pacemaker and anaesthetic implications

53
Pacemakers and their Anaesthetic Implications Dr. Rahul Chauhan PG III yr (Anaesthesia) Moderator-Dr. Subhro Mitra Date- 27/12/2014

Upload: rahul-chauhan

Post on 10-Aug-2015

42 views

Category:

Healthcare


5 download

TRANSCRIPT

Page 1: Pacemaker and anaesthetic implications

Pacemakers and their Anaesthetic Implications

Dr. Rahul Chauhan

PG III yr (Anaesthesia)

Moderator-Dr. Subhro Mitra

Date- 27/12/2014

Page 2: Pacemaker and anaesthetic implications

Introduction.

A pacemaker system is a device capable of generating artificial pacing impulses and delivering them to the heart.

Pacemakers are being used with greater frequency for both conduction and arrhythmia (specially bradyarrythmias).

Page 3: Pacemaker and anaesthetic implications

Pacemaker today!

Page 4: Pacemaker and anaesthetic implications

1980 Large Devices: abdominal site

First human implants Thoracotomy, multiple incisions Primary implanter= cardiac surgeon General anesthesia Long hospital stays Complications from major surgery Perioperative mortality up to 9% Nonprogrammable therapy High-energy shock only Device longevity 1.5 years Fewer than 1,000 implants/year

Page 5: Pacemaker and anaesthetic implications

Pacemaker today-Small devices pectoral site.

First-line therapy for VT/VF patients Treatment of atrial arrhythmias Cardiac resynchronization therapy for HF Transvenous, single incision Local anesthesia; conscious sedation Short hospital stays and few complications Perioperative mortality < 1% Programmable therapy options Single- or dual-chamber therapy Battery longevity up to 9 years More than 100,000 implants/year

Page 6: Pacemaker and anaesthetic implications
Page 7: Pacemaker and anaesthetic implications

Indications of permanent pacemakerimplantation (Acc. To ACC/AHA guidelines)

1) Acquired AV block:

A) Third degree AV block

Bradycardia with symptoms

After drug treatment that cause symptomatic bradycardia

Postoperative AV block not expected to resolve

Neuromuscular disease with AV block

Escape rhythm <40 bpm or asystole > 3s

B) Second degree AV block

Permanent or intermittent symptomatic bradycardia

2) After Myocardial infarction:

Persistent second degree or third degree block

Infranodal AV block with LBBB

Symptomatic second or third degree block

Page 8: Pacemaker and anaesthetic implications

3) Bifascicular or Trifascicular block:

Intermittent complete heart block with symptoms

Type II second degree AV block

Alternating bundle branch block

4) Sinus node dysfunction:

Sinus node dysfunction with symptoms as a result of long term drug therapy

Symptomatic chronotropic incompetence

5) Hypertensive carotid sinus and neurocardiac syndromes:

Recurrent syncope associated with carotid sinus stimulation

Asystole of >3s duration in absence of any medication

Page 9: Pacemaker and anaesthetic implications

Technique of Permanent Pacing

In permanent pacing, leads through the subclavian or cephalic vein.

Leads positioned in the right atrial appendage for atrial pacing and right ventricular apex for ventricular pacing.

The pulse generator lies in the subcutaneous pocket below the clavicle.

Epicardial lead placement is used when no transvenous or if the chest is open.

Page 10: Pacemaker and anaesthetic implications

Generic Codes of Pacemaker

To understand language of Pacemekers Coding system was developed originally by the international conference on heart disease and subsequently modified by the

- NASPE/BPEG (North American society of pacing and electrophysiology/British pacing and electrophysiology group) alliance.

Page 11: Pacemaker and anaesthetic implications

First letter-the chamber being paced Second letter-the chamber being sensed Third letter-response to sensing (I and T indicates inhibited or

triggered responses, respectively). Fourth and Fifth positions-programmable and

antitachyarrhythmia functions, but these two are rarely used. An R in fourth position indicates that the pacemaker

incorporates a sensor to modulate the rate independently of intrinsic cardiac activity such as with activity or respiration.

