PANIC NMR 2018
MARIA VICTORIA SILVA ELIPE*, NEIL DONOVAN, ROBERT KRULL, DONALD POOKE, KIM COLSON
PERFORMANCE OF NEW 400 MHZ HTS POWER-DRIVEN MAGNET NMR TECHNOLOGY ON CINACALCET HCL (TYPICAL API IN PHARMA)
2
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
• Historical highlights of magnet development
• Symbiotic scientific collaboration
• Scheme of the instrument
• Cinacalcet HCl
• NMR data (1D 1H, 2D COSY, 2D NOESY, 2D 1H-13C HSQC)
• HTS NMR system installed at Amgen chemistry laboratory
• Pros and cons of HTS magnet system
• Acknowledgements
• Thank you
OUTLINE
3
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
HISTORICAL HIGHLIGHTS OF MAGNET DEVELOPMENT
0
200
400
600
800
1000
1200
1951
1955
1958
1965
1970
1973
1970
1980
2010M
ag
neti
c f
ield
fo
r P
roto
n (
MH
z)
Year
Increase on MHz over the years
Cryogenic
magnets
Electromagnets
Permanent
magnets
HTS magnets
superconducting at ~18 K
0
50
100
150
200
250
300
350
400
450
1950s-60s 2013 2017
Ma
gn
eti
c F
ield
fo
r P
roto
n (
MH
z)
Year
Electromagnets
4
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
SYMBIOTIC SCIENTIFIC COLLABORATION
Amgen
Bruker BioSpin
HTS-110
Maria is happy!
5
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
SCHEMATICS OF THE INSTRUMENT
console
computer
chiller He compressor
power supply
magnet monitor
400 MHz
HTS magnet
sample
Silva Elipe, MV, Donovan, N, Krull, R, Pooke, D, Colson, K, 2018, submitted
6
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
• API of commercial oral formulated tablets of Sensipar®/Mimpara®
• Approved in 2004 by FDA as Sensipar® and by EMEA as Mimpara®
• Therapeutic indications:
– Secondary hyperparathyroidism (sHPT) for patients with chronic kidney disease
(CKD) on dialysis
– Primary hyperparathyroidism and parathyroid carcinoma
CINACALCET HCL
7
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
COMPARING 1D 1H NMR ALIPHATIC REGION OF HTS NMR SYSTEM WITH CONVENTIONAL NMR INSTRUMENT
Silva Elipe, MV, Donovan, N, Krull, R, Pooke, D, Colson, K, 2018, submitted
Magnet up time: days Magnet up time: years
8
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
COMPARING 1D 1H NMR AROMATIC REGION OF HTS NMR SYSTEM WITH CONVENTIONAL NMR INSTRUMENT
Silva Elipe, MV, Donovan, N, Krull, R, Pooke, D, Colson, K, 2018, submitted
9
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
COMPARING 2D COSY OF HTS NMR SYSTEM WITH CONVENTIONAL NMR INSTRUMENT
Silva Elipe, MV, Donovan, N, Krull, R, Pooke, D, Colson, K, 2018, submitted
10
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
COMPARING 2D NOESY OF HTS NMR SYSTEM WITH CONVENTIONAL NMR INSTRUMENT
Silva Elipe, MV, Donovan, N, Krull, R, Pooke, D, Colson, K, 2018, submitted
11
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
COMPARING 2D 1H-13C HSQC OF HTS NMR SYSTEM WITH CONVENTIONAL NMR INSTRUMENT
Silva Elipe, MV, Donovan, N, Krull, R, Pooke, D, Colson, K, 2018, submitted
12
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
COMPARING 1D 1H NMR OF HTS NMR SYSTEM (BBFO/BOSS3 AND QNP/BOSS1) WITH CONVENTIONAL NMR INSTRUMENT
Silva Elipe, MV, Donovan, N, Krull, R, Pooke, D, Colson, K, 2018, submitted
13
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
COMPARING 1D 1H NMR ALIPHATIC REGION HTS NMR SYSTEM (QNP/BOSS1) WITH CONVENTIONAL NMR INSTRUMENT
Silva Elipe, MV, Donovan, N, Krull, R, Pooke, D, Colson, K, 2018, submitted
14
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
HTS NMR SYSTEM INSTALLED AT AMGEN CHEMISTRY LAB
15
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
Pros:
• No need to fill with liquid cryogens (no liquid
N2 and He needed) and no O2 sensors
• Easy access to chemist (in the same lab with
no restricted access)
• Small footprint magnet (can site in hood,
easy sample insertion, small stray field)
• Quick installation (couple of days to field and
acquiring data)
• Ability to shutdown magnet during times of
non-use (easy to do by the user, no need for
engineer, magnet can be easy moved to
other location if needed, no need for extra
money)
PROS AND CONS OF HTS MAGNET SYSTEM Cons:
• Sensitive to power
outages (UPS/generator
needed)
• Compressor & power
supply must be nearby
(real-estate)
• Water cooling or air-
cooling needed
• Noisy equipment (He
compressor)
16
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
• Amgen Sponsors: Janet Cheetham & David Semin
• Cinacalcet HCl sample from Amgen: Minhui Ma & Robert
Munger
• Bruker BioSpin: Clemens Anklin
• Others: many people helped through the process
ACKNOWLEDGEMENTS
17
Provided March 6, 2018, as part of an oral presentation and is qualified by
such, contains forward-looking statements, actual results may vary
materially; Amgen disclaims any duty to update.
THANK YOU
Elvis is alive and has been working
with a cryogenic magnet for decades
No cryogens????
But now …
Oops!!!