pacemaker data collection from device

Upload: gavin-texeirra

Post on 02-Jun-2018

214 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/11/2019 Pacemaker Data Collection From Device

    1/6

    Automatic analysis of pacemakerdiagnostic data in the identification ofatrial tachyarrhythmias in patients withno prior history of them

    A. Schuchert a,*, S. Lepage b, J.J.Ostrander c,R.J. Bos d,

    M. Gwechenbergere

    , A. Nichollsf

    , G. Ringwaldg

    a Heart Center, University Hospital Hamburg-Eppendorf, Hamburg,

    Martinistrasse 52, D-20251 Hamburg, Germanyb Centre Hospitalier Universitaire de Sherbrooke, Sherbrooke, QC, Canadac Grey Bruce Health Services Corporation, Owen Sound, ON, Canadad Franciscus Ziekenhuis, Roosendaal, The Netherlandse AKH, Wien, Austriaf Leeds General Infirmary, Leeds, UKg Gemeinschafts Praxis, Bruchsal, Germany

    Submitted 27 September 2004, and accepted after revision 30 January 2005

    KEYWORDSpacemakerdiagnostics;expert system;atrial tachyarrhythmia

    Abstract Aims Modern pacemakers provide a large amount of diagnostic data.Given the limited time available during a pacemaker follow-up visit essentialinformation may be overlooked. This registry was conducted to assess the utility ofan expert system that analyses the diagnostic data collected by an implantedpacing device and notifies and advises the physician about suspected technicalissues and arrhythmias that need further attention.Methods Patients with various standard indications for pacing were included inthis registry and received single or dual chamber pacemakers. Data were collectedand analysed by the expert system during at least two subsequent follow-up visits.

    The evaluation of this system focused on data obtained from patients with a dualchamber pacing device without prior history of atrial arrhythmias.Results A total number of 239 patients without prior history of atrial tachyar-rhythmias were included in this analysis. Atrial tachyarrhythmias were detected in73 (31%) of these patients. The highest incidence of newly detected arrhythmiasoccurred in the group of patients with high-degree AV block and VDD pacemakers.

    * Corresponding author. Tel.: C49 404 717 5304; fax: C49 404 717 7420.E-mail address: [email protected](A. Schuchert).

    1099-5129/$30 2005 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved.doi:10.1016/j.eupc.2005.01.005

    Europace (2005) 7, 242e247

    http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-mailto:[email protected]:[email protected]://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-http://-/?-
  • 8/11/2019 Pacemaker Data Collection From Device

    2/6

    Furthermore, newly detected arrhythmias predominantly occurred in the periodshortly after implantation. Device programming suggestions by the expert systemwere adopted in 30% of the cases. Following detection of atrial tachyarrhythmias bythe expert system, pharmacological management was adjusted at 71% of the firstfollow-up visits and at 27% of later follow-up visits.Conclusion Results of this registry show that this expert system providesa valuable tool for the detection of atrial tachyarrhythmias during pacemakerfollow-up visits.

    2005 The European Society of Cardiology. Published by Elsevier Ltd. All rightsreserved.

    Introduction

    Modern pacemakers are capable of storing a largeamount of diagnostic data related to various ar-rhythmias. One of the most striking advantages ofhaving such diagnostic data is that it is collected ona 24-h day and 7-day week basis. This allowsphysicians to evaluate the appropriate function of

    the device, the status and progression of thepatients arrhythmias and the effects of pacingand pharmacological therapies[1e4]. However, italso means that the data have to first be retrievedand carefully analysed, which adds considerably tothe length of each follow-up visit. Furthermore,correct interpretation of the retrieved data re-quires extensive experience with pacemaker man-agement, especially since specialized informationin the form of textbooks or original reports is rare.

    One solution to this problem is to developsoftware that is capable of automatically analysing

    the data obtained during pacemaker interrogation.Such software, which is also referred to as an expertsystem, combines the expert knowledge that isembedded in the software with individual patientdiagnostic data. It is designed to provide relevantmessages if technical issues arise such as sensingfailures or if significant arrhythmic events occur.The progressively increasing functionality of modernpacemaker programmers allows for the introductionof such an expert system, known as DiagnosticObservations (DOs), into daily clinical routine.

