perfusing dentine with horse serum or physiologic saline: its effect on adhesion of dentine bonding...
TRANSCRIPT
596 © 1998 Blackwell Science Ltd
Journal of Oral Rehabilitation 1998 25; 596–602
Perfusing dentine with horse serum or physiologic saline:its effect on adhesion of dentine bonding agentsC . A U G U S T I N , S . J . PA U L , H . L U T H Y & P. S C H A R E R University of Zurich, Center for Dentaland Oral Medicine, Department of Fixed and Removable Prosthodontics and Dental Material Sciences, Zurich, Switzerland
SUMMARY Freshly prepared dentine specimens of
human teeth were perfused with either horse serum
or physiologic saline. After application of AllBond2®,
ART Bond®, Syntac® or an experimental dentine
bonding agent called P-Bond, a composite cylinder
was added and cured at the same time. After 1500
thermal cycles with constant imitation of intrapulpal
pressure, the shear bond strengths were measured.
Resulting shear bond strength values were analysed
with Students t-test or Mann–Whitney Test. The
Introduction
The testing of bond strength is a commonly used methodto evaluate dentine bonding agents (Jendresen et al.,1995). However, concern has been expressed if thistesting technique is a good indicator for the clinicalefficiency of dentinal adhesives because of thesometimes poor correlation between the in vitro resultsand the in vivo performance of these materials.
Intra-pulpal pressure seems to be very important inaltering the surface of vital teeth for bonding procedures.If the dentinal surface is acid-etched in the same wayas it is appropriate on enamel to obtain a good bond(Buonocore, 1955), dentinal fluid, under pressure, willleak out of numerous cut tubules. This may considerablychange the conditions of the chemical reaction of theadhesive resin with the dentine. Therefore, seepingdentinal fluid may not only prevent the bonding agentfrom entering the tubules, but may also prevent contactbetween the bonding agent and the solid dentinalsurface (Derkson, Pashley & Derkson, 1986; Terklaet al., 1987).
values for AllBond2® were not significantly different.
The values for ART Bond®(P , 0·05) and for Syntac®
(P , 0·05) were significantly higher if the dentine
was perfused with horse serum. For P-Bond
(P , 0·001) the values were significantly higher if
the dentine was perfused with physiologic saline.
According to these results it does not seem to be
appropriate to take a clear decision as to which of
the two perfusing media investigated might be more
suitable to imitate in vivo conditions.
The perfusion of the teeth with a solute that is similar
to dentinal liquid is a means of improving the imitation
of in vivo conditions of vital teeth for the in vitro
experiment. Physiologic saline is often used for this
purpose. However, dentinal liquid is a protein-
containing solution and therefore serum from animals
might be used as an improved substitute for the
perfusion of the dentinal samples through the pulpal
chamber.
Many in vitro studies report on effects on bond
strength and sealing properties of dentine bonding
agents using pressurized instead of dry dentine (Derkson
et al., 1986; Terkla et al., 1987; Mitchem, Terkla &
Gronas, 1988; Tao & Pashley, 1989; Tao, Tagami &
Pashley, 1991; Prati & Pashley, 1992; Davidson, Abdalla
& De Gee, 1993; Paul & Scharer, 1993a; Krejci, Kuster
& Lutz, 1993). The perfusing solutions used in these
studies were sterile phosphate-buffered saline (Derkson
et al., 1986; Terkla et al., 1987; Tao & Pashley, 1989; Tao
et al., 1991; Prati & Pashley, 1992), sterile physiologic
saline (Paul & Scharer, 1993a), horse serum (Krejci
P E R F U S I O N E F F E C T S O N A D H E S I O N O F D E N T I N E B O N D I N G A G E N T S 597
et al., 1993) and water (Mitchem et al., 1988; Davidson
et al., 1993).
Recent in vitro studies (Prati, 1994; Nikaido et al.,
1995) suggest that the use of a protein-containing fluid
for the perfusion of dentine instead of saline results in an
increase of bond strength values due to the interaction of
the bonding agents with these proteins.
