pacemaker: narrow pulses generation for design and ... · the pacemaker designer needs to allow for...

5
ARTICLE REPRINT Pacemaker: Narrow Pulses Generation for Design and Sensitivity Test A pacemaker is a small device that helps the heart to beat more regularly and properly. It does this with a small electric stimulation that controls the heartbeat. This paper outlines pulse generation and sensitivity test of a pacemaker. Designers require a generated signal to simulate the real pacemaker pulse signal to use for sensitivity testing of the pacemaker. Next generation of function generators offer the best signal generators that can simulated a pulse signal as narrow as 1 µsec pulse width and arbitrary cardiac signal. The new digital or mixed signal oscilloscope, with its high resolutions acquisition, allows designers to analyze both low amplitude (0.5 mV and 1 mV) ECG signal and pulse signal stimulus simultaneously, to make real- time adjustments to their pacemaker designs.

Upload: others

Post on 04-Jul-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Pacemaker: Narrow Pulses Generation for Design and ... · The pacemaker designer needs to allow for many adjustments, including the signal’s amplitude and pulse width. These are

A R T I C L E R E P R I N T

Pacemaker: Narrow Pulses Generation for Design and Sensitivity TestA pacemaker is a small device that helps the heart to beat more regularly and properly. It does this with a small electric stimulation that controls the heartbeat. This paper outlines pulse generation and sensitivity test of a pacemaker. Designers require a generated signal to simulate the real pacemaker pulse signal to use for sensitivity testing of the pacemaker. Next generation of function generators of fer the best signal generators that can simulated a pulse signal as narrow as 1 µsec pulse width and arbitrary cardiac signal. The new digital or mixed signal oscilloscope, with its high resolutions acquisition, allows designers to analyze both low amplitude (0.5 mV and 1 mV) ECG signal and pulse signal stimulus simultaneously, to make real-time adjustments to their pacemaker designs.

Page 2: Pacemaker: Narrow Pulses Generation for Design and ... · The pacemaker designer needs to allow for many adjustments, including the signal’s amplitude and pulse width. These are

Pacemaker: Narrow Pulses Generation for Design

and Sensitivity Test

Kah-Meng Chew

General Electronics Measurement Solutions

Keysight Technologies Sdn. Bhd.

Bayan Lepas, Malaysia

[email protected]

Doris Lau

General Electronics Measurement Solutions

Keysight Technologies Sdn. Bhd.

Bayan Lepas, Malaysia

[email protected]

Abstract— A pacemaker is a small device that helps the heart

to beat more regularly and properly. It does this with a small

electric stimulation that controls the heartbeat. This paper

outlines pulse generation and sensitivity test of a pacemaker.

Designers require a generated signal to simulate the real

pacemaker pulse signal to use for sensitivity testing of the

pacemaker. Next generation of function generators offer the best

signal generators that can simulated a pulse signal as narrow as

1usec pulse width and arbitrary cardiac signal. The new digital

or mixed signal oscilloscope, with its high resolutions acquisition,

allows designers to analyze both low amplitude (0.5 mV and 1

mV) ECG signal and pulse signal stimulus simultaneously

allowing designers to make real-time adjustments to their

pacemaker designs.

Keywords— pacemaker, pulse signal, pacing, ECG, narrow pulse,

pacemaker sensitivity, cardiac

Introduction

What is a pacemaker? According to the American Heart

Association, a pacemaker is a small device that helps the heart

beats more regularly and properly. It does this with a small

electric stimulation that helps control the heartbeat. An

engineer may refer to a pacemaker device as a “pulse

generator”. Pacemakers perform four major critical functions:

• Stimulate cardiac depolarization;

• Sense intrinsic cardiac function;

• Respond to increase metabolic demand by providing

rate responsive pacing; and

• Provide diagnostic information stored by the

pacemaker

In clinical use, three common types of pacemakers are [1]

1. Implantable,

• Long-term permanent use

2. External, and

• Miniaturized, transistorized, battery-

powered

3. Console

• Battery- or AC-powered defibrillators

In addition, other sub-options includes unipolar pacing,

bipolar pacing, asynchronous (fix rate), synchronous (on

demand), single or dual chamber, programmable and non-

programmable.

