classification of medical devices - tuev-sued.de. non invasive devices annex vii, chapter iii, rule...
TRANSCRIPT
Classification of medical devices medical devices MDR Annex VII
Norbert Stuiber
TÜV SÜD Product Service GmbH Slide 1Classification of medical devices 6-Jul-16
Overview
f f
1 Introduction & classification process
Note: The EU’s new regulations have not been finalised, and the information contained herein is subject to change.
1 Introduction & classification process
2 Non Invasive Devices
3 Invasive Devices
4 Active Devices
5 Special Rules
TÜV SÜD Product Service GmbH Slide 2Classification of medical devices 6-Jul-16
p
1 Introduction and classification process
MDD classificationMDD classification Rules Annex IX of the MDD
MEDDEV Guidelines e.g.2. 4/1, 2.1/6
Borderline ManualBorderline Manual
MDD amendments e.g. COMMISSION DIRECTIVE 2003/12/EC
TÜV SÜD Slide 3Classification of medical devices 6-Jul-16
1 Introduction and classification process
Changes in the classification rulesChanges in the classification rules Changes in the classification rules of medical devices
might lead to reclassification of certain medical devices.
Manufacturers shall check, if the respective classification rule is still applicable.
In case of disputes regarding the classification ofmedical devices, different scenarios may occur
TÜV SÜD Product Service GmbH Slide 4Classification of medical devices 6-Jul-16
1. Introduction & Classification Process
MDDManufacturer
ClassificationClassification
Notified Body
Not acceptedAccepted Dispute
Competent Authority to which the
TÜV SÜD Slide 5Classification of medical devices 6-Jul-16
notified body is subjected
1. Introduction & Classification Process Art. 41
Manufacturer (registered place of business) and notified body in the same Member State
Manufacturer (registered place of business) and notified body in different Member States
Manufacturer
Classification
NB
AcceptedDispute Accepted DisputeAcceptedDispute Accepted Dispute
Competent Competent Consultation with competent p
authority of the Member State of the manufacturer
(address of
Notification to MDCG
pauthority of the
Member State of the manufacturer
(address of
competent authority of the Member State
which designated the
TÜV SÜD Slide 6Classification of medical devices 6-Jul-16
(business)
(business) designated the
notified body
2. Non invasive Devices Annex VII, Chapter III, Rule 1-4
R l 2Rule 2Channeling or
storing for eventual infusion eventual infusion
administration
I
For use with blood, other body fluids, cellsorgans tissues
May be connected to an active medical device in
Class IIa or higherBlood bags
organs, tissues Class IIa or higher
IIa IIa IIb
TÜV SÜD Product Service GmbH Slide 716-07-06
IIa IIa IIb
2. Non invasive Devices Annex VII, Chapter III, Rule 1-4
R l 3Rule 3Modifying biological or chemical
composition of blood, body liquids other liquids intended forliquids, other liquids intended forimplantation or administration
IIb
Only filtration , centrifugation or
exchange of gas or heat
Substances intended to be used in vitro in direct contact with
human cells/tissues/organs or with human embryos before exchange of gas or heat
IIa
with human embryos before implantation/administration
III
TÜV SÜD Product Service GmbH Slide 816-07-06
IIa III
2. Non invasive Devices Annex VII, Chapter III, Rule 1-4
Rule 4In contact with injured skin or
mucous membrane
(mechanical barrier, ( ,compression, absorption)
II
Intended for injuries to skin which have breached the
dermis or mucous b d h l l b
Intended to manage the micro-environment of
injured skin or mucous membrane and heal only by secondary intent
jmembrane
TÜV SÜD Product Service GmbH Slide 916-07-06
IIb IIa
2. Non invasive Devices Annex VII, Chapter III, Rule 1-4
ConclusionsConclusions Different classification rules, without changing the class
Coverage of IVF / Tissue Engineering as Class III plus e.g. antibodies of human/synthetic origin
TÜV SÜD Product Service GmbH Slide 10Classification of medical devices 6-Jul-16
3. Invasive Devices Annex VII, Chapter III, Rule 5-8
Rule 6 / 7 (surgically invasive devices for transient/ short term use):
Devices intended specifically for use in direct contact with the heart or central circulatory system are class III, regardless of duration of application