Dentinal fluid clearly has a protein content (Stuben
& Spreter von Kreudenstein, 1957) but there is little
available information on the exact composition of this
protein fracture and the qualitative changes as a
function of the condition of the pulp. Recent studies
indicate that albumin and globulins are the main
components of the protein fracture of human dentinal
fluid (Pashley, 1992; Knutsson, Jontell & Bergenholtz,
1994).
In recent studies, 1·5% bovine serum albumin diluted
in water or in phosphate-buffered saline (Pashley,
Stewart & Galloway, 1984), undiluted horse serum
(Krejci et al., 1993), undiluted fetal calf serum (Prati,
1994) or bovine serum diluted 1 : 3 in physiologic
saline (Nikaido et al., 1995) were used to evaluate the
hydraulic conductance of dentine, marginal adaptation
of composite resins and bond strength of dentine
bonding systems.
The aim of the present study was to evaluate the
shear bond strength of AllBond2®, Art Bond®, the
experimental P-Bond and Syntac® on human dentine
which was perfused with horse serum and to compare
these results with values obtained with a standardized
perfusion of human dentine with physiologic saline.
Materials and methods
Seventy-two extracted human third molars were
prepared for a laboratory testing set up, allowing dentine
to be kept under intra-pulpal pressure and
simultaneously allowing the samples to be exposed to
thermal cycling.
The teeth were randomly divided into two groups of
eight teeth four times. One group was perfused with
physiologic saline (NaCl 0·9%)* the other group with
undiluted horse serum† (Fig. 1). Each tooth was
trimmed on one side with a model trimmer‡ parallel to
its longitudinal axis until the pulp chamber appeared.
* Hausmann, St. Gallen, Switzerland.† Nr. 16050098, Life Technologies, Basel, Switzerland.‡ Reco GMT 5350, LeLeux, Oberhausen, Switzerland.
© 1998 Blackwell Science Ltd, Journal of Oral Rehabilitation 25; 596–602
Fig. 1. A, A tooth is glued to an acrylic platelet to form a pressure-
proof chamber. B, The dentinal samples were perfused with either
horse serum or physiologic saline and they received a subsequent
surface treatment by applying a dentine bonding agent and a
composite luting resin (AllBond 2® 1 Porcelite®, ART
Bond® 1 Porcelite®, the experimental P-Bond 1 Porcelite® or
Syntac® 1 Dual Cement®). In addition, AllBond 2® 1 Porcelite®
were applied onto dentine which was not perfused (i.e. ‘no intra-
pulpal pressure). C, The resulting cylinder of cured material on
the dentine. D, Eight samples of the type seen in section C for
each combination of materials were submitted to 1500 thermal
cycles. E: Measurement of shear bond strength.
All pulpal tissue was removed carefully. Two channels
were drilled at opposing sides of the tooth (Fig. 1) with
a diamond bur§. One silicon tube for the inflow and
one for the outflow of the perfusing solution for each
tooth were glued into each channel. Then the open
(i.e. ground) side of each tooth with its silicon tubing
attached was closed by covering it with an acrylic
platelet (1·6 cm 3 3·0 cm) glued first with Superglue¶,
then with Araldite Rapid** to restore a leakfree pulp
chamber. The opposite untouched side of the tooth was
then also trimmed using the same model trimmer (300
grit), so that the remaining dentine thickness was
1·3 6 0·4 mm (5 dentine of medium depth; Paul &
§ Intensiv SA, Nr. 881315, Viganello-Lugano, Switzerland.¶Cyanacrylat, Dentex, Zurich, Switzerland.
** Epoxy resin, Ciba-Geigy, Basel, Switzerland.
598 C . A U G U S T I N et al.
Table 1. Dentine bonding agents used in this study
Acid Primer Unfilled resin
All Bond 2® 10% phosphoric acid A: NTG-GMA in ethanol and acetone B: BPDM in Bis-GMA, UDMA, HEMA
acetone
Syntac® None TEGDMA 1 4% Maleic acid in acetone and water Bis-GMA, TEGDMA
Adhesive: PEGDMA 1 5% glutaraldehyde in water
ART Bond® None A: 1·6% Maleic acid 1 sodium fluoride in water Bis-GMA, TEGDMA,
B: HEMA, HPMA, PMA-Maleic acid water PMA-Maleic acid
P-Bond 10% organic acid Liquid A Liquid B
Scharer, 1993b). This whole unit was finally embedded
into an aluminium frame (Fig. 1) with acrylic*. The
intrapulpal pressure was set at its physiological height of
36 cm water pressure (5 25 mm Hg, Van Hassel, 1971).