    The DORE registry (Diagnostic Observation REg-istry) was designed to carry out a prospective

    evaluation of the performance of an expert systemfor the automatic analysis of diagnostic pacemakerdata in a large patient population.

    Only patients with a newly implanted pace-maker that had the DOs feature were included inthe registry. Since newly implanted devices inparticular often require reprogramming, auto-matic systems that evaluate device performanceand alert the physician to significant arrhythmicevents are expected to have the most utility duringthe immediate post implantation period.

    Methods

    Patients and devices

    Patients with a wide variety of pacemaker indica-tions, who received both single and dual chamberpacemakers, were included in the registry. Nospecific inclusion criteria applied to this registry.

    For the study we used the Clarity DDDR, ClaritySSIR and the Collection 3 series of pacemakers(Saphir VVD/R, Diamond DDDR, Ruby DDD and TopazSSIR, Vitatron BV, Arnhem, The Netherlands). Theseare state-of-the-art pacemakers equipped with theDO function.

    These pacemakers continuously collect data oncardiac events and carry out measurements by, forexample, periodically measuring the amplitude ofatrial signals. This information is then stored in thepacemaker memory for later retrieval and auto-matic analysis by the DOs function. The diagnostic

    information used by the DOs function is identical tothe information provided to the physician; the DOsfunction interprets these data and presents con-densed information and recommendations to thephysician. Information related to possible atrialtachycardia is based on diagnostics, emanatingfrom the mode switching algorithm, such as theamount of atrial sensed rhythm that is classified aspathological and the percentage of atrioventricular(AV) dissociation. Mode switching itself is an algo-rithm that has been used for many years in thesedevices and is based on a beat-to-beat analysis ofthe atrial rhythm[5].

    Messages such as Low paced atrial rates(indicative of chronotropic incompetence), orSuspected atrial undersensing can be solvedby reprogramming sensing parameters or activat-ing rate adaptive pacing. Messages indicat-ing significant arrhythmic events may requireadditional intervention or adjustment of pacingparameters and/or pharmacological therapy.This is very important if the events are new orhave not previously been seen. Particularly rele-vant is any suggestion of the presence of atrial

    Automatic analysis of pacemaker diagnostic data 243

  • 8/11/2019 Pacemaker Data Collection From Device

    3/6

    tachyarrhythmias (AT) that the expert systemdetects on the basis of a relatively low percentageof AV synchrony (Less than 96% AV synchrony) incombination with mode switching (percentage ofpathological atrial rates).

    Design

    One year of postoperative pacemaker follow-updata were used for the DORE registry. This couldinclude up to four follow-up visits: one week afterimplantation, and after 3, 6 (as an option), and 12months. Unscheduled follow-up visits were sepa-rately reported.

    At the start of each follow-up visit, patientsymptoms were recorded and pacemaker datawere retrieved. Upon retrieval, the expert systemanalysed the diagnostic information and recog-nized and responded to several events. The phys-ician was alerted to any notable events by anappropriate message (a Diagnostic Observation),which was displayed on the programmer screen.All DOs were documented at each follow-up. Thephysician was asked to confirm the presentedmessage by reviewing the underlying diagnosticdata, such as 24 h heart rate trend and varioushistograms and event counters. Subsequently, anychanges in pacing and drug therapy and theirrelationship to the DOs were documented.

    Analysis

    The DORE registry included patients with a widevariety of pacemaker indications and includedsingle and dual chamber pacemakers. The dataanalysis presented in this manuscript focuses onthe detection of AT, the most frequent observationin dual chamber devices. Our particular interest wasin patients with no known history of AT and theproportion of these patients in whom the expertsystem detected the presence of this tachyarrhyth-mia. In addition we stratified the patients by in-dication for pacing.

    Results

    Patients

    Multiple centres from eight European countriesand from Canada contributed to the registry. Atotal of 358 pacemaker patients were entered,comprising 192 (53%) males and 166 (47%) females.