TÜV SÜD Product Service GmbH Slide 11Classification of medical devices 6-Jul-16
3. Invasive Devices Annex VII, Chapter III, Rule 5-8
Rule 6S i ll i i Surgically invasive – Transient use
IIa
Control/diagnose/monitor/ correct a defect of
Used in direct contact with central nervous Supply energy Reusable
Biological effect
Intended to administer r/ correct a defect of
heart or central circulatory system
through direct contact
with central nervous system or with heart
or central circulatory system
Supply energy /ionizing radiation
Reusable surgical
instrument
effect –mainly or wholled
absorbed
administer medicinal products
in a potential hazardous manner
TÜV SÜD Product Service GmbH Slide 1216-07-06
III IIII IIb IIb IIb
3. Invasive Devices Annex VII, Chapter III, Rule 5-8
Rule 7Rule 7Surgically invasive – Short-term use
IIa
Control/diagnose/ monitor / correct a
defect of heart or centralcirculatory system
Used in direct contactwith central nervoussystem or with heart
or central circulatory
Supplyenergy/ionizin
g radiation
Undergo chemicalchange in body or
adminstermedicines (NOT in
Biological effect –
mainly/wholly absorbed y y
through direct contacty
system
III
g(teeth) y absorbed
IIIIIb IIb III
TÜV SÜD Product Service GmbH Slide 1316-07-06
III IIIIIb IIb III
3. Invasive Devices Annex VII, Chapter III, Rule 5-8
Rule 8Rule 8Surgically invasive long-term use and
implantable devices
IIb
Total/Partial
Breast implants
and To be placed in teeth
Undergo chemical change in
body (NOT in
Used in direct contact with
heart or central
Biological effect or wholly or
Active implantable medical devices
and their
Total/Partial joint/spinal disc
replacements and implantable devices in contact with the surgical
meshes
in teeth y (teeth) or
administer medicines
circulatory / nervous system
ymainly
absorbed
and their accessories
in contact with the spinal column(NOT ancillary components )
TÜV SÜD Product Service GmbH Slide 1416-07-06
IIa IIIIII III III III III
3. Invasive Devices Annex VII, Chapter III, Rule 5-8
Conclusions
Some devices will have to be reclassified e.g.:
Devices intended specifically for use in direct contact with the heart or central icirculatory system are class III (e.g. icirculatory system are class III (e.g. catheters which were not falling in MDD rule 6/7 )
Active implantable devices and their accessories are class III
Total, partial joints, spinal disc replacements and implantable devices in
TÜV SÜD Product Service GmbH Slide 15Classification of medical devices 6-Jul-16
contact with the spinal column are class III
4. Active Devices Annex VII, Chapter III, Rule 9-12
Rule 9Active therapeutic
devices intended toadminster or
exchange energyexchange energy
IIa
Intended to emit Intended to
control/monitor or toinfluence directly the
performance of a class
Administer orexchange energy in
potentially
Intended to emit ionizing radiation for
therapeutic purposes, including devices which control or monitor such
Active devices intended for controlling,
monitoring, or directly influencing the performance of a class
IIb active therapeuticdevice
potentiallyhazardous way
control or monitor such devices, or which
directly influence their performance
influencing the performance of active
implantable devices
TÜV SÜD Product Service GmbH Slide 1616-07-06
IIb IIb IIb III
4. Active Devices Annex VII, Chapter III, Rule 9-12
Rule 10Active devices for diagnosis and
monitoring. If intended to supply energy absorbed by the human body, to image
in vivo distribution of radio pharmaceuticals, or for direct diagnosis
or monitoring of vital physiological or monitoring of vital physiological processes
IIa
Specifically intended for monitoring vital physiological parameters where variations could result in immediate danger of the patient or to
Devices intended to illuminate the
patient's body (in g pdiagnose clinical situations where the
patient is in immediate danger
p y (the visible spectrum)
TÜV SÜD Product Service GmbH Slide 1716-07-06
I IIb
4. Active Devices Annex VII, Chapter III, Rule 9-12
Rule 10aSoftwareSoftware
II
Software intended to Intended to provide information which is used to take decisions monitor physiological
processes
IIa