Pacing system

In the typical pacing system, the pacemaker provides the

voltage and the current (electrons) flows down the conductor

through the electrode, and towards the tip of the lead (positive

terminal, cathode). The tip of the lead touches the myocardium

(the muscular tissue of the heart) where the electrical

resistance is produced and stimulates the heart. Then the

current flows through the body tissue to the tip of the lead

(negative terminal, anode) and back to the pacemaker.

Electrical Testing of Pacemaker

Electrical testing parameters involve in testing a pacemaker,

include battery testing, pacing impedance and pulse

generation. Within pulse generation, there are pulse

generation, sensitivity test, timing test, mode switching and

rate adaptive sensor. The focus in this paper is on pulse

generation and sensitivity test of the pacemaker.

I. PULSE GENERATION

The pacemaker designer needs to allow for many adjustments,

including the signal’s amplitude and pulse width. These are

the most important settings used in the pacemaker pacing.

Pulse width is the time or duration of the pacing pulse and is

expressed in milliseconds. Pulse width must be wide enough

to allow for the delivery of each pacing pulse and must be

long enough for the depolarization to disperse into the

surrounding muscle tissues. Widening the pulse width

improves the ability of the pacemaker to capture the cardiac

signal and maintains synchronization. Narrowing the pulse

width improves battery life span and reduces side effects of

electrical signals; these side effects include twitching of the

pectoral muscles and interrupting the normal breathing pattern

by stimulating the diaphragm. The pacemaker produces a

narrow pulse signal approximately 71 beats per minute and a

pulse width around 0.5 msec. [2]

Page 3: Pacemaker: Narrow Pulses Generation for Design and ... · The pacemaker designer needs to allow for many adjustments, including the signal’s amplitude and pulse width. These are

Amplitude is the amount of voltage delivered to the heart by

the pacemaker. Amplitude reflects the strength or height of the

pulse signal. The amplitude of the pulse must be large enough

to cause depolarization (i.e. to “capture” the heart), “capture”

is the nomenclature for effective stimulation of cardiac

depolarization by the pacemaker. On top of that, the amplitude

of the pulse must be sufficient to provide an appropriate

pacing safety margin. The pacemaker produces a pulse burst

amplitude approximately 5.0V and this corresponds to a

narrow “spike” in the ECG.

Figure 1 [3]: This is the output pulse of the pacemaker. The

voltage of the pacemaker is referring to the amplitude of the

leading edge. The droop is influenced by many factors

including the electrode’s lead impedance. The output pulse

corresponds to a narrow pulse or “spike” on the ECG signal.

A. Pulse Generations Solutions

A function generator is the best solution for a pacemaker

designer to simulate the narrow pulses. Most of the new

function generators have built-in pulse capability to generate

basic pulses. For older function generators, the designer can

use the built-in square wave capability to generate pulse signal

by varying the duty cycle between 20 and 80 percent. Some

designers also use the arbitrary waveform generator (AWG) to

generate a pulse because it allows them to top up with

additional customizations such as adding overshoot or

undershoot into the pulse signal.

Questions might arise on why the designer does not use a PC

sound card to generate a pulse signal. The disadvantages of

using a PC sound card include poor waveform quality,

distortion, noise, signal ringing and imprecise amplitude.

These disadvantages far outweigh the main advantages of

sound card’s low cost and availability.

B. Narrower Pulses

Pacemaker designers require narrow pulses for testing. This is

achieved by using the built-in BURST mode capability in the

function generator or arbitrary function generator. BURST

mode capability allows the user to configure the instrument to

output a waveform for a specific number of cycles. The user

can control the amount of time that elapses between bursts

with the internal timer or external triggering.

For example, the designer can output a very narrow pulse

signal of 1us pulse width by adjusting the frequency range

value and duty cycle of the signal, then switch on the BURST

mode feature to adjust the burst rate to achieve the narrow

pulse.

Figure 2: 1µsec pulse signal generate from function generator

II. SENSITIVY TEST FOR PACEMAKER

What is the sensitivity of a pacemaker? [4] This is defined as

the minimum myocardial voltage required to be detected as a

P wave or R wave, measured in mV. Why is this important? If

the pacemaker is overly sensitive, any random fluctuations of

electrical signal from cardiac activity could be mistaken by the

pacemaker and leads the pacemaker to keep pacing

continuously. This can lead to ”madness”. On the other hand,

the pacemaker would not fire at all because the cardiac

activity is considered within normal range. Hence proper

timing and sensitivity need to be calibrated appropriately to

avoid unnecessary complications later.