    In the first year after implantation, all patientsunderwent at least two follow-up examinations,with 46% of them having at least three. Hospital

    discharge data were available for 351 patients. The1e3 months follow-up was reported for 276 pa-tients, the 5e7 months follow-up for 175 patientsand the 10e14 months follow-up for 146 patients.

    Of all included patients, 280 patients with a dualchamber device had at least two scheduled follow-up visits from which diagnostic pacemaker data

    were available. Of these 280 patients, 145 weremales (age 62.5G 12 years), and 135 were females(age 82.0G 8.4 years).

    At the time of pacemaker implantation, 36% ofpatients had bradycardiaetachycardia syndrome,33% were in sinus rhythm, 30% had sinus nodedysfunction and 1% had other atrial rhythms. Withrespect to the ventricular rhythm, 59% of patientshad normal ventricular activation, 25% presentedwith idioventricular rhythms, 6% had frequentpremature ventricular complexes and 10% hadother ventricular rhythms.

    The indication for pacemaker implantation was

    high-degree AV block in 52% of patients, sinus nodedysfunction in 47%, and other diagnoses in 1%.

    As mentioned earlier, we analysed only patientswithout a history of AT. Of the 280 dual chamberpatients with diagnostic data from at least twofollow-up visits, 239 patients had no prior history ofatrial tachyarrhythmias. These patients were strat-ified according to the indication for pacing. Weconsidered two main subgroups: those with a high-degree AV block (nZ 168) and those without AVblock, i.e. with intact AV conduction (nZ 71). Thefirst group was subdivided into patients receiving

    a dual lead (DDD,nZ

    128) and a single lead (VDD;nZ 40) dual chamber pacemaker.

    Frequency of diagnostic observations

    A total of 1433 follow-up visits were reported andin 323 (23%) one or more DOs were presented.During the follow-up period, 43% of all patientspresented with at least one DOs.

    Fig. 1shows that 43% of all DOs messages wererelated to a low percentage of AV synchrony (Lessthan 96% AV synchrony) and that 12% wererelated to High sensed atrial rates. Both mes-sages are indicative of the presence of AT. Thedevice responds to high sensed atrial rates bymeans of mode switching; mode switching itselfreduces the percentage of AV synchrony.

    Fig. 2illustrates the distribution of DOs over thevarious follow-up visits. Although the absolutenumber of DOs decreases over time, the percent-age of patients with DOs does not substantiallydecline. This may be explained by the fact thatduring early follow-ups many DOs are related to

    244 A. Schuchert et al.

  • 8/11/2019 Pacemaker Data Collection From Device

    4/6

    technical issues such as over- or undersensing.

    Once these issues are resolved, other issues thatare possibly related to the occurrence of AT remainand will trigger the expert system. Non-scheduledfollow-ups were usually due to significant symp-toms. However, DOs obtained during these non-scheduled follow-ups did not frequently result inthe detection of AT that was not previously known.

    Atrial tachyarrhythmia detection

    Patients with high-degree AV blockand a DDD(R) pacemaker

    Table 1 shows the proportion of patients witha DDD(R) pacemaker and high-degree AV blockand no prior history of AT in whom the expertsystem issued a message indicating that such anarrhythmia had indeed been detected. At the firstscheduled follow-up after approximately 3months, AT was detected in 24 patients (19%) andwithin 1 year of follow-up AT was first detected in36 of these patients (almost 30%).

    Patients with high-degree AV blockand a VDD pacemakerTable 2 shows the number of patients with high-degree AV block, a single lead VDD pacemaker andno prior history of AT in whom the expert systemissued a message indicating that such a tachyar-rhythmia had been detected. At the first scheduled

    follow-up visit after approximately 3 months, ATwas detected in 14 patients (35%). In 50% of thesepatients, the presence of AT was detected for thefirst time within 1 year.

