September 26, 2018 TCEQ Operator Training 1 PAMS AutoGC Operator Training Heath Hinojosa [email protected] Orsat, LLC www.orsat.com

Upload: others

Post on 20-May-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

September 26, 2018TCEQ Operator Training 1

PAMS

AutoGC

Operator Training

Heath [email protected]

Orsat, LLC

www.orsat.com

Page 2: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

AUTOGC SYSTEM OVERVIEW

AutoGC Overview of VOC Analysis

Description of Equipment Functionality

Overview

Air System

Thermal Desorber Method

Gas Chromatographic Method

Chromatographic Data System

Analytical Method

Collection using Sequences

Data Processing and File management

OPERATOR ACTIVITIES

Building Sequences

EZ Sequence sequence generator

Initiation of sequences

Modifying sequences

Calibration

Merlin MicroScience Dilution System

Site Spreadsheet

Optimizing Methods

Updating retention times

Using Overlay Chromatograms

2018 TCEQ Operator Training

September 26, 2018TCEQ Operator Training 2

Page 3: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

PerkinElmer AutoGC System

September 26, 2018TCEQ Operator Training 3

Totalchrom

Data System

NCI 902 Inte

rface

Turbomatrix

Thermal Desorber

Clarus Gas

Chromatograph

Page 4: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

PerkinElmer AutoGC System

September 26, 2018TCEQ Operator Training 4

DVI

TD

DVINafion

Ballast

Tank to backup air cylinder

Zero Air

Purifier

Flow meter

In Line Regulator

Ballast

Tank

TOC Gas

Generator

Compressor

Auto Drain

GC

Diluter

AutoGC Air SystemHC Free Zero Air

for FIDs

HC Free & Dry (-50º DP)

Zero Air for Pelier Cooler

and Nafion Drier

Page 5: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Temperatures

Transfer line 200 ºC

Valve 150 ºC

Tube 0 ºC

Trap – low -30 ºC

Trap – high 325 ºC

Rate 40 ºC/min

Times

Cycle 60 min.

Sampling 40 min.

Trap Hold 6 min.

Desorb 1 min.

Purge 1 min.

Options

Mode Online

Outlet split checked

Inlet split unchecked

Heated purge unchecked

Cycles 99

Pneumatics

Outlet split 1.0 – 3.0 mL/min.

Desorb vent ~ 10.0 mL/min.

Column 40.5 – 43.5 psig.

Pump 10-15 mL/min.

Inlet split N/A

Other

Nafion flow 0.15 – 0.30 Lpm.

AutoGC Methods - Turbomatrix

September 26, 2018TCEQ Operator Training 5

Page 6: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Oven Temperature Profile

Initial Temperature 45 ºC

Initial time 15 min.

Ramp rate 1 5 ºC /min.

Temperature 1 170 ºC

Hold time 1 0 min.

Ramp rate 2 15 ºC /min.

Temperature 2 200 ºC

Final hold time 6 min.

Configuration

Valve 2 initial Off

Events

Valve 2 on 0.1 min.

Valve 2 off 8.0 – 13.0 min.

FID Parameters

Temperature 250 ºC

mV signal 0.12 – 0.40 mV

Attenuation x1

Range x1

Output Integrator

Offset 5.0 mV

Filter 200 ms

Flows and Pressures

Carrier pressure 12-18 psig.

FID air flow 350 – 400 mL/min.

FID H2 flow 35-40 mL/min.

Preset Air regulator 50-80 psig.

AutoGC Methods – Clarus Gas Chromatograph

September 26, 2018TCEQ Operator Training 6

Page 7: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

SYSTEMS WITH NCI902 INTERFACE

Totalchrom only collects signals from each FID

Cut-time is set only from the GC

Instrument method contains relay events to control sample type

X-sample.mth - ambient sample collection

X-cal.mth - collects diluted standard from blender

X-blank.mth – collects only diluent from blender

X-lcs.mth – collects static dilution LCS

X-rts.mth – collects full 40 min. sample from RTS can

X-spike.mth – collects 2 min. from RTS can and 38 min. of diluent from blender

SYSTEMS WITH INTEGRAL INTERFACE

Totalchrom controls all setpoints in Clarus GC for temperatures and flows

Totalchrom controls the cut-time in the method.

