studies 25 million cycles

Wear rate of 0.05mm3 +/- 0.0.05 mm3/ mc. (volumes actually negative owing to fluid

absorption)

Surface roughness at start = 944nm, at end of test = 605nm with machine marks still

evident.

Page 42: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Independent Wear Tests

Page 43: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Results for 25 Million Cycles

Page 44: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

MITCH

C/F PEEK

Titanium

HA

Page 45: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Instrument Development

Mitch PCR™

Page 46: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Clinical Trial

Page 47: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

02 004 Pre-operative AP Radiograph

Page 48: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

02 004 6 Week Post-operative AP Radiograph

Page 49: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

02 004 3 Month Post-operative Lateral Radiograph

Page 50: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Clinical trial

Page 51: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Squeaking!

Page 52: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

REF squeaker After 1 year Did not improve – noise disruptive Arthroscopy – no abnormality Revision- excellent fixation minimal black material not tight Lab analysis - small patch of wear outside congruent zone some wear around keyhole

NR squeaker After 1 year Improving – not disruptive No surgery

Page 53: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside
Page 54: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

The original concept

was to develop a system

that was sympathetic

with the existing tissue

Page 55: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

*Modular system

*Interchangeable components

*Augments existing Stryker

components

Page 56: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

MITCH PER (Ceramic version) MITCH PCR

Page 57: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Some Lessons from

Recent Experiences

Page 58: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Transformation of Cells in

Response to Metal Ions

Page 59: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside
Page 60: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside
Page 61: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Treatment of Avascular Necrosis

Page 62: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Pseudotumour • Destructive, locally

invasive tumour

• Difficult to treat

• Class action

• Cessation of use

Page 63: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Capital Hip

• Made by 3M

• Designed to be cheap

• Copy of Charnley Hip

• High failure rate

• Class action

• 3M leave orthopaedic

business

Page 64: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Cosmetic Products • Cosmetic surgery is

going from boom to bust

• Competition to make it cheaper (or make more profit)

• Some Dow Corning implants failed

• Class action

• Dow Corning out of orthopaedics

Page 65: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

The True Story

• Class action

• Claimed to interfere

with immune defences

(not correct)

• Claimed to rupture more

frequently

(not correct)

• $4 billion paid in

settlements!

Page 66: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Swanson Prosthesis

Page 67: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Poly Implant Prosthese

• Made cheaply

• Non-medical

grade filler

• Poor quality

• High failure rate

• Company

vanished!

Page 68: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Poly Implant Prosthese

• Made cheaply

• Non-medical

grade filler

• Poor quality

• High failure rate

• Company

vanished!

Politicians and public do not differentiate between different prostheses!

Page 69: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Titanium Troubles

Page 70: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Trunnion Wear • Titanium and Ti alloys

are not suitable for some

Morse taper junctions

• Worse problems when

counterface is CoCr

• Leads to dramatic bone

loss

• Biological mechanism

not known

Page 71: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Some Messages

• Always think of the ultimate application

Page 72: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Some Messages

• Always think of the ultimate application

• How will the product be used?

• Are special instruments required?

Page 73: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Some Messages

• Always think of the ultimate application

• How will the product be used?

• Are special instruments required?

• How will the device fail?

• Is there a satisfactory solution when it has

failed?

Page 74: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Some Messages

• Always think of the ultimate application

• How will the product be used?

• Are special instruments required?

• How will the device fail?

• Is there a satisfactory solution when it has failed?

• What is the real definition of

biocompatibility?

Page 75: Perspective on Implant Design - BioTiNet · 2017-02-28 · Perspective on Implant Design Neil Rushton Emeritus Professor of Orthopaedics ... Lab analysis - small patch of wear outside

Some Messages

• Always think of the ultimate application

• How will the product be used?

• Are special instruments required?

• How will the device fail?

• Is there a satisfactory solution when it has failed?

• What is the real definition of biocompatibility?

• Do not believe your own marketing!