    Patients without high-degree AV blockTable 3shows the number of patients with intrinsicAV conduction and no prior history of AT in whomthe expert system issued a message indicating thatsuch a tachyarrhythmia had been detected. At thefirst scheduled follow-up visit after approximately3 months, AT was detected in 10 patients (14%). In22.5% of these patients, the presence of AT was

    detected for the first time within 1 year.The cumulative incidence of newly detected AT

    stratified according to the indication for pacing isshown inFig. 3. Combining all the data shows thatin 73/239 patients (31%) with no prior history of AT,such an arrhythmia was detected over a 1-yearfollow-up period. AT is predominantly detected inthe period shortly after implantation and dimin-ishes after the 3-month follow-up visit. Thesefindings are in line with earlier reports [6].

    Response to diagnostic observations

    Responses to DOs were either pacemaker reprog-ramming, adaptation of the drug regimen or di-agnostic strategy.

    In response to DOs, the suggested programmingrecommendations were followed in 30% of cases.In 71% of the first follow-ups and in 27% of thelater follow-ups, the pharmacological manage-ment was adjusted after atrial tachyarrhythmiashad been detected.

    < 96% AVsynchrony

    Low pacedatrial rates

    High sensedatrial rates

    Low pacedventricular rates

    Suspected

    atrial

    undersensing

    Others

    Figure 1 Incidence of diagnostic observations.

    Number of DO % of patients with DO

    Discha

    rge

    1-3month

    s

    5-7

    month

    s

    10-14month

    s

    Unsche

    dule

    dfollo

    w-up

    0

    20

    40

    60

    80100

    120

    140

    0%

    5%

    10%

    15%

    20%25%

    30%

    35%

    Figure 2 Distribution of DOs over the various follow-upvisits.

    Table 1 Frequency of newly detected atrialtachyarrhythmias in patients with AV block anda DDD(R) pacemaker and no known previous historyof atrial tachyarrhythmias (nZ 128)

    Follow-up Patients withnewly detected AT

    1e3 months 24 (19%)5e7 months 6 (5%)10e14 months 6 (5%)Unscheduled follow-up 1 (0.8%)

    Automatic analysis of pacemaker diagnostic data 245

  • 8/11/2019 Pacemaker Data Collection From Device

    5/6

    Discussion

    In 43% of all follow-up visits, the DOs functionreported significant diagnostic events.

    A major result of this study was the detection ofAT in 31% of patients with no known history ofthem. AT may be asymptomatic and unrecognized

    in many patients as has been reported in previousstudies. For instance, in the AIDA trial, 65% of thepatients in whom atrial tachyarrhythmias weredetected had no documented history of them.Detection of AT in asymptomatic patients is im-portant, given the risk of AT related stroke [7,8].

    Most first detections of AT occur shortly afterimplantation and diminish after the 3-month fol-low-up visit. These findings are in line with earlierreports[9].

    Since this registry was not designed to investi-gate the effects of atrial and ventricular pacing onthe incidence of AT we did not statistically analyseour data with respect to this issue. However,the trends observed in this study with regard tothe percentage of patients with AT in relationto the amount of atrial and ventricular pacing areconsistent with earlier reports. Patients treatedwith dual chamber pacing have been reported toshow a lower incidence of AT, compared withpatients receiving ventricular pacing[10]. Consis-tently, the percentage of patients with AT de-tected by the pacemaker in combination with theexpert system was considerably higher in the

    subgroup with a single lead VDD pacemaker,compared with patients with a dual lead dualchamber device. It is possible that the ventricularpacing which occurs at a lower rate in VDD mode

    contributed to this AT detection. In addition, inpatients with a high-degree AV block, who areexpected to receive a high amount of ventricularpacing, the percentage of newly discovered ATover a 1-year follow-up period was higher than inthe group with intact AV conduction (30% versus22.5%). This observation is in line with previousreports that conclude that frequent ventricularpacing correlates with a higher incidence of AT[11,12].

    Atrial undersensing is a common issue witha VDD lead and may have been related to in-

    appropriate detection of atrial tachyarrhythmias.This would suggest that the amount of AT detectedby the expert system in patients with a VDD deviceis an underestimation. The expert system notifiesthe user with regard to suspected undersensing,based on the available diagnostic information.