Totalchrom controls relay events to control samples selected

X-sample.mth - ambient sample collection

X-cal.mth - collects diluted standard from blender

X-blank.mth – collects only diluent from blender

X-lcs.mth – collects static dilution LCS

X-rts.mth – collects full 40 min. sample from RTS can

X-spike.mth – collects 2 min. from RTS can and 38 min. of diluent from blender

AutoGC Methods – Totalchrom Instrument Methods

September 26, 2018TCEQ Operator Training 7

Page 8: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

PLOT METHOD

Controls peak detection and

integration of all PAMS targets in the

PLOT column chromatogram

Detector signal = FID A

All components must go through the

BP1 column first

Last target component on the PLOT

column determines the cut-time on

the GC (2-methyl-1-pentene or 1-

hexane)

BP1 METHOD

Controls peak detection and

integration of all PAMS targets in the

BP1 column chromatogram

Detector signal = FID B

Connects to the Deans switch and

flows first to the PLOT column and

then to a deactivated restrictor after

the Deans switch time.

After Deans switch first peak to elute

from the BP1 column is n-Hexane

AutoGC Methods – Totalchrom Process Methods

September 26, 2018TCEQ Operator Training 8

Page 9: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

AutoGC Overlapping Sample Collection and Analysis

September 26, 2018TCEQ Operator Training 9

0 10 20 30 40 50 60

0 10 20 30 40 50 60

Collect Sample

Desorb

Heat Trap & Hold

Recycle to -30 degrees

Turbomatrix

System Ready

Start Analysis

Recycle

Gas Chromatograph

System Ready

Collecting Data

Process Data

Data System

0 10 20 30 40

48

48

40

101

7

1

Page 10: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Sample Collection and Analysis

September 26, 2018TCEQ Operator Training 10

Sample Collection

Sample AnalysisFID A

FID B

FID A

FID B

BP1 Column

PLOT Column

Restrictor

Outlet Split

Nafion® Drier

Sample Pump

Trap

Weak Strong

C2 C3–C10

Collection Desorption

Trap Heat

Page 11: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Sequences are used to automate sample collection

Allows for repeated unattended sample analysis

Specifies necessary instrument file for each hour

Applies processing method to each raw signal collection

Generates results files and ASCII text report files

EZ Sequence Visual Basic program

To generate a sequence which schedules daily check samples and weekly QC samples

Provides descriptive information of the collection hour for each sample

Generates unique filename based on site, sample type, month, day and hour of sample collection

September 26, 2018TCEQ Operator Training 11

AutoGC Totalchrom Sequences

ABSK25A.rst

hourday

month

Sample type

ColumnSite

Site designation: up to 3 alpha-numeric characters

Column: B = BP1; P = PLOT

Sample type: S = ambient; C = CVS; B = Blank; E = LCS; Q = RTS; D = MDL; M = multipoint

Month: A = Jan; B = Feb ….. L = Dec

Day: day-of-month

Hour: A = 00, B = 01, C = 03 …. X = 23

Page 12: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Building Sequences

AutoGC EZ Sequence

September 26, 2018TCEQ Operator Training 12

Page 13: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Building Sequences

September 26, 2018TCEQ Operator Training 13

Page 14: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Building Sequences

September 26, 2018TCEQ Operator Training 14

This import sequence is

properly configured for

the necessary 2-channel

operations

Page 15: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Save and Load Sequence

September 26, 2018TCEQ Operator Training 15

To Load Sequence:

• Find the row for the

starting sample hour

• Use Setup from the TC

Navigator

• Enter the sequence

filename and row to

start

Page 16: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Setting Up the Sequence to Run

September 26, 2018TCEQ Operator Training 16

Page 17: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Setting Up the Sequence to Run

September 26, 2018TCEQ Operator Training 17

Page 18: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Setting Up the Sequence to Run

September 26, 2018TCEQ Operator Training 18

Page 19: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Instrument Loading Sequence into GC Memory – Setting Up