Immediately after the preparation of the dentinal
surface four groups of eight teeth were randomly coated
with one of the four following dentine bonding agents
and the respective composite luting resin in each group:
(1) AllBond2® (Lot No. 039025/029235/049065)† plus
Porcelite U® (Lot No. 601282/608124)‡, (2) ART Bond®
(Lot No. 9708 EH851)§ plus Porcelite U® (Lot No.
601282/608124)‡, (3) P-Bond plus Porcelite U®
(experimental dentine bonding agent; Lot No. 601282/
608124)‡ and (4) Syntac® plus Dual Cement® (Lot Nos.
660030/602200/660021 and 617044/617042)¶. The
dentine bonding agents were applied according to
manufacturers instructions and a cylinder of luting
composite was added. Only then were the bonding
agent and the composite cylinder light-cured
simultaneously for 60 s, from above the samples and
parallel to their longitudinal axes, using an Optilux
400**. The light output of the light curing unit used in
this experiment was tested to be 631 mW/cm2 with a
Cure Rite, Model 8000††. The composition of the dentine
bonding agents used in this study is shown in Table 1.
The intra-pulpal pressure was reduced to 19·6 cm
H2O (13·6 mmHg) during the bonding process to
reproduce the effect of the vasoconstrictor in local
anaesthetics. The dentine bonding agents were applied
to the whole dentinal surface whereas the respective
* Fastcure-Rovix, R. Vix AG, Basel, Switzerland.† Bisco, Itasca, IL, U.S.A.‡ Kerr, Romulus, MI, U.S.A.§ Coltene-Whaledent, Mahwah, NJ, U.S.A.
¶ Vivadent, Amherst, NY, U.S.A.** Demetron, Danbury, CT, U.S.A.†† EFOS INC., Mississauga, Ontario, Canada.
© 1998 Blackwell Science Ltd, Journal of Oral Rehabilitation 25; 596–602
luting composite was applied into a piece of silicon tube
3 mm in diameter and 3 mm in height to generate
standardized cylinders of luting resin with a defined
area of bonded surface for the shear bond testing. After
1500 thermal cycles in water baths of 5°C and 55°C
(dwell time 30 s, time between water baths 30 s) shear
bond strength values (SBS) were measured with a
universal testing machine (crosshead speed: 0·5 mm/
min) using the ‘wire loop technique’ (Sorensen & Dixit,
1991; Prati, Pashley & Montanari, 1991). A Ni-Cr-
wire (Rexillium III)‡‡was adjusted at the base of the
composite cylinder near the dentinal surface. Then a
pulling force from the Instron was applied until a
fracture occurred. The fractured surfaces were
observed with a binocular microscope at 253
magnification.§§ The resulting values were statistically
analysed with the Students t-test (AllBond2®, ART
Bond® and P-Bond) and with the Mann–Whitney Test
(Syntac®; no similar variances between groups as shown
in the box plot in Fig. 2).
Results
The single values of the resulting shear bond strengths
are shown in Table 2 and they are graphically displayed
in Fig. 2. Vertical bars in the tables indicate cohesive
fractures in dentine. All other samples fractured
adhesively.
No statistically significant differences could be
detected between the values of AllBond2® 1 Porcelite
U®/Horse Serum and AllBond2® 1 Porcelite U®/NaCl.
A clearly significant difference (Fig. 2) towards increased
bond strength values was seen when AllBond2® 1
Porcelite U® was applied on dentine under zero pulpal
‡‡ Jeneric Industries Inc., Wallingford, CT, U.S.A.§§Wild, Hersbrugg, Switzerland.