which is used to take decisions with diagnosis or therapeutic
purposes
IIaIIa
If the decision may directly or If the decision may directly or
IIa
Software intended to monitor If the decision may directly or indirectly cause death or
irreversible deterioration of the state of health
If the decision may directly or indirectly cause deterioration of the state of health or a surgical
intervention
physiological parameters, where the nature of variations
could result in immediate danger to the patient
TÜV SÜD Product Service GmbH Slide 1816-07-06
IIIIIb
g p
IIb
5. Special rules Annex VII, Chapter III, Rule 13-23
Rule 17Devices manufactured
utilizing tissues or cells ofutilizing tissues or cells ofhuman or animal origin ortheir derivatives which arenot viable or rendered not
Rule 19Devices incorporating or
consisting of nanomaterialsviable
III
Devices manufactured utilizing ti ll f i l i i High/medium L t ti l f Negligible potential tissues or cells of animal origin or their derivatives, which are non-
viable or rendered non-viable that are intended to come into contact
ith i t t ki l
High/medium potential for internal
exposure
Low potential for internal exposure
Negligible potential for internal exposure
with intact skin only
To be classified according to other
III IIb IIa
TÜV SÜD Product Service GmbH Slide 1916-07-06
To be classified according to other applicable rules
5. Special rules Annex VII, Chapter III, Rule 13-23
Rule 21Devices that are composed of substances intended t b i t d d i t th b d i b d ifi to be introduced into the body via a body orifice, or applied on skin and that are absorbed by or locally
dispersed in the human body
The device, or its products of metabolism,
are systemically
The device achieves its intended purpose in the
stomach or lower
Device applied on skinor
in the nasal or oral cavity as far are systemically absorbed by the human body in order to achieve
the intended purpose
gastrointestinal tract and the device or its products of
metabolism are systemically absorbed by the human body
in the nasal or oral cavity as far as pharynx and achieving the
intended purpose on these cavities
All other cases
III IIa IIbIII
TÜV SÜD Product Service GmbH Slide 2016-07-06
5. Special rules Annex VII, Chapter III, Rule 13-23
Rule 22i i d i ith t t
Rule 23Active therapeutic devices with an
integrated or incorporated invasive devices with respect to body orifices, other than
surgically invasive devices, which are intended to administer
di i l d t b i h l ti
integrated or incorporated diagnostic function, which
significantly determinates the patient management by the device such as closed loop medicinal products by inhalation device, such as closed loop
systems or automated external defibrillators.
IIa III
Essential impact on the efficacy and safety of the administered medicinal
product , including the devices intended p gto treat life threatening conditions
IIb
TÜV SÜD Product Service GmbH Slide 2116-07-06
IIb
5. Special rules: Annex VII, Chapter III, Rule 13-23
C l iConclusions Nanomaterials shall be taken into consideration for the classification New wording “human origin” in addition to the already existing “animal origin” Open points regarding some rules (e.g. #10a, 21)
Early preparation is key. We urge manufacturers to check if the devices have to be reclassified according to the new medical device regulation.
TÜV SÜD Product Service GmbH Slide 22Classification of medical devices 6-Jul-16
Stay informed about the expected changes and prepare for transition
Prepare for the transitionPrepare for the transition We are closely following the regulation
developments and will provide updates to developments, and will provide updates to medical devices manufacturers to help them stay informed about the transition.
For the latest information, visit the following webpages:
Medical Device Regulation: www.tuv-sud.com/mdr
In Vitro Diagnostic Device Regulation:www.tuv-sud.com/ivdr
TÜV SÜD Product Service GmbH Slide 23Classification of medical devices 6-Jul-16
Contact us today for one-stop quality, safety and sustainability solutions. www.tuv-sud.com/medicaldevice
di ld i @t [email protected]
TÜV SÜD Product Service GmbH Slide 246-Jul-16 Classification of medical devices