September 26, 2018TCEQ Operator Training 19

Page 20: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Details: Clarus GC Integral Interface vs. NCI902 A/D Interface

September 26, 2018TCEQ Operator Training 20

Page 21: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Modifying the Active Sequence

September 26, 2018TCEQ Operator Training 21

Page 22: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Modifying the Active Sequence

September 26, 2018TCEQ Operator Training 22

Page 23: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Modifying Sequences

September 26, 2018TCEQ Operator Training 23

To Modify a sequence:

• Instrument Method

• Sample Name

• File Name

For both Channels

Page 24: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Modifying Sequences

September 26, 2018TCEQ Operator Training 24

The same process is used

to schedule Calibration

curves. Calibrations are

run manually by changing

the setpoints on the

dilution system after the

sequence has been setup.

Page 25: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Modifying Sequences

September 26, 2018TCEQ Operator Training 25

Audit samples or

additional retention time

check samples can be

scheduled to run when

needed.

Page 26: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Modifying Sequences

September 26, 2018TCEQ Operator Training 26

After saving the

modified sequence the

system prompts for the

next starting row to

resume.

Page 27: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Resetting the Sequence row

September 26, 2018TCEQ Operator Training 27

If your system gets off-

time, you can reset the

sequence by simply

using pause and resume

to choose a different

row.

Page 28: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Calibration Curves

September 26, 2018TCEQ Operator Training 28

Automatic introduction of QC

samples

Dynamically diluted check

standard

Analytical blank

Manual dilution of multipoint

calibration curve

Dilution from 100 ppbv or 1

ppmv multi-component standard

Merlin MicroScience Dilution System

Page 29: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Calibration Curves – MMSD Spreadsheet

September 26, 2018TCEQ Operator Training 29

The Blender Spreadsheet

is a useful tool for

tracking system

performance and

documenting calibration

curves. This workbook is

protected and highly

linked.

Page 30: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Calibration Curves – MMSD Spreadsheet

September 26, 2018TCEQ Operator Training 30

The current CVS Standard

is entered into the CVS

Worksheet. This allows

operators to enter the

measured concentrations

from their nightly check

samples and see the

recoveries.

Page 31: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Calibration Curves – MMSD Spreadsheet

September 26, 2018TCEQ Operator Training 31

The LCS Statically diluted

second source standard

concentrations and

dilution factor will be

entered on the LCS

worksheet for tracking

LCS recoveries.

Page 32: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Calibration Curves – MMSD Spreadsheet

September 26, 2018TCEQ Operator Training 32

A worksheet is also

provided for calculating

weekly precision checks.

Page 33: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Calibration Curves – MMSD Spreadsheet

September 26, 2018TCEQ Operator Training 33

The Blender Ratios worksheet shows the calibrated flows and allows operators to

enter pressures and see resulting flow rates. It also has a section for recording

annual flow checks.

Page 34: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Calibration Curves – MMSD Spreadsheet

September 26, 2018TCEQ Operator Training 34

The Curve worksheet shows the pressures and resulting blend ratios used for the calibration

curve. Operators must change pressure settings for each sample collection and record the

resulting area counts for propane and benzene.

Page 35: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

INSTRUMENT METHODS

A single method for control of the gas

chromatographic parameters

Includes temperatures and pressures

as well as valve event for Deans switch

Includes relay events to control

samples collected on the Turbomatrix

All instrument methods are designed

to be exactly the same except for

valving required to change sample

types

PROCESS METHODS

A single method for each signal, 2

methods for each sample collection

Contains all targets and their expected

retention times for identification

Contains response factor to be used to

quantitate all identified targets.

Also contains noise and area thresholds

and bunching factors used to facilitate

accurate peak identification

Totalchrom Methods – Dual Channel Methods

September 26, 2018TCEQ Operator Training 36

Page 36: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Totalchrom Methods

September 26, 2018TCEQ Operator Training 37

Instrument Methods

and Process Methods

are the same, but are

used differently

Page 37: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Instrument Methods – xx-Sample.mth

September 26, 2018TCEQ Operator Training 38

Methods which are

designated as

Instrument methods in

the sequence will

control the instrument

parameters.