    The expert system evaluated in this registryproved to be valuable in detecting new onset orasymptomatic episodes of AT. It needs to beemphasized that pacemaker DOs were particularlyfrequent during non-scheduled follow-up examina-tions of patients with significant symptoms. How-ever, apart from one follow-up (see Table 1), theexpert system did not detect AT in patients withouta history of atrial tachyarrhythmias during thesenon-scheduled follow-up visits. This registry wasconducted in order to assess the ability of theexpert system to alert the user of issues potentiallyrequiring adjustment in the therapy. The relation-ship between these adjustments in response to DOsand the effects on symptoms has not been in-vestigated during this study. Future studies shouldaddress to what extent the expert system may behelpful in effectively reducing symptoms.

    Table 2 Frequency of newly detected atrialtachyarrhythmias in patients with AV block anda VDD pacemaker and no known history of atrialtachyarrhythmias (nZ 40)

    Follow-up Patients withnewly detected AT

    1e3 months 14 (35%)

    5e7 months 4 (10%)10e14 months 2 (5%)Unscheduled follow-up 0 (0%)

    Table 3 Frequency of newly detected atrialtachyarrhythmias in patients without AV block andno known history of atrial tachyarrhythmias (nZ 71)

    Follow-up Patients withnewly detected AT

    3 months 10 (14%)6 months 5 (7%)12 months 1 (1.5%)Unscheduled follow-up 0 (0%)

    Incidenceo

    fnewlydetectedAT[%]

    AVB / VDD

    AVB / DDD

    Non-AVB

    0

    10

    20

    30

    40

    50

    1 - 3 mo 10 - 14 mo

    Follow up

    5 - 7 mo

    Figure 3 Cumulative incidence of newly detected ATin patients with no prior history of AT.

    246 A. Schuchert et al.

  • 8/11/2019 Pacemaker Data Collection From Device

    6/6

    Conclusion

    The expert system evaluated in this registryproved to be valuable in detecting new onset orasymptomatic episodes of AT in patients withoutprior history of AT.

    Acknowledgements

    The authors would like to acknowledge the collab-oration and commitment of all local investigatingphysicians and their staff.

    Austria: F. Rauscha, AKH Wien, Wien; A. Teubl,A.o. Krankenhaus Wr. Neustadt, Wr. Neustadt; K.H.Tscheliessnig, LKH Graz e Univ.-Klinikum, Graz; E.Vesely, A.o. Krankenhaus Waidhofen a.d. Thaya,Waidhofen a.d. Thaya; G. Zach, A.o. KrankenhausBruck a.d. Mur, Bruck a.d. Mur.

    Canada: R.I.G. Brown, Royal Columbian Hospi-tal, New Westminster, QC; R. Dong, Surrey Memo-rial Hospital, Surrey, BC; G. Evans, HamiltonH.S.C., Hamilton, ON; W.G. Hughes, The KawarthaCardiology Clinic, Peterborough, ON; P. Polasek,Kelowna General Hospital, Kelowna, BC.

    France: C. Besson, Clinique Saint-Joseph, Alen-con; H. Bonnet, Clinique dArgonnay, Pringy; J.Covillard, Clinique de Chenove, Dijon; Ch. Delaire,Centre Hospitalier, Cholet; E. Favre, Centre Hos-pitalier, Saint Vallier; J. Pellet, Clinique Belle-done, Grenoble; F. Senellart, Clinique des Cedres,Grenoble.

    Germany: U. Abdel-Rahman, Klin. d. J.-W.-Goethe Universitat, Frankfurt; A. Brattstrom, Kli-nikum Quedlinburg, Quedlinburg; H.-P. Rebeski,Kard. Gemeinschaftspraxis, Kiel; G. Ringwald,Gemeinschaftspraxis, Bruchsal; T. Unger, St. Sal-vator-Krankenhaus, Halberstadt.