P E R F U S I O N E F F E C T S O N A D H E S I O N O F D E N T I N E B O N D I N G A G E N T S 599
Fig. 2. Box plot of the resulting shear bond strength values on dentine which was either perfused with horse serum (5 Horse Ser.) or
physiologic saline (5 NaCl). Contraction: this line indicates the maximum contraction forces upon polymerization of resin composites
(Feilzer, De Gee & Davidson, 1989). GIC: this line indicates the reported bond strength of glass–ionomers on dentine (Prati & Pashley,
1992; Craig, 1993). The insert shows the distribution of values in a box plot drawing: 80% of the values lie between 1 and 2, 50% of
values between 3 and 4. Median: 50% of values lie above, the other 50% below this mark. (AllB 2 5 AllBond 2®; no IP 5 the adhesives
were bonded to dentine without the application of an intra-pulpal pressure).
Table 2. Shear bond strength values of AllBond 2, ART Bond, the experimental P-Bond, and Syntac after 1500 thermal cycles. Dentin
was perfused with either horse serum or physiologic saline.
All Bond 2® ART Bond® P-Bond Syntac®
Horse serum NaCl No IP Horse serum NaCl Horse serum NaCl Horse serum NaCl
0·38 4·88 11·16 11·30 10·68 11·76 13·52 7·42 6·78
3·78 8·76 11·22 13·18 11·46 11·62 17·64 10·26 8·38
5·54 7·20 11·50 12·98 12·72 13·74 18·02 13·30 7·12
4·94 0 9·78 13·94 11·54 12·96 18·22 11·76 9·34
4·22 7·46 9·98 13·80 12·10 14·04 18·24 16·96 7·20
3·80 3·62 10·44 13·42 10·22 13·66 18·52 13·94 8·56
7·28 0·98 10·64 13·08 11·00 13·74 19·60 20·02 7·64
8·64 5·34 10·96 17·86 13·12 13·04 21·72 8·34 7·92
Mean 6 SD
4·82 6 2·49 4·78 6 3·11 10·71 6 0·61 13·70 6 1·87 11·61 6 1·00 13·07 6 0·93 18·19 6 2·30 12·75 6 4·27 7·87 6 0·86
No IP, no intra-pulpal pressure was applied. Application of the dentin-bonding agent onto a wet dentinal surface with tubuli filled with
fluid but with zero pressure. Cohesive fractures in dentin are specified with a vertical bar.
© 1998 Blackwell Science Ltd, Journal of Oral Rehabilitation 25; 596–602
600 C . A U G U S T I N et al.
pressure. For ART Bond® 1 Porcelite U® (P , 0·05) and
for Syntac® 1 Dual Cement® (P , 0·05), significantly
higher bond strength values were obtained if horse
serum was used as the perfusing liquid. P-
Bond® 1 Porcelite U® demonstrated significantly higher
bond strength values (P , 0·001) if the dentine was
perfused with physiologic saline.
Discussion
In the discussion of the composition of dentinal fluid
and the options for finding a true substitute for in vitro
studies, the origin of this fluid might be of help. Dentinal
fluid must finally be an ‘ultra filtrate’ of pulpal
extracellular fluid (5 pulp interstitial fluid) which is an
ultra filtrate of blood plasma. The different barriers
plasma has to encounter until it seeps as ‘dentinal fluid’
from cut tubules are the capillary endothelium, the
odontoblast layer and structures at the inner surface of
tubular walls (peritubular hydroxyapatite, intratubular
collagenous fibrils). The question arises as to what
supposedly happens with the original blood plasma
after transition through these barriers.
Blood consists of cells suspended in a fluid called
plasma. Human plasma is composed of an aqueous
solution of proteins (70 g/L), electrolytes and small
organic molecules. The composition of this protein
fraction, its interaction with the barriers mentioned and
its role in generating osmotic pressure are of primary
interest in the context of this study. Albumin (MW
69 000) makes up over 40 g/L of the total protein
content followed by several globulin fractions and finally
a very small amount of fibrinogen (µ 3 g/L; MW
340 000) (West, 1985; Schmidt & Thews, 1990). Besides
having carrier functions these proteins (and among
them particularly albumin) are responsible for the
colloid osmotic pressure of plasma.