Page 38: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Instrument Methods – xx-Sample.mth

September 26, 2018TCEQ Operator Training 39

There is no carrier

designated in the

Instrument Methods

since the carrier flow

is being supplied by

the Turbomatrix

Page 39: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Instrument Methods – xx-Sample.mth

September 26, 2018TCEQ Operator Training 40

The Aux 2 carrier is

used to supply the

mid-point pressure

used by the Dean’s

switch.

Page 40: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Instrument Methods – xx-Sample.mth

September 26, 2018TCEQ Operator Training 41

Detector settings are

the same for both

channels

Page 41: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Instrument Methods – xx-Sample.mth

September 26, 2018TCEQ Operator Training 42

Instrument Methods

also contain the relay

events necessary to

determine which

sample is being

collected by the

Turbomatrix. This

means there must be a

different Instrument

method for each

sample type.

Page 42: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Instrument Methods – xx-Sample.mth

September 26, 2018TCEQ Operator Training 43

There should be

nothing entered into

the component list on

any instrument

method

Page 43: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Process Methods – xx-PLOT.mth

September 26, 2018TCEQ Operator Training 44

Process Methods contain

only the information

necessary for peak

identification and

quantitation. There is

one Process method for

each detector

signal/column.

Page 44: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Process Methods – xx-PLOT.mth

September 26, 2018TCEQ Operator Training 45

You can review the

Method Summary or

the Component List by

selecting which

window is shown.

Page 45: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Process Methods – xx-PLOT.mth

September 26, 2018TCEQ Operator Training 46

All the information for

the components in the

PLOT chromatogram can

be reviewed here,

however optimization is

more easily accomplished

graphically.

Page 46: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Process Methods – xx-PLOT.mth

September 26, 2018TCEQ Operator Training 47

The Instrument section

of the method has

default values that are

not used when the

method is designated

in the sequence as a

process method.

Page 47: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Process Methods – xx-PLOT.mth

September 26, 2018TCEQ Operator Training 48

Process parameters

such as integration

events can be reviewed

and defaults changed

globally. However,

optimization is better

done in Graphic Editor.

Page 48: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Process Methods – xx-PLOT.mth

September 26, 2018TCEQ Operator Training 49

The user programs tab is

used to designate the

use of a user program

like mmove to be

executed at the end of

a chromatographic run.

Page 49: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Process Methods – xx-PLOT.mth

September 26, 2018TCEQ Operator Training 50

While most

optimization occurs in

the Graphic Editor,

calibration information

must be entered

directly into the

method.

Page 50: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Process Methods – xx-PLOT.mth

September 26, 2018TCEQ Operator Training 51

Once the response

factor is determined

when running a

calibration curve it

must be entered in the

global information to

be used for all

unidentified peaks.

Page 51: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Process Methods – xx-PLOT.mth

September 26, 2018TCEQ Operator Training 52

Since the system uses

a carbon response

factor for all

components it also

must be entered

directly into each

method.

Page 52: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Process Methods – xx-PLOT.mth

September 26, 2018TCEQ Operator Training 53

You can review the

parameters for

each component by

double clicking on

any one in the list.

Page 53: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Process Methods – xx-BP1.mth

September 26, 2018TCEQ Operator Training 54

The BP1 Method is

setup the same as the

PLOT but with a

different component

list.

Page 54: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Process Methods – xx-PLOT.mth

September 26, 2018TCEQ Operator Training 55

The Mmove user

program is initiated

from the BP1 method

at the end of the

analysis. Mmove

makes copies of the

ASCII text file and

moves them to the

polling directory

hourly. It also archives

the daily data.

Page 55: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Modifying Methods

September 26, 2018TCEQ Operator Training 56

Method are protected

from editing by CAMS

locks when the system

is running. To modify

methods these CAMS

locks must be

removed.

Page 56: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Modifying Methods

September 26, 2018TCEQ Operator Training 57

Remove CAMS locks by

deleting the methods

from the CAMS

Administration list.

Page 57: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Modifying Methods

September 26, 2018TCEQ Operator Training 58

Once the methods are

removed from the list

you may edit them

without interruption.

