PACS PACS FM-ILT SPECTROMETER FM-ILT SPECTROMETER SPATIAL CALIBRATIONSPATIAL CALIBRATION
A. A. ContursiContursi
(H. Feuchtgruber)(H. Feuchtgruber)
PACS Science Verification Review – 8/9 November 2007PACS Science Verification Review – 8/9 November 2007MPE-GarchingMPE-Garching
![Page 2: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/2.jpg)
TESTS EXECUTEDTESTS EXECUTED
3 27x27 rasters, ¼ pixel step, 3 chopper 3 27x27 rasters, ¼ pixel step, 3 chopper position, @ 0', +-3' (AOT), same grating position, @ 0', +-3' (AOT), same grating positionposition
3 9x9 rasters,1/4 pixel step, around 3 9x9 rasters,1/4 pixel step, around central pixel FOV, 3 grating position, central pixel FOV, 3 grating position, same chopper position @ 0'same chopper position @ 0'
![Page 3: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/3.jpg)
TESTS EXECUTIONTESTS EXECUTION
External BB @ External BB @ 1000 1000 °°CC
Hole = 1.5 mmHole = 1.5 mm
![Page 4: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/4.jpg)
PSFs in every modulePSFs in every module
BLUEBLUE
Telemetry drop!Telemetry drop!no time for no time for re-executionre-execution
![Page 5: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/5.jpg)
PSFs in every modulePSFs in every module
REDRED
Cross Talks with Cross Talks with module 11. module 11. Present in all Present in all modules, all chopmodules, all choppositionposition
Module 11Module 11
![Page 6: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/6.jpg)
FMILT 1-2-3 comparisonFMILT 1-2-3 comparison
FMILT-2FMILT-2FMILT-1FMILT-1 FMILT-3FMILT-3
![Page 7: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/7.jpg)
COMPARISON WITH COMPARISON WITH THEORETICAL PSFTHEORETICAL PSF
Along chopper directionAlong chopper direction
BLUEBLUE
REDRED
Mod 0Mod 0 Mod 12Mod 12 Mod 24Mod 24
ObservedObserved
TheoreticalTheoretical
![Page 8: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/8.jpg)
COMPARISON WITH COMPARISON WITH THEORETICAL PSFTHEORETICAL PSForthogonal to chopper directionorthogonal to chopper direction
BLUEBLUE
REDRED
Mod 0Mod 0 Mod 12Mod 12 Mod 24Mod 24
ObservedObserved
TheoreticalTheoretical
![Page 9: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/9.jpg)
Spectrometer FOV fingerprintSpectrometer FOV fingerprintall spectral pixels averagedall spectral pixels averaged
![Page 10: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/10.jpg)
Rotation array-chopper directionRotation array-chopper directionNON LINEAR!NON LINEAR!
The dependence The dependence of the rotation angle of the rotation angle from the chopperfrom the chopperposition is not linear.position is not linear.
Consistent with Consistent with photometer results photometer results obtained with more obtained with more chopper positionchopper position
![Page 11: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/11.jpg)
Spectrometer FOV fingerprintSpectrometer FOV fingerprinteach spectral pixeleach spectral pixel
½ pixel½ pixel
![Page 12: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/12.jpg)
Spectrometer FOV fingerprintSpectrometer FOV fingerprinteach spectral pixeleach spectral pixel
½ pixel½ pixel
![Page 13: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/13.jpg)
Position of pixels in the skyPosition of pixels in the skychop +Lchop +L chop 0chop 0 chop -Lchop -L
Assuming XY Assuming XY stage to skystage to skyconversion:conversion:
1”= 0.18 mm1”= 0.18 mm
Assuming chopperAssuming choppercommanded units, commanded units, angle conversionangle conversion
![Page 14: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/14.jpg)
Dependence on grating: BLUEDependence on grating: BLUE
![Page 15: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/15.jpg)
Dependence on grating: REDDependence on grating: RED
![Page 16: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/16.jpg)
Conservation of the fluxConservation of the flux
Full dynamical rangeFull dynamical range High cutHigh cut
BLUE homogeneous: 20% maximum variationBLUE homogeneous: 20% maximum variation
Sum of 27x27 raster in each position after havingSum of 27x27 raster in each position after having
normalized each raster to the PSF peak valuenormalized each raster to the PSF peak value
![Page 17: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/17.jpg)
Conservation of the fluxConservation of the flux
Full dynamical rangeFull dynamical range High cutHigh cut
RED: high signal (up to a factor of 2)RED: high signal (up to a factor of 2)from module 11 cross talkfrom module 11 cross talk
Sum of 27x27 raster in each position after havingSum of 27x27 raster in each position after having
normalized each raster to the PSF peak valuenormalized each raster to the PSF peak value
![Page 18: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/18.jpg)
RASTER MAP RECONSTRUCTIONRASTER MAP RECONSTRUCTION
!! CAUTION: not Flat Fielded !!!! CAUTION: not Flat Fielded !!