Page 12: Pacemaker and anaesthetic implications

Generic codes for pacemaker

I II III IV VPacing Sensing Response Programmability TachycardiaO-None O- None O-None O-None O- NoneA-Atrium A-Atrium I-Inhibited C-Communicating P-PacingV-Ventricle V-Ventricle T-Triggered P-simple S-Shocks programmable D-Dual D-Dual D-dual M-multi D-Dual (A+V) (A+V) (I+T) programmable (P+S) S-Simple S-Simple R-Rate (A or V) (A or V) modulation

Page 13: Pacemaker and anaesthetic implications

Generic codes for Defibrillators

NASPE / BPEG Defibrillator (NBD) code

I II III IV Shock chamber Antitachycardia Tachycardia detection Antibradycardia pacing

pacing chamber chamber

O = None O = None E = Electrogram O = None

A = Atrium A = Atrium H = Hemodynamic A = Atrium

V = Ventricle V = Ventricle V = Ventricle

D = Dual (A+V) D = Dual (A+V) D = Dual (A+V)

Page 14: Pacemaker and anaesthetic implications

Important terms regarding pacemakers.

Pulse Generator Leads Electrode Unipolar Pacing Bipolar Leads Endocardial Pacing Epicardial Pacing Pacing Threshold R Wave Sensitivity Hysteresis Runaway Pacemaker

Page 15: Pacemaker and anaesthetic implications

Factors affecting pacing thresholds

Increase Decrease1-4 weeks after implantation Increased catecholaminesMyocardial ischaemia/infaction Stress, anxietyHypothermia, hypothyroidism Sympathomimetic drugsHyperkalaemia, acidosis/alkalosis AnticholinergicsAntiarrythmics (class Ic,3) GlucocorticoidesAntiarrythmics ( class IA/B,2)* HyperthyroidismSevere hypoxia/hypoglycaemia Hypermetabolic statusInhalation-local anaesthetics**

*possibly increase threasholds

**conflicting evidence, probably dose-related

Page 16: Pacemaker and anaesthetic implications

Hysteresis- It is the difference between intrinsic heart rate at which pacing begins (about 60 beats/min) and pacing rate (e.g.72 beats/min). Particularly useful in patients with sick sinus syndrome.

Runaway Pacemaker- It is the acceleration in paced rates due to aging of the pacemaker or damage produced by leakage of the tissue fluids into the pulse generator.

Treatment with antiarrhythmic drugs or cardioversion may be ineffective in such cases.

It is necessary to change the pacemaker to an asynchronous mode, or reprogram it to lower outputs.

If the patient is haemodynamically unstable temporary pacing should be done followed by changing of pulse generator

Page 17: Pacemaker and anaesthetic implications

Types of Pacing Modes

Asynchronous: (AOO, VOO, and DOO) Single Chamber Atrial Pacing (AAI, AAT) sinus arrest

and sinus bradycardia Single Chamber Ventricular Pacing (VVI, VVT) complete

heart block with chronic atrial flutter, atrial fibrillation and long ventricular pauses.

Dual Chamber AV Sequential Pacing (DDD, DVI, DDI, and VDD) AV block, carotid sinus syncope, and sinus node disease.

Programmable Pacemaker Rate responsive pacemaker.

Page 18: Pacemaker and anaesthetic implications

Asynchronous: (AOO, VOO, and DOO)

It is the simple form of fixed rate pacemaker which discharges at a preset rate irrespective of the inherent heart rate.

Can be used in cases with no ventricular activity.

Disavantage- it competes with the patient’s intrinsic rhythm and results in induction of tachyarrythmias.

Continuous pacing wastes energy and also decreases the half-life of the battery.

Page 19: Pacemaker and anaesthetic implications

Single Chamber Atrial Pacing (AAI, AAT)

Atrium is paced and the impulse passes down the conducting pathways, thus maintaining atrioventricular synchrony.

A single pacing lead with electrode is positioned in the right atrial appendage, which senses the intrinsic P wave and causes inhibition or triggering of the pacemaker.

Useful in patients with sinus arrest and sinus bradycardia provided atrioventricular conduction is adequate.

Inappropriate for chronic atrial fibrillation and long ventricular pauses.

Page 20: Pacemaker and anaesthetic implications

Single Chamber Ventricular Pacing (VVI, VVT)

VVI is the most widely used form of pacing in which ventricle is sensed and paced.

It senses the intrinsic R wave and thus inhibits the pacemaker function.

Indication- complete heart block with chronic atrial flutter, atrial fibrillation and long ventricular pauses.

Single chamber ventricular pacing is not recommended for patients with sinus node disease, as these patients are more likely to develop the pacemaker syndrome.

Page 21: Pacemaker and anaesthetic implications

Dual Chamber AV Sequential Pacing (DDD, DVI,DDI, and VDD)

Two leads used-unipolar or bipolar, one for the right atrial appendage and the other for right ventricular apex.