This applies to Process

methods. Removing

locks from instrument

methods or sequences

could result in errors.

Page 58: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Modifying Methods

September 26, 2018TCEQ Operator Training 59

Edit Methods using the

Build Method icon in

the Totalchrom

Navigator.

Page 59: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Modifying Methods

September 26, 2018TCEQ Operator Training 60

If you do not see your

method in the recently

edited method list you

may use Load method

stored on disk to

browse to the

D:\TC\xx-meth

directory where all

your active methods

should be.

Page 60: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Calibration Curves – Updating Process Methods for RF

September 26, 2018TCEQ Operator Training 61

When calibration curves

are run and new

Response Factors (RF)

are calculated they must

be entered into the

methods. This is done

from the Components

drop-down menu. RFs

are entered in the Global

Information and in the

Change Calibration Info

sections of the method.

Page 61: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Modifying Methods – Updating Process Methods for RF

September 26, 2018TCEQ Operator Training 62

In the Global

Information section of

the method the

response factor is used

to calculate the

concentration of all

unidentified peaks on

each column.

Page 62: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Modifying Methods – Updating Process Methods for RF

September 26, 2018TCEQ Operator Training 63

To change the response

factor for all target

compounds the new RF

must be entered using

the Change calibration

type check box and

entering the new RG

under use calibration

factor. Both the

change calibration

information and global

information must be

updated each time the

RF changes.

Page 63: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 64

Many factors can cause

peaks to be incorrectly

identified. Operators

must review

chromatograms often

to insure all targets

are being identified

correctly. Occasionally

method must be

“optimized” and

retention time

windows adjusted.

Page 64: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 65

After removing CAMS

locks and opening the

method use the Other

Graphic Editor to

graphically adjust

retention time

windows in each

method.

Page 65: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 66

The weekly RTS check is

a good place to start

since all the target

compounds will be in

this sample.

Page 66: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 67

Page 67: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 68

Page 68: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 70

Page 69: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 71

Page 70: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 72

Page 71: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 73

Page 72: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 74

Page 73: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 75

Page 74: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 76

Page 75: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 77

Page 76: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 78

Page 77: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 79

Page 78: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 80

Page 79: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 81

Page 80: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 82

Page 81: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 83

Page 82: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 84

Page 83: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Optimizing Methods for Peak Identification

September 26, 2018TCEQ Operator Training 85

Page 84: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

September 26, 2018TCEQ Operator Training 86

Analytical Tools and Software to Assess Site Performance

Page 85: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Site Reports (Orsat)

Rhone

Remote Connection/Site Visit

MAX - Merlin AutoGC Xplorer

Totalchrom Software – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 87

Analytical Tools and Software to Assess Site Performance:

Page 86: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

E:\Operator presentation\18_09_18_a_clinton.xls

September 26, 2018TCEQ Operator Training 88

Analytical Tools and Software to Assess Site Performance:

Site Reports

Page 87: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

September 26, 2018TCEQ Operator Training 89

Analytical Tools and Software to Assess Site Performance:

Rhone - Resources

Page 88: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

September 26, 2018TCEQ Operator Training 90

Analytical Tools and Software to Assess Site Performance:

Rhone – QC Report

Page 89: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

September 26, 2018TCEQ Operator Training 91

Analytical Tools and Software to Assess Site Performance:

Rhone – Monthly Report by Site

Page 90: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

TotalChrom Details: view sequence file,

current row, sample date and hour

D: Current folder – RST files, TXO files, TDS

D: Archive folder – Weekend files: LCS, RTS

spike, CVS, blanks, Daily Zip files

Site Visit: Instrument displays for splash

screens(power failures), TD event log,

equipment, support gas, and QC standards

for pressure/flow measurements

September 26, 2018TCEQ Operator Training 92

Analytical Tools and Software to Assess Site Performance:

Remote connection/Site Visit

Page 91: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

http://prod.orsatmax.com/login

September 26, 2018TCEQ Operator Training 93

Analytical Tools and Software to Assess Site Performance:

MAX – Merlin AutoGC Xplorer

Page 92: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Analytical Tools and Software to Assess Site Performance:

Chromatograms – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 94

The Chromatograms Icon

in the TC Navigator will

allow operators to overlay

multiple chromatograms.