GHOST!GHOST!~< 0.1%~< 0.1%of Bright of Bright SourceSource
GHOST!GHOST!
![Page 19: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/19.jpg)
GHOSTs in the BLUEGHOSTs in the BLUE?? 1 moving ghost or 2 fixed ghosts ???? 1 moving ghost or 2 fixed ghosts ??
![Page 20: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/20.jpg)
RASTER MAP RECONSTRUCTIONRASTER MAP RECONSTRUCTION
!! CAUTION: not Flat Fielded !!!! CAUTION: not Flat Fielded !!
Background structuresBackground structuresworse in the REDworse in the REDresult in alignment result in alignment problems:problems:
reconstructed PSF reconstructed PSF elongatedelongated
Possible contributionPossible contributionfrom cross talk offrom cross talk ofmodule 11module 11
![Page 21: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/21.jpg)
CONCLUSIONSCONCLUSIONS
Almost all optical misalignments Almost all optical misalignments successfully correctedsuccessfully corrected
Observed PSF widths as expectedObserved PSF widths as expected
Chopper-array rotated by ~3 degreeChopper-array rotated by ~3 degree
Cross talk module 11 in the REDCross talk module 11 in the RED
No dependency on grating positionNo dependency on grating position
Faint ghosts in the BLUE: potential Faint ghosts in the BLUE: potential concerns for weak extended sourcesconcerns for weak extended sources
![Page 22: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/22.jpg)
OPEN ISSUEsOPEN ISSUEs
Module 11 cross talks in the DUMMY Module 11 cross talks in the DUMMY and OPEN in RED and BLUE: sometime and OPEN in RED and BLUE: sometime only in 1 supply group, sometime in all only in 1 supply group, sometime in all modules.modules.
Number of ghosts in the BLUE: 1 or 2? Number of ghosts in the BLUE: 1 or 2? Which is their origin? Inside PACS or Which is their origin? Inside PACS or external window? (may be some external window? (may be some answers in FS)answers in FS)
![Page 23: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/23.jpg)
Spectrometer Spatial Calibration Spectrometer Spatial Calibration in PVin PV
External BB @ 1000 External BB @ 1000 °°C corresponds to:C corresponds to:
125 Jy at 77 micron125 Jy at 77 micron50 Jy at 154 micron50 Jy at 154 micron
This gives :This gives :
S/N ~ 100 in the BLUES/N ~ 100 in the BLUES/N ~ 12 in the REDS/N ~ 12 in the RED
Background levels:Background levels:
3000 Jy in the BLUE3000 Jy in the BLUE1700 Jy in the RED1700 Jy in the RED
![Page 24: PACS FM-ILT SPECTROMETER SPATIAL CALIBRATION A. Contursi (H. Feuchtgruber) PACS Science Verification Review – 8/9 November 2007 MPE-Garching](https://reader036.vdocuments.us/reader036/viewer/2022062409/56649ec65503460f94bd126b/html5/thumbnails/24.jpg)
Spectrometer Spatial Calibration Spectrometer Spatial Calibration in PVin PV
Expected Herschel background levelExpected Herschel background level
~450 Jy at 77 micron (factor 6 less)~450 Jy at 77 micron (factor 6 less)~175 Jy @ 154 micron (factor 10 less)~175 Jy @ 154 micron (factor 10 less)
Duration of 27x27 FMILT-3 raster ~2.5hDuration of 27x27 FMILT-3 raster ~2.5h
Doubled in PV due to chopping Doubled in PV due to chopping to correct to correct
for responsivity changes due to glitchesfor responsivity changes due to glitches