The atrium is stimulated first to contract, then after an adjustable PR interval ventricle is stimulated to contract.

They preserve the normal atrioventricular contraction sequence, and are indicated in patients with AV block, carotid sinus syncope, and sinus node disease.

Page 22: Pacemaker and anaesthetic implications

Dual Chamber AV Sequential Pacing (DDD)

Advantages- they are similar to sinus rhythm and are beneficial in patients, where atrial contraction is important for ventricular filling (e.g. aortic stenosis).

Disadvantage- pacemaker-mediated tachycardia (PMT) due to ventriculoatrial (VA) conduction in which ventricular conduction is conducted back to the atrium and sensed by the atrial circuit, which triggers a ventricular depolarization leading to PMT.

This problem can be overcome by careful programming of the pacemaker.

Page 23: Pacemaker and anaesthetic implications

Programmable Pacemaker

Pacemakers, which not only sense the atrial or ventricular activity but also sense various other stimuli and thus, increase the pacemaker rate.

The various factors, which can be programmed are pacing rate, pulse duration, voltage output, R wave sensitivity, refractory periods, PR interval, mode of pacing, hysteresis and atrial tracking rate.

Various types of sensors have been designed which respond to the parameters such as vibration, acceleration, minute ventilation, respiratory rate, central venous pressure, central venous pH, QT interval, preejection period, right ventricular stroke volume, mixed venous oxygen saturation, and right atrial pressure.

Page 24: Pacemaker and anaesthetic implications

Pacemaker Syndrome

Some individuals, particularly those with intact retrograde VA conduction, may not tolerate ventricular pacing and may develop a variety of clinical signs and symptoms resulting from deleterious haemodynamics induced by ventricular pacing

These include hypotension, syncope, vertigo, light headedness,fatigue, exercise intolerance, malaise, weakness,lethargy, dyspnoea. and even CHF.

Reason - complex interaction of haemodynamic, neurohumoral and vascular changes induced by the loss of AV synchrony

Page 25: Pacemaker and anaesthetic implications

Pacemaker Failure

Due to generator failure, lead failure, or failure to capture.

Failure to capture owing to a defect at the level of myocardium (i.e. the generator continues to fire but no myocardial depolarization takes place) remains the most difficult problem to treat

Page 26: Pacemaker and anaesthetic implications

Factors Important from AnaesthesiaPoint of View

Physiological Potassium Myocardial Infarction Antiarrhythmic Drug Therapy Acid Base Status Hypoxia Anaesthetic Drugs

Page 27: Pacemaker and anaesthetic implications

Physiological

Initial sharp increase in the pacing threshold during first 2 weeks i.e. up to ten times the acute level because of the tissue reaction around the electrode tip. Then It decreases to two to three times the acute level because of the scar formation.

In chronic state, it reaches the initial level in 80% of patients. Problem reduced with the introduction of steroid-eluting leads

and other refinements in the lead technology.

Page 28: Pacemaker and anaesthetic implications

Potassium

It determines the resting membrane potential (RMP) of cell mebrane.

In certain situations, If RMP is less negative less current density is required to initiate an action potential, making capture by the pacemaker easier and vice versa.

Page 29: Pacemaker and anaesthetic implications

Myocardial Infarction

Its scar tissue is unresponsive to electrical stimulation and may cause loss of pacemaker capture

Page 30: Pacemaker and anaesthetic implications

Antiarrhythmic Drug Therapy

Class Ia (quinidine, procainamide), Ib (lidocaine, diphenylhydrantoine), and Ic (flecainide, encainide, propafenone) drugs have been found to increase the pacing threshold

Page 31: Pacemaker and anaesthetic implications

Acid Base StatusAlkalosis and acidosis both cause increase in pacing threshold

HypoxiaIncrease in pacing threshold

Page 32: Pacemaker and anaesthetic implications

Anaesthetic Drugs

These drugs are not likely to change the pacing threshold.

It is notable that addition of equipotent halothane, enflurane, or isoflurane to opiate based anaesthesia after cardiopulmonary bypass did not increase pacing threshold.

Page 33: Pacemaker and anaesthetic implications

Preoperative Evaluation

Evaluation of the patient.- underlying cardiovascular disease responsible for pacemaker implantation and also other asso. problems e.g CAD.HTN,DM etc.