This allows evaluation of

the amount of retention

time shifting is occurring

as well as facilitate

determining if peaks at

low concentrations are

being correctly identified.

Page 93: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Analytical Tools and Software to Assess Site Performance:

Chromatograms – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 95

Page 94: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Analytical Tools and Software to Assess Site Performance:

Chromatograms – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 96

Page 95: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Analytical Tools and Software to Assess Site Performance:

Chromatograms – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 97

Page 96: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Analytical Tools and Software to Assess Site Performance:

Chromatograms – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 98

Page 97: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Analytical Tools and Software to Assess Site Performance:

Chromatograms – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 99

Page 98: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Analytical Tools and Software to Assess Site Performance:

Chromatograms – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 100

Page 99: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Analytical Tools and Software to Assess Site Performance:

Chromatograms – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 102

Page 100: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Analytical Tools and Software to Assess Site Performance:

Chromatograms – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 103

Page 101: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Analytical Tools and Software to Assess Site Performance:

Chromatograms – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 104

Page 102: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Analytical Tools and Software to Assess Site Performance:

Chromatograms – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 105

Page 103: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Analytical Tools and Software to Assess Site Performance:

Chromatograms – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 106

Page 104: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Analytical Tools and Software to Assess Site Performance:

Chromatograms – Overlaying Chromatograms

September 26, 2018TCEQ Operator Training 107

Page 105: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

System Status

Binding the Instrument

Method Loss

Port Comm Error

Unknown Instrument/Reconfigure Steps

September 26, 2018TCEQ Operator Training 108

Clarus GC Integral Interface: Understanding the TotalChrom

Software Appearance and Common Software Errors

Page 106: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

September 26, 2018TCEQ Operator Training 109

Clarus GC Integral Interface: System Status Ready

Page 107: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

September 26, 2018TCEQ Operator Training 110

Clarus GC Integral Interface: System Status Ready

Page 108: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Clarus GC Integral Interface: Binding the Instrument

September 26, 2018TCEQ Operator Training 111

Page 109: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Clarus GC Integral Interface: Port Comm Error

September 26, 2018TCEQ Operator Training 112

Ensure the GC interface cable is connected.

Page 110: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Clarus GC Integral Interface: Method and Sequence Loss

September 26, 2018TCEQ Operator Training 113

Due to power outages.

System will continue to run without

the sequence but will not create any

data files.

Reloading the sequence is necessary.

(Approximately 6 minutes to load)

CMD: None, Not Ready, Ready, Run

Page 111: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Clarus GC Integral Interface: Unknown Instrument/Reconfigure

September 26, 2018TCEQ Operator Training 114

Page 112: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Clarus GC Integral Interface: Unknown Instrument/Reconfigure

September 26, 2018TCEQ Operator Training 115

Page 113: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Clarus GC Integral Interface: Unknown Instrument/Reconfigure

September 26, 2018TCEQ Operator Training 116

Page 114: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Clarus GC Integral Interface: Unknown Instrument/Reconfigure

September 26, 2018TCEQ Operator Training 117

Page 115: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Clarus GC Integral Interface: Unknown Instrument/Reconfigure

September 26, 2018TCEQ Operator Training 118

Continue to select Yes and OK to each

command prompt until reconfiguration is

complete.

Page 116: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Clarus GC Integral Interface: Unknown Instrument/Reconfigure

September 26, 2018TCEQ Operator Training 119

Page 117: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Clarus GC Integral Interface: Unknown Instrument/Reconfigure

September 26, 2018TCEQ Operator Training 120

Page 118: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Clarus GC Integral Interface: Unknown Instrument/Reconfigure

September 26, 2018TCEQ Operator Training 121

Page 119: PAMS AutoGC€¦ · Temperatures Transfer line 200 ºC Valve 150 ºC Tube 0 ºC Trap –low -30 ºC Trap –high 325 ºC Rate 40 ºC/min Times Cycle 60 min. Sampling 40 min. Trap

Questions?

Concerns?

Feed Back?

Thank you all for your time and attention. Good luck!