The colloid osmotic pressure of human plasma is
reported to be 25 mmHg (generated by the protein
content of 70 g/L) whereas the colloid osmotic pressure
of the interstitial fluid is said to be 5 mmHg (according
to a protein concentration of µ 20 g/L) (Schmidt &
Thews, 1990). The distribution of extracellular fluid
across (pulpal) capillaries and venules is regulated by a
set of inter-related forces which are called the Starling
factors (Guyton, 1991). Roughly, hydrostatic pressures
and colloid osmotic pressures are the opposing forces
in this regulative system and their inter-relationship
with regard to the pulpodentine complex has been
© 1998 Blackwell Science Ltd, Journal of Oral Rehabilitation 25; 596–602
nicely reviewed by Pashley (1992). Due to a higher
than atmospheric interstitial tissue pressure (Heyeraas,
1985), fluid will flow from the pulp in a peripheral
direction when dentine is uncovered and tubules are
cut by, for instance, cavity preparation (Pashley, 1985).
Earlier results of Pashley et al. (1981) suggest that the
capillary endothelium is the major diffusion barrier for
plasma proteins whereas dentine itself (i.e. surface
structures inside tubular walls) does not seem to reduce
the outwardly directed flux of the plasma proteins
(Maita et al., 1991).
The reported data imply that dentinal fluid contains
protein and that albumin and globulins (and to a certain
degree fibrinogen) should be the principal components.
Knutsson et al. (1994) measured the amount of albumin
and IgG (immune globulins) in effluents from cavities
prepared in human teeth in vivo and came to the
conclusion that their quantity was larger in samples
from pulp exposures but that the ratio between albumin
and IgG was always similar. Unfortunately the ratio of
albumin and IgG concentrations between pulp
interstitial fluid and effluents of dentinal fluid are not
given numerically in this study and therefore, they can
only be estimated from the diagrams provided to be in
the range of 7 : 1 to 5 : 1. This would actually be in
correlation with the results of Maita et al. (1991) who
reported that the protein concentration of dentinal fluid
measured in vivo in dogs was about one-fifth that of
plasma under spontaneous conditions.
The horse serum which we used in our study
contained albumin and globulins in concentrations of
32 g/L and 38 g/L, respectively. This indicates a similar
total protein content but a different composition
compared to human serum as reported above. It is
unknown at this time if this different composition could
have a significant influence. However, according to the
above data a dilution of such horse serum in physiologic
saline at a ratio of 1 : 5 to 1 : 7 would seem to be
appropriate according to the above data to simulate
dentinal fluid for in vitro studies.
Despite these considerations we started with
undiluted horse serum for this first study because of
the variety of set-ups reported in the literature which
claim an increasing effect on bond strength values if a
protein containing solute instead of physiologic saline
was used (1·5% bovine serum albumin (Pashley et al.,
1984); undiluted horse serum (Krejci et al., 1993; Prati,
1994); diluted horse serum (1 : 3) (Nikaido et al., 1995).
Without giving a more detailed description Prati (1994)
P E R F U S I O N E F F E C T S O N A D H E S I O N O F D E N T I N E B O N D I N G A G E N T S 601
reported almost ‘doubled’ bond strength values for
Gluma or Scotchbond 2® under such conditions
compared to results obtained on dentine perfused with
phosphate-buffered saline. Nikaido et al. (1995)
reported significant differences between bond strength
values depending on the nature of the perfusing dentinal
fluid for Scotchbond MP® and for Clearfil Liner Bond
II®. These results were on a significance level (P , 0·05)
comparable with our results found for ART Bond® and
Syntac®. However, in our study P-Bond showed a
reverse relation (higher bond strengths if the dentine
was perfused with saline) whereas AllBond2® failed to
demonstrate any significant difference.
Causing a precipitation reaction between horse serum
and the freshly mixed primers of AllBond2® or ART
Bond® in a glass vial is a simple experiment and easy
to reproduce. The higher the protein content the more
obvious is the reaction. Very probably this reaction is
caused by mechanisms which are commonly known
for proteins: acids, alkaline agents and organic solvents
are able to coagulate globular protein resulting in a
destruction of the tertiary structure of the protein. It
thereby is irreversibly ‘denatured’ and its biologic
activity is lost.
A coagulation/precipitation of organic and inorganic
constituents seems to be responsible for the altered
permeability of dentine by blocking the dentinal tubules
(Pashley et al., 1984; Tagami, Nakajima & Hosoda, 1994).