    Sweden: E. Baselier, Malarsjukhuset, Eskilstuna.Switzerland: P. Dubach, Kantonsspital Chur,

    Chur.The Netherlands: L.H.M. Bouwens, Deventer

    Ziekenhuis, Deventer; D.J. van Doorn, SpaarneZiekenhuis, Heemstede; H. Drost, Slingeland Zie-

    kenhuis, Doetinchem; G.M. Jochemsen, ZiekenhuisTjongerschans, Heereveen; H.C. Klomps, Zorgnet-werk De Nieuweborgh, Weert; M.R. van der Linde,Ziekenhuis Nij Smellinghe, Drachten; F.J. deNooyer, Ziekenhuis Bethesda, Hoogeveen; M.H.H.de Vaan, IJsselmeer Ziekenhuizen, lokatie Zuider-zee, Lelystad; C.J.P.J. Werter, St. LaurentiusZiekenhuis, Roermond.

    United Kingdom: J. Campbell-Cowan, J. Mc-Lenachan, J. Perrins, G. Reynolds, Leeds GeneralInfirmary, Leeds; A. Cunnington, M. Payne, ManorHospital, Walsall; J. Flint, P. Forsey, WordsleyHospital, West Midlands; A.N. Sulke, EastbourneDistrict General Hospital, Eastbourne; W. McCrea,Great Western Hospital, Swindon; P. Rees, Great

    Ormond Street Hospital, London; N. Spyrou, BattleHospital, Reading.

    References

    [1] Fitts SM, Hill MR, Mehra R, Gillis AM. High rate atrialtachyarrhythmia detections in implantable pulse genera-tors: low incidence of false-positive detections. The PAclinical trial investigators. Pacing Clin Electrophysiol 2000;23:1080e6.

    [2] Sack S, Mouton E, Defaye P, et al. Automatic interpretationfor diagnosis assistance. Improved detection and analysisof sensed and paced events in dual chamber pacemakers

    with extended memory function. A prospective multicen-ter trial in 626 patients. Herz 2001;26:30e9.

    [3] Plummer CJ, Henderson S, Gardener L, McComb JM. Theuse of permanent pacemakers in the detection of cardiacarrhythmias. Europace 2001;3:229e32.

    [4] Glotzer TV, Hellkamp AS, Zimmermann J, Lamas GA. Atrialhigh Rrte episodes detected by pacemaker diagnosticspredict death and stroke. Circulation 2003;107:1614e9.

    [5] Lau CP, Leung SK, Tse HF, Barold SS. Automatic modeswitching of implantable pacemakers: II. clinical perfor-mance of current algorithms and their programming.Pacing Clin Electrophysiol 2002;25:1094e113.

    [6] Defaye P, Dournaux F, Mouton E. Prevalence of supraven-tricular arrhythmias from the automated analysis of datastored in the DDD pacemakers of 617 patients: the AIDA

    study. Pacing Clin Electrophysiol 1998;21:250e5.[7] Ezekowitch MD, Kenneth E, Nazarian SM, et al. Silent

    cerebral infarction in patients with nonrheumatical atrialfibrillation. Circulation 1995;92:2178e82.

    [8] Savelieva I, Camm AJ. Clinical relevance of silent atrialfibrillation: prevalence, prognosis, quality of life, andmanagement. J Int Card Electrophysiol 2000;4:369e82.

    [9] Garrigue S, Cazeau S, Ritter P, Lazarus A, Gras D, Mugica J.Incidence des arythmies atriales chez patients stimules aulong cours en mode double chamber. Arch Mal Coeur Vaiss1996;89:873e81.

    [10] Connolly SJ, Kerr C, Gent M, Yusuf S. Dual chamber versusventricular pacing. Critical appraisal of current data.Circulation 1996;94:578e83.

    [11] Kristensen L, Nielsen JC, Mortensen PT, Pedersen OL,

    Pedersen AK, Andersen HR. Incidence of atrial fibrillationand thromboembolism in a randomized trial of atrial versusdual chamber pacing in 177 patients with sick sinussyndrome. Heart 2004;90:661e6.

    [12] Sweeney M, Hellkamp A, Ellenbogen A, et al. Adverseeffect of ventricular pacing on heart failure and atrialfibrillation among patients with normal baseline QRSduration in a clinical trial of pacemaker therapy for sinusnode dysfunction. Circulation 2003;107:2932e7.

    Automatic analysis of pacemaker diagnostic data 247