Figure 3: Typical setup for testing the sensitivity of the

pacemaker

Figure 3 shows the typical test setup for testing the sensitivity

of the pacemaker. In this test, the designer sets the pacemaker

to output a signal of 70 beats per minute. The function

generator produces the cardiac signal to simulate the heart

signal to the pacemaker. The designer tunes the amplitude of

Page 4: Pacemaker: Narrow Pulses Generation for Design and ... · The pacemaker designer needs to allow for many adjustments, including the signal’s amplitude and pulse width. These are

the cardiac signal from the function generator (in mV level)

and the sensitivity of the pacemaker is measured using the

oscilloscope. This minimum sensitivity value is the sensitivity

threshold. Next-generation function generators does not need

an attenuator because they are able to generate an accurate

signal at low amplitude.

Figure 4: This is an oscilloscope screen capture showing both

the cardiac ECG signal and the pacemaker pulse signal.

Showing two signal at the same time allow the user to analyze

their design and make adjustments.

To automate a sensitivity test, the designer needs to write a

program to control the function generator to transmit the

signal burst to the pacemaker and measure the pacemaker

sensitivity level, output pulse width, and beats per minute;

these settings are customized for each patient.

III. SUMMARY

A. Solution for Narrow Pulse Signal Generation

Keysight’s 33622A function generator device is an example of

a state-of-the-art function generator to create pulses for many

applications and is a good solution for pacemaker pulse

simulation. The burst mode allows designers to generate low

duty-cycle pulses from a general-purpose function generator

rather than using a dedicated pulse generator. This device also

allows pacemaker designer to generate cardiac signal and

provides sequencing capability for seamlessly transitioning

through various cardiac signal conditions.

Keysight’s 33622A Waveform Generator (Figure 5) was used

to generate the narrow pulse signal in Figure 2 and the cardiac

signal plus narrow pulse in Figures 4.

Figure 5: Keysight’s 33622A Functional/Arbitrary Waveform

Generator

B. Solution for Sensitivity Test Analysis

Keysight’s InfiniiVision 4000 X-series Oscilloscope is an

example of a next generation oscilloscope in analyzing very

low amplitude signal (such as ECG signals) with the accuracy

needed for medical application. The high update rate allows

the designer to analyze the full spectrum of the signal in high

resolutions acquisition. The signal analysis is also economical

in that it does not require additional accessories to achieve

high vertical resolution for the ECG signal amplitude. In

additional to this, the oscilloscope allows the designer to view

the ECG cardiac signal and the pulse signal stimulus from the

pacemaker.

Keysight’s InfiniiVision MSO-X 4104A (Figure 6) was used

to make measurements in Figures 2 and 4.

Figure 6: Keysight’s InfiniiVision 4000 X-series Oscilloscope

REFERENCES

[1] Doris J. W. Escher, M.D., Types of Pacemakers and their Complications, 1973; 47: 1119-1131

[2] Generating Narrow Pulses with Function Generator, Agilent Technologies, 5898-8033EN, 2008; 1

[3] S. Serge Barold, Roland X. Stroobandt, Alfons F. Sinnaeve, Cardiac Pacemakers Step-by-Step: An Illustrated Guide, 2008; 23

[4] Sensitivity and output settings of the temporary pacemaker, http://www.derangedphysiology.com/main/core-topics-intensive-care/mechanical-haemodynamic-support/Chapter%202.5.3/sensitivity-and-output-settings-temporary-pacemaker

Sponsor by Keysight Technologies Inc.

Page 5: Pacemaker: Narrow Pulses Generation for Design and ... · The pacemaker designer needs to allow for many adjustments, including the signal’s amplitude and pulse width. These are

This information is subject to change without notice. © Keysight Technologies, 2018, Published in USA, July 24, 2018, 5992-3044EN

Page 5Find us at www.keysight.com

Learn more at: www.keysight.comFor more information on Keysight Technologies’ products, applications or services,

please contact your local Keysight office. The complete list is available at:

www.keysight.com/find/contactus

This article was originally published in © 2017 IEEE. Reprinted, with permission, from IEEE International Symposium on MeMeA (Medical Measurements and Applications).