A protein coagulating effect was reported (Tagami,
Hosoda & Imai, 1987; Watanabe et al., 1991) for primers
such as HEMA (hydroxyethylmethacrylate) and 5-
NMSA (N-methacryloyl 5-aminosalicylic acid). This
would imply that bond-strength values should be
significantly increased for HEMA-containing DBAs if
dentine is perfused with a protein containing liquid
instead of protein-free saline. However, the present
results show only a significance level of P , 0·05 for
the HEMA-containing ART Bond® despite the use of
even undiluted horse serum. On the other hand,
AllBond2® values were not significantly different
despite the very distinct coagulation reaction in the
glass vial and clearly higher values were only obtained
for this system if the pulpal pressure was zero. Therefore,
according to our results the effect of perfusing dentine
with protein containing liquids and the way this effect
relates to adhesive mechanisms on dentine still remains
a question which calls for further research.
© 1998 Blackwell Science Ltd, Journal of Oral Rehabilitation 25; 596–602
Acknowledgments
The authors thank Dr G. Menghini for his help with
the statistical analysis of the results in this study. We
would furthermore like to thank Bisco, Itasca, U.S.A.,
Coltene-Whaledent, Mahwah, U.S.A., Kerr, Romulus,
U.S.A. and Vivadent, U.S.A. for the donation of the
testing materials.
References
BUONOCUORE, M.G. (1955) A simple method of increasing the
adhesion of acrylic filling materials to enamel surfaces. Journal
of Dental Research, 35, 849.
CRAIG, R.G. (1993) Glass ionomer cement, In: Restorative Dental
Materials, p. 199. Mosby, St. Louis.
DAVIDSON, C.L., ABDALLA, A.I. & DE GEE, A.J. (1993) An
investigation into the quality of dentine bonding systems for
accomplishing a durable bond. Journal of Oral Rehabilitation,
20, 291.
DERKSON, G.D., PASHLEY, D.H. & DERKSON, M.E. (1986) Microleakage
measurement of selected restorative materials: a new in-vitro
method. Journal of Prosthetic Dentistry, 56, 435.
FEILZER, A.J., DE GEE, A.J. & DAVIDSON, C.L. (1989) Increased wall-
to-wall curing contraction in thin bonded resin layers. Journal
of Dental Research, 68, 48.
GUYTON, A.C. (1991) Textbook of Medical Physiology, 8th edn. WB
Saunders, Philadelphia, PA.
HEYERAAS, K.J. (1985) Pulpal, microvascular, and tissue pressure.
Journal of Dental Research, 64, 585.
JENDRESEN, M.D., ALLEN, E.P., BAYNE, S.C., DONOVAN, T.E., HANSSON,
T.L., KLOOSTER, J. & KOIS, J.C. (1995) Annual review of selected
dental literature: Report of the Committee on Scientific
Investigation of the American Academy of Restorative Dentistry.
Journal of Prosthetic Dentistry, 74, 60.
KNUTSSON, G., JONTELL, M. & BERGENHOLTZ, G. (1994) Determination
of plasma proteins in dentinal fluid from cavities prepared in
healthy young human teeth. Archives of Oral Biology, 39, 185.
KREJCI, I., KUSTER, M. & LUTZ, F. (1993) Influence of dentinal fluid
and stress on marginal adaptation of resin composites. Journal
of Dental Research, 71, 490– 494.
MAITA, E., SIMPSON, M.D., TAO, L. & PASHLEY, D.H. (1991) Fluid
and protein flux across the pulpodentine complex of the dog
in vivo. Archives of Oral Biology, 36, 103.
MITCHEM, J.C., TERKLA, L.G. & GRONAS, D.G. (1988) Bonding of
resin dentine adhesives under simulated physiological
conditions. Dental Materials, 4, 351.
NIKAIDO, T., BURROW, M.F., TAGAMI, J. & TAKATSU, T. (1995) Effect
of pulpal pressure on adhesion of resin composite to dentine:
bovine serum versus saline. Quintessence International, 26, 221.
PASHLEY, D.H. (1985) Dentine-predentine complex and its
permeability: physiologic overview. Journal of Dental Research,
64, 613.