Severity of the cardiac disease, the current functional status, medication of the patient, preimplantation symptoms( Consult cardiologist if present), presence of bruits or signs of CCF.

Routine Investigations esp. 12 lead ECG/CXR/ Sr. Electrolytes. Pacemaker Evaluation

Page 34: Pacemaker and anaesthetic implications

Pacemaker evaluation

Assistance from the cardiologist and the manufacturer’s representative. Location of the pulse generator. Type of pacemaker (fixed rate or demand rate),time since implanted,

pacemaker rate at the time of implantation, and half-life of the pacemaker battery.

Further evaluation of pacemaker function can be done by slowing down the heart rate by carotid sinus massage or by Valsalva manoeuvre.

Reprogramming the pacemaker is generally indicated to disable rate responsiveness.

The AICD/ Anti tachycardia therapy also needs to be disabled before anaesthesia as per ACC/ AHA guidelines.

If the risk of electromagnetic interference (EMI) is high, such as, when the electricity is in close proximity to the generator, alternative temporary cardiac pacing device should be available

Page 35: Pacemaker and anaesthetic implications

Effect of the Magnet Application on Pacemaker Function.

The magnet is placed over the pulse generator to trigger the reed switch present in the pulse generator resulting in a non-sensing

asynchronous mode with a fixed pacing rate (magnet rate). It shuts down the demand function so that the pacemaker stimulates asynchronous pacing. Thus, it protects the pacemaker dependent patient during EMI, such as diathermy or electrocautery.

Demonstrates remaining battery life and sometimes pacing thresholds.

The response varies with the model and the manufacturer so advisable to consult the manufacturer to know the magnet response before use.

Complications- ventricular asynchrony, altered programme

Page 36: Pacemaker and anaesthetic implications

36 Saturday, April 15, 2023

Page 37: Pacemaker and anaesthetic implications

u

Page 38: Pacemaker and anaesthetic implications

Intraoperative Management

Based on the patient’s underlying disease and the type of surgery.

Continuous ECG monitoring. Both electrical and mechanical evidence of the heart

function should be monitored by manual palpation of the pulse, pulse oximetry, precordial stethoscope and arterial line.

Careful during insertion of the guide wire or central venous catheter.

Multipurpose PA cathetor with pacing facilities is preferred.

Page 39: Pacemaker and anaesthetic implications

Anaesthetic technique

Narcotic and inhalational techniques can be used successfully. Anaesthetic agents do not alter current and voltage threhold of

pacemakers. Skeletal myopotentials, electroconvulsive therapy, succinylcholine

fasciculation, myoclonic movements, or direct muscle stimulation can inappropriately inhibit or trigger stimulation, depending on the programmed pacing modes.

Etomidate and ketamine should be avoided as these cause myoclonic movements.

Pacemaker function should be verified, before and after initiating mechanical ventilation as PPV/N2O entrapment can dislodge pacemaker.

If rate responsive mode it should be deactivated before surgery.

Page 40: Pacemaker and anaesthetic implications

Electromagnetic Interference

Direct sources of EMI: Strong Ionising radiation,Nuclear magnetic resonance imaging, surgical electrocautery or dental pulp vitality tester.

Indirect sources of EMI include radar, orthopedic saw, telemetric devices, mechanical ventilators, lithotriptors, cellular telephones, and whole body vibrations.

No effects of CT scan, X rays. Electrocautery is most important source of EMI. Involves

radiofrequency of 300-500 KHz.

Page 41: Pacemaker and anaesthetic implications

Measures to decrease the possibility of adverse effects due to electrocautery

Bipolar cautery unipolar cautery (grounding plate should be placed close

to the operative site and as far away as possible from the site of pacemaker)

Electrocautery should not be used within 15cm of pacemaker.

Programme to asynchronous mode. Provision of alternative temporary pacing. Drugs (isoproterenol and atropine). Careful with Defibrillation if required ( away paddles,

lowest energy required)

Page 42: Pacemaker and anaesthetic implications

Careful monitoring of pulse, pulse oximetry and arterial pressure is necessary during electrocautery, as ECG monitoring can also be affected by interference.

The device should always be rechecked after operation.