PASHLEY, D.H. (1992) Mechanistic analysis of fluid distribution
across the pulpodentine complex. Journal of Endodontics, 18, 72.
PASHLEY, D.H., NELSON, R., WILLIAMS, E.C. & KEPLER, E.E. (1981)
602 C . A U G U S T I N et al.
Use of dentine-fluid protein concentrations to measure pulp
capillary reflection coefficients in dogs. Archives of Oral Biology,
26, 703.
PASHLEY, D.H., STEWART, F.P. & GALLOWAY, S.E. (1984) Effects of
air-drying in vitro on human dentine permeability. Archives of
Oral Biology, 29, 379.
PAUL, S.J. & SCHARER, P. (1993a) Intrapulpal pressure and thermal
cycling: Effect on shear bond strength of eleven modern dentine
bonding agents. Journal of Esthetic Dentistry, 5, 179.
PAUL, S.J. & SCHARER, P. (1993b) Factors in dentine bonding: Part
I. A review of the morphology and physiology of human
dentine. Journal of Esthetic Dentistry, 5, 5.
PRATI, C. (1994) What is the clinical relevance of in vitro dentine
permeability tests? Journal of Dentistry, 22, 83.
PRATI, C. & PASHLEY, D.H. (1992) Dentine wetness, permeability
and thickness and bond strength of adhesive systems. American
Journal of Dentistry, 5, 33.
PRATI, C., PASHLEY, D.H. & MONTANARI, G. (1991) Hydrostatic
intrapulpal pressure and bond strength of dentine bonding
systems. Dental Materials, 7, 54– 58.
SCHMIDT, R.F. & THEWS, G. (eds) (1990) Physiologie des Menschen.
24th edn, pp. 422, 528. Springer Verlag, Berlin.
SORENSEN, J.A. & DIXIT, N.V. (1991) In vitro shear bond strength
of dentine adhesives. International Journal of Prosthodontics, 4, 117.
STUBEN, J. & SPRETER V. KREUDENSTEIN, T. (1957) Dentinstoffwechsel-
studien. VI. Mitteilung: Papierelektrophoretische Unter-
suchungen uber die Zusammen- setzung der Dentinliquor-
proteine. Deutsche Zahnarztliche Zeitschrift, 12, 500.
© 1998 Blackwell Science Ltd, Journal of Oral Rehabilitation 25; 596–602
TAGAMI, J., HOSODA, H. & IMAI, Y. (1987) Evaluation of a new
adhesive liner as an adhesive promoter and a desensitizer of
hypersensitive dentine. Dental Materials, 6, 201.
TAGAMI, J., NAKAJIMA, M. & HOSODA, H. (1994) Influence of dentine
primers on the flow of bovine serum through dentine. Archives
of Oral Biology, 39(Suppl.), 146
TAO, L. & PASHLEY, D.H. (1989) Dentine perfusion effects on the
shear bond strengths of bonding agents to dentine. Dental
Materials, 5, 181.
TAO, L., TAGAMI, J. & PASHLEY, D.H. (1991) Pulpal pressure and
bond strengths of SuperBond and Gluma. American Journal of
Dentistry, 4, 73.
TERKLA, L.G., BROWN, A.C., HAINISCH, A.P. & MITCHEM, J.C. (1987)
Testing sealing properties of restorative materials against moist
dentine. Journal of Dental Research, 66, 1758.
VAN HASSEL, H.J. (1971) Physiology of the human dental pulp.
Oral Surgery, 32, 126.
WATANABE, T., SANO, M., ITOH, K. & WAKUMOTO, S. (1991) The
effects of primers on the sensitivity of dentine. Dental Materials,
7, 148.
WEST, J.B.(ed.) (1985) Best and Taylor’s Physiological Basis of Medical
Practice. 11th edn, pp. 126, 334, 438. Williams & Wilkins,
Baltimore.
Correspondence: Dr S. J. Paul, University of Zurich, Center for
Dental and Oral Medicine, Department of Fixed and Removable
Prosthodontics and Dental Material Sciences, Plattenstr. 11, 8028
Zurich, Switzerland. E-mail: [email protected]