Page 43: Pacemaker and anaesthetic implications

Specific Perioperative Considerations

1) Transuretheral Resection of Prostate (TURP) and Uterine Hysteroscopy

-cutting current at high frequencies (up to 2500 kc/sec) can suppress the output of a bipolar demand ventricular pacemaker 2) Electroconvulsive Therapy

-Little current is safe due to high tissue impedence but seizure may generate myopotentials which may inhibit the pacemaker

Page 44: Pacemaker and anaesthetic implications

3) Radiation- radiotherapy for deep seated tumors, therapeutic radiation can damage the complementary metal oxide semiconductors (CMOS) that are the parts of most modern pacemakers. Generally, doses in excess of 5000 rads are required to cause pacemaker malfunction but as little as 1000 rads may induce pacemaker failure or cause runaway pacemaker.4) Nerve Stimulator Testing or Transcutaneous Electronic Nerve Stimulator Unit (TENS) -TENS unit consists of several electrodes placed on the skin and connected to a pulse generator that applies 20 μsec rectangular pulses of 1 to 200 V and 0 to 60 mA at a frequency of 20 to 110 Hz. This repeated frequency is similar to the normal range of heart rates, so it can create a far field potential that may inhibit a cardiac pacemaker.

Page 45: Pacemaker and anaesthetic implications

5) Lithotripsy-High-energy vibrations can cause closure of reed switch causing asynchronous pacing.Activity’ rate responsive pacemaker can be affected by damage caused to the piezoelectric crystals by ESWLOne should have cardiologist’s opinion, perioperative ECG monitoring, device programmer and a standby cardiologist to deal with any device malfunction.Rate responsive pacemaker should have their activity mode deactivated.Focal point of the lithotriptor should be kept at least six inches (15 cm) away from the pacemaker.Patient with abdominally placed pacemaker generators should not be treated with ESWL. Low shock waves (<16 kilovolts should be used initially followed by a gradual increase in the level of energy

Page 46: Pacemaker and anaesthetic implications

6) Magnetic Resonance Imaging (MRI) – MRI is an important diagnostic tool. But its use in patients with pacemaker is contraindicated due to lethal consequences and mortality.Three types of powerful forces exist in the MRI suite.

A) Static Magnetic Field:

-Ferromagnetic material of pacemakers gets attracted to the static magnetic field in the MRI and may exert a torque effect leading to discomfort at the pacemaker pocket.

-Reed switch activation by high static field of 0.5-1.5 T can result in switching of pacemaker to a nonsensing asynchronous pacing.

Page 47: Pacemaker and anaesthetic implications

B) Radiofrequency Field (RF): The radiofrequency signals can cause interference with pacemaker output circuits resulting in rapid pacing at multiple of frequency between 60-300 bpm causing rapid pacing rate.

- It may cause pacemaker reprogramming and destruction of electronic components and also causes thermal injury.

C) Gradient Magnetic field--used for spatial localization.-Can interact with reed - switch in pacemaker.-Causes inappropriate sensing and triggering because of the induced voltages. It may also induce negligible heating effect

Page 48: Pacemaker and anaesthetic implications

Postoperative care

Full telemetric check . Re-programming back to the original setting. Anti-tachycardia therapies of implantable

defibrillators should obviously be re-programmed to their original settings.

Page 49: Pacemaker and anaesthetic implications

Summary

Pacemaker and AICD implanted patients undergoing elective surgery:

1. Contact pacemaker or AICD clinic or manufacturer and also consult cardiologist.

2. Reprogram the device function in selective group of patients

Page 50: Pacemaker and anaesthetic implications

Summary

Pacemaker and AICD implanted patients undergoing elective surgery:

1. Monitoring and anesthesia technique with due considerations to patients CVS status

2. Avoid electrocautery use. If necessary consider use of bipolar or harmonic scalpel

3. Be ready for alternate mode of pacemaker and defibrillator if necessity arises

Page 51: Pacemaker and anaesthetic implications

Summary

Pacemaker and AICD implanted patients undergoing emergency surgery:

1. Try get information regarding pacemaker or AICD at earliest possible or if time permits

2. Initiate invasive arterial pressure monitoring in addition to standard monitoring

Page 52: Pacemaker and anaesthetic implications

52 Saturday, April 15, 2023

Summary

Pacemaker and AICD implanted patients undergoing emergency surgery:

1. If pacemaker failure is associated with haemodynamic instability consider magnet application

2. Consider magnet application with AICD only when EMI source cannot stopped and it is associated with haemodyanamically destabilising inappropriate shocks

3. Interrogate device following either elective or emergency surgery at the earliest possible and consider appropriate resetting if required

Page 53: Pacemaker and anaesthetic implications

THANK YOU…